Skip to main content
. 2026 Jan 19;34(2):16. doi: 10.32604/or.2025.068967

Table 2. Stability of paclitaxel concentration in cell lysate and culture medium (n = 5).

Analyte nominal concentration (ng/mL) Conditions Measured concentration (Mean ± SD) Accuracy (RE%) Precision (RSD%)
1 4 h, ambient 1.07 ± 0.12 7.00 10.84
4000 3946.38 ± 176.19 1.34 4.45
1 3 freeze-thaw cycles (−20°C/ambient) 1.09 ± 0.12 9.20 11.30
4000 3959.60 ± 206.41 1.01 5.21
1 −80°C/30 days 1.06 ± 0.13 6.40 11.94
4000 4042.02 ± 212.51 1.05 1.05

Note: SD, standard deviation; RE, relative error; RSD, relative standard deviation.