Roadmap towards personalized approaches and safety considerations in non-ionizing radiation: from dosimetry to therapeutic and diagnostic applications

Ilkka Laakso 1,*,, Margarethus Marius Paulides 2,3, Sachiko Kodera 4, Seungyoung Ahn 5, Christopher L Brace 6, Marta Cavagnaro 7, Ji Chen 8, Zhi-De Deng 9, Valerio De Santis 10, Yinliang Diao 11, Lourdes Farrugia 12, Mauro Feliziani 10, Serena Fiocchi 13, Francesco Fioranelli 14, Takashi Hikage 15, Sergey Makaroff 16,17, Maya Mizuno 18, Alexander Opitz 19, Emma Pickwell-MacPherson 20, Punit Prakash 21, Dario B Rodrigues 22, Kensuke Sasaki 18, Takuya Sakamoto 23, Zachary Taylor 24, Hubregt J Visser 25, Desmond T B Yeo 26, Akimasa Hirata 4,*,
PMCID: PMC12902664  PMID: 41270373

Abstract

This roadmap provides a comprehensive and forward-looking perspective on the individualized application and safety of non-ionizing radiation (NIR) dosimetry in diagnostic and therapeutic medicine. Covering a wide range of frequencies, i.e. from low-frequency to terahertz, this document provides an overview of the current state of the art and anticipates future research needs in selected key topics of NIR-based medical applications. It also emphasizes the importance of personalized dosimetry, rigorous safety evaluation, and interdisciplinary collaboration to ensure safe and effective integration of NIR technologies in modern therapy and diagnosis.

Keywords: brain stimulation, diagnostic applications, dosimetry, implant safety, personalized medicine, thermal therapy

1. Introduction

1.1. Overview

Akimasa Hirata1, Margarethus Marius Paulides2,3 and Ilkka Laakso4

1 Department of Electrical and Mechanical Engineering, Nagoya Institute of Technology, Nagoya, Japan

2 Care + CureLaboratory of the Electromagnetics Group (EM4Care + Cure), Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands

3 Department of Radiotherapy, Erasmus University Medical Center Cancer Institute, Rotterdam, The Netherlands

4 Department of Electrical Engineering and Automation, Aalto University, Espoo, Finland

The rapid advancement of non-ionizing radiation (NIR) technologies has resulted in transformative innovations in medical applications, especially in diagnostics, monitoring, and therapeutic interventions. NIR covers a wide spectrum of frequencies, ranging from static and low-frequency (LF) electromagnetic fields (EMFs) to radio frequencies, millimetre waves (MMWs), terahertz (THz), and optical radiation. Although optical techniques, such as optical coherence tomography (OCT) and multiphoton microscopy, have been extensively studied and reviewed in recent years (Hong et al 2017, Wang et al 2022a), this roadmap focuses on non-optical NIR technologies in the THz and lower frequency ranges. Given the distinct physical interactions and clinical applications of these modalities, we aim to address the complementary domain of non-optical NIR technologies, which are governed by electrical stimulation and EM heating, making them suitable for non-superficial tissue diagnostics, neuromodulation, thermal therapy, and advanced sensing. In particular, frequencies below 100 kHz predominantly involve neural and tissue stimulation, whereas frequencies above 100 kHz are associated with heating (IEEE-C95.1 2019, ICNIRP 2020). The transition frequency depends on the pulse shape and duration. To limit the scope, this roadmap does not consider the wide class of electroporation-based techniques (Mir 2008, Campelo et al 2023) which use strong pulsed electric fields to alter cell membrane permeability irreversibly for ablation or reversibly for DNA or drug delivery.

Progress in NIR technologies has shown great potential for improving healthcare outcomes. In addition to advances in personalized modelling, ongoing reductions in costs and improvements in device miniaturization and flexibility have facilitated the wider adoption of systems with smaller footprints and lower operational burdens. Nevertheless, individualized approaches are often required to ensure both efficacy and safety because of the complexity of biological interaction and the variability in tissue properties among individuals. Digitized human body phantoms (models) enable personalized modelling in the computational domain, which requires an accurate representation of tissue properties, that is, (i) human body morphology and tissue composition, and (ii) tissue physical constants (dielectric properties and thermal constants).

A foundational development in this area occurred in 1996, when a comprehensive database of the dielectric properties of different tissues was introduced by Gabriel et al (1996). This was later expanded to include measurements of malignant tissues (e.g. Lazebnik et al (2007)), as summarized in recent reviews by Sasaki et al (2022). Significant strides have also been made in human body modelling for safety and medical applications. For instance, Dimbylow’s normalized human voxel model in 1997 (Dimbylow 1997) based on the ICRP database and Nagaoka’s first female voxel model (Nagaoka et al 2003) were key milestones, followed by the development of a complete set of voxel models, that is, a Virtual Family, by the IT’IS Foundation (Gosselin et al 2014). These models have since been extended, deformed, and personalized for various applications, ranging from the evaluation of EM exposure to the planning of patient-specific treatment. The development of pipelines for generating personalized human models has evolved to support specific applications, such as hyperthermia treatment of the head and neck (Paulides et al 2010) and, more recently, brain stimulation (Laakso et al 2015, Saturnino et al 2019b).

In addition to personalized human head models, including data-driven models, recent advances in computational simulations have enabled the use of subject-specific (Homann et al 2011) and population-based human models for detailed dosimetry analysis, and population-based human models for a more sophisticated analysis of NIR–tissue interactions (Diao et al 2025). These models support application-specific design and safety evaluations, while accounting for anatomical variability. These developments have led to a shift from reliance on generic phantom-based evaluations to the integration of real-time physiological monitoring based on personalized modelling. Importantly, the accuracy requirements and modelling targets differ considerably across clinical applications, such as neuromodulation, thermal ablation, and implant safety (Winter et al 2021), which require tailored dosimetric strategies. Nevertheless, quantifying and controlling energy deposition at relevant spatial and temporal scales, as well as capturing individual differences in anatomy, physiology, and pathology, remain challenging. With the advancement of NIR technologies towards clinical translation and widespread adoption, establishing integrated frameworks that combine personalized anatomical modelling, real-time monitoring, and adaptive control strategies, as well as maintaining compliance with safety standards and regulatory guidelines, are important.

In addition to computational modelling and simulation, experimental dosimetry plays a key role in validating and translating models into clinically viable solutions. This includes the use of phantom measurements, magnetic resonance (MR) thermometry, probe-based validation. Moreover, based on fundamental studies on dielectric properties, MMW and THz technologies have been increasingly applied for monitoring physiological parameters in the human body.

This roadmap brings together international experts on dosimetry and associated techniques to summarize and highlight current knowledge and outline future directions across major application areas of NIR in medicine. Each section addresses the status, key challenges, and emerging scientific and technological advances aimed at realizing effective and safe individualized NIR-based medical and healthcare applications. As these technologies approach clinical translation, it is essential to address scientific and technical challenges as well as practical issues such as regulatory approval, interoperability with existing systems, and integration into clinical workflows.

To further illustrate how different applications relate to frequency ranges and dominant interaction mechanisms, figure 1 provides an overview of the technologies discussed in this roadmap. Various diagnostic and therapeutic techniques, such as stimulation, hyperthermia, thermal ablation, sensing, and implant safety, are aligned along the spectrum of NIR. This overview highlights the transition of dominant physical effects, from nerve stimulation at low frequencies to heating at radiofrequency and beyond, and thereby provides a common framework for subsequent sections. The roadmap is organized as follows.

Figure 1.

Figure 1.

Overview of the diagnostic and therapeutic applications discussed in this roadmap positioned along the non-ionizing radiation frequency spectrum. The figure illustrates that different modalities fall in relation to the frequency range and dominant biophysical effects, from nerve stimulation at low frequencies to heating mechanisms at higher frequencies. The effect of brain and nerve stimulation is dominant at <100 kHz, whereas that of heating is dominant at >100 kHz. Wavelength in the skin is also shown as a reference.

Section 2 introduces the foundational elements of dosimetry, focusing on tissue dielectric properties and computational EM methods across low-to-radiofrequency ranges. Variability in tissue characteristics and methodological limitations constrain the modelling accuracy. Recent advances in noninvasive imaging and computational approaches have provided some solutions. Further developments are required to achieve dosimetry that reflects individual anatomy and dynamic physiology for use in design, therapy planning, and safety evaluation.

Section 3 addresses noninvasive neuromodulation techniques in the brain, specifically transcranial electrical stimulation (TES) and transcranial magnetic stimulation (TMS). TES typically employs direct current (0 Hz), LF alternating current (in the range of a few Hz to several 100 Hz), or kilohertz carriers for temporal interference approaches, whereas TMS utilizes pulsed magnetic fields generated by kilohertz-range pulse currents to induce effective LF stimulation. Both techniques increasingly rely on individualized modelling to improve precision, although outcome variability across individuals remains a major challenge. Advances in coil and electrode configurations provide new technical options. Future work is needed to determine how such methods can be consistently applied in clinical practice.

Section 4 examines dosimetric considerations for thermal therapies including hyperthermia and thermal ablation. These modalities typically employ radiofrequency and microwave (MW) ranges, and hyperthermia is commonly performed at tens of megahertz to a few gigahertz, whereas thermal ablation uses either radiofrequency currents around several 100 kHz or MWs in the gigahertz range. Prediction of the thermal dose remains limited by uncertainties in tissue properties, blood perfusion, and patient-specific variability. Imaging-based monitoring and computational models are gradually improving, but real-time reliability is still lacking. Progress in adaptive modelling and monitoring techniques is important to ensure safe and accurate treatment.

Section 5 outlines radiofrequency (RF)-based diagnostic technologies for MR imaging (MRI), MMW sensing, and THz applications. These span a broad frequency spectrum: MRI operates at Larmor frequencies of tens to hundreds of megahertz depending on the field strength (e.g. ∼64 MHz at 1.5 T and ∼128 MHz at 3 T), MMW sensing utilizes bands in the 30–300 GHz range (commonly around 60–80 GHz), and THz applications extend from 0.1 to several THz. Each modality has demonstrated potential for noninvasive assessment while also facing challenges related to safety evaluation, robustness under practical conditions, and individual variability. The early applications of computational and measurement-based approaches have demonstrated their utility. Continuous refinement is necessary to enable wider adoption in clinical diagnostics.

Section 6 focuses on implantable medical devices (IMDs) including wireless power transfer, MRI safety, and EM interference (EMI). Wireless power technology (WPT) typically operates from hundreds of kilohertz to several megahertz, enabling contactless energy delivery, but facing challenges in efficiency, penetration depth, and safety. MRI safety and compatibility assessments are critical for implants exposed to RF fields in the tens to hundreds of megahertz range depending on the magnetic field strength. Electromagnetic interference further spans a wide frequency spectrum, particularly for devices incorporating communication or control circuits in the megahertz-to-gigahertz bands. Patient-specific modelling and standardized procedures play a key role in establishing reliable safety evaluations.

Based on this roadmap of the need for dosimetry in seminal applications, we hope to provide a good overview of the key challenges and future directions for the individualized and safe application of NIR in medicine and healthcare. The authors hope that this will serve as a valuable reference for researchers and practitioners in this evolving field, and hence that will accelerate the development and increase the impact of tailored NIR-based medical applications.

Acknowledgements

This roadmap article was organized by A H, I L, and M P The authors of each section are listed in the headings of the corresponding section. The document was edited and prepared by A H, I L, and S K. After the document was prepared, the authors reviewed the corresponding sections and approved the final version. The summary of each section has been reviewed by the authors of each section. A H received funding from JSPS KAKENHI 21H04956, which was partly related to this article. M P was financially supported by Eurostars Project 3868 (SENS-THERM) and Project P21-34 (CARES) of the research programme Perspectief, which is financed by the Dutch Research Council (NWO).

1.2. Basic physics and definition

Christopher L Brace1, Marta Cavagnaro2, Yinliang Diao3, Akimasa Hirata4, Ilkka Laakso5, Margarethus Marius Paulides6,7, Punit Prakash8 and Dario B Rodrigues9

1 Departments of Radiology and Biomedical Engineering, University of Wisconsin-Madison, Madison, WI, United States of America

2 Department of Information Engineering, Electronics, and Telecommunications, Sapienza University of Rome, Rome, Italy

3 College of Electronic Engineering, South China Agricultural University, Guangzhou, People’s Republic of China

4 Department of Electrical and Mechanical Engineering, Nagoya Institute of Technology, Nagoya, Japan

5 Department of Electrical Engineering and Automation, Aalto University, Espoo, Finland

6 Care + CureLaboratory of the Electromagnetics Group (EM4Care + Cure), Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands

7 Department of Radiotherapy, Erasmus University Medical Center Cancer Institute, Rotterdam, The Netherlands

8 Department of Biomedical Engineering, The George Washington University, Washington, DC, United States of America

9 Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, United States of America

Electromagnetics

To understand dosimetry in the context of NIR, it is essential to recall the fundamental physical equations that govern EM interactions with biological tissues. Full-wave Maxwell’s equations are required at frequencies where the wavelength of the EMF is comparable to or smaller than the dimensions of the human body (typically in the RF and higher ranges). Expressions in both the time and frequency domains are used in biomedical applications. In the frequency domain,

×H(ω)=σE(ω)+jωε0ε(ω)E(ω),
×E(ω)=jωμ(ω)H(ω),

where ω is the angular frequency (rad s−1), E the electric field vector (V m−1), H the magnetic field vector (A m−1), μ(H m−1) the magnetic permeability that can be considered equal to that of the vacuum μ0 (H m−1), ϵ0 the permittivity of vacuum (F m−1), ϵ(ω) the relative permittivity, and σ(S m−1) the ionic (DC) conductivity. The relative permittivity ϵ(ω) is a complex quantity, typically expressed as ε(ω)=ε(ω)jε(ω), while σ is real. When ω > 0, these can be combined into the relative complex permittivity ε^(ω) such that ε^(ω)=ε(ω)+σ/(jωε0). All variables are spatially dependent, but the explicit dependence has been omitted here for clarity.

Similarly, the term equivalent conductivity σeq(ω) is often used because it accounts for both dielectric and ionic losses, σeq(ω)=σ+ωε0ε(ω). To solve Maxwell’s equations, appropriate boundary and initial conditions must be used, depending on the computational method employed.

At low frequencies, i.e. in the quasistatic regime, the EM wavelength is much larger than the system under study, and the secondary magnetic fields induced by tissue currents are negligible, so simplified formulations are often sufficient (Cheng 2014). Consequently, the electric field can be expressed as E=φjωA, where ϕ is the scalar potential and A is the magnetic vector potential of the applied magnetic field. Depending on whether the problem is dominated by the electric or magnetic field, ϕ can be determined by solving either

σeq(ω)φ=jωq,or
σeq(ω)φ=jωσeq(ω)A,

where q is the charge density generated by an applied electric field or current. This formulation is widely used in LF dosimetry, such as in studies of nerve stimulation and in RF ablation. The applicable frequency range of the quasistatic approximation depends on factors such as the source dimensions and the exposure scenario. In practice, it is typically valid from direct current up to the lower MHz range (Hirata et al 2013, Wang et al 2024).

These two regimes—full-wave and quasistatic—form the physical basis for dosimetry models used throughout this roadmap. The appropriate formulation depends on both the exposure frequency and the relevant biological interaction mechanism (stimulation or heating). At MMW frequencies and above, where the penetration depth of EMFs is low, exposure is often characterized using the reflection coefficient or absorbed power ratio, rather than by directly solving Maxwell’s equations.

Bioheat transfer equations (BHTEs)

The temperature increase induced by the absorption of the EMF can be determined by solving the bioheat equation (Pennes 1948):

ρCT(t)t=(KT(t))B0(t)(T(t)TB(t))+M0(t)+ρSAR

where T is the temperature (°C), C the volumetric specific heat capacity (J kg−1 K−1), ρ the tissue density (kg m−3), K the thermal conductivity (W m−1 K−1), B0 is a quantity proportional to blood perfusion rate (W m−3 K−1), M0 the volumetric metabolic heat generation rate (W m−3), TB is the blood temperature (°C), and SAR is the specific absorption rate (W kg−1), defined below. All variables are spatially dependent, but the explicit dependence has been omitted here for clarity.

The boundary condition (BC) associated with the bioheat equation is typically a convective (Robin-type) condition:

KT(t)n=H(Ts(t)Ta(t)),

where H is the combined convective and radiative heat transfer coefficient (W m−2 K−1), Ts the temperature at tissue BC (typically at the skin level), Ta the temperature of the surrounding material, and n denotes the outward normal to the tissue BC.

The initial temperature distribution within biological tissue is obtained solving the steady-state bioheat equation:

(KT(r))+M0(r)B0(r)(T(r)TB(0))=0.

Definitions of related physical quantities

In this subsection, the physical quantities used in EM dosimetry are defined, as they are referenced in different sections. Thermal dose metrics are not included in this subsection because their definitions vary by application. Instead, they are introduced in the respective subsections.

A. Specific absorption rate

In the international exposure guidelines, SAR is the standard physical quantity used to characterize the heating at frequencies below 6 GHz, and it is widely applied in the evaluation of medical applications. This definition reflects its direct relationship to temperature rise and its unit resemblance to absorbed dose (grey, Gy = J kg−1), which facilitates interpretation in a biomedical context. SAR is defined as

SAR(r)=σeq(r)|E(r)|2ρ(r),

where E(r) is the complex electric field (V m−1 rms (root mean square)).

From the bioheat equation, it can be shown that in phantoms (no perfusion and no metabolic heat generation) and for short exposure times (initial linear temperature rise), the SAR is directly proportional to the temperature rise:

SAR(r)=C(r)dT(r)dt|t=0.

The two formulations are consistent, as the temperature rise reflects all losses included in σeq.

B. Power density

In international exposure guidelines, the relevant dosimetric quantities at frequencies above 6 GHz are the absorbed power density (APD) and incident power density (IPD) averaged over a specified area (1 or 4 cm2). APD is conceptually similar to SAR, but at these higher frequencies power absorption becomes superficial, leading to surface heating. Consequently, heating in the internal tissues is limited and primarily governed by heat conduction. APD is defined as follows,

Sab=1AsAsRe[E(r)×H(r)]ds=1AsAs0zmaxσeq(r)|E(r)|2ρ(r)dzds

where E(r) and H(r) denote the complex rms electric and magnetic fields inside the body model, respectively; * indicates the complex conjugate; ds is the integral variable vector; As is the averaging surface area; and zmax is the depth to which the field significantly penetrates. The area-averaged IPD is defined as follows:

Sinc=1AsAsRe[E(r)×H(r)]ds

where E(r) and H(r) are defined in free space, in contrast to equation (10). In the far-field regime, APD can be approximated from IPD (defined in free space) by including the power transmission factor, (1R), yielding Sab(1R)Sinc, where R is the reflection coefficient at the air–skin interface.

2. Methods for personalized and accurate NIR dosimetry

2.1. Tissue dielectric properties

Kensuke Sasaki1 and Lourdes Farrugia2

1 National Institute of Information and Communications Technology, Koganei, Japan

2 Department of Physics, University of Malta, Malta

Status

Tissue dielectric properties (i.e. permittivity and electrical conductivity) are fundamental physical constants that characterize the electrical response of biological materials. They are essential for EM modelling used across various research fields, such as NIR safety, medical/healthcare applications, and information and communication applications via the human body. The values of these tissue dielectric properties are both frequency- and temperature-dependent and vary with the tissue type and pathology. In the case of NIR dosimetry for the human body, dielectric measurements for each tissue and frequency have been performed; review studies have been reported by Foster and Schwan (1989), Gabriel et al (1996), Amin et al (2019), and Sasaki et al (2022). Because tissue dielectric properties are sensitive to tissue conditions (e.g. temperature and hydration), research in this area has also explored potential applications in medical diagnostics and therapeutic modalities.

Extensive datasets of tissue dielectric properties, including their frequency and temperature dependence and variability among tissues, have been reported in recent reviews (e.g. Sasaki et al (2022)). These works provide numerical values and graphical comparisons across a wide frequency range and also discuss the underlying physical mechanisms, such as dispersion regions. Although the full reproduction of these datasets is beyond the scope of this roadmap, readers are referred to these reviews for detailed numerical data and measurement comparisons.

Current and future challenges

NIR dosimetry covers a broad frequency range, from extremely low frequencies (approximately tens of hertz) to MMW frequencies (Hirata et al 2021). With ongoing advancements in wireless communication, particularly beyond 5G and 6G technologies, there is growing interest in sub-MMW frequencies extending up to a few THz. This introduces new research opportunities and challenges for characterizing the dielectric properties of biological tissues across a wider spectrum. Contrary to advances in high-resolution computational models, comprehensive dielectric property data below 1 MHz remain scarce and represent a key challenge for future research. This difficulty arises primarily from artefacts, such as electrode polarization and the dominance of ionic conduction and polarization effects (Kuang and Nelson 1998). In addition, dielectric properties at these frequencies may become anisotropic, further complicating their characterization. Unlike at higher frequencies, validation through temperature rise measurements is not feasible, making it even more challenging to obtain reliable and reproducible data below 1 MHz. As a result, reliable and reproducible data in this LF regime are still limited. Furthermore, investigating the dielectric response at extremely low frequencies (1–300 Hz) is also of interest, particularly for understanding the dielectric dispersion mechanisms in tissues.

The characteristics of tissue dielectric properties for medical and healthcare applications are topics of interest in the frequency range from tens of megahertz to a few gigahertz. Research in this field requires accurate knowledge of the dielectric properties of healthy and malignant tissues for diagnostic purposes, and the mechanism by which the dielectric properties change in response to various pathological conditions. Furthermore, therapeutic technologies such as hyperthermia and ablation rely on the thermal and dielectric properties of tissues, which vary owing to tissue degeneration and temperature changes during treatment. Although the time-variant nature of tissue dielectric properties is known, an efficient dosimetry method that accounts for this nature has not yet been fully developed.

At higher frequencies, i.e. MMW and sub-MMW frequencies, the penetration depth of EM waves becomes shallow. This makes the dielectric properties of superficial tissues (tissues that constitute the skin and eye) a special topic of interest. THz time-domain spectroscopy (THz-TDS) has extended the frequency range over which the tissue dielectric properties can be measured (Pickwell et al 2004). Future research should address the dominant factors (e.g. skin wrinkles and hydration are the potential factors) that influence the tissue dielectric properties, and the modelling of the dielectric properties, which embed the dominant factors as parameters, is essential, particularly in NIR dosimetry.

Advances in science and technology to meet the challenges

Dielectric measurement techniques have been developed over time, enabling the characterization of tissue dielectric properties over a broad frequency spectrum. At kilohertz frequencies or lower, the primary method is to use dielectric measurement cells or sensors constructed using pair(s) of electrodes (Schwan and Kay 1957). For higher frequencies, typically beyond hundreds of megahertz, coaxial line-based sensors are a common approach, as pioneered by Stuchly and Stuchly (1980). These conventional measurement techniques remain fundamental and are widely used because they enable broadband frequency measurements, thus demonstrating their effectiveness. Measurement guidelines have been developed to ensure proper usage and measurement reproducibility when using such techniques (Farrugia et al 2024).

Recent advancements in the determination of tissue dielectric properties apply artificial intelligence (AI) techniques to conventional measurement techniques (Bonello et al 2020) and in vivo MRI-based electrical property tomography (MR-EPT) (Mandija et al 2019, Rashed et al 2020, Kubota et al 2025, Zilberti et al 2025). The imaging approach resolves EM inverse problems and allows noninvasive estimation of tissue dielectric properties with the resolution of imaging technology. More recently, an innovative approach has been proposed to fuse computational dosimetry techniques and conventional imaging techniques to estimate dielectric properties (Kangasmaa and Laakso 2022). These advancements in accuracy and efficiency allow for the understanding of intratissue variability that are difficult to capture using conventional measurement techniques. It is also noteworthy that these imaging-based approaches enable non-uniform assignment of dielectric properties within tissues rather than relying solely on tissue-specific averaged values.

These methodologies continue to advance, paving the way for deeper insights into the variability of dielectric properties within the same tissue type. Conventional measurement techniques often struggle to quantify such variations, and in most methodologies, average dielectric properties are measured. However, modern AI-driven and computational imaging approaches have the potential to accumulate precise, high-resolution data on how dielectric constants fluctuate due to physiological and pathological changes. These advancements have enabled more detailed, personalized assessments of tissue dielectric properties, which could enhance applications in medical diagnostics, treatment planning, and NIR safety evaluations.

Concluding remarks

For the tissue dielectric properties, which are fundamental physical quantities essential for computational NIR dosimetry, the demand for data across different frequency and tissue type (including tissue state) tends to expand with the development of science and technology. Because exhaustive measurements of all tissues and frequencies are impractical, the development of advanced estimation approaches can become a core topic in the near future. To promote personalized NIR dosimetry, several studies have attempted to estimate personal dielectric properties (e.g. Rashed et al (2020), Kubota et al (2025)). However, these methods remain at an early stage and require further validation even when anatomical human models are available before they can provide truly personalized and accurate dosimetry. Therefore, in parallel with the development of advanced approaches, the continuous accumulation of high-quality experimental data using scientifically well-characterized measurement techniques, along with ensuring traceability, remains essential for validating the estimation methods and achieving reliable dosimetry.

2.2. LF and radiofrequency computational methods

Sachiko Kodera1 and Yinliang Diao2

1 Department of Electrical and Mechanical Engineering, Nagoya Institute of Technology, Nagoya, Japan

2 College of Electronic Engineering, South China Agricultural University, Guangzhou, People’s Republic of China

Status

A. LF dosimetry

Accurate modelling of the interaction between LF electric and magnetic fields and biological tissues is essential for applications such as bioelectric signal analysis, neuromodulation, nerve stimulation, and electrotherapy. Many living cells and tissues generate and respond to electrical signals, and the dominant mechanism of LF EMF exposure involves the stimulation of excitable tissues, such as nerves and muscles. Consequently, dosimetry focuses on induced electric fields, whether generated by external stimuli or biological activity. LF dosimetry is based on a quasistatic approximation (see section 1.2) in which the EMF wavelengths are significantly larger than the dimensions of the system under study.

Computational models are essential for optimizing electrotherapy techniques. These models include TES, TMS, spinal cord stimulation, and source localization for electroencephalography (EEG) and electrocardiography (ECG). Anatomical human models commonly used in dosimetric analyses are constructed via tissue segmentation based on medical images, mainly using MRI and computed tomography. For medical applications, such as brain stimulation, procedures have been developed to enable semiautomatic segmentation from individual MRI scans to provide personalized EMF dosimetry (Laakso et al 2015, Saturnino et al 2019b).

The boundary element method (BEM) (Sauter and Schwab 2010) and finite element method (FEM) (Wang and Eisenberg 1994) are widely used in electrophysiological source reconstruction techniques (e.g. EEG, magnetoencephalography (MEG), and ECG) and brain stimulation techniques (e.g. TES and TMS). The BEM calculates the potential of surface elements at the interface between tissue compartments. This is preferred when the surface potential and current density are the primary quantities of interest. However, the BEM cannot handle inhomogeneous and anisotropic tissues (Dayarian and Khadem 2024). In the FEM, the entire volume is discretized into small elements (typically tetrahedrons), and the potential at all nodes is calculated. The FEM is particularly advantageous for incorporating arbitrary geometries and heterogeneous and anisotropic tissue properties. However, the computational resources are higher, and singularities are encountered when using point dipoles as current sources in EEG forward models (Beltrachini 2018).

Unlike mesh-based methods, the scalar-potential finite-difference (SPFD) method (Dawson and Stuchly 1996) uses voxel-based uniform grids, which makes it suitable for anatomical models generated from medical images. The SPFD method is effective for heterogeneous and anisotropic models and has been widely used in numerical dosimetry, for evaluating LF EMF safety. Compared to the FEM, it usually requires less computational memory and time (Soldati and Laakso 2020). However, one major problems with voxel-based modelling is the computational artefacts arising from the staircasing approximation of curved tissue interfaces (Hirata et al 2021). Other approaches are the quasistatic finite-difference time-domain (FDTD) method (Gandhi and Chen 1992), which uses a step function to approximate the voltage source, and the impedance method (Deford and Gandhi 1985), which represents human tissue as a network of impedances.

B. RF dosimetry

In RF-based medical applications, such as hyperthermia therapy, thermal ablation, and MRI-related safety assessments, dosimetry is essential to estimate power absorption in biological tissues and the corresponding temperature rise. The SAR is a key parameter for evaluating the temperature increase in tissues, and it affects treatment and safety considerations. Numerical dosimetry for RF exposure often uses the FDTD method (Taflove and Hagness 2003) because of its ability to handle inhomogeneous tissues and geometric complexities, as well as its suitability for parallel computation using graphics processing units. Similar to LF dosimetry, the FEM and BEM are also widely used in RF applications. The FEM is beneficial for applications involving complex vascular geometries, such as tumour thermal therapy, although it typically requires higher computational resources. However, it is also used in cases in which computational efficiency is a priority; though, it suffers from scalability issues when applied to complex and inhomogeneous geometries (Hall and Hao 2012). Thermal modelling is essential for evaluating tissue heating for RF-based medical treatments. The Pennes BHTE (Pennes 1948) (see section 1.2) is widely used to estimate the temperature rise in the human body. The BHTE was originally proposed as a simplified equation for heat transfer between tissues, but it has been extended to take into account other factors such as evaporation, blood perfusion, convection, and temperature changes in arterial and venous blood (Fiala et al 2001, Tucci et al 2021, Radmilović-Radjenović et al 2022). Due to these extensions, the BHTE has become applicable to various medical applications, including RF-induced heating in patients within MRI systems (Ertürk et al 2015) and thermal therapy (Keangin et al 2011, Paulides et al 2013).

Challenges and future directions

A. LF dosimetry challenges

Accurate computational dosimetry is essential for enabling individual dosimetry for diagnosis and treatment planning, thereby ensuring the safety of individuals exposed to EMF and establishing safe exposure limits for various scenarios (Diao et al 2023). To achieve this, continuous improvements in computational methods and the development of more accurate human models considering tissue anisotropy are required. The fast computations are also important, especially when real-time feedback is required. Multiscale modelling combining EMF computations and nerve activation models (Reilly and Diamant 2011) has attracted attention. This approach can help us better understand the effects of LF EMF at cellular level. Furthermore, exploring the effects of non-sinusoidal waveforms on nerve interactions is critical because it provides deeper insights into safety guidelines and informs them of therapeutic strategies.

B. LF dosimetry challenges

In hyperthermia therapy, personalized dosimetry has attracted attention because of its potential to optimize patient-specific treatment outcomes (Verhaart et al 2015, Wells et al 2021). This trend is likely to be extended to other RF treatments. However, the accurate modelling of tissue structures, such as blood vessels and tumours, remains a challenge (Lu et al 2024). The variation in the dielectric properties further complicates the dosimetry accuracy in hyperthermia and thermal ablation treatments. In RF dosimetry for thermal ablation, the temperature dependence of the dielectric properties of biological tissues should be clarified (Rossmann and Haemmerich 2014). Exposure assessments should be standardized to ensure consistency in medical and safety applications. A challenge in thermal modelling is the limited availability of reliable data on the thermal properties of biological tissues. Tissue thermal properties were partly estimated by extrapolation from in vitro or animal data or empirical equations (Duck 1990, Mcintosh and Anderson 2010, Hasgall et al 2023). However, differences in water content among individuals affect not only dielectric properties but also thermal properties (Michel et al 2017). To improve model accuracy, integrating subject-specific data on thermal properties, including individual variations, is essential. Another key challenge is modelling thermoregulatory responses (Kodera et al 2018). Most computational studies do not consider the dynamic blood flow and temperature feedback mechanisms, which are critical for realistic temperature increase predictions (Fiala et al 2001). Integrating more detailed data on thermoregulatory responses is essential for improving the accuracy of temperature increase estimation and for assessing temperature increases, considering individual variabilities.

C. Validation of dosimetric computation

Early attempts used saline or phantom models (Chen et al 1982, Kaune and Forsythe 1985). Intercomparison studies across groups and numerical methods have also been performed for RF local exposures (Beard et al 2006). As direct invasive measurements in human tissues are practically impossible, comparisons with analytical models and different computational approaches are common practice to ensure reliability.

Advances in science and technology to meet challenges

Dosimetry can help us better understand the interactions between EMFs and the human body and guide the development of devices that use these interactions. Advancements in dosimetry are likely to emerge alongside new device designs and applications, combining sophisticated multiphysics simulations for design and optimization with swift prototyping and testing tools.

A. Advances in LF dosimetry

Innovations in computational methods, such as improved FEM algorithms and hybrid techniques, promise greater accuracy and efficiency in modelling the interactions between EMFs and the human body. For instance, subtraction methods have been proposed to address singularity issues in the FEM (Beltrachini 2018). The hybrid BEM–FEM method, which combines the advantages of both approaches by ensuring continuity of the current density across the interface between the isotropic domain (using the BEM) and the anisotropic domain (using the FEM), has been proposed (Dodig et al 2021). However, this approach generally requires longer simulation times. Recently, a BEM with fast multipole acceleration was proposed to speed up computations using high-resolution bioelectromagnetic models (Makarov et al 2018). For the real-time computation of the TMS-induced electric field in the brain, a reciprocity-based method was introduced (Stenroos and Koponen 2019). In the finite-difference method, the geometric multigrid (Laakso and Hirata 2012) and algebraic multigrid (Stroka et al 2024) methods are used to accelerate computation, particularly when integrated with multiple graphics processing units (Xiong et al 2015).

Regarding computational artefacts, postprocessing methods that adjust percentile values based on the statistical distribution of the top 1% of electric field strengths have been proposed (Gomez-Tames et al 2018, Arduino et al 2020). In Laakso and Hirata (2012), a preprocessing smoothing conductivity algorithm was effective in reducing staircasing artefacts. However, this approach results in the loss of anatomical model details. Furthermore, recent studies have proposed an effective tensor conductance model to reduce staircasing artefacts in the SPFD method (Diao et al 2022, 2023). Moreover, multiscale simulation techniques that incorporate more realistic nerve models are expected to improve the accuracy of threshold estimation, which would also be valuable for refining spatial averaging methods for induced electric fields.

AI, particularly deep learning, has recently seen rapid growth due to advancements in hardware and software. Studies have reported the feasibility of using deep learning for the fast estimation of TMS-induced electric fields, achieving an estimation time of <0.1 s (Yokota et al 2019, Moser et al 2024). However, large datasets are required for training, and the neural network architecture must be tailored to specific problems. Embedding physics in deep learning networks can improve interpretability and generalizability; however, challenges remain (Guo et al 2023).

Imaging techniques such as EIT and MR electrical impedance tomography (MREIT) provide non-invasive approaches to estimate conductivity and induced current distributions (Scott et al 1991, Hasanov et al 2008). These methods could indirectly support the validation of computational dosimetry, although their spatial resolution is limited, and they are sensitive to noise and modelling assumptions such as isotropy. Extensions such as DT-MREIT have been proposed to reconstruct anisotropic conductivity (Kwon et al 2014, Sadighi et al 2021), but further refinement is required before they can serve as reliable validation tools.

