Systematic review and meta-analysis of regulator-approved deep learning systems for fundus diabetic retinopathy detections

Ting-Wei Wang 1,2,3,4, Wei-Ting Luo 2,5, Yu-Kang Tu 6, Yu-Bai Chou 2,7, Yu-Te Wu 3,8,✉
PMCID: PMC12864761  PMID: 41420101

Abstract

To clarify the real-world performance of regulator-approved deep-learning (DL) systems for autonomous diabetic retinopathy (DR) screening, we systematically searched PubMed, Embase, and ClinicalTrials.gov to 3 April 2025, identifying 82 studies (887,244 examinations) covering 25 devices in 28 countries. Hierarchical bivariate meta-analysis yielded pooled sensitivity/specificity of 0.93/0.90 on a per-patient basis and 0.92/0.93 per eye, closely paralleling expert grading. Meta-regression showed that DR severity threshold, national-income level, image gradability, pupil dilation, reference standard, and diagnostic criteria collectively explained most between-study heterogeneity; any-DR screening, low-income settings, or ungradable images increased false-positive rates, whereas dilated pupils, portable cameras, and adjudicated references improved specificity. Publication bias was minimal. Overall, regulator-approved DL algorithms provide accurate, scalable DR detection, but programs must tailor deployment and reimbursement to disease threshold, image quality, and local resources, and post-market audits with standardized gradability metrics are needed to ensure safe, equitable global adoption.

Subject terms: Medical research, Translational research, Medical imaging, Public health

Introduction

Diabetic retinopathy (DR) is one of the most common microvascular complications of diabetes mellitus and a leading cause of preventable blindness among working-age adults worldwide1,2. A landmark meta-analysis estimated that 93 million individuals had DR in 2010, including 28 million with vision-threatening stages1. With global diabetes prevalence projected to reach 783 million by 20453, the number living with DR will rise commensurately. Timely identification and treatment can avert up to 90% of severe vision loss4,5.

Regular retinal screening is, therefore, essential. Current guidelines recommend annual or biennial dilated fundus examinations for people with diabetes6. Yet real-world coverage remains sub-optimal: a recent meta-analysis found that only about 66% of patients in high-income countries and 39% in low- and middle-income countries receive guideline-recommended eye examinations7. Barriers include limited awareness, travel costs, and persistent shortages of trained graders and ophthalmologists, particularly in resource-constrained settings8. In the United Kingdom, the national diabetic eye screening program has contributed to substantial reductions in DR-related blindness; while DR is no longer the leading cause of certifiable blindness in working-age adults, it remains a major contributor9.

Advances in artificial intelligence (AI) and deep learning offer a scalable solution. Gulshan et al. first reported expert-level performance, achieving sensitivity > 90% and specificity ≈ 98% for referable DR on >100 000 images10. Subsequent systems validated in multi-ethnic cohorts11 and prospective trials across Africa and Asia12 confirm that deep learning can detect DR lesions reliably. A 2025 meta-analysis pooling 21 studies estimated pooled sensitivity of 88% and specificity of 91% for deep-learning models versus human grading13, suggesting that autonomous AI could markedly expand screening capacity and accessibility11,12.

Real-world deployment, however, demands rigorous external validation and regulatory oversight. In 2018 the U.S. Food and Drug Administration (FDA) granted de-novo clearance to the first autonomous DR-screening device (IDx-DR) after a pivotal prospective trial demonstrated 87% sensitivity and 91% specificity for more-than-mild DR in primary-care settings14. Other systems—including EyeArt and SELENA+—have since obtained FDA 510(k) clearance or Conformité Européenne (CE), marking the strength of similarly robust trials15,16. Early implementation studies show that these tools integrate smoothly into routine workflows and maintain high diagnostic accuracy at scale16,17. Nevertheless, concerns persist about generalizability across populations, algorithmic bias, and post-market performance drift18,19. Regulator-approval is thus a critical milestone—but not the final step—in ensuring safe and equitable adoption.

Previous reviews have largely evaluated experimental AI algorithms, many of which are not yet available for clinical use13. Clinicians, payers, and policymakers, therefore, lack consolidated evidence focused exclusively on regulator-cleared, commercially available AI devices. A recent review centered on a single cleared system (IDx-DR) confirmed high diagnostic performance in multiple studies20, but no synthesis has yet examined the full suite of approved tools. This systematic review and meta-analysis, therefore, aim to (i) catalog every FDA-, CE-, or other authority-cleared AI system for diabetic-retinopathy detection that is currently marketed; (ii) quantify their pooled diagnostic accuracy across independent validation and real-world implementation studies; and (iii) probe study- and device-level factors—such as imaging protocol, disease prevalence, camera type, patient ethnicity, and management of ungradable images—that may influence diagnostic accuracy.

Results

Study identification and selection

Our systematic literature review, depicted in the PRISMA flowchart (Fig. 1), began with a search across PubMed, EMBASE, and ClinicalTrials.gov, yielding 8404 studies. After removing 1723 duplicates, we screened 6681 articles using EndNote, excluding 5518 for insufficient relevance. Further scrutiny of 1163 full-text articles led to additional exclusions for various reasons. Ultimately, this process resulted in selecting 82 studies10,11,14–16,21–97 for our meta-analysis.

Fig. 1.

Fig. 1

PRISMA flowchart of included studies.

Overview of study characteristics

Table 1 aggregates 82 evaluation datasets of regulator-approved deep-learning systems for diabetic-retinopathy screening conducted between 2016 and 2025 in 30 countries. Roughly two-thirds of studies arose from high-income economies—most often the United States14–17,41,49,55,56,66,70,92,96, Poland27,28,42,67, Italy32,34,35,45,82, the United Kingdom46,68,81,95, Germany57,62,76 and Australia38,91—while one-third came from middle-income settings such as India21,33,59,60,69,83,84,86,92,93, China24,31,39,43,51–54,58,65,71,73,78,80,85, Thailand22,61,89, Mexico25, Chile29,50, Turkey26 and the Philippines48, with only a single dataset from a low-income environment in Vietnam33. Just over half of the evaluations were prospective real-world or clinical-trial designs, the remainder retrospective or mixed. Primary-care and community screening clinics provided the largest share of data (≈58%), followed by tertiary (31%) and secondary (9%) centers, with a handful of mixed-site or nationwide implementations. Adult participants with type 1 or type 2 diabetes were typically in their sixth decade of life, female representation ranged from 31% to 100%, and sample sizes spanned four orders of magnitude—from a 33-patient pilot30 to a national registry of 207,228 examinations73. All clinically relevant endpoints were covered (any DR, referable DR and vision-threatening DR), with 26 datasets reporting on all three thresholds21,24,27,28,31,33,49,50,55,59,63,67,71,73,74,76–78,81–83,85–87,90,91, enabling comprehensive stratified performance analyses. For external validation, 36/82 studies (44%) used at least one fully external test set collected in a different country and/or ethnic population from the training data. These external cohorts spanned 22 countries across all World Bank income strata, including four lower-middle-income settings (India, Thailand, Philippines, Dominica) and one low-income cohort (Vietnam). The remaining 46 studies (56%) performed only internal or same-center testing.

Table 1.

