Abstract
Percutaneous kidney biopsies, including coaxial needle biopsy (CNB), have become the preferred method for obtaining tissue samples to supplement laboratory and imaging data in diagnosing or characterizing various renal conditions. Despite the advent of intraprocedural computed tomography scans or ultrasound image guidance, as well as various “plugging” techniques, bleeding remains a frequent and potentially highly morbid complication, particularly in patients with bleeding risks. As such, postbiopsy bleeding is also a primary driver of extended patient monitoring and, in significant cases, results in additional imaging, intervention, and even surgery. Death, although very rare, also occurs.
We reported a series of 5 image-guided kidney biopsy cases that used a new electrocautery hemostasis device called SinglePass Kronos for hemostasis and tract closure. All biopsy procedures were successful, and there were no complications through 30 days following CNB. Postprocedure imaging confirmed the absence of bleeding in all 5 cases.
The SinglePass Kronos device is a potentially valuable tool in minimizing bleeding complications in percutaneous solid organ biopsy procedures and has particular utility in high-risk biopsies such as kidney biopsies. Larger-scale studies of this new device may be warranted.
Index Words: Percutaneous kidney biopsy, medical renal biopsy, core needle biopsy, bleeding, hemostasis, image-guided biopsy, case report
Introduction
Image-guided coaxial needle biopsy (CNB) is a common technique used to obtain high-yield solid organ tissue samples in a percutaneous fashion. CNB has allowed for the minimally invasive diagnosis of various solid organ disorders without the need for surgery and is commonplace in the management and workup of kidney disease.1, 2, 3 However, percutaneous CNB procedures can also result in adverse events related to needle passes or capsular organ punctures during tissue collection. This may be further accentuated in high-risk patients who may have inherent coagulopathies, be on anticoagulation medication, or often have a combination of the 2.4, 5, 6, 7 These risks are especially heightened in renal patients who are often inherently coagulopathic because of uremia. As such, the most common major adverse events associated with these biopsies are related to bleeding, including significant hematoma, macroscopic hematuria, bleeding requiring transfusion, and intervention to stop bleeding, with combined event rates of 7%-16%.8,9 In addition, it is difficult to predict and detect which patients may bleed or develop hematomas after CNB procedures,10 and many patients are also taking antiplatelet and/or anticoagulant medications for various medical conditions. Because of these factors, there has been an increasing use of various techniques and agents among interventional radiologists to plug biopsy channels and promote hemostasis.11,12
Despite these methods, patients still frequently require manual pressure, pressure bandages, bed rest, and/or close nursing observation for several hours and even overnight. Serious hemorrhage requiring blood transfusions, endovascular intervention such as angiography and embolization, or surgical repair can become necessary; patient mortality is also known to occur.1,4,5,8 The SinglePass Kronos Electrocautery Device was developed to address these common and potentially serious complications of percutaneous CNB.13 In this case series, we introduced an initial experience of 5 kidney biopsies using the SinglePass Kronos Electrocautery Device following its introduction onto the market in the United States.
Methods
Ultrasound- or computed tomography (CT) scan-guided, percutaneous, medical kidney CNB was performed in both inpatient and outpatient settings for various kidney tissue assessments in 5 patients. At our institution, medical kidney biopsies are routinely performed with 17/18-gauge Bard Max-Core Disposable Core Biopsy Instruments (Becton Dickinson Inc) through a 17-gauge coaxial access needle used to maintain a single tract. The SinglePass Kronos Electrocautery Device (SinglePass Inc) was applied postbiopsy for hemostasis of the needle tract in each of these 5 cases. Kronos is a disposable, battery-operated electrocautery device used to control bleeding following CNB of solid organs (ie, kidney, lung, etc.). It consists of an ergonomic handle that activates a probe with an electrically activated tip that heats up to a maximum of 90 °C to cauterize tissue (Fig 1). Informed consent to perform the procedure was obtained from each patient. In each case, the SinglePass Kronos is adjusted to match the biopsy access needle length. More specifically, the operator measures the excursion of the biopsy device beyond the tip of the coaxial access needle. The SinglePass device excursion is then set to match this length.
Figure 1.
SinglePass Kronos Electrocautery Device.
