Abstract
Objectives: To evaluate the precision of both total %fat and all the regional %fat measures acquired from both the Lunar Prodigy and Lunar iDXA software. Design: Cross-sectional study. Setting: University-based research laboratory. Participants: The sample consisted of 300 individual test records from men and women who had volunteered to participate in dual-energy x-ray absorptiometer (DXA) technician precision training studies. Subjects ranged in age from 20–84 years and in body mass index from 15.7–52.0 kg.m−2. Measurements: A total of 27 different technicians performed three total body scan measurements on 10–15 different subjects. The Lunar Prodigy and Lunar iDXA were used for 253 and 47 precision training evaluations, respectively. The regions of interest (ROI) were automatically determined by the enCORE software (autoROI) for total body, android, gynoid, trunk, legs, and arms regions and the region %fat data were used for analyses. Results: The CV for total body %fat was 1.9% and 0.9% for the Prodigy and iDXA, respectively. CV's for %fat measures at regional sites ranged from 1.2–4.4% for the Prodigy measures and 0.9–2.4% for the iDXA measures. The ICC for both devices ranged from 0.990 to 0.999. Conclusion: Monitoring the status of body composition changes with age is gaining more clinical acceptance. Thus, it is important that practitioners use measures that are both precise and accurate. The findings from the current study add support that DXA measurements can be used with a high level of confidence for serial testing of patients.
Key words: Body composition, dual-energy x-ray absorptiometer, precision, body fat percentage
Introduction
Global trends show that as a whole the population of developed countries is aging (1). Aging is associated with a number of physiological body composition changes including increased amounts of fat tissue and decreased amounts of lean tissue and bone mineral density (BMD) 2, 3. These changes in body composition are associated with a number of unfavorable health consequences.
Measurement of body composition, particularly estimates of body fat percentage (%fat), have been utilized in both research and athletic settings for many years. The clinical importance of assessing body composition is gaining more appreciation, in part due to the increasing age of the population. Measurement of body composition can be obtained from a number of different methods including underwater weighing, anthropometrics, electrical impedance, plethysmography, ultrasound, and dual-energy x-ray absorptiometer (DXA) (4). Two recent studies evaluated the use of ultrasound and electrical impedance to assess fat-free mass 5, 6. In clinical settings, DXA has a number of advantages due to its capability to provide soft tissue measurements of both fat and lean components and to measure bone density, thus providing valuable information on adiposity, sarcopenia, and osteoporosis. Additionally, DXA provides a more comprehensive subset of regional fat measurements, which may prove to be quite useful in monitoring individuals’ body composition as they age.
Precision is one of the most important, yet surprisingly seldom studied characteristics of measurement. Indeed, a measurement cannot be accurate unless it can be obtained with a high degree of precision (7). Additionally, the stronger the reproducibility of a measurement, the more likely it is to detect a small, yet meaningful change, which is especially important in clinical environments when monitoring individuals as they age and also in response to prescribed therapies.
DXA methods were originally developed to assess bone density and there have been a number of reports about the precision of BMD assessments (8). Software developments led to expanding the DXA measurement capabilities to assess soft tissue. Unfortunately, with the exception of a few abstracts from professional conferences, there is little information on the precision of DXA measures of body fat percentage (9, 10, 11). Rothney et al. (12) (Lunar iDXA) and Santos et al. (13) (Hologic Explorer-W) evaluated precision of total body scans, however, only reported coefficient of variation (CV) data on fat (±1.0 and 3.7%, respectively) and lean (±0.5 and 1.1%, respectively) mass measures. Recently, Hind et al. (14), (Lunar iDXA) and Carver et al. (15) reported of CVs for total body %fat (0.9 and 0.8%) along with android (2.3 and 1.6%) and gynoid (1.0 and 1.4%) regional %fat values, respectively. Therefore, the purpose of this study was to evaluate the precision of both total %fat and all the regional %fat measures provided from both the Lunar Prodigy and iDXA software.
