Abstract
BACKGROUND:
The Osteoarthritis Initiative (OAI) collected extensive imaging data, including Multi-Echo Spin-Echo (MESE) sequences for measuring knee cartilage T2 relaxation times. Mono-exponential models are used in the OAI for T2 fitting, which neglects stimulated echoes and inhomogeneities. Extended Phase Graph (EPG) modeling addresses these limitations but has not been applied to the OAI dataset.
PURPOSE:
To assess how different fitting methods, including EPG-based and exponential-based approaches, affect the accuracy and reproducibility of cartilage T2 in the OAI dataset.
STUDY TYPE:
Retrospective.
POPULATION:
From OAI dataset, 50 subjects, stratified by osteoarthritis (OA) severity using Kellgren-Lawrence grades (KLG), and 50 subjects without OA diagnosis during OAI duration were selected (each group: 25 females, mean ages ~61 years).
FIELD STRENGTH/SEQUENCE:
3-T, two-dimensional (2D) MESE sequence.
ASSESSMENT:
Femoral and tibial cartilages were segmented from DESS images, subdivided into seven sub-regions, and co-registered to MESE. maps were obtained using three EPG-based methods (non-linear least squares, dictionary matching, and deep learning) and three mono-exponential approaches (linear least squares, non-linear least squares, and noise-corrected exponential). Average T2 values within sub-regions were obtained. Pair-wise agreement among fitting methods was evaluated using the stratified subjects, while reproducibility using healthy subjects. Each method’s T2 accuracy and repeatability varying signal-to-noise ratio (SNR) were assessed with simulations.
STATISTICAL TESTS:
Bland-Altman analysis, Lin’s concordance coefficient, and coefficient of variation assessed agreement, repeatability, and reproducibility. Statistical significance was set at p-value <0.05.
RESULTS:
EPG-based methods demonstrated superior T2 accuracy (mean absolute error below 0.5 ms at SNR>100) compared to mono-exponential methods (error >7 ms). EPG-based approaches had better reproducibility, with limits of agreement 1.5–5 ms narrower than exponential-based methods. T2 values from EPG methods were systematically 10–17 ms lower than those from mono-exponential fitting.
DATA CONCLUSION:
EPG modeling improved agreement and reproducibility of cartilage mapping in subjects from the OAI dataset.
Keywords: Osteoarthritis, Osteoarthritis Initiative, T2 mapping, MESE, knee imaging
Plain Language Summary
This study aimed to improve the accuracy and reproducibility of measuring the MRI relaxation time T2 in knee cartilage, which is related to its quality. The authors re-analyzed existing data from the osteoarthritis initiative using advanced modeling techniques that better account for the physics of the MRI acquisition process and compared it against simpler conventional modelling. The results showed that using advanced modeling provided more accurate and reproducible measurements of cartilage T2. The study highlights the importance of considering the complexities of MRI physics to enhance the quality of cartilage quantitative MRI measurements for osteoarthritis research.
Introduction
The Osteoarthritis Initiative (OAI) is a multicenter, longitudinal, prospective observational study of knee osteoarthritis (OA), with the goal of developing a public-domain research resource to facilitate the scientific evaluation of biomarkers for OA as potential surrogate endpoints for disease onset and progression (1). The OAI collected longitudinal radiography and MRI data from 4,796 participants (1).
One biomarker of interest was the cartilage transverse relaxation time , because of its potential to be sensitive to collagen hydration and degeneration(2,3). To compute cartilage relaxation time, the OAI MRI protocol included a two-dimensional (2D) Multi-Echo Spin-Echo (MESE) sequence (4,5). Studies analyzing data from the OAI have shown to be a promising OA biomarker (6–8). For example, elevated has been associated with increased risk of developing OA (9–12), and longitudinal change of 2 ms has been associated with future OA radiographic progression(13–16). Despite being a promising biomarker, cartilage relaxation times are still mostly disregarded in clinical evaluation. Multiple factors have contributed to the lack of consideration for cartilage in clinical settings. A major reason might be that even from a large study such as the OAI, the marginal value of in monitoring OA compared to other biomarkers, such as cartilage thickness, has not been well established (6,17). While there has been extensive development in advanced statistical and processing methods to exploit cartilage maps in the OAI, such as texture analysis (18,19), deep learning (DL) (20), and statistical parameter mapping (21), much less attention has been given to how the MESE data are modeled to generate the maps. Most studies relied on a simple mono-exponential modeling of the signal (8,9,20,21). Exponential modeling is inherently suboptimal as it does not account for stimulated echoes produced by radiofrequency (RF) slice profile and inhomogeneities, resulting in stimulated echoes affecting the MESE signal and making it deviate from a pure exponential decay (22,23). To mitigate errors due to stimulated echoes, the first echo, which is a pure spin echo, is often discarded before applying a mono-exponential fit (16). While this improves accuracy, the resulting is still biased, and comes at the cost of discarding the data point with the highest signal-to-noise ratio (SNR), which may reduce overall SNR efficiency.
A fitting procedure that addresses the limitations inherent with exponential fitting is based on the Extended Phase Graph (EPG) formalism (24,25), which accounts for slice profile and inhomogeneity. Current EPG fitting methods have been shown to improve accuracy in MESE mapping in multiple body regions (22,26–28).
Against this background, the main goal of this study was to compare the effect of the fitting method used on cartilage accuracy and, more importantly, reproducibility in simulations and using data from the OAI dataset. Furthermore, given the widespread use of mono-exponential fitting in cartilage mapping, we aimed to compare common approaches (linear, non-linear, with or without considering the first echo) to explore which type of fit should be preferred if a mono-exponential model is to be used.
