Abstract
Optically pumped magnetometers (OPMs) offer a promising advancement in noninvasive neuroimaging via magnetoencephalography (MEG), but establishing their reliability and comparability to existing methods remains an ongoing endeavor. Here, we evaluated OPM recordings by assessing their test-retest reliability and comparing them to functional magnetic resonance imaging (fMRI) and intracranial electroencephalography (iEEG) recordings. Data were collected from three independent participant groups during repeated viewings of a movie segment. In 7 canonical frequency bands (: 0.5–4 Hz, : 4–8 Hz, : 8–12 Hz, : 12–28 Hz, : 28–46 Hz, : 55–70 Hz, and HF: 64–116 Hz), as well as a broadband (BB: 0.5–116 Hz) signal, we quantified the signal consistency (1) within individuals, (2) across subjects, and (3) across modalities. OPM exhibited widespread reliability, particularly in lower frequency bands; spatial patterns resembled those of fMRI and iEEG in visual and auditory regions. Cross-modal analyses revealed robust correspondence between OPM and both fMRI and iEEG, including inverse correlations at low frequencies and positive correlations at higher frequencies in the OPM-fMRI comparison, consistent with known relationships between oscillatory power and BOLD responses. Comparisons of signal-to-noise (SNR) estimates further revealed that in some regions, the SNR of cross-modal alignment exceeded within-modality reliability, suggesting that bridging between modalities can sometimes enhance SNR by attenuating reliably shared noise. Our findings demonstrate that OPM consistently captures stimulus-driven neural dynamics that converge with established modalities.
Keywords: OPM, MEG, fMRI, iEEG, reliability, intersubject correlation, cross-modal correlation
1. Introduction
Advancing our understanding of the neural activity underlying human cognition depends on tools that reliably capture neurophysiological signals with high temporal precision. Optically pumped magnetometers (OPMs) represent a promising innovation in magnetoencephalography (MEG), a noninvasive technique with high temporal resolution (milliseconds) that can (under favorable conditions) measure cortical sources with up to 2–3 mm precision (Brookes et al., 2021; D. Cohen, 1968, 1972; Hämäläinen et al., 1993; Proudfoot et al., 2014). OPMs address many of the practical limitations associated with superconducting quantum interference devices (SQUIDs) used in conventional MEG systems, whose cryogenic requirements impose substantial operational costs and technical constraints (Andersen et al., 2017; Boto et al., 2017; D. Cohen, 1972; Hämäläinen et al., 1993; Hill et al., 2020): Unlike SQUIDs, OPMs operate at room temperature, eliminating the need for cryogenic cooling and the associated costs (Boto et al., 2017). Their lightweight, compact design allows for direct placement on the scalp, which minimizes the sensor-to-brain distance and may consequently enhance signal amplitude and spatial resolution (Hill et al., 2020; Iivanainen et al., 2017; Nugent et al., 2022). Additionally, OPMs can be configured into wearable, adjustable helmets, accommodating naturalistic movement during data acquisition, and enabling greater adaptability to different head sizes and shapes (Boto et al., 2018). These features not only broaden MEG’s accessibility and versatility in research but also have the potential to improve signal quality by ensuring consistent and close sensor-to-scalp proximity across individuals (Boto et al., 2016, 2017, 2018; Brookes et al., 2021; Hill et al., 2020; Nugent et al., 2022; Rea et al., 2022).
OPM–MEG (hereafter referred to as OPM) is consequently positioned to be a major advancement in neuroimaging, yet its utility hinges on validating whether it captures reliable and meaningful neural signals. While it has been shown that OPMs can detect canonical MEG signals such as evoked responses, oscillatory activity, functional localization, and large-scale connectivity (An et al., 2022; Boto et al., 2021; Rier et al., 2023; Safar et al., 2024; Sun et al., 2024; Tierney et al., 2018; Wang et al., 2024), their reliability has not been as thoroughly established as the reliability of SQUID based MEG (e.g., Lankinen et al., 2014; Lew et al., 2021; Liu et al., 2022; Martín-Buro et al., 2016).
Moreover, while OPMs have demonstrated strong concordance with SQUIDs across sensor- and source-level analyses, sometimes achieving comparable or greater signal quality with fewer sensors (Iivanainen et al., 2023; Marhl et al., 2022; Rhodes et al., 2023), comparative evidence with other neuroimaging and electrophysiological modalities remains limited. Because different measurement techniques are sensitive to distinct neural processes and noise sources, direct comparisons are needed to clarify both shared and modality-specific components. For instance, EEG and iEEG primarily detect radially oriented electrical currents in cortical gyri, while MEG is most sensitive to tangentially oriented sources located in the sulci. Furthermore, MEG is particularly sensitive in regions closer to the sensors because magnetic field strength diminishes rapidly as it travels away from the source (Hämäläinen et al., 1993). In contrast, fMRI measures vascular responses associated with neural activity, capturing slow hemodynamic changes rather than direct electrophysiological signals. As a novel technique, it is important to understand which neurophysiological signals OPM captures, and critically, to determine whether it does so reliably across individuals and in alignment with signals measured by other established neuroimaging methods.
