Skip to main content
. Author manuscript; available in PMC: 2026 Feb 3.
Published in final edited form as: Int J Hyperthermia. 2026 Jan 4;43(1):2606701. doi: 10.1080/02656736.2025.2606701

Table 2.

Maximum temperature rise at sonication 5, necrosis volume, and computation time for each model in cases 1 (60 W) and 2 (30 W).

Case Properties Model Maximum Temperature Change (°C) Necrosis Volume (mm3) Computation Time (mm:ss)

Case 1 Constant at 25 °C 45.90 1208 17:27
Constant at 37 °C 47.19 1275 19:08
Temperature-dependent 47.02 1466 78:06
Constant at 25 °C 22.55 96 17:00
Case 2 Constant at 37 °C 23.19 109 19:26
Temperature-dependent 22.96 102 77:18