Materials and Methods
This study utilized publicly available de-identified MRI data from the OAI dataset. The IRB approval and written informed consent from all participants was originally collected by the institutions participating in the OAI recruitment and data collection (29). Furthermore, the National Institute of Arthritis and Musculoskeletal and Skin Diseases (NIAMS) at the National Institutes of Health (NIH) appointed an independent Observational Study Monitoring Board (OSMB) to oversee the Osteoarthritis Initiative (OAI) study from 2002 to 2014 (30).
MESE parameters in the OAI dataset
The OAI MRI protocol was acquired by the institutions participating in the OAI using a 3T Siemens Trio (Siemens Medical Solutions, Erlangen, Germany). The full technical details can be found it the development report (31). Relevant to this study, the 2D MESE sequence acquired 7 echoes with an echo time spacing of 10 ms, using a repetition time (TR) equals to 2700 ms. The FOV was 120 mm (384 × 269 matrix size) with an in-plane resolution of 0.313 × 0.446 mm2, and a slice thickness of 3 mm with a 0.5 mm slice gap for a total of 21 slices.
EPG model, simulations, and fitting
The pixel-wise 2D MESE cartilage signal was modeled using the EPG algorithm considering the effects of slice profiles and inhomogeneities, following the work by Lebel and Wilman(22), according to eq. 1
| [1] |
where and are the longitudinal and transverse relaxation times, respectively; is the pulse efficiency ratio; TE is the echo time-space; EXprofile and REFprofile are the excitations and refocusing slice profiles, respectively; ETL is the echo train length (number of refocusing pulses in the MESE sequence); and is the signal amplitude at time equals zero. To simulate the cartilage signal, was fixed to 1200 ms (28). Following the technical specifications for the MESE sequence used in the OAI imaging protocol(32), TE and ETL were fixed to 10 ms and 7, respectively. There is no information about the excitation and refocusing pulse shapes in the technical documentation regarding the MESE acquisition in the OAI. Thus, Hanning-windowed Sinc pulses with a time-bandwidth product (TBW) of 2 and a refocusing factor of 1.2 were used for slice-profile simulations. The slice profiles were obtained using the Shinnar-Le Roux (SLR) algorithm(33). All simulations run using the EPG model were performed using an in-house MATLAB (MATLAB R2023b, MathWorks, Natick, MA, USA) library based on the StimFit toolbox (https://github.com/usc-mrel/StimFit) (34).
OAI MESE fitting methods
Three proposed EPG-based methods and three exponential (EXP)-based methods used in prior OAI literature were considered in this study (8,9,20,21,35). The methods are summarized in Figure 1 and explained individually in more detail in the following sections. The EPG-based methods proposed to estimate voxel-wise cartilage from the MESE data in the OAI dataset were non-linear least squared (EPG NLSQ), dictionary matching (EPG Dictionary), and a DL approach (EPG DL). The conventional EXP-based methods were linear least square(8) (LLSQ), NLSQ(9,20,21), and noise-corrected exponential(35) (NCEXP).
Figure 1:

Summary of fitting methods used in this work. (a) Mono-exponential based methods used in OAI literature. (b) Proposed EPG-based fitting methods have yet to be applied in the OAI dataset.
EPG Non-Linear Least Square and Dictionary fitting procedures
In the EPG NLSQ approach, the experimental decay curve is fitted according to the model described in eq 1 using an iterative non-linear least-squares procedure to minimize the sum of the squares of the difference between the experimental signal and the EPG-modeled theoretical signal(27,34).
The EPG Dictionary approach is summarized in Figure 2(a). The EPG model described in eq. 1 was used to create a dictionary of MESE simulated signals, using and values spanning [10, 110] ms and [0.4, 1.2], respectively, with 0.25 ms and 0.02 as step sizes. Please note that B1 values are dimensionless as they are expressed as the ratio between the actual and nominal flip angles, indicating the efficiency of the RF pulse. The experimental MESE decay curve was fitted by comparing it with the pre-computed dictionary of simulated signals(36). The dictionary entry that showed the highest similarity with the experimental signal was selected and the and values used to generate the dictionary entry were assigned to the pixel. The dot product was used as a metric of similarity, which is a standard metric for dictionary-based fitting in MRI(37,38). It is worth noting that, when using this metric, both the experimental signals and each dictionary entry must be normalized such that their norm is equal to one.
Figure 2:

Schematization of the EPG dictionary (a) and DL (b) fitting methods along with the training pipeline used for training the DL model (c). (b) 4 fully connected hidden layers with a progressively decreasing number of neurons. A bottle-neck structure was chosen as done for other qMRI applications. (C) During training, simulated data are augmented by injecting white Gaussian noise with different noise variances. Signals are normalized before being fed to the DL model.
Both EPG NLSQ and EPG dictionary were developed in MATLAB (MATLAB R2023b, MathWorks, Natick, MA, USA) and were optimized for computational efficiency through code vectorization and exploitation of MATLAB’s automatic parallel support for multicore CPU.
EPG DL approach
The DL approach trained a neural network (NN) to predict cartilage and values from 50000 EPG-simulated MESE signals generated by randomly sampling the parameter space according to a uniform distribution. As for the dictionary approach, and values spanned [10, 110] ms and [0.4, 1.2], respectively. The 50000 sampled (T2, B1) pairs served as ground-truth values for the supervised training procedure.
The NN model developed for fast muscle -water mapping with subject-specific -fat calibration from MESE data(38) was adapted for predicting and values in cartilage. The NN model, summarized in Figure 2(b), has 6 fully connected layers, the Rectified Linear Unit (ReLU) (39) was used as the activation function for the neurons in the first 5 hidden layers, while a linear activation function was used for the output layer. The bottle-neck structure was chosen as it has been successfully applied in other quantitative MRI applications(40,41). The application was developed in Python using Keras (42) with TensorFlow (43) backend.