B. Advances in RF dosimetry

In the field of EMF safety, high-resolution generic human models have been developed considering the detailed morphology of the skin (e.g. sweat ducts, nerves, epidermis, dermis, and subcutaneous tissue). Eye (e.g. cornea, iris, lens, sclera, vitreous body, and anterior chamber) have been developed to enhance high-precision dosimetry (Karampatzakis and Samaras 2010, Sasaki et al 2017, Haider et al 2022). These advances could be useful for medical applications such as thermal and retinal laser therapies. One of the key challenges in hyperthermia and ablation include improving the focus of the irradiated area and minimizing the residual heat in the area surrounding the tumour (Hassan et al 2021). Achieving this requires detailed and personalized EMF dosimetry that reproduces the complex structure of the tumour and surrounding tissue, statistical approaches that consider the variability of dielectric properties, and estimation of the time course of the temperature rise considering thermoregulation. Thermoregulation modelling could be further improved by clarifying the relationship between temperature, blood flow, and sweating using measured data obtained from the human subject experiments. A more accurate understanding of these physiological responses would improve the accuracy of estimations of temperature changes in biological tissues during treatments, especially under high-intensity exposure conditions. Furthermore, considering the interindividual variability of thermal properties such as specific heat capacity could help quantify uncertainties in estimated temperature rise. Future research should focus on individual differences in thermoregulatory responses, including those related to age, sex, environmental conditions, and pathology/diseases. The use of real-time physiological monitoring data obtained from wearable sensors may enable the validation of these models more simply and under various conditions. The advanced imaging techniques such as dynamic contrast-enhanced MRI combined with computational simulations may improve the depiction of microvascular blood flow, ultimately improving the accuracy of the evaluation of thermal therapy (Singh 2024).

Concluding remarks

Integrating LF and RF dosimetry into medical applications is important for optimizing the therapeutic outcomes and patient safety. Although computational modelling techniques have substantially improved the accuracy and resolution of dosimetric assessments, challenges related to physiological variations, tissue properties, and thermoregulation remain an active area of research. Accurate and individualized dosimetry is expected to improve the optimization of therapeutic techniques. Improved dosimetric models can enhance the precision of these treatments and lead to improved patient outcomes. Future advances in numerical modelling combined with machine learning and real-time physiological monitoring will further enhance personalized treatment planning. Furthermore, incorporating EM interaction modelling that accounts for tissue anisotropy and dynamic physiological responses will enable dosimetry evaluations that reflect real-world exposure scenarios and individual differences, such as age, sex, and morphology. These advancements represent safer and more effective medical therapies involving EMF exposure, thus benefiting a wide range of clinical applications.

Acknowledgements

Sachiko Kodera was supported by JSPS KAKENHI Grant Number JP24K00868.

3. Brain stimulation

3.1. TES

Ilkka Laakso1, Serena Fiocchi2 and Alexander Opitz3

1 Department of Electrical Engineering and Automation, Aalto University, Espoo, Finland

2 CNR—National Research Council, Institute of Electronics Computer and Telecommunication (IEIIT), Milan, Italy

3 Department of Biomedical Engineering, University of Minnesota, Minneapolis, MN, United States of America

Status

TES refers to a class of noninvasive neuromodulation techniques that apply weak electrical currents to the scalp to modulate brain activity. The applied current generates an electric field within the brain, which is thought to modulate the function of the cerebral cortex, without directly triggering action potentials. These methods have gained widespread attention in both research and clinical settings due to their safety (Antal et al 2017) and potential applications in the treatment of various neurological and psychiatric disorders (Kuo et al 2014).

Various forms of TES exist, each with specific mechanisms and potential advantages. The most commonly used technique is transcranial direct current stimulation (tDCS), in which a constant, low-intensity direct current is applied to alter cortical excitability depending on the current polarity (Nitsche and Paulus 2000). Another common technique is transcranial alternating current stimulation (tACS) (Antal et al 2008), which applies sinusoidal currents at specific frequencies to influence neural oscillations. In temporal interference stimulation (Grossman et al 2017, Violante et al 2023), kilohertz tACS at different frequencies is applied through two or more pairs of electrodes to produce patterns of two interfering electric fields, which allows for modulation of deeper brain structures compared to conventional tACS.

Personalized dosimetry in TES relies on advanced computational and imaging tools to optimize stimulation parameters for individual subjects. FEM-based models, often constructed using high-resolution structural MR images and diffusion-weighted imaging, allow for subject-specific electric field estimations by incorporating variations in skull conductivity, cerebrospinal fluid (CSF) distribution, and brain tissue properties. Freely available software platforms like SimNIBS (Saturnino et al 2019b) and ROAST (Huang et al 2019) provide automated pipelines for generating individualized electric field simulations, enabling clinicians and researchers to predict stimulation intensity at target brain regions.

These models have allowed new strategies for tuning the electric field dose through optimization of TES electrode montages. A multitude of possible electrode placements, using either classical bipolar montage or arrays of smaller electrodes, permit shaping current flow patterns through the head or targeted stimulation of cortical areas. In this context, electrode montage personalization has emerged as approach to enhance stimulation focality and ensure consistent electric field delivery across subjects (Evans et al 2020). Moreover, advanced strategies, such as focal tDCS setups and multi-electrode configurations (Kuo et al 2013, Parazzini et al 2017, Niemann et al 2024), have been investigated to optimize stimulation precision. More recently, MRI-free approach using EEG coordinate-based electrode placement has demonstrated to enhance both the intensity and focality of electric fields compared to conventional configurations (Caulfield and George 2022).

Challenges and future directions

Personalization of TES holds the potential to significantly improve the consistency and predictability of treatment outcomes by addressing the substantial interindividual variability observed in response to TES (e.g. (López-Alonso et al 2014, Wiethoff et al 2014, Therrien-Blanchet et al 2023). While the underlying causes of the variability are not yet established, several factors are thought to contribute to it, including individual head anatomy, brain features, ongoing brain state, and the resulting electric field distribution (Van Hoornweder et al 2023). By tailoring stimulation parameters to an individual’s specific characteristics, personalized TES aims to optimize the electric field strength and focality in the targeted brain regions (Simula et al 2022, Van Hoornweder et al 2022, 2023).

However, linking individual electric fields to neurophysiological outcomes has been challenging so far. For tDCS of the primary motor cortex, studies have examined the relationship between individual electric fields and motor evoked potential size modulations, but results have been inconsistent. They have reported negative effects (Laakso et al 2019), positive effects (Mosayebi-Samani et al 2021), no effects (Ahn and Fröhlich 2021), and participant-specific, nonlinear effects (Laakso et al 2024). Individual electric field strengths have also been associated with positive effects on GABA modulation and increases in sensorimotor network strength (Antonenko et al 2019). While prospective personalized dosing of motor cortical tDCS and tACS has been explored for reaction time and working memory, the results showed little to no improvement over conventional dosing (Joshi et al 2023). Studies in other brain regions also show mixed results: some found strong positive associations between electric fields and functional connectivity (Indahlastari et al 2021), while others reported weaker or more variable effects (Müller et al 2023). Higher electric field strength was related to faster reaction times in one of two working memory tasks but had no effect heart rate variability (Razza et al 2024a, 2024b). Over the visual cortex, individual tACS electric fields were associated with stronger alpha-power of the magnetoencephalogram (Kasten et al 2019). Overall, the dose-response relationship is still unclear due to conflicting results in various studies.

Furthermore, electric field simulations are only an approximation of physical reality and include various sources of uncertainty. Initially, spherical models were used to predict TES electric fields (Miranda et al 2009, Miranda 2013). While being only a coarse representation of the intricate head and brain morphology, spherical models nevertheless provided important insights into TES physical mechanisms. Today, anatomically realistic head models can be derived from individual MR images, allowing more precise estimation of electric fields, but incomplete validation remains a concern. To address this, indirect validation of dosimetric computations has been attempted using recordings in heterogeneous anatomically-shaped phantoms (Kim et al 2015), non-invasive voltage measurements in humans (Datta et al 2013), or MREIT (Mosayebi-Samani et al 2025). Complementarily, stereotactic EEG electrode recordings in both non-human primates and neurosurgical patients have revealed maximum field strengths around 0.5 V m−1 (Opitz et al 2016). Moreover, intracranial recordings at subcortical sites, such as the subthalamic nucleus in humans, yielded field estimates in the range of 0.08–5 V m−1 depending on target region and stimulation modality (Chhatbar et al 2018, Ruhnau et al 2018). Notably, these efforts converge with pioneering in vivo intracranial recordings (Huang et al 2017), which demonstrated strong spatial and quantitative correspondence between individualized finite-element models and patient measurements (Pearson’s r ≈ 0.86–0.89), following calibration of tissue conductivities. Collectively, these validation strategies from phantom studies to human and animal in vivo recordings reflect a steadily increasing commitment to grounding TES dosimetry in terms of empirical evidence and anatomical realism.

For TES applications, electrodes are often placed based on the EEG 10–20 system (Antal et al 2017, Wischnewski et al 2023) which can introduce consistent variability in placement with respect to anatomical targets. With the increasing popularity of take-home TES (Palm et al 2018) where stimulation electrodes are placed by participants themselves, localization errors are likely larger. Imprecise electrode location can have large effects on the electric fields in the brain (Opitz et al 2018) and lead to suboptimal outcomes.

Additionally, tissue conductivity variations can introduce further uncertainty and affect the predicted electric field strength. While the directionality of predicted electric fields is generally more robust, electric field strength is more affected by conductivity uncertainties (Saturnino et al 2018). For electric field simulations, tissue conductivities serve as input parameters, typically derived from ex vivo animal models (Gabriel et al 1996), with considerable variability reported across measurements (Gabriel et al 1996). Also, tissue conductivities can change during aging (Akhtari et al 2002). For TES, non-brain conductivities, such as of the skin, skull and CSF, play a crucial role in determining the amount of current shunting occurring before the current reaches the brain (Narayan et al 2023; Opitz et al 2015; Salvador et al 2012), thus affecting the electric field strength. Furthermore, the conductivity of white matter is anisotropic, which may lead to additional uncertainty, as its conductivity needs to be estimated from diffusion-weighted MRI data (Wolters et al 2006). However, modelling white matter as anisotropic has been shown to have only a small effect relative to overall conductivity uncertainties (Huang et al 2017, Mosayebi-Samani et al 2025), and both isotropic and anisotropic conductivity models have provided equally good agreement with the measured MREIT data (Mosayebi-Samani et al 2025). Related to the influence of tissue conductivity uncertainties is the effect of segmentation errors. If voxel tissues are assigned an incorrect segmentation label, the conductivity distribution in the head model is altered, leading to inaccurate local electric field estimation.

Science and technological advances addressing the challenges

Several scientific advances are needed to address the optimal electric field dose and its relationship with the response, which currently remain unclear, hindering personalization of TES. Presently available dose-response studies in volunteer participants are highly heterogeneous, having investigated various outcomes, applied different experimental protocols, variable modelling parameters, and dosing parameters for characterizing the individual electric fields (Van Hoornweder et al 2023). To address this heterogeneity, larger studies with standardized experimental protocols and consistent electric field dose measures are needed. Furthermore, computational and experimental strategies need to be developed to avoid errors caused by inaccurate electrode localization, and in vivo measurement of electrical conductivities in humans are needed to reduce uncertainties associated with tissue electrical properties.

The majority of TES studies have not reported individual electric field data but instead report the dose in terms of electrode size and montage and stimulation current intensity, resulting in variability in electric field distributions in the brain. This makes it challenging to compare outcomes across studies and establishing dose-response relationships. However, variations in stimulation parameters will result in different induced electric fields which can be accounted for using computational modelling, allowing for meta-analyses that compare and aggregate studies based on estimated TES electric field distributions. For instance, combining electric field modelling with a traditional meta-analysis has been shown to identify brain regions involved in performance changes in working memory following tDCS (Wischnewski et al 2021, Wischnewski et al 2024) as well as identify different brain regions causally altered by theta and gamma (tACS) (Wischnewski et al 2024). This meta-dosimetry approach could be a powerful tool to identify effective electric field dosing parameters across an increasingly growing TES literature.

In addition to computational approaches, new experimental studies in animals can provide detailed and accurate information into the TES dose-response relationship. Animal studies leveraging invasive physiological recordings have already demonstrated that TES physiological effects are dose-dependent (Johnson et al 2020, Alekseichuk et al 2019, Krause et al 2022). For example, Johnson et al (2020) showed that increasing tACS electric field strength leads to greater neuronal spike entrainment to external oscillations. Thus, recordings in animals can help identify TES parameters to achieve a desired physiological response. However, care has to be taken to translate findings in animals to human applications due to differences in head/brain anatomy leading to differences in TES electric field strengths (Alekseichuk et al 2019).

The difficulty in identifying the TES dose response relationship may also be related to the dependence of TES effects on the brain state when stimulation is delivered. A recently developed neural network (Zhao et al 2024) shows that the effects of tACS depend on the ongoing network oscillation and the applied stimulation frequency. For strong ongoing brain oscillations, tACS first disrupts existing spike field coupling, and only at higher intensities does it start to entrain spiking activity to the external oscillations. Additionally, real-time EEG-informed current steering has emerged as a potential approach to enhance targeting precision by adjusting TES in response to ongoing brain activity (Wischnewski et al 2024). These findings show that ongoing brain states will affect TES dose-response relationships, and it is beneficial to consider these effects in future studies on dose response.

Concluding remarks

Personalized computational dosimetry is a potential approach to reduce variability and improve the therapeutic effectiveness of TES. However, the optimal dosing of the brain electric fields is still unclear and there are several sources of dosimetric uncertainty. To overcome these challenges, we propose reducing the heterogeneity in study design, using standardized dosing parameters, systematic review and meta-analysis of TES literature to identify relevant dosing parameters, animal experiments on dose response, and considering brain-state dependence of TES in experimental design.

3.2. TMS

Zhi-De Deng1 and Sergey N Makaroff 2,3

1 Computational Neurostimulation Research Program, Noninvasive Neuromodulation Unit, Experimental Therapeutics and Pathophysiology Branch, National Institute of Mental Health, NIH 10 Center Drive, Bethesda, MD 20892, United States of America

2 Electrical and Computer Engineering Department, Worcester Polytechnic Institute, Worcester, MA 01609, United States of America

3 Athinoula A. Martinos Ctr. for Biomedical Imaging, Massachusetts General Hospital, Harvard Medical School, Charlestown, MA 02129, United States of America

Status

TMS is a noninvasive brain stimulation technique that uses rapidly alternating magnetic fields to induce electric fields within targeted brain regions. These electric fields are sufficiently intense to depolarize neuronal membranes and induce action potentials, enabling TMS to directly activate neurons in a focal, spatially constrained manner (Barker et al 1985). Unlike TES, TMS achieves direct cortical activation, making it a more potent tool for probing and modulating brain function.

Since its first demonstration in the mid-1980s, TMS has evolved from a laboratory tool into a clinically validated intervention, particularly in the treatment of neuropsychiatric disorders. Its capacity to modulate neural activity in a site-specific, frequency-dependent manner has led to regulatory approvals for a range of neuropsychiatric disorders such as major depressive disorder, obsessive-compulsive disorder, migraine, and smoking cessation (O’Reardon et al 2007, Carmi et al 2019, Lefaucheur et al 2020, Cohen et al 2022). It has also become a valuable adjunct in stroke rehabilitation and a tool for cognitive enhancement (Luber and Lisanby 2014, Luber and Deng 2017, Fisicaro et al 2019). The appeal of TMS lies in its versatility. As both a therapeutic modality and an investigative probe of brain function, TMS occupies a unique position within the broader field of neuromodulation, bridging basic neuroscience, clinical application, and personalized medicine.

Over the past two decades, several distinct forms of TMS have emerged. Among these modalities, repetitive TMS (rTMS) has traditionally been the most established and extensively studied, forming the basis of many clinical protocols. rTMS involves delivering a series of pulses at regular intervals and has demonstrated frequency-dependent effects on cortical excitability. LF rTMS, typically administered at 1 Hz, is generally associated with inhibitory outcomes, thought to reflect mechanisms akin to long-term depression (LTD). In contrast, high-frequency rTMS, often delivered at 5 Hz or higher, tends to facilitate cortical excitability, likely through mechanisms resembling long-term potentiation (LTP) (Fitzgerald et al 2006). A notable evolution of rTMS is theta-burst stimulation (TBS), which capitalizes on brief, high-frequency bursts delivered in theta-range temporal patterns. Intermittent TBS (iTBS) tends to enhance excitability, whereas continuous TBS has been shown to reduce it (Huang et al 2005). TBS protocols can induce lasting plasticity with considerably shorter administration times compared to rTMS, offering logistical advantages in both research and clinical settings.

Beyond these therapeutic paradigms, single-pulse and paired-pulse TMS are commonly employed as investigative tools in neurophysiology. Single-pulse TMS can be used to assess motor threshold and corticospinal integrity. Paired-pulse paradigms—where two stimuli are delivered in close succession—allow researchers to probe intracortical facilitation and inhibition (Kujirai et al 1993). These techniques have yielded critical insights into cortical circuitry, inhibitory/excitatory balance, and the neurophysiological basis of various brain disorders.

Additionally, more experimental paradigms such as quadri-pulse stimulation (Hamada et al 2007) and patterned burst protocols are under investigation for their capacity to induce highly specific and sustained neuroplastic changes. As stimulation technologies advance, the repertoire of TMS protocols continues to grow, offering increasingly nuanced control over how, where, and when neuromodulation occurs.

The effects of TMS unfold across multiple spatial and temporal scales, encompassing both local neurophysiological effects and network-level effects. At the site of stimulation, TMS can modulate neuronal excitability via mechanisms that resemble classical synaptic plasticity, including LTP and LTD mechanisms. These effects are believed to arise from activity-dependent modulation of synaptic strength, as well as from alterations in the balance of excitatory and inhibitory neurotransmitter systems—most notably glutamate and GABA (Ziemann et al 2008). Moreover, the direction and magnitude of these effects are strongly influenced by stimulation parameters, such as pulse frequency, intensity, and pattern. While the local impact of TMS is important, its true functional reach emerges through distributed network modulation. Neuroimaging studies employing functional MRI, EEG, and MEG consistently show that TMS induces changes in functionally connected brain regions. These network-level responses are particularly evident when stimulating cortical hubs that anchor large-scale networks, such as the default mode network, salience network, or frontoparietal control network (Eldaief et al 2011, Fox et al 2012). For example, targeting the dorsolateral prefrontal cortex can modulate activity not only in the prefrontal region itself, but also in the anterior cingulate, striatum, and thalamus—regions implicated in emotional regulation and executive function. Crucially, these network effects are not merely epiphenomenal. They appear to play a central role in the therapeutic efficacy of TMS, with mounting evidence suggesting that normalization of aberrant connectivity may underlie symptom improvement in conditions such as depression. Thus, understanding TMS as a tool that reshapes neural communication—rather than simply exciting or inhibiting individual nodes—offers a more accurate and powerful framework for both research and clinical application.

Safety considerations for TMS commonly highlight auditory effects, as the acoustic artefact of coil discharge can exceed comfortable sound levels. Beyond sound exposure, clinical and neuroscientific safety guidance (Rossi et al 2009) emphasizes adjusting coil design and stimulation parameters to minimize other potential adverse effects. Where relevant, engineering measures to reduce acoustic output may also be considered (e.g. low-noise coil designs).

Challenges and future directions

A major limitation of present TMS systems lies in the rigidity of coil geometry and placement during stimulation. Most coils are designed with fixed coil windings and require physical repositioning to target different brain regions. As a result, they are poorly suited for aligning stimulation with the rapidly evolving spatial-temporal dynamics of brain activity. This mismatch becomes especially critical when attempting to engage distributed or state-dependent networks, as commonly seen in higher-order cognitive tasks or neuropsychiatric conditions involving fluctuating brain states. Even protocols that target ‘hubs’ of large-scale networks assume static topographies and lack the flexibility to adapt stimulation trajectories within or across sessions. These constraints underscore the need for systems capable of dynamic, multifocal, and state-dependent stimulation.

The development of portable TMS systems raises both exciting opportunities and unique technical hurdles. These systems promise to increase access to TMS by enabling home-based treatments or ambulatory neuromodulation. However, portability often entails compromises in coil design, field strength, and power requirements. Integrating lightweight and energy-efficient components without sacrificing efficacy or safety remains a formidable engineering challenge. Moreover, the clinical use of portable TMS will demand new approaches to remote supervision, personalized calibration, and long-term adherence—all of which require coordination between hardware, software, and regulatory frameworks.

Finally, across both traditional and emerging applications, computational modelling remains a pivotal but underdeveloped tool. Most current frameworks terminate at the biophysical interface—predicting electric field strength or estimating neural activation thresholds (Siebner et al 2022)—without resolving how stimulation sequences influence neural circuits over time. As a result, they offer limited insight into how specific stimulation sequences shape neuronal population activity or how repeated pulses reorganize network function over time. Yet, such mechanistic understanding is critical for advancing beyond heuristic protocol design. Equally lacking are models that incorporate plasticity mechanisms capable of capturing the longer-term, cumulative effects of stimulation. While empirical findings suggest that certain TMS protocols induce LTP- or LTD-like changes (Ziemann et al 2008), these effects are typically inferred post hoc rather than predicted a priori. Biophysically grounded models that simulate synaptic plasticity (Robinson 2011, Fung et al 2013)—particularly in response to temporally patterned input—could help identify optimal treatment frequency, session duration, or cumulative dosing strategies for clinical applications. In this way, computational modelling could evolve from a tool for anatomical targeting into a predictive framework for both mechanistic insight and individualized protocol development.

Scientific and technological advances addressing the challenges

A significant frontier in TMS coil technology is the development of multi-locus and electronically steerable stimulation systems, which aim to overcome a long-standing limitation: the need to physically reposition the coil to stimulate different cortical targets. These systems use arrays of overlapping coils whose outputs can be digitally controlled to steer the electric field in real time across the cortical surface without mechanical movement. This innovation enables stimulation of multiple brain areas in rapid succession or along complex spatial trajectories, thereby facilitating more flexible, network-oriented neuromodulation (Koponen et al 2018, Nieminen et al 2022, Rissanen et al 2023). In research contexts, multicoil arrays can even be combined with neuroimaging to allow for causal mapping of functional connectivity (Navarro de Lara et al 2021), while in clinical settings it may enable more efficient engagement of distributed circuits implicated in neuropsychiatric disorders. As such, these systems bring TMS closer to the ideals of adaptive, precision-guided neuromodulation, aligning stimulation not just to anatomy, but to dynamically shifting brain states and network configurations. Another recent advancement in coil engineering focuses not on the shape or depth of stimulation, but on the acoustic features of TMS. A novel class of quiet TMS coils—such as the double-containment coil—has been developed to substantially reduce acoustic emissions (Koponen et al 2021).

One recent development in TMS hardware is a wearable, battery-powered systems capable of delivering suprathreshold repetitive stimulation. A notable example is rTMS-tiny, a lightweight (3 kg), fully portable rTMS system powered by a high-efficiency magnetic stimulator and an innovative bent double-T magnetic core coil (Qi et al 2025a). This system achieves E-field intensities and repetition frequencies comparable to commercial devices, while reducing power consumption to just 10% of conventional rTMS systems. Critically, rTMS-tiny supports delivery of 10 Hz rTMS and iTBS during free behaviours, enabling real-time neuromodulation in ambulatory and naturalistic settings. The ability to stimulate during movement—such as walking—has revealed state-dependent enhancement of cortical excitability, illustrating the potential for closed-loop or context-aware neuromodulation outside laboratory constraints. These breakthroughs not only address long-standing limitations in access and ecological validity, but also open new frontiers for longitudinal intervention, home-based therapy, and real-world causal neuroscience.

Personalization in TMS has become an essential component of both research and clinical applications, aiming to maximize efficacy by tailoring stimulation to individual brain anatomy. A cornerstone of this personalization effort lies in EM modelling—particularly under the quasistatic approximation—which enables prediction of the electric fields induced in brain tissue during stimulation. This approximation, rigorously justified and widely adopted (Wang et al 2024), underpins most modern modelling strategies and is foundational to determining which neuronal populations are likely to be activated by a given TMS pulse. The FEM remains the most prevalent approach for macroscopic modelling of TMS fields. As a powerful numerical technique, FEM excels at solving the complex geometries and heterogeneous conductivity profiles inherent to the human head. Among the open-source tools available, SimNIBS stands out as the leading FEM-based platform for individualized electric field modelling (Thielscher et al 2015, Saturnino et al 2019a, 2019c). It offers a streamlined pipeline for generating subject-specific head models from MRI, running simulations, and visualizing the induced electric fields.

An alternative to FEM is the BEM with fast multipole acceleration (BEM-FMM), a modernized variant of a technique long used in bioelectromagnetics (Makarov et al 2018, 2020). BEM-FMM accelerates calculations by leveraging surface integrals, thus avoiding the volumetric meshing required in FEM. This approach proves especially advantageous in high-resolution simulations or when computational resources are limited. Its use in TMS modelling has expanded in recent years due to improvements in numerical stability and toolkits tailored to neuroscience applications.

While FEM and BEM-FMM capture how electric fields distribute across cortical and subcortical structures, they do not directly address the neural response to stimulation. This gap has motivated the development of multiscale neuronal models that couple macroscopic field simulations with biophysically detailed neuron models (Aberra et al 2018, 2020, Weise et al 2023, Qi et al 2025b). These frameworks track the effects of TMS pulses on individual neurons—including action potential initiation and subthreshold membrane dynamics—and can be extended to study network-level recruitment. By resolving how fields interact with morphologically and directionally sensitive neural elements, multiscale models help bridge the divide between EM physics and neurophysiological outcome.

Recent work highlights how individualized E-field modelling is transforming TMS coil placement from a heuristic practice into a data-driven optimization problem. As reviewed by Dannhauer et al (2024), subject-specific modelling pipelines now support precise dosing by accounting for factors like scalp-cortex distance, gyral geometry, and coil orientation. These pipelines, often built on platforms such as SimNIBS and the targeting and analysis pipeline (Dannhauer et al 2022), integrate E-field simulations with structural and functional imaging to inform and verify coil placement. By leveraging EM reciprocity and auxiliary dipole methods, they enable efficient optimization without sacrificing physical realism. Empirical studies have demonstrated the clinical relevance of such approaches. In a study of depressed adolescents, Deng et al compared four common TMS targeting strategies: the 5 cm rule, Beam F3 method, MRI-guided targeting, and a computationally optimized approach. E-field modelling revealed substantial differences in target engagement, with the computational method delivering the strongest and most focal stimulation to the left dorsolateral prefrontal cortex (L-DLPFC). Moreover, among participants who completed treatment, median E-field magnitude at the L-DLPFC was linearly correlated with symptom improvement, underscoring the potential for model-informed dosing to enhance therapeutic outcomes (Deng et al 2023). These findings support broader integration of modelling pipelines into clinical workflows for personalized TMS.

Recent advances in real-time E-field modelling are pushing personalized TMS into the realm of dynamic, intra-session optimization. Traditional modelling approaches, though accurate, are computationally intensive and not practical for real-time clinical feedback. Addressing this gap, Li et al introduced a self-supervised deep learning framework capable of predicting high-resolution E-fields within seconds, bypassing the need for time-consuming partial differential equation solutions (2022). Trained on MRI-derived head models, the network achieved accuracies comparable to FEM solutions across realistic anatomical conditions, dramatically accelerating E-field estimation. Complementing this, Hasan et al developed a GPU-accelerated solver that integrates reciprocity and Huygens’ principles to compute E-fields in under 4 ms for full-head models (2025). By projecting coil-induced fields onto a reduced basis of cortical modes, their method maintains sub-3% error rates even at high spatial resolution. Importantly, these solutions are compatible with neuronavigation systems, enabling potential real-time updates to coil placement and intensity based on the evolving brain state. Together, these innovations demonstrate how modern computational strategies—ranging from physics-informed deep learning to high-efficiency reciprocal solvers—are reshaping the possibilities for adaptive, model-informed TMS delivery.

Concluding remarks

TMS has matured into a versatile platform for both clinical intervention and neuroscientific discovery. Yet, its full potential is only beginning to be realized. As advances in device engineering, pulse sequencing, and computational modelling converge, TMS is moving from a one-size-fits-all tool towards a precision-guided neuromodulation system. The integration of personalized electric field modelling—particularly with real-time capabilities—ushers in the possibility of adaptive TMS, where stimulation parameters are dynamically tailored to a patient’s unique brain anatomy, functional state, and clinical trajectory. Multiscale modelling continues to enhance our mechanistic understanding, connecting physics with physiology, while novel coil technologies and digitally steerable arrays promise unprecedented spatial and temporal control. In the coming years, the coupling of TMS with neuroimaging, electrophysiology, and machine learning may yield closed-loop stimulation paradigms that are both individualized and responsive, accelerating our capacity to treat complex brain disorders and to interrogate the neural substrates of cognition with causal precision.

Acknowledgments

ZDD and SNM are inventors of patents and patent applications on electrical and magnetic brain stimulation technology. ZDD is supported by the NIMH Intramural Research Program (ZIAMH002955). SNM is supported by the NIMH Grants R01MH130490 and NIBIB R01EB035484.

4. Thermal therapeutic techniques

4.1. RF and MW hyperthermia

Margarethus Marius Paulides1,2 and Dario B Rodrigues3

1 Care + CureLaboratory of the Electromagnetics Group (EM4Care + Cure), Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands

2 Department of Radiotherapy, Erasmus University Medical Center Cancer Institute, Rotterdam, The Netherlands

3 Department of Radiation Oncology, University of Maryland School of Medicine, Baltimore, MD, United States of America

Status

A. Background

Hyperthermia therapy (HT) has been long recognized as an effective adjuvant approach in cancer treatment. This therapy involves elevating tumour temperature to 39 °C–45 °C for 1 h, enhancing the efficacy of radiation therapy (RT), chemotherapy (ChT), or a combination of these modalities (Datta et al 2015). The radio-sensitization is mostly achieved via increased blood perfusion and oxygenation as well as inhibition of radiation-induced DNA damage repair (Elming et al 2019). The benefits from HT on CT relate to increased blood perfusion in tumours accommodated with an increased permeability of the tumour blood vessels, thus improving the leakage of drugs in the heated region (Issels 2008). Several studies also showed that hyperthermia boosts local and systemic immune response and enhance the effects of immunotherapies (Hurwitz 2019).

In this section, we will analyse dosimetry for HT by EM techniques involving RF and MW applicators, but the same principles are valid for conductive heating, ultrasound-based HT applicators or HT induced by nanoparticles (Kok et al 2020). HT can be applied invasively or noninvasively using externally applied power and RF/MW approaches can be further divided into local or regional HT. Local HT can be applied to both superficial and deep-seated tumours by external, intraluminal or interstitial MW methods. Regional hyperthermia is delivered using capacitive (8–27 MHz) or RF phased-array applicators (>65 MHz) to treat deep-seated pelvic or abdominal tumours. Capacitive systems generate broad, lens-shaped heating that is diffuse and strongly influenced by superficial fat, often leading to unwanted fat heating, while radiative systems provide a steerable, ellipsoidal focus in deeper tissues that enables more selective tumour heating. Despite these differences in heating pattern, both strategies can achieve effective regional heating (Ohguri et al 2018, Kok et al 2020). The key difference between local and regional HT approaches lies in the operating frequency. Localized heating is typically achieved with MW frequencies above 300 MHz, which allow for focal energy deposition but are limited by reduced penetration depth in biological tissues. In contrast, deep regional heating relies on RF energy below 300 MHz. For MW-based local HT, phased-array systems with a sufficient number of independently controlled channels are required to compensate for the shallow penetration and to achieve adequate focusing (Rodrigues et al 2022).

In the following sections, we will provide an overview of key HT dosimetric parameters, discuss current methodologies used for their assessment, and highlight the state-of-the-art techniques along with the main challenges faced in HT dosimetry.

B. Hyperthermia dosimetric parameters

A clear-cut relationship between thermometric parameters with treatment outcome is highly attractive because it would improve our understanding of tumour-specific mechanisms of interaction between HT and RT, CT and immunotherapy. HT dosimetric parameters can be divided in terms of the induced EM electric field component (E, units V m−1), resulting temperature (T, units °C) from the EM power absorption (SAR, units W kg−1), and thermal dose, a parameter that considers both temperature and heating time. SAR-based parameters include maximum SAR (SARmax) and SAR50, corresponding to the 50th percentile SAR distribution in tissue or phantom. Temperature-based parameters include maximum (Tmax), minimum (Tmin), and average Tave target temperature over the treatment duration, as well as T90, T50, and T10, with Ti representing the ith percentile temperature distribution in the target.

Since thermal dose distributions within a tumour volume are heterogenous, the calculation of tumour dose involves the thermal dose accumulated at each point with T90, resulting in CEM43T90 (Dewhirst et al 2003, Dewey 2009). CEM43 parameters were derived from cell death studies as a function of temperature, but they do not include the radio- or chemo-sensitization effect. Nonetheless, there is a strong correlation between thermal dose and clinical outcomes as has been shown by several clinical studies (Sherar et al 1997, Lee et al 1998, Ohguri et al 2018, Kroesen et al 2019). Thus, while it does not capture all hyperthermic effects, CEM43 has been identified as the an important measure of the treatment efficacy when combining radiation with HT to treat cancer (van Rhoon 2016). This correlative effect is expected in HT + ChT treatments, but not yet validated. Note that all these dosimetric parameters are relevant for treatment efficacy, but many are also used to assess treatment safety, for which Tmax in healthy tissue is often used (Bakker et al 2018). Finally, the biological equivalent dose is another dosimetric parameter currently under investigation to capture the combined effect of radiation and hyperthermia treatments (Kok et al 2022, 2023).

Challenges and future directions

The efficacy of HT combined with RT, ChT and/or immunotherapy has been explored across various tumour types. Comparing thermal dosimetry is a challenge, because consensus on HT delivery and reporting of dosimetric parameters is limited leading to substantial heterogeneity of HT treatment protocols and clinical results (Ademaj et al 2022). For instance, tumour Tmin was a prognostic factor in a few studies (Cox and Kapp 1992) and another study showed that T90, T50, and T10 in the target volume were more strongly correlated with cancer response than tumour Tmin (Leopold et al 1992). Moreover, a short time interval between HT and RT was shown to significantly predict treatment outcome in retrospective analyses of cervical cancer patients (van Leeuwen et al 2017). Note that conflicting results have been also reported (Kroesen et al 2019), which may be attributed to differences in time interval and tumour temperature achieved, and in the patient population included (Crezee et al 2020). Thermal dose was successfully tested in several clinical trials as predictor of RT + HT tumour response (Jones et al 2005, Franckena et al 2009, Bakker et al 2019, Ademaj et al 2022).

