Skip to main content
Lippincott Open Access logoLink to Lippincott Open Access
. 2025 Apr 9;283(2):184–191. doi: 10.1097/SLA.0000000000006723

Use of an Intraoperative Head, Neck, and Back Support Exoskeleton on Surgeons’ Pain and Posture

Joseph Y Kim *, Hamid Norasi *,†, Stephen D Cassivi ‡, Dalliah M Black ‡, M Susan Hallbeck *,†,‡,✉
PMCID: PMC12783344  PMID: 40201977

Abstract

Objective:

To evaluate the effectiveness of an intraoperative exoskeleton that supports surgeons’ heads, necks, and backs to reduce neck discomfort.

Background:

Surgeons are at a considerable risk of developing neck pain and related injuries. Passive exoskeletons are a potential intervention to support surgeons’ body parts and alleviate strain and discomfort.

Methods:

The NekSpine (a passive neck exoskeleton) was trialed on 12 surgeons (7 males and 5 females) across 6 specialties. Each surgeon performed 4 surgical procedures, 2 with the NekSpine (exoskeleton) and 2 without (baseline). Exoskeleton and baseline surgeries were paired primarily by surgical procedure and secondarily by duration. Surgeons completed surveys that included the NASA-TLX and usability questions before and after surgical procedures to evaluate body part discomfort, overall fatigue, workload, and potential disruptions to the surgical workflow. Surgeons also wore inertial measurement units to objectively record their upper arms, neck, and torso postures.

Results:

Use of the exoskeleton yielded significant decreases in discomfort in the neck, left shoulder, right shoulder, and left arm. Reductions in percent surgical duration in risk 4 (extreme risk postures) coupled with increases spent in risk 2 (moderate risk postures) for the neck and torso were noted. Surgeons reported overall favorable usability results with the exoskeleton not interfering with the surgical workflow, and most stated that they would use the exoskeleton again.

Conclusions:

The NekSpine is a promising intervention to alleviate surgeons’ neck discomfort and improve their neck and torso postures.

Key Words: musculoskeletal disorders, passive exoskeleton, surgeon wellness, surgical ergonomics


Surgeons are at risk for developing work-related neck pain and fatigue due to holding ergonomically risky postures during extended surgeries.1 This may lead to musculoskeletal disorders and can decrease surgeon’s performance, necessitate early retirement, increase burnout, and decrease career satisfaction.2–4 A survey of 775 surgeons from the Society for Vascular Surgery (SVS) revealed significant increases in musculoskeletal pain before and after each operation; 45.2% of those surgeons stated that the body segment in the most pain was their neck.5

Previous studies demonstrated that risky postures affect numerous surgical specialties and can increase fatigue and work-related neck pain.2,5–8 One study by the Mayo Clinic analyzed 116 surgical procedures and observed that surgeons held risky neck postures during 42% of laparoscopic surgery and 80% of open surgery duration.9 Similarly, a quantitative assessment of breast surgeons’ neck postural exposure showed high intraoperative mean flexion angles ranging between 27.3 and 32.9 degrees, which is considered risky for developing musculoskeletal neck pain.10

One promising intervention for mitigating this risk is passive exoskeletons. Passive exoskeletons are nonpowered devices that rely on energy storage in their elastic parts to support users’ body segments during awkward postures to reduce postural load, fatigue, pain, and physical demand, thereby reducing the likelihood of developing pain and, ideally, musculoskeletal disorders. This intervention has yielded positive results in non–health care environments, especially during physically demanding tasks focused on the upper extremities11,12 and the lower back,13–15 by reducing muscle activities at targeted body parts, subsequently decreasing perceived discomfort levels and increasing workers’ endurance times.

Due to the success of exoskeletons in non–health care settings, back and arm exoskeletons for surgeons16 and other health care professionals17,18 have been previously tested. However, only a few studies have trialed exoskeletons that support surgeons’ heads and necks, with one being a PhD thesis19 and the other a lab-based study on undergraduate students.20 While these studies are promising, the actual operating room (OR) has unique constraints that make implementing passive neck exoskeletons much more challenging.21

This project aims to assess the ergonomic characteristics and usability of a passive exoskeleton that supports the head-neck complex during surgical procedures with high-risk neck flexion postures. Specifically, this project will evaluate the exoskeleton’s ability to reduce surgeons’ neck discomfort and pain while performing open surgeries in various specialties without significantly restricting their range of motion. In addition, the exoskeleton’s usability will be evaluated to ensure that the exoskeleton is not impeding the surgeon and their surgical team.

