Table 1.

>Quality of ventricular fibrillation detection algorithms (sensitivity (Sns), specificity (Spc), integrated receiver operating characteristic (IROC)) in per cent, rounded on 3 significant digits, wl = window length in seconds. (*) ... no appropriate parameter exists.

Data Source MIT DB CU DB AHA DB overall result
Parameter wl Sns. Spc. Sns. Spc. Sns. Spc. Sns. Spc. IROC

TCI 3 82.5 78.1 73.5 62.6 75.2 78.3 75.0 77.5 82
TCI 8 74.5 83.9 71.0 70.5 75.7 86.9 75.1 84.4 82
ACF95 8 33.2 45.9 38.1 58.9 51.5 52.2 49.6 49.0 49
ACF99 8 59.4 30.1 54.7 49.3 71.5 40.3 69.2 35.0 49
VF 4 36.7 100 32.2 99.5 17.6 99.9 19.6 99.9 85
VF 8 29.4 100 30.8 99.5 16.9 100 18.8 100 87
SPEC 8 23.1 100 29.0 99.3 29.2 99.8 29.1 99.9 89
CPLX 8 6.3 92.4 56.4 86.6 60.2 91.9 59.2 92.0 87
STE 8 54.5 83.4 52.9 66.6 49.6 81.0 50.1 81.7 67
MEA 8 62.9 80.8 60.1 87.5 49.8 88.6 51.2 84.1 82
SCA 8 72.4 98.0 67.7 94.9 71.7 99.7 71.2 98.5 92
WVL1 8 28.7 99.9 26.2 99.4 26.8 99.5 26.7 99.7 80
WVL2 8 81.1 89.0 61.0 72.1 73.5 89.6 72.0 88.4 (*)
LI 8 3.1 95.1 7.5 94.8 9.3 92.0 9.0 93.9 58
TOMP 8 68.5 40.6 71.3 48.4 95.9 39.7 92.5 40.6 67