Table 1.
Operational characteristics of miECCO2R
| Time | Blood flow (mL/min) | Sweep gas flow (L/min) | aPTT (sec) | Post-membrane iCa2+ (mMol/L) | ||||
|---|---|---|---|---|---|---|---|---|
|
miECCO2R (N = 11) |
miECCO2R/CRRT (N = 9) |
miECCO2R (N = 11) |
miECCO2R/CRRT (N = 9) |
miECCO2R (N = 11) |
miECCO2R/CRRT (N = 9) |
miECCO2R (N = 11) |
miECCO2R/CRRT (N = 9) |
|
| 1 h | 400 (400–400) | 175 (100–200) | 6.0 (6.0–6.0) | 2.5 (1.5–3.0) | 71 (57–88) | 42 (34–46) | 1.13 (1.00–1.16) | 0.30 (0.26–0.34) |
| 24 h | 400 (390–400) | 175 (140–200) | 6.0 (6.0–6.0) | 2.5 (1.9–3.0) | 88 (65–116) | 44 (37–48) | 1.12 (1.03–1.16) | 0.28 (0.25–0.31) |
| 48 h | 400 (390–400) | 175 (140–200) | 6.0 (6.0–6.0) | 2.5 (1.9–3.0) | 69 (64–82) | 41 (36–48) | 1.13 (1.02–1.19) | 0.30 (0.28–0.31) |
Data are presented as median (25%–75% interquartile range)
CRRT: continuous renal replacement therapy, miECCO2R: minimally invasive extracorporeal CO2 removal, iCa2+: ionized calcium