It was hypothesized that (I) the exoskeleton would lead to significant reductions in self-reported neck discomfort and (II) reduce extreme, awkward, and ergonomically high-risk postures relative to baseline conditions. In addition, (III) we expected favorable usability scores, demonstrated by the exoskeleton not interfering with the surgeon’s or their team’s workflows and their desire to use it again during future procedures, and (IV) there would not be significant differences in surgeons’ perceived workloads between exoskeleton and baseline surgeries.

METHODS

This study evaluated a commercially available passive exoskeleton (NekSpine, Composite Manufacturing Inc.). The NekSpine is a biomedically engineered spinal support system designed to reduce stress on the cervical and thoracic spine. Its action supports the head-neck during awkward intraoperative flexed neck postures. As shown in Figure 1, the device consists of 3 main parts: (1) a carbon-fiber support beam running the length of the lumbar, thoracic, and cervical spine to transfer the load from the head to the hips, (2) an adjustable backpack-like component that houses the support beam, and (3) an adjustable headpiece which connects to the support beam via an inextensible cable fixed to a block that slides up and down the beam enabling head flexion and rotation; this headpiece can also support surgical headlights. When wearing the NekSpine, surgeons should feel that their head is supported, as if a hand is holding their forehead gently while they look down.

FIGURE 1.

FIGURE 1

A, The NekSpine exoskeleton. B, Surgeon (on the right) wearing the NekSpine exoskeleton during open surgery. Used with permission of Mayo Foundation for Medical Education and Research, all rights reserved.

Twelve attending surgeons from a single academic hospital were recruited and voluntarily participated in this study; specialties with extreme neck flexion were targeted,1 and those recruited subsequently suggested other surgeons who performed similarly applicable surgeries. Surgeons were then asked to identify open surgeries where they typically held risky neck postures, ideally lasted at least 2 hours, and would personally perform at least 50% of the surgery. Participants performed 2 pairs of 2 surgical procedures (4 completed in total), with each pair consisting of the same type of surgery (eg, skin-sparing mastectomy); 1 done with the exoskeleton and 1 without (baseline). These matching exoskeleton and baseline surgeries were then compared. If surgeons performed the same surgical procedure during both pairs of cases, they were matched based on duration (eg, if a surgeon performed 4 skin-sparing mastectomies, the longest baseline surgery would be paired with the longest exoskeleton surgery) during the data analysis. The order of exoskeleton and baseline surgeries were randomized by participant to reduce potential bias.

Surgeons were introduced to the exoskeleton outside of the OR to find the appropriate exoskeleton size and comfortable settings, as established by Wang et al.16 Each surgeon trialed the exoskeleton during a short (~1 h) portion of a surgery to ensure they were comfortable with the exoskeleton and did not anticipate any issues for themselves, staff, or their patients.

During surgeries with the exoskeleton, the researchers were present in the OR to fit, don, and doff the exoskeleton. After confirming the exoskeleton’s comfort, the researchers stayed outside the OR to be available for any adjustments the surgeon might need during the surgery, ensuring it remained supportive and comfortable throughout the procedure. Surgeons completed a survey and wore inertial measurement units (IMUs) by APDM Wearable Technologies Inc. Surveys were administered immediately before and after each surgery and evaluated overall fatigue and body part discomfort (modified Nordic Musculoskeletal Questionnaire)22 on a Borg CR-10 scale23 [0–10 scale, with 0=none and 10=extreme (maximum)] before, during, and after surgical operations; intraoperative workload based on NASA-Task Load Index (TLX)24 (0–20 scale, with 0 being very low and 20 being very high), difficulty relative to a “standard” procedure, actual difficulty relative to their initial expectations, and intraoperative exoskeleton usability.25 Usability questions asked surgeons to evaluate the exoskeleton on a 0 to 10 scale about its potential interference, improvements, changes in range of motion, and changes in physical comfort. Three yes/no questions asked whether surgeons would want to use the exoskeleton again, if the exoskeleton interfered with their workflow, or whether it interfered with surgical equipment or colleagues.

