Table 2.
Robust reference intervals and overlap percentages for ROTEM Sigma EXTEM parameters in dogs, compared with classical reference intervals previously published for ROTEM Delta.
| EXTEM | |||||||
|---|---|---|---|---|---|---|---|
| Parameter | Robust model | Barthélemy et al. (18) | Pereira et al. (19) | Schefer et al. (20) | |||
| Reference value | Overlap | Reference value | Overlap | Reference value | Overlap | ||
| CT (s) | [25.58–64.42] | [33.00–69.70] | 85.60% | [34.00–83.00] | 62.10% | [23.00–87.00] | 60.70% |
| MCF (mm) | [35.80–72.20] | [44.60–69.00] | 100.00% | [46.00–73.00] | 97.00% | [50.00–65.00] | 100.00% |
| A10 (mm) | [25.02–52.98] | [32.59–59.00] | 77.20% | – | – | [21.00–55.00] | 82.20% |
| ML (%) | [−3.00–35.00] | [0.00–41.00] | 85.40% | [8.00–59.00] | 52.90% | [0.00–12.00] | 100.00% |
| LI60 (%) | [100.00–100.00] | [72.60–100.00] | 100.00% | [70.00–100.00] | 100.00% | – | – |
| G (dyn·s) | [3941.47–10861.11] | – | – | – | – | [3–5,928] | 33.00% |