Basic characteristics of included studies

Author Country Study design Study Study period Healthcare setting Population Age Female % Sample size (patients) Economic status DR type External validation (country)
Brant et al. (2025)21 India Cross-sectional Retrospective 2019–2023 Mixed Adults with diabetes referred or self-presenting for DR screening Mean 55.2 ± 11.9 years 49.9 4537 Middle vtDR No
Chotcomwongse et al. (2025)22 Thailand Cross-sectional Prospective 2022 Mixed Adults with diabetes attending yearly DR screening; exclusions: already under eye-care follow-up or unable to be photographed Mean 63.0 ± 10.6 y (AI arm); 62.0 ± 10.9 y (manual arm) 61.7 1454 Middle rDR Yes
Tan-Torres III et al. (2025)33 India Cross-sectional Prospective 2021–2022 Tertiary Adults 18–45 y with diabetes ( ≈ 99% type 1) Younger cohort 21.3 ± 2.3 y; older cohort 31.7 ± 4.2 y (overall 18–45 y) 50.5 321 Middle rDR No
Wei et al. (2025)24 China Cross-sectional Retrospective 2024 Tertiary Consecutive patients (any reason for visit) undergoing routine fundus examination; exclusions: media opacity, mental disorder, pregnancy Median 45 y (range 6 – 95) 50.4 3076 Middle DR No
Wroblewski et al. (2025)25 Mexico Cross-sectional Retrospective 2019 Primary Adults with a known history of diabetes. Females: mean 56.4 years (range 5–80 years) Males: mean 55.9 years (range 12–73 years) 85.5 248 Middle DR Yes
Doğan et al. (2024)26 Turkey Cross-sectional Prospective NR Tertiary Adults with diabetes, no prior DR diagnosis Mean: 58.3 years 49.11 865 Middle rDR/vtDR No
Grzybowski et al. (2024)27 Poland Cross-sectional Retrospective 2020–2021 Secondary Adults with diabetes mellitus NR NR 758 High rDR Yes
Grzybowski et al. (2024)28 Poland Cross-sectional Retrospective 2020–2021 Secondary Adults with diabetes mellitus, no previous DR diagnosis Mean 57 years (range: 18–83 years) 58.5 695 High DR/rDR/vtDR Yes
Ibañez-Bruron et al. (2024)29 Chile Cross-sectional Retrospective 2019 Secondary Consecutive diabetic patients attending for DR screening Mean age: 61 years ( ± 14 years) 61 366 Middle DR/rDR/vtDR Yes
Karabeg et al. (2024)30 Norway Cross-sectional Retrospective 2017 Primary 33 women (66 eyes), self-reported diabetes mellitus (type 1 or type 2) ≥18 years 100 33 High rDR Yes
Li et al. (2024)31 China Cross-sectional Prospective 2020–2021 Primary Adults ≥30 years with type 2 diabetes Mean: 65 ± 12.5 years 55.78 1358 Middle DR/rDR/vtDR No
Musetti et al. (2024)32 Italy Cross-sectional Retrospective 2019 Tertiary Adults (age > 18 years) with diabetes mellitus (Type 1 or Type 2) Mean 65 ± 13 years 46.3 201 High rDR Yes
Nguyen et al. (2024)33 Vietnam Cross-sectional Retrospective 2021–2022 Primary 583 patients Mean age: 61.8 years (range 16–90) 59.3 583 Low DR/rDR/vtDR Yes
Piatti et al. (2024)34 Italy Cross-sectional Prospective 2021–2023 Primary Adults with type 1, type 2, or other diabetes (excluding gestational) Mean: 65 years (range 18–92); 40 637 High DR/rDR Yes
Piatti et al. (2024)35 Italy Cross-sectional Prospective 2022–2023 Primary Adults with diabetes (type 1, type 2, or other; excluding gestational) Mean 68.02 ± 11.78 years 43 506 High rDR Yes
Ritto et al. (2024)36 Switzerland Cross-sectional Retrospective 2021–2023 Tertiary Adults with diabetes mellitus age ≥18 years NR 1141 High rDR Yes
Šín et al. (2024)37 Czech Republic Cross-sectional Prospective 2022 Primary Adults (≥18 years) with diabetes (type 1 or 2) Mean 59.7 ± 14.2 years 48.5 1154 High DR Yes
Taylor et al. (2024)38 Australia Cross-sectional Prospective 2020–2022 Tertiary Adults with pre-existing DR and/or AMD diagnoses Mean: 64.0 ± 12.8 years 51 82 High rDR Yes
Yao et al. (2024)39 China Cross-sectional Prospective 2023–2024 Tertiary 474 adults (natural & diabetic cohorts) — no prior fundus-disease selection Mean 50.04 ± 11.98 y 65.4 474 Middle DR/rDR/vtDR No
Chia et al. (2023)40 Austria Cross-sectional Retrospective 2013–2020 Primary Indigenous Australians with diabetes Mean 54.9 years (SD ± 15.0) 52.4 864 High rDR/vtDR Yes
Dow et al. (2023)41 USA Cohort Retrospective NR Mixed Adults aged ≥18 years with diabetes, without prior DR diagnosis or exam within the past year NR NR 790 High rDR No
Grzybowski et al. (2023)42 Poland Cross-sectional Retrospective 2020–2021 Secondary Adults with diabetes mellitus (Type 1 or Type 2) NR NR 811 High rDR Yes
He et al. (2023)43 China Cross-sectional Prospective 2021–2022 Tertiary Adults > 18 y with/without DR, AMD or GON who agreed to double imaging NR NR 504 Middle rDR No
Kemp et al. (2023)44 Dominica Cross-sectional Prospective 2021 Primary Adults with diabetes ≥18 years Mean: 64 years (range 26–94) 72.6 587 Middle rDR No
Lim et al. (2023)17 USA Cross-sectional Prospective 2017–2018 Tertiary Adults aged ≥18 years with diabetes mellitus 71% aged <65 years, 29% aged ≥65 years 53.2 521 High rDR/vtDR No
Lupidi et al. (2023)45 Italy Cross-sectional Prospective 2022 Primary Adults >18 years; both diabetic and non-diabetic Mean: 60.1 ± 16.2 years NR 251 High DR Yes
Meredith et al. (2023)46 United Kingdom Cross-sectional Retrospective 2021 Primary Patients attending routine diabetic eye screening Mean: 62 years (range: 12–100) 46.6 9817 High DR/rDR Yes
Roubelat et al. (2023)47 Scotland Cross-sectional Retrospective 2006–2016 Primary All people with diabetes ≥12 years invited for screening; single 45° macula-centered image per eye NR NR 179944 High DR/rDR/vtDR No
Salongcay et al. (2023)48 Philippines Cross-sectional Prospective 2021–2022 Primary Adults ≥18 years with type 1 or 2 diabetes Mean: 59.1 ± 10.4 years 64.4 2793 Middle rDR/vtDR Yes
Vought et al. (2023)49 USA Cross-sectional Retrospective 2022 Primary Adults (age >18) with and without known diabetes Mean age: 64 years 55 62 High rDR/vtDR No
Arenas-Cavalli et al. (2022)50 Chile Cross-sectional Retrospective 2014–2016 Primary Adults with diabetes enrolled in the national DR program Mean: 63 years (SD: 12.7) 59.8 1123 Middle DR No
Dong et al. (2022)51 China Cross-sectional Prospective 2019–2021 Primary Adults (≥18 years) with type 1 or type 2 diabetes Mean: 52.09 ± 11.51 years 36.12 443 Middle DR/rDR/vtDR No
Han et al. (2022)52 China Cross-sectional Prospective 2018, 2020 Mixed Adults aged 23–86 years (community: healthy; hospital: varied diseases) Median: 63 (PUMCH); 27 (NSDE) NR 300 Middle rDR No
Hao et al. (2022)53 China Cross-sectional Prospective 2021 Primary Adults with diabetes (type 1 or 2), aged ≥18 years Mean: 63.03 ± 8.72 years 65.9 3796 Middle DR/rDR No
Li et al. (2022)54 China Cross-sectional Prospective 2020 Primary Adults ≥18 years able to cooperate with fundus photography Mean: 53.2 ± 15.6 years 43.2 1585 Middle rDR No
Mehra et al. (2022)55 USA Cross-sectional Retrospective 2019–2021 Primary Adult patients with DM age ≥18 years 40 1052 High rDR No
Mokhashi et al. (2022)56 USA Cross-sectional Retrospective 2019 Primary Adults (age ≥18) with diabetes mellitus Mean age: 60.7 years (range 21–90 years) 60.8 260 High rDR No
Paul et al. (2022)57 Germany Cross-sectional Prospective 2020 Tertiary Patients with diabetes mellitus (Type 1 and Type 2) Mean: 54.2 years (range: 9–92 years) 42 503 High DR/rDR/vtDR Yes
Pei et al. (2022)58 China Cross-sectional Retrospective 2018–2020 Tertiary Adults with type 2 diabetes mellitus (T2DM) Mean: 61.2 ± 11.8 years 50.5 549 Middle DR No
Rao et al. (2022)59 India Cross-sectional Retrospective 2015–2016 Tertiary Adults with diabetes, aged >21 years Mean: 54.1 ± 8.3 years 35 135 Middle DR/rDR/vtDR No
Rio et al. (2022)60 India Cross-sectional Prospective NR Primary Adults with known or screen-detected diabetes Mean: 57.7 years (SD: 11.1) 52.2 11199 Middle rDR Yes
Ruamviboonsuk et al. (2022)61 Thailand Cohort Prospective 2018–2020 Primary Adults (≥18) with diabetes, listed on national registry Mean: 59.9 ± 10.3 years 63.1 7651 Middle vtDR Yes
Skevas et al. (2022)62 Germany Cross-sectional Prospective 2020 Tertiary Adults ≥18 years, not pre-selected for DR/AMD NR NR 630 High rDR Yes
Sedova et al. (2022)63 Austria Cross-sectional Prospective NR Tertiary Adults with diabetes mellitus (Type 1 or Type 2), no previous DR diagnosis Mean 55 ± 15.5 years 38.9 54 High rDR No
Tsai et al. (2022)64 Taiwan Cross-sectional Retrospective NR Primary Adults with diabetes (no specific type given) NR NR 1407 High rDR No
Yang et al. (2022)65 China Cross-sectional Prospective 2018–2019 Tertiary Adults ≥ 18 y with type 1 or type 2 diabetes mellitus (one eye/participant) Median 60.6 y (range 20.2 – 85.8) 51.8 1001 Middle rDR No
Bhuiyan et al. (2021)66 USA Cross-sectional Prospective 2019–2020 Primary Adults ≥50 years; diabetic and non-diabetic (mixed) patients Over 50 years NR 340 High rDR No
Grzybowski et al. (2021)67 Poland Cross-sectional Retrospective 2011–2012 Secondary Diabetic patients who visited a community-based secondary care Adults with diabetes mellitus ( ≥ 18 years old) NR 170 High rDR Yes
Heydon et al. (2021)68 United Kingdom Cross-sectional Prospective 2017–2018 Primary All patients who attended routine diabetic eye screening in these programs, with no additional editing or selection of images 60/66/61 mean 46/43/45.6 30405 High rDR Yes
Ipp et al. (2021)15 USA Cross-sectional Prospective 2017–2018 Primary, Tertiary Adults (≥18 years) with diabetes (type 1 or type 2) Mean 53.9 ± 15.2 years 49 893 High rDR No
Jain et al. (2021)69 India Cross-sectional Prospective 2018–2019 Primary Adults with diabetes visiting public clinics Mean: 54.9 ± 10.4 years 46.87 1294 Middle DR/rDR No
Kim et al. (2021)70 USA Cross-sectional Prospective 2016–2017 Tertiary Patients with diabetes Mean 57.0 ± 15.7 years 37.7 69 High rDR No
Li et al. (2021)71 China Cross-sectional Prospective 2018–2019 Tertiary Adults with diabetes (Type 1 and Type 2) Mean: 50 ± 12 years 31.6 1147 Middle rDR No
Li et al. (2021)79 Taiwan Cross-sectional Prospective 2019 Tertiary Adults with type 2 diabetes NR NR 468 High rDR No
Lin et al. (2021)73 China Cross-sectional Prospective 2018–2019 Mixed General Chinese population undergoing fundus screening NR 47 207228 Middle rDR No
Ming et al. (2021)74 China Cross-sectional Prospective 2018–2019 Primary Adults with diabetes aged ≥40 years, no prior DR diagnosis Mean: 69.3 ± 9.0 years 54.3 173 Middle DR/rDR No
Shah et al. (2021)75 Spain Cross-sectional Prospective 2011–2012 Primary Adults with diabetes mellitus (Type 1 and Type 2) Medianage = 74 years 55 2680 High rDR Yes
Wintergerst et al. (2021)76 Germany Cross-sectional Prospective 2018 Primary Patients with DM, mean ~69 NR 72 High rDR Yes
Gonzalez-Gonzalo et al. (2020)77 Netherlands, Spain, Sweden, Denmark Cross-sectional Retrospective 2011–2016 Tertiary Adults undergoing fundus imaging (with or without DR/AMD) NR NR 600 High rDR No
He et al. (2020)78 China Cross-sectional Prospective 2018 Primary Adults with diabetes (Type 2 or Type 1), aged ≥18 years Mean: 68.5 ± 7.2 years 53 889 Middle DR/rDR No
Hsieh et al. (2020)79 Taiwan Cross-sectional Retrospective 2007–2017 Tertiary Adults with diabetes undergoing DR screening NR NR 7524 High DR/rDR No
Li et al. (2020)80 China Cross-sectional Prospective 2018 Primary Adults with diabetes (diagnosed by fasting glucose ≥7.0 mmol/L, RBG ≥ 11.1 mmol/L, or on diabetes treatment) Mean: 58.3 ± 10.6 years 59.8 8005 Middle DR/rDR No
Olvera-barrios et al. (2020)81 United Kingdom Cross-sectional Prospective NR Primary Final human grade of standard NDESP images NR NR 1257 High DR/rDR/vtDR Yes
Sarao et al. (2020)82 Italy Cross-sectional Prospective NR Tertiary diabetic participants (type 1 or 2) Mean ≈ 64.7 ± 14 47 165 High rDR Yes
Sosale et al. (2020)83 India Cross-sectional Prospective 2018 Tertiary Adults with type 1, type 2, or secondary diabetes Mean: 55 ± 11 years 42 297 Middle DR/rDR No
Sosale et al. (2020)84 India Cross-sectional Prospective 2018 Tertiary Adults (≥18 years) with type 1 or type 2 diabetes Mean: 55 ± 11 years 42 900 Middle DR/rDR No
Zhang et al. (2020)85 China Cross-sectional Prospective 2018–2019 Primary Adults ≥18 years with diabetes (mostly type 2, ~98%) Mean: 54.3 ± 11.6 years 42.6 15805 Middle DR/rDR/vtDR No
Bhaskaranand et al. (2019)16 United States Cohort Retrospective 2014–2015 Primary Adults with diabetes mellitus NR NR 107001 High rDR No
Gulshan et al. (2019)86 India Cross-sectional Prospective 2016–2017 Tertiary Adults >40 years with known diabetes Aravind: 56.6 ± 9.0; Sankara: 56.0 ± 10.0 years 41.9/32.8 3049 Middle rDR No
Natarajan et al. (2019)87 India Cross-sectional Prospective 2018 Primary Adults with diabetes visiting government clinics NR 51.6 213 Middle DR/rDR No
Quellec et al. (2019)88 France Cross-sectional Retrospective 2004–2017 Mixed Routine DR screening patients in OPHDIAT NR NR 164660 High rDR Yes
Ruamviboonsuk et al. (2019)89 Thailand Cross-sectional Retrospective 2015–2017 Primary Adults with diabetes (random sample from national registry) Mean: 61.1 ± 10.96 years 67.5 7515 Middle rDR Yes
Verbraak et al. (2019)90 Netherlands Cohort Study Retrospective 2015 Primary Adults with Type 2 diabetes mellitus Mean age: 63 years (SD 11.3) 47 1616 High rDR Yes
Abràmoff et al. (2018)14 USA Cross-sectional Prospective 2017 Primary Adults aged ≥22 years with diabetes, no known diabetic retinopathy Median age: 59 years (range 22–84 years) 52.5 819 High rDR No
Keel et al. (2018)91 Australia Cross-sectional Prospective 2017 Tertiary Adults ≥18 with diabetes Mean: 44.3 ± 16.6 years (range 20–90) 43 96 High rDR No
Krause et al. (2018)92 USA, India Cross-sectional Retrospective 2015–2016 Mixed Adults with diabetes mean 54.9 ± 10.9 years 60.5 1958 High, Middle rDR No
Rajalakshmi et al. (2018)93 India Cross-sectional Prospective NR Tertiary Adults (≥18 years) with type 2 diabetes (T2DM) ≥18 NR 301 Middle DR/rDR/vtDR Yes
van et al. (2018)94 Netherlands Cross-sectional Retrospective NR Primary Adults with Type 2 diabetes Mean 65 ± 11.9 years 43.9 1415 High rDR Yes
Ting et al. (2017)11 Singapore, China, USA, Australia, Mexico, Hong Kong Cross-sectional Retrospective 2010–2015 Primary Adults with diabetes mellitus Mean age in primary validation set: 60.2 years 45.4 38256 High, Middle rDR Yes
Tufail et al. (2017)95 United Kingdom Cross-sectional Prospective 2012–2013 Primary Adults with known diabetes mellitus (any type), attending their annual DR screening Median 60 years (range 10–98 NR 20258 High rDR No
Abràmoff et al. (2016)96 USA, France Cross-sectional Retrospective NR Secondary Adults with diabetes mellitus Mean 57.6 years ( ± 15.9 years) 43 874 High rDR Yes
Gulshan et al. (2016)10 USA, India Cross-sectional Retrospective 2015–2016 Mixed Adults with diabetes EyePACS-1: mean 54.4; Messidor-2: mean 57.6 EyePACS-1: 62.2%; Messidor-2: 42.6% 9963/1748 High, Middle rDR No

DR diabetic retinopathy, rDR referable diabetic retinopathy, vtDR vision-threatening diabetic retinopathy, T1DM type 1 diabetes mellitus, T2DM type 2 diabetes mellitus, DM diabetes mellitus, AMD age-related macular degeneration, SD standard deviation, NR not reported.

*A study may appear in more than one panel when it reports distinct 2 × 2 tables for different disease thresholds or datasets—counts remain mutually exclusive and weights are calculated within each panel only.