Methods
Subjects
Approval for the project was obtained from Ball State University's Institutional Review Board. Data were obtained from the Ball State University Clinical Exercise Physiology program database in the form of a de-identified dataset. The request included data on individuals > 20 years of age who had volunteered to participate in DXA technician precision training studies between 2007–2010. The list of variables provided in the dataset were: age, gender, height, weight, and DXA measurements of %fat for the total body, android, trunk, gynoid, leg, and arm regions. A total of 300 individual's test records were obtained from subjects that completed 3 DXA total body scans on the same day. The characteristics of the subject population are provided in Table 1.
Table 1.
Descriptive Characteristics of Subjects (mean±SD)
| Age (y) | Height (in) | Weight (lb) | BMI (kg.m-2) | |
|---|---|---|---|---|
| Prodigy (n=253) | 39.1±18.5* | 66.8±3.7 | 162.2±33.4 | 25.5±4.7 |
| Range | 20-84 | 52.0-77.0 | 96.2-268.0 | 15.7-52.0 |
| iDXA (n=47) | 31.0±14.3 | 67.5±3.9 | 163.9±36.4 | 25.3±5.4 |
| Range |
20-71 |
60.0-77.0 |
100.0-266.4 |
19.2-45.0 |
p<0.01
DXA Measurement Procedures
Measurements were made with the GE Lunar Prodigy (n=253) between 2007-2009 and with the GE iDXA (n=47) in 2010. Prior to all measurement sessions the DXA was evaluated with the GE Lunar block calibration phantom to ensure that the unit was operating within the manufacturer's specifications. A total of 27 different technicians, each of whom had completed a DXA operator's training program, performed three total body scan measurements on 10-15 different subjects. All subjects were measured without shoes, jewelry and had removed anything containing metal from their clothing. Subjects were positioned appropriately in the supine position for the total body scan and after each scan they stood up and were then re-positioned on the DXA platform prior to the second and third measurements. Subject's legs were held together by a cloth strap around the ankles to prevent movement during the scans.
Prodigy scans were analyzed with the enCORE software version 11.4 and iDXA scans with version 13.4 (GE Healthcare, Madison, WI). The ROIs were automatically determined by the enCORE software (autoROI) for total body, android, gynoid, trunk, legs, and arms regions. Region % fat, which is determined by fat mass divided by the sum of fat, lean, and bone masses, was used for all analyses.
Statistical analysis
SAS version 9.2 (SAS Institute, Cary, NC) was used to perform the statistical analyses with p-value of <0.05, as the criterion for statistical significance. Independent t-tests were used to compare differences in subject characteristics between the two samples measured by different devices (Prodigy vs. iDXA).
Intraclass correlations (ICC) using variance from one-way repeated measures ANOVA were calculated to compare the three DXA scans performed within each device for each regional measure.
Precision of DXA at different regions were calculated using the root mean square SD (RMS-SD), %CV, and least significant change (LSC) using the calculation tool of the International Society for Clinical Densitometry guideline.16
Table 2B.
Region %fat of 3 repeated measures with Lunar iDXA (n=47; Mean ± SD)
| First Scan | Second Scan | Third Scan | ICC | |
|---|---|---|---|---|
| Total Body | 27.7 ± 10.7 | 27.8 ± 10.6 | 27.8 ± 10.6 | 0.999 |
| Android | 28.0 ± 14.3 | 27.9 ± 14.0 | 28.1 ± 14.0 | 0.999 |
| Gynoid | 31.9 ± 11.6 | 31.9 ± 11.5 | 31.9 ± 11.5 | 0.998 |
| Trunk | 27.7 ± 12.4 | 27.7 ± 12.3 | 27.8 ± 1232 | 0.999 |
| Legs | 29.2 ± 10.0 | 29.2 ± 10.0 | 29.2 ± 10.0 | 0.999 |
| Arms |
26.1 ± 11.2 |
26.0 ± 11.1 |
26.1 ± 11.0 |
0.990 |
Results
Data for the precision studies of both the Prodigy and iDXA were collected from different subject populations. The subject pool for the Prodigy precision studies were slightly older, however there were no other mean differences. Both subject pools had a wide range of values of descriptive characteristics, especially in terms of age and body mass index (BMI), which is similar to those seen by clinicians in adult practice patient populations.