After procedure, patients underwent immediate imaging for hemorrhagic complication assessment and were observed for 4 hours for pain, bleeding, adjacent organ injury, cardiopulmonary compromise, and any other etiology for clinical decompensation. Absent complications, patients were discharged (outpatients) or released from bed rest (inpatients). Patient follow-up occurred within 24-48 hours with emergency department referral for concerning signs or symptoms. Our typical post-CNB regimen has patients visiting their referring physician within 1 month, and postprocedure complications are routinely seen in-network. A chart review performed 30 days postprocedure was then used to identify complications.
Details of Case Series
Case 1
A 76-year-old female patient underwent outpatient CNB to work up medical kidney disease with hematuria and proteinuria. She was not taking antiplatelet or anticoagulant medications. Her preprocedure laboratory tests identified a low hemoglobin (Hb) level (7.64 g/dL) and estimated glomerular filtration rate (29 mL/min/1.73 m2), along with elevated levels of serum urea nitrogen (35 mg/dL), creatinine (Cr, 1.8 mg/dL), and urine protein/Cr ratio (10.20).
Three core samples were obtained under ultrasound guidance. The biopsy procedure, followed by Kronos electrocautery, was completed without complications (Fig 2). The patient was discharged 4 hours postprocedure. Immediately postprocedure, ultrasound imaging confirmed the absence of bleeding complications and a clearly defined cautery tract.
Figure 2.
Case 1 intraprocedural ultrasound image shows SinglePass Kronos in active use as shown by the clearly defined echogenic cautery site (long white arrow) and access needle (short white arrows).
There were no complications, including pain or discomfort, reported in the 30 days following the procedure.
The pathology results included a finding of moderate arterio- and arteriolonephrosclerosis and thickening of glomerular basement membranes suggestive of diabetic nephropathy.
Case 2
A 75-year-old woman underwent an inpatient kidney CNB procedure because of a history of systemic lupus erythematosus, chronic kidney disease, and worsening kidney function. A review of the record showed that a prior biopsy in 2021 showed chronic active membranous nephropathy. The patient was not taking antiplatelet or anticoagulant medications. Preprocedure laboratory work identified a low Hb level (10.9 g/dL) and estimated glomerular filtration rate (15 mL/min/1.73 m2). Laboratory tests were notable for elevated levels of international normalized ratio (1.4), serum urea nitrogen (34 mg/dL), Cr (1.73 mg/dL), and urine protein/Cr ratio (24.517).
The biopsy procedure, followed by Kronos electrocautery, was completed under CT scan guidance with 3 core samples obtained. During the procedure, a self-limited perinephric hematoma was noted, but after Kronos electrocautery, the perinephric hematoma remained stable and did not enlarge in size (Fig 3A and B). Thereafter, the patient was placed on bed rest for 4 hours after procedure and ultimately discharged from the hospital uneventfully. There were no complications, including pain or discomfort, reported in the 30 days following the procedure.
Figure 3.
Case 2 intraprocedure computed tomography (CT) scan image (panel A) demonstrates biopsy needle in place (white arrows) and self-limited pararenal hematoma adjacent to the kidney parenchyma (white circle). Following SinglePass Kronos electrocautery, the hematoma remained unchanged on subsequently obtained CT scan images (panel B). Case 3 intraprocedure CT scan image (panel C) demonstrates the biopsy needle in place (white arrows). The postprocedure CT scan image (panel D) shows the cauterized needle tract (long white arrow) and no evidence of hematoma.
The pathology found membranous nephropathy with global glomerulosclerosis, segmental glomerulosclerosis, and severe tubulointerstitial scarring in keeping with diffuse chronic and membranous lupus nephritis.
Case 3
A 42-year-old woman underwent an outpatient kidney CNB because of a history of proteinuria. She was not taking antiplatelet or anticoagulant medication. Her preprocedure blood work was unremarkable. Using CT scan guidance, 6 core samples were obtained, followed by Kronos electrocautery (Fig 3C). There were no complications during the procedure, and the patient was discharged 4 hours later. CT scan imaging confirmed no bleeding following the procedure (Fig 3D). No pain, discomfort, or other complications were reported in the 30 days following the procedure.