IMUs were worn on the back of the head, right and left upper arms, and upper torso to track surgeons’ intraoperative postures. These postures were then interpreted using the methodology developed by Davila et al26 and Morrow et al27,28 to calculate mean body part angles and the percentage of surgical duration spent in specific risky postures and based on the Rapid Upper Limb Assessment of McAtamney and Nigel Corlett.29 Angles of each body part were quantified into 1 of 4 risk categories, with 1 being minimal risk, 2 as mild, 3 as moderate, and 4 being extreme risk, and the percent surgical duration in each risk category was calculated.26 A list of dependent variables is shown in Table 1.

TABLE 1.

List of Dependent Variables

Difficulty/workload metrics Discomfort/fatigue metrics Posture metrics
Difficulty based on procedure type
Difficulty based on patient factors
NASA-TLX Workload (0–20 scale)
• Mental demand
• Physical Demand
• Temporal demand
• Performance
• Effort
• Frustration
Usability (exoskeleton trials only; Borg CR-10 Scale)
• Did the exoskeleton interfere with your ability to perform surgery?
• Did the exoskeleton improve your ability to perform surgery?
• Was your range of motion compromised?
• Did the use of the exoskeleton increase your physical comfort when performing surgical procedures?
Usability (exoskeleton trials only; binary questions)
• Given the choice, would you want to use the exoskeleton in future surgical procedures?
• Did wearing the exoskeleton interfere with your normal workflow in any way?
• Did the exoskeleton interfere with surgical equipment or the work of your colleagues?
Changes in discomfort/overall fatigue (Borg CR-10 Scale)
• Neck
• Left shoulder
• Right shoulder
• Left upper arm
• Right upper arm
• Left wrist/hand
• Right wrist/hand
• Upper back
• Lower back
• Overall fatigue
Mean angle (degrees)
• Neck
• Torso
• Right upper arm
• Left upper arm
Duration in risky postures 1–4 (%)
• Neck
• Torso
• Right upper arm
• Left upper arm

Matching procedures were paired, and the pairs’ differences in workload, discomfort, difficulty, and postures were calculated. Body segment discomfort and overall fatigue were evaluated at 3 separate time points: before, during, and after the surgical procedure (“during” and “after” fatigue/discomfort scores were simultaneously and retroactively reported after the surgery). Changes in these scores were defined as the difference between the discomfort level “before” and the larger discomfort level from “during” or “after” the surgery. These changes in body part discomfort and overall fatigue will subsequently be referred to as “(specific body part) discomfort” and “overall fatigue” throughout the rest of the paper.

Statistical analyses were performed using JMP Pro 17 (SAS). For each dependent variable, a Shapiro-Wilk test was used to assess normality. Statistically significant differences between exoskeleton and baseline conditions were then evaluated using paired Student t tests for normally distributed data, while Wilcoxon signed-rank tests evaluated data that violated normality. Statistical significance was defined as having a P value≤0.05. Thematic analyses were used to identify common themes from the usability evaluations.

This study was approved by this academic medical center’s Institutional Review Board (IRB) before recruiting participants or collecting data. All participants consented to the study and were not compensated for participating.

RESULTS

Demographic and Surgery Breakdown

Twelve surgeons (7 males and 5 females) specializing in breast (4), cardiovascular (2), gynecologic (1), plastic (2), thoracic (2), and vascular/endovascular (1) surgery performed a total of 48 surgical procedures (4 per surgeon); specific surgery types and their matching pairs are listed in Appendix Table I, Supplemental Digital Content 1, http://links.lww.com/SLA/F458. All surgeons were right-handed. On average, the surgical duration was 2:42 (H:M); exoskeleton surgeries lasted 2:32, and baseline surgeries lasted 2:52. These durations did not differ significantly (Student t test P value=0.06), with the baseline procedure 0:20 longer on average. Surgical procedures ranged between 0:40 and 5:12.

NASA-TLX Workload

There were no statistically significant differences in perceived procedure difficulty between exoskeleton and paired baseline cases (Student t test P value=0.29) or difficulty relative to initial expectations (Student t test P value=0.52).

Exoskeleton procedures were scored at higher mental and physical demand levels than paired baseline surgeries by 2.6 (Student t test P value=0.01) and 1.7 (Student t test P value=0.05) on a 20-point scale, respectively. No statistically significant differences existed in temporal demand, performance, effort, or frustration between exoskeleton and baseline surgeries. Average reported workloads for exoskeleton and baseline surgeries are shown in Figure 2.

FIGURE 2.

