. 2025 Oct 22;41(5):e70211. doi: 10.1002/joa3.70211

Explosion Risk of Cardiac Implantable Electronic Devices During Cremation: Experimental and Survey Findings From Japan

Takanori Arimoto 1, Yasushi Oginosawa 2, Katsuya Ebe 3,4, Masatomo Takahashi 5, Eiichi Watanabe 6,✉, Haruhiko Abe 2,7, Masahiko Takagi 8, Hiroshi Tada 9; the Japanese Heart Rhythm Society
PMCID: PMC13001675  PMID: 41868154

ABSTRACT

Background

The widespread use of cardiac implantable electronic devices (CIEDs), combined with increasing global cremation rates, has raised concerns regarding potential explosion risks during cremation. Lithium batteries within these devices may rupture due to heat‐induced internal pressure, potentially threatening crematorium infrastructure and staff safety. Despite these concerns, real‐world data on explosion behavior and its consequences remain limited. This study aimed to assess the explosion characteristics of CIEDs during cremation and evaluate their impact on facility safety, with the goal of informing evidence‐based guidelines.

Methods

Combustion experiments were performed using full‐scale cremation furnaces set at 1100°C and 800°C. A total of 13 CIEDs (7 pacemakers and 6 implantable cardioverter‐defibrillators) were tested, with synchronized audio‐video monitoring. Additionally, nationwide survey data were collected from 2044 crematoria across Japan, focusing on cremation practices, reported damage, and staff injuries.

Results

All tested devices exploded, with mean ignition‐to‐explosion times of 4.3 ± 1.1 min at 1100°C and 5.4 ± 2.0 min at 800°C (p = 0.19). Seventeen explosion events were observed, yet no structural damage to the chambers occurred. Survey data revealed a decline in cremation refusal due to CIEDs—from 11.0% in 2012 to 4.6% in 2013 and 3.0% in 2014 (p < 0.001). According to respondents, this trend likely reflects an increased collaboration with medical associations, modernization of crematorium facilities, and dissemination of practical safety information.

Conclusions

Cremation of CIEDs without removal appears safe under modern, regulated facilities. However, region‐specific guidelines remain necessary due to international variability in infrastructure and practices.

Keywords: cardiac resynchronization therapy, cremation, implantable cardioverter‐defibrillators, pacemakers


This study provides the first full‐scale experimental and national survey data on CIED explosions during cremation, demonstrating no structural damage and highlighting reduced cremation refusals in Japan. Findings support risk‐based, region‐specific guidelines allowing safe cremation without routine device removal.

graphic file with name JOA3-41-e70211-g005.jpg

1. Introduction

Cremation is the predominant form of body disposition in many countries, with rates exceeding 60% in the United States and reaching 99.97% in Japan [1]. The global increase in cremation, together with the growing number of individuals with cardiac implantable electronic devices (CIEDs)—such as pacemakers and implantable cardioverter‐defibrillators (ICDs)—has raised concern over explosion risks during cremation [2, 3, 4, 5, 6]. These devices, particularly those powered by lithium‐based batteries, pose a risk of explosion during cremation, as heat exposure can trigger gas generation and increased internal pressure, leading to device rupture. The risk is supported by both experimental evidence—showing lithium‐ion battery explosions between 620°C and 800°C and pacemaker casing rupture between 600°C and 800°C—and by real‐world incident reports [7, 8]. Consequently, many countries recommend or mandate device removal prior to cremation [9, 10, 11, 12]. However, universal removal raises legal, logistical, and cultural issues.

International guidelines vary widely in their approach to this issue. In the United Kingdom and Brazil, device removal before cremation is mandatory due to explosion risks, whereas in the United States and Europe, removal or deactivation—particularly of ICDs—is generally recommended but not legally required (Table S1) [9, 10, 11, 12]. In contrast, Japan follows a distinct policy. Since 1989, the Japanese Heart Rhythm Society (JHRS) has advised that routine removal of CIEDs before cremation is not necessary, citing limitations such as lack of legal authority to open caskets, difficulty verifying device presence, and the absence of reimbursement for post‐mortem explantation (Table S2) [13]. Japan also conducts over 1.6 million cremations annually, and routine device removal would pose a significant logistical burden under such high procedural volume [1].

Although a recent benchtop study has shown that CIEDs can explode at temperatures as low as 500°C, real‐world data from full‐scale crematoria remain limited [14]. Few studies have systematically examined the thermal behavior of CIEDs under actual cremation conditions, the structural resilience of modern furnaces, or safety protocols across international settings [15].