The mean %fat values for the three separate total body scans per subject on both the Prodigy and iDXA are presented in Table 2. The ICC of the measures for total body fat and for the regional indices within the Prodigy and within the iDXA ranged from 0.990 to 0.999.
Table 2A.
Region %fat of 3 repeated measures with Lunar Prodigy (n=253; Mean ± SD)
| First Scan | Second Scan | Third Scan | ICC | |
|---|---|---|---|---|
| Total Body | 30.4 ± 10.9 | 30.4 ± 10.9 | 30.4 ± 10.8 | 0.994 |
| Android | 35.7 ± 12.1 | 35.7 ± 12.1 | 35.8 ± 12.1 | 0.992 |
| Gynoid | 37.0 ± 11.4 | 37.1 ± 11.4 | 37.1 ± 11.4 | 0.995 |
| Trunk | 32.4 ± 11.2 | 32.4 ± 11.1 | 32.4 ± 10.9 | 0.990 |
| Legs | 31.7 ± 12.2 | 31.7 ± 12.1 | 31.7 ± 12.2 | 0.997 |
| Arms |
25.9 ± 11.7 |
26.0 ± 11.8 |
26.3 ± 12.1 |
0.991 |
Precision data for all regional measures of %fat are provided in Table 3. The CV for total body %fat was 1.9% and 0.9% for the Prodigy and iDXA, respectively. CV's for %fat measures at regional sites ranged from 1.2-4.4% for the Prodigy measures and 0.9-2.4% for the iDXA measures.
Table 3A.
Total Body and Regional Fat (%) Precision Acquired by Lunar Prodigy (n=253)
| Region Variables | Mean | (range) | RMS-SD | CV (%) | LSC |
|---|---|---|---|---|---|
| Total body | 30.4 | (6.6 - 56.2) | 0.60 | 2.0 | 1.7 |
| Android Fat | 35.7 | (6.6 - 62.9) | 0.58 | 1.6 | 1.6 |
| Gynoid Fat | 37.1 | (9.5- 61.7) | 0.43 | 1.2 | 1.2 |
| Trunk Fat | 32.4 | (7.0 - 58.3) | 1.00 | 3.1 | 2.8 |
| Leg Fat | 31.7 | (6.6 -67.0) | 0.56 | 1.8 | 1.6 |
| Arm Fat |
26.1 |
(4.8 - 65.7) |
1.13 |
4.4 |
3.1 |
Discussion
These data support that total body %fat obtained with both the Lunar Prodigy and iDXA can be made with a high level of precision (CV <2%). Our results of total body CV from the Prodigy of 1.9% are similar to those reported by Wacker et al. (10) and Chen et al. (11) (1.7 and 1.2%, respectively). Table 4 summarizes the CV reported in the current and other studies, which demonstrates the excellent, and even higher precision (<1%), of total body %fat measures from the iDXA. The findings from the current study included a larger subject pool and wider range of adult ages, BMIs, which extends the application to a more heterogeneous patient population. The current study also made three repeated total body scans on each subject, compared to only two measures in all of the other studies cited, and also involved a total of 27 different technicians performing the measurements. This later procedural characteristic of the current study, of multiple technicians over many years, makes the case for excellent precision of DXA total body %fat values even stronger.
Table 4.
