Table 2.
Maximum temperature rise at sonication 5, necrosis volume, and computation time for each model in cases 1 (60 W) and 2 (30 W).
| Case | Properties Model | Maximum Temperature Change (°C) | Necrosis Volume (mm3) | Computation Time (mm:ss) |
|---|---|---|---|---|
|
| ||||
| Case 1 | Constant at 25 °C | 45.90 | 1208 | 17:27 |
| Constant at 37 °C | 47.19 | 1275 | 19:08 | |
| Temperature-dependent | 47.02 | 1466 | 78:06 | |
| Constant at 25 °C | 22.55 | 96 | 17:00 | |
| Case 2 | Constant at 37 °C | 23.19 | 109 | 19:26 |
| Temperature-dependent | 22.96 | 102 | 77:18 | |