FIGURE 2

Average changes in workloads during exoskeleton and baseline surgeries. *P value ≤0.05, **P value ≤0.01. Workloads were evaluated on a 0 to 20 scale, with 0 being minimal workloads (ie, very low mental demand) and 20 being maximal workloads (ie, very high mental demand). Performance scores were inverted in this figure, meaning 0 meant failure and 20 meant perfection. The error bar shows the standard deviation of the average workload score.

Body Part Discomfort/Overall Fatigue

There were statistically significant decreases in discomfort when using the exoskeleton for the neck (Student t test P value=0.05), left shoulder (Wilcoxon signed-rank test P value <0.001), right shoulder (Wilcoxon signed-rank test P value=0.02), and left upper arm (Wilcoxon signed-rank test P value=0.01). There were no differences in discomfort between the right upper arm, either wrist, upper or lower back, or overall fatigue. These results can be seen in Figure 3.

FIGURE 3.

FIGURE 3

Average changes in discomfort by body part and overall fatigue during exoskeleton and baseline cases. Surgeons evaluated pain on a 0 to 10 scale, with 0=no pain and 10=extreme pain; differences between the beginning of the surgery and during/after were calculated. Positive meant an increase in pain from before to during/after; negative meant a decrease in pain from before to during/after. *P value ≤0.05, **P value ≤0.01, ***P value ≤0.001. The error bar shows the standard deviation of the average workload score.

IMUs

Statistically significant decreases in mean angles were seen in the neck (Wilcoxon signed-rank test P value <0.001) when using the exoskeleton compared with baseline surgeries, as seen in Figure 4. The mean angle of the neck reduced from 33.4 degrees during baseline to 28.6 degrees during exoskeleton surgeries, with an average difference of 4.8 degrees. There were no statistically significant decreases in mean angles on either the left or the right arms or torso.

FIGURE 4.

FIGURE 4

Average changes in postural angles during exoskeleton and baseline surgeries. ***P value ≤0.001. The error bar shows the standard deviation of the average workload score.

Changes in mean angles relative to potential discomfort can be better explained using the 4 risk categories.1,26,28 Statistically significant changes in ergonomic risk were noted in the neck and torso; the exoskeleton led to significant ergonomically favorable decreases in risk 4 (the most extreme, riskiest posture; neck Wilcoxon signed-rank test P value <0.0001, torso Wilcoxon signed-rank test P value <0.0001) and ergonomically favorable increases in risk 2 (moderate risk posture; neck Student t test P value <0.01, torso Student t test P value <0.01) relative to baseline surgeries. There were also ergonomically favorable decreases in risk 3 for both the neck and the torso while wearing the exoskeleton, but these findings were not statistically significant, as shown in Figure 5.

FIGURE 5.

FIGURE 5

Average percentage of surgery duration in high-risk postures. A, Percentage of surgery in high-risk neck postures. B, Percentage of surgery in high-risk torso postures. C, Percentage of surgery in high-risk left shoulder postures. D, Percentage of surgery in high-risk right shoulder postures. *P value ≤0.05, **P value ≤0.01, ***P value ≤0.001. Used with permission of Mayo Foundation for Medical Education and Research, all rights reserved.

Usability

Surgeons responded favorably overall to usability questions about the exoskeleton, as the exoskeleton minimally interfered with surgeons’ abilities to perform surgery and did not compromise their range of motion. In addition, the exoskeleton was reported to slightly increase surgeons’ physical comfort and ability to perform surgery. The results of these usability questions are reported in Table 2.

TABLE 2.

Exoskeleton Usability Questions

Average (SD) Range IQR
Did the exoskeleton interfere with your ability to perform surgery?
(0=no interference, 10=greatly interfered)
1.3 (2.2) (0, 10) (0, 1.25)
Did the exoskeleton improve your ability to perform surgery?
(0=no improvement, 10=great improvement)
3.7 (2.6) (0, 8) (1, 5.25)
Was your range of motion compromised?
(0=no limitation, 10=extremely limiting)
1.5 (2.5) (0, 10) (0, 2)
Did the use of the exoskeleton increase your physical comfort when performing surgical procedures?
(−5=decreased comfort, 0=no change, 5=increased comfort)
1.6 (1.9) (−4, 5) (0, 3)

During one atypical surgical procedure, the exoskeleton restricted the surgeon’s mobility during the second half of the surgery and had to be removed; the first half was completed without incident. In the survey, the surgeon reported 10 for exoskeleton interference and range of motion (greatly interfered, extremely limiting) and −4 for change in physical comfort. If this surgery were to be removed, the ranges in Table 2 for interference with surgical ability, range of motion, and changes in physical comfort would change to (0 to 4), (0 to 6), and (−1 to 5), respectively. In addition, the surgeon performed this same procedure again with the exoskeleton but completed the entire surgery successfully.