The thermal dose parameters used to date did not lead to established thresholds for different cancer sites, even though the European Society for Hyperthermic Oncology (ESHO) guidelines recommend superficial HT treatment maintain T50 ⩾ 41 °C and T90 ⩾ 40 °C (Dobšíček Trefná et al 2017). Similarly, safety thresholds are yet to be defined. However, several reports indicate that if temperatures reach levels above 44 °C–45 °C, patients can experience pain, formation of oedema, and thermal blistering for local applications and solid organ damage for regional therapies (Linthorst et al 2015, Longo et al 2016). Temperatures within 43 °C–44 °C may be safe for short periods of time if sufficient thermometry is provided to maximize tumour monitoring coverage (Bakker et al 2018).

A. Probe-based thermometry

Invasive or intraluminal temperature measurements performed directly within the tumour are still regarded by many clinicians to be the gold standard for monitoring HT treatments. Temperature probes used to monitor HT treatment include thermistors, optical fibre probes and thermocouples (Schooneveldt et al 2016, Dobšíček Trefná et al 2017). These are inserted into preplaced closed-tip catheters and can be arranged in multi-sensor probes, or cyclically pulled though the catheter, to increase spatial resolution. Thermocouples include a metal component that makes them sensitive to EM fields and requires power to be momentarily stopped (few seconds) to reduce the effects of self-heating (De Leeuw et al 1993, Bakker et al 2020). In superficial HT, probes are positioned on the surface and sometimes invasive probes are used to measure temperature at depth. In deep HT, temperature is measured at the surface (for control) and via probes inserted into closed-tip catheters that are placed in the rectum, vagina and bladder. Only occasionally, due to the risk of bleeding and tumour spreading, interstitial catheters are inserted directly in the tumour for direct temperature readings. Note that interstitial HT uses brachytherapy catheters for both placement of MW antennas and temperature probes so there is no need to add extra invasive measurements.

The main challenge with point-measurement approaches is the limited number of monitoring positions, which result in under sampling of both tumour and healthy tissue temperatures (Bakker et al 2018). More frequent insertion of intratumoral probes, on the other hand, will restrict the number of patients being treated with HT due to potential complications associated with the invasive procedure. Hence, a noninvasive and 3D monitoring approach is thus warranted.

B. MR thermometry (MRT)

Noninvasive MRT, or MR thermal imaging is emerging as an alternative to invasive and point measurements (Winter et al 2016). Most commonly, the shift of the proton resonance frequency (PRF) is used due to its linearity with temperature and independence of tissue type (Rieke and Butts Pauly 2008). In this approach, baseline MR images are acquired before treatment to map the PRF. During HT, the PRF is measured at set time-intervals, which is subtracted from the baseline PRF map to monitor the local temperature change in each voxel by the PRF shift. Despite few companies providing MR-guided thermal therapy treatments at ablative temperatures, there is only one device for MR-guided hyperthermia commercially available for deep-regional HT treatments (Curto et al 2019).

As the PRF shift with temperature is a very small effect, measurement noise is severe and other variables like patient or physiological motion (cardiac or respiratory) create inaccuracies and/or severe artefacts (Feddersen et al 2020). Advanced algorithms and techniques are employed to correct these inaccuracies and artefacts. For example, while the PRF shift provides temperature change in water-rich tissues, the limited PRF shift in low-water tissues like fat can be used to correct the images for other sources of PRF changes (Baron et al 2014), e.g. motion effects but also MRI scanner drift (Bing et al 2016). Still, these techniques are mostly at the research phase (VilasBoas-Ribeiro et al 2021).

C. Hyperthermia treatment modelling

Pre-treatment planning by HT modelling involves using simulation tools to optimize the dose distribution before treatment (Paulides et al 2013). First, imaging slices are used to segment tissue regions, which is often a semi-automatic process (Ribeiro et al 2018). Second, each region is assigned EM and thermal tissue properties to obtain 3D EM and thermal representations of the patient. Third, EM field simulation is conducted to predict how the energy will be distributed within the tissues for each antenna, which will be used to calculate 3D SAR from multiple antennas in step four. Following this, temperature calculations can be made to estimate the tissue’s thermal response to the applied energy. This workflow is illustrated in figure 2. Pre-treatment planning is generally used for phased-array applicators, for which phase-amplitude fine-tuning of the signals applied to the antennas is carried out to achieve maximum heating, i.e. predicted SAR or temperature, while minimizing the impact on surrounding healthy tissues (Paulides et al 2021).

Figure 2.

Figure 2.

Schematic representation of the hyperthermia treatment planning workflow, shown for a head-and-neck applicator (MRcollar) operating at 434 MHz (Paulides et al 2018). Adapted from Rodrigues et al 2022, Copyright (2022), with permission from Elsevier.

One of the main challenges in patient-specific HT modelling is the unknown EM properties of the patient tissues, so literature-based values are assigned for all cases. This implies that tissue homogeneity is assumed within normal and tumour tissue, and that there are no heating-induced changes, both of which are oversimplifications. Another challenge is that accurate patient positioning is required for predictive simulations, but recent studies found that patient’s position and anatomy during imaging and treatment differ up to 2 cm (VilasBoas-Ribeiro et al 2023) under MR guidance, and likely even more without it. Moreover, thermal tissue properties are temperature and blood perfusion dependent, and the body counteracts tissue heating using strong, local (immediate) and global (after approximately 15 min for significant heat loads) thermoregulation responses. Consequently, temperature-based models currently involve substantial uncertainties. Therefore, although pre-treatment planning is advised in ESHO protocols (Bruggmoser et al 2011), these uncertainties have limited its use in wide-scale routine clinical practice. Besides temperature, modelling the biological effects of HT combined with other modalities is crucial to understand how tissues respond to different levels of exposure (Kok et al 2022). However, this type of modelling has even greater uncertainties and only provides indicative values. Finally, the high variability and uncertainty of patient tissue properties result in marked differences between HT and radiotherapy dosimetry. Due to uncertainties in both tissue parameters and thermoregulatory responses, patient-specific thermal dosimetry can achieve adequate accuracy only when temperature measurements are incorporated to calibrate the simulations.

D. Phantom dosimetry

Phantom dosimetry plays a major role in assuring that HT applicators function correctly, as well as for experimental validation of novel heating and monitoring approaches. Moreover, phantom dosimetry is crucial for comprehensive assessment of the applicator modelling implementation, both at installation and periodically thereafter, in view of HT modelling. To quickly assess the heating symmetry and steering abilities of phased-array applicators, quality assurance procedures use light-emitting diode (LED) or lamp phantoms with cylindrical or elliptical shape (Schneider and Van Dijk 1991, Wust et al 1994, Aklan et al 2019). These consist of a LED diode or lamp matrix immersed in a saline solution contained in a cylindrical/elliptical plastic shell. The saline is calibrated to emulate tissue electrical conductivity enabling the LEDs/lamps to respond to the generated EM field. The fast assessments can be effectively carried out qualitatively by using LED/Lamp phantoms or quantitatively either by using the Schottky diode sheet (Van Rhoon et al 2003) or even with a tissue-mimicking gel phantom with multiple thermal mapping probes covering axial and longitudinal measurements in catheters.

An important challenge impeding widespread application of hyperthermia treatment modelling, for both pre-planning and real-time treatment guidance, is the current lack of rigorous device performance verification. The limited commercial solutions to measure the heating delivered by the devices generally fail in terms of the quantitative SAR assessment required to assure simulation accuracy.

Scientific and technological advances addressing the challenges

MRT is expected to become a crucial technique in hyperthermia treatment research and possibly also in the clinical routine, as new advances enable it to ensure heating as well as a sound validation of treatment modelling. Hereto, new MR compatible MW devices are under development that enable more precise focus steering possibilities and can contain integrated coils to reduce the noise of the imaging, leading to better motion correction possibilities (Drizdal et al 2021, Sumser et al 2021). In addition, new correction approaches are under development with various (Kok et al 2022) levels of complexity, addressing challenges such as motion artefacts and improving the reliability of temperature measurements (Nouwens et al 2022). Additionally, MR enables improved tissue modelling, allowing for more accurate predictions of thermal responses and better treatment modelling. This treatment modelling in turn enables hybrid thermal assessment for filling in the gaps of imaging in both space and time (VilasBoas-Ribeiro et al 2022). Together, these innovations are set to significantly advance the field of HT.

Adaptive treatment modelling allows for adjustments to be made during treatment based on real-time data like from temperature sensors or patient complaints, i.e. the patient generally senses overheating normal tissue through pain before the heating leads to (severe) side effects. Adaptive HT based on model guidance is currently used routinely in a few specialized centres. Given that relative agreement between model and measurements are already adequate, this approach is already providing clinical benefits (Franckena et al 2010, Kok et al 2014). In addition, as patient positioning and posture control is improved and our knowledge on EM and thermal tissue properties expands, robust planning and re-planning approaches can be exploited. Furthermore, the treatment model also enables us to include real-time supportive measurements from sources like E-field sensors or flow sensors that are placed inside the device. This helps to enhance the predictive value of the model, which is expected in turn to translate into improved benefits of adaptive HT.

Generally, it is assumed that the developments in simulation and MR-guided treatment will reinforce each other since MRT forms the tool required to validate treatment planning, patient positioning and device performance in 3D (VilasBoas-Ribeiro et al 2024). Note that validated treatment planning simulations not only provide the tool for guiding treatment, but also provides the tool to design better applicators, create treatment and quality assurance guidelines, conduct superior training, develop new treatment approaches, and provide more robust thermal dose assessments of patient treatments (Paulides et al 2021). The latter are crucial to establish correlations between thermal dose and treatment outcome or toxicity per indication and per multi-modality treatment approach.

Concluding remarks

In summary, clinical hyperthermia is advancing through improvements in dosimetry, real-time monitoring, and adaptive modelling. The integration of MRT with computational simulation tools is essential for enhancing both treatment efficacy and patient safety. Ongoing research and international standardization efforts will be critical for establishing robust, reproducible treatment protocols and for supporting broader clinical adoption.

Acknowledgements

MMP was financially supported by Eurostars Project 3868 (SENS-THERM) and Project P21-34 (CARES) of the research programme Perspectief, which is financed by the Dutch Research Council (NWO).

4.2. RF and MW ablation

Christopher L Brace1, Marta Cavagnaro2 and Punit Prakash3

1 Departments of Radiology and Biomedical Engineering, University of Wisconsin-Madison, Madison, WI, United States of America

2 Department of Information Engineering, Electronics, and Telecommunications, Sapienza University of Rome, Rome, Italy

3 Department of Biomedical Engineering, The George Washington University, Washington, DC, United States of America

Status

A. Background

Thermal ablation is a medical treatment where power from an external source is delivered to tissue via an applicator. The applied power is locally absorbed leading to increased tissue temperatures and tissue destruction. Thermal ablation systems are used for various applications in oncology, cardiac electrophysiology, dermatology, gastroenterology, neurosurgery, and others. In this section of the roadmap article, we will restrict our attention to thermal ablation employing RF (∼500 kHz) and MW (∼500 MHz–10 GHz) energy.

The temperature field created during an ablation procedure is a function of the EM power loss density around the applicator and factors contributing to bioheat transfer, notably: thermal conduction; microvascular blood perfusion and heat transfer due to flow in discrete vessels; active cooling of ablation applicators; and others. Advances in experimental and computational assessment of transient temperature profiles have contributed to the development and advancement of thermal ablation devices and systems, and have potential to inform treatment delivery strategies in a variety of tissue types.

As previously introduced, thermal ablation employing radiofrequency (RFA) and MW (MWA) energy use two different frequency ranges. This difference translates into a different mechanism of energy release, which in turn leads to different possible outcomes of the two techniques.

During RFA, an electric current is released by an electrode into the tissue to be treated. While propagating into the tissue, the current is dissipated through the Joule effect: the electric power lost by the current is transformed into heat that increases the tissue temperature. At about 500 kHz, the wavelength is much larger than the electrode dimension or the tissue to be treated (e.g. in muscle the wavelength is about 10 m at 500 kHz). Accordingly, the electrode can be modelled as a point source, and the current follows radial paths while moving away from the electrode. This means that the current spreads out on concentric spheres, i.e. the current density amplitude decreases according to a law inversely proportional to the square of the distance from the electrode. Given that the heat is proportional to the dissipated power, which in turn is proportional to the square of the current density, the heat source into the tissue has an amplitude that depends on the distance from the electrode (r) as 1/r4. The direct consequence of this dependence is that the electric current directly heats (thermally ablates) only a few mm of tissue around the electrode, while the overall RF ablation zone is mainly achieved by thermal conduction, which is a slow phenomenon, easily impeded by blood flow (Huang 1995).

Still related to the physics of RFA, electric current flows through tissues with higher water and ionic content. However, if the tissue’s temperature reaches about 100 °C, water vaporization and tissue desiccation increase the circuit impedance, causing electric power to drop (Chu and Dupuy 2014). Expandable electrodes, power cycling, and saline infusion are common techniques to help distribute the electrical energy and control heating by RFA devices.

At the frequencies of MWA, an EMF is radiated by an antenna within the MWA applicator. The most commonly used frequencies for MWA in clinical use are 2.45 GHz and 915 MHz, though systems operating at higher frequencies have also been explored (Jones et al 2012, Curto et al 2015, Bottiglieri et al 2022). In contrast to the electric current in RFA, the EMF propagates through dielectric materials of any loss factor so that thermal effects such as tissue desiccation are not an impediment to continued MWA treatment. While propagating, the EMF dissipates due to losses into the tissue; the lost power is transformed into heat. In MWA, the temperature increase is directly proportional to the EM power dissipation, roughly decreasing according to a law inversely proportional to the square of the distance from the antenna. Accordingly, the thermally ablated area is mainly due to the EM dissipation, a phenomenon much faster than thermal conduction (Brace 2009). We note that the performance of the different devices in terms of shape and dimension of the thermally ablated zones depend also on the electrode (RFA) or antenna (MWA) design. A full analysis of the different applicators is well beyond the scope of this manuscript, and has been reviewed elsewhere (Kok et al 2020). To this end, we focused our attention on the energy-tissue interaction physics and elements of dosimetry needed to assess an arbitrary device under varying tissue/clinical conditions.

B. Dosimetric parameters and quantities of interest

Section 1.2 presents the pertinent physics and equations for characterizing time-averaged EMFs, power absorption, and bioheat transfer during RFA and MWA. The biological effects of thermal therapy are a function of the time-temperature history during heating (Pearce 2013). In vitro studies of thermal exposure to cells in culture to identify the kinetics of thermal injury have demonstrated that the time taken to observe a bioeffect (e.g. cell death) decreases with increasing temperature. Cell death occurs rapidly at temperatures above ∼55 °C, and as such threshold temperatures in the range ∼50 °C–60 °C are widely used to approximate the extent of cell death. Thus, the primary quantity of interest for dosimetry of ablation system is the transient temperature profile, T(r, t) (°C) in tissue. Tissue temperature can be measured directly with a variety of invasive temperature sensors (e.g. thermocouples, thermistors, fibreoptic sensors), or estimated via medical imaging techniques (Zaltieri et al 2021).

The time-averaged EM power loss density profile Q(r) (W m−3) is another quantity of interest as it represents the locally absorbed power in tissue and provides insight into the region of tissue where the applied power directly affects tissue.

Q(r)=ρ(r)SAR(r).

As previously reported, the SAR (SAR(r)) represents the power absorbed in tissue per unit mass, and it is typically used to represent the EM heat source into the bio-heat equation (see section 1.2). It can be evaluated from measurements of the electric field once the conductivity and density are known, or from the transient temperature profile, when measured using very short time durations in phantoms (Deshazer et al 2017).

One of the main difficulties in using these equations is the high dose rate during RFA and MWA. In contrast to HT, thermal ablation is of relatively short duration, typically on the order of ∼5–15 min for oncologic applications, and often less for other applications where the treatment zone is small. In this time temperatures can exceed 100 °C, or even much higher during MWA. Therefore, the dosimetry problem is largely focused on achieving an ablative threshold exceeding ∼50 °C throughout the targeted tissue, while limiting temperature in non-targeted tissue beneath safety thresholds.

Such dynamic heating also presents a dosimetry problem in the assumptions about EM and thermal parameters. All tissue properties depend on the tissue temperature, so the EM and thermal problems become interrelated. However, knowledge about the behaviour of tissues at ablative temperatures (up to about 120 °C) is still quite scattered. Thermal dosimetry has been important for guiding studies aimed at investigating how tissue physical properties and state (e.g. dielectric properties, thermal properties) vary as function of temperature and rate of heating (Ji and Brace 2011, Lopresto et al 2012, 2019, Deás Yero et al 2018, Etoz and Brace 2019, Vidjak et al 2024). Improved accounting of these temperature dependencies may enable more accurate dosimetry predictions in future studies.

C. Routine clinical use

In routine clinical use, careful thermal and/or EM dosimetry are rarely used intra-procedurally. During RF ablation procedures, electrode tip temperature is often monitored with temperature sensors incorporated into the applicator, though these measurements are not directly indicative of extent of the ablation zone. For tumour ablation procedures, the outcome is typically assessed with contrast imaging to assess non-enhancing regions, indicative of tissue regions devoid of perfusion following the ablation (Lin et al 2024). Recently, image analysis tools have been developed and are in clinical use for assessing the extent of the ablation zone relative to the tumour boundary prior to ablation, to facilitate estimation of treatment margins (Laimer et al 2025). Several clinical systems accommodate the use of auxiliary temperature sensors that clinicians may choose to position at regions at risk to monitor for thermal damage or insufficient treatment.

Volumetric thermometry with MRI during RFA and MWA has been demonstrated in the clinical setting and has potential to provide transient temperature profiles in multiple planes (Lepetit-Coiffé et al 2010, Öcal et al 2024). However, this technique is not routinely used due to technical complexity of integrating ablation systems within the MRI environment, the considerable added cost of conducting clinical procedures in the MRI suite, and the potential for tissue motion and water content changes to corrupt the thermographic data. This remains an area of ongoing research and development.

Careful thermal dosimetry has played a key role in guiding the development of ablation applicators and energy delivery strategies. For example, several studies over the last three decades measured temperature at select distances from ablation applicators to develop insight into how various device and/or energy delivery parameters affected the overall treatment zone (Yang et al 2007, Cavagnaro et al 2011, Andreano and Brace 2013, Curto et al 2015, Namakshenas et al 2024). Thermal dosimetry has also been important for understanding how the physical quantity that ablation procedures ultimately modify (i.e. temperature) relate to quantities measured/observed in the clinical setting (e.g. imaging appearance of tissue exposed to ablation with/without contrast) (Strigari et al 2019). In each case, limitations in spatial or temporal resolution, or a limited range of tissue types and/or physiological variables such as perfusion, impacted their extrapolation into broader clinical utility.

Challenges/opportunities

Several challenges persist in verifying adequate thermal coverage of the target zones across clinical indications.

While most device manufacturers provide guidelines for expected ablation size compared to power and delivery time, they are often characterized in ex vivo, unperfused, animal tissue (Hoffmann et al 2013). Ex vivo ablations are not necessarily representative of ablation zones achieved in the clinical setting due to the influence of blood perfusion heat sinks, organ boundaries, and heterogeneity in tissue physical characteristics as a function of disease state (Amabile et al 2017). Accordingly, there remains a gap between the expected and realized treatment effects for similar equipment settings.

Computational approaches for estimating tissue temperature profiles during ablation are under investigation for potential use in planning treatments and/or assessing treatment outcomes. Physics-based models that employ numerical approaches to solve the differential equations pertinent to EM absorption and bioheat transfer have been applied towards predicting extents of the treatment zone, as estimated from computed transient temperature profiles. Recent efforts in the field have integrated these models with pre- and peri-procedural imaging data to inform model geometry and facilitate model predictions of treatment zone overlaid with imaging. A few clinical studies evaluating the technical feasibility and performance of these computational approaches have been reported (Moche et al 2020, Hoffer et al 2024).

Real-time or immediate post-treatment feedback from thermal sensors or medical imaging can help ensure that the proper thermal dose was used for a full treatment effect, but this approach may not be practical for all clinical situations. Invasive temperature probes can provide continuous feedback with limited spatial resolution. In applications where the ablation zone is relatively small (⩽1 cm maximum dimension), such limited spatial resolution may be sufficient; however, for large ablations (⩾3 cm maximum dimension) such pointwise data cannot give reliable feedback about the entire treatment volume (Zaltieri et al 2021). Local heterogeneities in tissue perfusion, structure or other properties can lead to irregular ablation shapes that deviate from expectations. Substantial variation is expected between patients as well (Mathy et al 2024). Imaging-based dosimetry can fill some of the gaps to provide more full-volume thermal data, but at the cost of reduced temporal resolution and generally greater noise figures than direct thermal profiling. Recent studies have explored alternative solutions to determine noninvasively temperature distributions in real time, such as the use of MW imaging (Wang et al 2022b) or analysing tissue dielectric property changes from the antenna applicator (Wang and Brace 2012, Vidjak et al 2023).

Concluding remarks

The thermal profile induced in tissue during thermal ablation can vary considerably across procedures using the same applied energy settings, potentially contributing to the variability in treatment outcomes. Currently, there are few practical means for thermal dosimetry throughout targeted tissue volumes in routine clinical use. Computational and experimental approaches for estimating volumetric thermal profiles are the subject of active research investigation and have potential to contribute to improved outcomes of thermal ablation procedures.

Acknowledgments

CB was financially supported by NIH R01AR082375. PP was financially supported by National Institutes of Biomedical Imaging and Bioengineering, grant no. R01EB028848.

5. RF-based modalities for diagnosis and sensing

5.1. RF dosimetry in ultra-high field (UHF) parallel transmit (PTx) MRI

Desmond T B Yeo

GE HealthCare, Technology and Innovation Center, Niskayuna, NY, United States of America

Status

Due to its superior soft tissue contrast, diverse functional and structural information content, and absence of ionizing radiation, MRI has had significant impact in the management of multiple diseases and disorders. MRI requires the application of i) a strong static magnetic field (B0) to create non-zero net magnetization vectors, (ii) circularly polarized RF transmit magnetic fields (B1+) at the Larmor frequency (f0) to nutate spins and (iii) gradient fields for spatial localization. While high B0 fields may induce reversible physiological effects (Schenck 2000), and strong, fast-switching gradient fields may cause peripheral nerve stimulation (PNS) (Mansfield and Harvey 1993) and high sound pressure levels, thermal risks in an MRI scan arise primarily from RF-induced concomitant and conservative electric fields (E-fields). The increasing proliferation of very high B0 field MRI systems (⩾5 T) with multichannel RF transmit capabilities warrants increased rigor in global/ local RF heating assessment.

As B0 increases, MRI signal-to-noise ratio increases, which improves image quality and/or spatial resolution. Since the Larmor frequency, f0, scales linearly with B0, higher-frequency RF transmit magnetic fields are needed to excite spins as B0 increases. As f0 increases, the wavelengths of EM transmit fields decrease and approach the dimensions of body structures, which induce inhomogeneous B1+ fields that lead to image shading (Röschmann 1987, Bernstein et al 2006) that can confound diagnosis. To address this, multichannel RF transmit architectures and imaging techniques have been proposed, i.e. RF shimming (Hoult 2000, Ibrahim et al 2000) and PTx (Katscher et al 2003, Zhu 2004).

The goal of RF shimming (figure 3(b)) is to create a homogeneous B1+ field by linearly combining B1+ fields generated by N transmit elements. Because RF pulse envelopes in all channels are similar in RF shimming, the composite B1+ and E-field/ local SAR distributions do not change appreciably during RF excitation, and throughout an exam for a given set of complex weights. Conversely, in PTx MRI (figure 3(c)), each transmit channel’s signal is independent of other channels. Instead of a static, amplitude-weighted B1+ distribution, PTx RF pulses produce time-varying B1+ distributions, which induces time-varying concomitant E-field/ local SAR distributions. While the additional degrees of freedom afforded by PTx MRI can potentially reduce local SAR, the absence of accurate personalized information can increase risks of tissue heating instead (Zhu 2004).

Figure 3.

Figure 3.

Three RF transmit architectures used in MRI with increasing uncertainties in assessing RF-induced local hot spots: (a) quadrature-driven, (b) RF shimming and (c) parallel transmit.

In 1.5 T and 3 T MRI scanners that utilize quadrature-driven RF transmit volume coils (figure 3(a)), RF power deposition in the exposed mass is predicted prior to a scan using knowledge of the RF pulse envelope and patient weight. During the scan, global average SAR in the exposed mass is then measured with power monitors in the RF transmit chain, typically in tandem with pre-characterized information that account for coil loss, exposed mass, etc. The predicted and measured power deposition rates are time-averaged (6 min and 10 s) and assessed against whole body and/or partial body global average SAR limits stipulated in IEC 60601-2-33. However, for a multichannel transmit coil, both global and local SAR limits may be pertinent (IEC 60601-2-33 2022, Fiedler et al 2025), and the same limits apply to both RF shimming and PTx MRI. Compared to quadrature-driven coils, the higher uncertainty and potential severity of local SAR hot spots in PTx MRI present a more challenging RF safety problem, which often leads to implementation of high local SAR safety margins at the expense of imaging performance, e.g. worst-case local SAR predictions (Neufeld et al 2011). Conversely, a framework that consistently under-predicts local and global SAR would be unacceptable for patient safety. In PTx MRI, it is desirable to develop frameworks that minimize over-prediction of SAR, with safeguards that prevent under-prediction. In the long term, since thermal dose and thermal damage thresholds are more predictive of heat-induced tissue damage (Yarmolenko et al 2011, van Rhoon et al 2013), frameworks that measure subject-specific local temperature rise and thermal dose over time are expected to increase confidence when assessing RF heating risks in UHF PTx systems.

Current and future challenges

To widen adoption of UHF PTx MRI, it is imperative to increase confidence in the validity of SAR/thermal risk assessment methodologies. Figure 4 depicts a forward-looking, personalized local SAR/thermal dosimetry workflow for PTx MRI that incorporates individual functional modules that have been studied for local SAR management in PTx MRI (Homann et al 2011, Graesslin et al 2012, Boulant et al 2016). To render this workflow feasible for routine use in PTx MRI though, advancements in the speed, precision and accuracy of these functional modules are critical. Specifically, the key requirements can be categorized as follows: (i) accurate PTx coil array EM model validation (particularly for E-fields), (ii) accurate personalized EM/ thermal modelling to minimize prediction errors, (iii) real-time personalized local SAR prediction/ measurement and (iv) real-time thermal dose measurement.

Figure 4.

Figure 4.

Forward-looking framework for personalized RF dosimetry in PTx MRI.

The validation of EM coil models in inhomogeneous tissue is limited by the lack of direct high-resolution E-field measurements, often necessitating indirect approaches such as phantom-based MRT (Ishihara et al 1995, Poorter et al 1995, Rieke and Butts Pauly 2008).

In UHF PTx, a safety margin derived from EM modelling with generic human body models (HBMs) is commonly applied to over-predict local SAR (Neufeld et al 2011). This is because it is very challenging to accurately replicate the salient input parameters for EM simulations pertaining to the subject, RF transmit coil, landmark position, etc. To create a personalized HBM, whole or partial body MRI datasets with appropriate contrast types need to be acquired efficiently prior to the actual PTx MRI scan (e.g. a UHF scan with derated RF power or a non-UHF MRI scan), followed by accurate and fast tissue segmentation. This entire process can present workflow and computational challenges. MR-EPT can potentially provide personalized estimates of electrical properties but improvements in robustness and computational efficiency are needed (Katscher and van den Berg 2017, Leijsen et al 2021). Subsequently, after the HBM is digitally inserted into a validated PTx coil EM model, fast EM solvers are needed to generate local SAR/Q matrices (Bardati et al 1995), which can be applied to PTx RF pulse design algorithms with local/global SAR constraints (Graesslin et al 2012, Boulant et al 2016). Techniques that compress the set of Q matrices can significantly reduce computation time in local SAR-constrained RF pulse design but higher levels of compression are often accompanied by greater over-prediction of local SAR (Eichfelder and Gebhardt 2011). Solving Pennes’ BHTE (see section 1.2) across transmit channel combinations yields a Q matrix-equivalent for temperature rise (Das et al 1999), useful for predicting heating with different PTx waveforms. However, accurate in vivo temperature prediction requires subject-specific thermal property maps (e.g. perfusion, conductivity) and modelling of nonlinear thermoregulation, both of which remain challenging (Murbach et al 2014).

In transitioning from pre-scan local SAR prediction to intra-scan local SAR measurement, the problem becomes more challenging as MRI cannot directly measure E-field distributions. Even if personalized local SAR/ temperature predictions are accurate, they do not account for in-scan events like transmit channel failures, which can produce E-field distributions that differ from predictions. A local SAR monitoring approach might consist of channel-specific sniffer coils that partially measure each channel’s transmit field information, which is then applied to precomputed, simulated E-fields derived from individually excited channels that irradiate a personalized HBM. This indirect monitoring of local SAR will need to account for uncertainties in the field measurement process. Volumetric MRT (Ishihara et al 1995, Poorter et al 1995, Rieke and Butts Pauly 2008) is a promising monitoring solution that can measure temperature and thermal dose maps during a scan, which may obviate the reliance on personalized HBMs or sniffer coils. However, it is challenging to implement fast, motion robust, volumetric MRT techniques that are accurate and precise enough to detect tissue heating across a wide field of view (FOV) in real-time.

Advances in science and technology to meet challenges

Advances that lead to accurate and precise estimates of subject-specific parameters for EM/ thermal modelling can reduce mismatches between actual scans and simulation results. Fast segmentation of multi-contrast MRI datasets, combined with atlas-based nonlinear registration, can efficiently generate personalized HBMs for local SAR prediction (Cabezas et al 2011, Brink et al 2022). In addition, these segmentation approaches can be used to generate large numbers of realistic, non-subject-specific HBMs with diverse characteristics that impact local/ global SAR risks. Data-driven algorithms that learn salient features from tissue-segmented imaging datasets may conceivably be developed to generate synthetic HBMs to augment existing repositories. If widely accessible and thoroughly validated by experts, the repositories can help reduce uncertainties when determining safety margins for a broad population.

Direct MR EPT methods that estimate electrical property maps from measured transmit/ receive fields are computationally less expensive than forward methods (Leijsen et al 2021). However, if forward methods, which require tissue-segmented data and EM numerical computation, can be accelerated sufficiently, they can potentially produce more robust estimates of electrical property maps. Another approach, image-based EPT, reconstructs electrical properties directly from MR images, which obviates the need for transmit/receive fields (Leijsen et al 2021), i.e. reduces scan time and increases SNR. Emerging data-driven EPT approaches have also shown promise in estimating electrical property maps that are more robust to noise (Mandija et al 2019, Leijsen et al 2022).

The use of accurate personalized HBMs and electrical properties within a carefully validated PTx coil model will increase confidence in the validity of output from modules in figure 4. To minimize the odds of under-predicting local SAR, formulations like the virtual observation points that compress Q matrices (derived from personalized HBMs) can provide upper limits for fast local SAR estimates that can be adjusted to trade off the degree of compression and the level of local SAR over-prediction allowed (Eichfelder and Gebhardt 2011). Alternatively, an efficient PTx RF pulse optimization algorithm that can operate on uncompressed Q matrices within a practical timeframe can be employed to minimize over-prediction and obviate the compression process (Pendse et al 2019). It has also been shown that fast local SAR prediction with low over-prediction errors is feasible with supervised deep learning networks trained with B1+ and local SAR maps (Meliadò et al 2020, Gokyar et al 2023).

Volumetric thermal dose measurements require fast, motion-robust MRT techniques. With proton frequency resonance MRT (Poorter et al 1995), motion-induced B0 perturbations can create non-temperature-dependent phase accruals that result in MRT errors. These increase in severity as B0 increases. A recent study that utilized motion-compensated MRT demonstrated the feasibility of sub-degree accuracy when monitoring RF-induced heating in the brain at 7 T (Le Ster et al 2022). Other motion robust MRT approaches that are applied to thermotherapy applications can also be assessed for online thermal dosimetry (Kim et al 2024). To overcome the FOV limitations in MRT for RF thermal dose assessment, it is conceivable that carefully calibrated, hybrid measurement-simulation frameworks that combine local thermal dose measurements and EM/thermal simulation results may ameliorate the shortcomings of both approaches.

Concluding remarks

The assessment of RF heating risks in UHF PTx MRI is confounded by multiple sources of uncertainties. More accessible, standardized phantoms for EM/thermal validation, fast/accurate tissue segmentation and EPT techniques, and fast EM/thermal computation solvers are important drivers that can reduce these uncertainties. When integrated within an efficient processing pipeline, the workflow should be rigorously optimized for online personalized local RF dosimetry. In the long term, accurate MRT techniques that directly measure temperature rise/thermal dose will provide the most direct risk assessment of RF heat-induced tissue damage. Advancement in these areas will facilitate safe UHF PTx MRI scans, and hopefully, deliver benefit to more patients.

5.2. MMW sensing

Takuya Sakamoto1 and Francesco Fioranelli2

1 Department of Electrical Engineering, Kyoto University, Kyoto, Japan

2 Department of Microelectronics, Delft University of Technology, Delft, The Netherlands

Status

The use of MWs and MMW waves for human body measurements has attracted significant attention, with many research groups actively investigating radar-based, noncontact human sensing (Fioranelli et al 2019, Gurbuz and Amin 2019, Le Kernec et al 2019, Paterniani et al 2023). These technologies offer important advantages over more conventional contact sensors, such as wearable devices, by reducing patient discomfort and restrictions. Moreover, fewer privacy concerns are raised than those observed in widely used camera-based systems. Radar systems for human measurements can address not only physical quantities, such as distance, angle, position, and velocity, which are the main targets of classical radar systems, but also human-specific shapes and movements, including posture, limb motion, gesture, and activity (Seyfioğlu et al 2018, Chen et al 2021, Gorji et al 2021, Zhao et al 2022, Zhu et al 2022, Kruse et al 2023). These capabilities open promising applications in fields such as healthcare and entertainment, with specific use cases including in-vehicle or outdoor person presence detection, activity and gesture recognition, fall detection, security screening, and gait analysis (Wang et al 2014, 2019, Seifert et al 2019, Abedi et al 2022, 2023, Gharamohammadi et al 2023, He et al 2024).

Radar systems can also detect small-scale body movements associated with physiological processes, including respiration, heartbeats, and pulse waves (Hong et al 2019, Piriyajitakonkij et al 2020, Oyamada et al 2021, Rong et al 2021, Uddin et al 2023, Chowdhury et al 2024). This has great potential for diagnostic applications, such as respiratory-rate and heart-rate monitoring, heart-rate-variability analysis, pulse-wave-velocity measurement, autonomic nervous system monitoring, and even emotional-state assessment (Sakamoto et al 2015, Sakamoto 2020, Zhang et al 2021, 2023, Yin et al 2022, Koda et al 2024). Such technologies have been applied to medical diagnostics (e.g. irregular breathing, cardiac arrhythmias, and respiratory infections) and health-related monitoring and event detection (e.g. sleep apnea, sleep stages, autonomic nervous system activity, animal monitoring, and sudden changes in health status, including those related to sudden infant death syndrome) (Koda et al 2024, Sakamoto et al 2024). By enabling noncontact monitoring, MMW sensing inherently reduces invasiveness and improves patient safety. MMWs are deposited in the superficial skin layer, resulting in heating effects. International guidelines for RF exposure have recently added a limit for whole-body exposure (ICNIRP 2020), highlighting the importance of safety considerations even in this frequency range. The development of advanced algorithms to achieve high-accuracy imaging may also help reduce the exposure.