Comparison of CV of %fat values reported from the Lunar iDXA
The current study affirmed that the most clinically valued regional body composition measures (android and gynoid) also had excellent precision on both the Prodigy and IDXA (CV <2%). It is well accepted that excess body fat accumulated in the android region is associated with increased risks for cardiovascular and metabolic diseases (17). Thus, having the ability to obtain highly precise readings of android and gynoid %fat will provide useful information for tracking subjects as they age.
The current study also provides the first reports of precision of other regional % fat indices. Although leg %fat CV were <2% for both the Prodigy and IDXA, values for the trunk on the Prodigy (3.1%) and arms on both the Prodigy (4.1%) and iDXA (2.4%) were higher. Although there are no accepted clinical uses of these additional regional %fat measures, the ability to track body composition changes in these different regions would allow researchers to investigate whether these changes relate to any useful clinical or functional measures. As this is the first study to report the precision %fat measures of the trunk, legs and arms, other studies are needed prior to making conclusions about the possible lower precision for arm measures. However, our results suggest the iDXA provide higher precision for both trunk and arm %fat than those obtained previously from the Prodigy.
Table 3B.
Total Body and Regional Fat (%) Precision Acquired by Lunar iDXA (n=47)
| Region Variables | Mean | (range) | RMS-SD | CV (%) | LSC |
|---|---|---|---|---|---|
| Total body | 27.8 | (8.6 - 54.2) | 0.26 | 0.9 | 0.7 |
| Android Fat | 28.0 | (4.8 - 62.6) | 0.54 | 1.5 | 1.5 |
| Gynoid Fat | 31.9 | (9.1-56.8) | 0.51 | 1.4 | 1.4 |
| Trunk Fat | 27.7 | (6.0 - 60.0) | 0.43 | 1.3 | 1.2 |
| Leg Fat | 29.2 | (11.4 - 53.6) | 0.37 | 1.2 | 1.0 |
| Arm Fat |
26.1 |
(5.7 - 56.9) |
0.62 |
2.4 |
1.7 |
As the importance of assessing body composition continues to gain more clinical acceptance, practitioners will require measures that are both precise and accurate. The findings from the current study add support that DXA measurements can be used with a high level of confidence for serial testing of patients. Indeed, the LSC data from the iDXA suggests that meaningful responses to interventions for total body fat can be captured with changes as small as 0.7%, and changes in all other regional measures of 1.7% or less. The excellent precision of the DXA measures of %fat will allow determination of small changes as a person ages or in response to a therapeutic or lifestyle change.
Disclosures: The authors report no conflicts of interest
References
- 1.United Nations 2001 World Population Ageing. 1950 [Google Scholar]
- 2.Hughes VA, Frontera WR, Roubenoff R, Evans WJ, Singh MA. Longitudinal changes in body composition in older men and women: role of body weight change and physical activity. Am J Clin Nutr. 2002;76:473–481. doi: 10.1093/ajcn/76.2.473. PubMed PMID: 12145025. [DOI] [PubMed] [Google Scholar]
- 3.Guo SS, Zeller C, Chumlea WC, Siervogel RM. Aging, body composition, and lifestyle: the Fels Longitudinal Study. Am J Clin Nutr. 1999;70:405–411. doi: 10.1093/ajcn/70.3.405. PubMed PMID: 10479203. [DOI] [PubMed] [Google Scholar]
- 4.Heymsfield SB, Lohman TG, Wang Z, Going SB. Human Body Composition. Human Kinetics; Champaign, IL: 2005. 2nd edn. [Google Scholar]