When asked whether they would be interested in using the exoskeleton during future surgeries, surgeons responded “Yes” during 91.7% (22/24) of the surgeries. Surgeons handwrote “Maybe” during 2 surgeries instead of selecting the provided “yes” or “no” options. Favorable responses were also seen when asked whether the exoskeleton interfered with their workflow or their colleagues’ workflows/surgical equipment, with surgeons reporting that the exoskeleton did not interfere during 87.5% (21/24) and 100% (23/23; 1 questionnaire was accidentally missed), respectively.

DISCUSSION

In this study, we set out to prove 4 hypotheses: the exoskeleton, relative to baseline, would lead to (I) reductions in neck discomfort, (II) reductions in extreme, ergonomically unfavorable neck postures, (III) overall positive usability scores with minimal interference from the exoskeleton on the surgical workflow, and (IV) minimal differences in surgeons’ workloads. Our data supported all 4 hypotheses overall.

Demographic and Surgery Breakdown

We sought to recruit an equal number of male and female surgeons and ended with 7 males (58.3%) and 5 females (41.7%). Although there are more males than females in the surgical population,30 we oversampled female surgeons to ensure that the exoskeleton would be usable across both sex and anthropometric ranges. In addition, we trialed the exoskeleton across 6 surgical subspecialties. It should be noted that the exoskeleton was only assessed during open surgeries, as these surgeries typically require a greater degree of neck flexion compared with other approaches (eg, laparoscopic or robotic).

Hypothesis (I) Body Part Discomfort and Overall Fatigue

Surgeons experienced significantly lower neck discomfort when using the exoskeleton relative to baseline. Considering the purpose of the NekSpine is to support the head-neck complex, this reduction was unsurprising and supported our first hypothesis regarding decreases in neck discomfort. Surgeons stated that it felt like the exoskeleton was “passively supporting” them and that they forgot they were wearing it. They also noted that after the exoskeleton’s tension was disabled, they felt as if their “head was heavy.”

Along with the reductions in neck discomfort, the exoskeleton significantly reduced the discomforts of the left arm, right shoulder, and left shoulder. These findings were surprising as the exoskeleton did not support the surgeon’s arms, and the effect was seen more on surgeons’ left sides than right sides despite all the surgeons being right-handed. Surgeons may have been altering their operating techniques, leading to this reduced discomfort, though further inquiry is necessary. Considering the exoskeleton was not increasing surgeons’ discomfort, these findings were a beneficial, yet unexplained, side effect.

Hypothesis (II) Postural Changes

Postural data objectively assessed the exoskeleton’s effects. Using the exoskeleton significantly improved surgeons’ neck and torso postures, as shown through the reductions in the percentage of surgical duration spent in risk 4 (high-risk) and increases in duration spent in risk 2 (mild risk).

In the case of the neck, statistically significant improvements were observed by increasing the surgical duration spent in risk 2 and decreasing the duration spent in risk 4. In addition, while not statistically significant, surgeons spent 76.2% of their baseline surgeries on average in risk levels 3 and 4, which was reduced to 69.4% while wearing the exoskeleton. As the 2 highest-risk postures, both risk categories should lead to the greatest discomfort experienced and are the most crucial to mitigate. Surgeons repeatedly stated that the exoskeleton was a subtle reminder not to go to the extremes of postures. These reductions are ergonomically beneficial and may explain surgeons’ subjective decreases in neck discomfort.

Torso posture also improved by reducing the duration spent in risk 4 and increasing the duration spent in risk 2. However, surgeons did not note any significant decreases in upper or lower back discomfort between the exoskeleton and baseline conditions despite the improvements in torso posture. More importantly, torso posture improvements meant that although surgeons reduced their average neck flexion angle, they were not compensating with increased torso flexion.

The exoskeleton had a 2-fold effect on surgeons’ ergonomics: first, the exoskeleton supported surgeons’ body segments, and second, the exoskeleton reminded surgeons to reduce extreme angles. The support for surgeons’ heads was variable, so surgeons could calibrate it to support their heads at any operating posture they desired, even high-risk postures, theoretically reducing neck pain. Simultaneously, setting the tension established an upper limit on the degree of neck flexion allowed without fighting against the exoskeleton. Surgeons were instructed to calibrate the exoskeleton to a “comfortable” level but not necessarily favorable ergonomic postures. Despite the lack of instructions, surgeons naturally set the exoskeleton to support them at less flexed postures, and the upper limit reminded them to avoid the higher-risk postures typically assumed during baseline surgeries. While this study examines short-term effects, long-term implications may exist, such as whether the exoskeleton could be a training tool to help improve surgeons’ postures even during baseline surgeries.