To address these gaps, we conducted combustion experiments using explanted CIEDs in full‐sized cremation furnaces. Although these experiments were performed in 2005, they remain the only published full‐scale cremation trials using operational furnaces, and the findings are increasingly relevant due to the continued rise in CIED use and cremation rates. As part of the Social Issues Committee of the JHRS, these historical data were consolidated with the intention of presenting Japan's cremation experience and safety practices to the international community. In addition to evaluating the physical risks via combustion experiments, we sought to understand real‐world crematorium practices and safety measures currently in place across Japan. We therefore analyzed nationwide survey data from over 2000 Japanese crematoria to assess adherence to CIED‐related protocols and the evolution of facility‐level safety standards. The study aims to assess the explosion behavior of CIEDs and their impact on crematorium infrastructure and staff safety, with the goal of informing evidence‐based cremation policies.

2. Methods

2.1. Crematoria Experiment

In 2005, combustion experiments were performed using a full‐scale cremation furnace designed for human remains, under the supervision of on‐site crematorium personnel with professional expertise (Figure 1). Two temperature settings were evaluated: 1100°C, representing the standard protocol for adult cremation, and 800°C, commonly used for individuals with lower body mass, such as children or small adults. A total of 13 explanted CIEDs were tested, seven transvenous pacemakers and six ICDs. All devices had reached end‐of‐life status at the time of explantation. The devices were placed directly on refractory bricks within the cremation chamber and then the chamber was heated to the designated temperature. Explosion events were monitored using synchronized video and audio recording equipment installed in front of the furnace monitoring window. These windows, typically present in Japanese crematoria, are used for visual confirmation of the condition of bone fragments to support culturally important bone collection practices. The timing from ignition to explosion was manually reviewed and recorded. Detailed methods are presented in Supporting Information: Full‐scale combustion experiment protocol. The cremation facility used for the combustion trials was equipped with standard Japanese furnace architecture, including an inner combustion chamber lined with high‐performance ceramic refractory materials (alumina, mullite, zirconia) and an outer insulating wall structure. This dual‐layer configuration allows sustained exposure to temperatures exceeding 1600°C while minimizing thermal deformation. Post‐combustion inspections were carried out by on‐site crematorium personnel with professional expertise to assess any structural damage to the chamber.

FIGURE 1.

FIGURE 1

Experimental Setup for Combustion Study. (A) Wide view of the crematorium interior, showing devices placed on refractory bricks. The chamber has internal dimensions of approximately 2.5 m (length) × 0.75 m (width) × 0.6 m (height), with a total internal volume of approximately 1100 L. (B) Schematic diagram corresponding to (A). (C) Interior view of the crematorium as seen through the monitoring window. (D) Synchronized audio (explosion sounds) and video recording of the crematorium's interior through the monitoring window. White arrow indicates a monitoring window, which is typically present in Japanese crematoria. The window can be opened and closed to adjust the position of the body or burial accessories during the cremation process, using a specialized stick inserted through the opening.

2.2. Nationwide Survey Analysis

Three nationwide surveys were conducted between 2012 and 2014 to assess crematorium practices and evaluate adherence to CIED‐related management protocols. The 2012 survey was administered by the Japanese Heart Rhythm Society (JHRS), while the 2013 and 2014 surveys were conducted by the Japan Society of Environmental Crematory (JSEC). The surveys were sent by fax or email to crematorium directors listed in a national directory maintained by the Ministry of Health, Labour, and Welfare of Japan. The questionnaires included items on environmental practices, the handling of infectious remains, occupational safety procedures, and other aspects of crematorium operations.

The present analysis focused on three key outcomes.

  1. The frequency of refusal to cremation when the CIED had not been removed.

  2. The incidence and severity of crematorium damage attributed to CIED explosions.

  3. Reports of personnel injuries from device fragments or explosion debris have been linked to an inadvertent opening of the monitoring window.

In January 2020, the JHRS administered a follow‐up telephone survey to 71 crematoria that had previously reported requiring the removal of CIED to assess whether their policies had changed subsequently.

2.3. Statistical Analysis

Continuous variables are presented as mean ± standard deviation and categorical variables as absolute numbers and percentages. The 95% confidence intervals (CI) were calculated for key categorical outcomes. Chi‐square tests were used to compare proportions (e.g., cremation refusal rates throughout survey years), and Student's t tests were used to compare explosion times between temperature protocols. All tests were two‐sided, with a p‐value of < 0.05 considered statistically significant. Statistical analyses were performed using JMP version 17.0 (SAS Institute, Cary, NC).

2.4. Ethical Considerations

This study involved physical combustion testing of nonfunctional, explanted medical devices and did not involve human or animal subjects. As such, the Institutional Review Board determined that formal ethical approval was not required.

