Table 2. Count rate parameters – Si sensor.
| Beam energy (keV) | Threshold energy (keV) | 10% loss (Mcts s−1 pixel−1) | τR (ns) | τP (ns) | Cut-off (Mcts s−1 pixel−1) |
|---|---|---|---|---|---|
| 8 | 4 | 19.96 | 22.61 ± 0.07 | 22.31 ± 1.24 | 64.67 |
| 12 | 6 | 16.36 | 20.89 ± 0.01 | 17.21 ± 0.02 | 72.28 |
| 16 | 8 | 21.02 | 18.60 ± 0.10 | 16.71 ± 0.52 | 80.07 |
| 20 | 10 | 21.50 | 17.12 ± 0.06 | 14.80 ± 0.51 | 87.54 |
| 25 | 12.5 | 24.32 | 16.66 ± 0.01 | 13.54 ± 0.01 | 90.28 |