Current and future challenges

Radar systems based on wideband signals achieve a high range resolution, enabling the precise measurement of the distance between objects in the FOV and the radar system. Radar systems with two-dimensional (2D) array antennas achieve high angular resolution, enabling the measurement of the azimuth and elevation angles of objects. Doppler radar systems can estimate the line-of-sight velocity of an object via frequency shifts. Therefore, radar data are typically represented as complex-valued functions in five dimensions: time, range, azimuth, elevation, and Doppler frequency.

Although Doppler-time spectrograms are widely used in activity recognition because of their ability to capture velocity changes, they often neglect a wealth of information from other dimensions. For physiological sensing, phase-based analysis is often used by extracting the phase of a signal over time and converting it into displacement as a function of time. This enables the detection of subtle motions, such as those caused by respiration and heartbeats, which are often smaller than deliberate body movements. However, both approaches underutilize the full multidimensional potential of radar data, particularly from modern 2D multiple-input and multiple-output radars.

Another major challenge lies in the diversity of data characteristics driven by individual differences in patients/users and varying measurement conditions. Most studies have been conducted in controlled environments, instructing participants to perform specific activities or remain stationary. However, in real-world scenarios, activities are continuous and transitions seamless, and multiple actions often overlap. Stationary individuals exhibit unconscious movements influenced by factors such as age, physical condition, and cultural background. These individual differences can be exploited for identification and authentication (Alrawili et al 2024, Kobayashi et al 2024, Peng et al 2024); however, they complicate radar-based human sensing and demand robust signal processing techniques. For example, body displacement waveforms from physiological signals vary depending on which part of the human body is measured; respiratory displacements at the chest and abdomen are different, as do heartbeat displacements at the head and legs. In the radar measurements of a person, both the position and posture of the target person and the radar installation position affect the measurement accuracy (Li et al 2006, Wang et al 2013, Nahar et al 2018, Koshisaka and Sakamoto 2024). Therefore, an important future challenge is the development of a technology that can handle data diversity depending on person- and environment-specific factors while combining multiple radar echoes to achieve accurate measurement of physiological signals.

Another challenge is the integration of radar-based sensing with existing systems and services in medicine and healthcare. Sensor fusion (i.e. combining radar with other sensors) can enhance system performance by leveraging the distinct advantages of different modalities (Zhang et al 2021). Networks of multiple radar sensors, whether identical systems are positioned to observe a target from various angles or diverse systems with varying frequencies and modulation techniques, can further improve the performance through synergistic effects. The use of radar networks is also crucial for measuring the physiological signals and activities of multiple individuals simultaneously (Li et al 2020, Singh et al 2020, Iwata et al 2021, Koda et al 2021). This is because a single radar alone cannot constantly capture all echoes from multiple individuals due to blocking and shadowing. These integrations are expected to address the current limitations and enable practical applications of radar-based sensing technologies.

Finally, machine learning is expected to play an important role in addressing these challenges (Nocera et al 2024). Through pattern recognition and feature extraction, diverse and dynamic radar data can be processed and interpreted (Gurbuz and Amin 2019). However, the most advanced techniques in the field of deep learning require large, diverse, and representative datasets, which are often unavailable in radar applications. Therefore, publicly available databases and effective data-labelling approaches are crucial for the widespread adoption of machine learning in this field (Gusland et al 2021; Fioranelli et al 2022, Yang et al 2023).

Advances in science and technology to meet the challenges

Radar-based measurements of human activities and physiological signals have benefited from recent advancements in radar system design and processing capabilities. Some recent advances are summarized as follows:

1. Exploiting high-dimensional radar data with machine learning

To address the challenge of the underutilization of high dimensionality in radar signals, recent advances in radar signal processing techniques and applied machine learning provide a promising solution to improve the performance of radar-based human monitoring by simultaneously exploiting information in multiple domains (i.e. range, velocity, azimuth and elevation angles, and time). Advanced algorithms, particularly deep learning methods, can automatically extract relevant features from large-scale datasets. These approaches are adept at handling multidimensional data inputs, thereby enabling the simultaneous analysis of temporal, spatial, and velocity-related data. By leveraging large amounts of data, machine learning can discover complex patterns and correlations that traditional signal processing may overlook. In this context, the development of well-curated open-access radar datasets can support the training and validation of machine learning models, thus making radar-based sensing more scalable.

2. Addressing data diversity and environmental variability

Real-world radar measurements vary widely due to individual differences and fluctuating environmental conditions. Although machine learning can adapt to diverse training data, advanced signal processing methods, such as adaptive filtering, multidimensional transforms, and other non-machine-learning-based signal separation methods (e.g. independent vector analysis, nonnegative matrix factorization, sparse-representation-based optimization, robust principal component analysis, and Bayesian inference-based methods) are essential for extracting subtle physiological features. Moreover, integrating multiple radar systems follows the same principle as multi-camera setups in computer vision, combining multiple viewpoints to compensate for occlusions, poor angular resolution, and unfavourable aspect angles, thereby enhancing the overall measurement accuracy. These combined strategies enable robust and practical radar-based sensing applications.

3. Integrating radar with multimodal sensing systems

The integration of radar-based sensing with other systems in medicine and healthcare is another frontier of technological advancement. Ongoing research aims to integrate radar-based health monitoring with telemedicine and smart healthcare systems to enable continuous remote physiological sensing. These advancements are expected to increase the clinical viability of radar-based sensing technologies, making them suitable for real-world deployment in hospitals, homes, and public spaces. Sensor fusion techniques—where radar data are combined with information from other modalities, such as optical, acoustic, or inertial sensors—can help develop synergistic systems that overcome the limitations of any single sensor. Moreover, radar sensor networks deployed at multiple locations/angles or using different frequencies and modulation techniques can improve performance by mitigating issues, such as signal blockage and occlusions, particularly when monitoring multiple individuals with arbitrary positions and postures.

Concluding remarks

Radar-based human sensing technologies offer transformative potential in healthcare, diagnostics, and entertainment by enabling noncontact measurements with minimal privacy concerns. These systems measure physical and physiological parameters and support applications such as activity recognition, fall detection, and heart-rate monitoring. However, current methods simplify radar data, thereby neglecting the full multidimensional potential of radar-based sensing. Addressing this issue requires advanced signal processing to achieve greater accuracy and robustness. Real-world scenarios add complexity because of continuous overlapping activities and individual differences. Machine learning holds promise for overcoming these challenges by enabling advanced pattern recognition and feature extraction; however, progress has been hindered by a lack of large-scale, labelled, and representative datasets. Therefore, the development of public databases and the exploration of unsupervised and transfer learning methods are necessary. Integrating radars with complementary sensors or using multiple radar systems offers opportunities to enhance the radar performance. With continued innovation, radar sensing is expected to play a key role in the creation of smarter and more connected solutions to meet societal needs.

5.3. THz sensing

Zachary Taylor1, Emma Pickwell-MacPherson2 and Maya Mizuno3

1 Department of Electronics and Nanoengineering, Aalto University, Espoo, Finland

2 Department of Physics, University of Warwick, Coventry CV4 7AL, United Kingdom

3 National Institute of Information and Communications Technology, Koganei, Japan

Status

Since the 1990s, the development of optical sources and detectors has accelerated in THz frequency region. Biomedical applications of THz spectroscopic and imaging systems using these technologies have benefited from increased radiance and sensitivity and have been applied to tissue assessment via the probing tissue water content, tissue birefringence and high order structure of protein. (Pickwell and Wallace 2006, Taylor et al 2015, Markelz and Mittleman 2022). Small changes in tissue hydration and protein structure often lead to measurable variations in THz refractive index and absorption coefficient, making THz spectroscopy a promising tool for diagnosing e.g. protein denaturation and early-stage cancers. High tissue water content and scattering from turbid tissue properties limit in vivo penetration depth. Therefore, a significant body of current work is focused on the cornea (Ozheredov et al 2018) and skin. Research efforts have explored and advanced implementation, including system development, theoretical analysis, in vitro and in vivo experimental studies, and a small but growing body of human subject measurements (Hernandez-Cardoso et al 2022, Qi et al 2024, Young et al 2024). At the same time, safety considerations remain important. The ICNIRP (2020) guidelines currently extend up to 300 GHz as well as optical radiation (ICNIRP 2013a), whereas their application to THz biomedical uses is still under discussion. Despite these advancements, the clinical implementation of THz technology remains challenging.

Current and future challenges

Historically, much of THz biomedical research has been performed with general purpose spectroscopy systems that are configured to maximize spectral coverage and dynamic range. However, clinical and biomedical utilization of THz waves necessitate systems engineered end-to-end specifically for a target application which takes into consideration both the expected contrast mechanisms and the possible in vivo surface topography. For example, detecting abnormal endothelial water content as related to corneal disease benefits from system optics matched to the near-spherical profile of human cornea and a spectral band that can probe the resonant-like backscatter due to distinct layer structure. Conversely, detection, monitoring, and management of early-stage skin cancer, psoriasis, and dermatitis should leverage systems with large illumination bandwidths to increase to maximize contrast. Skin features a large range of layer thickness and hydration gradient profiles which can limit contrast. Increased bandwidth can increase contrast and enable temporal separation of return pulses from dielectric window front and back surfaces which aid in calibration.

On the data processing side, the development of numerical correction and measurement techniques to compensate for clutter due to curved uneven topography is ongoing but challenging as THz system optics render detection systems highly directive and dense, and user-friendly focal plane arrays do not exist. Additionally, THz waves are highly susceptible to diffraction as corneal radii of curvature are on the order of a 10’s of wavelengths, skin surface roughness is in the sub-wavelength regime, and optical apertures are typically only 10’s of wavelength in diameter.

Finally, it is often difficult to correlate THz imaging and spectroscopy data with other related data sets such as visual appearance, layer structure and thickness, pressure, or any other information that could improve detection/diagnosis if it was taken into consideration with THz measurements. Improvements to multimodal fiducial markers and integration of adjunct imaging/sensing modalities into the THz system aperture may improve the clinical utility of THz data.

Advances in science and technology to meet the challenges

Novel imaging methods have been developed for scanning the near-spherical surface of human cornea (Sung et al 2018, Virk et al 2023). The system in figure 5(a) takes advantage of an off-axis parabolic mirror’s focusing properties. As demonstrated by the ray-tracing diagram scanning a collimated beam transverse to the mirror’s clear aperture plane produces an angular scan at the focal point. If the cornea is positioned such that its centre of curvature is coincident with the mirror focal point, then the cornea surface can be scanned at normal incidence while keeping the source, detector, and target stationary.

Figure 5.

Figure 5.

In vivo sensing systems and measurement examples for (a) human eyes and (b) skin and (c) a hybrid THz–optical analysis example where optical coherence tomography (OCT) is combined with THz measurements for rabbit eyes. © 2018 IEEE. Reprinted, with permission, from Sung et al 2018. Reproduced from Hernandez-Serrano et al 2024. CC BY 4.0. Adapted with permission from Mizuno et al 2021. © Optical Society of America.

Diffraction artefacts were observed at the edges of the eyeball in all images of the characterization targets, either increasing or decreasing the signal in areas where the known target reflectivity did not vary. Ongoing research is focused on mitigating diffraction effects by optimizing the aperture size. Similar diffraction artefacts were observed in a THz analysis of rabbit eyes used for MMW exposure experiments (Mizuno et al 2021), and its effects were suppressed by extracting specular reflection components using a time-domain method.

For the accurate diagnosis of skin with surface roughness, a portable THz handheld scanner was developed, as shown in figure 5(b) (Hernandez-Serrano et al 2024). A femtosecond pulse train is split into two arms. One arm generates ultrabroadband THz radiation (emitter) and second arm is used as an optical gate at the detector. Data can be analysed directly in the time domain or analysed in the frequency domain via the Fourier Transform (figure 5(b)). The human body’s non-uniform topography prevents system optical design to a specific shape a priori (unlike cornea) and thus it is typical to use a low loss, dielectric window (e.g. quartz) to flatten the field as seen in figure 5(b). This configuration produces two distinct reflections; one from the air–quartz interface (labelled as ‘A’) and the other from the quartz–skin interface labelled as ‘B’). The refractive index and absorption coefficient of the skin were calculated from the reflection signals. To ensure consistent force application on the skin during the measurement, two force-sensitive resistors were integrated into the tip of the probe. Furthermore, these optical properties were derived from data measured at 55 and 60 s, when the air gap between the quartz and skin of each subject was minimal. Additionally, the occlusion effect (blocking the skin) stabilizes after ∼30 s of contact. This measurement technique enables the quantitative evaluation of human skin hydration gradients and stratum corneum thickness, demonstrating its potential for noninvasive dermatological assessment.

For measurement of skin and eyes, significant recent work has focused on integrating OCT with THz spectroscopic imaging. OCT utilizes a low-coherence light source to perform tomography by cohering a tissue backscattered beam with a copy of itself. The system design and data analysis methodologies are very similar to THz TDS and thus form a natural adjunct. An example of OCT integration with THz TDS experiments is shown in figure 5(c) where OCT imagery confirms increased corneal thickness 1 d after 162 GHz irradiance at 360 mW cm−2; which is ∼36x more than current safety standards dictate. On the other hand, THz reflectance from the cornea decreased just after the 162 GHz irradiance due to the acute drying. The integrated analysis enables discussion that the acute drying and heat cause a trigger for corneal epithelium damage.

Concluding remarks

THz sensing represents a promising approach for overcoming current measurement limitations, offering noninvasive, high-sensitivity detection of tissue hydration and structural changes. Advances in imaging systems and data correction methods are expected to enable clinical translation of THz technology, ophthalmology and dermatology, in the near future.

To ensure human protection in these cases, experimental clarification and data accumulation, particularly in the eyes (Kojima et al 2020) and skin (Sasaki et al 2017), are critical issues. This experimental data is necessary to revise the current safety standards that the THz band spans. RF guidelines of the International Commission on NIR Protection (ICNIRP 2020) sources from low frequency up to 300 GHz while the guidelines on laser radiation (ICNIRP 2013b) covers sources from ultraviolet down to 300 GHz. These standards do not agree at 300 GHz and it is clear that the success of future clinical applications will rely on clarifying application and frequency specific safety margins in the THz band.

According to ICNIRP (2020), incident/APD averaged over 1 or 4 cm2 in square shape is the relevant quantity up to 300 GHz, whereas assessment in terms of irradiance (W m−2) with an aperture with an area of about 1 cm2 circle is suggested above 300 GHz. Although the typical output levels used in current THz sensing systems are far below these limits, explicit reference to the guideline quantities is essential to assess potential adverse effects for future applications.

6. Wireless charging and safety in medical devices

6.1. Wireless charging of implanted medical devices

Mauro Feliziani1, Valerio De Santis1, and Hubregt J Visser2

1 Department of Industrial and Information Engineering and Economics, University of L’Aquila, Italy

2 Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands

Status

The global IMDs market has been experiencing significant growth and is projected to expand further.

IMDs can be categorized by device type, with the most common ones being:

  • 1.

    Orthopaedic implants: devices such as joint replacements, trauma fixation, and spinal implants;

  • 2.

    Cardiovascular implants: including pacemakers (PMs), implantable cardioverter-defibrillators, cardiac resynchronization therapy devices, and cardiac pumps/total artificial hearts;

  • 3.

    Neurostimulators: devices such as deep brain stimulators (DBSs) that deliver electrical stimulation to specific parts of the nervous system;

  • 4.

    Cochlear implants: devices that provide a sense of sound to individuals with severe hearing loss;

  • 5.

    Ophthalmic implants: devices such as intraocular lenses replacing natural lenses in cataract surgery;

  • 6.

    Capsules and microrobots: although not implanted, these small ingestible devices with sensors and cameras travel through the digestive tract for diagnostic or therapeutic purposes.

Electric/electronic IMDs, generally known as active IMDs (AIMDs), are powered by an autonomous battery that must be replaced once exhausted. This needs surgical intervention for patients with potential health risks from complications (infections or failures).

WPT has revolutionized AIMDs by enabling the transfer of electrical energy without the need for physical connectors (Jow and Ghovanloo 2007, Kurs et al 2007, Si et al 2008, Ho et al 2013, Agarwal et al 2017). Recent advancements in WPT for AIMDs has led to improving patient comfort, reducing infection risks, extending device longevity, and enabling AIMD functions that were previously inactive to save energy. These benefits motivated the transition from traditional power supplies to WPT systems.

The current EM WPT technologies for AIMDs can be classified based on various criteria:

  • Number of transmitters (TX) and receivers (RX): the most common setup features a single external TX wirelessly powering an implanted RX. However, advanced configurations may incorporate multiple TXs and/or RXs to enhance the WPT performance.

  • Operating frequency: LF, intermediate-frequency (IF), and RF;

  • Technology: near-, mid-, and far-field

  • Coupling mechanisms: capacitive, inductive, resonant inductive, radio-frequency/MW, and optical coupling;

  • Power levels: ultra-low power, low power, and high power;

  • Implantation depth (range): deep, subcutaneous, and wearable;

  • Powering system: battery-powered and battery-free implants.

For AIMD applications, the key indicators of WPT performance include power transfer efficiency (PTE), power level, range, size, and biocompatibility.

The typical performance of different WPT categories is summarized in table 1.

Table 1.

WPT performance vs WPT technology.

Feature Near-field WPT Mid-field WPT Far-field WPT
Operating principle Inductive/resonant couplING Evanescent/quasi-static fields Full-wave propagation
Frequency bands IF (100 kHz–13.56 MHz) IF (85 kHz–10 MHz) RF/microwave/optical waves (2.45, 5.8 GHz/ 24–60 GHz THz−1)
Range Small (mm to cm) Medium (cm to 10 cm) High (10 cm to m)
Efficiency High (∼80%–90%) Moderate (∼50%–70%) Low (∼10%–50%)
Power level High (mW to W) Moderate (mW to W) Low (μW to mW)
Application suitability Pacemakers, insulin pumps, neuro-stimulators Deep-tissue implants, wireless biosensors Smart implants, bioelectronic sensors

This classification provides a clearer understanding of how different WPT technologies are applied in AIMDs with varying trade-offs in efficiency, range, power, and suitability for specific applications.

WPT applications have been investigated for common IMDs such as cochlear implants (Zeng et al 2008, Qian et al 2019), retinal implants (RamRakhyani and Lazzi 2012, Weiland and Humayun 2014, Mashhadi et al 2018), neuro-stimulators (Mirbozorgi et al 2017, Tanabe et al 2017, Barbruni et al 2020), PMs (Monti et al 2015, Campi et al 2016, Wang et al 2020a), cardiac defibrillators (Tang et al 2017, Campi et al 2021a), artificial heart, insulin pump (Ballardini et al 2023), drug pump (Rhee et al 2021), endoscopic capsules (Basar et al 2014, Campi et al 2023, Zhang et al 2024). Although IMDs with WPT technology are currently under development and experimentation worldwide, they are not already widely available in the market, even though some commercial applications exist, for example, Medtronics PERCEPT™ RC NEUROSTIMULATOR Rechargeable for DBS (www.medtronic.com/) and Cochlear Implants for Hearing Loss, MED-EL (www.medel.com/uk).

WPT technologies offer several advantages over the traditional solutions:

  • Reduced maintenance: AIMDs with rechargeable batteries can be wirelessly powered by an external source, eliminating the need for frequent battery or device replacements by surgical procedures.

  • Improved device longevity: WPT allows for continuous power delivery and extends the operational life of the devices.

  • Minimized invasiveness: WPT eliminates the need for external cables and reduces the capacity, size, and weight of the internal battery, making it less expensive and easier to implant.

  • Enhanced patient comfort: WPT eases the management of external power sources, improves patient comfort, and reduces complications.

  • Safer operation: WPT reduces the risk of mechanical failure and infection by eliminating external power connections.

  • Advanced functionality: WPT supports more advanced IMDs, enabling higher power and real-time data transmission with continuous power.

Owing to these advantages, WPT has the potential to improve the overall functionality, safety, and patient experience with IMDs, particularly in devices that require continuous operation inside the human body.

In terms of operating frequency, medical WPT systems for AIMDs typically use LF (hundreds of kHz) to RF ranges (up to several MHz or beyond, depending on the coupling mechanism). Unlike environmental exposure, ICNIRP or IEEE guidelines for public EMF limits are not directly applied to implanted medical devices; instead, safety is primarily ensured through device-specific standards (e.g. ISO, IEC, and FDA guidance). Nevertheless, as WPT systems become more powerful and widely adopted, there is a growing need for harmonization between exposure guidelines and medical device regulations to ensure both patient protection and reliable device performance.

Challenges and future directions

Although the potential of WPT for IMDs is evident, several challenges remain in the development and widespread adoption of these technologies, including efficiency improvements, EMF safety, EMI mitigation, biocompatibility, standardization, and market acceptance. However, from a physical and biological perspective, the main problems mainly depend on the power requested from the AIMD and the depth of the implants (Poon et al 2010).

The power levels of AIMDs vary according to their functionality and design. Ultra-low-power devices, typically consuming less than 10 μW, include PMs and continuous monitoring sensors designed for long-term operation with minimal energy. Low-power devices, ranging from 10 μW to a few milliwatts, include implantable drug delivery systems, hearing implants, and cardiac devices that require low energy for specific tasks. High-power devices, which consume several milliwatts to watts, include neurostimulation devices, muscle stimulators, cardiac pumps, and MW-based implants that require higher energy for therapeutic functions.

The EM challenges of wirelessly powering deep medical implants mainly involve the PTE, EMF safety, and field penetration (Campi et al 2019). As the implant depth increases, energy transfer from the external source becomes less efficient owing to tissue attenuation, which increases with frequency. The need to ensure safe power transmission is critical, as higher power levels can cause adverse health effects such as tissue damage, neurostimulation, or overheating. Additionally, field penetration becomes more difficult at greater depths, and the surrounding tissue can absorb or scatter EM energy, reducing the efficiency of power transfer. Implant size has also become an issue, as deep implants require compact power receivers, which are harder to design for efficient deep power transfer. EMI with other devices and power transfer reliability can be problematic (Hikage et al 2016, Campi et al 2021b).

Scientific and technological advances addressing the challenges

A range of scientific and technological advancements are helping to address the challenges of WPT in medical devices. However, the main challenge in applying WPT to IMDs is the delivery of high power to deep implants, while ensuring compliance with EMF safety standards (Campi et al 2019, Lan et al 2025). To address this challenge, the following sections present three examples related to capsules, high-power implants, and future developments.

Electronic capsules rely on integrated batteries, but their limited capacity often restricts functions, such as drug release and electric stimulation. WPT offers a potential solution using a TX coil placed outside the human body and an RX coil inside the capsule. The PTE depends on the capsule’s position, with distances from the skin ranging from 2 cm to over 10 cm, a considerable distance given the attenuation of magnetic fields in biological tissues. The key design challenges include selecting the operational frequency, configuring the receiving coil, and designing the transmitting coil. The operational frequency must balance the field attenuation in biological tissues with the WPT performance. The receiving coil must be sufficiently small to fit the capsule (typically 2–2.5 cm) while ensuring adequate power transfer. A triaxial RX coil is more effective than a uniaxial RX coil because it can capture magnetic fields in/from any direction. However, because they are larger and heavier, research has focused on using an uniaxial RX coil in combination with a TX coil system capable of generating a spatially varying magnetic field in the human body. Various solutions have been proposed to address this issue (Campi et al 2023, Fadhel and Kamal 2024, Zhang et al 2024, Zheng et al 2025).

One of the AIMDs characterized by high power consumption and deep implantation that could greatly benefit from WPT is the left ventricular assist device (LVAD), an advanced electrically powered mechanical pump that supports patients with severe heart failure. Currently, LVADs receive power externally via a percutaneous driveline (DL), which significantly increases the risk of severe infection at the exit site. LVADs require an average power of approximately 5 W, with peak demands reaching 20–25 W in the pulsatile mode. These devices are implanted directly attached to the heart, meaning that they are positioned several centimetres beneath the skin. As a result, wirelessly powering these implants using conventional architectures, where an external TX coil transfers energy to an internal RX coil mounted on the LVAD, presents significant challenges. To overcome these limitations, a simple yet highly efficient WPT architecture has been developed for implanted systems (Campi et al 2020). In this design, the RX coil is implanted subcutaneously, whereas the TX coil remains external, forming a well-established transcutaneous energy transfer (TET) coupler. The subcutaneous RX unit is connected to the LVAD through a fully internal DL. In addition, a rechargeable battery is implanted in the RX to provide supplemental energy. The internal battery supplies power during peak LVAD demands in pulsatile mode. Therefore, the WPT system is sized for average power consumption, whereas the battery compensates for short bursts of higher power. The WPT configuration with a battery allows patients to remain untethered for 1–2 h, significantly improving their mobility and quality of life. By eliminating the need for a constant connection, WPT-integrated LVADs offer greater freedom, reduced infection risk, and enhanced patient comfort (Liu et al 2018, Seshadri et al 2018, Cao and Tang 2018, Campi et al 2021b).

Finally, only a few medical devices, such as artificial organs and cardiac pumps, require significant power; however, the demand for high-power AIMDs is expected to increase. Next-generation devices require continuous operation and substantial power consumption. Mainstream research typically focuses on powering each AIMD individually, with each AIMD equipped with its own WPT receiver. However, when multiple AIMDs are implanted in the same patient, EM coupling and detuning between receivers can reduce PTE. In addition, mutual coupling may lead to localized tissue heating or EMI, which could compromise device functionality and patient safety (Campi et al 2021b, Maradei et al 2024). Recently, a new concept that implants a wired network within the human body to link multiple AIMDs was proposed (Campi et al 2021a). This network can be powered externally by a high-power TET system that distributes energy to multiple AIMDs via wired connections. Additionally, the internal batteries of AIMDs can serve as shared backup power sources for all devices connected to the network. The proposed solution offers several benefits, including a reduced number of implanted WPT systems, higher efficiency, continuous high-power delivery, fewer surgical interventions for battery or device replacement, elimination of percutaneous cables, lower risk of infection, and the potential to deliver exceptionally high power to the human body, enabling novel and previously unimagined applications.

Concluding remarks

WPT of medical devices is on the brink of revolutionizing healthcare by enabling the continuous operation of implanted devices without the need for invasive procedures or frequent battery replacement. This offers vast potential to improve the management of chronic medical conditions and enhance patients’ quality of life.

Current technologies, such as inductive and resonant inductive coupling, have demonstrated promise in powering small- and medium-power medical devices; however, challenges related to power efficiency, EMF safety, and standardization still remain. Research and technological advances are addressing these challenges, particularly the development of more efficient energy transfer methods, improved biocompatibility, miniaturization, and creation of universal charging standards.

6.2. Safety for IMDs

Ji Chen

Department of Electrical and Computer Engineering, University of Houston, Houston, TX, United States of America

Status

Implantable medical devices are increasingly vital in supporting various organ functions, thereby enhancing overall societal well-being (Bazaka and Jacob 2012, Khanna 2016, Wilson et al 2021, Magisetty and Park 2022). Based on their operational characteristics (see table 2), these devices can be classified into passive IMDs (PIMDs), which do not require a battery for functionality, and AIMDs, which rely on a power source or wireless charging to operate (European Parliament and Council 2017). When patients with these devices undergo MRI scans, significant safety concerns arise—such as RF induced heating and gradient field-induced PNS (Schueler et al 1999, Nyenhuis et al 2005, Woods 2007, Dal Molin and Hecker 2013, Davids et al 2019, Jabehdar Maralani et al 2020, Wang et al 2022c, Yang et al 2025). Typically, AIMD systems include devices such as PMs, various neuromodulation systems (e.g. spinal cord stimulators, DBSs), cochlear implants, and brain-machine interface systems. In contrast, PIMD systems often comprise stents, orthopaedic plates, pedicle screw systems, hip replacements, knee replacements, and similar devices.

Table 2.

Key features of passive and active implantable devices.

Feature Passive implantable medical devices (PIMDs) Active implantable medical devices (AIMDs)
Power source No external power required Requires an internal/external power source
Functionality Primarily structural or mechanical Electronic or electromechanical
Interaction with body Limited direct interaction, relies on body mechanics Actively regulates or influences body functions

To ensure patient safety, regulatory bodies have developed guidance documents that integrates international standards to address these concerns. For all AIMDs, the following MRI evaluations should be performed (ASTM F2213 2017, ISO/TS 10974 2018, ASTM F2182 2020, ASTM F2052 2022, U.S. FDA 2023, ASTM F2119 2024, Lidgate et al 2024):

  • 1.

    RF-induced heating in the patient

  • 2.

    Gradient-induced device heating and RF-induced device heating in the patient

  • 3.

    Gradient-induced vibration in the patient and device

  • 4.

    B0-induced force in the patient

  • 5.

    B0-induced torque in the patient

  • 6.

    Gradient-induced extrinsic electric potential in the patient

  • 7.

    B0-induced malfunction in the device

  • 8.

    RF-induced malfunction in the device and RF rectification in the patient

  • 9.

    Gradient-induced malfunction in the device

  • 10.

    Combined fields test for both patient and device

  • 11.

    Image artefacts

For all PIMDs, the required evaluations are items 1, 4, 5, and 11 from the list above.

Based on the results of these tests, appropriate MRI conditions can be established to safely scan patients with implantable devices.

Challenges and future directions

The safety of implantable devices under MRI exposure can be characterized as a standard EM compatibility (EMC) problem, with challenges arising from the EMC source, coupling pathways to the implant, and the device’s immunity to external EM emissions. Likewise, solutions also lie within these domains. Despite advancements in current testing methods to address these issues, significant challenges persist.

1. Large uncertainties exist in the RF safety assessment

The state-of-the-art method for AIMD RF related evaluations involve the device model development using the transfer function (TF) method as described in the ISO/TS 10974 (2018)g. This procedure requires one to use both numerical modelling and experimental measurement to estimate the RF related heating near the device tip. Due to the variations in human body type, different implantation pathways inside human body, tissue properties near implantation pathways, measurement equipment accuracy, and numerical modelling precision, there is a large uncertainty budget typically around 25% needs to be accessed on top of the results (Yao et al 2018, Zeng et al 2018a, Wang et al 2020b, Winter et al 2021). Consequently, the developed conditions for safe scan of patients are often very conservative and can prohibit patients from scanning under some aggressive and high-resolution MR sequence.

2. UHF systems 5 T, 7 T and beyond

With the development of whole-body 5 T and head-imaging 7 T systems, the current methodology on device evaluation under 1.5 T and 3 T needs to be revised and updated (Barisano et al 2019, Fagan et al 2020). It was observed that the conventional method of using TF models to characterize device behaviour under 3 T is no longer applicable for high fields system over 7 T and 10.4 T (Abedi et al 2022, Zuo 2025). For UHF system, the safety evaluation may require more rigorous method, such as full-wave EM characterization tools.

3. Fast switching dB/dt lead to device neuron stimulation

For MR system related safety concerns, the RF-induced heating is typically the bottle neck that restricts the condition for safe MR scans. However, as the fast gradient switch of dB/dt increases, the unintended PNS stimulation will have pronounced effect due to the implants. Conventionally, only the AIMD needs to consider the gradient induced intended stimulation, however, recently a study found out that large PIMD can also lead unintended nerve stimulation near the ends of these devices (Yang et al 2025).

4. Implant interactions with other medical equipment, such as RF ablation, electrical surgical knife, and transcutaneous electrical nerve stimulation systems (TENSs)

With the operation of nearby RF ablation system, electrical surgical knife, and TENS systems, due to the antenna effect of the implantable devices (both AIMD and PIMD), the implantable devices can attract surrounding EM energy and redistribute these energies based on the factors such as shape and insulation thickness around them. Depending on the operation frequency of the surrounding medical equipment, such energy redistribution can lead to significant heating concerns or neural stimulation (Pfeiffer et al 1995, Holmgren et al 2008, Paniccia et al 2014).

Scientific and technological advances addressing the challenges

  • 1.

    Large uncertainties in the MRI RF safety assessment can potentially be addressed through patient specific approach in the safety evaluation. With the AI assisted segmentation method, it is feasible to quickly develop a patient model using low energy MR scan or x-ray method. Based on the developed HBM and pathway identified inside the HBM, one can quickly access the patient specific MR safety, especially the RF related safety concerns. With more and more human models developed using this approach, a large AI based imaging library can quickly assess the MR safety for various implantable devices (Aissani et al 2019, Vu et al 2021, Zheng et al 2021, Chang et al 2022).

  • 2.

    UHF systems, such as 5 T, 7 T, and beyond, pose significant challenges but also offer innovative solutions. These systems typically require multichannel methods to achieve B1+ field homogeneity for imaging purposes. However, this multichannel approach can also be leveraged to enhance safety in the presence of implants. For instance, a multichannel high-field system can be optimized to maintain field homogeneity in the region of imaging interest while simultaneously minimizing emission power in areas where an implanted device is located, reducing potential safety risks (van den Bergen et al 2009, Tang et al 2011, Zeng et al 2018b, Aberra et al 2020).

  • 3.

    Fast-switching dB/dt can lead to neuronal stimulation, which can be better addressed through more accurate neuron modelling. Traditionally, safety limits have been established based on induced charge limits, derived from experimental studies. While this approach is conservative to ensure patient safety, it may not fully capture real physiological responses. With advances in EM modelling and the use of the NEURON code, it is now possible to conduct patient-specific EM simulations. These simulations can provide a realistic electric field distribution along nerve fibres, which can then be integrated into the NEURON modelling tool. This approach enables the identification of actual neuronal activation rather than relying solely on the accumulated charge limit, leading to more precise and individualized safety assessments (Carnevale and Hines 2006, Han 2018, Tan et al 2020, Zilberti et al 2024).

  • 4.

    Interactions between implants and other medical equipment, such as RF ablation, electrosurgical knives, and TENS, pose significant safety concerns. However, with recent advancements in multiphysics and multi-scale modelling, neuron physiology, and GPU/cloud-based computational power, it is now feasible to conduct in-silico studies to better understand and mitigate these interactions. These advanced simulations enable more accurate assessments, thereby improving both device safety and treatment efficacy (Kenjereš 2014, Han 2018, Akbari 2021).

Concluding remarks

While significant advancements have been made in assessing MR safety for patients with medical implants, the growing landscape of other medical equipment—such as TENS devices and RF ablation systems—introduces new challenges for implant safety. Additionally, many everyday applications involve EM sources that may emit unintended signals. While these emissions are generally considered safe for individuals without implants, they can pose potential risks for patients with implanted medical devices. To ensure patient safety, it is imperative to thoroughly investigate the interactions between implants and various EM emission sources. A comprehensive understanding of these interactions will help mitigate risks and improve safety standards for individuals with medical implants.