The mean absolute error (MAE) between NN estimated and parameters and ground-truth parameters was used as a loss function. The Adam optimizer was used to train the NN model with 500 epochs and 1000 gradient steps for each epoch, with a fixed batch size of 500. During training, as depicted in Figure 2(c), to improve generalizability and robustness to noise, simulated noiseless MESE signals from the training set were corrupted by white Gaussian noise with different variances, which were randomly sampled from a uniform distribution with . A wide range of noise variances was the best noise addition strategy when a similar NN architecture was applied to pixel-wise parameter mapping in MRI fingerprinting (40). After noise addition, the magnitude of each input was taken and normalized to the unit norm. When applied to real acquired data for pixel-wise processing, the experimental MESE data was first normalized to the unit norm, by dividing the signal with its norm and then fed to the pre-trained network model.
Conventional exponential-based fitting methods
All conventional EXP-based methods were based on mono-exponential modeling according to eq. 2
| [2] |
In the linear least squared approach (EXP LLSQ), eq. 2 was linearized by taking its natural logarithm and obtaining a problem of the form following eq. 3:
| [3] |
where , and . The problem was then solved using a LLSQ minimization procedure, which can be computed quickly and efficiently using linear algebraic operations.
In the non-linear least square approach (EXP NLSQ), the experimental decay curve was fitted according to the mono-exponential model of eq. 2 using an iterative non-linear least squares procedure to minimize the sum of the squares of the difference between the experimental signal and the mono-exponential theoretical signal.
In the NCEXP method, the experimental decay curves are fitted using an NLSQ iterative procedure using a modified mono-exponential model accounting for Gaussian noise (35) following eq 4:
| [4] |
where and is the noise standard deviation and is the modified Bessel function of the first kind of -th order.
For the EXP LLSQ and EXP NLSQ, the fitting was performed both by dropping the first echo and without dropping the first echo. When the first echo was not excluded in the fit, the EXP LLSQ and EXP NLSQ methods were referenced as EXP LLSQ ndrp and EXP NLSQ ndrp, respectively.
Computer configuration and computational time
All processing was run on a laptop computer (Dell XPS 15 9530, Dell, Round Rock, TX, USA) equipped with 32 GB of Random Access Memory (RAM), using only Central Processing Units (CPUs): Intel Core i7–13700H. With such a configuration, training of the NN model took 10 s per epoch, requiring 1 h and 23 min to complete the training. For the dictionary approach, the creation of the dictionary required 0.25 s.
Regardless of the model used (EPG, EXP, or NCEXP) the NLSQ iterative procedure was implemented in MATLAB (MATLAB R2023b, MathWorks, Natick, MA, USA) using the lsqcurvefit function using the default trust-regionreflective optimization algorithm.
Experiment with simulated data
To assess the performance of each fitting method as a function of noise, 2000 MESE signals were simulated using the EPG framework. Ground truth values ranged from 20 to 80 ms, and values ranged from 0.9 to 1.1. Both were randomly sampled from a uniform distribution. Each fitting method was used to fit values after adding white Gaussian noise with increasing variance. The procedure was repeated 10 times with resampling of the noise. Performance was assessed in terms of accuracy, using the MAE, and repeatability, estimated by the mean repeatability coefficient (RC), defined as , where is the standard deviation of values from the 10 repetitions.
Experiments with data from the OAI dataset
Agreement among fitting methods
To assess agreement among fitting methods, 50 subjects (mean age: 62.1 ± 7.8 years ) were randomly selected from the OAI dataset, with 10 subjects (5 females and 5 males) per OA severity, classified according to the KLG: 0 (none), 1 (doubtful), 2 (minimal), 3 (moderate), and 4 (severe). The publicly available OAI dataset reported the KLG for each participant at each timepoint, which was performed by the institutions participating in the initiative according to their study description (29).
Tibial and femoral cartilages were automatically segmented from the Double Echo Steady State (DESS) images and subdivided into 7 regions: medial tibia (MT), lateral tibia (LT), trochlea, central and posterior regions for the medial (MF) and lateral (LF) femoral cartilage. A DL-based cartilage segmentation algorithm previously validated on the OAI DESS data was used to perform automatic segmentation(44). Femur sub-regions were determined using the opensource python Musculoskeletal Toolkit (pyMSKT, https://github.com/gattia/pymskt) (21). To obtain segmentations for the MESE images, DESS was registered to the MESE data using the open-source Elastix toolbox (https://elastix.dev/) (45). For each subject, pixel-wise mapping was performed with all the described fitting methods, and the average within the 7 subregions was computed. Pooling all subjects together,
Effect of fitting method on repeatability and reproducibility
The OAI dataset does not contain scan-rescan repeatability for the MESE data. Thus, to estimate the repeatability of each fitting method, MESE data from 5 subjects in the OAI dataset (1 in each KLG) were corrupted by injecting white Gaussian noise into the MESE images twice, with noise resampling, simulating a scan-rescan test. The operation was then repeated increasing the noise variance to assess the robustness to noise of each fitting method. For each variance, the two MESE repetitions are inherently coregistered.
Lastly, to assess the effect of fitting methods on reproducibility, 50 healthy subjects (mean age: 60.7 ± 9.3 years) (KLG = 0) who did not experience onset of OA for the entire duration of the study (no change in KLG) were randomly selected.