Hypothesis (III) Usability

Although the findings of the exoskeleton’s effects on discomfort and posture appeared to provide ergonomic benefits, they would be meaningless if they were considered unusable to the surgeon or if they interfered with the surgical workflow. Surgeons were overall favorable regarding the exoskeleton’s usability, as reported by surgeons saying “yes” (22/24) and “maybe” (2/24) when asked whether they would want to use the exoskeleton again. In addition, the exoskeleton typically did not interfere with surgeons’ workflows, as there were no issues during 21 of the 24 surgeries. During the 3 surgeries with interference, (1) the exoskeleton’s headgear pulled the surgeon’s goggles upward and had to be readjusted, (2) the surgeon could not find a comfortable/familiar angle for the headlight while it was attached to the headgear, and (3) the surgeon noted that it took an extra 3 to 5 minutes to scrub in due to application of the exoskeleton and sensors. These issues were resolved during the surgeons’ other exoskeleton surgeries, and the delay was due to the study design and not the exoskeleton itself.

Finally, surgeons were asked whether the exoskeleton interfered with their surgical colleagues or equipment. This question was accidentally missed during 1 surgery, but during all other surgeries, there were no instances of the exoskeleton interfering with their colleagues or surgical equipment.

Surgeons also evaluated the exoskeleton interfering with surgical ability, improvement to surgical abilities, compromises to range of motion, and changes in physical comfort. Most surgeons reported that the exoskeleton did not interfere with their surgical ability or compromise their range of motion. The exoskeleton provided slightly increased comfort and improved surgical ability. It should be noted that during one surgery, the surgeon had to remove the exoskeleton during the second half due to it restricting his mobility, which led to the surgeon answering, “greatly interfered” and “extremely limiting.” These issues did not occur during the first half of this surgery or the entire second surgery and appeared to be an outlier in the exoskeleton’s performance, but this instance of the exoskeleton limiting mobility is worth considering.

Hypothesis (IV) NASA-TLX Workload

The exoskeleton increased surgeons’ mental and physical demands but did not affect the other workloads. The lack of significant differences between exoskeleton and baseline cases when comparing procedure difficulty suggests that the exoskeleton may have contributed to this workload instead of inherent differences in case difficulty. The novelty of the exoskeleton may explain this increase in these demands and having to adjust to atypical, albeit improved, postures. This increase in mental demand may indicate a potential learning curve, which was out of the scope of this study as surgeons only performed 2 exoskeleton surgeries, though assessing the change in mental demand in a longitudinal study would be worth considering.

Surgeons also reported higher physical demands during exoskeleton cases despite their reported decreases in discomfort and improvements in posture while wearing the exoskeleton. The learning curve may again explain this increase in physical demand as surgeons had to operate in unfamiliar postures. Although these postures were improved, these novel postures could have felt more physically demanding to the surgeons due to the lack of familiarity.

LIMITATIONS AND CONCLUSIONS

There are plans to incorporate additional participants, such as residents and fellows from additional institutions, in a subsequent study, but this study itself is limited by the small, voluntary sample size from a single institution, which may affect its generalizability. The individual characteristics of each surgeon may have also affected the results, such as varying operative volume and schedules. Paired Student t tests were used to mitigate these additional factors, though their effects may have still been present. In addition, the exoskeleton was only assessed during a select number of procedures, most notably only during open surgeries, so these results may not be generalizable to other approaches like laparoscopic, microscopic, or robotic; however, some surgeons did express interest in using the exoskeleton during these untested approaches. Finally, participants volunteered to participate and were not blinded to the effects of the exoskeleton, creating the possibility of a self-selection bias and placebo effect. Despite these possibilities, our objective postural data indicate that actual differences could be measured separately from and corroborating with the subjective questionnaire results.

In conclusion, this study evaluated the effectiveness of the NekSpine in reducing surgeons’ neck discomfort, improving postures, and its overall favorable usability results. While this exoskeleton is not expected to be a solution to solve every surgical ergonomic issue, it appears to be a step in the right direction in potentially alleviating neck discomfort and may be beneficial in reducing long-term injuries and improving surgical ergonomics.