6.3. Pacemaker malfunction: interference of EMFs

Takashi Hikage1 and Seungyoung Ahn2

1 Faculty of Information Science and Technology, Hokkaido University, Sapporo, Japan

2 Cho Chun Shik Graduate School of Mobility, Korea Advanced Institute of Science and Technology, Daejeon, Republic of Korea

Status

In contrast to passive implants (section 6.2), active implants such as PMs and neurostimulators require safety evaluations that address not only RF-induced heating but also the risk of malfunction and EMI. Pacemakers are critical medical devices that regulate cardiac rhythms and provide life-saving support to patients with arrhythmias or other heart conditions. However, the functionality of PMs can be compromised by exposure to EMFs, which are becoming increasingly ubiquitous due to the proliferation of electronic devices and communication systems. Numerous studies have identified specific frequency ranges and intensities of EMFs that can interfere with PM operations and cause malfunctions, such as inappropriate pacing, inhibition, and mode switching (Barbaro et al 1995, Irnich et al 1996, Toyoshima 1996, Hayes et al 1997, Grant and Schlegel 2000, Wang et al 2000, Mattei et al 2014). Regulatory bodies, including the International Commission on NIR Protection and Institute of Electrical and Electronics Engineers, have established guidelines (IEEE-C95.1 2019, ICNIRP 2020) for EMF exposure to minimize risks; however, real-world incidents highlight persistent vulnerabilities near devices, such as electronic article surveillance and wireless power transfer systems that generate strong magnetic fields in relatively LF bands, where EMI effects can occur (ISO/IEC TR 20017 2011, Hikage et al 2016, Driessen et al 2019, Campi et al 2021b, Maradei et al 2024). Furthermore, particularly in environments with high-intensity or unexpected EMF exposure, such as industrial zones, during medical imaging procedures, such as MRI, or near devices that generate strong magnetic fields, the potential impact of metal heating should not be overlooked.

Challenges and future directions

Key challenges and future directions for managing EMI risks in PMs are summarized below. Despite considerable advancements in PM technologies, several challenges remain.

  • 1.

    Increased EMF sources: The growing deployment of wireless power transfer systems, and wearable devices has increased the likelihood of exposure to diverse EMFs. Because individuals with PMs are likely to come into the close proximity to these devices, careful investigation and research are essential.

  • 2.

    Comprehensive testing standards: Current standards for PM testing under EMF exposure may not fully replicate real-world scenarios, leading to gaps in safety assessments (ANSI/AAMI PC69 2000, ANSI/AAMI PC69 2007, ISO/IEC 14117d 2012, ISO 14708-2 2019, ISO 14708-6 2019). The development of testing methods that address various frequencies, signal waveforms, and applications, along with achieving global standardization, remains a critical challenge.

  • 3.

    Complex interactions: Pacemaker susceptibility to EMFs varies according to device design, patient anatomy, and environmental factors, complicating the development of universal mitigation strategies.

  • 4.

    Enhancing risk communication: Further promotion of risk communication is essential. These efforts should target not only PM users but also electronic device users, thereby fostering a broader understanding of PM EMI-related risks.

Future research should focus on developing more robust EMF shielding technologies for PMs, thereby enhancing device programming to detect and mitigate EMF interference and establishing more rigorous and realistic testing protocols. Interdisciplinary collaboration among medical device manufacturers, regulatory bodies, and researchers is essential to effectively address these challenges.

Scientific and technological advances addressing the challenges

Recent scientific and technological advancements offer promising solutions to mitigate EMF-related risks for PM users.

  • 1.

    Enhanced shielding and filtering: Advances in materials and technologies have improved the resistance of PMs to high-frequency EMFs, particularly those emitted by modern communication devices, such as mobile phones and cellular devices (Anjitha and Sunitha 2024).

  • 2.

    MRI-compatible devices: Innovations in MRI-safe PMs and leads have reduced the risks associated with one of the most intense sources of EMF exposure in clinical settings (Mitka 2011, Indik et al 2017).

  • 3.

    Simulation-based testing: Sophisticated computational models of the interactions between human anatomy and EMFs can accurately assess PM behaviour under various exposure conditions. These models contributed to the development of safer device designs (Hikage et al 2020, Waki et al 2023). Nevertheless, the development of methods for verifying the accuracy and validity of the simulation results remains an important challenge.

Concluding remarks

Although significant strides have been made in addressing the effects of EMFs on PM functionality, challenges persist due to the rapidly evolving EMF environment and the diverse needs of PM users. Future research should prioritize multidisciplinary approaches that integrate technological innovation, patient education, regulatory updates, and the promotion of effective risk communication to ensure reliability and safety of PMs in increasingly complex EM environments. Addressing these challenges will contribute to improving the quality of life of PM users and maintaining their safety in a technology-driven world.

Contributor Information

Ilkka Laakso, Email: ilkka.laakso@aalto.fi.

Akimasa Hirata, Email: ahirata@nitech.ac.jp.

Data availability statement

The data that support the findings of this study are available upon reasonable request from the authors.