Statistical analysis
Bland-Altmann (BA) analysis was used to assess pair-wise agreement among fitting methods in mean values. Limits of agreement (LOA) and mean bias were the primary BA statistics. The Lin’s concordance coefficient () and coefficient of variation (CV) were also used as metrics of agreement. For the simulated scan-rescan experiment, for each fitting method, repeatability was assessed using a BA analysis considering all the pixels within the cartilage segmentation, instead of averaging across the 7 sub-regions. The LOA and CV were also used as metrics of repeatability. Regarding the reproducibility experiment, the difference in average values in the 7 cartilage subregions between 12-month follow-up and baseline exams was assessed using BA analysis. All the statical analyses were performed using in-house scripts developed in Python based on the statistical library built into the open-source Python package SciPy (https://scipy.org/). A p-value <0.05 was considered statistically significant.
Results
Experiment with simulated data
The MAE (accuracy) and RC (repeatability) for different fitting methods from the simulated data are reported in Figure 3 as a function of SNR. All methods showed degradation of accuracy and repeatability as SNR decreased. The EPG-based methods showed the best performance in terms of accuracy, as MAE was below 0.5 ms when SNR was greater than 100 and below 5 ms for SNR below that level. On the contrary, EXP-based methods showed an average bias above 7 ms. Specifically, when the first echo was considered in the fit, the bias was greater than or equal to 13 ms, with the highest bias when was fitted with the LLSQ procedure. Regarding repeatability, the EPG DL method showed the lowest RC across different noise levels, whereas similar performances in terms of repeatability were observed for all of the other methods.
Figure 3:

accuracy (a) and repeatability (b) for different fitting methods as a function of noise form the simulation experiment. SNR is defined as the ratio between the power of the signal and the power of the simulated Gaussian noise. MAE: mean absolute error; RC: repeatability coefficient.
Experiments with data from the OAI dataset
Representative parameter maps reconstructed using the proposed EPG-based approaches and the conventional EXP-based methods are shown in Figure 4. The average time required by each fitting method to fit the maps within the femorotibial cartilage region is reported in Table 1. By visual inspection, maps obtained using EPG-based methods showed lower values than the maps obtained using EXP-based methods. Furthermore, empirically the maps obtained with the EPG-based methods appeared less noisy than those obtained using exponential-based methods. This is particularly pronounced in the tibia when the first echo was not used for the computation (Figure 4(d), (e) and (f)).
Figure 4:

maps with different fitting methods. Representative slice from maps computed using EPG-based methods (a), and exponential-based approaches (b) from MESE data taken from a subject within the OAI dataset.
Table 1:
Average computational time required by each method to fit T2 values in femoral and tibial cartilages in a subject from the OAI dataset1.
| EPG fitting |
Exponential fitting |
|||||
|---|---|---|---|---|---|---|
| DL | Dictionary | NLSQ | LLSQ | NLSQ | NCEXP | |
|
| ||||||
| Time | 2 s | 1 s | 200 s | 1 s | 200 s | 150 s |
Agreement among fitting methods
To quantitatively characterize the agreement between fitting methods, the BA statistics for the pair-wise method comparisons were as follows: MD, , and are reported in Figure 5(a), (b) and (c), respectively. The EPG-based methods (EPG DL, EPG Dict, and EPG NLSQ) showed the highest inter-method agreement with a mean difference ranging from −0.01 ms to 0.4 ms, ≤ 1.55 ms, and equals to 1. When looking at values obtained using EXP-based fitting methods without including the first echo (EXP LLSQ, EXP NLSQ, and NCEXP), the inter-method agreement was inferior to the one observed for EPG-based methods (−3.59 ms ≤ MD ≤ 4.63 ms, 3.19 ms ≤ LoAwidth ≤ 8.89 ms, 0.83 ≤ ≤ 0.99). Inter-method agreement between exponential fitting methods, performed without dropping the first echo, was also inferior compared to EPG-based methods (MD = −2.84 ms, LoAwidth = 9.27 ms and = 0.91). Poor agreement was found between EPG-based and EXP-based methods (−17.2 ms ≤ MD ≤−10.53 ms, 5.84 ms ≤ LoAwidth ≤ 9.87 ms, 0.26 ≤ ≤ 0.48). On average, values evaluated with the EXP-based methods were higher than those obtained with EPG-based methods, with the highest bias observed for the LLSQ exponential method performed without dropping the first echo, 17 ms, while the lowest bias, about 10 ms, was observed for the NCEXP method.
Figure 5:

fitting method pair-wise comparison from cross-sectional analysis. Bland Altman statistics for average in cartilage regions among different pair-wise method comparisons by pooling average in 50 subjects with 8 cartilage regions per subject: mean difference (a), limits of agreement (b) and Lin’s concordance coefficient (c). Mono exponential methods where the fitting was performed without dropping the first echo are labeled with the suffix ndrp.
Effect of fitting method on repeatability and reproducibility
The Lin’s concordance coefficient (agreement) and coefficient of variation (repeatability) for different fitting methods from the simulated scan-rescan experiment using OAI data are reported in Figure 6 as a function of noise variance. All methods showed degradation agreement and repeatability as noise increased. The EPG-based methods yielded the best performance, as the concordance coefficients and CV were consistently higher, and lower, respectively, than those obtained with the EXP-based methods. Among the EPG-based methods, using the DL approach yielded better performance than using both EPG dictionary and EPG NLSQ methods, which had almost identical performance. Among the EXP-based methods, the LLSQ method with dropping the first echo had the worst repeatability across noise levels, while the NLSQ fitting considering the first echo had the best performance in terms of repeatability.
Figure 6:

Cartilage repeatability of different fitting methods as a function of noise from the simulated scan-rescan experiment using in vivo data from the OAI dataset. (a) Concordance correlation coefficient as a function of noise variance. (b) Coefficient of variation as a function of noise variance.