Here, we systematically compare neural activity recorded with OPM, fMRI, and iEEG during repeated viewings of a continuous audio-visual movie stimulus. We (i) quantify OPM signal reliability within individuals, (ii) quantify shared OPM signal between individuals, (iii) determine the extent to which OPM captures the same stimulus-driven neural activity as fMRI and iEEG, and (iv) characterize how its signal reliability varies across these comparisons. To answer these questions, we compute test–retest reliability within individual participants, between different subjects (intersubject correlations, ISC; Hasson et al., 2004), and between recording modalities (see Fig. 1 for an overview of our analysis framework). Finally, we compare signal-to-noise ratios (SNR) within OPM and between OPM and fMRI to characterize their variation across levels of comparison.
Fig. 1.
Schematic of correlation analyses. Within-subject correlations (blue) were computed between the first and second viewing for each subject and averaged across the respective cohort; for iEEG, electrodes were treated as a single subject. Between-subject correlations (green) were computed across viewings, averaging correlations from viewing 1 to viewing 2 (solid) and vice versa (dashed), comparing each subject to the rest of their cohort. Between-method correlations (red) were computed similarly, comparing each subject i in one modality to each subject j in another, or to the single-subject iEEG dataset; fMRI to iEEG comparisons were conducted but are not depicted here for simplicity. Within-viewing analyses were also conducted and are reported in the Supplementary Materials. Notes: Movie was presented in color. Diagrams created in https://BioRender.com/epg2wkb. Image used in this schematic: Warner Bros (1974). [Al Pacino and Penelope Allen in Dog Day Afternoon] [Photograph]. Wikimedia Commons.
2. Methods
Neural activity was recorded from three independent subject groups as they watched the same segment of the feature film Dog Day Afternoon (DDA; Lumet, 1975) twice. The datasets used in this study comprised recordings from 10 OPM subjects (Rier et al., 2023), 11 fMRI subjects (Haufe et al., 2018), and five iEEG subjects (Honey et al., 2012). Comprehensive details on study subjects, data acquisition, experimental settings, and preprocessing steps can be found in the corresponding articles; however, a succinct description of each study’s methodology is provided below. Unless otherwise specified, all data were processed using custom MATLAB scripts and the FieldTrip toolbox (Oostenveld et al., 2010).
2.1. OPM
2.1.1. Subjects and experimental settings
OPM data were originally collected at the University of Nottingham as part of a separate study (Rier et al., 2023). The OPM cohort consisted of 10 subjects (4 female, 6 male, all right-handed; age: = 31 years, SD = 8) who provided written informed consent to participate in the study. The study protocol was approved by the University of Nottingham Medical School Research Ethics Committee.
Subjects were presented with a 600 s clip of DDA twice, with approximately 1–2 min between each viewing. The movie clips were presented at a visual angle of ~13° horizontally and ~9° vertically, approximately 80 cm from the subjects’ heads, and audio was delivered via a waveguide connected to speakers outside of the scanning room (Rier et al., 2023).
2.1.2. Signal acquisition
OPM data were recorded using triaxial OPM sensors (QuSpin, Inc., Colorado, USA) mounted in a 3D-printed helmet (Cerca Magnetics Ltd., Nottingham, UK), selected to best fit each participant. Each sensor has a noise floor of ~13 fT/sqrt (Hz), a bandwidth of ~150 Hz, and provides three independent channels. Sensors were distributed evenly across the scalp, enabling approximately whole-head coverage, which resulted in an average of ~168 channels per subject (range: 165–174). Recordings took place in an OPM-optimized magnetically shielded room (Cerca Magnetics Ltd., Nottingham, UK) equipped with background magnetic field control via reference OPMs, biplanar electromagnetic coils, and a motion tracking camera system. Once magnetic field stabilization procedures were complete, data were recorded as participants viewed the movie segment (600 s per run, sampled at 1,200 Hz). Co-registration of sensors to anatomy was performed by aligning 3D digitizations of participants’ heads, collected with and without the helmet on, to MRI-derived surfaces (Rier et al., 2023).
2.1.3. Preprocessing
OPM data were minimally preprocessed. Channels were excluded based on the initial preprocessing performed by Rier et al. (2023), which removed sensors with near-zero variance, resulting in an average of 164 usable channels per subject (range: 161–166). To minimize edge artifacts, MEG data were padded with 1 s before and after the 600 s movie segment, then bandpass filtered (0.5–150 Hz), notch filtered at 50, 100, and 120 Hz, and down-sampled to 256 Hz.
2.1.4. Source reconstruction
Source reconstruction was performed using linear constrained minimum variance (LCMV) beamforming (Veen et al., 1997). Individual T1-weighted MRIs were segmented and used to construct single-shell head models (Nolte, 2003), co-registered with subject-specific sensor locations. An 8 mm-spaced template grid in MNI space was nonlinearly warped to each participant’s anatomy to obtain individualized source positions for corresponding MNI coordinates. Sources were estimated using the data covariance computed separately for each movie segment across the 600 s time window, with 5% regularization applied to the covariance matrix to stabilize the beamformer solution. The dipole orientation was fixed to the direction of maximum amplitude. Separate filters were computed for each movie repetition in each of the 10 participants.
For comparison to iEEG, source reconstruction of the OPM data was repeated using a shared grid defined by the MNI coordinates of 656 iEEG electrodes. These coordinates were nonlinearly warped to each participant’s anatomical space, and source activity was reconstructed directly at these locations using the same preprocessed MEG data, head models, and beamforming procedure described above.