A summary of the pathology results included moderate arterio- and mild arteriolonephrosclerosis with segmental glomerular immunoglobulin-lambda deposits suspicious for early cryoglobulinemic glomerulonephritis secondary to a type 1 cryoglobulin.
Case 4
A 36-year-old woman underwent an outpatient kidney CNB because of a history of systemic lupus erythematosus, Sjogren’s syndrome, and proteinuria. She was not taking any antiplatelet or anticoagulant medication. Preprocedure laboratory tests demonstrated a low Hb level (9.9 mg/dL). With CT scan guidance, 4 core samples were obtained (Fig 4A). There were no complications with the biopsy procedure followed by Kronos electrocautery, and postprocedure imaging confirmed no bleeding (Fig 4B). The patient was discharged 4 hours following the procedure, and there were no reported complications, including pain or discomfort, in the 30 days after the procedure.
Figure 4.
Case 4 intraprocedure computed tomography (CT) scan image (panel A) demonstrates biopsy needle well-positioned within the kidney parenchyma (white arrows). The postprocedure CT scan image (panel B) demonstrates no evidence of bleeding or hematoma. Case 5 intraprocedure CT scan image (panel C) demonstrates the biopsy needle in a good position within the renal cortex (white arrows). The postprocedure CT scan image (panel D) demonstrates no evidence of bleeding or hematoma.
Pathology was read out as diffuse lupus nephritis.
Case 5
A 66-year-old man with a history of chronic kidney disease, hypertension, and proteinuria underwent a kidney CNB during an inpatient stay. He was not taking antiplatelet or anticoagulant medication. Preprocedure laboratory work was significant with a low Hb level (11.0 mg/dL) and estimated glomerular filtration rate (18 mL/min/1.73 m2) coupled with elevated levels of serum urea nitrogen (101 mg/dL), Cr (3.6 mg/dL), and urine protein/Cr ratio (12.96). A CT scan-guided biopsy yielded 3 core samples and was immediately followed by Kronos electrocautery (Fig 4C). There were no complications from the procedure, and postbiopsy imaging confirmed needle tract hemostasis (Fig 4D). The patient was kept on bed rest for 4 hours following the procedure. No pain, discomfort, or other complications were reported in the subsequent 30 days.
Pathology results indicated an acute and chronic immune complex-mediated tubulointerstitial nephritis, with differential considerations including Sjogren syndrome, systemic lupus erythematosus, and drug exposure.
Discussion
Bleeding complications remain a concern in percutaneous needle biopsies despite recent equipment and technique advances. SinglePass Kronos is a novel electrocautery device that was well-tolerated and has demonstrated rapid and durable needle tract hemostasis during this early case series of kidney CNB procedures. Although a perinephric hematoma occurred during one of the biopsy procedures presented, it stabilized following electrocautery with SinglePass Kronos and led to no further sequelae during the standard postprocedure observation period and through 30 days postprocedure.
Limitations
This report is limited by the small number of cases presented (5), the relatively short follow-up duration (30 days), and a single instance of imaging immediately post-CNB versus serial imaging. Larger-scale, comparative evaluations with more extensive follow-up to determine the extent of Kronos’ ability to significantly reduce bleeding events are warranted.
Conclusion
The Kronos device is relatively simple to use, with a rapid learning curve that allows most users to become expert users within 10 cases. Routine Kronos use has the potential to overcome bleeding unpredictability, thereby becoming a valuable tool in minimizing potentially life-threatening hemorrhagic complications.
Article Information
Authors’ Names and Academic Degrees
Alexander S. Misono, MD, MBA, Tust Techasith, MD, Avinash Mesipam, MD, Trushar Patel, MD, Christopher Baker, MD, Behroz Oftadeh, PA-C, Vincent Nguyen, PA-C, Quan Dang, PA-C, Lindsey Young, PA-C, and Thomas Velling, MD
Support
Financial support for publication fees was provided by SinglePass, Inc.
Financial Disclosure
Dr Misono is a consultant and speaker for SinglePass Inc and holds equity options of SinglePass. All other authors have no conflict of interests to declare.
Patient Protections
The authors declare that they have obtained informed consent from the patients reported in this article for publication of the information about them that appears within this Case Report.
Peer Review
Received July 21, 2025. Evaluated by 1 external peer reviewer, with direct editorial input from the Editor-in-Chief. Accepted in revised form October 08, 2025.