The BA plot for the reproducibility experiment using 50 healthy subjects in the OAI dataset is reported in Figure 7, along with the width of LoA and Lin’s concordance coefficient, for each fitting method. The EPG-based methods showed higher reproducibility in a population of healthy non-progressing patients compared to the EXP-based methods: LOA width associated with EPG-based methods (6.1 ms ≤ LoAwidth ≤ 6.3 ms) were from 1.5 to 5 ms lower than those obtained with EXP-based methods (7.7 ms ≤ LoAwidth ≤ 11.1 ms). Among the EXP-based methods, using the LLSQ method computed dropping the first echo yielded the worst reproducibility (LoAwidth = 11.1 ms), while adopting the NCEXP produced the best reproducibility (LoAwidth = 7.7 ms). In the LLSQ fitting method, considering the first echo increased reproducibility (LoAwidth = 10.2 ms < 11.1 ms), whereas it slightly worsened reproducibility when using the NLSQ fitting method (LoAwidth = 8.1 ms > 7.8 ms).
Figure 7:

Cartilage reproducibility in non-progressing healthy patients. Bland-Altman plots between baseline and 12 months follow-up mean values in femoral cartilage (FC) and tibia (T) compartments in a population of 50 healthy (KLG=0) patients that did not present onset of OA for the entire duration of the OAI study. (a) BA plots for EPG-based methods. (b) BA plots for exponential-based methods. The width of the limits of agreement () and Lin’s concordance coefficient are reported for each BA plot.
Discussion
This study investigated the efficacy of EPG modeling for improving cartilage mapping accuracy and reproducibility in the OAI dataset. Three EPG-based techniques – NLSQ, dictionary matching, and DL – were evaluated against conventional mono-exponential approaches through simulations and in-vivo data analysis. In terms of accuracy, simulation results demonstrated the superiority of EPG-based methods across various SNR levels, with mean absolute error consistently below 0.5 ms at SNR > 100, compared to over 7 ms for exponential-based methods. The substantial bias observed with exponential fitting, particularly when including the first echo (MAE > 13 ms), underscores the limitations of mono-exponential models and highlights the importance of accounting for stimulated echoes in MESE quantification, consistent with previous studies that showed improved accuracy in MESE mapping using EPG-based methods in other body regions (22,26,27). Cross-sectional analysis of 50 OAI subjects supported results from simulations, showing improved agreement among EPG-based methods, compared with exponential-based methods. The systematic difference between EPG and exponential estimates (10–17 ms) suggests that previous studies on the OAI dataset using mono-exponential fitting may have overestimated cartilage values (8,9,20,21,35). Additionally, variations in exponential fitting methods (either LLSQ, NLSQ, or NCEXP) may have resulted in different degrees of overestimation. This could have potentially impacted the use and interpretation of elevated cartilage values in the context of OA, especially in a previous meta-analysis (17).
It is important to note that the effect of the fitting method on repeatability and reproducibility is more critical than its impact on systematic bias. This is because despite evidence that elevated is associated with the risk of OA (9–12), longitudinal change in cartilage has been shown to be more informative on OA onset and progression than values at a single timepoint (13–16). The results from simulations and the simulated scan-rescan experiments demonstrated overall superior repeatability of EPG-based fitting compared to exponential-based methods. Specifically, among EPG methods, the DL model showed the highest repeatability, followed closely by the EPG dictionary and NLSQ approaches. Among the EXP-based methods, the LLSQ approach yielded the worst performance, particularly when the first echo was not included. The NCEXP method provided better reproducibility than other EXP-based methods but was still inferior to EPG-based techniques. The reproducibility experiments with 50 healthy, non-progressing subjects from the OAI dataset further demonstrated the advantage of EPG-based methods for cartilage mapping. The EPG-based approaches demonstrated narrower LoA compared to conventional methods, with LoA widths being 1.5–5 ms lower than those observed using exponential-based approaches. This represents a substantial relative improvement given the magnitude of changes typically associated with OA progression, which is on the order of 2 ms (13,14). The improved reproducibility shown by EPG-based methods may have several implications both in the context of analysis performed with the OAI dataset and, more generally, on each longitudinal study that utilizes 2D MESE to investigate the role of cartilage in OA. The narrower LoA translates directly to reduced measurement variability, which in the end increases the statistical power of longitudinal studies. This means that smaller sample sizes may be sufficient to detect statistically significant changes in , or that more subtle changes can be identified with the same sample size. The ability to reliably detect smaller changes could potentially allow for meaningful results to be obtained from shorter follow-up periods. In the context of the OAI dataset, the substantial improvement in reproducibility suggests that reanalysis of existing OAI data using EPG-based methods could yield new insights. For example, studies that previously found borderline or statistically non-significant results might reveal statistically significant findings when reanalyzed with EPG-based mapping.