Because source reconstruction parameters can influence the spatial distribution of reconstructed activity, and no single approach is uniquely correct, we performed additional analyses to demonstrate the robustness of our findings. Specifically, we recomputed all main OPM analyses using a minimum-norm-estimation (MNE) solution that did not consider the data covariance structure or include regularization. These results, reported in the Supplementary Materials, show quantitatively consistent reliability estimates, albeit with broader spatial extent, demonstrating that the effects reported here are not driven by beamformer-specific assumptions.
2.1.5. Amplitude of the analytic signal
To extract frequency-specific activity, the source-reconstructed OPM data were bandpass filtered according to the following frequency bands: (0.5–4 Hz), (4–8 Hz), (8–12 Hz), (12–28 Hz), two ranges within the band (: 28–46 Hz, : 55–70 Hz), a high frequency band (HF: 64–116 Hz), and a broadband (BB: 0.5–116 Hz), using a 4th-order Butterworth IIR filter. We then estimated the log-transformed amplitude at each source location by taking the logarithm of the absolute value of the analytic signal obtained via the Hilbert transform, following standard recommendations for oscillatory amplitude analysis (M. X. Cohen, 2014). This computation was performed separately for each subject, movie viewing, and frequency band.
2.1.6. Cross-modal temporal alignment
OPM data were trimmed to match the adjusted 297 s time axis of the fMRI data (see Section 2.2) by removing the first 16.74 s and the final 286.26 s of the original 600 s recording. Time courses of log-transformed amplitude were resampled to 20 Hz to enable subsequent cross-modal analyses and ensure temporal alignment across modalities.
2.2. fMRI
2.2.1. Subjects and experimental settings
fMRI data were originally collected at Princeton University for the purpose of another study (Haufe et al., 2018). The fMRI sample consisted of 11 subjects (6 female, 5 male; age range: 20–35 years) who provided written informed consent to participate in the study, approved by the Princeton University Institutional Review Board. All subjects were fluent in English, with normal or corrected-to-normal vision, and no history of psychiatric or neurological disorders.
Subjects viewed a 325 s clip from DDA twice, with approximately 11 min between repetitions, as part of a larger study. Movie clips were displayed (subtended 20° horizontally and 16° vertically) using the Psychophysics Toolbox in MATLAB, synchronized with MRI data acquisition, and presented with audio delivered via in-ear headphones (Haufe et al., 2018).
2.2.2. Signal acquisition
MRI data were collected using a 3T Skyra scanner (Siemens, Munich, Germany) with a 16-channel head coil. Functional images were acquired with a gradient-echo EPI sequence (3 3 4 mm resolution, 27 interleaved slices, TR = 1.5 s, TE = 30 ms, flip angle = 72°, FOV = 192 192 mm2, GRAPPA iPAT = 2); high-resolution structural scans were obtained in each session using an MPRAGE sequence (0.9375 mm isotropic, TR = 1.9 s, TE = 2.1 ms, flip angle = 9°, FOV = 240 240 mm2) (Haufe et al., 2018).
2.2.3. Spatial registration
Each subject’s data were aligned to MNI space using AFNI’s linear (3dAllineate) and nonlinear (3dQWarp) registration tools (Cox, 1996), resampled to a 4 mm grid using linear interpolation, and spatially smoothed using a 10 mm full-width-at-half-maximum Gaussian kernel, resulting in the retained volumetric data (32,798 voxels; Haufe et al., 2018).
2.2.4. Preprocessing
Functional data were preprocessed by the original authors as follows: slice-time- and motion-corrected (AFNI’s 3dvolreg) data were linearly detrended, and the discrete Fourier transform (and its inverse) were used to apply a high-pass filter at 0.01 Hz. To ensure signal stability and exclude the blank screen period at the conclusion of the recording, the first 15 s and final 13 s were removed, yielding 297 s of data (Haufe et al., 2018).
2.2.5. Cross-modal alignment
To align the fMRI data to the OPM source grid, we interpolated fMRI activity from MNI-space coordinates onto the template source model used in the OPM beamforming analysis. For each grid point inside the brain volume, we identified the nearest fMRI coordinate(s) based on the minimum Euclidean distance and averaged the time series of all voxels assigned to each corresponding grid point. This yielded spatially aligned fMRI data on the same cortical grid used for OPM source projection.
For analyses comparing fMRI and iEEG, we aligned the fMRI data to the MNI-space coordinates of the 656 iEEG electrodes. For each electrode, we identified all fMRI voxels within a 6 mm radius based on Euclidean distance and averaged their time series. This yielded a version of the fMRI dataset spatially down-sampled to the iEEG grid, allowing direct comparison across modalities at matched anatomical locations. Voxels outside this radius were excluded from the average, and voxel counts per electrode were tracked to normalize the resulting time series.
To ensure temporal alignment across imaging modalities, the fMRI time axis was adjusted to account for hemodynamic delay. The signal was shifted by the lag that maximized the cross-correlation between the amplitude envelope of the movie audio and the time series of the 20 locations nearest to an anatomical mask of auditory cortex regions defined by the AAL atlas (Tzourio-Mazoyer et al., 2002). Auditory cortex was selected as the reference region because its BOLD response has been shown to reliably track continuous auditory input during naturalistic movie viewing, providing a robust, stimulus-locked signal for temporal alignment (Hasson et al., 2010; Mukamel et al., 2005). Preprocessed data were then linearly interpolated to a resolution of 20 Hz.