References

  1. Abedi H, Ansariyan A, Morita P P, Wong A, Boger J, Shaker G. AI-powered noncontact in-home gait monitoring and activity recognition system based on mm-wave FMCW radar and cloud computing. IEEE Int. Things J. 2023;10:9465–81. doi: 10.1109/JIOT.2023.3235268. [DOI] [Google Scholar]
  2. Abedi H, Boger J, Morita P P, Wong A, Shaker G. Hallway gait monitoring using novel radar signal processing and unsupervised learning. IEEE Sens. J. 2022;22:15133–45. doi: 10.1109/JSEN.2022.3184188. [DOI] [Google Scholar]
  3. Aberra A S, Peterchev A V, Grill W M. Biophysically realistic neuron models for simulation of cortical stimulation. J. Neural Eng. 2018;15:066023. doi: 10.1088/1741-2552/aadbb1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Aberra A S, Wang B, Grill W M, Peterchev A V. Simulation of transcranial magnetic stimulation in head model with morphologically-realistic cortical neurons. Brain Stimul. 2020;13:175–89. doi: 10.1016/j.brs.2019.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Ademaj A, et al. Clinical evidence for thermometric parameters to guide hyperthermia treatment. Cancers. 2022;14:625. doi: 10.3390/cancers14030625. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Agarwal K, Jegadeesan R, Guo Y-X, Thakor N V. Wireless power transfer strategies for implantable bioelectronics. IEEE Rev. Biomed. Eng. 2017;10:136–61. doi: 10.1109/RBME.2017.2683520. [DOI] [PubMed] [Google Scholar]
  7. Ahn S, Fröhlich F. Pinging the brain with transcranial magnetic stimulation reveals cortical reactivity in time and space. Brain Stimul. 2021;14:304–15. doi: 10.1016/j.brs.2021.01.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  8. Aissani S, Laistler E, Felblinger J. MR safety assessment of active implantable medical devices. Der Radiologe. 2019;59:40–45. doi: 10.1007/s00117-019-0541-6. [DOI] [PubMed] [Google Scholar]
  9. Alekseichuk I, Mantell K, Shirinpour S, Opitz A. Comparative modeling of transcranial magnetic and electric stimulation in mouse, monkey, and human. NeuroImage. 2019;194:136–8. doi: 10.1016/j.neuroimage.2019.03.044. [DOI] [PMC free article] [PubMed] [Google Scholar]
  10. Akbari A. Efficient Solver for Multi-physics Modeling of Radiofrequency Tumor Ablation Using High Performance Computing. McGill University; 2021. [Google Scholar]
  11. Akhtari M, Bryant H C, Mamelak A N, Flynn E R, Heller L, Shih J J, Mandelkern M, Matlachov A, Ranken D M, Best ED, DiMauro M A, Lee R R, Sutherling W W. Conductivities of three-layer live human skull. Brain Topogr. 2002;14:151–67. doi: 10.1023/a:1014590923185. [DOI] [PubMed] [Google Scholar]
  12. Aklan B, Zilles B, Paprottka P, Manz K, Pfirrmann M, Santl M, Abdel-Rahman S, Lindner L H. Regional deep hyperthermia: quantitative evaluation of predicted and direct measured temperature distributions in patients with high-risk extremity soft-tissue sarcoma. Int. J. Hyperth. 2019;36:170–85. doi: 10.1080/02656736.2018.1545098. [DOI] [PubMed] [Google Scholar]
  13. Alrawili R, AlQahtani A A S, Khan M K. Comprehensive survey: biometric user authentication application, evaluation, and discussion. Comput. Electr. Eng. 2024;119:109485. doi: 10.1016/j.compeleceng.2024.109485. [DOI] [Google Scholar]
  14. Amabile C, Ahmed M, Solbiati L, Meloni M F, Solbiati M, Cassarino S, Tosoratti N, Nissenbaum Y, Ierace T, Goldberg S N. Microwave ablation of primary and secondary liver tumours: ex vivo, in vivo, and clinical characterisation. Int. J. Hyperth. 2017;33:34–42. doi: 10.1080/02656736.2016.1196830. [DOI] [PubMed] [Google Scholar]
  15. Amin B, Elahi M A, Shahzad A, Porter E, O’Halloran M. A review of the dielectric properties of the bone for low frequency medical technologies. Biomed. Phys. Eng. Express. 2019;5:022001. doi: 10.1088/2057-1976/aaf210. [DOI] [PubMed] [Google Scholar]
  16. Andreano A, Brace C L. A comparison of direct heating during radiofrequency and microwave ablation in ex vivo liver. Cardiovasc. Interv. Radiol. 2013;36:505–11. doi: 10.1007/s00270-012-0405-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Anjitha V, Sunitha K. Handbook of Nano-Metamaterials. Springer; 2024. pp 1–43. [Google Scholar]
  18. ANSI/AAMI PC69 Active Implantable Medical Devices—Electromagnetic Compatibility—EMC Test Protocols for Implantable Cardiac Pacemakers and Implantable Cardioverter Defibrillators 2000.
  19. ANSI/AAMI PC69 Active Implantable Medical Devices—Electromagnetic Compatibility—EMC Test Protocols for Implantable Cardiac Pacemakers and Implantable Cardioverter 2007.
  20. Antal A, Alekseichuk I, Bikson M, Brockmöller J, Brunoni A R, Chen R, Cohen L G, Dowthwaite G, Ellrich J, Flöel A. Low intensity transcranial electric stimulation: safety, ethical, legal regulatory and application guidelines. Clin. Neurophysiol. 2017;128:1774–809. doi: 10.1016/j.clinph.2017.06.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Antal A, Boros K, Poreisz C, Chaieb L, Terney D, Paulus W. Comparatively weak after-effects of transcranial alternating current stimulation (tACS) on cortical excitability in humans. Brain Stimul. 2008;1:97–105. doi: 10.1016/j.brs.2007.10.001. [DOI] [PubMed] [Google Scholar]
  22. Antonenko D, Thielscher A, Saturnino G B, Aydin S, Ittermann B, Grittner U, Flöel A. Towards precise brain stimulation: is electric field simulation related to neuromodulation? Brain Stimul. 2019;12:1159–68. doi: 10.1016/j.brs.2019.03.072. [DOI] [PubMed] [Google Scholar]
  23. Arduino A, Bottauscio O, Chiampi M, Giaccone L, Liorni I, Kuster N, Zilberti L, Zucca M. Accuracy assessment of numerical dosimetry for the evaluation of human exposure to electric vehicle inductive charging systems. IEEE Trans. Electromagnet. Compat. 2020;62:1939–50. doi: 10.1109/TEMC.2019.2954111. [DOI] [Google Scholar]
  24. ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment. 2022. (available at: www.astm.org/f2052-21.html)
  25. ASTM F2119 Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants. 2024. (available at: www.astm.org/f2119-24.html)
  26. ASTM F2182 Standard Test Method for Measurement of Radio Frequency-Induced Heating on or Near Passive Implants During Magnetic Resonance Imaging. 2020. (available at: www.astm.org/f2182-19e02.html)
  27. ASTM F2213 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment. 2017. (available at: www.astm.org/f2213-17.html) [DOI] [PMC free article] [PubMed]
  28. Bakker A, Holman R, Rodrigues D B, Dobsicek Trefna H, Stauffer P R, van Tienhoven G, Rasch C R N, Crezee H. Analysis of clinical data to determine the minimum number of sensors required for adequate skin temperature monitoring of superficial hyperthermia treatments. Int. J. Hyperth. 2018;34:910–7. doi: 10.1080/02656736.2018.1466000. [DOI] [PubMed] [Google Scholar]
  29. Bakker A, van der Zee J, van Tienhoven G, Kok H P, Rasch C R N, Crezee H. Temperature and thermal dose during radiotherapy and hyperthermia for recurrent breast cancer are related to clinical outcome and thermal toxicity: a systematic review. Int. J. Hyperth. 2019;36:1024–39. doi: 10.1080/02656736.2019.1665718. [DOI] [PubMed] [Google Scholar]
  30. Bakker A, Zweije R, Kok H P, Kolff M W, van den Bongard H, Schmidt M, van Tienhoven G, Crezee H. Clinical feasibility of a high-resolution thermal monitoring sheet for superficial hyperthermia in breast cancer patients. Cancers. 2020;12:3644. doi: 10.3390/cancers12123644. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Ballardini G, et al. Controlling and powering a fully implantable artificial pancreas refillable by ingestible pills. Conf. IEEE Eng. in Medicine & Biology Society; 2023. pp. 1–7. [DOI] [PubMed] [Google Scholar]
  32. Barbaro V, Bartolini P, Donato A, Militello C, Altamura G, Ammirati F, Santini M. Do European GSM mobile cellular phones pose a potential risk to pacemaker patients? Pacing Clin. Electrophysiol. 1995;18:1218–24. doi: 10.1111/j.1540-8159.1995.tb06961.x. [DOI] [PubMed] [Google Scholar]
  33. Barbruni G L, Ros P M, Demarchi D, Carrara S, Ghezzi D. Miniaturised wireless power transfer systems for neurostimulation: a review. IEEE Trans. Biomed. Circuits Syst. 2020;14:1160–78. doi: 10.1109/TBCAS.2020.3038599. [DOI] [PubMed] [Google Scholar]
  34. Bardati F, Borrani A, Gerardino A, Lovisolo G A. SAR optimization in a phased array radiofrequency hyperthermia system. IEEE Trans. Biomed. Eng. 1995;42:1201–7. doi: 10.1109/10.476127. [DOI] [PubMed] [Google Scholar]
  35. Barisano G, Culo B, Shellock F G, Sepehrband F, Martin K, Stevens M, Wang D J, Toga A W, Law M. 7-Tesla MRI of the brain in a research subject with bilateral, total knee replacement implants: case report and proposed safety guidelines. Magn. Reson. Imaging. 2019;57:313–6. doi: 10.1016/j.mri.2018.11.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Barker A T, Jalinous R, Freeston I L. Non-invasive magnetic stimulation of human motor cortex. Lancet. 1985;1:1106–7. doi: 10.1016/S0140-6736(85)92413-4. [DOI] [PubMed] [Google Scholar]
  37. Baron P, Deckers R, de Greef M, Merckel L G, Bakker C J, Bouwman J G, Bleys R L, van den Bosch M A, Bartels L W. Correction of proton resonance frequency shift MR-thermometry errors caused by heat-induced magnetic susceptibility changes during high intensity focused ultrasound ablations in tissues containing fat. Magn. Reson. Med. 2014;72:1580–9. doi: 10.1002/mrm.25063. [DOI] [PubMed] [Google Scholar]
  38. Basar M R, Ahmad M Y, Cho J, Ibrahim F. Application of wireless power transmission systems in wireless capsule endoscopy: an overview. Sensors. 2014;14:10929–51. doi: 10.3390/s140610929. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. Bazaka K, Jacob M V. Implantable devices: issues and challenges. Electronics. 2012;2:1–34. doi: 10.3390/electronics2010001. [DOI] [Google Scholar]
  40. Beard B B, Kainz W, Onishi T, Iyama T, Watanabe S, Fujiwara O, Wang J, Bit-Babik G, Faraone A, Wiart J. Comparisons of computed mobile phone induced SAR in the SAM phantom to that in anatomically correct models of the human head. IEEE Trans. Electromagnet. Compat. 2006;48:397–407. doi: 10.1109/TEMC.2006.873870. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Beltrachini L. A finite element solution of the forward problem in EEG for multipolar sources. IEEE Trans. Neural Syst. Rehabil. Eng. 2018;27:368–77. doi: 10.1109/TNSRE.2018.2886638. [DOI] [PubMed] [Google Scholar]
  42. Bernstein M A, Huston I I I J, Ward H A. Imaging artifacts at 3.0 T. J. Magn. Reson. Imaging. 2006;24:735–46. doi: 10.1002/jmri.20698. [DOI] [PubMed] [Google Scholar]
  43. Bing C, Staruch R M, Tillander M, Kohler M O, Mougenot C, Ylihautala M, Laetsch T W, Chopra R. Drift correction for accurate PRF-shift MR thermometry during mild hyperthermia treatments with MR-HIFU. Int. J. Hyperth. 2016;32:673–87. doi: 10.1080/02656736.2016.1179799. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Bonello J, Demarco A, Farhat I, Farrugia L, Sammut C V. Application of artificial neural networks for accurate determination of the complex permittivity of biological tissue. Sensors. 2020;20:4640. doi: 10.3390/s20164640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  45. Bottiglieri A, Brace C, O’Halloran M, Farina L. MWA performed at 5.8 GHz through ‘Side Firing’approach: an exploratory study. Sensors. 2022;22:9320. doi: 10.3390/s22239320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  46. Boulant N, Wu X, Adriany G, Schmitter S, Uğurbil K, Van de Moortele P F. Direct control of the temperature rise in parallel transmission by means of temperature virtual observation points: simulations at 10.5 Tesla. Magn. Reson. Med. 2016;75:249–56. doi: 10.1002/mrm.25637. [DOI] [PMC free article] [PubMed] [Google Scholar]
  47. Brace C L. Radiofrequency and microwave ablation of the liver, lung, kidney, and bone: what are the differences? Curr. Probl. Diagn. Radiol. 2009;38:135–43. doi: 10.1067/j.cpradiol.2007.10.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. Brink W M, Yousefi S, Bhatnagar P, Remis R F, Staring M, Webb A G. Personalized local SAR prediction for parallel transmit neuroimaging at 7T from a single T1‐weighted dataset. Magn. Reson. Med. 2022;88:464–75. doi: 10.1002/mrm.29215. [DOI] [PMC free article] [PubMed] [Google Scholar]
  49. Bruggmoser G, et al. Quality assurance for clinical studies in regional deep hyperthermia. Strahlenther. Onkol. 2011;187:605–10. doi: 10.1007/s00066-011-1145-x. [DOI] [PubMed] [Google Scholar]
  50. Cabezas M, Oliver A, Lladó X, Freixenet J, Cuadra M B. A review of atlas-based segmentation for magnetic resonance brain images. Comput. Methods Programs Biomed. 2011;104:e158–e77. doi: 10.1016/j.cmpb.2011.07.015. [DOI] [PubMed] [Google Scholar]
  51. Campelo S N, Huang P-H, Buie C R, Davalos R V. Recent advancements in electroporation technologies: from bench to clinic. Annu. Rev. Biomed. Eng. 2023;25:77–100. doi: 10.1146/annurev-bioeng-110220-023800. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. Campi T, Cruciani S, De Santis V, Feliziani M. EMF safety and thermal aspects in a pacemaker equipped with a wireless power transfer system working at low frequency. IEEE Trans. Microw. Theory Tech. 2016;64:375–82. doi: 10.1109/TMTT.2015.2514087. [DOI] [Google Scholar]
  53. Campi T, Cruciani S, De Santis V, Maradei F, Feliziani M. Near field wireless powering of deep medical implants. Energies. 2019;12:2720. doi: 10.3390/en12142720. [DOI] [Google Scholar]
  54. Campi T, Cruciani S, Maradei F, Feliziani M. A new transmitting coil for powering endoscopic capsules using wireless power transfer. Electronics. 2023;12:1942. doi: 10.3390/electronics12081942. [DOI] [Google Scholar]
  55. Campi T, Cruciani S, Maradei F, Montalto A, Musumeci F, Feliziani M. Wireless powering of next-generation left ventricular assist devices (LVADs) without percutaneous cable driveline. IEEE Trans. Microw. Theory Tech. 2020;68:3969–77. doi: 10.1109/TMTT.2020.2992462. [DOI] [Google Scholar]
  56. Campi T, Cruciani S, Maradei F, Montalto A, Musumeci F, Feliziani M. Centralized high power supply system for implanted medical devices using wireless power transfer technology. IEEE Trans. Med. Robot. Bionics. 2021a;3:992–1001. doi: 10.1109/TMRB.2021.3123404. [DOI] [Google Scholar]
  57. Campi T, Cruciani S, Maradei F, Montalto A, Musumeci F, Feliziani M. EMI in a cardiac implantable electronic device (CIED) by the wireless powering of a left ventricular assist device (LVAD. IEEE Trans. Electromagn. Compat. 2021b;63:988–95. doi: 10.1109/TEMC.2020.3047465. [DOI] [Google Scholar]
  58. Cao K, Tang W. Application of wireless power transfer in totally implantable heart pump. IEEE Wireless Power Transfer Conf.; 2018. pp. 1–4. [Google Scholar]
  59. Carmi L, et al. Efficacy and safety of deep transcranial magnetic stimulation for obsessive-compulsive disorder: a prospective multicenter randomized double-blind placebo-controlled trial. Am. J. Psychiatry. 2019;176:931–8. doi: 10.1176/appi.ajp.2019.18101180. [DOI] [PubMed] [Google Scholar]
  60. Carnevale N T, Hines M L. The NEURON Book. Cambridge University Press; 2006. [Google Scholar]
  61. Caulfield K A, George M S. Optimized APPS-tDCS electrode position, size, and distance doubles the on-target stimulation magnitude in 3000 electric field models. Sci. Rep. 2022;12:20116. doi: 10.1038/s41598-022-24618-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  62. Cavagnaro M, Amabile C, Bernardi P, Pisa S, Tosoratti N. A minimally invasive antenna for microwave ablation therapies: design, performances, and experimental assessment. IEEE Trans. Biomed. Eng. 2011;58:949–59. doi: 10.1109/TBME.2010.2099657. [DOI] [PubMed] [Google Scholar]
  63. Chang J, Lan Q, Guo R, Zheng J, Romero R, Kainz W, Long S A, Chen J. MRI RF-induced heating prediction of complex-shaped passive implantable medical devices using mesh-based convolutional neural network. IEEE Trans. Microw. Theory Tech. 2022;71:2207–14. doi: 10.1109/TMTT.2022.3224944. [DOI] [Google Scholar]
  64. Chen K M, Rukspollmuang S, Nyquist D. Measurement of induced electric fields in a phantom model of man. Radio Sci. 1982;17:49S–59S. doi: 10.1029/RS017i05Sp0049S. [DOI] [Google Scholar]
  65. Chen P, Wang X, Wang M, Yang X, Guo S, Jiang C, Cui G, Kong L. Multi-view real-time human motion recognition based on ensemble learning. IEEE Sens. J. 2021;21:20335–47. doi: 10.1109/JSEN.2021.3094548. [DOI] [Google Scholar]
  66. Cheng D K. Field and Wave Electromagnetics. Pearson Eduication Limited; 2014. [Google Scholar]
  67. Chhatbar P Y, Kautz S A, Takacs I, Rowland N C, Revuelta G J, George M S, Bikson M, Feng W. Evidence of transcranial direct current stimulation-generated electric fields at subthalamic level in human brain in vivo. Brain Stimul. 2018;11:727–33. doi: 10.1016/j.brs.2018.03.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  68. Chowdhury J H, Shihab M, Pramanik S K, Hossain M S, Ferdous K, Shahriar M, Islam S M. Separation of heartbeat waveforms of simultaneous two-subjects using independent component analysis and empirical mode decomposition. IEEE Microw. Wirel. Technol. 2024;34:1059–62. doi: 10.1109/LMWT.2024.3420253. [DOI] [Google Scholar]
  69. Chu K F, Dupuy D E. Thermal ablation of tumours: biological mechanisms and advances in therapy. Nat. Rev. Cancer. 2014;14:199–208. doi: 10.1038/nrc3672. [DOI] [PubMed] [Google Scholar]
  70. Cohen S L, Bikson M, Badran B W, George M S. A visual and narrative timeline of US FDA milestones for transcranial magnetic stimulation (TMS) devices. Brain Stimul. 2022;15:73–75. doi: 10.1016/j.brs.2021.11.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  71. Cox R S, Kapp D S. Correlation of thermal parameters with outcome in combined radiation therapy-hyperthermia trials. Int. J. Hyperth. 1992;8:719–32. doi: 10.3109/02656739209005020. [DOI] [PubMed] [Google Scholar]
  72. Crezee J, Oei A L, Franken N A P, Stalpers L J A, Kok H P. Response: commentary: the impact of the time interval between radiation and hyperthermia on clinical outcome in patients with locally advanced cervical cancer. Front. Oncol. 2020;10:528. doi: 10.3389/fonc.2020.00528. [DOI] [PMC free article] [PubMed] [Google Scholar]
  73. Curto S, et al. Quantitative, multi-institutional evaluation of MR thermometry accuracy for deep-pelvic MR-hyperthermia systems operating in multi-vendor MR-systems using a new anthropomorphic phantom. Cancers. 2019;11:1709. doi: 10.3390/cancers11111709. [DOI] [PMC free article] [PubMed] [Google Scholar]
  74. Curto S, Taj‐Eldin M, Fairchild D, Prakash P, Crezee J, Oei A L. Microwave ablation at 915 MHz vs 2.45 GHz: a theoretical and experimental investigation. Med. Phys. 2015;42:6152–61. doi: 10.1118/1.4931959. [DOI] [PubMed] [Google Scholar]
  75. Dal Molin R, Hecker B. Implantable Medical Devices MRI Safe, pHealth 2013. IOS Press; 2013. pp. 96–100. [PubMed] [Google Scholar]
  76. Dannhauer M, Gomez L J, Robins P L, Wang D, Hasan N I, Thielscher A, Siebner H R, Fan Y, Deng Z-D. Electric field modeling in personalizing transcranial magnetic stimulation interventions. Biol. Psychiatry. 2024;95:494–501. doi: 10.1016/j.biopsych.2023.11.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  77. Dannhauer M, Huang Z, Beynel L, Wood E, Bukhari-Parlakturk N, Peterchev A V. TAP: targeting and analysis pipeline for optimization and verification of coil placement in transcranial magnetic stimulation. J. Neural Eng. 2022;19:026050. doi: 10.1088/1741-2552/ac63a4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  78. Das S K, Clegg S T, Samulski T V. Computational techniques for fast hyperthermia temperature optimization. Med. Phys. 1999;26:319–28. doi: 10.1118/1.598519. [DOI] [PubMed] [Google Scholar]
  79. Datta A, Zhou X, Su Y, Parra L C, Bikson M. Validation of finite element model of transcranial electrical stimulation using scalp potentials: implications for clinical dose. J. Neural Eng. 2013;10:036018. doi: 10.1088/1741-2560/10/3/036018. [DOI] [PubMed] [Google Scholar]
  80. Datta N R, Ordóñez S G, Gaipl U S, Paulides M M, Crezee H, Gellermann J, Marder D, Puric E, Bodis S. Local hyperthermia combined with radiotherapy and-/or chemotherapy: recent advances and promises for the future. Cancer Treat. Rev. 2015;41:742–53. doi: 10.1016/j.ctrv.2015.05.009. [DOI] [PubMed] [Google Scholar]
  81. Davids M, Guérin B, Vom Endt A, Schad L R, Wald L L. Prediction of peripheral nerve stimulation thresholds of MRI gradient coils using coupled electromagnetic and neurodynamic simulations. Magn. Reson. Med. 2019;81:686–701. doi: 10.1002/mrm.27382. [DOI] [PMC free article] [PubMed] [Google Scholar]
  82. Dawson T W, Stuchly M A. Analytic validation of a three-dimensional scalar-potential finite-difference code for low-frequency magnetic induction. Appl. Comput. Electromagn. J. 1996;11:72–81. [Google Scholar]
  83. Dayarian N, Khadem A. A hybrid boundary element-finite element approach for solving the EEG forward problem in brain modeling. Front. Syst. Neurosci. 2024;18:1327674. doi: 10.3389/fnsys.2024.1327674. [DOI] [PMC free article] [PubMed] [Google Scholar]
  84. De Leeuw A A, Crezee J, Lagendijk J J. Temperature and SAR measurements in deep-body hyperthermia with thermocouple thermometry. Int. J. Hyperth. 1993;9:685–97. doi: 10.3109/02656739309032056. [DOI] [PubMed] [Google Scholar]
  85. Deás Yero D, González F G, Van Troyen D, Vandenbosch G A. Dielectric properties of ex vivo porcine liver tissue characterized at frequencies between 5 and 500 kHz when heated at different rates. IEEE Trans. Biomed. Eng. 2018;65:2560–8. doi: 10.1109/TBME.2018.2807981. [DOI] [PubMed] [Google Scholar]
  86. Deford J, Gandhi O. An impedance method to calculate currents induced in biological bodies exposed to quasi-static electromagnetic fields. IEEE Trans. Electromagn. Compat. 1985;EMC-27:168–73. doi: 10.1109/TEMC.1985.304281. [DOI] [Google Scholar]
  87. Deng Z-D, Robins P L, Dannhauer M, Haugen L M, Port J D, Croarkin P E. Optimizing TMS coil placement approaches for targeting the dorsolateral prefrontal cortex in depressed adolescents: an electric field modeling study. Biomedicines. 2023;11:2320. doi: 10.3390/biomedicines11082320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  88. Deshazer G, Prakash P, Merck D, Haemmerich D. Experimental measurement of microwave ablation heating pattern and comparison to computer simulations. Int. J. Hyperth. 2017;33:74–82. doi: 10.1080/02656736.2016.1206630. [DOI] [PMC free article] [PubMed] [Google Scholar]
  89. Dewey W C. Arrhenius relationships from the molecule and cell to the clinic. Int. J. Hyperth. 2009;25:3–20. doi: 10.1080/02656730902747919. [DOI] [PubMed] [Google Scholar]
  90. Dewhirst M W, Viglianti B L, Lora-Michiels M, Hanson M, Hoopes P J. Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermia. Int. J. Hyperth. 2003;19:267–94. doi: 10.1080/0265673031000119006. [DOI] [PubMed] [Google Scholar]
  91. Diao Y, et al. Recent advances and future perspective in computational bioelectromagnetics for exposure assessments. Bioelectromagnetics. 2025;46:e70002. doi: 10.1002/bem.70002. [DOI] [PubMed] [Google Scholar]
  92. Diao Y, Liu L, Deng N, Lyu S, Hirata A. Tensor-conductance model for reducing the computational artifact in target tissue for low-frequency dosimetry. Phys. Med. Biol. 2023;68:205014. doi: 10.1088/1361-6560/acfae0. [DOI] [PubMed] [Google Scholar]
  93. Diao Y, Zhang L, Shi D, Hirata A. An effective edge conductivity for reducing staircasing error in induced electric field computation for low-frequency magnetic field dosimetry. Phys. Med. Biol. 2022;67:215011. doi: 10.1088/1361-6560/ac944b. [DOI] [PubMed] [Google Scholar]
  94. Dimbylow P J. FDTD calculations of the whole-body averaged SAR in an anatomically realistic voxel model of the human body from 1 MHz to 1 GHz. Phys. Med. Biol. 1997;42:479–90. doi: 10.1088/0031-9155/42/3/003. [DOI] [PubMed] [Google Scholar]
  95. Dobšíček Trefná H, et al. Quality assurance guidelines for superficial hyperthermia clinical trials: I. Clinical requirements. Int. J. Hyperth. 2017;33:471–82. doi: 10.1080/02656736.2016.1277791. [DOI] [PubMed] [Google Scholar]
  96. Dodig H, Poljak D, Cvetković M. On the edge element boundary element method/finite element method coupling for time harmonic electromagnetic scattering problems. Int. J. Numer. Methods Eng. 2021;122:3613–52. doi: 10.1002/nme.6675. [DOI] [Google Scholar]
  97. Driessen S, Napp A, Schmiedchen K, Kraus T, Stunder D. Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting electromagnetic fields in the intermediate frequency range: a systematic review. Ep Europace. 2019;21:219–29. doi: 10.1093/europace/euy155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  98. Drizdal T, Sumser K, Bellizzi G G, Fiser O, Vrba J, Rhoon G C V, Yeo D T B, Margarethus M P. Simulation guided design of the MRcollar: a MR compatible applicator for deep heating in the head and neck region. Int. J. Hyperth. 2021;38:382–92. doi: 10.1080/02656736.2021.1892836. [DOI] [PubMed] [Google Scholar]
  99. Duck F A. Physical Properties of Tissues: A Comprehensive Reference Book. Academic Press; 1990. [Google Scholar]
  100. Eichfelder G, Gebhardt M. Local specific absorption rate control for parallel transmission by virtual observation points. Magn. Reson. Med. 2011;66:1468–76. doi: 10.1002/mrm.22927. [DOI] [PubMed] [Google Scholar]
  101. Eldaief M C, Halko M A, Buckner R L, Pascual-Leone A. Transcranial magnetic stimulation modulates the brain’s intrinsic activity in a frequency-dependent manner. Proc. Natl Acad. Sci. USA. 2011;108:21229–34. doi: 10.1073/pnas.1113103109. [DOI] [PMC free article] [PubMed] [Google Scholar]
  102. Elming P B, Sorensen B S, Oei A L, Franken N A P, Crezee J, Overgaard J, Horsman M R. Hyperthermia: the optimal treatment to overcome radiation resistant hypoxia. Cancers. 2019;11:60. doi: 10.3390/cancers11010060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  103. Ertürk M A, El‐Sharkawy A M M, Bottomley P A. Monitoring local heating around an interventional MRI antenna with RF radiometry. Med. Phys. 2015;42:1411–23. doi: 10.1118/1.4907960. [DOI] [PMC free article] [PubMed] [Google Scholar]
  104. Etoz S, Brace C L. Computed tomography-based modeling of water vapor-induced changes in permittivity during microwave ablation. IEEE Trans. Biomed. Eng. 2019;67:2427–33. doi: 10.1109/TBME.2019.2962363. [DOI] [PubMed] [Google Scholar]
  105. European Parliament and Council . Official Journal L. 117 edn. European Parliament and Council; 2017. Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices; p. p 175. (available at: http://data.europa.eu/eli/reg/2017/745/oj) [Google Scholar]
  106. Evans C, Bachmann C, Lee J S, Gregoriou E, Ward N, Bestmann S. Dose-controlled tDCS reduces electric field intensity variability at a cortical target site. Brain Stimul. 2020;13:125–36. doi: 10.1016/j.brs.2019.10.004. [DOI] [PubMed] [Google Scholar]
  107. Fadhel Y B, Kamal A-H. Wireless power transfer system using Helmholtz coils for powering Gastrointestinal Capsule biomedical implants. IEEE Int. Multi-Conf. Systems, Signals & Devices; 2024. pp. 309–13. [Google Scholar]
  108. Fagan A J, Amrami K K, Welker K M, Frick M A, Felmlee J P, Watson R E. Magnetic resonance safety in the 7T environment. Magn. Reson. Imaging Clin. 2020;28:573–82. doi: 10.1016/j.mric.2020.07.002. [DOI] [PubMed] [Google Scholar]
  109. Farrugia L, et al. The complex permittivity of biological tissues: a practical measurement guideline. IEEE Access. 2024;12:10296–314. doi: 10.1109/ACCESS.2024.3352728. [DOI] [Google Scholar]
  110. Feddersen T V, Hernandez-Tamames J A, Franckena M, van Rhoon G C, Paulides M M. Clinical performance and future potential of magnetic resonance thermometry in hyperthermia. Cancers. 2020;13:31. doi: 10.3390/cancers13010031. [DOI] [PMC free article] [PubMed] [Google Scholar]
  111. Fiala D, Lomas K J, Stohrer M. Computer prediction of human thermoregulatory and temperature responses to a wide range of environmental conditions. Int. J. Biometeorol. 2001;45:143–59. doi: 10.1007/s004840100099. [DOI] [PubMed] [Google Scholar]
  112. Fiedler T M, Ladd M E, Orzada S. Local and whole‐body SAR in UHF body imaging: implications for SAR matrix compression. Magn. Reson. Med. 2025;93:842–9. doi: 10.1002/mrm.30306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  113. Fioranelli F, Le Kernec J, Shah S A. Radar for health care: recognizing human activities and monitoring vital signs. IEEE Potentials. 2019;38:16–23. doi: 10.1109/MPOT.2019.2906977. [DOI] [Google Scholar]
  114. Fioranelli F, Zhu S, Roldan I. Benchmarking classification algorithms for radar-based human activity recognition. IEEE Aerosp. Electron. Syst. Mag. 2022;37:37–40. doi: 10.1109/MAES.2022.3216262. [DOI] [Google Scholar]
  115. Fisicaro F, Lanza G, Grasso A A, Pennisi G, Bella R, Paulus W, Pennisi M. Repetitive transcranial magnetic stimulation in stroke rehabilitation: review of the current evidence and pitfalls. Ther. Adv. Neurol. Disord. 2019;12:1756286419878317. doi: 10.1177/1756286419878317. [DOI] [PMC free article] [PubMed] [Google Scholar]
  116. Fitzgerald P B, Fountain S, Daskalakis Z J. A comprehensive review of the effects of rTMS on motor cortical excitability and inhibition. Clin. Neurophysiol. 2006;117:2584–96. doi: 10.1016/j.clinph.2006.06.712. [DOI] [PubMed] [Google Scholar]
  117. Foster K, Schwan H P. Dielectrical properties of tissues and biological materials. A critical review. Crit. Rev. Biomed. Eng. 1989;17:25–104. [PubMed] [Google Scholar]
  118. Fox M D, Halko M A, Eldaief M C, Pascual-Leone A. Measuring and manipulating brain connectivity with resting state functional connectivity magnetic resonance imaging (fcMRI) and transcranial magnetic stimulation (TMS) Neuroimage. 2012;62:2232–43. doi: 10.1016/j.neuroimage.2012.03.035. [DOI] [PMC free article] [PubMed] [Google Scholar]
  119. Franckena M, Canters R, Termorshuizen F, Van Der Zee J, Van Rhoon G. Clinical implementation of hyperthermia treatment planning guided steering: a cross over trial to assess its current contribution to treatment quality. Int. J. Hyperth. 2010;26:145–57. doi: 10.3109/02656730903453538. [DOI] [PubMed] [Google Scholar]
  120. Franckena M, Fatehi D, de Bruijne M, Canters R A, van Norden Y, Mens J W, van Rhoon G C, van der Zee J. Hyperthermia dose-effect relationship in 420 patients with cervical cancer treated with combined radiotherapy and hyperthermia. Eur. J. Cancer. 2009;45:1969–78. doi: 10.1016/j.ejca.2009.03.009. [DOI] [PubMed] [Google Scholar]
  121. Fung P K, Haber A L, Robinson P A. Neural field theory of plasticity in the cerebral cortex. J. Theor. Biol. 2013;318:44–57. doi: 10.1016/j.jtbi.2012.09.030. [DOI] [PubMed] [Google Scholar]
  122. Gabriel S, Lau R W, Gabriel C. The dielectric properties of biological tissues: II. Measurements in the frequency range 10 Hz to 20 GHz. Phys. Med. Biol. 1996;41:2251–69. doi: 10.1088/0031-9155/41/11/002. [DOI] [PubMed] [Google Scholar]
  123. Gandhi O P, Chen J Y. Numerical dosimetry at power‐line frequencies using anatomically based models. Bioelectromagnetics. 1992;13:43–60. doi: 10.1002/bem.2250130706. [DOI] [PubMed] [Google Scholar]
  124. Gharamohammadi A, Khajepour A, Shaker G. In-vehicle monitoring by radar: a review. IEEE Sens. J. 2023;23:25650–72. doi: 10.1109/JSEN.2023.3316449. [DOI] [Google Scholar]
  125. Gokyar S, Zhao C, Ma S J, Wang D J. Deep learning‐based local SAR prediction using B1 maps and structural MRI of the head for parallel transmission at 7 T. Magn. Reson. Med. 2023;90:2524–38. doi: 10.1002/mrm.29797. [DOI] [PMC free article] [PubMed] [Google Scholar]
  126. Gomez-Tames J, Laakso I, Haba Y, Hirata A, Poljak D, Yamazaki K. Computational artifacts of the in situ electric field in anatomical models exposed to low-frequency magnetic field. IEEE Trans. Electromagnet. Compat. 2018;60:589–97. doi: 10.1109/TEMC.2017.2748219. [DOI] [Google Scholar]
  127. Gorji A, Bourdoux A, Sahli H. On the generalization and reliability of single radar-based human activity recognition. IEEE Access. 2021;9:85334–49. doi: 10.1109/ACCESS.2021.3088452. [DOI] [Google Scholar]
  128. Gosselin M-C, et al. Development of a new generation of high-resolution anatomical models for medical device evaluation: the virtual population 3.0. Phys. Med. Biol. 2014;59:5287. doi: 10.1088/0031-9155/59/18/5287. [DOI] [PubMed] [Google Scholar]
  129. Graesslin I, Homann H, Biederer S, Börnert P, Nehrke K, Vernickel P, Mens G, Harvey P, Katscher U. A specific absorption rate prediction concept for parallel transmission MR. Magn. Reson. Med. 2012;68:1664–74. doi: 10.1002/mrm.24138. [DOI] [PubMed] [Google Scholar]
  130. Grant F H, Schlegel R E. Effects of an increased air gap on the in vitro interaction of wireless phones with cardiac pacemakers. Bioelectromagnetics. 2000;21:485–90. doi: 10.1002/1521-186X(200010)21:7&#x0003c;485::AID-BEM1&#x0003e;3.0.CO;2-S. [DOI] [PubMed] [Google Scholar]
  131. Grossman N, et al. Noninvasive deep brain stimulation via temporally interfering electric fields. Cell. 2017;169:1029–41.e16. doi: 10.1016/j.cell.2017.05.024. [DOI] [PMC free article] [PubMed] [Google Scholar]
  132. Guo R, Huang T, Li M, Zhang H, Eldar Y C. Physics-embedded machine learning for electromagnetic data imaging: examining three types of data-driven imaging methods. IEEE Signal Process. Mag. 2023;40:18–31. doi: 10.1109/MSP.2022.3198805. [DOI] [Google Scholar]
  133. Gurbuz S Z, Amin M G. Radar-based human-motion recognition with deep learning: promising applications for indoor monitoring. IEEE Signal Process. Mag. 2019;36:16–28. doi: 10.1109/MSP.2018.2890128. [DOI] [Google Scholar]
  134. Gusland D, Christiansen J M, Torvik B, Fioranelli F, Gurbuz S Z, Ritchie M. Open radar initiative: large scale dataset for benchmarking of micro-doppler recognition algorithms. 2021 IEEE Radar Conf. (RadarConf21); 2021. pp. 1–6. [Google Scholar]
  135. Haider Z, Le Dréan Y, Sacco G, Nikolayev D, Sauleau R, Zhadobov M. High-resolution model of human skin appendages for electromagnetic dosimetry at millimeter waves. IEEE Microw. Mag. 2022;2:214–27. doi: 10.1109/JMW.2021.3126712. [DOI] [Google Scholar]
  136. Hall P S, Hao Y. Antennas and Propagation for Body-Centric Wireless Communications. Artech House; 2012. [Google Scholar]
  137. Hamada M, Hanajima R, Terao Y, Arai N, Furubayashi T, Inomata-Terada S, Yugeta A, Matsumoto H, Shirota Y, Ugawa Y. Quadro-pulse stimulation is more effective than paired-pulse stimulation for plasticity induction of the human motor cortex. Clin. Neurophysiol. 2007;118:2672–82. doi: 10.1016/j.clinph.2007.09.062. [DOI] [PubMed] [Google Scholar]
  138. Han Z. Real-Time Multi-physics Modeling of Radio-frequency Electrosurgical Proceduress. Rensselaer Polytechnic Institute; 2018. [Google Scholar]
  139. Hasan N I, Dannhauer M, Wang D, Deng Z-D, Gomez L J. Real-time computation of brain E-field for enhanced transcranial magnetic stimulation neuronavigation and optimization. Imaging Neurosci. 2025;3:imag_a_00412. doi: 10.1162/imag_a_00412. [DOI] [PMC free article] [PubMed] [Google Scholar]
  140. Hasanov K F, Ma A W, Nachman A I, Joy M L. Current density impedance imaging. IEEE Trans. Med. Imaging. 2008;27:1301–9. doi: 10.1109/TMI.2008.922691. [DOI] [PubMed] [Google Scholar]
  141. Hasgall P, Di Gennaro F, Baumgartner C, Neufeld E, Gosselin M, Payne D, Klingenböck A, Kuster N. IT’IS database for thermal and electromagnetic parameters of biological tissue 2023.
  142. Hassan M, Lias K, Buniyamin N, Naimullah B, Jobli A. Sar performance of rectangular microstrip antenna for breast cancer hyperthermia treatment with different period of treatment procedure. J. Phys.: Conf. Ser. 2021;2071:012048. doi: 10.1088/1742-6596/2071/1/012048. [DOI] [Google Scholar]
  143. Hayes D L, Wang P J, Reynolds D W, Estes N M, Griffith J L, Steffens R A, Carlo G L, Findlay G K, Johnson C M. Interference with cardiac pacemakers by cellular telephones. N. Engl. J. Med. 1997;336:1473–9. doi: 10.1056/NEJM199705223362101. [DOI] [PubMed] [Google Scholar]
  144. He Q, Yang E, Fang S. A survey on human profile information inference via wireless signals. IEEE Commun. Surv. Tutor. 2024;26:2577–610. doi: 10.1109/COMST.2024.3373397. [DOI] [Google Scholar]
  145. Hernandez-Cardoso G G, Amador-Medina L F, Gutierrez-Torres G, Reyes-Reyes E S, Benavides Martínez C A, Cardona Espinoza C, Arce Cruz J, Salas-Gutierrez I, Murillo-Ortíz B O, Castro-Camus E. Terahertz imaging demonstrates its diagnostic potential and reveals a relationship between cutaneous dehydration and neuropathy for diabetic foot syndrome patients. Sci. Rep. 2022;12:3110. doi: 10.1038/s41598-022-06996-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  146. Hernandez-Serrano A I, Ding X, Young J, Costa G, Dogra A, Hardwicke J, Pickwell-MacPherson E. Terahertz probe for real time in vivo skin hydration evaluation. Adv. Photon. 2024;3:016012. doi: 10.1117/1.APN.3.1.016012. [DOI] [Google Scholar]
  147. Hikage T, Ito S, Ohtsuka A. Novel interference voltage measurement for beam-type wireless power transfer using an electro-optical converter for EMI assessment of active implantable medical devices. URSI Radio Sci. Lett. 2020;2:1–3. [Google Scholar]
  148. Hikage T, Nojima T, Fujimoto H. Active implantable medical device EMI assessment for wireless power transfer operating in LF and HF bands. Phys. Med. Biol. 2016;61:4522. doi: 10.1088/0031-9155/61/12/4522. [DOI] [PubMed] [Google Scholar]
  149. Hirata A, Diao Y, Onishi T, Sasaki K, Ahn S, Colombi D, De Santis V, Laakso I, Giaccone L, Joseph W. Assessment of human exposure to electromagnetic fields: review and future directions. IEEE Trans. Electromagn. Compat. 2021;63:1619–30. doi: 10.1109/TEMC.2021.3109249. [DOI] [Google Scholar]
  150. Hirata A, Ito F, Laakso I. Confirmation of quasi-static approximation in SAR evaluation for a wireless power transfer system. Phys. Med. Biol. 2013;58:N241. doi: 10.1088/0031-9155/58/17/N241. [DOI] [PubMed] [Google Scholar]
  151. Ho J S, Kim S, Poon A S. Midfield wireless powering for implantable systems. Proc. IEEE. 2013;101:1369–78. doi: 10.1109/JPROC.2013.2251851. [DOI] [Google Scholar]
  152. Hoffer E K, Drinane M C, Bhatnagar V, Mehta R, Munger D P, Borsic A. Radiofrequency ablation of hepatocellular carcinoma guided by real-time physics-based ablation simulation: a prospective study. Int. J. Hyperth. 2024;41:2331704. doi: 10.1080/02656736.2024.2331704. [DOI] [PubMed] [Google Scholar]
  153. Hoffmann R, Rempp H, Erhard L, Blumenstock G, Pereira P L, Claussen C D, Clasen S. Comparison of four microwave ablation devices: an experimental study in ex vivo bovine liver. Radiology. 2013;268:89–97. doi: 10.1148/radiol.13121127. [DOI] [PubMed] [Google Scholar]