- 5.Ramsey R, Isenring E, Daniels L. Comparing measures of fat-free mass in overweight older adults using three different bioelectrical impedance devices and three prediction equations. J Nutr Health Aging. 2012;16:26–30. doi: 10.1007/s12603-011-0085-6. 10.1007/s12603-011-0085-6 PubMed PMID: 22237998. [DOI] [PubMed] [Google Scholar]
- 6.Abe T, Bemben MG, Kondo M, Kawakami Y, Fukunaga T. Comparison of skeletal muscle mass to fat-free mass ratios among different ethnic groups. J Nutr Health Aging. 2012;16:534–538. doi: 10.1007/s12603-012-0015-2. 10.1007/s12603-012-0015-2 PubMed PMID: 22659992. [DOI] [PubMed] [Google Scholar]
- 7.Grubbs FE. Errors of measurement, precision, accuracy and the statistical comparison of measuring instruments. Technometrics. 1973;15:53–66. 10.1080/00401706.1973.10489010 [Google Scholar]
- 8.Forsen L, Berntsen GK, Meyer HE, Tell GS, Fonnebo V. Differences in precision in bone mineral density measured by SXA and DXA: the NOREPOS study. Eur J Epidemiol. 2008;23:615–624. doi: 10.1007/s10654-008-9271-1. 10.1007/s10654-008-9271-1 PubMed PMID: 18594994. [DOI] [PubMed] [Google Scholar]
- 9.Huizenga R, Barden HS, Oates MK. Precision of Lunar iDXA total body BMD and composition measurements on obese subjects. J Clin Densitom. 2007;10:S208. 10.1016/j.jocd.2007.03.036 [Google Scholar]
- 10.Wacker WK, Barden HS, Franz JR, Ergun DL. Precision comparison between the Lunar iDXA and Lunar Prodigy bone densitometers. J Bone Miner Res. 2006;21:S355. [Google Scholar]
- 11.Chen R, Yang Q, Chen F, Chen Y, Lin S. Precision of total body BMD and body composition using the GE Lunar Prodigy densitometer. J Bone Miner Res. 2011;20:SU132. [Google Scholar]
- 12.Rothney MP, Martin FP, Xia Y, Beaumont M, Davis C, Ergun D, Fay L, Ginty F, Kochhar S, Wacker W, Rezzi S. Precision of GE Lunar iDXA for the measurement of total and regional body composition in nonobese adults. J Clin Densitom. 2012;15:399–404. doi: 10.1016/j.jocd.2012.02.009. 10.1016/j.jocd.2012.02.009 PubMed PMID: 22542222. [DOI] [PubMed] [Google Scholar]
- 13.Santos DA, Gobbo LA, Matias CN, Petroski EL, Gonçalves EM, Cyrino ES, Minderico CS, Sardinha LB, Silva AM. Body composition in taller individuals using DXA: A validation study for athletic and non-athletic populations. J Sports Sci. 2013;31:405–413. doi: 10.1080/02640414.2012.734918. 10.1080/02640414.2012.734918 PubMed PMID: 23092580. [DOI] [PubMed] [Google Scholar]
- 14.Hind K, Oldroyd B, Truscott JG. In vivo precision of the GE Lunar iDXA densitometer for the measurement of total body composition and fat distribution in adults. Eur J Clin Nutr. 2011;65:140–142. doi: 10.1038/ejcn.2010.190. 10.1038/ejcn.2010.190 PubMed PMID: 20842171. [DOI] [PubMed] [Google Scholar]
- 15.Carver TE, Christou NV, Andersen RE. In vivo precision of the GE iDXA for the assessment of total body composition and fat distribution in severely obese patients. Obesity (Silver Spring) 2013;21:1367–1369. doi: 10.1002/oby.20323. 10.1002/oby.20323 [DOI] [PubMed] [Google Scholar]
- 17.Despres JP, Arsenault BJ, Cote M, Cartier A, Lemieux I. Abdominal obesity: the cholesterol of the 21st century? Can J Cardiol. 2008;24(SupplD):7D–12D. doi: 10.1016/s0828-282x(08)71043-2. 10.1016/S0828-282X(08)71043-2 PubMed PMCID 2794449; PMID 18787730. [DOI] [PMC free article] [PubMed] [Google Scholar]
Uncited references
- 16.International Society for Clinical Densitometry Calculators. 2013 [Google Scholar]