Table 2.

Quality of ventricular fibrillation detection algorithms (positive predictivity (PP), accuracy (Ac), calculation time(ct)) for a window length of 8 seconds. Positive predictivity and accuracy in per cent, rounded on 3 digits; calculation time in per cent of the real time of the data, rounded on 2 digits, wl = window length in seconds.

Data Source MIT DB CU DB AHA DB overall result
Parameter wl PP. Ac. PP. Ac. PP. Ac. PP. Ac. ct.

TCI 3 0.6 78.1 33.9 64.8 41.7 77.8 23.7 77.2 1.5
TCI 8 0.8 83.9 38.9 70.6 54.4 84.9 31.1 83.6 2.1
ACF95 8 0.1 45.9 19.7 54.5 18.2 52.1 8.3 49.0 3.6
ACF99 8 0.1 30.2 22.2 50.4 19.9 45.7 9.1 37.9 3.6
VF 4 91.3 99.9 94.0 85.5 98.0 85.8 97.0 93.1 1.4
VF 8 82.4 99.9 94.5 85.2 98.9 85.7 97.7 93.0 1.9
SPEC 8 60.6 99.8 92.0 84.6 97.3 87.7 96.1 93.8 1.9
CPLX 8 0.1 92.3 52.7 80.3 60.7 86.5 40.8 89.2 2.5
STE 8 0.5 83.4 29.5 63.8 35.1 75.6 20.4 79.0 1.9
MEA 8 0.5 80.8 56.0 81.8 47.5 81.9 23.2 81.3 2.5
SCA 8 5.6 97.9 77.8 89.2 98.0 94.9 81.6 96.2 5.9
WVL1 8 38.9 99.8 92.1 84.1 91.8 87.0 90.5 93.5 1.9
WVL2 8 1.2 88.9 36.6 69.8 59.3 86.8 36.8 87.0 40
LI 8 0.1 94.9 27.5 76.5 19.4 77.8 12.1 86.6 15
TOMP 8 0.2 40.6 26.7 53.2 24.8 49.4 12.7 45.0 0.84