  154. Holmgren C, Carlsson T, Mannheimer C, Edvardsson N. Risk of interference from transcutaneous electrical nerve stimulation on the sensing function of implantable defibrillators. Pacing Clin. Electrophysiol. 2008;31:151–8. doi: 10.1111/j.1540-8159.2007.00962.x. [DOI] [PubMed] [Google Scholar]
  155. Homann H, Börnert P, Eggers H, Nehrke K, Dössel O, Graesslin I. Toward individualized SAR models and in vivo validation. Magn. Reson. Med. 2011;66:1767–76. doi: 10.1002/mrm.22948. [DOI] [PubMed] [Google Scholar]
  156. Hong G, Antaris A L, Dai H. Near-infrared fluorophores for biomedical imaging. Nat. Biomed. Eng. 2017;1:0010. doi: 10.1038/s41551-016-0010. [DOI] [Google Scholar]
  157. Hong H, Zhang L, Zhao H, Chu H, Gu C, Brown M, Zhu X, Li C. Microwave sensing and sleep: noncontact sleep-monitoring technology with microwave biomedical radar. IEEE Microw. Mag. 2019;20:18–29. doi: 10.1109/MMM.2019.2915469. [DOI] [Google Scholar]
  158. Hoult D I. Sensitivity and power deposition in a high‐field imaging experiment. J. Magn. Reson. Imaging. 2000;12:46–67. doi: 10.1002/1522-2586(200007)12:1&#x0003c;46::AID-JMRI6&#x0003e;3.0.CO;2-D. [DOI] [PubMed] [Google Scholar]
  159. Huang S K. Radiofrequency Catheter Ablation of Cardiac Arrhythmias : Basic Concepts and Clinical Applications. Futura Publishing Company; 1995. [Google Scholar]
  160. Huang Y, Datta A, Bikson M, Parra L C. Realistic volumetric-approach to simulate transcranial electric stimulation—ROAST—a fully automated open-source pipeline. J. Neural Eng. 2019;16:056006. doi: 10.1088/1741-2552/ab208d. [DOI] [PMC free article] [PubMed] [Google Scholar]
  161. Huang Y, Liu A A, Lafon B, Friedman D, Dayan M, Wang X, Bikson M, Doyle W K, Devinsky O, Parra L C. Measurements and models of electric fields in the in vivo human brain during transcranial electric stimulation. eLife. 2017;6:e18834. doi: 10.7554/eLife.18834. [DOI] [PMC free article] [PubMed] [Google Scholar]
  162. Huang Y-Z, Edwards M J, Rounis E, Bhatia K P, Rothwell J C. Theta burst stimulation of the human motor cortex. Neuron. 2005;45:201–6. doi: 10.1016/j.neuron.2004.12.033. [DOI] [PubMed] [Google Scholar]
  163. Hurwitz M D. Hyperthermia and immunotherapy: clinical opportunities. Int. J. Hyperth. 2019;36:4–9. doi: 10.1080/02656736.2019.1653499. [DOI] [PubMed] [Google Scholar]
  164. Ibrahim T S, Lee R, Baertlein B A, Kangarlu A, Robitaille P-M L. Application of finite difference time domain method for the design of birdcage RF head coils using multi-port excitations. Magn. Reson. Imaging. 2000;18:733–42. doi: 10.1016/S0730-725X(00)00143-0. [DOI] [PubMed] [Google Scholar]
  165. ICNIRP ICNIRP guidelines on limits of exposure to incoherent visible and infrared radiation. Health Phys. 2013a;105:74–96. doi: 10.1097/HP.0b013e318289a611. [DOI] [PubMed] [Google Scholar]
  166. ICNIRP ICNIRP guidelines on limits of exposure to laser radiation of wavelengths between 180 nm and 1,000 μm. Health Phys. 2013b;105:271–95. doi: 10.1097/HP.0b013e3182983fd4. [DOI] [PubMed] [Google Scholar]
  167. ICNIRP Guidelines for limiting exposure to electrromagnetic fields (100 kHz to 300 GHz) Health Phys. 2020;118:483–524. doi: 10.1097/HP.0000000000001210. [DOI] [PubMed] [Google Scholar]
  168. IEC 60601-2-33 . Medical Electrical equipment—Part 2-33: Particular Requirements for the Basic Safety and Essential Performance of Magnetic Resonance Equipment for Medical Diagnosis. International Electrotechnical Commission (IEC); 2022. [Google Scholar]
  169. IEEE-C95.1 . IEEE Standard for Safety Levels with respect to Human Exposure to Radio Frequency Electromagnetic Fields, 0 Hz to 300 GHz. IEEE; 2019. [Google Scholar]
  170. Indahlastari A, Albizu A, Kraft J N, O’Shea A, Nissim N R, Dunn A L, Carballo D, Gordon M P, Taank S, Kahn A T. Individualized tDCS modeling predicts functional connectivity changes within the working memory network in older adults. Brain Stimul. 2021;14:1205–15. doi: 10.1016/j.brs.2021.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  171. Indik J, Gimbel J, Abe H, Alkmim-Teixeira R, Birgersdotter-Green U, Clarke G, Dickfeld T, Froelich J, Grant J, Hayes D. HRS expert consensus statement on magnetic resonance imaging and radiation exposure in patients with cardiovascular implantable electronic devices. Hear Rhythm. 2017;14:e97–e153. doi: 10.1016/j.hrthm.2017.04.025. [DOI] [PubMed] [Google Scholar]
  172. Irnich W, Batz L, Müller R, Tobisch R. Electromagnetic interference of pacemakers by mobile phones. Pacing Clin. Electrophysiol. 1996;19:1431–46. doi: 10.1111/j.1540-8159.1996.tb03155.x. [DOI] [PubMed] [Google Scholar]
  173. Ishihara Y, Calderon A, Watanabe H, Okamoto K, Suzuki Y, Kuroda K, Suzuki Y. A precise and fast temperature mapping using water proton chemical shift. Magn. Reson. Med. 1995;34:814–23. doi: 10.1002/mrm.1910340606. [DOI] [PubMed] [Google Scholar]
  174. ISO/IEC 14117d Active Implantable Medical Devices—Electromagnetic Compatibility—EMC Test Protocols for Implantable Cardiac Pacemakers, Implantable Cardioverter Defibrillators and Cardiac Resynchronization Devices. 2012 International Organization for Standardization.
  175. ISO/IEC TR 20017 Information Technology—Radio Frequency Identification for Item Management—Electromagnetic Interference Impact of ISO/IEC 18000 Interrogator Emitters on Implantable Pacemakers and Implantable Cardioverter Defibrillators. 2011 International Organization for Standardization.
  176. ISO/TS 10974 Assessment of the Safety of Magnetic Resonance Imaging for Patients with an Active Implantable Medical Device (International Organization for Standardization) 2018 (available at: www.iso.org/obp/ui/en/#iso:std:iso:ts:10974:ed-2:v1:en)
  177. ISO 14708-2 Implants For Surgery—Active Implantable Medical Devices—Part 2: Cardiac Pacemakers. 2019 International Organization for Standardization.
  178. ISO 14708-6 Implants for Surgery—Active Implantable Medical Devices—Part 6: Particular Requirements for Active Implantable Medical Devices Intended to Treat Tachyarrhythmia (Including Implantable Defibrillators) 2019 International Organization for Standardization.
  179. Issels R D. Hyperthermia adds to chemotherapy. Eur. J. Cancer. 2008;44:2546–54. doi: 10.1016/j.ejca.2008.07.038. [DOI] [PubMed] [Google Scholar]
  180. Iwata S, Koda T, Sakamoto T. Multiradar data fusion for respiratory measurement of multiple people. IEEE Sens. J. 2021;21:25870–9. doi: 10.1109/JSEN.2021.3117707. [DOI] [Google Scholar]
  181. Jabehdar Maralani P, Schieda N, Hecht E M, Litt H, Hindman N, Heyn C, Davenport M S, Zaharchuk G, Hess C P, Weinreb J. MRI safety and devices: an update and expert consensus. J. Magn. Reson. Imaging. 2020;51:657–74. doi: 10.1002/jmri.26909. [DOI] [PubMed] [Google Scholar]
  182. Ji Z, Brace C L. Expanded modeling of temperature-dependent dielectric properties for microwave thermal ablation. Phys. Med. Biol. 2011;56:5249. doi: 10.1088/0031-9155/56/16/011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  183. Johnson L, Alekseichuk I, Krieg J, Doyle A, Yu Y, Vitek J, Johnson M, Opitz A. Dose-dependent effects of transcranial alternating current stimulation on spike timing in awake nonhuman primates. Sci. Adv. 2020;6:1–9. doi: 10.1126/sciadv.aaz2747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  184. Jones E L, Oleson J R, Prosnitz L R, Samulski T V, Vujaskovic Z, Yu D, Sanders L L, Dewhirst M W. Randomized trial of hyperthermia and radiation for superficial tumors. J. Clin. Oncol. 2005;23:3079–85. doi: 10.1200/JCO.2005.05.520. [DOI] [PubMed] [Google Scholar]
  185. Jones R P, Kitteringham N R, Terlizzo M, Hancock C, Dunne D, Fenwick S W, Poston G J, Ghaneh P, Malik H Z. Microwave ablation of ex vivo human liver and colorectal liver metastases with a novel 14.5 GHz generator. Int. J. Hyperth. 2012;28:43–54. doi: 10.3109/02656736.2011.610428. [DOI] [PubMed] [Google Scholar]
  186. Joshi R, Murali S, Thirugnanasambandam N. Behavioral validation of individualized low-intensity transcranial electrical stimulation (tES) protocols. Eneuro. 2023;10:ENEURO.0374–22.2023. doi: 10.1523/ENEURO.0374-22.2023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  187. Jow U-M, Ghovanloo M. Design and optimization of printed spiral coils for efficient transcutaneous inductive power transmission. IEEE Trans. Biomed. Circuits Syst. 2007;1:193–202. doi: 10.1109/TBCAS.2007.913130. [DOI] [PubMed] [Google Scholar]
  188. Kangasmaa O, Laakso I. Estimation method for the anisotropic electrical conductivity of in vivo human muscles and fat between 10 kHz and 1 MHz. Phys. Med. Biol. 2022;67:225002. doi: 10.1088/1361-6560/ac9a1e. [DOI] [PubMed] [Google Scholar]
  189. Karampatzakis A, Samaras T. Numerical model of heat transfer in the human eye with consideration of fluid dynamics of the aqueous humour. Phys. Med. Biol. 2010;55:5653. doi: 10.1088/0031-9155/55/19/003. [DOI] [PubMed] [Google Scholar]
  190. Kasten F H, Duecker K, Maack M C, Meiser A, Herrmann C S. Integrating electric field modeling and neuroimaging to explain inter-individual variability of tACS effects. Nat. Commun. 2019;10:5427. doi: 10.1038/s41467-019-13417-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  191. Katscher U, Börnert P, Leussler C, Van Den Brink J S. Transmit sense. Magn. Reson. Med. 2003;49:144–50. doi: 10.1002/mrm.10353. [DOI] [PubMed] [Google Scholar]
  192. Katscher U, van den Berg C A T. Electric properties tomography: biochemical, physical and technical background, evaluation and clinical applications. NMR Biomed. 2017;30:e3729. doi: 10.1002/nbm.3729. [DOI] [PubMed] [Google Scholar]
  193. Kaune W, Forsythe W. Current densities measured in human models exposed to 60‐Hz electric fields. Bioelectromagnetics. 1985;6:13–32. doi: 10.1002/bem.2250060103. [DOI] [PubMed] [Google Scholar]
  194. Keangin P, Rattanadecho P, Wessapan T. An analysis of heat transfer in liver tissue during microwave ablation using single and double slot antenna. Int. J. Heat Mass Transfer. 2011;38:757–66. doi: 10.1016/j.icheatmasstransfer.2011.03.027. [DOI] [Google Scholar]
  195. Kenjereš S. Modeling and simulation of multi-physics multi-scale transport phenomenain bio-medical applications. J. Phys.: Conf. Ser. 2014;530:012006 [Google Scholar]
  196. Khanna V K. Implantable Medical Electronics. vol 10. Springer; 2016. [Google Scholar]
  197. Kim D, Jeong J, Jeong S, Kim S, Jun S C, Chung E. Validation of computational studies for electrical brain stimulation with phantom head experiments. Brain Stimul. 2015;8:914–25. doi: 10.1016/j.brs.2015.06.009. [DOI] [PubMed] [Google Scholar]
  198. Kim K, Narsinh K, Ozhinsky E. Technical advances in motion‐robust MR thermometry. Magn. Reson. Med. 2024;92:15–27. doi: 10.1002/mrm.30057. [DOI] [PMC free article] [PubMed] [Google Scholar]
  199. Kobayashi H, Tanaka Y, Sakamoto T. Individual identification using radar-measured respiratory and heartbeat features. IEEE Access. 2024;12:190972–87. doi: 10.1109/ACCESS.2024.3517668. [DOI] [Google Scholar]
  200. Koda T, Okumura S, Taki H, Hamada S, Sunadome H, Sato S, Chin K, Sakamoto T. Noncontact detection of sleep apnea using radar and expectation-maximization algorithm. IEEE Sens. J. 2024;24:32748–56. doi: 10.1109/JSEN.2024.3450890. [DOI] [Google Scholar]
  201. Koda T, Sakamoto T, Okumura S, Taki H. Noncontact respiratory measurement for multiple people at arbitrary locations using array radar and respiratory-space clustering. IEEE Access. 2021;9:106895–906. doi: 10.1109/ACCESS.2021.3099821. [DOI] [Google Scholar]
  202. Kodera S, Gomez-Tames J, Hirata A. Temperature elevation in the human brain and skin with thermoregulation during exposure to RF energy. Biomed. Eng. Online. 2018;17:1. doi: 10.1186/s12938-017-0432-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  203. Kojima M, Suzuki Y, Tasaki T, Tatematsu Y, Mizuno M, Fukunari M, Sasaki H. Clinical course of high-frequency millimeter-wave (162 GHz) induced ocular injuries and investigation of damage thresholds. J. Infrared Millim. Terahertz Waves. 2020;41:1–12. doi: 10.1007/s10762-020-00714-1. [DOI] [Google Scholar]
  204. Kok H P, et al. Toward online adaptive hyperthermia treatment planning: correlation between measured and simulated specific absorption rate changes caused by phase steering in patients. Int. J. Radiat. Oncol. Biol. Phys. 2014;90:438–45. doi: 10.1016/j.ijrobp.2014.05.1307. [DOI] [PubMed] [Google Scholar]
  205. Kok H P, Cressman E N K, Ceelen W, Brace C L, Ivkov R, Grull H, Ter Haar G, Wust P, Crezee J. Heating technology for malignant tumors: a review. Int. J. Hyperth. 2020;37:711–41. doi: 10.1080/02656736.2020.1779357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  206. Kok H P, Herrera T D, Crezee J. The relevance of high temperatures and short time intervals between radiation therapy and hyperthermia: insights in terms of predicted equivalent enhanced radiation dose. Int. J. Radiat. Oncol. Biol. Phys. 2023;115:994–1003. doi: 10.1016/j.ijrobp.2022.10.023. [DOI] [PubMed] [Google Scholar]
  207. Kok H P, van Rhoon G C, Herrera T D, Overgaard J, Crezee J. Biological modeling in thermoradiotherapy: present status and ongoing developments toward routine clinical use. Int. J. Hyperth. 2022;39:1126–40. doi: 10.1080/02656736.2022.2113826. [DOI] [PubMed] [Google Scholar]
  208. Koponen L M, Goetz S M, Peterchev A V. Double-containment coil with enhanced winding mounting for transcranial magnetic stimulation with reduced acoustic noise. IEEE Trans. Biomed. Eng. 2021;68:2233–40. doi: 10.1109/TBME.2020.3048321. [DOI] [PMC free article] [PubMed] [Google Scholar]
  209. Koponen L M, Nieminen J O, Ilmoniemi R J. Multi-locus trancranial magnetic stimulation—theory and implementation. Brain Stimul. 2018;11:849–55. doi: 10.1016/j.brs.2018.03.014. [DOI] [PubMed] [Google Scholar]
  210. Koshisaka T, Sakamoto T. Radar-based pulse wave sensing at multiple sites using a 3D human model. IEEE Sens. J. 2024;25:5241–48. doi: 10.1109/JSEN.2024.3519571. [DOI] [Google Scholar]
  211. Krause M R, Vieira P G, Thivierge J P, Pack C C. Brain stimulation competes with ongoing oscillations for control of spike timing in the primate brain. PLOS Biol. 2022;20:e3001650. doi: 10.1371/journal.pbio.3001650. [DOI] [PMC free article] [PubMed] [Google Scholar]
  212. Kroesen M, et al. The effect of the time interval between radiation and hyperthermia on clinical outcome in 400 locally advanced cervical carcinoma patients. Front. Oncol. 2019;9:134. doi: 10.3389/fonc.2019.00134. [DOI] [PMC free article] [PubMed] [Google Scholar]
  213. Kruse N C, Fioranelli F, Yarovoy A. Radar point cloud processing methods for human activity classification with point transformer networks. IEEE Trans. Radar Syst. 2023;2:1–12. doi: 10.1109/TRS.2023.3341230. [DOI] [Google Scholar]
  214. Kuang W, Nelson S. Low-frequency dielectric properties of biological tissues: a review with some new insights. Trans. ASAE. 1998;41:173–84. doi: 10.13031/2013.17142. [DOI] [Google Scholar]
  215. Kubota Y, Kodera S, Hirata A. A novel transfer learning framework for non-uniform conductivity estimation with limited data in personalized brain stimulation. Phys. Med. Biol. 2025;70:105002. doi: 10.1088/1361-6560/add105. [DOI] [PubMed] [Google Scholar]
  216. Kujirai T, Caramia M D, Rothwell J C, Day B L, Thompson P D, Ferbert A, Wroe S, Asselman P, Marsden C D. Corticocortical inhibition in human motor cortex. J. Physiol. 1993;471:501–19. doi: 10.1113/jphysiol.1993.sp019912. [DOI] [PMC free article] [PubMed] [Google Scholar]
  217. Kuo H-I, Bikson M, Datta A, Minhas P, Paulus W, Kuo M-F, Nitsche M A. Comparing cortical plasticity induced by conventional and high-definition 4 × 1 ring tDCS: a neurophysiological study. Brain Stimul. 2013;6:644–8. doi: 10.1016/j.brs.2012.09.010. [DOI] [PubMed] [Google Scholar]
  218. Kuo M-F, Paulus W, Nitsche M A. Therapeutic effects of non-invasive brain stimulation with direct currents (tDCS) in neuropsychiatric diseases. Neuroimage. 2014;85:948–60. doi: 10.1016/j.neuroimage.2013.05.117. [DOI] [PubMed] [Google Scholar]
  219. Kurs A, Karalis A, Moffatt R, Joannopoulos J D, Fisher P, Soljačić M. Wireless power transfer via strongly coupled magnetic resonances. Science. 2007;317:83–86. doi: 10.1126/science.1143254. [DOI] [PubMed] [Google Scholar]
  220. Kwon O I, Jeong W C, Sajib S Z, Kim H J, Woo E J. Anisotropic conductivity tensor imaging in MREIT using directional diffusion rate of water molecules. Phys. Med. Biol. 2014;59:2955. doi: 10.1088/0031-9155/59/12/2955. [DOI] [PubMed] [Google Scholar]
  221. Laakso I, Hirata A. Fast multigrid-based computation of the induced electric field for transcranial magnetic stimulation. Phys. Med. Biol. 2012;57:7753–65. doi: 10.1088/0031-9155/57/23/7753. [DOI] [PubMed] [Google Scholar]
  222. Laakso I, Mikkonen M, Koyama S, Hirata A, Tanaka S. Can electric fields explain inter-individual variability in transcranial direct current stimulation of the motor cortex? Sci. Rep. 2019;9:626. doi: 10.1038/s41598-018-37226-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  223. Laakso I, Tanaka S, Koyama S, De Santis V, Hirata A. Inter-subject variability in electric fields of motor cortical tDCS. Brain Stimul. 2015;8:906–13. doi: 10.1016/j.brs.2015.05.002. [DOI] [PubMed] [Google Scholar]
  224. Laakso I, Tani K, Gomez-Tames J, Hirata A, Tanaka S. Small effects of electric field on motor cortical excitability following anodal tDCS. iScience. 2024;27:108967. doi: 10.1016/j.isci.2024.108967. [DOI] [PMC free article] [PubMed] [Google Scholar]
  225. Laimer G, Verdonschot K H, Kopf L, van der Lei S, Scharll Y, Hannink G, Jenniskens S F, Meijerink M R, Bale R, Overduin C G. Multicenter and inter-software evaluation of ablative margins after thermal ablation of colorectal liver metastases. Eur. Radiol. 2025;35:1046–56. doi: 10.1007/s00330-024-10956-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  226. Lan J, Wei R, Xu X, Hirata A. Novel inductively coupled wireless power transfer system based on modified birdcage coils for torso implanted medical devices. IEEE Trans. Microw. Theory Tech. 2025;73:641–9. doi: 10.1109/TMTT.2024.3417317. [DOI] [Google Scholar]
  227. Lazebnik M, et al. A large-scale study of the ultrawideband microwave dielectric properties of normal, benign and malignant breast tissues obtained from cancer surgeries. Phys. Med. Biol. 2007;52:6093. doi: 10.1088/0031-9155/52/20/002. [DOI] [PubMed] [Google Scholar]
  228. Le Kernec J, Fioranelli F, Ding C, Zhao H, Sun L, Hong H, Lorandel J, Romain O. Radar signal processing for sensing in assisted living: the challenges associated with real-time implementation of emerging algorithms. IEEE Signal Process. Mag. 2019;36:29–41. doi: 10.1109/MSP.2019.2903715. [DOI] [Google Scholar]
  229. Le Ster C, Mauconduit F, Mirkes C, Vignaud A, Boulant N. Measuring radiofrequency field‐induced temperature variations in brain MRI exams with motion compensated MR thermometry and field monitoring. Magn. Reson. Med. 2022;87:1390–400. doi: 10.1002/mrm.29058. [DOI] [PubMed] [Google Scholar]
  230. Lee H K, Antell G A, Perez C A, Straube W L, Ramachandran G, Myerson R J, Emami B, Molmenti E P, Buckner A, Lockett M A. Superficial hyperthermia and irradiation for recurrent breast carcinoma of the chest wall: prognostic factors in 196 tumors. Int. J. Radiat. Oncol. Biol. Phys. 1998;40:365–75. doi: 10.1016/S0360-3016(97)00740-2. [DOI] [PubMed] [Google Scholar]
  231. Lefaucheur J-P, et al. Evidence-based guidelines on the therapeutic use of repetitive transcranial magnetic stimulation (rTMS): an update (2014–2018. Clin. Neurophysiol. 2020;131:474–528. doi: 10.1016/j.clinph.2014.05.021. [DOI] [PubMed] [Google Scholar]
  232. Leijsen R, Brink W, van den Berg C, Webb A, Remis R. Electrical properties tomography: a methodological review. Diagnostics. 2021;11:176. doi: 10.3390/diagnostics11020176. [DOI] [PMC free article] [PubMed] [Google Scholar]
  233. Leijsen R, van den Berg C, Webb A, Remis R, Mandija S. Combining deep learning and 3D contrast source inversion in MR‐based electrical properties tomography. NMR Biomed. 2022;35:e4211. doi: 10.1002/nbm.4211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  234. Leopold K A, et al. Relationships among tumor temperature, treatment time, and histopathological outcome using preoperative hyperthermia with radiation in soft tissue sarcomas. Int. J. Radiat. Oncol. 1992;22:989–98. doi: 10.1016/0360-3016(92)90798-M. [DOI] [PubMed] [Google Scholar]
  235. Lepetit-Coiffé M, Laumonier H, Seror O, Quesson B, Sesay M-B, Moonen C T, Grenier N, Trillaud H. Real-time monitoring of radiofrequency ablation of liver tumors using thermal-dose calculation by MR temperature imaging: initial results in nine patients, including follow-up. Eur. Radiol. 2010;20:193–201. doi: 10.1007/s00330-009-1532-1. [DOI] [PubMed] [Google Scholar]
  236. Li C, Xiao Y, Lin J. Experiment and spectral analysis of a low-power Ka-band heartbeat detector measuring from four sides of a human body. IEEE Trans. Microw. Theory Tech. 2006;54:4464–71. doi: 10.1109/TMTT.2006.884652. [DOI] [Google Scholar]
  237. Li H, Deng Z-D, Oathes D, Fan Y. Computation of transcranial magnetic stimulation electric fields using self-supervised deep learning. Neuroimage. 2022;264:119705. doi: 10.1016/j.neuroimage.2022.119705. [DOI] [PMC free article] [PubMed] [Google Scholar]
  238. Li H, Mehul A, Le Kernec J, Gurbuz S Z, Fioranelli F. Sequential human gait classification with distributed radar sensor fusion. IEEE Sens. J. 2020;21:7590–603. doi: 10.1109/JSEN.2020.3046991. [DOI] [Google Scholar]
  239. Lidgate S, Griffin G, Martin G. Magnetic resonance conditionality applications: EU requirements and supporting standards BSI White Paper 24. 2024. (available at: www.bsigroup.com/globalassets/meddev/localfiles/en-gb/documents/bsi-md-mri-whitepaper-uk-en.pdf)
  240. Lin Y-M, et al. Intraprocedural versus initial follow-up minimal ablative margin assessment after colorectal liver metastasis thermal ablation: which one better predicts local outcomes? Invest. Radiol. 2024;59:314–9. doi: 10.1097/RLI.0000000000001023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  241. Linthorst M, Baaijens M, Wiggenraad R, Creutzberg C, Ghidey W, van Rhoon G C, van der Zee J. Local control rate after the combination of re-irradiation and hyperthermia for irresectable recurrent breast cancer: results in 248 patients. Radiother. Oncol. 2015;117:217–22. doi: 10.1016/j.radonc.2015.04.019. [DOI] [PubMed] [Google Scholar]
  242. Liu Y, Li Y, Zhang J, Dong S, Cui C, Zhu C. IEEE PELS Workshop on Emerging Technologies: Wireless Power Transfer,2018); IEEE; pp. 1–5. [Google Scholar]
  243. Longo T A, Gopalakrishna A, Tsivian M, Van Noord M, Rasch C R, Inman B A, Geijsen E D. A systematic review of regional hyperthermia therapy in bladder cancer. Int. J. Hyperth. 2016;32:381–9. doi: 10.3109/02656736.2016.1157903. [DOI] [PMC free article] [PubMed] [Google Scholar]
  244. López-Alonso V, Cheeran B, Río-Rodríguez D, Fernández-del-olmo M. Inter-individual variability in response to non-invasive brain stimulation paradigms. Brain Stimul. 2014;7:372–80. doi: 10.1016/j.brs.2014.02.004. [DOI] [PubMed] [Google Scholar]
  245. Lopresto V, Argentieri A, Pinto R, Cavagnaro M. Temperature dependence of thermal properties of ex vivo liver tissue up to ablative temperatures. Phys. Med. Biol. 2019;64:105016. doi: 10.1088/1361-6560/ab1663. [DOI] [PubMed] [Google Scholar]
  246. Lopresto V, Pinto R, Lovisolo G A, Cavagnaro M. Changes in the dielectric properties of ex vivo bovine liver during microwave thermal ablation at 2.45 GHz. Phys. Med. Biol. 2012;57:2309. doi: 10.1088/0031-9155/57/8/2309. [DOI] [PubMed] [Google Scholar]
  247. Lu Y, et al. Design of the distribution of iron oxide (Fe3O4) nano-particle drug in realistic cholangiocarcinoma model and the simulation of temperature increase during magnetic induction hyperthermia. Pharmacol. Res. 2024;207:107333. doi: 10.1016/j.phrs.2024.107333. [DOI] [PubMed] [Google Scholar]
  248. Luber B, Deng Z-D. In: Handbook of Psychophysiology. Cacioppo J T, et al., editors. Cambridge University Press; 2017. pp. 116–50. [Google Scholar]
  249. Luber B, Lisanby S H. Enhancement of human cognitive performance using transcranial magnetic stimulation (TMS. Neuroimage. 2014;85:961–70. doi: 10.1016/j.neuroimage.2013.06.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
  250. Magisetty R, Park S-M. New era of electroceuticals: clinically driven smart implantable electronic devices moving towards precision therapy. Micromachines. 2022;13:161. doi: 10.3390/mi13020161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  251. Makarov S N, Noetscher G M, Raij T, Nummenmaa A. A quasi-static boundary element approach with fast multipole acceleration for high-resolution bioelectromagnetic models. IEEE Trans. Biomed. Eng. 2018;65:2675–83. doi: 10.1109/TBME.2018.2813261. [DOI] [PMC free article] [PubMed] [Google Scholar]
  252. Makarov S N, Wartman W A, Daneshzand M, Fujimoto K, Raij T, Nummenmaa A. A sofftware toolkit for TMS electric-field modeling with boundary element fast multipole method: an efficient MATLAB implementation. J. Neural Eng. 2020;17:046023. doi: 10.1088/1741-2552/ab85b3. [DOI] [PubMed] [Google Scholar]
  253. Mandija S, Meliadò E F, Huttinga N R F, Luijten P R, van den Berg C A T. Opening a new window on MR-based electrical properties tomography with deep learning. Sci. Rep. 2019;9:8895. doi: 10.1038/s41598-019-45382-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  254. Mansfield P, Harvey P R. Limits to neural stimulation in echo-planar imaging. Magn. Reson. Med. 1993;29:746–58. doi: 10.1002/mrm.1910290606. [DOI] [PubMed] [Google Scholar]
  255. Maradei F, Ahn S, Campi T, Covic G, Cruciani S, Hirata A, Wang S, Feliziani M. EMC and EMF safety aspects of wireless power transfer systems for e-mobility. IEEE Electromagn. Compat. Mag. 2024;13:74–92. doi: 10.1109/MEMC.2024.10834399. [DOI] [Google Scholar]
  256. Markelz A G, Mittleman D M. Perspective on terahertz applications in bioscience and biotechnology. ACS Photonics. 2022;9:1117–26. doi: 10.1021/acsphotonics.2c00228. [DOI] [Google Scholar]
  257. Mashhadi I A, Pahlevani M, Hor S, Pahlevani H, Adib E. A new wireless power-transfer circuit for retinal prosthesis. IEEE Trans. Power Electron. 2018;34:6425–39. doi: 10.1109/TPEL.2018.2872844. [DOI] [Google Scholar]
  258. Mathy R M, Giannakis A, Franke M, Winiger A, Kauczor H-U, Chang D-H. Factors impacting microwave ablation zone sizes: a retrospective analysis. Cancers. 2024;16:1279. doi: 10.3390/cancers16071279. [DOI] [PMC free article] [PubMed] [Google Scholar]
  259. Mattei E, Censi F, Triventi M, Calcagnini G. Electromagnetic immunity of implantable pacemakers exposed to wi-fi devices. Health Phys. 2014;107:318–25. doi: 10.1097/HP.0000000000000113. [DOI] [PubMed] [Google Scholar]
  260. Mcintosh R L, Anderson V. A comprehensive tissue properties database provided for the thermal assessment of a human at rest. Biophys. Rev. Lett. 2010;5:129–51. doi: 10.1142/S1793048010001184. [DOI] [Google Scholar]
  261. Meliadò E, Raaijmakers A, Sbrizzi A, Steensma B, Maspero M, Savenije M, Luijten P, van den Berg C. A deep learning method for image‐based subject‐specific local SAR assessment. Magn. Reson. Med. 2020;83:695–711. doi: 10.1002/mrm.27948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  262. Michel E, Hernandez D, Lee S Y. Electrical conductivity and permittivity maps of brain tissues derived from water content based on T1‐weighted acquisition. Magn. Reson. Med. 2017;77:1094–103. doi: 10.1002/mrm.26193. [DOI] [PubMed] [Google Scholar]
  263. Mir L M. Electroporation Protocols: Preclinical and Clinical Gene Medicine. Humana Press; 2008. Application of electroporation gene therapy: past, current, and future; pp. 3–17. [DOI] [PubMed] [Google Scholar]
  264. Miranda P C. Physics of effects of transcranial brain stimulation. Handb. Clin. Neurol. 2013;116:353–66. doi: 10.1016/b978-0-444-53497-2.00029-2. [DOI] [PubMed] [Google Scholar]
  265. Miranda P C, Faria P, Hallett M. What does the ratio of injected current to electrode area tell us about current density in the brain during tDCS? Clin. Neurophysiol. 2009;120:1183–7. doi: 10.1016/j.clinph.2009.03.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  266. Mirbozorgi S A, Yeon P, Ghovanloo M. Robust wireless power transmission to mm-sized free-floating distributed implants. IEEE Trans. Biomed. Circuits Syst. 2017;11:692–702. doi: 10.1109/TBCAS.2017.2663358. [DOI] [PubMed] [Google Scholar]
  267. Mitka M. First MRI-safe pacemaker receives conditional approval from FDA. JAMA. 2011;305:985–6. doi: 10.1001/jama.2011.273. [DOI] [PubMed] [Google Scholar]
  268. Mizuno M, Kitahara H, Sasaki K, Tani M, Kojima M, Suzuki Y, Tasaki T, Tatematsu Y, Fukunari M, Wake K. Dielectric property measurements of corneal tissues for computational dosimetry of the eye in terahertz band in vivo and in vitro. Biomed. Opt. Express. 2021;12:1295–307. doi: 10.1364/BOE.412769. [DOI] [PMC free article] [PubMed] [Google Scholar]
  269. Moche M, et al. Clinical evaluation of in silico planning and real-time simulation of hepatic radiofrequency ablation (ClinicIMPPACT trial) Eur. Radiol. 2020;30:934–42. doi: 10.1007/s00330-019-06411-5. [DOI] [PubMed] [Google Scholar]
  270. Monti G, Arcuti P, Tarricone L. Resonant inductive link for remote powering of pacemakers. IEEE Trans. Microw. Theory Tech. 2015;63:3814–22. doi: 10.1109/TMTT.2015.2481387. [DOI] [Google Scholar]
  271. Mosayebi-Samani M, Cunha T, Eroğlu H H, Siebner H R, Nitsche M A, Thielscher A. The effect of brain tissue anisotropy on the electric field caused by transcranial electric stimulation: sensitivity analysis and magnetic resonance electrical impedance tomography. Imaging Neurosci. 2025;3:imag_a_00481. doi: 10.1162/imag_a_00481. [DOI] [PMC free article] [PubMed] [Google Scholar]
  272. Mosayebi-Samani M, Jamil A, Salvador R, Ruffini G, Haueisen J, Nitsche M A. The impact of individual electrical fields and anatomical factors on the neurophysiological outcomes of tDCS: a TMS-MEP and MRI study. Brain Stimul. 2021;14:316–26. doi: 10.1016/j.brs.2021.01.016. [DOI] [PubMed] [Google Scholar]
  273. Moser P, Reishofer G, Prückl R, Schaffelhofer S, Freigang S, Thumfart S, Mahdy Ali K. Real-time estimation of the optimal coil placement in transcranial magnetic stimulation using multi-task deep learning. Sci. Rep. 2024;14:19361. doi: 10.1038/s41598-024-70367-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
  274. Müller D, Habel U, Brodkin E S, Clemens B, Weidler C. HD-tDCS induced changes in resting-state functional connectivity: insights from EF modeling. Brain Stimul. 2023;16:1722–32. doi: 10.1016/j.brs.2023.11.012. [DOI] [PubMed] [Google Scholar]
  275. Murbach M, Neufeld E, Capstick M, Kainz W, Brunner D O, Samaras T, Pruessmann K P, Kuster N. Thermal tissue damage model analyzed for different whole‐body SAR and scan durations for standard MR body coils. Magn. Reson. Med. 2014;71:421–31. doi: 10.1002/mrm.24671. [DOI] [PubMed] [Google Scholar]
  276. Nagaoka T, Watanabe S, Sakurai K, Kunieda E, Watanabe S, Taki M, Yamanaka Y. Development of realistic high-resolution whole-body voxel models of Japanese adult males and females of average height and weight, and application of models to radio-frequency electromagnetic-field dosimetry. Phys. Med. Biol. 2003;49:1. doi: 10.1088/0031-9155/49/1/001. [DOI] [PubMed] [Google Scholar]
  277. Nahar S, Phan T, Quaiyum F, Ren L, Fathy A E, Kilic O. An electromagnetic model of human vital signs detection and its experimental validation. IEEE J. Emerg. Sel. Top. Circuits Syst. 2018;8:338–49. doi: 10.1109/JETCAS.2018.2811339. [DOI] [Google Scholar]
  278. Namakshenas P, Arcaini T, Cesare B, Dorato A, Durante E, Ricci M, Santucci D, Saccomandi P, Faiella E. Performance of an anti-phase technology-powered microwave ablation system on ex vivo liver, lung and kidney: analysis of temperature trend, ablation size and sphericity. Cardiovasc. Interv. Radiol. 2024;47:1392–401. doi: 10.1007/s00270-024-03811-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  279. Narayan A, Liu Z, Bergquist J A, Charlebois C, Rampersad S, Rupp L, Brooks D, White D, Tate J, MacLeod R S. UncertainSCI: uncertainty quantification for computational models in biomedicine and bioengineering. Comput. Biol. Med. 2023;152:106407. doi: 10.1016/j.compbiomed.2022.106407. [DOI] [PMC free article] [PubMed] [Google Scholar]
  280. Navarro de Lara L I, Daneshzand M, Mascarenas A, Paulson D, Pratt K, Okada Y, Raij T, Makarov S N, Nummenmaa A. A 3-axis coil design for multichannel TMS arrays. Neuroimage. 2021;224:117355. doi: 10.1016/j.neuroimage.2020.117355. [DOI] [PMC free article] [PubMed] [Google Scholar]
  281. Neufeld E, Gosselin M-C, Murbach M, Christ A, Cabot E, Kuster N. Analysis of the local worst-case SAR exposure caused by an MRI multi-transmit body coil in anatomical models of the human body. Phys. Med. Biol. 2011;56:4649. doi: 10.1088/0031-9155/56/15/002. [DOI] [PubMed] [Google Scholar]
  282. Niemann F, Riemann S, Hubert A-K, Antonenko D, Thielscher A, Martin A K, Unger N, Flöel A, Meinzer M. Electrode positioning errors reduce current dose for focal tDCS set-ups: evidence from individualized electric field mapping. Clin. Neurophysiol. 2024;162:201–9. doi: 10.1016/j.clinph.2024.03.031. [DOI] [PubMed] [Google Scholar]
  283. Nieminen J O, et al. Multi-locus transcranial magnetic stimulation system for electronically targeted brain stimulation. Brain Stimul. 2022;15:116–24. doi: 10.1016/j.brs.2021.11.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  284. Nitsche M, Paulus W. Excitability changes induced in the human motor cortex by weak transcranial direct current stimulation. J. Physiol. 2000;527:633–9. doi: 10.1111/j.1469-7793.2000.t01-1-00633.x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  285. Nocera A, Senigagliesi L, Raimondi M, Ciattaglia G, Gambi E. Machine learning in radar-based physiological signals sensing: a scoping review of the models, datasets and metrics. IEEE Access. 2024;12:156082–117. doi: 10.1109/ACCESS.2024.3482690. [DOI] [Google Scholar]
  286. Nouwens S A N, Paulides M M, Folker J, VilasBoas-Ribeiro I, de Jager B, Heemels W. Integrated thermal and magnetic susceptibility modeling for air-motion artifact correction in proton resonance frequency shift thermometry. Int. J. Hyperth. 2022;39:967–76. doi: 10.1080/02656736.2022.2094475. [DOI] [PubMed] [Google Scholar]
  287. Nyenhuis J A, Park S-M, Kamondetdacha R, Amjad A, Shellock F G, Rezai A R. MRI and implanted medical devices: basic interactions with an emphasis on heating. IEEE Trans. Device Mater. Reliab. 2005;5:467–80. doi: 10.1109/TDMR.2005.859033. [DOI] [Google Scholar]
  288. O’Reardon J P, et al. Efficacy and safety of transcranial magnetic stimulation in the acute treatment of major depression: a multisite randomized controlled trial. Biol. Psychiatry. 2007;62:1208–16. doi: 10.1016/j.biopsych.2007.01.018. [DOI] [PubMed] [Google Scholar]
  289. Öcal O, et al. Predicting liver ablation volumes with real-time MRI thermometry. JHEP Rep. 2024;6:101199. doi: 10.1016/j.jhepr.2024.101199. [DOI] [PMC free article] [PubMed] [Google Scholar]
  290. Ohguri T, Harima Y, Imada H, Sakurai H, Ohno T, Hiraki Y, Tuji K, Tanaka M, Terashima H. Relationships between thermal dose parameters and the efficacy of definitive chemoradiotherapy plus regional hyperthermia in the treatment of locally advanced cervical cancer: data from a multicentre randomised clinical trial. Int. J. Hyperth. 2018;34:461–8. doi: 10.1080/02656736.2017.1352105. [DOI] [PubMed] [Google Scholar]
  291. Opitz A, et al. Spatiotemporal structure of intracranial electric fields induced by transcranial electric stimulation in humans and nonhuman primates. Sci. Rep. 2016;6:31236. doi: 10.1038/srep31236. [DOI] [PMC free article] [PubMed] [Google Scholar]
  292. Opitz A, Paulus W, Will S, Antunes A, Thielscher A, Will A, Thielscher A, Will S, Antunes A, Thielscher A. Determinants of the electric field during transcranial direct current stimulation. NeuroImage. 2015;109:140–50. doi: 10.1016/j.neuroimage.2015.01.033. [DOI] [PubMed] [Google Scholar]
  293. Opitz A, Yeagle E, Thielscher A, Schroeder C, Mehta A D, Milham MP. On the importance of precise electrode placement for targeted transcranial electric stimulation. NeuroImage. 2018;181:560–7. doi: 10.1016/j.neuroimage.2018.07.027. [DOI] [PMC free article] [PubMed] [Google Scholar]
  294. Oyamada Y, Koshisaka T, Sakamoto T. Experimental demonstration of accurate noncontact measurement of arterial pulse wave displacements using 79-GHz array radar. IEEE Sens. J. 2021;21:9128–37. doi: 10.1109/JSEN.2021.3052602. [DOI] [Google Scholar]
  295. Ozheredov I, Prokopchuk M, Mischenko M, Safonova T, Solyankin P, Larichev A, Angeluts A, Balakin A, Shkurinov A. In vivo THz sensing of the cornea of the eye. Laser Phys. Lett. 2018;15:055601. doi: 10.1088/1612-202X/aaac76. [DOI] [Google Scholar]
  296. Palm U, Kumpf U, Behler N, Wulf L, Kirsch B, Wörsching J, Keeser D, Hasan A, Padberg F. Home use, remotely supervised,and remotely controlled transcranial direct current stimulation: a systematic review of the available evidence. Neuromodulation Technol. Neural Interface. 2018;21:323–33. doi: 10.1111/ner.12686. [DOI] [PubMed] [Google Scholar]
  297. Paniccia A, Rozner M, Jones E L, Townsend N T, Varosy P D, Dunning J E, Girard G, Weyer C, Stiegmann G V, Robinson T N. Electromagnetic interference caused by common surgical energy-based devices on an implanted cardiac defibrillator. Am. J. Surg. 2014;208:932–6. doi: 10.1016/j.amjsurg.2014.09.011. [DOI] [PubMed] [Google Scholar]
  298. Parazzini M, Fiocchi S, Cancelli A, Cottone C, Liorni I, Ravazzani P, Tecchio F. A computational model of the electric field distribution due to regional personalized or nonpersonalized electrodes to select transcranial electric stimulation target. IEEE Trans. Biomed. Eng. 2017;64:184–95. doi: 10.1109/TBME.2016.2553177. [DOI] [PubMed] [Google Scholar]
  299. Paterniani G, Sgreccia D, Davoli A, Guerzoni G, Di Viesti P, Valenti A C, Vitolo M, Vitetta G M, Boriani G. Radar-based monitoring of vital signs: a tutorial overview. Proc. IEEE. 2023;111:277–317. doi: 10.1109/JPROC.2023.3244362. [DOI] [Google Scholar]
  300. Paulides M, Bakker J, Linthorst M, Van der Zee J, Rijnen Z, Neufeld E, Pattynama P, Jansen P, Levendag P, Van Rhoon G. The clinical feasibility of deep hyperthermia treatment in the head and neck: new challenges for positioning and temperature measurement. Phys. Med. Biol. 2010;55:2465. doi: 10.1088/0031-9155/55/9/003. [DOI] [PubMed] [Google Scholar]
  301. Paulides M M, Drizdal T, Van Rhoon G C, Yeo D. Novel applicator design for MR guided RF hyperthermia in head and neck cancers: heating performance and RF coupling. International Society for Magnetic Resonance in Medicine (ISMRM); Paris, France. 2018. [Google Scholar]