3. Results

3.1. Combustion Experiment in Full‐Scale Crematoria

The details of the CIED types, models, and battery chemistries used are summarized in Table 1.

TABLE 1.

List of devices used in combustion and explosion testing.

No. CIED type Manufacturer Device Model Battery
1100°C cremation protocol
1 PM SIEMENS BILOG IV 2019 T Lithium‐iodine battery
2 PM SIEMENS DIALOG AV 704 Lithium‐iodine battery
3 ICD St. Jude Medical EPIC+ DR V‐236 Lithium silver vanadium oxide
4 PM SIEMENS DIALOG II 2038 T Lithium‐iodine battery
5 PM Medtronic Kappa DR900 KDR901 Lithium‐iodine battery
6 PM Medtronic Kappa DR700 KDR721 Lithium‐iodine battery
7 ICD Medtronic Marquis DR 7274 Lithium silver vanadium oxide
8 ICD Medtronic GEM III DR 7275 Lithium silver vanadium oxide
800°C cremation protocol
1 PM St. Jude Medical Fidelity 5388 Lithium‐iodine battery
2 PM SIEMENS DIALOG II 688 Lithium‐iodine battery
3 ICD St. Jude Medical EPIC+ DR V‐236 Lithium‐iodine battery
4 ICD Medtronic Micro Jewel 7223 Lithium silver vanadium oxide
5 ICD Medtronic GEM III DR 7275 Lithium silver vanadium oxide

Abbreviations: ICD, implantable cardioverter‐defibrillator; PM, pacemaker.

3.1.1. 1100°C Protocol

Eight devices (five pacemakers and three ICDs) were placed directly on refractory bricks in the cremation chamber (Figure 2A). A total of eight discrete explosion sounds were captured by audiovisual monitoring, with a mean time from ignition to explosion of 4.3 ± 1.1 min (range: 2.8 to 5.7 min) (Video 1). Each explosion was associated with a sharp sound, generally less intense than that of a typical firecracker.

FIGURE 2.

FIGURE 2

Placement of devices in the cremation chamber. CIEDs were positioned symmetrically on refractory bricks, spaced approximately 20 cm apart. (A) Arrangement for the 1100°C cremation protocol: Five pacemakers and three ICDs. (B) Arrangement for the 800°C cremation protocol: Two pacemakers and three ICDs. Device numbers correspond to those listed in Table 1.

VIDEO 1.
Download video file (4MB, mp4)

Thermal explosion experiment of CIEDs in 1100 °C, normal speed recording. Video content can be viewed at https://onlinelibrary.wiley.com/doi/10.1002/joa3.70211.

3.1.2. 800°C Protocol

Five devices (two pacemakers and three ICDs) were tested using the same setup (Figure 2B). Nine explosion sounds were recorded, with a mean explosion time of 5.4 ± 2.0 min (range: 1.5 to 7.6 min) from ignition. Time to explosion did not significantly differ between the two temperature protocols (4.3 ± 1.1 min vs. 5.4 ± 2.0 min, p = 0.19; Figure 3). All devices were destroyed during cremation, leaving metallic fragments including wires, discs, and plates. Although some external casing remnants were partially preserved, they were severely deformed (Figure 4). No visible structural damage to the crematorium—including walls, floors, or burners—was observed after incineration following each trial, as confirmed by on‐site crematorium personnel with professional expertise. Differences in explosion timing or severity between pacemakers and ICDs could not be evaluated, as devices were combusted simultaneously.

FIGURE 3.

FIGURE 3

Time interval from ignition to projectile sound. There was no significant difference in the time to the first explosion between the 1100°C protocol (4.3 ± 1.1 min) and the 800°C protocol (5.4 ± 2.0 min; p = 0.19).

FIGURE 4.

FIGURE 4

Post‐cremation condition of tested devices under 1100°C protocol. (A–C) All tested CIEDs disintegrated during the cremation, leaving behind metal discs, wires, mesh, and metal plates. The larger metal plates were identified as components of an ICD, but the pacemaker models could not be determined from the remaining fragments.

3.2. Nationwide Crematorium Survey

A total of 2044 responses were obtained from 4152 distributed questionnaires, which yielded a response rate of 49.2%. According to the JSEC 2013 data, the crematoria that responded conducted an average of 19 cremations related to CIED annually, which corresponds to a mean prevalence of 2.4% ± 6.0%. Table 2 summarizes the key findings from the 2012 and 2014 JHRS and JSEC surveys.

TABLE 2.

Nationwide crematorium survey.

