Assessment of automated loop-mediated isothermal amplification-(LAMP-)based xenomonitoring for Plasmodium spp. in Anopheles mosquitoes

Albert Eisenbarth 1,2,✉, Ayôla A Adegnika 3,4, Stravensky T Boussougou-Sambe 3,4, Lynn Glyschewski 2, Thomas Jacobs 2, Simone Kann 5, Andreas Krüger 5, Ralf M Hagen 5, Michael Ramharter 2,6, Esther Mehmel 2, Ange G Doumba Ndalembouly 3, Hagen Frickmann 1,7
PMCID: PMC13126929  PMID: 42050513

Abstract

Background

Xenomonitoring is an approach of epidemiological infection risk assessment addressing vector-transmitted infections like malaria in hematophagous arthropods. Standardization and automation can facilitate its use even in remote areas. In this study, the suitability of an automated commercial loop-mediated isothermal amplification (LAMP) assay, originally designed for the detection of malaria parasite DNA in human blood, was assessed for its applicability for xenomonitoring purposes.

Methods

An automated generic LAMP assay for malaria detection in human blood was applied with Anopheles spp. samples. The results were compared with commercial generic as well as species-specific real-time PCR.

Results

LAMP identified 15/43 (34.9%, 95% CI (20.1%; 50.9%)) mosquitoes artificially exposed to Plasmodium berghei, while this was the case for 22/43 (51.2%, 95% CI (35.5%; 66.7%)) samples applying real-time PCR. Considerably less discrepancy was observed with Anopheles spp. imported to Germany from field studies with 12/172 pools (7.0%, 95% CI (3.7%; 11.9%)) for LAMP and 13/172 pools (7.6%, 95% CI (4.1%; 12.6%)) for real-time PCR. Single testing and typing of DNA isolates indicated an overall infection rate of 1.2% (20/1711) with Plasmodium falciparum, P. malariae and P. ovale. Low rates of invalid results during a field exercise proved the general suitability of the LAMP approach for use at tropical settings.

Conclusions

Imperfect sensitivity in case of artificial exposure of mosquitoes to non-human plasmodial species was detected for both compared molecular approaches with slightly lower sensitivity of the LAMP approach. For Anopheles spp. infected with Plasmodium spp. causing human malaria, comparable diagnostic reliability could be shown for both approaches and the LAMP assay was shown to be suitable for application under tropical field conditions.

Supplementary Information

The online version contains supplementary material available at 10.1186/s12936-026-05917-w.

Keywords: Malaria, Anopheles, Plasmodium, Vector, Xenomonitoring

Background

Xenomonitoring is an epidemiological risk assessment approach targeting microorganisms in their transmitting vectors, usually applying molecular assays [1–3]. Some evidence exists on correlation between pathogen detection rates in vectors and associated human infection rates [4], although such data is not available for all vector-pathogen-associations. Next to general prevalence and transmission assessments [5–7], xenomonitoring has been successfully applied to monitor intervention approaches targeting vector-borne infectious disease transmission [8–10]. Sensitive molecular xenomonitoring techniques allow broad epidemiological sampling and testing without a need of usually more difficult-to-acquire human specimens [2], although the choice of the most appropriate vector collection methods remains an issue to consider [11–15]. Further, requirement of sophisticated laboratory equipment may be a challenge for xenomonitoring in resource-limited tropical settings where vector-borne infectious disease transmission frequently occur [16].

For xenomonitoring targeting Plasmodium spp. [17, 18], the causative agents of malaria, various molecular assays have been assessed and applied [19, 20]. They comprise conventional PCR [4] and real-time PCR [20, 21], but also other more sophisticated nucleic acid amplification techniques (NAT) like lateral flow recombinase polymerase amplification assays [19], digital PCR [20] and sequencing [22]. Sequence analysis-based xenomonitoring approaches for malaria have been described to allow analyses even beyond mere species identification, also including the identification of molecular resistance determinants of Plasmodium spp. [22]. For the latter purpose, target specific PCRs have been reported as well [23]. While Anopheles spp. are assessed for the xenomonitoring of causative agents of human malaria, DNA of avian Plasmodium spp. can also be detected in other vector species like Culex pipiens sensu lato [24]. Apart from mosquito samples, successful amplification of Plasmodium spp. DNA from vector feces/excreta has also been reported [16, 25–28].

Pathogen DNA stability within vectors is an issue to be considered when interpreting xenomonitoring results, because pathogen nucleic acid degradation readily occurs in the gut of arthropod vectors [29]. In spite of such limitations, correlations between xenomonitoring of Plasmodium spp. and malaria infection risk in human individuals is considered as well established [30]. Furthermore, infections in competent Anopheles spp. harbor generally higher amounts of detectable parasite DNA due to sexual reproduction and propagation of Plasmodium spp., inferring the malaria transmission risk at a given time point and location.

While variants of PCR represent the most commonly applied molecular detection approach for malaria [31–34], successful application of loop-mediated isothermal amplification (LAMP) has also been demonstrated for this purpose [35–39]. In the study presented here, we assessed a commercial automated LAMP approach, which detects Plasmodium spp. DNA in human blood samples, for its suitability for xenomonitoring of Plasmodium spp. in Anopheles spp., including its feasibility in the field setting.

Methods

Study design and sample materials

The assessment was conducted as a comparative test assessment between a molecular POCT (point-of-care-testing) assay and a benchtop device-based real-time PCR for the detection of Plasmodium spp. in Anopheles spp. mosquitoes. Whereas the LAMP instrument gave only three possible readouts positive, negative and invalid without further details, the real-time PCR indicated the cycle threshold (Ct) value as a proxy for initial template concentration of Plasmodium spp. parasite material.