Despite their known limitations, processing of MESE data has largely relied on exponential modeling of the signal because of their simplicity compared to EPG methods, which require deeper technical knowledge of MRI simulations (22,25). Thus, it is important to discuss some of their strengths and limitations, especially given their prevalence in the clinical research setting. Not surprisingly, the LLSQ approach was computationally efficient but showed the poorest accuracy and repeatability. Further, NLSQ and NCEXP, while more accurate, require considerably more computational time (200 times slower in this study). An additional limitation of the NCEXP is the need to estimate the noise variance, which is usually evaluated in a background region of interest (ROI) (35). However, when applied to large datasets, an automatic procedure is needed to process large batches of data, and errors in background region selection can severely impact fit performance. For example, we had to manually adjust the background ROI selection in our analysis. Further, the impact of including the first echo in mono-exponential fitting merits special consideration, as it is a common mitigation strategy to reduce errors in estimation. Our results were in accordance with previous literature, which demonstrated that dropping the first echo improved accuracy(46). However, its impact on repeatability has been less discussed. We observed that dropping the first echo negatively impacted repeatability, especially when using the LLSQ fit. This trade-off between accuracy and precision can be explained by two competing factors. First, excluding the first echo eliminates the only data point not affected by stimulated echoes, thus improving adherence to the exponential decay assumption, but at the cost of discarding an echo with high SNR. This effectively reduces the overall SNR of the data, which is particularly problematic when using a LLSQ approach. In NLSQ, the fitting is performed in the original signal intensity domain. This inherently gives more weight to high-intensity (typically higher SNR) data points, which are usually the earlier echoes in a MESE decay curve. In contrast, LLSQ is performed after logarithmic transformation of the data, which tends to equalize the weighting of all data points, giving disproportionate influence on later, low SNR echoes. The EPG methods solve this trade-off by accurately modeling MESE signal evolution, allowing the utilization of all acquired echoes, and further reducing the variability induced by varying B1 across the field of view, which has been shown to play a role in mapping (47,48). All the EPG-based methods proposed in this work performed similarly in terms of accuracy and reproducibility. Regarding computational efficiency, the DL and dictionary matching approaches were about 200 times faster than the NLSQ iterative procedure, similar to the simple exponential LLSQ approach. This efficiency makes them suitable for processing large datasets and potential candidates for online reconstruction at the scanner. However, the DL approach requires retraining when MESE sequence parameters change, while the dictionary matching approach needs pre-computation of a new dictionary, but with limited computational cost. It is worth noting that the high efficiency of the dictionary matching approach was achievable only because, for cartilage mapping, a single-component model is sufficient to describe the MESE signal evolution. When a multi-component model is required, such as for muscles, the DL approach has been shown to remarkably outperform the dictionary approach (38).
Given our findings, we recommend that researchers consider re-analyzing MESE data in the OAI dataset using the EPG method. If the exponential fitting is to be used, we discourage the use of a linear least squares exponential fit and believe that including the first echo is helpful for reproducibility. Using the EPG method improves both accuracy and reproducibility, which has the potential to uncover previously undetected trends, confirm borderline findings, and possibly lead to new insights into the usefulness of cartilage in OA by better exploiting the OAI dataset.
Limitations
The lack of a reference standard for in-vivo cartilage values makes it challenging to definitively assess accuracy. However, the consistent performance advantages of EPG methods in both simulations and in-vivo experiments provide strong evidence for their superiority against exponential modeling. Further, we could not retrieve the original pulses used in the MESE sequence employed in the OAI study. Therefore, we considered standard SINC pulses with a TBW of 2 and a refocusing factor of 1.2. Although we do not expect our results to be particularly different in terms of repeatability and reproducibility, using the actual pulse played during the MESE sequence would likely increase accuracy. Finally, while we demonstrated improved reproducibility in small sample of healthy, non-progressing subjects, future works on the OAI dataset should investigate a bigger sample size and evaluate the impact of EPG-based mapping on the sensitivity to detect OA-related changes in longitudinal studies.
Conclusion
Using the EPG method may improve both accuracy and reproducibility of MESE T2 mapping in cartilage. To facilitate the adoption of the EPG method in the community, we open-sourced the code to perform the EPG fitting for cartilage at https://github.com/barma7/EPGfit_for_cartilage_T2_mapping. Furthermore, to allow the community working on the OAI dataset to readily use cartilage maps fitted using the proposed EPG method, we have created an open-source repository on Hugging Face to store maps of the entire OAI dataset (https://huggingface.co/datasets/barma7/oai-t2maps-epgfit). At the time of writing, only maps relative to the baseline and 12-month follow-up visits have been uploaded to the repository, and the remaining visits will be uploaded in the following weeks. Hence, we invite the community to check the repository frequently.