JHRS 2012 JSEC 2013 JSEC 2014
Total number of distributed questionnaires 1539 1519 1094
Number of responses 656 857 531
Response rate (%) 42.6 56.4 48.5
Questionnaire
1. Have you ever declined to perform cremation for individuals whose CIEDs were not removed? (Yes response rate, %) 71 (11) 39 (4.6) 16 (3.0)
2. Have you ever encountered crematorium damage caused by CIEDs? (Yes response rate, %) 92 (14) 51 (6.0) —
3. Have you ever experienced staff injuries caused by CIEDs? (Yes response rate, %) 39 (6.0) 15 (1.8) —

Abbreviations: CIEDs, cardiac implantable electrical devices; JHRS, Japanese Heart Rhythm Society; JSEC, Japan Society of Environmental Crematory.

3.2.1. Refusal of Cremation With CIEDs

The proportion of crematoria refusing cremation in the absence of device removal decreased significantly over time: from 11.0% (95% CI 8.6%–13.4%) in the JHRS 2012 survey, to 4.6% (95% CI 3.4%–6.1%) in JSEC 2013, and 3.0% (95% CI 1.9%–4.8%) in JSEC 2014 (p < 0.001 for trend). This trend was reported to be the result of increased collaboration with medical associations, modernization of crematorium facilities, and improved information sharing led by the JSEC.

A targeted follow‐up in 2020 was conducted among 71 facilities that had previously reported requiring device removal in the 2012 JHRS survey (Figure 5). The number of crematoria that enforce mandatory CIED removal fell from three in 2012 to one in 2020. The number of facilities that requested device removal but proceeded with cremation even when removal was not completed decreased from 68 to 45. In particular, the number of cremation facilities regardless of the presence of CIED increased from zero in 2012 to 19 in 2020 (p < 0.001).

FIGURE 5.

FIGURE 5

Follow‐up survey of facilities requesting device removal from the 2012 JHRS Study. The figure illustrates the evolution of crematorium policies concerning mandatory removal of CIED before cremation, based on follow‐up surveys. The blue section represents facilities that strictly enforced CIED removal before cremation in 2012. The yellow section denotes facilities that requested device removal but still proceeded with cremation even if removal was not performed. The green section represents facilities that allowed cremation with CIED without requiring removal. Between 2012 and 2020, the proportion of crematoria enforcing mandatory removal declined by 66%, while the number of crematoria permitting cremation without removal increased significantly. JHRS, Japanese Heart Rhythm Society.

3.2.2. Facility Damage

Reported incidents of crematorium damage attributed to CIED explosions were 14.0% (95% CI: 11.6%–16.8%) in the JHRS 2012 survey and 6.0% (95% CI: 4.6%–7.7%) in the JSEC 2013 survey. All reported cases were minor in severity, including wall discoloration and surface cracks in monitoring windows. Respondents suggested that improvements such as controlled heating rates and the use of newer refractory materials may have contributed to the lower frequency of reported damage, although methodological variation cannot be excluded.

3.2.3. Staff Safety and Injury

Most staff‐related incidents described below were attributable to the use of the monitoring window—a procedure that has since been largely discontinued with the implementation of automated combustion systems and standardized safety protocols. Staff‐related incidents were reported in 6.0% (95% CI: 4.4%–8.0%) of crematoria in JHRS 2012 and 1.8% (95% CI: 1.1%–2.8%) in JSEC 2013 (p < 0.001). Documented incidents included facial trauma and lacerations caused by metallic fragments ejected through the monitoring window, particularly in cases where CIED implantation was not known prior to cremation. Personal protective equipment, such as goggles, face shields, and hearing protection, has since become standard practice. Additionally, many facilities instituted a 20‐ to 30‐min waiting period before opening the monitoring window, timed to follow the typical explosion interval. Although most of the staff reported being startled by the explosion sounds, none indicated a need to adjust standard operating procedures or experienced psychological distress.

4. Discussion

4.1. Major Findings

This study evaluated the explosion behavior of end‐of‐life CIEDs through combustion experiments in full‐scale cremation furnaces and analyzed data from nationwide survey from Japanese crematoria. All devices tested exploded, but no major structural damage to the furnace was observed under standard Japanese operating conditions. The survey data demonstrated a significant decline in cremation refusal rates, reduced reports of injury to the facility and staff, and a greater adoption of safety protocols. Although earlier reports linked staff injury to monitoring window use, such incidents are now rare due to system upgrades that eliminate the need to open the chamber during combustion. These findings support the feasibility and safety of Japan's practice of permitting cremation without routine removal of CIED.