In a first step, the proof-of-principle assessment was conducted with laboratory-bred An. stephensi artificially infected with the rodent malaria parasite Plasmodium berghei in the laboratories of the Bernhard Nocht Institute for Tropical Medicine Hamburg (Hamburg, Germany). Hereby, females which were fed with P. berghei containing mouse blood and incubated for 14–21 days to permit the development of oocysts and sporozoites. In a second step, the protocol was applied to Anopheles mosquitoes from hypo- and hyperendemic regions in Colombia, South America, and Gabon, Central Africa, respectively. The Gabonese mosquitoes were caught with CDC (John W. Hock, Gainesville, USA) and BG Pro (Biogents, Regensburg, Germany) light traps in the provinces Moyen-Ogooué and Ngounié between October 2021 and November 2024. The traps were operated in variable configurations, including light of the visible and ultraviolet spectrum, and the supplementation with carbon dioxide and mosquito attractant (BG Lure, Biogents, Regensburg, Germany). They were positioned in the vicinity of human settlements predominantly inside occupied houses, thus increasing the likelihood of the anthropophagic mosquitoes for being infected with Plasmodium spp. associated with human malaria. An available limited number of Colombian Anopheles mosquitoes from the Cesar department—collected in August 2021—was directly assessed, while the Gabonese mosquitoes were investigated as 10 × pooled homogenates and only positive pools were retested with sample materials from individual arthropods. Transfer of biological material to Germany occurred by air freight either with dry ice (Colombia) or with cooling pads (Gabon). Transport duration varied between 2 and 18 days. In a final step, applicability of the POCT assay under tropical field conditions was tested in a basic entomological laboratory in a hyperendemic Gabonese region during the rainy season with freshly caught Anopheles spp. from the adjacent area. During this final step, no confirmatory testing with real-time PCR was done. A workflow indicating the different steps is shown in Fig. 1.

Fig. 1.

Fig. 1

Workflow as applied during the comparative study. inf. = infected

Identification of vector mosquitoes for the study

Blood-seeking female Anopheles spp. were morphologically identified according to taxonomic keys [40, 41]. For representatives of identified species, DNA barcoding of the mitochondrial cytochrome c oxidase (COI) locus was applied [42, 43].

Laboratory diagnostics

Sample preparation

Female Anopheles spp. mosquitoes were manually mechanically homogenized in 100 µL PBS solution using reaction tube pestles (Eppendorf, Hamburg, Germany) and 1.5 ml Safe seal reaction tubes (Sarstedt, Nümbrecht, Germany). Subsequently, lysis was performed with 12 units proteinase K (Qiagen, Hilden, Germany) for 5 h or overnight at a temperature of 56 °C.

For subsequent loop-mediated amplification, 33.5 pg in 5 µL volume (6.7 pg/ µL) of an internal control plasmid (pEX-A258 vector backbone, Eurofins Genomics, Ebersberg, Germany) containing an insert of the bases 3307 to 3784 of human mitochondrial NADH (nicotinamide adenine dinucleotide dehydrogenase) subunit 1 (ND1, GenBank accession number NC_012920) were added to each reaction.

For subsequent real-time PCR, the lysed mosquito homogenates were subjected to DNA extraction applying the commercial DNeasy Blood and Tissue kit (Qiagen, Hilden, Germany) as described by the manufacturer’s protocol for insects [44].

Loop-mediated isothermal amplification-based point-of-care testing

For the LAMP-based POCT approach applying the commercially available Alethia Malaria LAMP kit (Meridian Biosciences, Cincinnati, USA), 50 µL volumes of lysed mosquito homogenates were included in each LAMP reaction as described by the manufacturer with a minor modification. This modification consisted of the inclusion of the internal control plasmid (approx. 0.61 pg/ µL reaction volume, as indicated by preceding titration tests) as stated in the sample preparation chapter above in order to compensate for lacking human cells within the mosquito samples. When applied with human blood as intended by the manufacturer, this step is obsolete, because human mitochondrial NADH subunit 1 naturally occurs in this type of specimen. After pooling of the mosquito lysates no further dilution was done before chemical extraction. Each 50 µL of extract was added to the lyophilized reagents for Plasmodium and internal control LAMP and incubated at 63 °C for 45 min. Results were automatically assigned by the instrument either as positive, negative or invalid. In case of an invalid test result of the POCT assay, the assay was repeated once. In case of persistent invalidity in the second run as well, this was accepted as the final result. No positive or negative controls were used.

With human blood, diagnostic accuracy of the assay, which was formerly known as illumigene Malaria LAMP, has been estimated to be in the 96.4–98.7% sensitivity range with 99.6% specificity compared to composite reference standards as detailed elsewhere [45, 46].

Confirmatory testing applying real-time PCR

Commercial generic malaria real-time PCR (RealStar Malaria PCR Kit 1.0, Altona Diagnostics, Hamburg, Germany) targeting Plasmodium spp. as well as commercial real-time PCR targeting the causative agents of human malaria P. falciparum, P. vivax, P. ovale complex, P. malariae and P. knowlesi in a multiplex approach (RealStar Malaria Screen and Type PCR Kit 1.0, Altona Diagnostics, Hamburg, Germany) were applied with the mosquito samples as applicable. Of note, the multiplex PCR did not allow a further discrimination of P. ovale complex into P. ovale wallikeri and P. ovale curtisi. In particular, the generic assay was required to detect P. berghei for the proof-of-principle assessment within the artificially infected mosquitoes, because this species was not included in the multiplex assay. After sample preparation and DNA extraction as stated above, the real-time PCR reactions were run exactly as described by the manufacturer on Corbett Q cyclers (Qiagen, Hilden, Germany). Briefly, 10 µL sample or control was mixed with 20 µL Mastermix per reaction. DNA-grade water as no template control and the proprietary positive controls were used for each run, as well as a no template control containing premixed inhibition control (20 µL Mastermix + 1 µL internal control per reaction). Cycling conditions were 95 °C for 2 min (RealStar Malaria PCR Kit 1.0) or 10 min (RealStar Malaria Screen & Type PCR Kit 1.0), followed by 45 cycles of 95 °C, 15 s, 58 °C, 45 s and 72 °C, 15 s. Results were considered valid when positive and internal controls were detected above threshold levels and the negative control gave no signal above the threshold.