Comparative overview of deep learning algorithms in diagnostic studies

Table S4 synthesizes the methodological foundations of the 82 diagnostic-accuracy studies of regulator-approved deep-learning screeners for diabetic retinopathy. Human-graded color fundus photography remains the gold standard: approximately 75 datasets (91%) used double-masked or consensus reads by retina specialists, with a third expert arbitrating any discordance. Clinically anchored reference standards are rarer: only about one study in eight verified ground truth through in-person dilated biomicroscopy or optical coherence tomography (OCT)-aided examination25,29,41,45,58,59,70. Image acquisition is remarkably homogeneous—most studies relied exclusively on 45° desktop fundus cameras—yet 12 incorporated wide-field, ultra-widefield or OCT inputs to probe peripheral or macular disease missed by single fields14,25,28,32,36,38,57,59,62,63,69,70,76,83,93. Pupil-management strategies varied: 15 datasets were fully mydriatic25,29,33,44,48,59,68–70,75,81–83,87,93, 39 strictly non-mydriatic24,27,28,30,31,33,34,37,40–43,45,46,49,51–54,56,57,60,62–64,66,67,71–74,77–80,84,86,91,94, and 26 adopted “dilate-if-needed” contingency protocols10,11,14,15,17,21,22,26,32,35,36,38,39,47,50,55,58,61,65,76,88–90,92,95,96. Single-field imaging predominated (≈60%), while two-field schemes were routine in hospital-based or smartphone studies and ≥3-field/montage approaches appeared in 12 investigations16,25,28,57,59,62,69,70,76,83,87,93. Vendor participation was declared in 30 datasets (37%), chiefly for IDx-DR, EyeArt, ARDA and RetCAD10,14,16,21,22,30,33,37,40,43,50,52,54,59,61,62,64,66,71,73,75,85,86,88–90,92,94–96, whereas independent validations—particularly of Medios, VeriSee, CARE and smartphone implementations—reported no developer involvement25,42,44,49,69,72,79,83,84,87. Collectively, the evidence demonstrates that cleared DL algorithms have been benchmarked across a spectrum of camera platforms, dilation protocols and adjudication workflows, yet still anchor their performance against rigorous expert interpretation.

Regulatory clearances and underlying deep-learning architectures of AI-based diabetic-retinopathy screening systems

Table 2 maps the current regulatory and technical landscape for the 25 leading commercial DL algorithms for diabetic retinopathy. EyeArt (510(k) K200667/K223357 + EU-MDR Class IIb) is the only product that holds both FDA clearance and an EU-MDR Class IIb certificate, making it the most broadly approved solution worldwide. IDx-DR (De Novo DEN180001 + successive 510(k) updates) holds FDA clearance and CE-MDD Class IIa certification. Europe’s CE roster has expanded to include Medios AI™, LuxIA, and Aireen® under EU-MDR Class IIb, plus RetCAD™, EyeCheckup, VISUHEALTH-AI DR, iPredict®, Retmarker, OphtAI, and the VGG-based SELENA + ™ under MDR/MDD Class IIa certifications, as well as Autograder (CE Class I, deployed in NHS Scotland DES since 2011). Additional international clearances include Health Canada approvals for IDx-DR, EyeArt, and OphtAI, and Australia TGA certification for Retmarker and iPredict®. Asian regulators have been equally active: China’s NMPA has issued Class III licenses for CARE, EyeWisdom, VoxelCloud Retina, AIDRScreening, Smart Eye Sugar Net, and AI-100, while Taiwan’s TFDA has cleared VeriSee DR™, and Singapore’s HSA has approved Medios AI™ and SELENA + ™; Malaysia (MDA), Thailand (Thai FDA), and Saudi Arabia (SFDA) have each granted additional market authorizations.

Table 2.

Regulatory clearances and underlying deep-learning architectures of AI-based diabetic-retinopathy screening systems

AI system Vendor/Manufacturer Key regulatory clearances DL architecture Training Data (Countries/Ethnicities) Primary Retinal Disease(s) Detected/Indication
ARDA® DL Verily (Google Health), USA CE-MDD IIa (2018) 10 × Inception-v3 ensemble ≈130k EyePACS (USA, Hispanic/White/Black/Asian) + 3 Indian hospitals (South Asian) DR incl. DME
EyeArt® Eyenuk, Inc., USA FDA 510(k): K200667 (2020, EyeArt) & K223357 (2023, EyeArt v2.2.0) · EU-MDR IIb (2023) · Health Canada MDL 104401 (2021) multi-CNN ensemble >78k EyePACS (USA; ~55% Hispanic, White, Black, Asian) >mild DR, vtDR; also AMD, GON
Medios AI™ Remidio Innovative Solutions, India EU-MDR IIb (2023) · Singapore HSA Class B (2022) · India CDSCO license (2024) 2 × Inception-v3 + MobileNet QC 50k: EyePACS + proprietary Indian smartphone images Referable DR (mobile/desktop)
LumineticsCore®(IDx-DR) Digital Diagnostics Inc., USA FDA De Novo: DEN180001 (2018 IDx-DR) + 510(k): K213037 (2021 IDX-DR) & K203669 (2022 IDX-DR v2.3) · CE-MDD IIa · Health Canada MDL 104238 (2021) 5-CNN Inception-like ensemble US + Indian data; pivotal trial (60% White, 28% Black, 12% Hispanic) Autonomous >mild DR ± DME
RetCAD™ Thirona / Delft Imaging, Netherlands CE-MDR IIa (NB 0344, 2022) 5-CNN deep-conv ensemble Messidor (France) + Retisalud (Spain); European ancestry DR + AMD
Retmarker Retmarker SA (formerly Critical Health), Portugal CE‑MDD Class IIa (2010); manufacturer notes TGA (Australia) approval deep-learning classifier for DR screening DL variant training data not publicly specified. DR screening (disease/no disease) + MA‑turnover biomarker for progression/DME risk
EyeCheckup EyeCheckup Ltd / URAL Telekom, Türkiye CE-MDR IIa (2023) single Inception-v4-style CNN Turkish clinic data; presumed Anatolian ethnicity DR, AMD, glaucoma + other lesions
LuxIA RetinAI & Fundación Ver Salud (Ikerian AG), Switzerland/Spain EU-MDR Class IIb certificate (May 2025) EfficientNet-based CNN (single) (Modern Retina, PubMed) Retisalud (Canary Islands); Spanish/N-African admixture >mild DR
EyeWisdom® Visionary Intelligence Ltd, China China NMPA Class III SaMD (CQD2200108, 2024) ResNet / EfficientNet mix (ensemble) Chinese hospital images; likely majority Han Chinese DR (all grades)
Aireen® Aireen Health s.r.o., Czech Rep. EU-MDR IIb (2023) single EfficientNet-V2 CNN Czech cohort: training data not disclosed DR screening (Czech setting)
SELENA + ™ EyRIS Pte Ltd, Singapore CE-MDD IIa · Singapore HSA Class B · Malaysia MDA Class B 8 × VGG-19 family ensemble 274k Singaporean (Chinese, Malay, Indian) + 200k global (multi-ethnic) DR, AMD, GON (tri-disease)
DART Chile MoH screening program Chile MoH adoption (2018); clinical validation accepted 2020 two fully convolutional networks Chile national program; Hispanic/Mapuche admixture DR (national program)
CARE Beijing Airdoc Tech, China China NMPA Class III SaMD (2020) single Inception-ResNet-v2 CNN >400k from 301 Chinese clinics; majority Han Chinese DR (auxiliary diagnosis)
VeriSee DR™ Acer Medical Inc., Taiwan TFDA Class II · Thai FDA Class 2 · Philippines CMDR 2024-02319 modified Inception-v4 CNN 7.5k from Taiwan (Han Chinese) + EyePACS Referable DR; “VeriSee AMD” in pilot
VISUHEALTH-AI DR VisuHealth Medical Tech Co., China CE-MDR IIa (Rule 11 SaMD) single Inception-v3 CNN Likely Chinese multi-center dataset DR screening
VoxelCloud Retina VoxelCloud Inc., China/USA China NMPA Class III · Saudi SFDA (2024) 6 × Inception-ResNet-v2 + FPN 15k+ patients from 155 China clinics (multi-regional Han) DR, DME, AMD, glaucoma
Smart Eye Sugar Net Shenzhen Silicon Intelligence Tech., China China NMPA Class III SaMD (2020) single proprietary CNN Not disclosed; presumed Chinese data DR (NMPA Class III)
iPredict® (DR module) iHealthScreen Inc., USA CE-MDR IIa (2021) · TGA IIa (2022) · UAE MoH (2022) FDA 510(k) pending single CNN (undisclosed) US urban datasets: ethnicity not published DR; also AMD, GON (regional)
BigVision MIAS 2000 BigVision Medical Tech Co., China China Class II SaMD (2018) – assistive single proprietary CNN No public info Assistive DR, AMD, glaucoma (18 lesions)
AIDRScreening Shenzhen SiBright Co. Ltd, China China NMPA Class III (innovative device, 2019) 3-CNN ensemble >40k Chinese images; ethnic details not given Referable DR only
Eyetelligence Assure™ Eyetelligence Pty Ltd, Australia CE-MDD Class I (2019) · Australia ARTG Class I (IDs 325343, 355170) single Inception-v4 derivative EyePACS, Messidor + Melbourne data (Caucasian & Indigenous Australians) DR, AMD, glaucoma, CV risk markers
AI-100 algorithm Baidu Intelligent Healthcare Unit, China China NMPA Class III SaMD (2022) — DR + RMD + GCS DenseNet-121 backbone with bilinear-pooling & quality-assessment modules >700k images from 179k Chinese patients (multi-regional Han) DR, RMD, GON
Huiyan AI (SG-DR System) Shang Gong Medical Info Tech Co., China Guangdong MPA Class II SaMD (May 2022) proprietary CNN (undisclosed) Not disclosed; presumed Chinese community data DR (regulatory indication); details not published
OphtAI Evolucare & ADCIS (France) CE Class IIa (EU); Health Canada; UKCA (per vendor) CNN‑based (proprietary) with heat‑map lesion localization AP‑HP Ophdiat tele‑screening network (France) as primary data source (per vendor) DR (severity grading), AMD, Glaucoma
Autograder Medalytix / EMIS Health, UK (NHS Scotland DES) CE Class I (MDD); routine use in Scottish DES since 2011 Dual deep neural networks using ResNet-101 architecture Scottish DES datasets; validation on 78,601 images / 33,535 patients DR triage: “no DR & gradable” vs “manual grading required”; ungradable detection; DLAG aims to further reduce manual grading

AI artificial intelligence, CE Conformité Européenne, CDSCO Central Drugs Standard Control Organization (India), CNN convolutional neural network, DL deep learning, FDA Food and Drug Administration (USA), HSA Health Sciences Authority (Singapore), MDD Medical Device Directive, MDR Medical Device Regulation (EU), ML machine learning, NMPA National Medical Products Administration (China), SaMD software as a medical device, SFDA Saudi Food and Drug Authority, TFDA Taiwan Food and Drug Administration, TGA Therapeutic Goods Administration (Australia).

Technically, the ecosystem remains dominated by Inception-family backbones (v3, v4, Inception-ResNet-v2) deployed either as single models (EyeCheckup, CARE, VeriSee DR, VISUHEALTH-AI DR) or as multi-CNN ensembles (IDx-DR with 5-CNN, EyeArt multi-CNN, ARDA® DL with 10× Inception-v3, RetCAD™ with 5-CNN, AIDRScreening with 3-CNN). Newer products increasingly adopt EfficientNet (Aireen®, LuxIA, EyeWisdom) or EfficientNet-V2 hybrids for improved parameter-efficiency, whereas Medios AI™ employs a MobileNet-lite branch to support on-device smartphone inference. Alternative strategies include DenseNet-121 + bilinear pooling (AI-100), VGG-19 ensemble (SELENA + ™ with 8× VGG-19), ResNet-101 dual networks (Autograder), feature-pyramid ensembles (VoxelCloud Retina with 6× Inception-ResNet-v2 + FPN), and proprietary CNN architectures with heat-map lesion localization (OphtAI). Retmarker employs a deep-learning classifier with MA-turnover biomarker analysis for progression and DME risk assessment. In short, although cleared screeners now span a broad geography and a widening range of regulators, most still converge on convolutional backbones from the Inception/EfficientNet families, often boosted by ensemble voting to maximize robustness across diverse imaging hardware and care settings.

Quality assessment

According to QUADAS-2 criteria, 56.1% of studies were rated low risk of bias for patient selection, 57.3% for the index test, 52.4% for the reference standard, and 41.5% for flow and timing. Flow and timing remained a frequent issue, with 17.1% of studies showing high risk—often due to substantial image exclusion or delays between clinical evaluation and AI testing. Unclear risk of bias was also common, particularly for the index test (42.7%) and patient selection (41.5%), with the reference standard showing 37.8% unclear ratings. Applicability concerns were generally minor: 76.8% of studies had low concern for patient spectrum, 92.7% for the index test, and 52.4% for the reference standard. However, 45.1% had unclear applicability for the reference standard, reflecting inconsistency with ETDRS or Wisconsin Reading Center protocols. In summary, most evaluations were methodologically sound and clinically representative, though approximately 1 in 6 exhibited high risk in either flow/timing (17.1%) or reference standard (9.8%) domains. These results underscore the need for prospectively registered, fully masked diagnostic-accuracy studies (Supplementary Fig. 1; Table 5).