4.2. Behavior of CIEDs During Cremation

In our full‐scale combustion experiments, eight devices were tested at 1100°C and five devices at 800°C. All devices exploded, producing a total of 17 discrete explosion sounds. Explosions occurred consistently within the first 5 min after ignition, although in some cases multiple bursts were observed over a period of up to 7.6 min. These explosions were estimated to occur when the furnace temperature reached approximately 400°C–600°C, consistent with previous benchtop studies that demonstrated CIED rupture in a similar thermal range [14].

Despite multiple explosion events, no visible structural damage was observed to the crematorium chamber, including walls, floor, or burners, in either trial. This is in contrast to benchtop models, where structural damage has been reported [14]. The lack of damage in our study may be attributed to the structural features of Japanese cremation chambers, which typically incorporate refractory bricks and ceramic linings, as well as larger chamber volumes and efficient heat dispersion (Supporting Information, Structural Materials of Modern Cremation Furnaces). These factors may have collectively attenuated the explosive force. The observation that some devices exploded more than once highlights the need for sufficient waiting periods before crematorium staff interact with the chamber after ignition.

While our combustion experiments did not show structural damage, the survey data reported a low incidence of facility damage. This discrepancy may reflect the variability in the crematorium infrastructure. The experiment was conducted in a modern and well‐maintained facility, equipped with refractory linings and standardized operating protocols, whereas the nationwide survey included many crematoria with heterogeneous infrastructure, some of which may have been older or less well equipped. In addition, our combustion experiments specifically evaluated end‐of‐life CIEDs, whereas survey reports may also have included damage caused by other items placed in the casket. The nationwide survey was retrospective and relied on the memory of staff, which may have introduced recall bias. Reported damage could have been confounded by the presence of other medical devices (e.g., orthopedic prostheses, spinal cord stimulators) or bulky burial items such as gas cylinders, golf clubs, prosthetic limbs, or metallic tools, as well as operational factors including coffin material, body size and composition, loading density, heating profile, and furnace maintenance status. Therefore, observed differences in reported damage frequency are more plausibly explained by differences between facilities, reporting scope, and potential recall bias, and should be interpreted with caution before attributing them to a temporal trend. These considerations highlight the limitations of retrospective survey data and the need for prospectively designed studies that control such confounders.

4.3. Explosion Risk and Mitigation in Japan

Since 1989, JHRS has recommended that removal of CIED before cremation is not always necessary (Table S2) [13]. This position reflects several practical challenges in Japan, including legal restrictions on corpse handling, cultural sensitivities regarding body preparation, and unreliable family‐reported information on device presence. Furthermore, postmortem device removal is not reimbursed, making routine removal difficult to implement within existing healthcare frameworks. Our survey analysis suggests an increase in acceptance of cremation without mandatory removal of CIEDs in recent years. According to respondents, this change reflects both the practical experience accumulated by crematory staff in managing CIED‐related risks and the broader implementation of risk mitigation practices, supported by certification programs and continuing education initiatives by the JSEC [16]. Importantly, advances in cremation technology—such as the introduction of sealed combustion chambers and computer‐controlled furnaces—have significantly reduced staff exposure to risk. These systems eliminate the need for manual repositioning of the body or accessories during cremation, thereby improving operator safety. While older facilities sometimes implemented a 20–30 min waiting period before opening the monitoring window, this precaution has become largely obsolete with modern automated systems. Nonetheless, some crematoria with outdated infrastructure or a history of explosion‐related incidents continue to enforce device removal. This highlights the need for facility‐specific and tailored policies in post‐mortem CIED management.

4.4. Comparison With International Cremation Practices

While the modernization of cremation technology has contributed to safer practices in Japan, cremation protocols and infrastructure differ significantly between countries [1]. In Japan, cremation is typically performed at approximately 1200°C using furnaces equipped with refractory linings and automated controls. On the contrary, Western crematoria may operate at higher maximum temperatures (up to 1600°C), often to meet environmental regulations [17], and many lack features such as monitoring windows. To mitigate the known risk of CIED explosions, Japanese crematoria have adopted pragmatic safety strategies. These include the use of personal protective equipment, such as face shields, goggles, and hearing protection, as well as operational adjustments based on the suspected presence of CIEDs. Notably, monitoring windows remain a standard feature in Japanese cremation chambers, serving culturally significant functions such as facilitating bone collection and permitting visual confirmation of the combustion process. On the contrary, Western cremation practices generally prioritize complete incineration over post‐cremation bone retrieval. As a result, the monitoring window is rarely present, making real‐time visual assessment unfeasible. Accordingly, safety protocols in such settings must be adapted to the available infrastructure and local cultural norms.