Diagnostic accuracy of the generic malaria real-time assay with human blood was formerly estimated to be 98.4% sensitivity and 99.9% specificity [47], while sensitivity dropped to a range of 94.0%-95.1% in case of subsequent species-level differentiation with the multiplex assay in human blood samples [45, 47] compared to composite reference standards.

Pooling approach

For the mosquito specimens collected in Gabon and shipped to Germany for the assessments, 10 × pooling of the homogenates was conducted. Optimal pooling size was determined by mixture of positively tested individuals with uninfected An. stephensi specimens at differing proportions (1–10 infected per pool). In case of a Plasmodium-positive result in any of the reactions, assessments were repeated with sample material of individual mosquitoes from the respective pool.

Statistics

Due to the low numbers of positive test results in the analyzed samples, concordant and discordant results of the compared assays were descriptively presented with only basic analytic statistical assessment. Considering the lack of a true “gold standard”, i.e., a method with a hypothetical sensitivity and specificity of 100% each [48], the presentation of diagnostic accuracy values in terms of sensitivity and specificity was omitted. Inter-rater agreement kappa was calculated with the software MedCalc version 23.5.0 (MedCalc Software Ltd., Ostende, Belgium), 95% confidence intervals of detection rates with the software GraphPad InStat version 3.06 (GraphPad Software, Inc., San Diego, CA, USA).

Results

Identification of vector mosquitoes from the wild catches

Overall, 2349 female Anopheles spp. were included in the study (Table 1), of which the majority (97.1%, n = 2282) was collected during field collections in Gabon. Anopheles stephensi was only present in the lab-reared colony based in Germany. Out of the remaining eleven field collected species, five (93.9%, n = 2165) are known to be medically important malaria vectors in Gabon (An. gambiae s.l., An. moucheti s.s., An. nili s.s., An. ziemanni) and Colombia (An. triannulatus s.l.).

Table 1.

List of Anopheles spp. included in the study. COI barcoding data on selected species are indicated

Anopheles species Country n [%] COI gene Acc.-No
An. carnevalei* Gabon 24 1.0
An. demeilloni Gabon 1 0.04
An. gabonensis Gabon 1 0.04
An. gambiae sensu lato* Gabon 1,295 54.8 PX307520–PX307521
An. moucheti sensu stricto* Gabon 741 30.9 PX307522
An. namibiensis* Gabon 69 2.9 PX307523
An. nili sensu stricto* Gabon 18 0.8 -
An. paludis* Gabon 87 3.7 PX307524
An. ruarinus Gabon 1 0.04
An. stephensi* Germany** 43 1.8
An. triannulatus sensu lato* Colombia 24 1.0 PX307525–PX307526
An. ziemanni* Gabon 3 0.1 PX307527
unidentified Anopheles spp. Gabon 42 1.8
Total 2,349 100.0

*Competent vectors for Plasmodium spp. transmission

**Laboratory-reared population. Endemic on the Indian subcontinent and Eastern Africa

Proof of principle with mosquitoes artificially infected with P. berghei

A total of 43 female Anopheles stephensi, which had been fed with blood containing P. berghei under controlled conditions in an insectary, were subjected to molecular assessment applying LAMP and generic real-time PCR. Generic real-time PCR led to amplification curves of Plasmodium spp. in 22 cases (51.2%, 95% CI (35.5%; 66.7%)). From these, 15 (68.2%) showed a positive LAMP result (34.9%, 95% CI (21.0%; 50.9%), while invalid and negative results were observed in 2 (9.1%) and 5 (22.7%) instances, respectively. Inter-rater agreement kappa was 0.68 (0.47, 0.88). No invalid LAMP results were recorded for the mosquito samples which remained negative in real-time PCR. Mean value and standard deviation (SD) of the Ct values measured by real-time PCR were 28.9 (± 3.3). Details are provided in Table 2.

Table 2.

Comparison of the test results of the LAMP-based POCT assay and real-time PCR, including the classification of Plasmodium spp.