Footnotes
Complete author and article information provided before references.
References
- 1.Sheth R.A., Baerlocher M.O., Connolly B.L., et al. Society of Interventional Radiology Quality Improvement Standards on Percutaneous Needle Biopsy in Adult and Pediatric Patients. J Vasc Interv Radiol. 2020;31(11):1840–1848. doi: 10.1016/j.jvir.2020.07.012. [DOI] [PubMed] [Google Scholar]
- 2.Hatfield M.K., Beres R.A., Sane S.S., Zaleski G.X. Percutaneous imaging-guided solid organ core needle biopsy: coaxial versus noncoaxial method. AJR Am J Roentgenol. 2008;190(2):413–417. doi: 10.2214/AJR.07.2676. [DOI] [PubMed] [Google Scholar]
- 3.Rathod K., Deshmukh H., Sundaram S., et al. Plugged percutaneous liver biopsy using Tru-cut needle and coils: a retrospective study. Indian J Gastroenterol. 2022;41(5):424–429. doi: 10.1007/s12664-022-01271-2. [DOI] [PubMed] [Google Scholar]
- 4.Buckley A., Petrunia D. Practice guidelines for liver biopsy. Canadian Association of Gastroenterology. Can J Gastroenterol. 2000;14(6):481–482. doi: 10.1155/2000/872560. [DOI] [PubMed] [Google Scholar]
- 5.Nagata H., Sato M., Ogura M., et al. Coagulopathy as a complication of kidney biopsies in paediatric systemic lupus erythematosus patients with antiphospholipid syndrome. Nephrology (Carlton) 2018;23(6):592–596. doi: 10.1111/nep.13175. [DOI] [PubMed] [Google Scholar]
- 6.Nybo M., Dieperink H., Kristensen S.R. Prolonged aPTT after kidney transplantation due to transient lupus anticoagulants. Nephrol Dial Transplant. 2006;21(4):1060–1065. doi: 10.1093/ndt/gfk017. [DOI] [PubMed] [Google Scholar]
- 7.Vassalli P., McCluskey R.T. The pathogenetic role of the coagulation process in glomerular diseases of immunologic origin. Adv Nephrol Necker Hosp. 1971;1:47–63. [PubMed] [Google Scholar]
- 8.Li D., Syriani D.A., Gupta S., et al. Safety and efficacy of different transplant kidney biopsy techniques: comparison of two different coaxial techniques and needle types. Abdom Radiol (NY) 2025;50(6):2619–2625. doi: 10.1007/s00261-024-04722-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Poggio E.D., McClelland R.L., Blank K.N., et al. Systematic review and meta-analysis of native kidney biopsy complications. Clin J Am Soc Nephrol. 2020;15(11):1595–1602. doi: 10.2215/CJN.04710420. doi:10.2215/CJN.04710420 Erratum in: Clin J Am Soc Nephrol. 2021;16(2):293. doi:10.2215/CJN.18391120. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Potretzke T.A., Gunderson T.M., Aamodt D., et al. Incidence of bleeding complications after percutaneous core needle biopsy in hypertensive patients and comparison to normotensive patients. Abdom Radiol (NY) 2016;41(4):637–642. doi: 10.1007/s00261-016-0653-7. [DOI] [PubMed] [Google Scholar]
- 11.Handke N.A., Koch D.C., Muschler E., et al. Bleeding management in computed tomography-guided liver biopsies by biopsy tract plugging with gelatin sponge slurry. Sci Rep. 2021;11(1) doi: 10.1038/s41598-021-04155-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.McDonald J., Amirabadi A., Farhat Z., et al. Experience with compressed Gelfoam Plugs in children during liver biopsies and other IR procedures: A retrospective single-center case series. J Vasc Interv Radiol. 2019;30(11):1855–1862. doi: 10.1016/j.jvir.2019.04.004. [DOI] [PubMed] [Google Scholar]
- 13.Misono A.S., Oftadeh B., Nguyen V., et al. Single Pass Kronos electrocautery device for closure after percutaneous medical liver biopsy: five case reports. World J Gastroenterol. 2025;31(40) doi: 10.3748/wjg.v31.i40.111158. [DOI] [PMC free article] [PubMed] [Google Scholar]