Overall meta-analysis of patient-wise regulator-approved deep learning system

The hierarchical random-effects bivariate model revealed high between-study heterogeneity, particularly for specificity (τ² = 1.02; I² = 86.81%) and, to a lesser extent, for sensitivity (τ² = 0.80; I² = 31.77%). The negative covariance (–0.31) and correlation (ρ = –0.34) indicate the expected trade-off: studies reporting higher sensitivity tended to show lower specificity and vice versa. A highly significant likelihood-ratio test (χ² = 29,231, df = 3, p < 0.001) confirms that the random-effects (RE) specification fits the data far better than a fixed-effects (FE) approach, validating the choice to model heterogeneity explicitly.

After multivariable adjustment in a bivariate random-effects meta-regression, the pooled summary absolute accuracy of 115 tables of regulator-approved AI screeners remained high: summary sensitivity 0.93 (95% CI 0.91–0.95) and summary specificity 0.90 (95% CI 0.87–0.92) (Fig. 2a, Supplementary Fig. 2). These figures suggest that, on average, cleared deep-learning systems correctly identify more than nine out of ten true cases of diabetic retinopathy while correctly classifying roughly nine out of ten non-cases—performance levels broadly comparable to expert human graders. Deeks’ funnel-plot test revealed marginal evidence of publication bias across studies (p = 0.04, Supplementary Fig. 3).

Fig. 2. Hierarchical summary receiver operating characteristic curves of regulator-approved deep-learning systems for diabetic retinopathy detection.

Fig. 2

a Patient-level HSROC curve showing the pooled model fit, 95% confidence and prediction regions, and study points scaled to weight. b Eye-level HSROC curve with the same visual elements, summarizing performance per eye.

Meta-regression of patient-wise regulator-approved deep learning system

Supplementary Table 6 presents the univariate meta-regression, revealing that while the overall diagnostic accuracy of regulator-approved deep-learning systems was high, several prespecified moderators contributed significantly to between-study heterogeneity. Studies conducted in low- and middle-income countries reported significantly higher specificity than those from high-income settings (0.93 vs 0.86; p < 0.01). Algorithm target influenced performance: systems detecting any DR showed reduced sensitivity and specificity (0.92 and 0.91, respectively) compared to those focused on referable or vision-threatening DR (p < 0.01 for both). Datasets with low image gradability yielded higher sensitivity (0.95 vs 0.92; p < 0.01) but lower specificity (0.82 vs 0.92; p < 0.01). The absence of vendor involvement produced a similar trade-off, with higher sensitivity (0.95 vs 0.87) but reduced specificity (0.85 vs 0.95; p < 0.01 for both). Single-field imaging was associated with higher specificity than two-field protocols (0.95 vs 0.88; p < 0.01). Portable fundus cameras also achieved higher specificity than desktop devices (0.95 vs 0.90; p < 0.01), though this comparison was underpowered (n = 3). Use of diagnostic criteria other than International Clinical Diabetic Retinopathy (ICDR), Early Treatment Diabetic Retinopathy Study (ETDRS) or National Health Service (NHS) was associated with higher specificity (0.92 vs 0.89; p < 0.01). Although healthcare setting and reference image modality significantly explained heterogeneity in sensitivity and specificity, respectively, they did not show statistically significant differences between subgroup means. Other factors—including grader discipline, study design, pupil dilation protocol, and article type—were not significantly associated with accuracy (p > 0.05 for all).

The multivariate meta-regression (Fig. 3a, Table 3) confirmed the independent contributions of DR category, national income level, diagnostic criteria, reference image modality, and image gradability to diagnostic performance, substantially improving model fit compared to the base model (AIC 3935 vs 5105 Supplementary Table 7). Portable fundus cameras and use of non-ICDR diagnostic criteria were consistently associated with higher specificity, while non-vendor-sponsored studies demonstrated a trade-off of higher sensitivity at the expense of lower specificity. In contrast, high-gradability datasets showed the opposite pattern, with slightly reduced sensitivity but improved specificity.

Fig. 3. Subgroup forest plots by diabetic-retinopathy subtype for regulator-approved deep-learning systems.

Fig. 3

a Patient-level analysis displaying effect sizes and 95% confidence intervals for each DR-type subgroup, with pooled estimates and weights shown on the right; b Eye-level analysis built and displayed in the same manner.

Table 3.

Multivariate meta-regression of patient-wise regulator-approved deep learning system

Moderator Category N Absolute Sensitivity Relative Sensitivity p-value (group-diff) p-value (LR test) Absolute Specificity Relative Specificity p-value (group-diff) p-value (LR test)
Economic status High income 61 0.94 (0.91 ~ 0.96) 1.01 (0.97 ~ 1.06) 0.52 <0.01* 0.86 (0.80 ~ 0.91) 0.94 (0.87 ~ 1.02) 0.05 <0.01*
Low Middle income 54 0.93 (0.89 ~ 0.96) 1.00 (1.00 ~ 1.00) - 0.92 (0.87 ~ 0.95) 1.00 (1.00 ~ 1.00) -
Healthcare setting Primary 54 0.91 (0.88 ~ 0.94) 0.95 (0.90 ~ 1.02) 0.15 0.43 0.91 (0.88 ~ 0.94) 1.18 (0.87 ~ 1.61) 0.29 0.46
Secondary 19 0.97 (0.94 ~ 0.99) 1.02 (0.96 ~ 1.08) 0.64 0.81 (0.62 ~ 0.92) 1.05 (0.73 ~ 1.51) 0.79
Tertiary 39 0.93 (0.90 ~ 0.96) 0.97 (0.92 ~ 1.04) 0.46 0.89 (0.83 ~ 0.94) 1.16 (0.85 ~ 1.58) 0.35
Mixed 3 0.96 (0.89 ~ 0.99) 1.00 (1.00 ~ 1.00) - 0.77 (0.47 ~ 0.93) 1.00 (1.00 ~ 1.00) -
Diabetic retinopathy type Any DR 35 0.92 (0.90 ~ 0.94) 0.98 (0.96 ~ 1.00) 0.03* 0.01* 0.90 (0.87 ~ 0.93) 0.99 (0.98 ~ 0.99) <0.01* <0.01*
Referable DR 58 0.94 (0.93 ~ 0.96) 1.00 (0.99 ~ 1.02) 0.61 0.87 (0.84 ~ 0.90) 0.95 (0.94 ~ 0.97) <0.01*
Vison threatening DR 22 0.94 (0.91 ~ 0.96) 1.00 (1.00 ~ 1.00) - 0.92 (0.89 ~ 0.94) 1.00 (1.00 ~ 1.00) -
Camera type Desktop 97 0.94 (0.92 ~ 0.95) 1.00 (1.00 ~ 1.00) - 0.90 0.89 (0.85 ~ 0.92) 1.00 (1.00 ~ 1.00) - 0.18
Smartphone 15 0.92 (0.83 ~ 0.97) 0.98 (0.91 ~ 1.06) 0.65 0.91 (0.79 ~ 0.96) 1.03 (0.92 ~ 1.14) 0.63
Portable 3 0.91 (0.83 ~ 0.95) 0.97 (0.91 ~ 1.03) 0.33 0.94 (0.91 ~ 0.97) 1.06 (1.03 ~ 1.10) <0.01*
Diagnostic criteria ICDR 75 0.93 (0.91 ~ 0.95) 1.00 (1.00 ~ 1.00) - 0.84 0.89 (0.85 ~ 0.92) 1.00 (1.00 ~ 1.00) - <0.01*
ETDRS 19 0.94 (0.91 ~ 0.96) 1.02 (0.97 ~ 1.07) 0.56 0.87 (0.75 ~ 0.94) 0.98 (0.87 ~ 1.10) 0.76
NHS 7 0.95 (0.84 ~ 0.98) 1.02 (0.95 ~ 1.09) 0.94 0.81 (0.53 ~ 0.94) 0.91 (0.71 ~ 1.17) 0.47
Other 14 0.94 (0.91 ~ 0.96) 1.01 (0.98 ~ 1.04) 0.46 0.94 (0.91 ~ 0.96) 1.05 (1.03 ~ 1.08) <0.01
Reference image Color fundus image 102 0.94 (0.92 ~ 0.95) 1.00 (1.00 ~ 1.00) - 0.44 0.90 (0.87 ~ 0.92) 1.00 (1.00 ~ 1.00) - <0.01*
With advanced image 13 0.93 (0.83 ~ 0.97) 1.00 (0.3 ~ 1.07) 0.85 0.80 (0.58 ~ 0.92) 0.89 (0.72 ~ 1.11) 0.30
Field strategy 1 19 0.95 (0.90 ~ 0.97) 1.00 (1.00 ~ 1.00) - 0.07 0.92 (0.85 ~ 0.96) 1.00 (1.00 ~ 1.00) - 0.16
2 74 0.93 (0.90 ~ 0.95) 0.98 (0.94 ~ 1.02) 0.33 0.90 (0.85 ~ 0.93) 0.97 (0.90 ~ 1.05) 0.61
>2 20 0.95 (0.90 ~ 0.98) 1.01 (0.96 ~ 1.06) 0.77 0.84 (0.69 ~ 0.93) 0.94 (0.80 ~ 1.09) 0.34
Vendor involved No 79 0.95 (0.93 ~ 0.96) 1.09 (1.03 ~ 1.15) <0.01* 0.61 0.86 (0.82 ~ 0.89) 0.90 (0.87 ~ 0.93) <0.01* 0.13
Yes 36 0.87 (0.82 ~ 0.91) 1.00 (1.00 ~ 1.00) - 0.96 (0.94 ~ 0.97) 1.00 (1.00 ~ 1.00) -
Gradable High 73 0.92 (0.90 ~ 0.94) 0.97 (0.94 ~ 0.99) 0.01* <0.01* 0.92 (0.89 ~ 0.94) 1.12 (1.08 ~ 1.17) <0.01* <0.01*
Low 42 0.95 (0.93 ~ 0.97) 1.00 (1.00 ~ 1.00) - 0.82 (0.77 ~ 0.86) 1.00 (1.00 ~ 1.00) -

Values represent pooled sensitivity and specificity with 95% confidence intervals. P-values refer to differences in performance between subgroups within each moderator (Wald-type tests). Significant comparisons (p < 0.05) are marked with an asterisk ().*

DR diabetic retinopathy, ICDR International Clinical Diabetic Retinopathy severity scale, ETDRS Early Treatment Diabetic Retinopathy Study, NHS National Health Service, LR test Likelihood Ratio Test, p-value probability value, N number of studies, CI confidence interval.

Overall meta-analysis of eye-wise regulator-approved deep learning system

The hierarchical random-effects bivariate model showed considerable between-study heterogeneity—greater for specificity (τ² = 0.62; I² = 85.53%) and somewhat lower for sensitivity (τ² = 0.79; I² = 37.63%). A small covariance (0.04) and weak correlation (ρ = 0.05) illustrate the familiar trade-off: studies achieving higher sensitivity tended to report lower specificity, and vice versa. A likelihood-ratio test strongly favored the random-effects specification over a fixed-effects alternative (χ² = 127,870, df = 3, p < 0.001), confirming the need to model heterogeneity explicitly. After multivariable adjustment in a bivariate random-effects meta-regression, the pooled summary of 88 tables remained high. The pooled sensitivity was 0.92 (95% CI 0.89–0.94) and the pooled specificity was 0.93 (95% CI 0.91–0.94) (Fig. 2b, Supplementary Fig. 4), indicating that cleared DL algorithms correctly detect more than nine out of ten true cases of referable diabetic retinopathy while correctly excluding roughly nine out of ten non-cases—performance comparable to expert human graders. Deeks’ funnel-plot test suggested no evidence of publication bias for these regulator-approved systems (p = 0.23; Supplementary Fig. 5).

Meta-regression of eye-wise regulator-approved deep learning system

Most moderators resulted in only modest shifts in diagnostic accuracy, but several effects reached statistical significance. Sensitivity was significantly higher in studies using mixed pupil-dilation protocols—where pharmacologic dilation was applied selectively—showing an 8% increase (p = 0.01), and when the reference standard included advanced imaging modalities such as ultra-wide-field photography or OCT, with a 5% gain (p = 0.04). In contrast, specificity was lower in studies conducted in high-income countries (−4%), in those targeting any DR rather than referable or vision-threatening DR (−2%; p < 0.01), and in datasets with high image gradability (−6%; p < 0.01), consistent with the sensitivity–specificity trade-offs observed in patient-level evaluations. Other moderators—including healthcare setting, camera form factor, study design, diagnostic criteria, index-test grader, field strategy, and vendor involvement—did not significantly contribute to heterogeneity (all p > 0.05; Supplementary Table 8).