Supplementary Material

sla-283-184-s001.docx (26.1KB, docx)

ACKNOWLEDGMENTS

The authors thank all the surgeons who volunteered their time as participants in this research.

Footnotes

Data access statement: The data supporting this study’s findings are available from the corresponding author upon reasonable request. Due to privacy and ethical restrictions, the data are not publicly available.

This study was supported in part by Mayo Clinic’s Robert D. and Patricia E. Kern Center for the Science of Health Care Delivery, Rochester, MN, and a grant from Composite Manufacturing Inc (CMI), San Clemente, CA. CMI was not involved in the study design, study execution, or interpretation of the results.

Mayo Clinic and M.S.H. have a financial interest related to this research. The remaining authors report no conflicts of interest. This research has been reviewed by the Mayo Clinic Conflict of Interest Review Board and is being conducted in compliance with Mayo Clinic Conflict of Interest policies.

Supplemental Digital Content is available for this article. Direct URL citations are provided in the HTML and PDF versions of this article on the journal's website, www.annalsofsurgery.com.

Contributor Information

Joseph Y. Kim, Email: Kim.Joseph@mayo.edu.

Hamid Norasi, Email: Norasi.Hamid@mayo.edu.

Stephen D. Cassivi, Email: cassivi.stephen@mayo.edu.

Dalliah M. Black, Email: Black.Dalliah@mayo.edu.