2.3. iEEG
2.3.1. Subjects and experimental settings
iEEG data were collected at New York University for the purpose of a separate study (Honey et al., 2012). The iEEG cohort consisted of five subjects (4 female, 1 male; age range: 20–47 years) from the Comprehensive Epilepsy Center of the New York University School of Medicine experiencing pharmacologically refractory complex partial seizures. All patients provided informed consent both before and after electrode implantation in accordance with National Institutes of Health guidelines administered by the local Institutional Review Board.
Patients viewed the same intact 325 s clip of DDA twice, each followed by a 30 s fixation period, with approximately 12 min between intact repetitions. Stimuli presented using PsychToolbox Extensions for MATLAB were displayed approximately 40–60 cm from the subjects’ eyes, on a bedside laptop.
2.3.2. Signal acquisition
Recordings were obtained from 922 subdural electrodes in implanted arrays (8 8, 4 8, or 1 8 configurations) spaced 10 mm apart and 2.3 mm in exposed diameter; signals were referenced to skull screws and acquired at 30 kHz using a custom system built on the NSpike framework (L.M. Frank and J. MacArthur, Harvard University Instrument Design Laboratory, Cambridge, MA), which incorporates a 0.6 Hz hardware high-pass filter (Honey et al., 2012).
2.3.3. Spatial registration
Electrode positions were determined using a combination of T1 images (collected pre- and post-implantation), intra-operative photos, and a MATLAB tool incorporating known electrode geometry (Yang et al., 2012); SPM’s DARTEL algorithm (Ashburner, 2007) was then used to warp each subject’s anatomy and individual electrode coordinates to MNI space (Honey et al., 2012). Note that, unlike Haufe et al. (2018) and Honey et al. (2012), we did not exclude electrodes based on hemisphere location; thus, all 656 artifact-free, labeled, and MNI-registered electrodes were retained (see Fig. 2). Given the availability of MNI coordinates as precise estimates of contact locations, we analyzed iEEG data in MNI space rather than estimating anatomical sources.
Fig. 2.
Location of iEEG electrodes on cortical surface. iEEG electrode placements are color-coded by subject: Subject 1 (red), Subject 2 (green), Subject 3 (blue), Subject 4 (orange), Subject 5 (purple).
2.3.4. Preprocessing
Data from all five subjects were first aggregated into a single dataset. Signals were high-pass (0.5 Hz), low-pass (150 Hz), and notch filtered at 60 and 120 Hz, following (Honey et al., 2012), using 4th-order Butterworth IIR filters implemented with FieldTrip. The resulting data were then down-sampled to 256 Hz.
2.3.5. Amplitude of the analytic signal
To estimate the time-resolved amplitude of iEEG activity, broadband signals were bandpass filtered into the same frequency bands as the OPM data using 4th-order Butterworth IIR filters, and the log-transformed amplitude was computed as described in Section 2.1.5.
2.3.6. Cross-modal temporal alignment
For the 338 s of iEEG data, we validated the temporal alignment with the stimulus based on the lag that maximized the cross-correlation between the envelope of the movie’s audio-trace and auditory responses in the high-frequency band (64–116 Hz). This signal was selected because it is a well-established signature of temporally precise, stimulus-locked auditory engagement (Nourski et al., 2009; Sinai et al., 2009), and alignment was further confirmed by visual inspection of peaks in the cross-correlogram across frequency bands. To this end, high-frequency activity from the aggregated dataset was averaged across the 20 locations nearest to an anatomical mask of auditory cortex regions defined by the AAL atlas (i.e., we followed the same procedure used to temporally align the fMRI to the movie). Data were then trimmed to the same 297 s as the OPM and fMRI. Time courses of log-transformed amplitude were, again, resampled to 20 Hz.
2.4. Analyses
2.4.1. Overview and methodological considerations
To identify the shared neural responses elicited across repeated movie viewings, we evaluated the consistency of activity within individuals, between individuals, and across recording modalities. All analyses were performed on the data segments temporally aligned to the movie stimulus and trimmed to identical duration across modalities. All correlation analyses were performed within separate frequency bands; for fMRI analyses the BOLD signal was used. For OPM and fMRI, coefficients were computed using Pearson product-moment correlations, whereas for iEEG, Spearman rank correlations were used to account for potential outliers in signal amplitudes due to epileptic activity. Because iEEG electrode locations varied across subjects, between-subject comparisons could not be performed within the iEEG dataset. Primary analyses compared reliability between viewings of the film, however, within-viewing comparisons are reported in the Supplementary Materials.
2.4.2. Within-subject comparisons
For each of the three modalities, subject-level empirical correlations were computed between the time series from the first and second movie viewings.
2.4.3. Between-subject comparisons
Pairwise correlations were computed bidirectionally between one viewing from each subject (first/second) with the alternate viewings (second/first) of the remaining subjects in the same cohort (OPM or fMRI).
2.4.4. Between-method comparisons
2.4.4.1. OPM to fMRI
Cross-modal consistency between OPM and fMRI was assessed by computing Pearson correlations between each movie viewing (first/second) from each OPM subject and the alternate viewing from each fMRI subject (second/first). Fisher -transformed correlations were then averaged across OPM and fMRI subjects.