4.5. Risk‐Based Approach and Policy Implications

Our study supports a risk‐based approach to post‐mortem CIED management. As shown in Table 3, devices can be classified by explosion risk according to battery type, device structure, and observed combustion behavior. Leadless pacemakers and implantable loop recorders (ILRs), which present minimal risk, may be safely cremated without removal [18, 19, 20]. Transvenous pacemakers, while capable of explosive disintegration, contain smaller batteries and may also be cremated without removal under appropriate precautions. In contrast, transvenous ICDs and extravascular ICDs typically contain larger‐capacity batteries and are theoretically associated with higher explosion risk. However, our combustion experiments did not demonstrate a significant difference in explosion characteristics between pacemakers and ICDs. Given their battery size and energy capacity, these devices may warrant case‐by‐case evaluation depending on local infrastructure and safety protocols. Subcutaneous ICDs, for which combustion data remain unavailable, should be considered for precautionary removal until further evidence becomes available. Development of standardized classification systems and incorporation of post‐mortem CIED guidance at the time of implantation would support safer and more consistent clinical and operational decision‐making.

TABLE 3.

Risk classification framework for CIED removal prior to cremation.

Risk level Device type Battery type Explosion likelihood Recommended action
Low risk Leadless pacemakers, ILRs Small lithium‐iodine batteries Very low Cremation without removal
Moderate risk Transvenous pacemakers Lithium‐iodine or Lithium‐silver vanadium oxide Disintegrated explosively Cremation without removal
High risk Transvenous ICDs, EV‐ICD Large lithium‐silver vanadium oxide or lithium‐manganese dioxide Disintegrated explosively Cremation without removal
Unknown risk S‐ICD Large‐capacity battery Data unavailable Case‐by‐case evaluation

Note: CIEDs are categorized by explosion risk based on device type, battery chemistry, and observed thermal behavior. Recommended actions range from no removal to precautionary removal depending on risk level. Until further combustion data are available, S‐ICD should be considered for removal as a precaution. In modern Japanese crematoria, automated combustion systems are standard, and observation windows are now rarely opened during the cremation process. As a result, the risk of fragment‐related injury to staff—previously associated with manual opening of the chamber during active combustion—has been markedly reduced. Nevertheless, in older crematoria within Japan, or in other countries with differing cremation protocols or facility infrastructure, additional safety measures may still be necessary.

Abbreviations: EV‐ICD, extravascular ICD; ILR, implantable loop recorder; S‐ICD, subcutaneous implantable cardioverter‐defibrillator.

Although the combustion experiments were conducted in 2005, they remain the first—and to our knowledge, only—full‐scale crematorium‐based investigation into the thermal behavior of CIEDs under real‐world conditions. While more recent benchtop studies, such as those by Maher et al., have explored explosion risks in laboratory settings, they were limited by small chamber volumes and artificial environments that do not accurately reflect operational crematoria [14]. In contrast, our experiments were performed in full‐sized cremation furnaces under standardized protocols, offering empirical data more applicable to actual cremation practices and infrastructure. The combustion and nationwide survey components of this study were conducted independently but were integrated due to their strong thematic relevance.

Importantly, these findings must be understood in the context of major advances in Japanese cremation infrastructure over the past two decades. The widespread implementation of computer‐controlled heating systems, sealed combustion chambers, and refractory ceramic linings has enhanced thermal stability and blast resistance. Concurrently, operational reforms—including the adoption of delayed chamber entry, protective equipment use, and staff training programs—have improved overall safety. These technical and procedural improvements likely contributed to the observed decline in injury reports and cremation refusals in the nationwide survey. Over the past two decades, pacemakers have generally become smaller, while ICDs, particularly S‐ICDs, often contain larger batteries. Although precise comparative data are limited, such changes could potentially influence explosion dynamics and underscore the need for contemporary experimental validation. Collectively, this study bridges the gap between laboratory modeling and real‐world safety outcomes, reinforcing the feasibility of Japan's risk‐adapted approach to post‐mortem CIED management.