M. Susan Hallbeck, Email: hallbeck.susan@mayo.edu.

REFERENCES

  • 1.Meltzer AJ, Hallbeck SM, Morrow MM, et al. Measuring ergonomic risk in operating surgeons by using wearable technology. JAMA Surg. 2020;155:444–446. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Davila VJ, Meltzer AJ, Hallbeck MS, et al. Physical discomfort, professional satisfaction, and burnout in vascular surgeons. J Vasc Surg. 2019;70:913–920.e2. [DOI] [PubMed] [Google Scholar]
  • 3.Howarth AL, Hallbeck MS, Lemaine V, et al. Work-related musculoskeletal discomfort and injury in craniofacial and maxillofacial surgeons. J Craniofac Surg. 2019;30:1982–1985. [DOI] [PubMed] [Google Scholar]
  • 4.Wells AC, Kjellman M, Harper SJF, et al. Operating hurts: a study of EAES surgeons. Surg Endosc. 2019;33:933–940. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Wohlauer M, Coleman DM, Sheahan M, et al. SS12. I feel your pain—a day in the life of a vascular surgeon: results of a national survey. J Vasc Surg. 2019;69:e189. [Google Scholar]
  • 6.Wohlauer M, Coleman DM, Sheahan MG, et al. Physical pain and musculoskeletal discomfort in vascular surgeons. J Vasc Surg. 2021;73:1414–1421. [DOI] [PubMed] [Google Scholar]
  • 7.Norasi H, Tetteh E, Money SR, et al. Intraoperative posture and workload assessment in vascular surgery. Appl Ergon. 2021;92:103344. [DOI] [PubMed] [Google Scholar]
  • 8.Wauben LSGL, Van Veelen MA, Gossot D, et al. Application of ergonomic guidelines during minimally invasive surgery: a questionnaire survey of 284 surgeons. Surg Endosc Other Interv Tech. 2006;20:1268–1274. [DOI] [PubMed] [Google Scholar]
  • 9.Yang L, Money SR, Morrow MM, et al. Impact of procedure type, case duration, and adjunctive equipment on surgeon intraoperative musculoskeletal discomfort. J Am Coll Surg. 2020;230:554–560. [DOI] [PubMed] [Google Scholar]
  • 10.Hallbeck MS, Law KE, Lowndes BR, et al. Workload differentiates breast surgical procedures: NSM associated with higher workload demand than SSM. Ann Surg Oncol. 2020;27:1318–1326. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kim S, Nussbaum MA, Mokhlespour Esfahani MI, et al. Assessing the influence of a passive, upper extremity exoskeletal vest for tasks requiring arm elevation: Part II – “Unexpected” effects on shoulder motion, balance, and spine loading. Appl Ergon. 2018;70:323–330. [DOI] [PubMed] [Google Scholar]
  • 12.Alabdulkarim S, Kim S, Nussbaum MA. Effects of exoskeleton design and precision requirements on physical demands and quality in a simulated overhead drilling task. Appl Ergon. 2019;80:136–145. [DOI] [PubMed] [Google Scholar]
  • 13.Alemi MM, Geissinger J, Simon AA, et al. A passive exoskeleton reduces peak and mean EMG during symmetric and asymmetric lifting. J Electromyogr Kinesiol. 2019;47:25–34. [DOI] [PubMed] [Google Scholar]
  • 14.Weston EB, Alizadeh M, Knapik GG, et al. Biomechanical evaluation of exoskeleton use on loading of the lumbar spine. Appl Ergon. 2018;68:101–108. [DOI] [PubMed] [Google Scholar]
  • 15.Bosch T, van Eck J, Knitel K, et al. The effects of a passive exoskeleton on muscle activity, discomfort and endurance time in forward bending work. Appl Ergon. 2016;54:212–217. [DOI] [PubMed] [Google Scholar]
  • 16.Wang T, Mendes BC, Hallbeck MS, et al. Introducing exoskeletons into the operating room: a pilot study with vascular surgeons. Proc Hum Factors Ergon Soc Annu Meet. 2021;65:1376–1380. [Google Scholar]
  • 17.Tetteh E, Koenig J, Wang T, et al. A quantitative assessment of the effects of passive upper extremity exoskeletons on expert cardiovascular sonographers’ muscle activity and posture while performing transthoracic echocardiograms (TTE). Int J Ind Ergon. 2023;94:103421. [Google Scholar]
  • 18.Cha JS, Athanasiadis DI, Asadi H, et al. Evaluation of a passive arm-support exoskeleton for surgical team members: results from live surgeries. J Safety Res. 2024;89:322–330. [DOI] [PubMed] [Google Scholar]
  • 19.Zhang C. A passive exoskeleton to prevent low back and neck overload in surgeons. Published online July 10, 2024. https://research.rug.nl/en/publications/a-passive-exoskeleton-to-prevent-low-back-and-neck-overload-in-su. doi: 10.33612/DISS.1028262643 [DOI]
  • 20.Tetteh E, Hallbeck MS, Mirka GA. Effects of passive exoskeleton support on EMG measures of the neck, shoulder and trunk muscles while holding simulated surgical postures and performing a simulated surgical procedure. Appl Ergon. 2022;100(November 2021):103646. [DOI] [PubMed] [Google Scholar]
  • 21.Cha JS, Monfared S, Stefanidis D, et al. Supporting surgical teams: identifying needs and barriers for exoskeleton implementation in the operating room. Hum Factors. 2019;62:377–390. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kuorinka I, Jonsson B, Kilbom A, et al. Standardised Nordic questionnaires for the analysis of musculoskeletal symptoms. Appl Ergon. 1987;18:233–237. [DOI] [PubMed] [Google Scholar]
  • 23.Borg G. Psychophysical scaling with applications in physical work and the perception of exertion. Scand J Work Environ Health. 1990;16(suppl 1):55–58. [DOI] [PubMed] [Google Scholar]
  • 24.Hart SG, Staveland LE. Development of NASA-TLX (Task Load Index): results of empirical and theoretical research. Adv Psychol. 1988;52:139–183. [Google Scholar]
  • 25.Brooke J. SUS: a “quick and dirty” usability scale. Usability Eval Ind. 1996;189:207–212. [Google Scholar]
  • 26.Davila VJ, Meltzer AJ, Fortune E, et al. Intraprocedural ergonomics of vascular surgeons. J Vasc Surg. 2021;73:301–308. [DOI] [PubMed] [Google Scholar]
  • 27.Davila VJ, Weidner TK, Morrow M, et al. RS10. Real-time ergonomic issues in vascular surgeons: how we are hurting ourselves while operating. J Vasc Surg. 2018;67:e242. [Google Scholar]
  • 28.Morrow MMB, Lowndes B, Fortune E, et al. Validation of inertial measurement units for upper body kinematics. J Appl Biomech. 2017;33:227–232. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.McAtamney L, Nigel Corlett E. RULA: a survey method for the investigation of work-related upper limb disorders. Appl Ergon. 1993;24:91–99. [DOI] [PubMed] [Google Scholar]
  • 30.American Medical Association. Active Physicians by Sex and Specialty; 2021. https://www.aamc.org/data-reports/workforce/data/active-physicians-sex-specialty-2021

Articles from Annals of Surgery are provided here courtesy of Wolters Kluwer Health