Table 2.
Average technical parameters and radiation dose metrics across the included CT examinations
| Examinations | Protocols | |||
|---|---|---|---|---|
| Included units | 907,992 | 1767 | ||
| Mean | Standard deviation | Mean | Standard deviation | |
| Patient diameter (mm) | 308 | 45 | 309 | 23 |
| Technical parameters | ||||
| mAs | 178 | 87 | 178 | 63 |
| Pitch | 1.0 | 0.3 | 1.0 | 0.3 |
| kV | 118 | 8 | 118 | 7 |
| Scan length (mm) | 483 | 85 | 450 | 82 |
| Number of phases | 1.3 | 0.6 | 1.4 | 0.6 |
| Collimation | 37 | 17 | 35 | 17 |
| Radiation dose metrics | ||||
| Volumetric CT dose index (CTDIvol), mGy | 13 | 7 | 13 | 5 |
| Size-adjusted CTDIvol, mGy | 11 | 5 | 12 | 5 |
| Dose length product (DLP), mGy-cm | 756 | 496 | 775 | 418 |
| Size-adjusted DLP (mGy-cm) | 657 | 352 | 669 | 361 |
The six technical parameters listed here were used as inputs in the k-means clustering analysis. DLP dose length product