Table 2.

Comparison of infusion efficiency and device safety metrics.

Group Resuscitation time (min) Total infusion volume (mL) Intraluminal pressure (mmHg) Infusion rate (mL/min)
Experimental Group 103.0 ± 16.97 (81.93, 124.07) 951.4 ± 124.07 (797.35, 1105.44) 58.4 ± 2.41 (55.41, 61.39) 9.29 ± 0.50 (8.67, 9.91)
Control Group 122.4 ± 8.02 (112.44, 132.36) 942.4 ± 106.67 (809.5, 1074.85) 56.4 ± 2.07 (53.83, 58.97) 7.68 ± 0.40 (7.18, 8.18)
t value −2.311 0.123 1.407 5.588
p value 0.0496 0.905 0.197 <0.001

Experimental group: the portable sealed positive pressure infusion device. Control group: Clear-Cuff pressure infuser bag. Data are expressed as mean ± SD, 95% CI.