Table 4.
Robust reference intervals and overlap percentages for ROTEM Sigma FIBTEM parameters in dogs, compared with classical reference intervals previously published for ROTEM Delta.
| FIBTEM | |||||||
|---|---|---|---|---|---|---|---|
| Parameter | Robust | Barthélemy et al. (18) | Pereira et al. (19) | Schefer et al. (20) | |||
| Reference value | Overlap | Reference value | Overlap | Reference value | Overlap | ||
| MCF (mm) | [1.09–6.91] | [3.00–14.00] | 17.10% | – | – | – | – |
| A10 (mm) | [1.09–6.91] | [3.00–12.70] | 16.50% | – | – | [2.00–9.00] | 70.10% |
| ML (%) | [0.00–0.00] | [11.90–47.70] | 0.00% | – | – | [1.00–99.00] | 0.00% |
| G (dyn·s) | [117.66–408.74] | – | – | – | – | [113.00–509.00] | 73.50% |