Multivariate meta-regression incorporating five moderators (Fig. 3b, Table 4)—economic status, DR category, pupil dilation, reference-image modality, and image gradability—substantially improved model fit over the base model (AIC 78570 vs 6550; Supplementary Table 9). Based on likelihood ratio tests, economic status (specificity, p < 0.01) and reference-image modality (both sensitivity and specificity, p < 0.01) showed overall significant effects, while DR category, pupil dilation, and gradability did not reach significance in the omnibus tests. However, individual group comparisons revealed important patterns: Compared to systems targeting vision-threatening DR, algorithms detecting any DR had 4% higher specificity (relative specificity 1.04; p = 0.01) with a corresponding 17% lower sensitivity (relative sensitivity 0.83; p < 0.01), suggesting threshold dispersion across severity definitions. Referable DR systems showed performance similar to vision-threatening DR systems (relative sensitivity 0.99, relative specificity 1.00; both p < 0.01). Mixed dilation protocols showed a trend toward 3-6% higher sensitivity compared to non-mydriatic or mydriatic imaging alone, though these differences did not reach statistical significance (p = 0.15). Advanced imaging methods (including OCT) showed a 4% sensitivity advantage (relative sensitivity 1.04; p = 0.18), though this was not statistically significant after adjustment. High-gradability datasets maintained significantly higher specificity than low-gradability datasets (6% difference; p < 0.01), with no difference in sensitivity (p = 0.98). High-income settings showed slightly lower specificity than low-middle-income settings (relative specificity 0.99; p < 0.01 by LR test).

Table 4.

Multivariate meta-regression of eye-wise regulator-approved deep learning system

Moderator Category N Absolute Sensitivity Relative Sensitivity p-value (group-diff) p-value (LR test) Absolute Specificity Relative Specificity p-value (group-diff) p-value (LR test)
Economic status High income 35 0.93 (0.89 ~ 0.96) 1.01 (0.96 ~ 1.06) 0.69 0.98 0.94 (0.91 ~ 0.96) 0.99 (0.95 ~ 1.02) 0.50 <0.01*
Low Middle income 53 0.92 (0.88 ~ 0.95) 1.00 (1.00 ~ 1.00) - 0.95 (0.92 ~ 0.96) 1.00 (1.00 ~ 1.00) -
Diabetic retinopathy type Any DR 14 0.79 (0.72 ~ 0.84) 0.83 (0.78 ~ 0.87) <0.01* 0.31 0.97 (0.96 ~ 0.98) 1.04 (1.03 ~ 1.05) 0.01* 0.29
Referable DR 64 0.94 (0.92 ~ 0.96) 0.99 (0.98 ~ 1.01) <0.01* 0.94 (0.92 ~ 0.95) 1.00 (1.00 ~ 1.00) <0.01*
Vison threatening DR 10 0.95 (0.93 ~ 0.96) 1.00 (1.00 ~ 1.00) - 0.93 (0.91 ~ 0.95) 1.00 (1.00 ~ 1.00) -
Pupil dilation Yes 12 0.90 (0.81 ~ 0.95) 0.94 (0.87 ~ 1.02) 0.15 0.22 0.90 (0.83 ~ 0.95) 0.97 (0.90 ~ 1.04) 0.33 0.19
No 57 0.92 (0.88 ~ 0.95) 0.97 (0.92 ~ 1.01) 0.15 0.95 (0.93 ~ 0.97) 1.02 (0.97 ~ 1.06) 0.48
Mixed 19 0.95 (0.92 ~ 0.98) 1.00 (1.00 ~ 1.00) - 0.94 (0.89 ~ 0.96) 1.00 (1.00 ~ 1.00) -
Reference image Color fundus image 84 0.93 (0.90 ~ 0.95) 1.00 (1.00 ~ 1.00) - <0.01* 0.94 (0.93 ~ 0.96) 1.00 (1.00 ~ 1.00) - <0.01
With advanced image 4 0.96 (0.87 ~ 0.99) 1.04 (0.98 ~ 1.10) 0.18 0.90 (0.76 ~ 0.97) 0.96 (0.86 ~ 1.06) 0.41
Gradable High 38 0.93 (0.89 ~ 0.95) 1.00 (1.00 ~ 1.00) - 0.69 0.90 (0.87 ~ 0.93) 1.00 (1.00 ~ 1.00) - 0.50
Low 50 0.93 (0.89 ~ 0.95) 1.00 (0.96 ~ 1.04) 0.98 0.96 (0.95 ~ 0.97) 0.94 (0.92 ~ 0.96) <0.01

Values represent pooled sensitivity and specificity with 95% confidence intervals. P-values refer to differences in performance between subgroups within each moderator (Wald-type tests). Significant comparisons (p < 0.05) are marked with an asterisk ().*

DR diabetic retinopathy, LR test Likelihood ratio test, p-value probability value, N number of studies, CI confidence interval.

Discussion

Our review screened 8404 records and ultimately meta-analyzed 82 eligible studies evaluating regulator-approved deep-learning systems for diabetic retinopathy screening. Pooled accuracy was high—patient-wise sensitivity 0.93 (95% CI 0.91–0.95) and specificity 0.90 (95% CI 0.87–0.92); eye-wise 0.92 (95% CI 0.89–0.94) and specificity was 0.93 (95% CI 0.91–0.94), respectively—yet substantial between-study heterogeneity (τ² ≈ 0.5–0.9) justified random-effects modeling. Multivariate meta-regressions revealed six consistent performance modifiers: DR severity tier, national income level, image gradability, pupil dilation status, diagnostic criteria, and reference image. In particular, “any-DR” tasks, low-income settings, and high-gradability datasets reduce sensitivity, while low-income settings, “vtDR” task, standard color fundus as a reference image, other diagnostic criteria than ICDR, ETDRS, NHS, and mixed dilated pupil protocol, boosted specificity. Eye-level analyses echoed these patterns but with smaller effect sizes; once confounding was controlled, only DR tier and gradability retained independent influence. Multivariate models halved unexplained variance (AIC 5439 → 3909; τ² 0.077 → 0.041). Deeks’ tests indicated marginal publication bias for patient-level results (p = 0.04) and none for eye-level data (p = 0.23). Together, these findings confirm that cleared DL algorithms perform comparably to expert graders while highlighting context-specific trade-offs that should guide real-world deployment.

Despite the uniformly high accuracy, we observed substantial heterogeneity in specificity across studies. Added evidence on co-pathology performance. A further source of real-world uncertainty is the presence of retinal co-morbidities—most commonly age-related macular degeneration (AMD)—whose lesions can mimic or obscure diabetic lesions. Four prospective or retrospective evaluations38,43,62,77 deliberately enrolled mixed DR + AMD cohorts and reported AUCs of 0.88–0.98 with sensitivities ≥84%. However, specificities fell to 66–93%, highlighting a consistent false-positive penalty when exudates or hemorrhages overlap with drusen or geographic atrophy. He et al.44 further showed that cross-camera domain shifts degrade agreement more for AMD than for DR, underscoring the need for multi-disease, multi-device validation before deployment. These findings confirm that current DL algorithms remain feasible in heterogeneous case-mix but that additional multi-center trials—including other look-alike pathologies such as vein occlusion or myopic choroidal neovascularization—are required to quantify workflow impact and cost-effectiveness. In practice, the reported sensitivities and specificities varied, reflecting differences in populations, settings, and implementation. A classic threshold trade-off was evident: studies that reported extremely high sensitivity tended to have lower specificity, and vice versa98. This negative correlation suggests that some implementations tuned the algorithm to prioritize catching all possible DR (yielding very few false negatives at the expense of more false positives), whereas others set a stricter threshold to reduce false positives98. Such threshold effects are common in diagnostic tests—if you “cast a wider net” to catch every case, you inevitably pull in a few healthy eyes as positive. In the UK, the UK National Screening Committee (UK NSC) emphasizes that service load is highly sensitive to specificity and supports risk-stratified recall intervals within the English Diabetic Eye Screening Program (DESP).

Meta-regression analyses revealed substantial heterogeneity in the diagnostic performance of regulator-approved deep learning (DL) systems for diabetic retinopathy (DR) detection across clinical, technical, and contextual moderators. Among these, disease severity tier emerged as a consistent and statistically significant source of variation. Systems evaluated for any DR demonstrated lower sensitivity (by 2 percentage points [pp]) and lower specificity (by 2 pp) compared to those targeting referable or vision-threatening DR (p < 0.01), with an additional 4 pp specificity gain at the eye level with 17% corresponding sensitivity loss. These findings reflect an inherent trade-off between early detection and false-positive burden. While broader screening thresholds may increase case-finding for mild DR, they also heighten the risk of false positives, potentially overwhelming follow-up services. Programs aiming to detect mild disease must therefore implement second-reader workflows, risk-based triage, or structured recall systems to prevent unnecessary escalation of low-risk findings. In the UK, UK NSC guidance recommends risk-stratified recall intervals (e.g., 12–24 months vs 6 months) within DESP to absorb the additional false-positive burden when “any-DR” thresholds are used99.

The analysis also identified clear performance discrepancies across geographic and health system contexts. Studies conducted in low- and middle-income countries (LMICs) exhibited modestly higher sensitivity (+1 pp) but substantially reduced specificity (–6 pp; p ≤ 0.01). This likely reflects higher background disease prevalence, a wider spectrum of disease presentation, and variable image quality, all of which challenge systems primarily trained on high-fidelity datasets from high-income countries. In such contexts, the increased false-positive burden can pose a serious threat to already constrained healthcare infrastructures. This underscores the need for context-aware deployment strategies—including tiered referral systems, integration with teleophthalmology, and infrastructure reinforcement to accommodate the downstream impact of automated detection systems.

Image gradability moderated DL performance differently depending on the unit of analysis. At the patient level, low-gradability datasets (N = 42) showed +3-percentage-point higher sensitivity (0.95 vs 0.92; p = 0.01) but—9-percentage-point lower specificity (0.84 vs 0.93; p < 0.01), suggesting systems err toward caution with poor-quality images—a “safety-positive” behavior that may inflate referrals. At the eye level (Table 4), high-gradability datasets (N = 38) paradoxically showed—6-percentage-point lower specificity (0.90 vs 0.96; p < 0.01) with no significant sensitivity difference, likely reflecting that datasets excluding more ungradable images retain harder borderline cases. These findings support real-time image quality assessment and recapture prompts at point of care to ensure optimal DL performance and minimize unnecessary escalations. The UK NSC calls for standardized gradability reporting to enable cross-site benchmarking and automated audit pipelines99.

Vendor involvement emerged as a significant moderator at the patient level: non-industry studies (N = 79) reported +7-percentage-point higher sensitivity (0.95 vs 0.88; p < 0.01) but—10-percentage-point lower specificity (0.86 vs 0.96; p < 0.01) compared to vendor-involved evaluations (N = 36). However, at the eye level (Table 4), vendor involvement showed no significant effect: non-vendor studies (N = 46) had only—3-percentage-point lower sensitivity (0.90 vs 0.93; p = 0.21) and identical specificity (0.93 vs 0.93; p = 0.61) compared to vendor-involved studies (N = 42). This discrepancy between patient-level and eye-level analyses suggests the patient-level differences may reflect study design factors rather than vendor influence per se. These findings underscore the need for preregistered, independently conducted validations to complement regulatory submissions. The UK National Screening Committee (UK NSC) lists ‘independent, head-to-head evaluations on identical datasets’ as a core evidence domain for adoption99.

On the technical front, portable fundus cameras delivered significantly higher specificity (+5 pp; p ≈ 0.01) without a loss in sensitivity, affirming their value in mobile or community-based screening programs, particularly in resource-limited environments. Smartphone-based systems, in contrast, showed no significant deviation in performance relative to desktop platforms (p > 0.50), suggesting functional equivalence across capture modalities in adequately trained models. Reference imaging modality was significantly associated with between-study heterogeneity, although within-group differences were not statistically significant, suggesting that certain advanced imaging techniques (e.g., ultra-widefield [UWF], OCT) may offer marginal sensitivity benefits that dissipate after adjustment for confounders. Similarly, single-field imaging protocols demonstrated higher specificity in univariate analysis (+7 pp; p < 0.01), although this benefit did not persist in multivariate models, indicating that field strategy alone is not a dominant performance driver.

Other factors—including study design, healthcare tier, pupil dilation, diagnostic criteria, index test grader, field strategy, camera brand, and article type—were not significantly associated with performance in adjusted models (all p > 0.05). This suggests a reassuring robustness of pooled estimates across a wide range of operational conditions and affirms the validity of aggregated performance metrics for regulatory and policy decision-making. However, interaction modeling provided further nuance: performance gains associated with non-vendor studies were concentrated in high-gradability datasets, while specificity losses were most pronounced in low-gradability scenarios. Portable camera advantages persisted across interaction strata, highlighting their resilience to variable imaging conditions.