Anopheles species An. stephensi An. triannulatus Anopheles spp. Anopheles spp. An. gambiae s.l
Country of origin Germany* Colombia Gabon Gabon Gabon
Infection status artificial natural natural natural natural
Diagnostic laboratory Germany Germany Germany Gabon Gabon
Collected 43 24 1,729 18 535
Analyzed 43 (100.0) 24 (100.0) 1,711 (99.0) 18 (100.0) 535 (100.0)
Pools 172 54
LAMP  +  15 (34.9) 0 (0.0) 12 (7.0) 1 (5.6) 13 (24.1)
 −  26 (60.5) 24 (100.0) 160 (93.0) 16 (88.9) 39 (72.2)
inv 2 (4.7) 0 (0.0) 0 (0.0) 1 (5.6) 2 (3.7)
real-time PCR  +  22 (51.2) 0 (0.0) 13 (7.6)
Ct [SD] 28.9 [± 3.3] 27.7 [± 2.0]
 −  21 (48.8) 24 (100.0) 159 (92.4)
inv 0 (0.0) 0 (0.0) 0 (0.0)
Test comparison LAMP real-time PCR
 +   +  15 (34.9) 0 (0.0) 11 (6.4)
Ct [SD] 29.2 [± 3.3] 27.4 [± 1.9]
 −   −  21 (48.8) 24 (100.0) 158 (91.9)
inv  +  2 (4.7) 0 (0.0) 0 (0.0)
Ct [SD] 24.5 [± 3.5]
inv  −  0 (0.0) 0 (0.0) 0 (0.0)
 +   −  0 (0.0) 0 (0.0) 1 (0.6)
 −   +  5 (11.6) 0 (0.0) 2 (1.2)
Ct [SD] 29.6 [± 2.1] 29.5 [± 2.1]
Single retesting LAMP real-time PCR 176
 +   +  17/106 (16.0)
Ct [SD] 27.8 [± 2.8]
 −   −  0/40 (0.0)
 +   −  3/10 (30.0)
 −   +  7/20 (35.0)
Ct [SD] 31.0 [± 3.8]
specific real-time PCR +  14/141 (10.0)
Plasmodium species P. berghei

12 × P. falciparum

3 × P. malariae

2 × P. ovale

Percentages in brackets. LAMP = Malaria LAMP (Alethia); real-time PCR = Malaria PCR Kit 1.0 (Altona Diagnostics); specific real-time PCR = Malaria Screen & Type PCR Kit 1.0 (Altona Diagnostics); inv = invalid test result. +  = positive test result.  − = negative test result. Ct = cycle threshold value. SD = standard deviation. *laboratory-reared population. Endemic on the Indian subcontinent and Eastern Africa

Comparative testing with LAMP and real-time PCR in mosquitoes imported from endemic regions

In 24 Anopheles spp. from the low-endemicity setting Colombia, real-time PCR and LAMP showed concordantly negative results in all instances. Focusing on samples from Gabon, 172 pools from a total of 1711 Anopheles spp. mosquitoes were included in the analysis. From those pools, 11 (6.4%) were concordantly positive and 158 (91.9%) were concordantly negative for both methods. From the remaining 3 (1.8%) discrepant results, 2 (1.2%) were only tested positive in the real-time PCR, whereas 1 (0.6%) sample showed only a positive LAMP test result, with kappa of 0.87 (0.72, 1.00) for LAMP (12/172, 7.0%, 95% CI (3.7%; 11.9%) and real-time PCR (13/172, 7.6%, 95% CI (4.1%; 12.6%)). The affected pools (n = 14) and a subset of negatively tested specimens both by LAMP and real-time PCR (n = 4) were subjected to further investigation of the respective DNA isolates (n = 166), including a fraction of available homogenates of single individuals (n = 10). Concordant negative and positive results were confirmed by the corresponding absence or presence, respectively, of at least one real-time PCR-positive mosquito per pool (17 positives in 146 isolates). Plasmodium spp. identification by the species-specific real-time PCR failed in 3/17 real-time PCR-positives derived from two pools, of which one pool had no other confirmed species. The sub-analyses of discordant results indicated 3/10 LAMP-positive and 7/20 real-time PCR-positive specimens, of which only 1/20 real-time PCR sample was confirmed by specific real-time PCR as P. falciparum.

In total, multiplex real-time PCR identified P. falciparum in 12 mosquitoes (An. gambiae s.l.: n = 6; An. moucheti s.s.: n = 5; Anopheles sp.: n = 1), P. malariae in three mosquitoes (An. gambiae s.l.: n = 2; An. moucheti s.s.: n = 1) and P. ovale complex in two (one each in An. gambiae s.l. and An. moucheti s.s.). Thereby, co-infections occurred in three individuals, comprising P. falciparum—P. malariae (n = 2, one each in An. gambiae s.l. and An. moucheti s.s.) and P. falciparum—P. ovale complex (n = 1, An. moucheti s.s.). Mean value and standard deviation (SD) of the Ct values measured by real-time PCR were 27.5 (± 2.4) and 31.0 (± 3.8) for pool samples and discordantly tested single samples, respectively. Details are provided in Table 2.

Applicability assessment of the LAMP approach in a tropical area of endemicity

During two field studies in the Gabonese high-endemicity setting for human malaria, 553 female Anopheles spp. were caught and assessed individually (n = 18) or in pools (n = 54) for causative agents of malaria with the LAMP device. Valid results were achieved from 69/72 samples (95.8%). One (5.6%) Plasmodium sp. was detected in the single-tested samples, and 13 (24.1%) were detected in the pooled samples. Details are provided in Table 2. Comparative real-time PCR testing was not available during this part of the study.

Discussion

The study was conducted to assess the suitability of an automated LAMP-based POCT-assay for the detection of malaria in vectors via xenomonitoring, since the assay was originally designed for malaria detection in human blood samples. Commercial real-time PCR targeting the Plasmodium genus and human-pathogenic malaria species on traditional benchtop devices was used for the comparison.