2.4.4.2. OPM to iEEG
Spearman correlations were computed between each viewing from each OPM subject (first/second viewing) and the alternate viewing from the single-subject iEEG dataset (second/first viewing). Fisher -transformed correlation coefficients were averaged across OPM subjects.
2.4.4.3. fMRI to iEEG
To comprehensively compare all three imaging modalities, we evaluated the consistency between iEEG and fMRI to reproduce the findings of Haufe et al. (2018) with our analysis approach. Spearman correlations were computed bidirectionally between one viewing from each fMRI subject and the alternate viewing from the single-subject iEEG dataset and averaged across fMRI subjects.
2.4.5. Statistical significance and analysis framework
Correlation estimates are sensitive to the autocorrelational structure of neural time series (Schaworonkow et al., 2015; Woolrich et al., 2001), which may be magnified by temporal down-sampling and smoothing (James et al., 2019; Mejia et al., 2015; Pajula & Tohka, 2014; Zarahn et al., 1997). To account for these dependencies, we assessed statistical significance using standardized metrics derived from modality-specific null distributions (Haufe et al., 2018). Unlike correlation coefficients, which quantify the magnitude of association between two signals, -scores reflect how extreme an observed correlation is relative to the expected distribution under the null, given the temporal structure and noise characteristics of the data. As a result, modest correlation values may yield large -scores if the null distribution is narrow, whereas larger correlation coefficients may not translate into large -scores when the null distribution is broader. We further illustrate the influence of temporal resolution on correlation estimates by computing Pearson correlations of OPM and iEEG at the same sampling rate as fMRI. To this end, we first downsampled the OPM and iEEG data to 0.67 Hz (the temporal frequency of the 1.5 s TR in fMRI) and then repeated our within-subject correlation analysis (see Supplementary Table S1). Crucially, because we aimed to quantify signal consistency rather than to maximize correlation coefficients, we resampled all estimates to 20 Hz, a resolution commonly used in the analysis of power spectra in human electrophysiology (e.g., Michelmann et al., 2016; see also: M. X. Cohen, 2014). We then computed surrogate null distributions using the iterative amplitude-adjusted Fourier transform (iAAFT; Schreiber & Schmitz, 1996), and standardized empirical correlations by deriving -scores relative to the null distribution. We further assessed the significance of observed correlations based on the distribution of averages of surrogate correlation. Specifically, we performed the following steps:
Empirical correlations were computed from pairs of real-time series at a given channel (i.e., source position, voxel, or electrode) in a comparison of interest (between viewings, subjects, or modalities). Subject-level empirical correlations were then Fisher -transformed and averaged across subjects, producing a grand-averaged (GA) empirical correlation coefficient for each channel.
We generated 100 surrogate datasets for each time series (movie viewing, subject, frequency band, and modality) and computed surrogate correlations, where one of the empirical time series in a given comparison was replaced with a phase-randomized surrogate time series. The 100 surrogate correlations that correspond to an empirical correlation were also normalized using the Fisher -transform.
-
Null distributions of averages were then bootstrapped by randomly sampling one surrogate correlation per subject and averaging across subjects to form a single set of GA surrogate correlation coefficients analogous to the empirical GA correlations. This averaging was repeated 10,000 times per comparison, randomly resampling surrogates (i.e., selecting 1 out of 100 options per subject for each surrogate GA). This yielded a null distribution of 10,000 GA surrogate correlation coefficients per channel (Stelzer et al., 2013).
3a. For the iEEG within-subject analysis, subject-level bootstrapping could not be performed. Instead, we repeated step 2 with 1,000 surrogate datasets per channel, using the resulting surrogate correlations directly as the null distribution.
-values were calculated for each channel as the proportion of the surrogate GA correlations that exceeded the empirical GA coefficient. The Benjamini-Hochberg correction was applied to the empirically derived -values across channels in each analysis, effectively controlling the false discovery rate (FDR) at a level of = .05.
For visualization and interpretation, the empirical GA correlations at each channel were standardized relative to the channel’s surrogate distributions by computing -scores: , where and denote the and of the surrogate null distribution. In other words, for each channel we compared a single empirical value, corresponding to the average of subject-level correlations, to a null distribution in which each sample reflects a surrogate value computed as the average of randomly selected subject-level surrogate correlations, thereby preserving the relationship between empirical and surrogate estimates. These -scores reflect estimates of SNR and enable comparison of SNR on a common scale across modalities and analyses.
2.4.6. Signal-to-noise (SNR) comparisons
To evaluate the relative SNR of OPM signals within and across subjects, we compared the channel-wise SNR estimates by subtracting the within-subject -scores from the between-subject -scores for each frequency band. Positive values in the resulting maps indicate greater consistency of neural responses across individuals than within, whereas negative values indicate greater consistency within than across individuals. Cluster-based thresholding was used to identify contiguous channels exceeding the 97.5th or falling below the 2.5th percentile of the observed SNR distribution. The sum of -scores for each cluster was used to identify the maximum clusters for both the positive and negative tails. We then assessed the significance of the maximum clusters relative to 10,000 surrogate SNR maps. These were computed as the difference between randomly sampled surrogate-within- and between-subject correlation sets derived from the null distributions computed in previous analyses. Empirical -values were calculated as the proportion of surrogate absolute cluster sums that were bigger than the empirical maximum cluster sums. We controlled the false discovery rate at = .025 per frequency band.