Footnotes
DL: Deep Learning; LLSQ: Linear Least Squared; NCEXP: Noise Corrected Expontential; NLSQ: Non-Linear Least Squared.
References
- 1.Branch NSC and O Osteoarthritis Initiative [Internet]. National Institute of Arthritis and Musculoskeletal and Skin Diseases. 2017. Available from: https://www.niams.nih.gov/grants-funding/funded-research/osteoarthritis-initiative [Google Scholar]
- 2.Eckstein F, Burstein D, Link TM. Quantitative MRI of cartilage and bone: degenerative changes in osteoarthritis. NMR Biomed. 2006. Nov;19(7):822–54. [DOI] [PubMed] [Google Scholar]
- 3.Liu F, Chaudhary R, Hurley SA, Munoz Del Rio A, Alexander AL, Samsonov A, et al. Rapid multicomponent T2 analysis of the articular cartilage of the human knee joint at 3.0T. Journal of Magnetic Resonance Imaging. 2014. May 1;39(5):1191–7. [DOI] [PubMed] [Google Scholar]
- 4.Jung BA, Weigel M. Spin echo magnetic resonance imaging. Journal of Magnetic Resonance Imaging. 2013;37(4):805–17. [DOI] [PubMed] [Google Scholar]
- 5.Feinberg DA, Mills CM, Posin JP, Ortendahl DA, Hylton NM, Crooks LE, et al. Multiple spin-echo magnetic resonance imaging. Radiology. 1985. May;155(2):437–42. [DOI] [PubMed] [Google Scholar]
- 6.Jungmann PM, Kraus MS, Nardo L, Liebl H, Alizai H, Joseph GB, et al. T2 relaxation time measurements are limited in monitoring progression, once advanced cartilage defects at the knee occur: Longitudinal data from the osteoarthritis initiative. Journal of Magnetic Resonance Imaging. 2013. Dec;38(6):1415–24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Eckstein F, Kwoh CK, Link TM. Imaging research results from the Osteoarthritis Initiative (OAI): a review and lessons learned 10 years after start of enrolment. Annals of the Rheumatic Diseases. 2014. Jul 1;73(7):1289–300. [DOI] [PubMed] [Google Scholar]
- 8.Zhong H, Miller DJ, Urish KL. T2 map signal variation predicts symptomatic osteoarthritis progression: data from the Osteoarthritis Initiative. Skeletal Radiology. 2016. Jul 1;45(7):909–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wirth W, Maschek S, Beringer P, Eckstein F. Subregional laminar cartilage MR spin–spin relaxation times (T2) in osteoarthritic knees with and without medial femorotibial cartilage loss – data from the Osteoarthritis Initiative (OAI). Osteoarthritis and Cartilage. 2017. Aug 1;25(8):1313–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Kretzschmar M, Nevitt MC, Schwaiger BJ, Joseph GB, McCulloch CE, Link TM. Spatial distribution and temporal progression of T2 relaxation time values in knee cartilage prior to the onset of cartilage lesions – data from the Osteoarthritis Initiative (OAI). Osteoarthritis and Cartilage. 2019. May 1;27(5):737–45. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Razmjoo A, Caliva F, Lee J, Liu F, Joseph GB, Link TM, et al. T2 analysis of the entire osteoarthritis initiative dataset. Journal of Orthopaedic Research®. 2021. Jan 1;39(1):74–85. [DOI] [PubMed] [Google Scholar]
- 12.Fuerst D, Wirth W, Gaisberger M, Hunter DJ, Eckstein F. Association of Superficial Cartilage Transverse Relaxation Time With Osteoarthritis Disease Progression: Data From the Foundation for the National Institutes of Health Biomarker Study of the Osteoarthritis Initiative. Arthritis Care and Research. 2022;74(11):1888–93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Kretzschmar M, Heilmeier U, Yu A, Joseph GB, Liu F, Solka M, et al. Longitudinal analysis of cartilage T2 relaxation times and joint degeneration in African American and Caucasian American women over an observation period of 6 years – data from the Osteoarthritis Initiative. Osteoarthritis and Cartilage. 2016. Aug 1;24(8):1384–91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Pan J, Pialat JB, Joseph T, Kuo D, Joseph GB, Nevitt MC, et al. Knee cartilage T2 characteristics and evolution in relation to morphologic abnormalities detected at 3-T MR imaging: A longitudinal study of the normal control cohort from the osteoarthritis initiative. Radiology. 2011. Nov;261(2):507–15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Liebl H, Joseph G, Nevitt MC, Singh N, Heilmeier U, Subburaj K, et al. Early T2 changes predict onset of radiographic knee osteoarthritis: data from the osteoarthritis initiative. Annals of the Rheumatic Diseases. 2015. Jul 1;74(7):1353–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Edd SN, Omoumi P, Jolles BM, Favre J. Longitudinal femoral cartilage t2 relaxation time and thickness changes with fast sequential radiographic progression of medial knee osteoarthritis—data from the osteoarthritis initiative (Oai). Journal of Clinical Medicine. 2021. Mar 2;10(6):1–13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.MacKay JW, Low SBL, Smith TO, Toms AP, McCaskie AW, Gilbert FJ. Systematic review and meta-analysis of the reliability and discriminative validity of cartilage compositional MRI in knee osteoarthritis. Osteoarthritis and Cartilage. 2018. Sep 1;26(9):1140–52. [DOI] [PubMed] [Google Scholar]
- 18.Carballido-Gamio J, Joseph GB, Lynch JA, Link TM, Majumdar S. Longitudinal analysis of MRI T2 knee cartilage laminar organization in a subset of patients from the osteoarthritis initiative: A texture approach. Magnetic Resonance in Medicine. 2011. Apr 1;65(4):1184–94. [DOI] [PubMed] [Google Scholar]
- 19.Joseph GB, Baum T, Carballido-Gamio J, Nardo L, Virayavanich W, Alizai H, et al. Texture analysis of cartilage T2 maps: individuals with risk factors for OA have higher and more heterogeneous knee cartilage MR T2 compared to normal controls--data from the osteoarthritis initiative. Arthritis Res Ther. 2011;13(5):R153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Pedoia V, Lee J, Norman B, Link TM, Majumdar S. Diagnosing osteoarthritis from T2 maps using deep learning: an analysis of the entire Osteoarthritis Initiative baseline cohort. Osteoarthritis and Cartilage. 2019. Jul 1;27(7):1002–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Gatti AA, Keir PJ, Noseworthy MD, Maly MR. Investigating acute changes in osteoarthritic cartilage by integrating biomechanics and statistical shape models of bone: data from the osteoarthritis initiative. Magnetic Resonance Materials in Physics, Biology and Medicine. 2022. Oct 1;35(5):861–73. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Lebel RM, Wilman AH. Transverse relaxometry with stimulated echo compensation. Magnetic Resonance in Medicine. 2010;64(4):1005–14. [DOI] [PubMed] [Google Scholar]