5. Limitation

This study has several limitations. First, all tested devices were explanted at end‐of‐life status; the behavior of fully charged or recently implanted devices remains unknown. Second, experiments were conducted without enclosing devices in coffins or surrounding soft tissue, which can influence heat transfer and explosion dynamics. To our knowledge, there are no published experimental studies evaluating the impact of surrounding tissue or coffins on the risk or dynamics of CIED explosions. Third, only lithium‐iodine and lithium‐silver vanadium oxide battery chemistries were tested; newer chemistries, such as lithium‐manganese dioxide, were not evaluated. We acknowledge that the devices tested were older‐generation models; however, the fundamental battery chemistry—lithium‐based cells—remains consistent in modern CIEDs. As such, the core mechanisms of thermal rupture and explosion remain unchanged, supporting the ongoing relevance of our findings. Nevertheless, further testing with contemporary devices would strengthen generalizability. Fourth, although no emission gas analysis was performed in this study, modern Japanese crematoria are equipped with advanced exhaust treatment systems, including secondary combustion chambers, gas cooling units, dust collectors, and denitrification devices that process combustion gases. Manufacturer data suggest that emissions of toxic substances such as sulfur oxides, hydrogen chloride, nitrogen oxides, particulate matter, and dioxins remain within strict regulatory limits [21]. Fifth, cremation conditions in Japan—such as standardized furnace design and monitoring windows—may not be generalizable to other countries, where infrastructure and operating temperatures vary considerably. Sixth, our dataset lacked detailed geographic identifiers, which precluded secondary analysis of regional variation. Future nationwide surveys are warranted to specifically examine potential differences across regions. Seventh, we could not compare ignition‐to‐explosion times between pacemakers and ICDs, and no prior studies have systematically addressed this question. Further controlled experiments are needed. Finally, the nationwide survey had a response rate of 49.2%, raising the possibility of response bias. Facilities that experienced CIED‐related incidents or adopted stricter cremation protocols may have been more motivated to respond to the survey, potentially skewing the overall findings toward overreporting of adverse events or proactive practices.

6. Conclusion

This study provides real‐world evidence from Japan that end‐of‐life CIEDs consistently explode during cremation, but do not typically cause major structural damage under standard operating conditions. National survey data suggest that crematoria have increasingly accepted cremation in CIED without removal, supported by improvements in facility design, operational protocols, and staff training. Historical risks associated with monitoring window access have been largely eliminated through system upgrades. In this context, Japan's longstanding practice of allowing cremation without mandatory device removal appears both feasible and safe for transvenous pacemakers and ICDs. However, international variations in cremation infrastructure require region‐specific guidelines informed by local risk profiles and technical capabilities. Nevertheless, although historical data suggest a low incidence of accidents, the lack of recent nationwide data prevents us from affirming safety with certainty. Updated surveys are urgently needed to validate current safety practices.

Conflicts of Interest

Mr. Masatomo Takahashi is an employee of Fukuda Denshi, a company with a professional interest in the subject of this study. Dr. Eiichi Watanabe serves as a consultant to Fukuda Denshi. The remaining authors declare no conflicts of interest.

Supporting information

Table S1: International guidelines on the management of cardiac implantable electronic devices during cremation.

Table S2: Handling of lithium battery pacemakers during cremation.

JOA3-41-e70211-s001.docx (40.3KB, docx)

Acknowledgments

We gratefully acknowledge Dr. Isamu Yokota (Shizuoka University), Dr. Masateru Takaoka, Dr. Takashi Fujimori, Dr. Yasuto Matsui (Kyoto University Graduate School of Global Environmental Studies), and Dr. Minoru Yoneda (Kyoto University Graduate School of Engineering) for their valuable contributions to nationwide research on crematorium practices. We also thank Mr. Kiyotomo Sudo (Fukuda Denshi), Mr. Keiichi Kitamura (Fukuda Denshi), Mr. Takafumi Kimura (Medtronic), and Mr. Dai Watanabe (Medtronic) for their technical assistance during the combustion experiments, and Ms. Chiaki Takahara for preparing the graphical abstract and video material. We are especially grateful to the crematorium operators for generously sharing their expertise and supporting the experimental procedures.

Arimoto T., Oginosawa Y., Ebe K., et al., “Explosion Risk of Cardiac Implantable Electronic Devices During Cremation: Experimental and Survey Findings From Japan,” Journal of Arrhythmia 41, no. 5 (2025): e70211, 10.1002/joa3.70211.

Funding: The authors received no specific funding for this work.