We performed the same analysis to compare between-subject correlations in OPM to between-method correlations in OPM and fMRI. For each frequency band, we computed channel-wise SNR maps by subtracting the OPM’s between-subject -scores from the corresponding -scores in the OPM to fMRI comparison. Because cross-modal comparisons may yield meaningful inverse relationships, we used the absolute value of the between-method -scores to represent the magnitude of the correlation, regardless of its direction. Thus, resulting positive values indicate greater consistency between OPM and fMRI than across OPM subjects alone, whereas negative values indicate greater consistency across OPM subjects than between OPM and fMRI. Cluster-based permutation testing and significance evaluation followed the same procedure as described above. It is important to note that clusters reflect regions with significant differences, not areas where reliability was equally strong across comparisons.
In addition to these primary SNR analyses, we conducted supplemental contrasts examining modality-specific and cross-modal reliability patterns, which are reported in the Supplementary Materials.
3. Results
3.1. Within-subject comparisons
Across all three methods, we observed significant reliability in spatially distinct but functionally meaningful areas. For OPM, reliable signal, indicated by channels surviving correction for multiple comparisons (MC), was observed across all frequency bands and spanned regions involved in visual processing, higher-order associative functions, and motor control (Fig. 3A). Peak -scores across frequency bands ranged from 5.28 (HF) to 16.04 (), with the highest proportion of channels surviving FDR correction in the band (82.0%; = 5.26, = 2.09). Signal reliability was broadly distributed in the lower frequency bands (, , , and ), with maxima localized to the left and right cuneus; in contrast, higher frequency bands (, , and HF) displayed more focal distributions, with peaks in the precentral gyrus, medial frontal gyrus, and middle frontal gyrus, respectively (see Table 1, OPM). Critically, these patterns were robust to alternative source reconstruction parameters (compare Supplementary Table S4, within-subject and Supplementary Fig. S4A). Interestingly, in the 28–116 Hz range, -scores decreased systematically with increasing frequency, but the spatial distributions of reliable effects followed an inverse pattern. For instance, while the HF band had a lower maximum reliability (peak = 5.28) it also demonstrated the greatest proportion of channels surviving MC correction (26.6%) compared to the (peak = 6.01, % MC = 18.4) and bands (peak = 5.85, % MC = 8.7). This pattern is consistent with prior work indicating that activity in higher-frequencies is driven by distinct neural processes, with band activity reflecting narrow band network oscillations and HF activity indicating broader population spiking; as such, oscillations may be locally strong, but spatially sparse, whereas HF activity can appear more spatially distributed but weak relative to noise (Ray & Maunsell, 2011). Finally, the broadband signal exhibited the highest overall reliability, with peak -scores exceeding those of any individual frequency band and a spatial distribution largely overlapping with low-frequency visual regions (Fig. 3A).
Fig. 3.
Within-subject reliability for different imaging modalities and frequency bands. FDR-corrected -scores are shown for: (A) OPM, 3,581 source locations across 10 subjects; (B) fMRI, 3,581 voxels across 11 subjects; (C) iEEG, 656 electrodes pooled across five subjects. For iEEG, -scores were computed relative to a null distribution of 1,000 surrogate correlations, without bootstrapping (see Methods). (D) Schematic of within-subject correlation analysis. Data were bandpass filtered to seven frequency bands: (0.5–4 Hz), (4–8 Hz), (8–12 Hz), (12–28 Hz), (28–46 Hz), (55–70 Hz), and HF (64–116 Hz), as well a broadband (0.5–116 Hz). Color scales reflect -scores surviving correction for MC of within-subject correlations, plotted on sensor-specific locations (iEEG) or MNI coordinates (OPM, fMRI).
Table 1.
Within-subject reliability statistics across modalities and frequency bands.