- 23.Hennig J Multiecho imaging sequences with low refocusing flip angles. Journal of Magnetic Resonance (1969). 1988. Jul 1;78(3):397–407. [Google Scholar]
- 24.Hennig J Echoes—how to generate, recognize, use or avoid them in MR-imaging sequences. Part I: Fundamental and not so fundamental properties of spin echoes. Concepts in Magnetic Resonance. 1991. Jul 1;3(3):125–43. [Google Scholar]
- 25.Weigel M Extended phase graphs: Dephasing, RF pulses, and echoes - pure and simple. Journal of Magnetic Resonance Imaging. 2015. Feb 1;41(2):266–95. [DOI] [PubMed] [Google Scholar]
- 26.Prasloski T, Mädler B, Xiang QS, MacKay A, Jones C. Applications of stimulated echo correction to multicomponent T2 analysis. Magnetic Resonance in Medicine. 2012. Jun 1;67(6):1803–14. [DOI] [PubMed] [Google Scholar]
- 27.Marty B, Baudin PY, Reyngoudt H, Azzabou N, Araujo ECA, Carlier PG, et al. Simultaneous muscle water T2 and fat fraction mapping using transverse relaxometry with stimulated echo compensation. NMR in Biomedicine. 2016. Apr 1;29(4):431–43. [DOI] [PubMed] [Google Scholar]
- 28.Sveinsson B, Chaudhari AS, Gold GE, Hargreaves BA. A simple analytic method for estimating T2 in the knee from DESS. Magnetic Resonance Imaging. 2017. May 1;38:63–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Osteoarthritis Initiative (OAI) Study Protocol - OAI [Internet]. [cited 2024 Oct 7]. Available from: https://nda.nih.gov/oai/study-details
- 30.Ethical Conduct of Research - OAI [Internet]. [cited 2024 Oct 7]. Available from: https://nda.nih.gov/oai/ethical_conduct
- 31.Peterfy CG, Schneider E, Nevitt M. The osteoarthritis initiative: report on the design rationale for the magnetic resonance imaging protocol for the knee. Osteoarthritis and Cartilage. 2008. Dec 1;16(12):1433–41. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Peterfy CG, Schneider E, Nevitt M. The osteoarthritis initiative: report on the design rationale for the magnetic resonance imaging protocol for the knee. Osteoarthritis and Cartilage. 2008. Dec 1;16(12):1433–41. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging). IEEE Transactions on Medical Imaging. 1991. Mar;10(1):53–65. [DOI] [PubMed] [Google Scholar]
- 34.Lebel RM. StimFit: A toolbox for robust T2 mapping with stimulated echo compensation. In: Proc Intl Soc Mag Reson Med 20. 2012. [Google Scholar]
- 35.Raya JG, Dietrich O, Horng A, Weber J, Reiser MF, Glaser C. T2 measurement in articular cartilage: Impact of the fitting method on accuracy and precision at low SNR. Magnetic Resonance in Medicine. 2010. Jan 1;63(1):181–93. [DOI] [PubMed] [Google Scholar]
- 36.Santini F, Deligianni X, Paoletti M, Solazzo F, Weigel M, de Sousa PL, et al. Fast Open-Source Toolkit for Water T2 Mapping in the Presence of Fat From Multi-Echo Spin-Echo Acquisitions for Muscle MRI. Frontiers in Neurology. 2021. Feb 26;0:248. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, et al. Magnetic resonance fingerprinting. Nature. 2013. Mar;495(7440):187–92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Barbieri M, Hooijmans MT, Moulin K, Cork TE, Ennis DB, Gold GE, et al. A deep learning approach for fast muscle water T2 mapping with subject specific fat T2 calibration from multi-spin-echo acquisitions. Sci Rep. 2024. Apr 8;14(1):8253. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Fukushima K Cognitron: A self-organizing multilayered neural network. Biol Cybernetics. 1975. Sep 1;20(3):121–36. [DOI] [PubMed] [Google Scholar]
- 40.Barbieri M, Brizi L, Giampieri E, Solera F, Manners DN, Castellani G, et al. A deep learning approach for magnetic resonance fingerprinting: Scaling capabilities and good training practices investigated by simulations. Physica Medica. 2021;89:80–92. [DOI] [PubMed] [Google Scholar]
- 41.Barbieri M, Lee PK, Brizi L, Giampieri E, Solera F, Castellani G, et al. Circumventing the curse of dimensionality in magnetic resonance fingerprinting through a deep learning approach. NMR in Biomedicine. 2022. Apr 1;35(4). [DOI] [PubMed] [Google Scholar]
- 42.Chollet F, others. Keras [Internet]. 2015. Available from: https://keras.io
- 43.Abadi Martín, Agarwal Ashish, Barham Paul, Brevdo Eugene, Chen Zhifeng, Citro Craig, et al. TensorFlow: Large-Scale Machine Learning on Heterogeneous Systems [Internet]. 2015. Available from: https://www.tensorflow.org/
- 44.Gatti AA, Maly MR. Automatic knee cartilage and bone segmentation using multi-stage convolutional neural networks: data from the osteoarthritis initiative. MAGMA. 2021. Dec;34(6):859–75. [DOI] [PubMed] [Google Scholar]
- 45.Klein S, Staring M, Murphy K, Viergever MA, Pluim JPW. Elastix: A toolbox for intensity-based medical image registration. IEEE Transactions on Medical Imaging. 2010. Jan;29(1):196–205. [DOI] [PubMed] [Google Scholar]
- 46.McPhee KC, Wilman AH. Limitations of skipping echoes for exponential T2 fitting. Journal of Magnetic Resonance Imaging. 2018;48(5):1432–40. [DOI] [PubMed] [Google Scholar]
- 47.Li H, Daniel AJ, Buchanan CE, Nery F, Morris DM, Li S, et al. Improvements in Between-Vendor MRI Harmonization of Renal T2 Mapping using Stimulated Echo Compensation. J Magn Reson Imaging. 2024. Feb 21; [DOI] [PubMed] [Google Scholar]
- 48.Barbieri M, Watkins L, Desai AD, Mazzoli V, Rubin E, Schmidt A, et al. B1 Field Inhomogeneity Correction for qDESS T2 Mapping: Application to Rapid Bilateral Knee Imaging. In: Proc of ISMRM 29th Annual Meeting and Exibition. [DOI] [PMC free article] [PubMed] [Google Scholar]