References

  • 1. “International Cremation Statistics 2023: The Cremation Society of Great Britain,” (2024), https://www.cremation.org.uk/International‐cremation‐statistics‐2023.
  • 2. Hawley C., “Crisis in Cremation. Poor Form Filling Makes Medical Referees Essential,” BMJ 318, no. 7186 (1999): 811. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Kurita T., Ueda S., Okamura H., et al., “Destructive Device Removal—Sparks and Deletion of Therapy History From an Implantable Cardioverter Defibrillator,” International Heart Journal 50, no. 6 (2009): 823–827. [DOI] [PubMed] [Google Scholar]
  • 4. “Annual Statistics of ICD, CRT‐D Implantation in Japan (in Japanese): Japan Arrhythmia Device Industry Association,” (2024), http://www.jadia.or.jp/medical/crt‐d.html.
  • 5. “Annual Statistics of Pacemaker, CRT‐P Implantation in Japan (in Japanese): Japan Arrhythmia Device Industry Association,” (2024), http://www.jadia.or.jp/medical/crt‐p.html.
  • 6. Iwasaki Y.‐k., Noda T., Akao M., et al., “JCS/JHRS 2024 Guideline Focused Update on Management of Cardiac Arrhythmias,” Journal of Arrhythmia 41, no. 3 (2025): e70033. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Dedinot J. L., “Inceneration Des Stimulateurs Cardiaques,” Stimucoeur Medicine 10 (1982): 63–68. [Google Scholar]
  • 8. Morrell P. J., “The Exploding Body,” Practitioner 219, no. 1309 (1977): 109. [PubMed] [Google Scholar]
  • 9. Lampert R., Hayes D. L., Annas G. J., et al., “HRS Expert Consensus Statement on the Management of Cardiovascular Implantable Electronic Devices (CIEDs) in Patients Nearing End of Life or Requesting Withdrawal of Therapy,” Heart Rhythm 7, no. 7 (2010): 1008–1026. [DOI] [PubMed] [Google Scholar]
  • 10. Pitcher D., Soar J., Hogg K., et al., “Cardiovascular Implanted Electronic Devices in People Towards the End of Life, During Cardiopulmonary Resuscitation and After Death: Guidance From the Resuscitation Council (UK), British Cardiovascular Society and National Council for Palliative Care,” Heart 102, no. 7 (2016): A1–A17. [DOI] [PubMed] [Google Scholar]
  • 11. Glikson M., Nielsen J. C., Kronborg M. B., et al., “2021 ESC Guidelines on Cardiac Pacing and Cardiac Resynchronization Therapy,” European Heart Journal 42, no. 35 (2021): 3427–3520. [DOI] [PubMed] [Google Scholar]
  • 12. Oliveira J. C., Fagundes A. A., Alkmim‐Teixeira R., et al., “Recommendations for the Post‐Mortem Management of Cardiac Implantable Electronic Devices,” Arquivos Brasileiros de Cardiologia 115, no. 6 (2020): 1178–1179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Japanese Heart Rhythm Society , “Handling of Lithium Battery Pacemakers During Cremation,” Cardiac Pacing (in Japanese) 5 (1989): 409. [Google Scholar]
  • 14. Maher T., Kong N., Spray R., et al., “Safety and Behavior of Implantable Electronic Devices During Cremation,” Heart Rhythm 22, no. 4 (2025): 1073–1079. [DOI] [PubMed] [Google Scholar]
  • 15. Weber F., “Cremation Temperature,” https://eulogyassistant.com/cremation‐temperature/.
  • 16. “Japan Society of Environmental Crematory (in Japanese),” (2003), https://www.j‐sec.jp/index.html.
  • 17. “Legislation.gov.uk: Environmental Protection Act 1990,” (1990), https://www.legislation.gov.uk/ukpga/1990/43/contents.
  • 18. McArthur A., “Micra Transcatheter Pacing System End of Service/Postmortem Considerations,” (2015), https://www.aaptuk.org/downloads/micra‐EOS‐post‐mortem‐white‐paper.pdf.
  • 19. Cheung L., Chan G. C. P., Chan J. Y. S., and Lau C. P., “Cremation of Leadless Pacemaker,” Pacing and Clinical Electrophysiology 40, no. 6 (2017): 629–631. [DOI] [PubMed] [Google Scholar]
  • 20. MinistÈre Des SolidaritÉs et de la SantÉ , “Arrêté du 19 Décembre 2017 Fixant la Liste Des Prothèses à Pile Exonérées de L'obligation D'explantation Avant Mise en Bière Prévue à L'article R. 2213–15 du Code Général Des Collectivités Territoriales,” (2017), https://www.legifrance.gouv.fr/jorf/id/JORFTEXT000036259211?utm_source=chatgpt.com.
  • 21. Taiyo Chikuro Industries Co. L ., “Auxiliary Equipment Specifications—Nagoya City Crematorium Example,” https://taiyochikuro.jp/product/option.html.

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Table S1: International guidelines on the management of cardiac implantable electronic devices during cremation.

Table S2: Handling of lithium battery pacemakers during cremation.

JOA3-41-e70211-s001.docx (40.3KB, docx)

Articles from Journal of Arrhythmia are provided here courtesy of Japanese Heart Rhythm Society

RESOURCES