  302. Paulides M M, Rodrigues D B, Bellizzi G G, Sumser K, Curto S, Neufeld E, Montanaro H, Kok H P, Dobsicek Trefna H. ESHO benchmarks for computational modeling and optimization in hyperthermia therapy. Int. J. Hyperth. 2021;38:1425–42. doi: 10.1080/02656736.2021.1979254. [DOI] [PubMed] [Google Scholar]
  303. Paulides M M, Stauffer P R, Neufeld E, Maccarini P F, Kyriakou A, Canters R A, Diederich C J, Bakker J F, Van Rhoon G C. Simulation techniques in hyperthermia treatment planning. Int. J. Hyperth. 2013;29:346–57. doi: 10.3109/02656736.2013.790092. [DOI] [PMC free article] [PubMed] [Google Scholar]
  304. Pearce J A. Comparative analysis of mathematical models of cell death and thermal damage processes. Int. J. Hyperth. 2013;29:262–80. doi: 10.3109/02656736.2013.786140. [DOI] [PubMed] [Google Scholar]
  305. Pendse M, Stara R, Mehdi Khalighi M, Rutt B. IMPULSE: a scalable algorithm for design of minimum specific absorption rate parallel transmit RF pulses. Magn. Reson. Med. 2019;81:2808–22. doi: 10.1002/mrm.27589. [DOI] [PMC free article] [PubMed] [Google Scholar]
  306. Peng J, Hao Z, Zhang Z, Wang R, Li M, Dang X. Utter innocence: contactless authentication method based on physiological signals. IEEE Sens. J. 2024;24:7740–51. doi: 10.1109/JSEN.2024.3353256. [DOI] [Google Scholar]
  307. Pennes H H. Analysis of tissue and arterial blood temperatures in the resting human forearm. J. Appl. Physiol. 1948;1:93–122. doi: 10.1152/jappl.1948.1.2.93. [DOI] [PubMed] [Google Scholar]
  308. Pfeiffer D, Tebbenjohanns J, Schumacher B, Jung W, Lüderitz B. Pacemaker function during radiofrequency ablation. Pacing Clin. Electrophysiol. 1995;18:1037–44. doi: 10.1111/j.1540-8159.1995.tb04746.x. [DOI] [PubMed] [Google Scholar]
  309. Pickwell E, Cole B, Fitzgerald A, Pepper M, Wallace V. In vivo study of human skin using pulsed terahertz radiation. Phys. Med. Biol. 2004;49:1595–607. doi: 10.1088/0031-9155/49/9/001. [DOI] [PubMed] [Google Scholar]
  310. Pickwell E, Wallace V. Biomedical applications of terahertz technology. J. Phys. D: Appl. Phys. 2006;39:R301. doi: 10.1088/0022-3727/39/17/R01. [DOI] [Google Scholar]
  311. Piriyajitakonkij M, Warin P, Lakhan P, Leelaarporn P, Kumchaiseemak N, Suwajanakorn S, Pianpanit T, Niparnan N, Mukhopadhyay S C, Wilaiprasitporn T. SleepPoseNet: multi-view learning for sleep postural transition recognition using UWB. IEEE J. Biomed. Health Inform. 2020;25:1305–14. doi: 10.1109/JBHI.2020.3025900. [DOI] [PubMed] [Google Scholar]
  312. Poon A S, O’Driscoll S, Meng T H. Optimal frequency for wireless power transmission into dispersive tissue. IEEE Trans. Antennas Propag. 2010;58:1739–50. doi: 10.1109/TAP.2010.2044310. [DOI] [Google Scholar]
  313. Poorter J D, Wagter C D, Deene Y D, Thomsen C, Ståhlberg F, Achten E. Noninvasive MRI thermometry with the proton resonance frequency (PRF) Magn. Reson. Med. 1995;33:74–81. doi: 10.1002/mrm.1910330111. [DOI] [PubMed] [Google Scholar]
  314. Qi X, Bertling K, Torniainen J, Kong F, Gillespie T, Primiero C, Stark M, Dean P, Indjin D, Li L. Terahertz in vivo imaging of human skin: toward detection of abnormal skin pathologies. APL Bioeng. 2024;8:016117. doi: 10.1063/5.0190573. [DOI] [PMC free article] [PubMed] [Google Scholar]
  315. Qi Z, et al. A wearable repetitive transcranial magnetic stimulation device. Nat. Commun. 2025a;16:2731. doi: 10.1038/s41467-025-58095-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  316. Qi Z, et al. Enabling electric field model of microscopically realistic brain. Brain Stimul. 2025b;18:77–93. doi: 10.1016/j.brs.2024.12.1192. [DOI] [PMC free article] [PubMed] [Google Scholar]
  317. Qian X-H, Wu Y-C, Yang T-Y, Cheng C-H, Chu H-C, Cheng W-H, Yen T-Y, Lin T-H, Lin Y-J, Lee Y-C. Design and in vivo verification of a CMOS bone-guided cochlear implant microsystem. IEEE Trans. Biomed. Eng. 2019;66:3156–67. doi: 10.1109/TBME.2019.2901374. [DOI] [PubMed] [Google Scholar]
  318. Radmilović-Radjenović M, Bošković N, Radjenović B. Computational modeling of microwave tumor ablation. Bioengineering. 2022;9:656. doi: 10.3390/bioengineering9110656. [DOI] [PMC free article] [PubMed] [Google Scholar]
  319. RamRakhyani A K, Lazzi G. Multicoil telemetry system for compensation of coil misalignment effects in implantable systems. IEEE Antennas Wireless Propag. Lett. 2012;11:1675–8. doi: 10.1109/LAWP.2013.2238993. [DOI] [Google Scholar]
  320. Rashed E A, Gomez-Tames J, Hirata A. Deep learning-based development of personalized human head model with non-uniform conductivity for brain stimulation. IEEE Trans. Med. Imaging. 2020;39:2351–62. doi: 10.1109/TMI.2020.2969682. [DOI] [PubMed] [Google Scholar]
  321. Razza L B, De Smet S, Cornelis X, Nikolin S, Pulopulos M M, De Raedt R, Brunoni A R, Vanderhasselt M A. Dose‐dependent response of prefrontal transcranial direct current stimulation on the heart rate variability: an electric field modeling study. Psychophysiology. 2024a;61:e14556. doi: 10.1111/psyp.14556. [DOI] [PubMed] [Google Scholar]
  322. Razza L B, De Smet S, Van Hoornweder S, De Witte S, Luethi M S, Baeken C, Brunoni A R, Vanderhasselt M-A. Investigating the variability of prefrontal tDCS effects on working memory: an individual E-field distribution study. Cortex. 2024b;172:38–48. doi: 10.1016/j.cortex.2023.10.025. [DOI] [PubMed] [Google Scholar]
  323. Reilly J P, Diamant A M. Electrostimulation: Theory, Applications, and Computational Model. Artech House; 2011. [Google Scholar]
  324. Rhee J, Shin Y, Woo S, Lee C, Kim D, Ahn J, Kim H, Ahn S. Wireless torque and power transfer using multiple coils with LCC-S topology for implantable medical drug pump. Sensors. 2021;21:8150. doi: 10.3390/s21238150. [DOI] [PMC free article] [PubMed] [Google Scholar]
  325. Ribeiro I V, van Holthe N, van Rhoon G C, Paulides M M. Impact of segmentation detail in hyperthermia treatment planning: comparison between detailed and clinical tissue segmentation. 2018 EMF-Med 1st World Conf. on Biomedical Applications of Electromagnetic Fields; 2018. pp. 1–2. [Google Scholar]
  326. Rieke V, Butts Pauly K. MR thermometry. J. Magn. Reson. Imaging. 2008;27:376–90. doi: 10.1002/jmri.21265. [DOI] [PMC free article] [PubMed] [Google Scholar]
  327. Rissanen I J, Souza V H, Nieminen J O, Koponen L M, Ilmoniemi R J. Advanced pipeline for designing multi-locus TMS coils with current density constraints. IEEE Trans. Biomed. Eng. 2023;70:2025–34. doi: 10.1109/TBME.2023.3234119. [DOI] [PubMed] [Google Scholar]
  328. Robinson P A. Neural field theory of synaptic plasticity. J. Theor. Biol. 2011;285:156–63. doi: 10.1016/j.jtbi.2011.06.023. [DOI] [PubMed] [Google Scholar]
  329. Rodrigues D B, Dobsicek-Trefna H, Curto S, Winter L, Molitoris J K, Vrba J, Vrba D, Sumser K, Paulides M M. In: Principles and Technologies for Electromagnetic Energy Based Therapies. Prakash P, Srimathveeravalli G, editors. Academic Press; 2022. pp. 281–311. [Google Scholar]
  330. Rong Y, Gutierrez R, Mishra K V, Bliss D W. Noncontact vital sign detection with UAV-borne radars: an overview of recent advances. IEEE Veh. Technol. Mag. 2021;16:118–28. doi: 10.1109/MVT.2021.3086442. [DOI] [Google Scholar]
  331. Röschmann P. Radiofrequency penetration and absorption in the human body: limitations to high‐field whole‐body nuclear magnetic resonance imaging. Med. Phys. 1987;14:922–31. doi: 10.1118/1.595995. [DOI] [PubMed] [Google Scholar]
  332. Rossi S, Hallett M, Rossini P M, Pascual-Leone A, Safety of TMS Consensus Group Safety, ethical considerations, and application guidelines for the use of transcranial magnetic stimulation in clinical practice and research. Clin. Neurophysiol. 2009;120:2008–39. doi: 10.1016/j.clinph.2009.08.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  333. Rossmann C, Haemmerich D. Review of temperature dependence of thermal properties, dielectric properties, and perfusion of biological tissues at hyperthermic and ablation temperatures. Crit. Rev. Biomed. Eng. 2014;42:467–92. doi: 10.1615/critrevbiomedeng.2015012486. [DOI] [PMC free article] [PubMed] [Google Scholar]
  334. Ruhnau P, Rufener K, Heinze H-J, Zaehle T. Sailing in a sea of disbelief: in vivo measurements of transcranial electric stimulation in human subcortical structures. Brain Stimul. 2018;11:241–3. doi: 10.1016/j.brs.2017.09.015. [DOI] [PubMed] [Google Scholar]
  335. Sadighi M, Şişman M, Açıkgöz B C, Eroğlu H H, Eyüboğlu B M. Low-frequency conductivity tensor imaging with a single current injection using DT-MREIT. Phys. Med. Biol. 2021;66:055011. doi: 10.1088/1361-6560/abddcf. [DOI] [PubMed] [Google Scholar]
  336. Sakamoto T. Signal separation using a mathematical model of physiological signals for the measurement of heart pulse wave propagation with array radar. IEEE Access. 2020;8:175921–31. doi: 10.1109/ACCESS.2020.3026539. [DOI] [Google Scholar]
  337. Sakamoto T, Imasaka R, Taki H, Sato T, Yoshioka M, Inoue K, Fukuda T, Sakai H. Feature-based correlation and topological similarity for interbeat interval estimation using ultrawideband radar. IEEE Trans. Biomed. Eng. 2015;63:747–57. doi: 10.1109/TBME.2015.2470077. [DOI] [PubMed] [Google Scholar]
  338. Sakamoto T, Iwata I, Minami T, Matsumoto T. Noncontact sensing of animals using radar. IEICE Trans. Electron. 2024;107:279–87. doi: 10.1587/transele.2024MMI0002. [DOI] [Google Scholar]
  339. Sasaki K, Mizuno M, Wake K, Watanabe S. Monte Carlo simulations of skin exposure to electromagnetic field from 10 GHz to 1 THz. Phys. Med. Biol. 2017;62:6993–7010. doi: 10.1088/1361-6560/aa81fc. [DOI] [PubMed] [Google Scholar]
  340. Sasaki K, Porter E, Rashed E A, Farrugia L, Schmid G. Measurement and image-based estimation of dielectric properties of biological tissues—past, present, and future. Phys. Med. Biol. 2022;67:14TR01. doi: 10.1088/1361-6560/ac7b64. [DOI] [PubMed] [Google Scholar]
  341. Saturnino G B, Madsen K H, Thielscher A. Electric field simulations for transcranial brain stimulation using FEM: an efficient implementation and error analysis. J. Neural Eng. 2019a;16:066032. doi: 10.1088/1741-2552/ab41ba. [DOI] [PubMed] [Google Scholar]
  342. Saturnino G B, Puonti O, Nielsen J D, Antonenko D, Madsen K H, Thielscher A. SimNIBS 2.1: a comprehensive pipeline for individualized electric field modelling for transcranial brain stimulation. Brain and Human Body Modeling: Computational Human Modeling at EMBC 2018; 2019b. pp. 3–25. [PubMed] [Google Scholar]
  343. Saturnino G B, Thielscher A, Madsen K H, Knösche T R, Weise K. A principled approach to conductivity uncertainty analysis in electric field calculations. NeuroImage. 2018;188:821–34. doi: 10.1016/j.neuroimage.2018.12.053. [DOI] [PubMed] [Google Scholar]
  344. Saturnino G B, Thielscher A, Madsen K H, Knösch T R, Weise K. A principled approach to conductivity uncertainty analysis in electric field calculations. Neuroimage. 2019c;188:821–34. doi: 10.1016/j.neuroimage.2018.12.053. [DOI] [PubMed] [Google Scholar]
  345. Sauter S A, Schwab C. Boundary Element Methods. Springer; 2010. pp. 183–287. [Google Scholar]
  346. Schenck J F. Safety of strong, static magnetic fields. J. Magn. Reson. Imaging. 2000;12:2–19. doi: 10.1002/1522-2586(200007)12:1&#x0003c;2::AID-JMRI2&#x0003e;3.0.CO;2-V. [DOI] [PubMed] [Google Scholar]
  347. Schneider C, Van Dijk J D. Visualization by a matrix of light-emitting diodes of interference effects from a radiative four-applicator hyperthermia system. Int. J. Hyperth. 1991;7:355–66. doi: 10.3109/02656739109005001. [DOI] [PubMed] [Google Scholar]
  348. Schooneveldt G, et al. Thermal dosimetry for bladder hyperthermia treatment. An overview. Int. J. Hyperth. 2016;32:417–33. doi: 10.3109/02656736.2016.1156170. [DOI] [PubMed] [Google Scholar]
  349. Schueler B A, Parrish T B, Lin J C, Hammer B E, Pangrle B J, Ritenour E R, Kucharczyk J, Truwit C L. MRI compatibility and visibility assessment of implantable medical devices. J. Magn. Reson. Imaging. 1999;9:596–603. doi: 10.1002/(SICI)1522-2586(199904)9:4&#x0003c;596::AID-JMRI14&#x0003e;3.0.CO;2-T. [DOI] [PubMed] [Google Scholar]
  350. Schwan H, Kay C. The conductivity of living tissues. Ann. N.Y. Acad. Sci. 1957;65:1007–13. doi: 10.1111/j.1749-6632.1957.tb36701.x. [DOI] [PubMed] [Google Scholar]
  351. Scott G, Joy M, Armstrong R, Henkelman R. Measurement of nonuniform current density by magnetic resonance. IEEE Trans. Med. Imaging. 1991;10:362–74. doi: 10.1109/42.97586. [DOI] [PubMed] [Google Scholar]
  352. Seifert A-K, Amin M G, Zoubir A M. Toward unobtrusive in-home gait analysis based on radar micro-Doppler signatures. IEEE Trans. Biomed. Eng. 2019;66:2629–40. doi: 10.1109/TBME.2019.2893528. [DOI] [PubMed] [Google Scholar]
  353. Seshadri S, Kavitha M, Bobba P B. Effect of coil structures on performance of a four-coil WPT powered medical implantable devices. IEEE Int. Conf. Power, Instrumentation, Control and Computing; 2018. pp. 1–6. [Google Scholar]
  354. Seyfioğlu M S, Özbayoğlu A M, Gürbüz S Z. Deep convolutional autoencoder for radar-based classification of similar aided and unaided human activities. IEEE Trans. Aerosp. Electron. Syst. 2018;54:1709–23. doi: 10.1109/TAES.2018.2799758. [DOI] [Google Scholar]
  355. Sherar M, et al. Relationship between thermal dose and outcome in thermoradiotherapy treatments for superficial recurrences of breast cancer: data from a phase III trial. Int. J. Radiat. Oncol. Biol. Phys. 1997;39:371–80. doi: 10.1016/S0360-3016(97)00333-7. [DOI] [PubMed] [Google Scholar]
  356. Si P, Hu A P, Malpas S, Budgett D. A frequency control method for regulating wireless power to implantable devices. IEEE Trans. Biomed. Circuits Syst. 2008;2:22–29. doi: 10.1109/TBCAS.2008.918284. [DOI] [PubMed] [Google Scholar]
  357. Siebner H R, et al. Transcranial magnetic stimulation of the brain: what is stimulated?—A consensus and critical position paper. Clin. Neurophysiol. 2022;140:59–97. doi: 10.1016/j.clinph.2022.04.022. [DOI] [PMC free article] [PubMed] [Google Scholar]
  358. Simula S, Daoud M, Ruffini G, Biagi M C, Benar C-G, Benquet P, Wendling F, Bartolomei F. Transcranial current stimulation in epilepsy: a systematic review of the fundamental and clinical aspects. Front. Neurosci. 2022;16:909421. doi: 10.3389/fnins.2022.909421. [DOI] [PMC free article] [PubMed] [Google Scholar]
  359. Singh A, Rehman S U, Yongchareon S, Chong P H J. Multi-resident non-contact vital sign monitoring using radar: a review. IEEE Sens. J. 2020;21:4061–84. doi: 10.1109/JSEN.2020.3036039. [DOI] [Google Scholar]
  360. Singh M. Modified Pennes bioheat equation with heterogeneous blood perfusion: a newer perspective. Int. J. Heat Mass Transfer. 2024;218:124698. doi: 10.1016/j.ijheatmasstransfer.2023.124698. [DOI] [Google Scholar]
  361. Soldati M, Laakso I. Computational errors of the induced electric field in voxelized and tetrahedral anatomical head models exposed to spatially uniform and localized magnetic fields. Phys. Med. Biol. 2020;65:015001. doi: 10.1088/1361-6560/ab5dfb. [DOI] [PubMed] [Google Scholar]
  362. Stenroos M, Koponen L M. Real-time computation of the TMS-induced electric field in a realistic head model. Neuroimage. 2019;203:116159. doi: 10.1016/j.neuroimage.2019.116159. [DOI] [PubMed] [Google Scholar]
  363. Strigari L, Minosse S, D’Alessio D, Farina L, Cavagnaro M, Cassano B, Pinto R, Vallati G, LoPresto V. Microwave thermal ablation using CT-scanner for predicting the variation of ablated region over time: advantages and limitations. Phys. Med. Biol. 2019;64:115021. doi: 10.1088/1361-6560/ab1a67. [DOI] [PubMed] [Google Scholar]
  364. Stroka S, Haussmann N, Kasolis F, Clemens M. 2024 Int. Symp. on Electromagnetic Compatibility–EMC Europe,2024) vol. Series); IEEE; pp. 762–7. [Google Scholar]
  365. Stuchly M A, Stuchly S S. Coaxial line reflection methods for measuring dielectric properties of biological substances at radio and microwave frequencies-a review. IEEE Trans. Instrum. Meas. 1980;29:176–83. doi: 10.1109/TIM.1980.4314902. [DOI] [Google Scholar]
  366. Sumser K, Bellizzi G G, Forner R, Drizdal T, Tamames J A H, van Rhoon G C, Paulides M M. Dual-function MR-guided hyperthermia: an innovative integrated approach and experimental demonstration of proof of principle. IEEE Trans. Biomed. Eng. 2021;68:712–7. doi: 10.1109/TBME.2020.3012734. [DOI] [PubMed] [Google Scholar]
  367. Sung S, Selvin S, Bajwa N, Chantra S, Nowroozi B, Garritano J, Goell J, Li A D, Deng S X, Brown E R. THz imaging system for in vivo human cornea. IEEE Trans. Terahertz Sci. Technol. 2018;8:27–37. doi: 10.1109/TTHZ.2017.2775445. [DOI] [PMC free article] [PubMed] [Google Scholar]
  368. Taflove A, Hagness S. 3rd edn. Artech House; 2003. Computational Electrodynamics: the Finite-Difference Time-Domain Method. [Google Scholar]
  369. Tan E T, Hua Y, Fiveland E W, Vermilyea M E, Piel J E, Park K J, Ho V B, Foo T K. Peripheral nerve stimulation limits of a high amplitude and slew rate magnetic field gradient coil for neuroimaging. Magn. Reson. Med. 2020;83:352–66. doi: 10.1002/mrm.27909. [DOI] [PMC free article] [PubMed] [Google Scholar]
  370. Tanabe Y, Ho J S, Liu J, Liao S-Y, Zhen Z, Hsu S, Shuto C, Zhu Z-Y, Ma A, Vassos C. High-performance wireless powering for peripheral nerve neuromodulation systems. PLoS One. 2017;12:e0186698. doi: 10.1371/journal.pone.0186698. [DOI] [PMC free article] [PubMed] [Google Scholar]
  371. Tang L, Hue Y K, Ibrahim T S. Studies of RF shimming techniques with minimization of RF power deposition and their associated temperature changes. Concepts Magn. Reson. B. 2011;39:11–25. doi: 10.1002/cmr.b.20185. [DOI] [PMC free article] [PubMed] [Google Scholar]
  372. Tang S C, Lun T L T, Guo Z, Kwok K-W, McDannold N J. Intermediate range wireless power transfer with segmented coil transmitters for implantable heart pumps. IEEE Trans. Power Electron. 2017;32:3844–57. doi: 10.1109/TPEL.2016.2584558. [DOI] [Google Scholar]
  373. Taylor Z D, Garritano J, Sung S, Bajwa N, Bennett D B, Nowroozi B, Tewari P, Sayre J, Hubschman J-P, Deng S. THz and mm-wave sensing of corneal tissue water content: electromagnetic modeling and analysis. IEEE Trans. Terahertz Sci. Technol. 2015;5:170–83. doi: 10.1109/TTHZ.2015.2392619. [DOI] [PMC free article] [PubMed] [Google Scholar]
  374. Therrien-Blanchet J-M, Ferland M C, Badri M, Rousseau M-A, Merabtine A, Boucher E, Hofmann L H, Lepage J-F, Théoret H. The neurophysiological aftereffects of brain stimulation in human primary motor cortex: a Sham-controlled comparison of three protocols. Cereb. Cortex. 2023;33:7061–75. doi: 10.1093/cercor/bhad021. [DOI] [PubMed] [Google Scholar]
  375. Thielscher A, Antunes A, Saturnino G B. Field modeling for transcranial magnetic stimulation: a useful tool to understand the physiological effects of TMS?. Conf. Proc. -IEEE Eng. in Medicine and Biology Society; 2015. pp. 222–5. [DOI] [PubMed] [Google Scholar]
  376. Toyoshima T. Electromagnetic interference of implantable cardiac pacemaker by portable telephones. Cardiac. Pacing. 1996;12:488. [Google Scholar]
  377. Tucci C, Trujillo M, Berjano E, Iasiello M, Andreozzi A, Vanoli G P. Pennes’ bioheat equation vs. porous media approach in computer modeling of radiofrequency tumor ablation. Sci. Rep. 2021;11:5272. doi: 10.1038/s41598-021-84546-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  378. U.S. FDA Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment: Guidance for Industry and FDA Staff. 2023 FDA-2019-D-2837. [Google Scholar]
  379. Uddin S D, Hossain M S, Islam S M, Lubecke V. Heart rate variability-based obstructive sleep apnea events classification using microwave Doppler radar. IEEE J. Electromagn. RF Microw. Med. Biol. 2023;7:416–24. doi: 10.1109/JERM.2023.3317304. [DOI] [Google Scholar]
  380. van den Bergen B, van den Berg C A, Klomp D W, Lagendijk J J. SAR and power implications of different RF shimming strategies in the pelvis for 7T MRI. J. Magn. Reson. Imaging. 2009;30:194–202. doi: 10.1002/jmri.21806. [DOI] [PubMed] [Google Scholar]
  381. Van Hoornweder S, Caulfield K A, Nitsche M, Thielscher A, Meesen R L. Addressing transcranial electrical stimulation variability through prospective individualized dosing of electric field strength in 300 participants across two samples: the 2-SPED approach. J. Neural Eng. 2022;19:056045. doi: 10.1088/1741-2552/ac9a78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  382. Van Hoornweder S, Nuyts M, Frieske J, Verstraelen S, Meesen R L, Caulfield K A. Outcome measures for electric field modeling in tES and TMS: a systematic review and large-scale modeling study. Neuroimage. 2023;281:120379. doi: 10.1016/j.neuroimage.2023.120379. [DOI] [PMC free article] [PubMed] [Google Scholar]
  383. van Leeuwen C M, Oei A L, Chin K, Crezee J, Bel A, Westermann A M, Buist M R, Franken N A P, Stalpers L J A, Kok H P. A short time interval between radiotherapy and hyperthermia reduces in-field recurrence and mortality in women with advanced cervical cancer. Radiat. Oncol. 2017;12:75. doi: 10.1186/s13014-017-0813-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  384. van Rhoon G, Samaras T, Yarmolenko P, Dewhirst M, Neufeld E, Kuster N. CEM43°C thermal dose thresholds: a potential guide for magnetic resonance radiofrequency exposure levels? Eur. Radiol. 2013;23:2215–27. doi: 10.1007/s00330-013-2825-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  385. van Rhoon G C. Is CEM43 still a relevant thermal dose parameter for hyperthermia treatment monitoring? Int. J. Hyperth. 2016;32:50–62. doi: 10.3109/02656736.2015.1114153. [DOI] [PubMed] [Google Scholar]
  386. Van Rhoon G C, Van Der Heuvel D J, Ameziane A, Rietveld P J, Volenec K, Van Der Zee J. Characterization of the SAR-distribution of the Sigma-60 applicator for regional hyperthermia using a Schottky diode sheet. Int. J. Hyperth. 2003;19:642–54. doi: 10.1080/0265673031000140813. [DOI] [PubMed] [Google Scholar]
  387. Verhaart R F, Verduijn G M, Fortunati V, Rijnen Z, van Walsum T, Veenland J F, Paulides M M. Accurate 3D temperature dosimetry during hyperthermia therapy by combining invasive measurements and patient-specific simulations. Int. J. Hyperth. 2015;31:686–92. doi: 10.3109/02656736.2015.1052855. [DOI] [PubMed] [Google Scholar]
  388. Vidjak K, Farina L, Challapalli R S, Quinn A M, O’Halloran M, Lowery A, Ruvio G, Cavagnaro M. Histology-validated electromagnetic characterization of ex-vivo ovine lung tissue for microwave-based medical applications. Sci. Rep. 2024;14:5940. doi: 10.1038/s41598-024-55035-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  389. Vidjak K, Hessinger C, Cavagnaro M. Broadband dielectric spectroscopy with a microwave ablation antenna. Sensors. 2023;23:2579. doi: 10.3390/s23052579. [DOI] [PMC free article] [PubMed] [Google Scholar]
  390. VilasBoas-Ribeiro I, Curto S, van Rhoon G C, Franckena M, Paulides M M. MR thermometry accuracy and prospective imaging-based patient selection in MR-guided hyperthermia treatment for locally advanced cervical cancer. Cancers. 2021;13:3503. doi: 10.3390/cancers13143503. [DOI] [PMC free article] [PubMed] [Google Scholar]
  391. VilasBoas-Ribeiro I, Franckena M, van Rhoon G C, Hernandez-Tamames J A, Paulides M M. Using MRI to measure position and anatomy changes and assess their impact on the accuracy of hyperthermia treatment planning for cervical cancer. Int. J. Hyperth. 2023;40:2151648. doi: 10.1080/02656736.2022.2151648. [DOI] [PubMed] [Google Scholar]
  392. VilasBoas-Ribeiro I, Nouwens S A N, Curto S, Jager B, Franckena M, van Rhoon G C, Heemels W, Paulides M M. POD-Kalman filtering for improving noninvasive 3D temperature monitoring in MR-guided hyperthermia. Med. Phys. 2022;49:4955–70. doi: 10.1002/mp.15811. [DOI] [PMC free article] [PubMed] [Google Scholar]
  393. VilasBoas-Ribeiro I, Sumser K, Nouwens S, Feddersen T, Heemels W, van Rhoon G C, Paulides M M. Adapting temperature predictions to MR imaging in treatment position to improve simulation-guided hyperthermia for cervical cancer. IEEE Open J. Eng. Med. Biol. 2024;5:99–106. doi: 10.1109/OJEMB.2023.3321990. [DOI] [PMC free article] [PubMed] [Google Scholar]
  394. Violante I R, Alania K, Cassarà A M, Neufeld E, Acerbo E, Carron R, Williamson A, Kurtin D L, Rhodes E, Hampshire A. Non-invasive temporal interference electrical stimulation of the human hippocampus. Nat. Neurosci. 2023;26:1994–2004. doi: 10.1038/s41593-023-01456-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  395. Virk A S, Harris Z B, Arbab M H. Design and characterization of a hyperbolic-elliptical lens pair in a rapid beam steering system for single-pixel terahertz spectral imaging of the cornea. Opt. Express. 2023;31:39568–82. doi: 10.1364/OE.496894. [DOI] [PubMed] [Google Scholar]
  396. Vu J, Nguyen B T, Bhusal B, Baraboo J, Rosenow J, Bagci U, Bright M G, Golestanirad L. Machine learning-based prediction of MRI-induced power absorption in the tissue in patients with simplified deep brain stimulation lead models. IEEE Trans. Electromagn. Compat. 2021;63:1757–66. doi: 10.1109/TEMC.2021.3106872. [DOI] [PMC free article] [PubMed] [Google Scholar]
  397. Waki S, Nishikawa T, Hikage T, Yamamoto M. Interference voltage estimation for implantable cardiac pacemaker electromagnetic interference at fourth-generation (4G) and fifth-generation (5G) Sub-6 GHz cellular frequency bands. URSI Radio Sci. Lett. 2023;4:47. doi: 10.46620/22-0047. [DOI] [Google Scholar]
  398. Wang B, Peterchev A V, Gaugain G, Ilmoniemi R J, Grill W M, Bikson M, Nikolayev D. Quasistatic approximation in neuromodulation. J. Neural Eng. 2024;21:041002. doi: 10.1088/1741-2552/ad625e. [DOI] [PMC free article] [PubMed] [Google Scholar]
  399. Wang F, Skubic M, Rantz M, Cuddihy P E. Quantitative gait measurement with pulse-Doppler radar for passive in-home gait assessment. IEEE Trans. Biomed. Eng. 2014;61:2434–43. doi: 10.1109/TBME.2014.2319333. [DOI] [PMC free article] [PubMed] [Google Scholar]
  400. Wang J, Fujiwara O, Nojima T. A model for predicting electromagnetic interference of implanted cardiac pacemakers by mobile telephones. IEEE Trans. Microw. Theory Tech. 2000;48:2121–5. doi: 10.1109/22.884203. [DOI] [Google Scholar]
  401. Wang J, Wang X, Zhu Z, Huangfu J, Li C, Ran L. 1-D microwave imaging of human cardiac motion: an ab-initio investigation. IEEE Trans. Microw. Theory Tech. 2013;61:2101–7. doi: 10.1109/TMTT.2013.2252186. [DOI] [Google Scholar]
  402. Wang L, Fu R, Xu C, Xu M. Methods and applications of full-field optical coherence tomography: a review. J. Biomed. Opt. 2022a;27:050901. doi: 10.1117/1.JBO.27.5.050901. [DOI] [PMC free article] [PubMed] [Google Scholar]
  403. Wang M, Crocco L, Costanzo S, Scapaticci R, Cavagnaro M. A compact slot-loaded antipodal vivaldi antenna for a microwave imaging system to monitor liver microwave thermal ablation. IEEE Open J. Antennas Propag. 2022b;3:700–8. doi: 10.1109/OJAP.2022.3183750. [DOI] [Google Scholar]
  404. Wang M, Liu H, Zhang P, Zhang X, Yang H, Zhou G, Li L. Broadband implantable antenna for wireless power transfer in cardiac pacemaker applications. IEEE J. Electromagn. RF Microw. Med. Biol. 2020a;5:2–8. doi: 10.1109/JERM.2020.2999205. [DOI] [Google Scholar]
  405. Wang P, Brace C L. Tissue dielectric measurement using an interstitial dipole antenna. IEEE Trans. Biomed. Eng. 2012;59:115–21. doi: 10.1109/TBME.2011.2167622. [DOI] [PMC free article] [PubMed] [Google Scholar]
  406. Wang W, Eisenberg S R. A three-dimensional finite element method for computing magnetically induced currents in tissues. IEEE Trans. Mag. 1994;30:5015–23. doi: 10.1109/20.334289. [DOI] [Google Scholar]
  407. Wang Y, Song S, Zheng J, Wang Q, Long S A, Kainz W, Chen J. A novel device model validation strategy for 1.5- and 3-T MRI heating safety assessment. IEEE Trans. Instrum. Meas. 2020b;69:6381–9. doi: 10.1109/TIM.2020.2971292. [DOI] [Google Scholar]
  408. Wang Y, Zheng J, Guo R, Wang Q, Kainz W, Long S, Chen J. A technique for the reduction of RF‐induced heating of active implantable medical devices during MRI. Magn. Reson. Med. 2022c;87:349–64. doi: 10.1002/mrm.28953. [DOI] [PubMed] [Google Scholar]
  409. Wang Z, Chang T, Cui H-L. Review of active millimeter wave imaging techniques for personnel security screening. IEEE Access. 2019;7:148336–50. doi: 10.1109/ACCESS.2019.2946736. [DOI] [Google Scholar]
  410. Weiland J D, Humayun M S. Retinal prosthesis. IEEE Trans. Biomed. Eng. 2014;61:1412–24. doi: 10.1109/TBME.2014.2314733. [DOI] [PMC free article] [PubMed] [Google Scholar]
  411. Weise K, Worbs T, Kalloch B, Souza V H, Jacqier A T, Van Geit W, Thielscher A, Knösche T R. Directional sensitivity of cortical neurons towards TMS induced electric fields. Imaging Neurosci. 2023;1:1–22. doi: 10.1162/imag_a_00036. [DOI] [PMC free article] [PubMed] [Google Scholar]
  412. Wells J, et al. Challenges and recommendations for magnetic hyperthermia characterization measurements. Int. J. Hyperth. 2021;38:447–60. doi: 10.1080/02656736.2021.1892837. [DOI] [PubMed] [Google Scholar]
  413. Wiethoff S, Hamada M, Rothwell J C. Variability in response to transcranial direct current stimulation of the motor cortex. Brain Stimul. 2014;7:468–75. doi: 10.1016/j.brs.2014.02.003. [DOI] [PubMed] [Google Scholar]
  414. Wilson N A, Reich A J, Graham J, Bhatt D L, Nguyen L L, Weissman J S. Patient perspectives on the need for implanted device information: implications for a post‐procedural communication framework. Health Expect. 2021;24:1391–402. doi: 10.1111/hex.13273. [DOI] [PMC free article] [PubMed] [Google Scholar]
  415. Winter L, Oberacker E, Paul K, Ji Y, Oezerdem C, Ghadjar P, Thieme A, Budach V, Wust P, Niendorf T. Magnetic resonance thermometry: methodology, pitfalls and practical solutions. Int. J. Hyperth. 2016;32:63–75. doi: 10.3109/02656736.2015.1108462. [DOI] [PubMed] [Google Scholar]
  416. Winter L, Seifert F, Zilberti L, Murbach M, Ittermann B. MRI‐related heating of implants and devices: a review. J. Magn. Reson. Imaging. 2021;53:1646–65. doi: 10.1002/jmri.27194. [DOI] [PubMed] [Google Scholar]
  417. Wischnewski M, Alekseichuk I, Opitz A. Neurocognitive, physiological, and biophysical effects of transcranial alternating current stimulation. Trends Cogn. Sci. 2023;27:189–205. doi: 10.1016/j.tics.2022.11.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  418. Wischnewski M, Mantell K E, Opitz A. Identifying regions in prefrontal cortex related to working memory improvement: a novel meta-analytic method using electric field modeling. Neurosci. Biobehav. Rev. 2021;130:147–61. doi: 10.1016/j.neubiorev.2021.08.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  419. Wischnewski M, Shirinpour S, Alekseichuk I, Lapid M I, Nahas Z, Lim K O, Croarkin P E, Opitz A. Real-time TMS-EEG for brain state-controlled research and precision treatment: a narrative review and guide. J. Neural Eng. 2024;21:061001. doi: 10.1088/1741-2552/ad8a8e. [DOI] [PMC free article] [PubMed] [Google Scholar]
  420. Wolters C H, Anwander A, Tricoche X, Weinstein D, Koch M A, Macleod R S. Influence of tissue conductivity anisotropy on EEG/MEG field and return current computation in a realistic head model: a simulation and visualization study using high-resolution finite element modeling. NeuroImage. 2006;30:813–26. doi: 10.1016/j.neuroimage.2005.10.014. [DOI] [PubMed] [Google Scholar]
  421. Woods T O. Standards for medical devices in MRI: present and future. J. Magn. Reson. Imaging. 2007;26:1186–9. doi: 10.1002/jmri.21140. [DOI] [PubMed] [Google Scholar]
  422. Wust P, Fahling H, Jordan A, Nadobny J, Seebass M, Felix R. Development and testing of SAR-visualizing phantoms for quality control in RF hyperthermia. Int. J. Hyperth. 1994;10:127–42. doi: 10.3109/02656739409009338. [DOI] [PubMed] [Google Scholar]
  423. Xiong Z, Feng S, Kautz R, Chandra S, Altunyurt N, Chen J. Multi-GPU accelerated admittance method for high-resolution human exposure evaluation. IEEE Trans. Biomed. Eng. 2015;62:2920–30. doi: 10.1109/TBME.2015.2453169. [DOI] [PubMed] [Google Scholar]
  424. Yang D, Converse M C, Mahvi D M, Webster J G. Measurement and analysis of tissue temperature during microwave liver ablation. IEEE Trans. Biomed. Eng. 2007;54:150–5. doi: 10.1109/TBME.2006.884647. [DOI] [PubMed] [Google Scholar]
  425. Yang L, Yang X, Ye H, Kaula N, Jin Y, Zheng J, Kainz W, Chen J. Computational study of the effects of orthopedic plates on gradient‐induced peripheral nerve stimulation under MRI using electromagnetic and neurophysiological modeling. Magn. Reson. Med. 2025;93:2583–94. doi: 10.1002/mrm.30470. [DOI] [PubMed] [Google Scholar]
  426. Yang S, Le Kernec J, Romain O, Fioranelli F, Cadart P, Fix J, Ren C, Manfredi G, Letertre T, Sáenz I D H. The human activity radar challenge: benchmarking based on the ‘radar signatures of human activities’ dataset from glasgow university. IEEE J. Biomed. Health Inform. 2023;27:1813–24. doi: 10.1109/JBHI.2023.3240895. [DOI] [PubMed] [Google Scholar]
  427. Yao A, Zastrow E, Kuster N. Data-driven experimental evaluation method for the safety assessment of implants with respect to RF-induced heating during MRI. Radio Sci. 2018;53:700–9. doi: 10.1029/2017RS006433. [DOI] [Google Scholar]
  428. Yarmolenko P S, Moon E J, Landon C, Manzoor A, Hochman D W, Viglianti B L, Dewhirst M W. Thresholds for thermal damage to normal tissues: an update. Int. J. Hyperth. 2011;27:320–43. doi: 10.3109/02656736.2010.534527. [DOI] [PMC free article] [PubMed] [Google Scholar]
  429. Yin H, Yu S, Zhang Y, Zhou A, Wang X, Liu L, Ma H, Liu J, Yang N. Let iot know you better: user identification and emotion recognition through millimeter-wave sensing. IEEE Int. Things J. 2022;10:1149–61. doi: 10.1109/JIOT.2022.3204779. [DOI] [Google Scholar]
  430. Yokota T, Maki T, Nagata T, Murakami T, Ugawa Y, Laakso I, Hirata A, Hontani H. Real-time estimation of electric fields induced by transcranial magnetic stimulation with deep neural networks. Brain Stimul. 2019;12:1500–7. doi: 10.1016/j.brs.2019.06.015. [DOI] [PubMed] [Google Scholar]
  431. Young J J, Hernandez-Serrano A I, Hardwicke J, Pickwell-MacPherson E. In vivo terahertz sensing of psoriasis and eczema patients. Sci. Rep. 2024;14:17546. doi: 10.1038/s41598-024-68106-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  432. Zaltieri M, Massaroni C, Cauti F M, Schena E. Techniques for temperature monitoring of myocardial tissue undergoing radiofrequency ablation treatments: an overview. Sensors. 2021;21:1453. doi: 10.3390/s21041453. [DOI] [PMC free article] [PubMed] [Google Scholar]
  433. Zeng F-G, Rebscher S, Harrison W, Sun X, Feng H. Cochlear implants: system design, integration, and evaluation. IEEE Rev. Biomed. Eng. 2008;1:115–42. doi: 10.1109/RBME.2008.2008250. [DOI] [PMC free article] [PubMed] [Google Scholar]
  434. Zeng Q, Liu J, Angelone L M, Lloyd T, Wedan S, Chen J, Kainz W. Investigation of RF-induced heating near interventional catheters at 1.5 T MRI: a combined modeling and experimental study. IEEE Trans. Electromagn. Compat. 2018a;61:1423–31. doi: 10.1109/TEMC.2018.2862249. [DOI] [Google Scholar]
  435. Zeng Q, Wang Q, Kainz W, Chen J. Impact of RF shimming on RF-induced heating near implantable medical electrodes in a 3T MRI coil. IEEE Trans. Electromagn. Compat. 2018b;62:52–64. doi: 10.1109/TEMC.2018.2877526. [DOI] [Google Scholar]
  436. Zhang H, Li Z, Lee C-K. Transmitter adaptation and wireless power control for capsule endoscopy. IEEE Trans. Power Electron. 2024;39:4884–94. doi: 10.1109/TPEL.2023.3345277. [DOI] [Google Scholar]
  437. Zhang L, Fu C-H, Hong H, Xue B, Gu X, Zhu X, Li C. Non-contact dual-modality emotion recognition system by CW radar and RGB camera. IEEE Sens. J. 2021;21:23198–212. doi: 10.1109/JSEN.2021.3107429. [DOI] [Google Scholar]
  438. Zhang Y, Yang R, Yue Y, Lim E G, Wang Z. An overview of algorithms for contactless cardiac feature extraction from radar signals: advances and challenges. IEEE Trans. Instrum. Meas. 2023;72:1–20. doi: 10.1109/TIM.2023.3239925. [DOI] [PMC free article] [PubMed] [Google Scholar]
  439. Zhao Z, Shirinpour S, Tran H, Wischnewski M, Opitz A. Intensity- and frequency-specific effects of transcranial alternating current stimulation are explained by network dynamics. J. Neural Eng. 2024;21:026024. doi: 10.1088/1741-2552/ad37d9. [DOI] [PubMed] [Google Scholar]
  440. Zhao Y, Yarovoy A, Fioranelli F. Angle-insensitive human motion and posture recognition based on 4D imaging radar and deep learning classifiers. IEEE Sens. J. 2022;22:12173–82. doi: 10.1109/JSEN.2022.3175618. [DOI] [Google Scholar]
  441. Zheng C, Chen X, Nguyen B T, Sanpitak P, Vu J, Bagci U, Golestanirad L. Predicting RF heating of conductive leads during magnetic resonance imaging at 1.5 T: a machine learning approach. 43rd Annual Int’l Conf. IEEE Eng. in Medicine & Biology Society; 2021. pp. 4204–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  442. Zheng T, Kang N, Lee C H, Shao L. Wireless powered capsule robots with a wide locomotion range and random orientation via planar transmitting coils. IEEE Robot. Autom. Lett. 2025;10:2502–9. doi: 10.1109/LRA.2025.3531149. [DOI] [Google Scholar]
  443. Zhu S, Guendel R G, Yarovoy A, Fioranelli F. Continuous human activity recognition with distributed radar sensor networks and CNN–RNN architectures. IEEE Trans. Geosci. Remote Sens. 2022;60:1–15. doi: 10.1109/TGRS.2022.3230829. [DOI] [Google Scholar]
  444. Zhu Y. Parallel excitation with an array of transmit coils. J. Magn. Reson. Imaging. 2004;51:775–84. doi: 10.1002/mrm.20011. [DOI] [PubMed] [Google Scholar]
  445. Ziemann U, Paulus W, Nitsche M A, Pascual-Leone A, Byblow W D, Berardelli A, Siebner H R, Classen J, Cohen L G, Rothwell J C. Consensus: motor cortex plasticity protocols. Brain Stimul. 2008;1:164–82. doi: 10.1016/j.brs.2008.06.006. [DOI] [PubMed] [Google Scholar]
  446. Zilberti L, Arduino A, Torchio R, Zanovello U, Baruffaldi F, Sanchez‐Lopez H, Bettini P, Alotto P, Chiampi M, Bottauscio O. Orthopedic implants affect the electric field induced by switching gradients in MRI. Magn. Reson. Med. 2024;91:398–412. doi: 10.1002/mrm.29861. [DOI] [PubMed] [Google Scholar]
  447. Zilberti L, Arduino A, Zanovello U, Martinez J A, Moulin K, Troia A, Bottauscio O. Magnetic resonance-based electric properties tomography via Green’s integral identity. IEEE Access. 2025;13:42029–44. doi: 10.1109/ACCESS.2025.3546036. [DOI] [Google Scholar]
  448. Zuo Z. Validity of transfer function (TF) model for partial in partial out (PIPO) device under different MR radio frequencies. Proc. ISMRM 2025 Annu. Meeting.2025. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available upon reasonable request from the authors.


Articles from Physics in Medicine and Biology are provided here courtesy of IOP Publishing