Taken together, these findings emphasize that the diagnostic performance of DL systems is not static, but rather deeply context-sensitive—shaped by disease target, image quality, geographic setting, technical infrastructure, and transparency of evaluation. As such, regulatory frameworks should move beyond one-size-fits-all benchmarks and adopt dynamic, context-aware approval and surveillance models that reflect real-world deployment conditions. Clinically, implementation should be accompanied by workflow adaptations to address false-positive burden and optimize human-AI collaboration. For health systems—especially those in LMICs—investment in triage capacity, image quality assurance, and portable imaging technology will be essential to realizing the full benefits of AI-assisted DR screening while maintaining efficiency and equity.

Our findings are broadly consistent with and extend previous meta-analyses of deep learning algorithms for diabetic retinopathy (DR) detection. Earlier reviews focusing on experimental, non-regulatory systems reported pooled sensitivities between 87–90% and specificities around 88–92%100. In contrast, our analysis—restricted to regulator-approved systems—demonstrates slightly higher sensitivity (93%) with comparable specificity (90%), suggesting that the regulatory pathway may favor top-performing models. For instance, our pooled sensitivity surpasses the 87.4% reported in the pivotal trial of the first FDA-cleared system, IDx-DR14, and aligns closely with EyeArt’s U.S. FDA trial (96% sensitivity, 88% specificity)15 and UK real-world deployments showing 95–97% sensitivity68.

Autonomous AI screening offers a scalable solution for expanding DR detection beyond ophthalmology clinics. It enables point-of-care use in primary care, endocrinology, and community settings—reaching patients who might otherwise miss annual exams. In a randomized trial, 100% of pediatric diabetes patients completed screening with autonomous AI compared to 22% under standard referral pathways101. In the UK, AI deployment halved the grading workload while maintaining >95% sensitivity68. These findings support a safe shift toward upstream screening with manageable trade-offs. However, regulatory clearance is only the first step. The UK NSC sets out seven evidence domains—pathway-level accuracy, prospective clinical impact, cost-effectiveness, equity analysis, vendor-independent validation, interoperability testing, and post-market surveillance—for the UK context99. These domains map onto our moderator findings and research gaps. Early engagement between developers and screening agencies, plus staged pilots using stepped-wedge or test-treat RCT designs, are highlighted as preferred evaluation routes99. Policy shifts, such as U.S. Medicare reimbursement for Current Procedural Terminology (CPT) 92229, signal growing recognition of AI’s clinical value102. Scaling these tools globally will require national screening integration and reimbursement models that support sustainability. Importantly, our review highlights a persistent evidence gap in low-resource settings. Only one included study was from a low-income country, underscoring the need for greater geographic representation. Smartphone-based fundus imaging combined with AI shows promise for underserved regions26,45,87, but robust validation, public-sector procurement, and international pilot programs will be essential. To avoid widening health inequities, global AI deployment must be accompanied by investments in image quality assurance, usability training, and infrastructure.

This is the first systematic review and meta-analysis focused exclusively on regulator-approved AI systems for diabetic retinopathy (DR) screening. Key strengths include the largest synthesis to date of diagnostic accuracy data for these systems and a uniquely detailed moderator analysis. Notably, we examined previously underreported variables such as vendor involvement and image gradability, which the latter contributed significantly to between-study heterogeneity. Our emphasis on these dimensions directly addresses four of the NSC’s seven evidence domains (accuracy, equity, independence, and image quality), underscoring the translational value of our findings. We also separated patient-wise and eye-wise outcomes to avoid conflating correlated units of analysis—a limitation seen in prior studies. Importantly, we applied a hierarchical bivariate random-effects model with extensions to account for within-study correlation between sensitivity and specificity, inter-moderator dependencies, and study-level clustering, representing one of the most comprehensive modeling frameworks applied to this field.

A key limitation is the inconsistent reporting of image gradability; future studies should adopt a standardized gradability metric to facilitate cross-study comparisons. In addition, patient-level factors that plausibly affect gradability—such as age, cataract, or neurocognitive disorders—could not be examined because most publications lacked individual-patient data; incorporating these covariates in upcoming datasets would enable a more granular exploration of accuracy modifiers. Software versioning was not uniformly documented across studies. Because different versions (e.g., v1 vs v2.2) may materially affect performance and thresholds, incomplete reporting prevented version-specific subgroup or sensitivity analyses. Also, Our publication-based search strategy may have missed some regulator-approved algorithms that lack peer-reviewed validation studies. While publication bias could not be ruled out, we observed no significant asymmetry in funnel plots. Additionally, although the inclusion of a broad range of algorithms enhances generalizability, it also introduces heterogeneity. Fewer than half of the included systems have been stress-tested outside the setting in which they were developed, underscoring the need for prospective, multi-regional validation, particularly in low-resource and ethnically diverse environments. Despite these limitations, our findings are robust and have strong implications for real-world deployment. Future work should include head-to-head trials of multiple AI systems, implementation research assessing clinical outcomes and workflow impact, and the development of automated audit pipelines. Standardized Automated Retinal Image Analysis Systems (ARIAS) application programming interfaces (APIs) and open audit dashboards, as advocated by the UK NSC in the UK program, would also streamline ongoing performance monitoring. Further studies should evaluate cost-effectiveness, prioritize low-resource settings, and explore multi-disease screening with fundus photography.

In conclusion, regulator-approved deep-learning systems have reached—and, in many contexts, surpassed—expert-grader accuracy for diabetic-retinopathy screening, correctly identifying ≈93% of referable cases while excluding ≈87–92% of non-cases. Yet their performance is not monolithic: disease-severity threshold, economic context, image gradability, vendor sponsorship, camera portability, and grader expertise each shift the sensitivity–specificity balance, explaining roughly half of the observed heterogeneity. Programs that target any DR, operate in low-/middle-income settings, or work with low-quality images must plan for a heavier false-positive load, whereas those using portable cameras and retina-specialist adjudication can expect higher specificity. Independent, preregistered evaluations and post-market surveillance are essential to mitigate sponsor bias and dataset drift, and future studies should adopt a standardized gradability metric and report patient-level covariates (e.g., age, cataract) to enable finer-grained moderator analyses. With robust quality-assurance loops, harmonized regulatory oversight, and equitable reimbursement models, autonomous AI can extend high-quality DR screening to the millions who currently lack access, helping to avert preventable vision loss on a global scale.

Methods

General guideline

This systematic review was prospectively registered with INPLASY (registration number: INPLASY202550099)103. We followed the relevant items of the Preferred Reporting Items for Systematic Reviews and Meta-Analyses of Diagnostic Test Accuracy Studies (PRISMA-DTA) checklist (see Supplementary Table 1,2 for item-by-item compliance)104,105 to quantify the diagnostic accuracy of all regulator-approved deep-learning systems for fundus-based diabetic retinopathy detection and identify key factors influencing their performance across diverse global settings. Title and abstract screening, full-text eligibility assessment, data extraction, and risk-of-bias appraisal were each conducted independently and in duplicate by two reviewers (T-WW, L-WT); disagreements were resolved through discussion with a third reviewer. Because the review relied solely on previously published, de-identified data, no institutional ethical approval or informed consent was required.

Database searches and identification of eligible manuscripts

Two reviewers (T-W.W. and L-W.T.) systematically searched PubMed, Embase, and ClinicalTrials.gov from database inception to 3 April 2025 for studies that validated a regulator-approved deep-learning system for detecting diabetic retinopathy on color fundus photographs. For each database, we used a structured search strategy combining three concept blocks—condition (diabetic retinopathy and related retinal diseases), technology (artificial intelligence, machine learning, deep learning, automated detection), and system/regulatory context (named commercial algorithms and regulator-approval terms)—with database-specific subject headings (e.g., MeSH, Emtree) and title/abstract fields, without language restrictions. The comprehensive search strategy is detailed in Supplementary Table 3. Records were imported into the Nested Knowledge™ platform for automatic de-duplication and audit-trailed screening. Titles/abstracts and full texts were assessed in duplicate against prespecified criteria; disagreements were resolved by consensus.

We included prospective or retrospective diagnostic-accuracy studies of deep-learning algorithms for DR screening using color fundus photographs in adults (≥18 years); mixed-age cohorts were eligible if ≥85% were adults or if adult-only data allowed construction of a 2 × 2 table. Any clinical screening/diagnostic setting was eligible; country/region refers to the validation dataset location. Index test: DL autonomous or DL-assisted classifier on fundus photographs (excluded: non-DL methods, non-fundus modalities such as OCT/OCTA, and non-image risk models). Reference standard: human graders using recognized DR schemes (excluded: algorithm-only references). Target conditions: any DR grade; analyses prioritized prespecified thresholds for referable DR (≥ moderate non-proliferative diabetic retinopathy (NPDR) ± diabetic macular edema (DME)) and vision-threatening DR, using author-defined cut-points. Study design: must report, or permit calculation of, TP/FP/FN/TN; excluded: development-only studies without an external or distinct hold-out validation. Primary pooling used independent validation only, defined a priori as evaluations on datasets not used for model development. Report characteristics: searches from inception to 3 April 2025, no language restrictions; we meta-analyzed peer-reviewed full texts only, using preprints/abstracts for identification and citation chasing unless superseded before the cutoff. We also performed hand searching of references and backward/forward citation chasing. Where studies reported distinct strata with unique 2 × 2 tables (e.g., by camera, threshold, mydriasis, model version, or grader/reference image), each stratum was treated as an independent sub-study.

Data extraction and management

Titles and abstracts were screened independently by two reviewers, followed by a full-text eligibility assessment. Data were captured with a prespecified electronic extraction form, and every exclusion—together with its rationale—was documented in the PRISMA flow diagram. For each eligible study, we recorded (i) bibliographic details (first author, publication year, country (geographic location of the validation dataset)); (ii) study design and context (prospective/retrospective, recruitment period, clinical setting, regulatory-clearance pathway, external validation in different country from training set); (iii) participant characteristics (sample size, mean age, proportion female, economic status (World-Bank income category106)); (iv) index-test specifications (device name, training data country and ethics, fundus-camera model, mydriatic vs non-mydriatic protocol, number and field of images, algorithm name, vendor involvement (participation in coauthor ship or provision of funding support by the algorithm manufacturer), regulatory certification, model architecture, image gradability); (v) reference standard (grading protocol and grader expertise); and (vi) diagnostic-accuracy data (2 × 2 confusion matrix, sensitivity, specificity, 95% CIs). We confirmed status in primary regulator databases (FDA De Novo/510(k), EU CE-MDR/MDD, UKCA, TGA-ARTG, HSA-SMDR, Philippines FDA, NMPA; MDALL/PMDA as needed). If no primary record was available, we corroborated manufacturer pages and press releases. When results were stratified by data source, camera type, AI model, DR severity, mydriasis status, and index test grader or reference image, we treated each stratum as a separate sub-study and entered it as an independent table in the meta-analysis.

Quality assessment

We appraised methodological quality using the QUADAS-2 framework107. Two reviewers (T-W.W. and L-W.T.) first piloted the QUADAS-2 signaling questions on a random set of five studies to calibrate interpretation, then independently applied the tool to each included record. After independent scoring, disagreements were resolved through discussion; if consensus was not reached, a third author adjudicated. Each study was rated for risk of bias and applicability concerns in four domains: patient selection, index test, reference standard, and flow & timing, classified as “low,” “high,” or “unclear.” Where studies included multiple algorithm versions or imaging protocols, we evaluated the analysis with the highest detail level available to capture AI-specific bias, such as data leakage or threshold optimization.

Statistical analysis

We first reconstructed 2 × 2 contingency tables for each eligible dataset by extracting the counts of true positives, false positives, true negatives, and false negatives from the numerators and denominators reported in the original studies. Patient- and eye-level analyses were separated because the correlation between fellow eyes if treated as independent, can underestimate variance, inflate precision, and artificially heighten heterogeneity in pooled accuracy estimates. Diagnostic accuracy was synthesized using a hierarchical bivariate random-effects model, the recommended framework for meta-analyses of diagnostic test accuracy. Specifically, we used a bivariate hierarchical logistic-normal model that preserves the intrinsic correlation between sensitivity and specificity while allowing both within- and between-study variability.

Level1(Within−Study):TPj~Bin(Sej,TPj+FNj)andTNj~Bin(Spj,TNj+FPj)
Level2(Between−Study):logit(Sej)logit(Spj)~Nμ1μ2,Σbetween,andΣbetween=τSe2ρτSeτSpρτSeτSpτSp2

Fixed Effects parameters (βSe, βSp) represent the pooled logit-sensitivity and logit-specificity, estimated in Stata 18 (metadta) by maximum-likelihood with adaptive Gauss–Hermite quadrature (nine nodes); Laplace approximation was used as a sensitivity check. This model produced pooled sensitivity, and specificity estimates with corresponding 95% confidence intervals (CIs) and supported the construction of hierarchical summary receiver operating characteristic (HSROC) curves. Each HSROC plot displays the summary operating point surrounded by a 95% confidence ellipse and an outer 95% prediction ellipse, representing the range within which a future study’s accuracy is expected to fall. Study-level estimates were visualized using forest plots. Individual-study CIs used exact Clopper–Pearson limits. Pooled 95% CIs were derived analytically (Wald method) from the multivariate-normal sampling distribution of (βSe, βSp) and back-transformed to the probability scale using the delta method. Prediction ellipses were generated in closed form from the random-effects variance–covariance matrix ΣRE. as the locus x−β^T∑RE−1x−β^X2,0.952. Because all variance components have analytic ML estimates, no bootstrapping was required.