The assessment with the artificially infected mosquitoes indicated that neither LAMP nor real-time PCR identified all mosquitoes which had been exposed to P. berghei-containing blood (51.2% and 34.9% for real-time PCR and LAMP, respectively), even when considering that not every exposed individual develops an infection [49]. This is well in line with previous concerns on sensitivity limitations even when highly sensitive molecular methods are applied for xenomonitoring [29] and should be kept in mind when interpreting respective molecular surveillance data. Moderate Ct-values (mean: 28.9 ± 3.3), as observed for the artificially infected mosquitoes, did not suggest the abundance of very low target DNA amounts close to the detection limit of the assays and so, merely stochastic effects based on critically low Plasmodium spp. DNA quantities does not seem plausible. In fact, two PCR-positive but LAMP-negative samples showed nearly the same mean Ct values (29.6 ± 2.1) compared with the concordantly positively tested samples (29.2 ± 3.3). Thereby, an identical exposition setting of the artificially infected mosquitoes makes effects of target sequence variance unlikely. Interestingly, real-time PCR still showed higher sensitivity compared to LAMP, although neither the real-time PCR assay nor the LAMP assay were specifically designed to detect P. berghei, which is not associated with human disease and may show sequence variations at the oligonucleotide bindings sites of the target sequences. The reason for the likely discrepancy in the binding affinity of the used oligonucleotides remain elusive, as the producers did not provide information on sequences of the oligonucleotides used in their assays. Anyway, it remains unresolved whether or not animal-associated Plasmodium spp., which infrequently cause spill-over infections in humans [50], can be reliably detected in mosquitoes with the assessed molecular approaches, as respective samples were not available for the analysis. More than this, it is difficult to explain why invalid LAMP-results occurred exclusively in real-time PCR-positive artificially infected mosquitoes with considerably lower Ct-values (24.5 ± 3.5) and not in negatively tested ones. However, considering the low overall number of only two cases, random effects cannot be excluded here.

Entomological assessment of the Anopheles spp. mosquitoes caught in tropical areas of endemicity showed a very high proportion of anthropophilic vector species of medical relevance (Table 1). This was expected, since the trapping setup was optimized for mosquitoes with human host preference mostly during the seasonal peaks of local malaria incidences. Remarkably, three species, namely An. demeilloni, An. namibiensis and An. ruarinus were not included in the geo-coded inventory of Afrotropical anophelines [51], and only An. demeilloni has been previously reported in Gabon in one entomological study [52]. Although our study is the first which confirms the occurrence of An. namibiensis in Gabon, it has been identified in the neighboring country Cameroon in at times high abundance and is known to attack humans [53]. Here, we provide the first genomic entry of the COI locus at GenBank (Acc. No. PX307523) for this species, which DNA sequence is almost identical with an unidentified Anopheles sp. entry from Kenya (Acc No. MK533643, 99.8% pairwise identity) and closely clustering with entries of An. coustani, An. tenebrosus and An. ziemanni.

Focusing on the molecular xenomonitoring with the Anopheles spp., excellent specificity was shown for both real-time PCR and LAMP used for the analysis of the specimens from a low-endemicity focus in Colombia. The assessment of the Gabonese mosquitoes showed similar results for real-time PCR and LAMP with regards to the detection of Plasmodium spp. causing human malaria. Further, the recorded infection rates and plasmodial species distribution matched previously made observations in Gabon [54, 55]. Nonetheless, partial degradation of parasite DNA during storage and transport below the limit of detection could lead to a reduced sensitivity. Interestingly, several cases of co-infections in the same mosquito could be confirmed by real-time PCR. Positive LAMP results without positive real-time PCR, which occurred in 0.6% (n = 1) of the dataset from Gabon (Table 2), remained difficult to interpret. Both imperfect specificity of malaria LAMP [45, 46] or imperfect sensitivity of real-time PCR [45, 47] could explain the observation. In particular, non-human plasmodial species not covered by the real-time PCR oligonucleotides would make the latter option likely; but the true reason remained unresolved.

Altogether, applicability of LAMP-based POCT-xenomonitoring in the tropical high-endemicity setting Gabon was considered as successful, because the rate of invalid test results was similarly low as observed under laboratory conditions in Germany. Although the setup forbade the inclusion of confirmatory real-time PCR in the field, the moderate rate of plasmodial detections of 14/553 (2.5%), assuming each positively tested pool contained one infected individual, was well in line with expected infection rates in Gabonese mosquitoes [54]. All positive and a subset of negative PCR pools were confirmed by single testing by real-time PCR. Whereas the higher Ct-values in the two LAMP discordant results (29.5 ± 2.1 vs. 27.4 ± 1.9 LAMP concordant results) may be an indicator for lower LAMP sensitivity, the aggregation of 6/8 (75%) of unconfirmed specimens by specific PCR in a single PCR-positive pool with negative LAMP result challenges this assumption. To increase the field applicability of the assay for xenodiagnosis of malaria, providing an Anopheles-specific internal control would remove the need for manual addition of human DNA and improve assay usability for xenomonitoring.

The study is not without limitations. First of all, mosquitoes with standardized exposure to Plasmodium spp. causing human malaria were not available. For this reason, mosquitoes infected with P. berghei had to be used in the first phase of the study, for which the applied diagnostic test assays were not optimized. It remains uncertain whether sensitivity limitations, as observed with the mosquitoes artificially exposed to P. berghei, would have been recorded with plasmodial species causing human malaria as well. Second, due to the overall low infection rates in the collected mosquitoes, discrepancies were predominantly descriptively assessed. Third, reasons for individual discrepancies between the compared molecular assays remained unresolved in the absence of a reference standard with perfect diagnostic accuracy. Future studies should address these limitations and assess the practical applicability of the LAMP-based POCT-approach for on-site xenomonitoring in the tropics.

Conclusions

In spite of the above-mentioned limitations, the LAMP-based POCT-approach may be a useful tool for field xenomonitoring of human Plasmodium spp., but differences in sensitivity (notably for non-human Plasmodium) and the lack of a perfect reference standard warrant further validation.

Supplementary Information

Acknowledgements

Annett Michel and Simone Priesnitz are gratefully acknowledged for excellent technical assistance.

Abbreviations

Acc.-No.

Accession number

An.

Anopheles

COI

Cytochrome c oxidase subunit I

Ct

Cycle threshold

DNA

Deoxyribonucleic acid

Inf.

