. 2005 Jun 28;9(5):465–471. doi: 10.1186/cc3763

Figure 2.

Figure 2

Optimization of CO2 elimination in patients with severe acute respiratory distress syndrome (ARDS). Open circles, reduction of arterial partial pressure of CO2 (PaCO2) obtained by replacing the heat and moisture exchanger (HME) placed between the Y piece and the proximal tip of the endotracheal tube by a conventional heated humidifier (HH) on the initial part of the inspiratory limb in 11 patients with ARDS (reproduced from [27] with the permission of the publisher); filled circles, reduction of PaCO2 obtained by combining the increase in respiratory rate (without generating intrinsic end-expiratory pressure) and the replacement of the HME by a conventional HH in six patients with ARDS [16]. ConMV, conventional mechanical ventilation (low respiratory rate with HME); OptiMV, optimized mechanical ventilation (optimized respiratory rate with HH). Published with kind permission of Springer Science and Business Media [27].