To explore sources of heterogeneity, we performed univariate meta-regression analyses using prespecified moderators: study design (prospective vs retrospective), national income classification (high, upper-middle, lower-middle), healthcare setting (primary, secondary, tertiary), DR category (any diabetic retinopathy (aDR), defined as the presence of any detectable DR lesion including mild non-proliferative DR or worse; referable DR (rDR), defined as moderate NPDR or worse, diabetic macular edema (DME) or proliferative DR; and vision-threatening DR (vtDR), defined as severe NPDR, proliferative diabetic retinopathy (PDR), and/or clinically significant macular edema (CSME))(mapping of grading scales to these categories is provided in Supplementary Table 10), pupil dilation protocol (yes, no, mixed), camera form factor (desktop vs smartphone vs portable), diagnostic criteria (ICDR, ETDRS, NHS, other), index-test grader (ophthalmologist, retina specialist, other), reference-image modality (color fundus vs advanced imaging such as wide-field fundus), retinal field strategy (one, two, more than two), vendor involvement (yes vs no), and image gradability (high [≥95%] vs low [<95%]108). Each covariate was simultaneously entered into both the sensitivity and specificity equations, and nested models were compared with likelihood-ratio tests; overall fit was evaluated by differences in Akaike and Bayesian information criteria (lower AIC/BIC = better fit). Between-study heterogeneity was summarized by τSe2, τSp2, their correlation ρ, and the bivariate analog of I2 (output by metadta). Moderators that significantly explained between-study variance (p < 0.05) in univariate models were included in a multivariate meta-regression to assess their combined effect. Publication bias was assessed for each AI system using Deeks’ funnel plot asymmetry test109; robustness was evaluated by stratifying analyses by article type (peer-reviewed vs conference, abstract or supplement). All statistical analyses were conducted in Stata 18.0 using the metadta (bivariate DTA) and midas (HSROC visualization) packages. No resampling (e.g., bootstrapping) was required because closed-form variance expressions were available for all model components and transformations.

Supplementary information

Supplementary Information (16.4MB, docx)

Acknowledgements

There was no funding provided for this study. We utilized ChatGPT for assistance with parts of the English proofing.

Author contributions

Wang T.-W.: Conceptualization, Methodology, Formal analysis, Data Curation, Writing – Original Draft, Visualization, Project administration; Luo W.-T.: Investigation, Data Curation, Writing – Review & Editing; Tu Y.-K.: Methodology, Validation, Formal analysis, Supervision, Writing – Review & Editing; Chou Y.-B.: Supervision, Writing – Review & Editing; Wu Y.-T.: Supervision, Resources, Writing – Review & Editing.

Data availability

The corresponding author had full access to all data in the study and final responsibility for the decision to submit the report for publication. The data used and analyzed during the current study are available from the corresponding author upon reasonable request.

Code availability

Study protocol and metadata are available from the corresponding author.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary information

The online version contains supplementary material available at 10.1038/s41746-025-02223-8.

