Table 1.
Common Events* That Can Cause Less Than 25% of Delivered Gas to Return to the Ventilator and Generate “Disconnect” Alarms
| 1.Disconnection of the ventilator circuit |
| a.Circuit becomes detached from the patient's artificial airway |
| b.Pieces of the circuit become disconnected from each other |
| 2.During a medical procedure, the ventilator is disconnected intentionally. However, the staff forget to ventilate the patient or forget to turn off the ventilator as alternative ventilation is being provided |
| 3.Obstruction of the tubing connecting the patient to the ventilator |
| a.Condensed water filling the tubing |
| b.Kink or blockage in the tubing |
| c.Leak in the tubing |
| 4.The patient's artificial airway comes out accidentally (extubation) |
| 5.Leak around the artificial airway (cuff leak) |
| 6.Obstruction of the patient's artificial airway |
| 7.Leakage through the outside surface of the lung (bronchopleural fistula) |
| 8.The patient takes multiple breaths from the ventilator in a row without exhaling between breaths; This generally occurs when the ventilator is delivering only small-sized breaths to the patient, but the patient is trying to get much larger breaths (breath-stacking or patient-ventilator dyssynchrony) |
All these events are classified by the ventilator manufacturer as “ventilator disconnections.”