Infected

LAMP

Loop-mediated isothermal amplification

NADH

Nicotinamide adenine dinucleotide hydrogenase

NAT

Nucleic acid amplification technique

PBS

Phosphate buffered saline

PCR

Polymerase chain reaction

P.

Plasmodium

Pg

Picogram

POCT

Point-of-care-testing

SD

Standard deviation

spp.

Species

Author contributions

Conceptualization: HF, AK, AE; methodology: AE, HF; software: AE; validation: AE; formal analysis: AE; investigation: AE; resources: HF, AE, SK; data curation: AE; writing—original draft preparation: HF, AE; writing—review and editing: HF, AE, AF, AAA, STBS, LG, EM, TJ, SK, RH, MR, AGDN; visualization: AE; supervision: HF, AE; project administration: HF, AE; funding acquisition: AK, AE. All authors have read and approved the final manuscript.

Funding

The present study was supported by the German Ministry of Defense (grant 38K2-S-43 2123) “Evaluation of automated malaria LAMP for the detection of parasites in vectors”.

Data availability

Data is provided within the manuscript or supplementary information files. Sequence data that support the findings of this study have been deposited in the GenBank Nucleotide Archive with the primary accession code PX307520-307527.

Declarations

Ethics approval and consent to participate

In line with National German laws, ethical clearance was not needed. The reason is that neither humans nor human-related samples were included in the study. Further, German animal protection laws do not apply for mosquitoes. Local research permit has been granted by the Ministry of Research and Technological Development in Gabon.

Consent for publication

Not applicable.

Competing interest

The authors declare no competing interests.