References

  • 1.Yau, J. W. et al. Global prevalence and major risk factors of diabetic retinopathy. Diab. Care35, 556–64 (2012). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Teo, Z. L. et al. Global prevalence of diabetic retinopathy and projection of burden through 2045: systematic review and meta-analysis. Ophthalmology128, 1580–1591 (2021). [DOI] [PubMed] [Google Scholar]
  • 3.International Diabetes Federation. IDF Diabetes Atlas, 10th ed.; 2021.
  • 4.Flaxel, C. J. et al. Diabetic retinopathy preferred practice pattern®. Ophthalmology127, P66–P145 (2020). [DOI] [PubMed] [Google Scholar]
  • 5.Cheung, N. et al. Diabetic retinopathy. Lancet376, 124–36 (2010). [DOI] [PubMed] [Google Scholar]
  • 6.Solomon, S. D. et al. Diabetic retinopathy: a position statement by the American Diabetes Association. Diab. Care40, 412–418, 10.2337/dc16-2641 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Rahmati, M. et al. Factors affecting global adherence for the uptake of diabetic retinopathy screening: a systematic review and meta-analysis. Am. J. Ophthalmol.268, 94–107 (2024). [DOI] [PubMed] [Google Scholar]
  • 8.Teo, Z. L. et al. Do we have enough ophthalmologists to manage vision-threatening diabetic retinopathy? A global perspective. Eye34, 1255–1261 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Liew, G., Michaelides, M. & Bunce, C. A comparison of the causes of blindness certifications in England and Wales in working age adults (16–64 years), 1999–2000 with 2009–2010. BMJ Open4, e004015 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Gulshan, V. et al. Development and validation of a deep learning algorithm for detection of diabetic retinopathy in retinal fundus photographs. JAMA316, 2402–2410 (2016). [DOI] [PubMed] [Google Scholar]
  • 11.Ting, D. S. W. et al. Development and validation of a deep-learning system for diabetic retinopathy in multi-ethnic populations. JAMA318, 2211–2223 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Bellemo, V. et al. Artificial intelligence using deep learning to screen for referable and vision-threatening diabetic retinopathy in Africa: a clinical validation study. Lancet Digit. health1, e35–e44 (2019). [DOI] [PubMed] [Google Scholar]
  • 13.Alqahtani, A. S. et al. “The efficacy of artificial intelligence in diabetic retinopathy screening: a systematic review and meta-analysis.” Int. J. Retina Vitreous 1148. 2025. [DOI] [PMC free article] [PubMed]
  • 14.Abràmoff, M. D. et al. Pivotal trial of an autonomous AI system for detection of diabetic retinopathy. npj Digit. Med.1, 39 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Ipp, E. et al. Pivotal evaluation of an AI system for autonomous detection of diabetic retinopathy. JAMA Netw. Open4, e2134254 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Bhaskaranand, M. et al. Automated DR screening with the EyeArt system: >100 000 encounters. Diab. Technol. Ther.21, 635–643 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Lim, J. I. et al. Artificial intelligence detection of diabetic retinopathy: subgroup comparison of the EyeArt System with ophthalmologists’ dilated examinations. Ophthalmol. Sci.3, 100228 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Kaushal, A., Altman, R. & Langlotz, C. Geographic and demographic bias in AI diagnostic performance. Nat. Med. 28, 1328–1334 (2022). [Google Scholar]
  • 19.Khinvasara, T., Tzenios, N. & Shankar, A. Post-market surveillance of medical devices using AI. J. Complement. Altern. Med. Res.25, 108–22 (2024). [Google Scholar]
  • 20.Khan, Z. et al. Diagnostic accuracy of IDx-DR: a systematic review and meta-analysis. Am. J. Ophthalmol.273, 192–204 (2025). [DOI] [PubMed] [Google Scholar]
  • 21.Brant, A. et al. Performance of a deep learning diabetic retinopathy algorithm in India. JAMA Netw. Open8, e250984 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Chotcomwongse, P. et al. Transforming non-digital, clinical workflows to detect and track vision-threatening diabetic retinopathy via a digital platform integrating artificial intelligence: implementation research. Ophthalmol. Ther.14, 447–460 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Tan-Torres, A. et al. Validation oF A Deep Learning Model For Diabetic Retinopathy On Patients With Young-onset Diabetes. Ophthalmol. Ther.14, 1147–1155 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Wei, Q. et al. Practical applications of artificial intelligence diagnostic systems in fundus retinal disease screening. Int. J. Gen. Med.18, 1173–1180 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Wroblewski, J. J. et al. Diabetic retinopathy screening using smartphone-based fundus photography and deep-learning artificial intelligence in the Yucatan Peninsula: a field study. J. Diab. Sci. Technol.19, 370–376 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Doğan, M. E. et al. Head-to-head comparison of diagnostic performance of three non-mydriatic cameras for diabetic retinopathy screening with artificial intelligence. Eye38, 1694–1701 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Grzybowski, A. et al. Diagnostic accuracy of automated diabetic retinopathy image assessment software: IDx-DR and RetCAD. Ophthalmol. Ther.14, 73–84 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Grzybowski, A. et al. Evaluating the efficacy of AI systems in diabetic retinopathy detection: a comparative analysis of Mona DR and IDx-DR. Acta Ophthalmol.103, 388–395 (2025). [DOI] [PubMed] [Google Scholar]
  • 29.Ibañez-Bruron, M. C. et al. Uso de inteligencia artificial en el tamizaje de retinopatía diabética: experiencia en un servicio de salud de Santiago, Chile” [Use of artificial intelligence in diabetic retinopathy screening: experience in a health service in Santiago, Chile]. Rev. Med. Chile152, 1148–1153 (2024). [DOI] [PubMed] [Google Scholar]
  • 30.Karabeg, M. et al. A pilot cost-analysis study comparing AI-based EyeArt® and ophthalmologist assessment of diabetic retinopathy in minority women in Oslo, Norway. Int. J. Retin. Vitreous10, 40 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Li, Z. et al. Application of artificial intelligence remote screening system for diabetic retinopathy in Yinchuan community of Ningxia. Int. Eye Sci.12, 1365–1368 (2022). (2022). [Google Scholar]
  • 32.Musetti, D. et al. Autonomous artificial intelligence versus teleophthalmology for diabetic retinopathy. Eur. J. Ophthalmol.35, 232–238 (2025). [DOI] [PubMed] [Google Scholar]
  • 33.Espeut, J., Molina, F. & Van, T. N. N. Effectiveness of artificial intelligence for diabetic retinopathy screening in community in Binh Dinh Province, Vietnam. Taiwan J. Ophthalmol.14, 394–402 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Piatti, A. et al. Feasibility and accuracy of the screening for diabetic retinopathy using a fundus camera and an artificial intelligence pre-evaluation application. Acta Diabetol.61, 63–68 (2024). [DOI] [PubMed] [Google Scholar]
  • 35.Piatti, A. et al. Diabetic retinopathy screening with confocal fundus camera and artificial intelligence - assisted grading. Eur. J. Ophthalmol.35, 679–688 (2025). [DOI] [PubMed] [Google Scholar]
  • 36.Riotto, E. et al. Accuracy of autonomous artificial intelligence-based diabetic retinopathy screening in real-life clinical practice. J. Clin. Med.13, 4776 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Šín, M. et al. Comparison of the Aireen system with telemedicine evaluation by an ophthalmologist - a real-world study. Clin. Ophthalmol.19, 957–964 (2025). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Taylor, J. R. et al. Real-world evaluation of RetCAD deep-learning system for the detection of referable diabetic retinopathy and age-related macular degeneration. Clin. Exp. Optomet. 1-6 (2024). [DOI] [PubMed]
  • 39.Yao, L. et al. Screening and evaluation of diabetic retinopathy via a deep learning network model: a prospective study. World J. Diab.15, 2302–2310 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Chia, M. A. et al. Validation of a deep learning system for the detection of diabetic retinopathy in Indigenous Australians. Br. J. Ophthalmol.108, 268–273 (2024). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Dow, E. R. et al. AI-human hybrid workflow enhances teleophthalmology for the detection of diabetic retinopathy. Ophthalmol. Sci.3, 100330 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Grzybowski, A. et al. Diagnostic accuracy of automated diabetic retinopathy image assessment softwares: IDx-DR and medios artificial intelligence. Ophthalmic Res.66, 1286–1292 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.He, S. et al. Cross-camera performance of deep learning algorithms to diagnose common ophthalmic diseases: a comparative study highlighting feasibility to portable fundus camera use. Curr. Eye Res.48, 857–863 (2023). [DOI] [PubMed] [Google Scholar]
  • 44.Kemp, O. et al. Real-world evaluation of smartphone-based artificial intelligence to screen for diabetic retinopathy in Dominica: a clinical validation study. BMJ Open Ophthalmol.8, e001491 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Lupidi, M. et al. Artificial intelligence in diabetic retinopathy screening: clinical assessment using handheld fundus camera in a real-life setting. Acta Diabetolog.60, 1083–1088 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Meredith, S. et al. “Performance of an artificial intelligence automated system for diabetic eye screening in a large English population.” Diabetic Med. 40, e15055 (2023). [DOI] [PubMed]
  • 47.Fleming, A. D. et al. Deep learning detection of diabetic retinopathy in Scotland’s diabetic eye screening programme. Br. J. Ophthalmol.108, 984–988 (2024). [DOI] [PubMed] [Google Scholar]
  • 48.Salongcay, R. P. et al. Accuracy of integrated artificial intelligence grading using handheld retinal imaging in a Community Diabetic Eye Screening Program. Ophthalmol. Sci.4, 100457 (2023). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Vought, R. et al. EyeArt artificial intelligence analysis of diabetic retinopathy in retinal screening events. Int. Ophthalmol.43, 4851–4859 (2023). [DOI] [PubMed] [Google Scholar]
  • 50.Arenas-Cavalli, J. T. et al. Clinical validation of an artificial intelligence-based diabetic retinopathy screening tool for a national health system. Eye36, 78–85 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Dong, X. et al. Evaluation of an artificial intelligence system for the detection of diabetic retinopathy in chinese community healthcare centers. Front. Med.9, 883462 (2022). [DOI] [PMC free article] [PubMed]
  • 52.Han, R. et al. Validating automated eye disease screening AI algorithm in community and in-hospital scenarios. Front. Public Health10, 944967 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Hao, S. et al. Clinical evaluation of AI-assisted screening for diabetic retinopathy in rural areas of midwest China. PloS One17, e0275983 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Li, F. et al. A multicenter clinical study of the automated fundus screening algorithm. Transl. Vis. Sci. Technol.11, 22 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Mehra, A. A. et al. Diabetic retinopathy telemedicine outcomes with artificial intelligence-based image analysis, reflex dilation, and image overread. Am. J. Ophthalmol.244, 125–132 (2022). [DOI] [PubMed] [Google Scholar]
  • 56.Mokhashi, N. et al. A comparison of artificial intelligence and human diabetic retinal image interpretation in an urban health system. J. Diab. Sci. Technol.16, 1003–1007 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Paul, S. et al. Einsatz von künstlicher Intelligenz im Screening auf diabetische Retinopathie an einer diabetologischen Schwerpunktklinik” [Use of artificial intelligence in screening for diabetic retinopathy at a tertiary diabetes center]. Die Ophthalmol.119, 705–713 (2022). [DOI] [PubMed] [Google Scholar]
  • 58.Pei, X. et al. Efficacy of artificial intelligence-based screening for diabetic retinopathy in type 2 diabetes mellitus patients. Diab. Res.Clin. Pract.184, 109190 (2022). [DOI] [PubMed] [Google Scholar]
  • 59.Rao, D. P. et al. Towards a device agnostic AI for diabetic retinopathy screening: an external validation study. Clin. Ophthalmol.16, 2659–2667 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Nunez do Rio, J. M. et al. Evaluating a deep learning diabetic retinopathy grading system developed on mydriatic retinal images when applied to non-mydriatic community screening. J. Clin. Med.11, 614 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Ruamviboonsuk, P. et al. Real-time diabetic retinopathy screening by deep learning in a multisite national screening programme: a prospective interventional cohort study. Lancet Digit. Health4, e235–e244 (2022). [DOI] [PubMed] [Google Scholar]
  • 62.Skevas, C. et al. Simultaneous screening and classification of diabetic retinopathy and age-related macular degeneration based on fundus photos—a prospective analysis of the RetCAD system. Int. J. Ophthalmol.15, 1985–1993 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Sedova, A. et al. Comparison of early diabetic retinopathy staging in asymptomatic patients between autonomous AI-based screening and human-graded ultra-widefield colour fundus images. Eye36, 510–516 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Tsai, M.-J. et al. Cross-camera external validation for artificial intelligence software in diagnosis of diabetic retinopathy. J. Diab. Res.2022, 5779276 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Yang, Y. et al. Performance of the AIDR Screening system in detecting diabetic retinopathy in the fundus photographs of Chinese patients: a prospective, multicenter, clinical study. Ann. Transl. Med.10, 1088 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Bhuiyan et al. Combined automated screening for age-related macular degeneration and diabetic retinopathy in primary care settings. Ann. Eye Sci.6, 12 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Grzybowski, A. & Brona, P. Analysis and comparison of two artificial intelligence diabetic retinopathy screening algorithms in a pilot study: IDx-DR and retinalyze. J. Clin. Med.10, 2352 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Heydon, P. et al. Prospective evaluation of an artificial intelligence-enabled algorithm for automated diabetic retinopathy screening of 30 000 patients. Br. J. Ophthalmol.105, 723–728 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 69.Jain, A. et al. Use of offline artificial intelligence in a smartphone-based fundus camera for community screening of diabetic retinopathy. Indian J. Ophthalmol.69, 3150–3154 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Kim, T. N. et al. Comparison of automated and expert human grading of diabetic retinopathy using smartphone-based retinal photography. Eye35, 334–342 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 71.Li, N. et al. A stratified analysis of a deep learning algorithm in the diagnosis of diabetic retinopathy in a real-world study. J. Diab.14, 111–120 (2022). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 72.Li, Y.-H. et al. The clinical influence after implementation of convolutional neural network-based software for diabetic retinopathy detection in the primary care setting. Life11, 200 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Lin, D. et al. Application of Comprehensive Artificial intelligence Retinal Expert (CARE) system: a national real-world evidence study. Lancet Digit. health3, e486–e495 (2021). [DOI] [PubMed] [Google Scholar]
  • 74.Ming, S. et al. Evaluation of a novel artificial intelligence-based screening system for diabetic retinopathy in community of China: a real-world study. Int. Ophthalmol.41, 1291–1299 (2021). [DOI] [PubMed] [Google Scholar]
  • 75.Shah, A. et al. Validation of automated screening for referable diabetic retinopathy with an autonomous diagnostic artificial intelligence system in a Spanish population. J. Diab. Sci. Technol.15, 655–663 (2021). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Wintergerst, M. W. M. et al. Telemedical Diabetic Retinopathy Screening in a Primary Care Setting: Quality of Retinal Photographs and Accuracy of Automated Image Analysis. Ophthalmic Epidemiol.29, 286–295 (2022). [DOI] [PubMed] [Google Scholar]
  • 77.González-Gonzalo, C. et al. Evaluation of a deep learning system for the joint automated detection of diabetic retinopathy and age-related macular degeneration. Acta Ophthalmol.98, 368–377 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 78.He, J. et al. Artificial intelligence-based screening for diabetic retinopathy at community hospital. Eye34, 572–576 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 79.Hsieh, Y.-T. et al. Application of deep learning image assessment software VeriSee™ for diabetic retinopathy screening. J. Formos. Med. Assoc. = Taiwan yi zhi120, 165–171 (2021). [DOI] [PubMed] [Google Scholar]
  • 80.Li, Z. X. et al. Zhonghua Yi Xue Za Zhi100, 3835–3840 (2020). [DOI] [PubMed] [Google Scholar]
  • 81.Olvera-Barrios, A. et al. Diagnostic accuracy of diabetic retinopathy grading by an artificial intelligence-enabled algorithm compared with a human standard for wide-field true-colour confocal scanning and standard digital retinal images. Br. J. Ophthalmol.105, 265–270 (2021). [DOI] [PubMed] [Google Scholar]
  • 82.Sarao, V. et al. Automated diabetic retinopathy detection with two different retinal imaging devices using artificial intelligence: a comparison study. Graefe’s. Arch. Clin. Exp. Ophthalmol. = Albrecht Graefes Arch. Klinische Exp. Ophthalmol.258, 2647–2654 (2020). [DOI] [PubMed] [Google Scholar]
  • 83.Sosale, B. et al. Medios- An offline, smartphone-based artificial intelligence algorithm for the diagnosis of diabetic retinopathy. Indian J. Ophthalmol.68, 391–395 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 84.Sosale, B. et al. Simple, Mobile-based Artificial Intelligence Algorithm in the detection of Diabetic Retinopathy (SMART) study. BMJ Open Diab. Res. Care8, e000892 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 85.Zhang, Y. et al. Artificial intelligence-enabled screening for diabetic retinopathy: a real-world, multicenter and prospective study. BMJ Open Diab. Res. Care8, e001596 (2020). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 86.Gulshan, V. et al. Performance of a deep-learning algorithm vs manual grading for detecting diabetic retinopathy in India. JAMA Ophthalmol.137, 987–993 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 87.Natarajan, S. et al. Diagnostic accuracy of community-based diabetic retinopathy screening with an offline artificial intelligence system on a smartphone. JAMA Ophthalmol.137, 1182–1188 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 88.Quellec, G. et al. Instant Automatic Diagnosis of Diabetic Retinopathy (Cornell University, 2019).
  • 89.Raumviboonsuk, P. et al. Deep learning versus human graders for classifying diabetic retinopathy severity in a nationwide screening program. NPJ Digit. Med.2, 25 (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Verbraak, F. D. et al. Diagnostic accuracy of a device for the automated detection of diabetic retinopathy in a primary care setting. Diab. Care42, 651–656 (2019). [DOI] [PubMed] [Google Scholar]
  • 91.Krause, J. et al. Grader variability and the importance of reference standards for evaluating machine learning models for diabetic retinopathy. Ophthalmology125, 1264–1272 (2018). [DOI] [PubMed] [Google Scholar]
  • 92.Rajalakshmi, R. et al. Automated diabetic retinopathy detection in smartphone-based fundus photography using artificial intelligence. Eye32, 1138–1144 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 93.van der Heijden, A. A. et al. Validation of automated screening for referable diabetic retinopathy with the IDx-DR device in the Hoorn Diabetes Care System. Acta Ophthalmol.96, 63–68 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 94.Tufail, A. et al. Automated diabetic retinopathy image assessment software: diagnostic accuracy and cost-effectiveness compared with human graders. Ophthalmology124, 343–351 (2017). [DOI] [PubMed] [Google Scholar]
  • 95.Abràmoff, M. D. et al. Improved automated detection of diabetic retinopathy on a publicly available dataset through integration of deep learning. Investig. Ophthalmol. Vis. Sci.57, 5200–5206 (2016). [DOI] [PubMed] [Google Scholar]
  • 96.Keel, S. et al. Feasibility and patient acceptability of a novel artificial intelligence-based screening model for diabetic retinopathy at endocrinology outpatient services: a pilot study. Sci. Rep.8, 4330 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 97.Higgins, J. P. T., Thompson, S. G., Deeks, J. J. & Altman, D. G. Measuring inconsistency in meta-analyses. BMJ327, 557–560 (2003). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 98.Rutter, C. M. & Gatsonis, C. A. A hierarchical regression approach to meta-analysis of diagnostic test accuracy evaluations. Stat. Med.20, 2865–2884 (2001). [DOI] [PubMed] [Google Scholar]
  • 99.Macdonald, T. et al. Generating evidence to support the role of AI in diabetic eye screening: considerations from the UK National Screening Committee. Lancet Digit. Health S2589-750000271-1 (2025). [DOI] [PubMed]
  • 100.Li, Z. et al. Deep-learning-based detection of diabetic retinopathy: a systematic review and meta-analysis. PLoS One16, e0250033 (2021).33882095 [Google Scholar]
  • 101.Wolf, R. M. et al. Autonomous artificial intelligence increases screening and follow-up for diabetic retinopathy in youth: the ACCESS randomized control trial. Nat. Commun.15, 421 (2024). [DOI] [PMC free article] [PubMed]
  • 102.Centers for Medicare & Medicaid Services. HCPCS/CPT Code 92229 – Imaging of retina for detection or monitoring of disease; point-of-care automated analysis. Final Rule CY (2021).
  • 103.Wang, T.W., Luo, W., Tu, Y. Chou, Y., & Wu, Y. (2025). Diagnostic accuracy of regulatory-approved deep-learning systems for fundus-based detection of diabetic retinopathy—a systematic review and meta-analysis. 10.37766/inplasy2025.5.0099.
  • 104.McInnes, M. D. F. et al. Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies: the PRISMA-DTA statement. JAMA319, 388–396 (2018). [DOI] [PubMed] [Google Scholar]
  • 105.Moher, D., Liberati, A., Tetzlaff, J. & Altman, D. G. & PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ339, b2535 (2009). [PMC free article] [PubMed] [Google Scholar]
  • 106.World Bank. World Bank country classifications by income level for 2024–2025. World Bank Blogs. https://blogs.worldbank.org/opendata/world-bank-country-classifications-by-income-level-for-2024-2025.
  • 107.Whiting, P. F. et al. QUADAS-2 Group. QUADAS-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann. Intern. Med.155, 529–36 (2011). [DOI] [PubMed] [Google Scholar]
  • 108.International Council of Ophthalmology (ICO). ICO Guidelines for Diabetic Eye Care. Updated 2017.
  • 109.Deeks, J. J., Macaskill, P. & Irwig, L. The performance of tests of publication bias and other sample size effects in systematic reviews of diagnostic test accuracy was assessed. J. Clin. Epidemiol.58, 882–893 (2005). [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Information (16.4MB, docx)

Data Availability Statement

The corresponding author had full access to all data in the study and final responsibility for the decision to submit the report for publication. The data used and analyzed during the current study are available from the corresponding author upon reasonable request.

Study protocol and metadata are available from the corresponding author.