| Maxima | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Modality | Frequency | % MC | MNI Coord. | AAL Label (Left/Right) | |||||
| OPM | 0.5–4 Hz | 5.00 | 2.19 | 81.2% | 0.08 | 13.42 | 8, –96, 16 | Cuneus (R) | |
| 4–8 Hz | 4.86 | 2.54 | 77.0% | 0.08 | 16.04 | 8, –96, 16 | Cuneus (R) | ||
| 8–12 Hz | 5.26 | 2.09 | 82.0% | 0.08 | 13.57 | 0, –88, 32 | Cuneus (L) | ||
| 12–28 Hz | 4.45 | 1.89 | 79.3% | 0.04 | 12.28 | –8, –96, 24 | Cuneus (L) | ||
| 28–46 Hz | 3.15 | 0.67 | 18.4% | 0.02 | 6.01 | –40, –16, 64 | Precentral (L) | ||
| 55–70 Hz | 3.31 | 0.64 | 8.7% | 0.02 | 5.85 | 0, 56, 0 | Frontal Sup Medial (L) | ||
| HF | 64–116 Hz | 2.96 | 0.59 | 26.6% | 0.02 | 5.28 | 40, 56, 8 | Frontal Mid (R) | |
| BB | 0.5–116 Hz | 7.56 | 3.34 | 94.2% | 0.07 | 20.76 | 0, –96, 24 | Cuneus (L) | |
| fMRI | – | – | 4.41 | 2.44 | 55.5% | 0.35 | 13.75 | –56, –16, 0 | Temporal Mid (L) |
| iEEG | 0.5–4 Hz | 3.63 | 1.59 | 30.9% | 0.29 | 12.35 | –58, –14, 48 | Postcentral (L) | |
| 4–8 Hz | 3.57 | 1.48 | 34.8% | 0.22 | 9.79 | 65, –18, 25 | SupraMarginal (R)a | ||
| 8–12 Hz | 3.80 | 1.63 | 23.3% | 0.25 | 10.75 | –68, –11, 8 | Temporal Sup (L) | ||
| 12–28 Hz | 3.97 | 1.42 | 17.4% | 0.11 | 9.74 | –67, –21, 16 | SupraMarginal (L) | ||
| 28–46 Hz | 3.54 | 1.13 | 11.0% | 0.09 | 7.80 | –48, –54, –24 | Temporal Inf (L) | ||
| 55–70 Hz | 4.08 | 1.44 | 11.7% | 0.12 | 8.30 | 65, –20, 15 | Temporal Sup (R) | ||
| HF | 64–116 Hz | 5.69 | 3.82 | 26.7% | 0.34 | 21.26 | –67, –21, 16 | SupraMarginal (L) | |
| BB | 0.5–116 Hz | 3.83 | 1.86 | 40.9% | 0.25 | 12.30 | –68, –11, 8 | Temporal Sup (L) | |
Note. All statistics are reported for data surviving correction for multiple comparisons (MC). For each modality and frequency band, we report the mean and standard deviation of -scores (, ), the percentage of FDR-surviving virtual channels/voxels/electrodes (% MC), the maximum absolute correlation coefficient (), the peak -score, its MNI coordinate, and the corresponding anatomical label. Superscript a denotes that the first label was found for the electrode with the second highest -score.
For fMRI, reliable signal was more spatially localized, with 55.5% of voxels surviving MC correction (Table 1, fMRI). Significant effects were observed primarily in regions associated with auditory and visual processing (see Fig. 3B), with a peak -score of 13.75 observed in the left middle temporal gyrus ( = 4.41, = 2.44; see also Table 1, fMRI).
iEEG showed strong within-subject reliability, though spatial extent was more limited, with 11.0–34.8% (see Table 1, iEEG) of electrodes surviving correction for MC across frequency bands. In the lower frequency bands, reliability was strongest in temporal, parietal, and sensorimotor areas (see Fig. 3C). Similar to OPM, the broadband iEEG signal yielded the highest proportion of electrodes surviving MC correction (40.9%; Table 1, iEEG) relative to the frequency-decomposed signal; however, the highest FDR corrected -score was observed for the high-frequency band concentrated in the left supramarginal gyrus ( = 21.26, = 5.69, = 3.82).
Although absolute correlation values were highest for fMRI ( = .35) and iEEG ( = .34) compared to OPM ( = .08, compare Supplementary Table S1), we found that both temporally down-sampling the data and “grand averaging” time series across subjects substantially increased OPM correlations (see Supplementary Table S1, OPM). When re-computed at 0.67 Hz, matching the temporal resolution of fMRI, OPM peak correlations more than tripled ( = .30), and further increased after grand averaging measurements across subjects before computing correlations ( = .66). These findings illustrate that high test-retest correlations can be achieved with OPM by down-sampling and averaging across subjects. Importantly, estimating the SNR via -scores (see above) confirmed robust statistical reliability.
3.2. Between-subject comparisons
OPM exhibited reliable signal across all frequency bands, with effects most pronounced in occipitotemporal and cerebellar regions implicated in perceptual and linguistic processing (see Fig. 4A; MacEvoy & Epstein, 2011; Nakatani et al., 2022). Across frequency bands, peak -scores ranged from 5.53 (HF) to 15.29 (), with the greatest proportion of channels surviving correction for MC in the band (92.8%; = 6.09, = 2.38; see Table 2, OPM). In lower frequencies, regions of strongest reliability were concentrated in the temporal pole (), lingual gyrus (), precuneus (), and superior occipital gyrus (), suggesting pronounced reliability in posterior visual and associative regions. High-frequency bands exhibited more focal effects, with peak values in the cerebellum (), precuneus (), and inferior frontal gyrus (HF). Widespread effects were further observed in the broadband (raw) signal, with peak reliability localized to the calcarine fissure (peak -score = 22.13), indicating stable responses across subjects in early visual regions. Similar patterns were observed with alternative source reconstruction parameters (see Supplementary Table S4, between-subject and Supplementary Fig. S4B).
Fig. 4.
Between-subject reliability for OPM and fMRI. FDR-corrected -scores are shown for: (A) OPM, 3,581 source locations across 10 subjects for 8 frequency bands; (B) fMRI, 3,581 voxels, across 11 subjects. (C) Schematic of between-subject correlation analysis. Color bars reflect -scores (surviving correction for MC).
Table 2.
