Table 2.
Clinical characteristics of the patient in case 2.
| Characteristics | Baseline | 1 month | 3 months | 6 months | |
|---|---|---|---|---|---|
| BCVA | Right eye | 20/50 | 20/32 | 20/30 | 20/25 |
| Left eye | 20/30 | 20/25 | 20/25 | 20/20 | |
| Deviation degree | −30 | −10 | −5 | ||
| PEP | Horizontal | 34 | 5 | 24 | 13 |
| Vertical | 5 | 8 | 8 | 4 | |
| Stereopsis | S0 | 0 | 0 | 1 | 1 |
| S1 | 0 | 0 | 1 | 1 | |
| S2 | 1 | 1 | 1 | 1 | |
| Cobb angle | 13 | – | 0 | 0 |
BCVA, best-corrected visual acuity; PEP, perceptual eye position.
S0 (zero-order stereopsis) was graded as follows: 0 = failed test, 1 = 100 arcseconds, 2 = 200 arcseconds, 3 = 400 arcseconds, 4 = 800 arcseconds.
S1 (first-order stereopsis) was graded as follows: 0 = failed test, 1 = passing at a speed of 50 arcs/ms, and 2 = passing at a speed of 600 arcs/ms.
S2 (second-order stereopsis) was graded as follows: 0 = failed test, 1 = 100% accuracy, 2 = 75% accuracy, 3 = 50% accuracy, 4 = 25% accuracy.
PEP and stereopsis were assessed with a visual and perceptual evaluation system.