Footnotes

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Vieira CJSP, Onn MB, Shivas MA, Shearman D, Darbro JM, Graham M, et al. Long-term co-circulation of multiple arboviruses in southeast Australia revealed by xeno-monitoring and viral whole-genome sequencing. Virus Evolution. 2024;10:0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Howlett M, Mayfield HJ, McPherson B, Rigby L, Thomsen R, Williams SA, et al. Molecular xenomonitoring as an indicator of microfilaraemia prevalence for lymphatic filariasis in Samoa in 2019. Parasit Vectors. 2024;17:382. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Cameron MM, Ramesh A. The use of molecular xenomonitoring for surveillance of mosquito-borne diseases. Philos Trans R Soc Lond B Biol Sci. 2021;376:20190816. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.McPherson B, Mayfield HJ, McLure A, Gass K, Naseri T, Thomsen R, et al. Evaluating molecular xenomonitoring as a tool for lymphatic filariasis surveillance in Samoa, 2018–2019. Trop Med Infect Dis. 2022;7:203. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Riches N, Badia-Rius X, Mzilahowa T, Kelly-Hope LA. A systematic review of alternative surveillance approaches for lymphatic filariasis in low prevalence settings: Implications for post-validation settings. PLoS Negl Trop Dis. 2020;14:e0008289. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.da Costa CF, Dos Passos RA, Lima JBP, Roque RA, de Souza SV, Campolina TB, et al. Transovarial transmission of DENV in Aedes aegypti in the Amazon basin: a local model of xenomonitoring. Parasit Vectors. 2017;10:249. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Moustafa MA, Salamah MMI, Thabet HS, Tawfik RA, Mehrez MM, Hamdy DM. Molecular xenomonitoring (MX) and transmission assessment survey (TAS) of lymphatic filariasis elimination in two villages, Menoufyia Governorate, Egypt. Eur J Clin Microbiol Infect Dis. 2017;36:1143–50. [DOI] [PubMed] [Google Scholar]
  • 8.Ramesh A, Oliveira P, Cameron M, Castanha PMS, Walker T, Lenhart A, et al. Postintervention immunological and entomological survey of lymphatic filariasis in the city of Olinda, Brazil, 2015–2016. Am J Trop Med Hyg. 2024;110:470–82. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Lau CL, Won KY, Lammie PJ, Graves PM. Lymphatic filariasis elimination in American Samoa: evaluation of molecular xenomonitoring as a surveillance tool in the endgame. PLoS Negl Trop Dis. 2016;10:e0005108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Schmaedick MA, Koppel AL, Pilotte N, Torres M, Williams SA, Dobson SL, et al. Molecular xenomonitoring using mosquitoes to map lymphatic filariasis after mass drug administration in American Samoa. PLoS Negl Trop Dis. 2014;8:e3087. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Coulibaly S, Sawadogo SP, Nikièma AS, Hien AS, Bamogo R, Koala L, et al. Assessment of Culicidae collection methods for xenomonitoring lymphatic filariasis in malaria co-infection context in Burkina Faso. PLoS Negl Trop Dis. 2024;18:e0012021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Verhulst NO, Juurlink M, Wondwosen B, Rugaimukamu S, Hill SR, Ignell R, et al. Fermenting molasses and a synthetic odour blend to attract blood-fed Anopheles coluzzii. Med Vet Entomol. 2023;37:228–37. [DOI] [PubMed] [Google Scholar]
  • 13.Coulibaly YI, Sangare M, Dolo H, Doumbia SS, Coulibaly SY, Dicko I, et al. Comparison of different sampling methods to catch lymphatic filariasis vectors in a Sudan savannah area of Mali. Am J Trop Med Hyg. 2022;106:1247–53. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Pi-Bansa S, Osei JHN, Joannides J, Woode ME, Agyemang D, Elhassan E, et al. Implementing a community vector collection strategy using xenomonitoring for the endgame of lymphatic filariasis elimination. Parasit Vectors. 2018;11:672. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Opoku M, Minetti C, Kartey-Attipoe WD, Otoo S, Otchere J, Gomes B, et al. An assessment of mosquito collection techniques for xenomonitoring of anopheline-transmitted lymphatic filariasis in Ghana. Parasitology. 2018;145:1783–91. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Higgins M, Kristan M, Collins EL, Messenger LA, Dombrowski JG, Vanheer LN, et al. A pan Plasmodium lateral flow recombinase polymerase amplification assay for monitoring malaria parasites in vectors and human populations. Sci Rep. 2024;14:20165. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Farlow R, Russell TL, Burkot TR. Nextgen vector surveillance tools: sensitive, specific, cost-effective and epidemiologically relevant. Malar J. 2020;19:432. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Moin-Vaziri V, Djadid ND, Hoosh-Deghati H, Atta H, Raz AA, Seyyed-Tabaei SJ, et al. Molecular detection of Plasmodium infection among anophelinae mosquitoes and differentiation of biological forms of Anopheles stephensi collected from malarious areas of Afghanistan and Iran. Ethiop J Health Sci. 2022;32:269–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Garrigós M, Veiga J, Garrido M, Marín C, Recuero J, Rosales MJ, et al. Avian Plasmodium in invasive and native mosquitoes from southern Spain. Parasit Vectors. 2024;17:40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Pilotte N, Cook DAN, Pryce J, Zulch MF, Minetti C, Reimer LJ, et al. Laboratory evaluation of molecular xenomonitoring using mosquito and tsetse fly excreta/feces to amplify Plasmodium, Brugia, and Trypanosoma DNA. Gates Open Res. 2020;3:1734. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Irish SR, Al-Amin HM, Paulin HN, Mahmood ASMS, Khan RK, Muraduzzaman AKM, et al. Molecular xenomonitoring for Wuchereria bancrofti in Culex quinquefasciatus in two districts in Bangladesh supports transmission assessment survey findings. PLoS Negl Trop Dis. 2018;12:e0006574. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Nkemngo FN, Mugenzi LMJ, Tchouakui M, Nguiffo-Nguete D, Wondji MJ, Mbakam B, et al. Xeno-monitoring of molecular drivers of artemisinin and partner drug resistance in P. falciparum populations in malaria vectors across Cameroon. Gene. 2022;821:146339. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Sorosjinda-Nunthawarasilp P, Bhumiratana A. Ecotope-based entomological surveillance and molecular xenomonitoring of multidrug resistant malaria parasites in Anopheles vectors. Interdiscip Perspect Infect Dis. 2014;2014:969531. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Martínez-de la Puente J, Soriguer R, Senar JC, Figuerola J, Bueno-Mari R, Montalvo T. Mosquitoes in an urban zoo: identification of blood meals, flight distances of engorged females, and avian malaria infections. Front Vet Sci. 2020;7:460. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Minetti C, Pilotte N, Zulch M, Canelas T, Tettevi EJ, Veriegh FBD, et al. Field evaluation of DNA detection of human filarial and malaria parasites using mosquito excreta/feces. PLoS Negl Trop Dis. 2020;14:e0008175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Cook DAN, Pilotte N, Minetti C, Williams SA, Reimer LJ. A superhydrophobic cone to facilitate the xenomonitoring of filarial parasites, malaria, and trypanosomes using mosquito excreta/feces. Gates Open Res. 2018;1:7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Pilotte N, Zaky WI, Abrams BP, Chadee DD, Williams SA. A novel xenomonitoring technique using mosquito excreta/feces for the detection of filarial parasites and malaria. PLoS Negl Trop Dis. 2016;10:e0004641. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Pryce J, Pilotte N, Menze B, Sirois AR, Zulch M, Agbor JP, et al. Integrated xenosurveillance of Loa loa, Wuchereria bancrofti, Mansonella perstans and Plasmodium falciparum using mosquito carcasses and faeces: a pilot study in Cameroon. PLoS Negl Trop Dis. 2022;16:e0010868. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Drummond C, Gebhardt ME, Sáenz Robles MT, Carpi G, Hoyer I, Pastusiak A, et al. Stability and detection of nucleic acid from viruses and hosts in controlled mosquito blood feeds. PLoS ONE. 2020;15:e0231061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Nascimento J, Sampaio VS, Karl S, Kuehn A, Almeida A, Vitor-Silva S, et al. Use of anthropophilic culicid-based xenosurveillance as a proxy for Plasmodium vivax malaria burden and transmission hotspots identification. PLoS Negl Trop Dis. 2018;12:e0006909. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Yalley AK, Ocran J, Cobbinah JE, Obodai E, Yankson IK, Kafintu-Kwashie AA, et al. Advances in malaria diagnostic methods in resource-limited settings: a systematic review. Trop Med Infect Dis. 2024;9:190. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Gómez-Hoyos R, Cardona-Arias JA, Higuita Gutiérrez LF, Salas-Zapata W, Carmona-Fonseca J. Systematic review of the diagnostic accuracy of thick smear compared to polymerase chain reaction for pregnancy-associated malaria, 2010-2022. Rev Peru Med Exp Salud Publica. 2022;39:302–11. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Costa GL, Alvarenga DAM, Aguiar ACC, Louzada J, Pereira DB, de Oliveira TF, et al. Improving the molecular diagnosis of malaria: droplet digital PCR-based method using saliva as a DNA source. Front Microbiol. 2022;13:882530. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Oyegoke OO, Maharaj L, Akoniyon OP, Kwoji I, Roux AT, Adewumi TS, et al. Malaria diagnostic methods with the elimination goal in view. Parasitol Res. 2022;121:1867–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Antinori S, Ridolfo AL, Grande R, Galimberti L, Casalini G, Giacomelli A, et al. Loop-mediated isothermal amplification (LAMP) assay for the diagnosis of imported malaria: a narrative review. Infez Med. 2021;29:355–65. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Feleke DG, Alemu Y, Yemanebirhane N. Performance of rapid diagnostic tests, microscopy, loop-mediated isothermal amplification (LAMP) and PCR for malaria diagnosis in Ethiopia: a systematic review and meta-analysis. Malar J. 2021;20:384. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Morris U, Aydin-Schmidt B. Performance and application of commercially available loop-mediated isothermal amplification (LAMP) kits in malaria endemic and non-endemic settings. Diagnostics (Basel). 2021;11:336. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Picot S, Cucherat M, Bienvenu AL. Systematic review and meta-analysis of diagnostic accuracy of loop-mediated isothermal amplification (LAMP) methods compared with microscopy, polymerase chain reaction and rapid diagnostic tests for malaria diagnosis. Int J Infect Dis. 2020;98:408–19. [DOI] [PubMed] [Google Scholar]
  • 39.Imai K, Tarumoto N, Misawa K, Runtuwene LR, Sakai J, Hayashida K, et al. A novel diagnostic method for malaria using loop-mediated isothermal amplification (LAMP) and MinION™ nanopore sequencer. BMC Infect Dis. 2017;17:621. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Coetzee M. Key to the females of Afrotropical Anopheles mosquitoes (Diptera: Culicidae). Malar J. 2020;19:1–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Sallum MAM, Obando RG, Carrejo N, Wilkerson RC. Identification keys to the Anopheles mosquitoes of South America (Diptera: Culicidae). IV. Adult females. Parasit Vectors. 2020;13:584. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Hebert PD, Cywinska A, Ball SL, deWaard JR. Biological identifications through DNA barcodes. Proc Biol Sci. 2003;270:313–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Krüger A, Balczun C, Scheid PL, Hagen RM, Eisenbarth A. Molecular detection of Leishmania (Sauroleishmania) adleri (Trypanosomatida: Trypanosomatidae) in Sergentomyia sp. sand flies (Diptera: Psychodidae) in Mali and Niger. Acta Trop. 2023;243:106936. [DOI] [PubMed] [Google Scholar]
  • 44.Purification of total DNA from insects using the DNeasy® Blood & Tissue Kit (DY14 Aug-06). Qiagen Supplementary Protocol. 2006. https://www.qiagen.com/us/resources/resourcedetail?id=cabd47a4-cb5a-4327-b10d-d90b8542421e&lang=en.com. Accessed 9 Sep 2025.
  • 45.Altangerel E, Frickmann H. Meta-analysis of the diagnostic performance characteristics of three commercial and one in-house nucleic acid amplification tests for malaria screening. J Lab Med. 2020;44:47–53. [Google Scholar]
  • 46.Frickmann H, Hinz R, Rojak S, Bonow I, Ruben S, Wegner C, et al. Evaluation of automated loop-mediated amplification (LAMP) for routine malaria detection in blood samples of German travelers - a cross-sectional study. Travel Med Infect Dis. 2018;24:25–30. [DOI] [PubMed] [Google Scholar]
  • 47.Frickmann H, Wegner C, Ruben S, Behrens C, Kollenda H, Hinz R, et al. Evaluation of the multiplex real-time PCR assays RealStar malaria S&T PCR kit 1.0 and FTD malaria differentiation for the differentiation of Plasmodium species in clinical samples. Travel Med Infect Dis. 2019;31:101442. [DOI] [PubMed] [Google Scholar]
  • 48.Hahn A, Podbielski A, Meyer T, Zautner AE, Loderstädt U, Schwarz NG, et al. On detection thresholds-a review on diagnostic approaches in the infectious disease laboratory and the interpretation of their results. Acta Trop. 2020;205:105377. [DOI] [PubMed] [Google Scholar]
  • 49.Esperança PM, Blagborough AM, Da DF, Dowell FE, Churcher TS. Detection of Plasmodium berghei infected Anopheles stephensi using near-infrared spectroscopy. Parasit Vectors. 2018;11:377. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Mewara A, Sreenivasan P, Khurana S. Primate malaria of human importance. Trop Parasitol. 2023;13:73–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Irish SR, Kyalo D, Snow RW, Coetzee M. Updated list of Anopheles species (Diptera: Culicidae) by country in the Afrotropical Region and associated islands. Zootaxa. 2020;4747:401–49. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Makanga B, Yangari P, Rahola N, Rougeron V, Elguero E, Boundenga L, et al. Ape malaria transmission and potential for ape-to-human transfers in Africa. Proc Natl Acad Sci USA. 2016;113:5329–34. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Manga L, Mbingue S, Nkou Etoundi M, Ngollo M. Anopheles namibiensis is anthropophilic and widespread in Cameroon. Med Vet Entomol. 1997;11:409. [DOI] [PubMed] [Google Scholar]
  • 54.Boussougou-Sambe ST, Woldearegai TG, Doumba-Ndalembouly AG, Ngossanga B, Mba RB, Edoa JR, et al. Assessment of malaria transmission intensity and insecticide resistance mechanisms in three rural areas of the Moyen Ogooué Province of Gabon. Parasit Vectors. 2022;15:217. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Lalremruata A, Jeyaraj S, Engleitner T, Joanny F, Lang A, Bélard S, et al. Species and genotype diversity of Plasmodium in malaria patients from Gabon analysed by next generation sequencing. Malar J. 2017;16:398. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Data Availability Statement

Data is provided within the manuscript or supplementary information files. Sequence data that support the findings of this study have been deposited in the GenBank Nucleotide Archive with the primary accession code PX307520-307527.


Articles from Malaria Journal are provided here courtesy of BMC

RESOURCES