Between-subject reliability statistics for OPM and fMRI.
| Maxima | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Modality | Frequency Band | % MC | MNI Coord. | AAL Label (Left/Right) | |||||
| OPM | 0.5–4 Hz | 4.64 | 1.98 | 87.2% | 0.04 | 15.18 | –32, 24, –32 | Temporal Pole Sup (L) | |
| 4–8 Hz | 5.36 | 2.78 | 81.6% | 0.03 | 15.29 | 16, –88, –8 | Lingual (R) | ||
| 8–12 Hz | 6.11 | 2.70 | 78.9% | 0.03 | 13.59 | 8, –72, 56 | Precuneus (R) | ||
| 12–28 Hz | 6.09 | 2.38 | 92.8% | 0.02 | 14.16 | 16, –96, 24 | Occipital Sup (R) | ||
| 28–46 Hz | 3.29 | 0.63 | 12.4% | 0.01 | 5.55 | 16, –64, –32 | Cerebellum VI (R) | ||
| 55–70 Hz | 3.47 | 0.61 | 5.5% | 0.01 | 5.67 | 16, –48, 16 | Precuneus (R) | ||
| HF | 64–116 Hz | 2.99 | 0.64 | 26.2% | 0.01 | 5.53 | –8, –16, –48 | Frontal Inf Tri (R)a | |
| BB | 0.5–116 Hz | 9.07 | 3.70 | 99.3% | 0.03 | 22.13 | 24, –96, 0 | Calcarine (R) | |
| fMRI | – | – | 4.51 | 1.97 | 24.5% | 0.15 | 11.84 | –48, –24, 8 | Temporal Sup (L) |
Note. All statistics are reported for data surviving MC correction. For each modality and frequency band, we report the mean and standard deviation of -scores (, ), the percentage of FDR-surviving voxels/channels/electrodes (% MC), the maximum absolute correlation coefficient (), and the peak -score with its MNI coordinate and anatomical label.
Interestingly, although reliability estimates were generally lower in the between-subject relative to the within-subject comparison for OPM, certain frequency bands exhibited the opposite pattern. Greater between-subject reliability can emerge because individual-specific noise contributes to within-subject test-retest correlations. Idiosyncratic noise in between subject correlations may decrease to a larger extent than shared signal, effectively improving the SNR. For instance, regions reflecting a mixture of stimulus-driven and subject-specific processes may fail to reach significance within subjects yet emerge as significant between subjects once individual variance is attenuated. Most notably, we observed greater between-subject reliability in the and bands, where peaks shifted from the cuneus to superior temporal and occipital areas, respectively. This is consistent with prior work demonstrating that ISCs during naturalistic movie viewing extend well beyond primary sensory regions, particularly to the superior temporal sulcus (Hasson et al., 2004). Moreover, these results align with previous findings of increased cross-subject synchronization in the band during emotionally salient scenes (Maffei, 2020).
For fMRI, although fewer voxels survived correction for MC (24.5%; Table 2, fMRI) compared to the within-subject analysis, the spatial pattern was largely preserved, with effects localized to auditory and visual regions ( = 4.51, = 1.97; see Fig. 4B). The maximum was observed in the left superior temporal gyrus (peak = 11.84), consistent with its role in processing complex audio-visual stimuli (Ghinst et al., 2016; Park et al., 2018; Ye et al., 2017). Within-viewing analyses converged on these results, yielding highly similar spatial and statistical patterns (compare Supplementary Table S2 and Supplementary Fig. S1).
3.3. Between-method comparisons
3.3.1. OPM to fMRI
Consistent cross-modal alignment between OPM and fMRI was observed across all frequencies, with the most pronounced effects localized to visual processing regions in the mid-range and high-frequency bands (see Fig. 5A). After correction for MC, peak -scores ranged from -12.42 () to 5.19 (HF), with the largest proportion of surviving channels observed in the band (17.8%; = -4.37, = 3.05; Table 3, OPM to fMRI). Importantly, we observed negative cross-modal correlations in the , , , and bands, and positive correlations in the , , and high-frequency bands, in line with previous demonstrations of the relationship between MEG power spectra and BOLD signal (Zumer et al., 2010). Broadband reliability between fMRI and OPM closely resembled that of the lower frequencies, with peak effects observed in the superior occipital gyrus (peak -score = -12.73; Table 3, OPM to fMRI).
Fig. 5.
Consistent between-method alignment. (A) OPM to fMRI: -scores from OPM subjects (n = 10) and fMRI subjects (n = 11) are shown for 3,581 locations. (B) OPM to iEEG: -scores from OPM subjects (n = 10) and single-subject iEEG from 656 electrodes pooled across five subjects. (C) fMRI to iEEG: -scores from fMRI subjects (n = 11) and single-subject iEEG from 656 electrodes pooled across five subjects. Color bars reflect -scores (see above). Within-viewing comparisons are shown in Supplementary Figure S2.
Table 3.
Between-method reliability statistics across frequency bands for OPM, fMRI, and iEEG.
| Maxima | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Comparison | Frequency Band | % MC | MNI Coord. | AAL Label (Left/Right) | |||||
| OPM to fMRI | 0.5–4 Hz | –2.30 | 3.62 | 5.5% | –0.02 | –8.05 | –24, –56, 56 | Parietal Sup (L) | |
| 4–8 Hz | –3.19 | 3.32 | 9.5% | –0.02 | –9.51 | –24, –56, 72 | Parietal Sup (L) | ||
| 8–12 Hz | –4.37 | 3.05 | 17.8% | –0.03 | –12.42 | –40, –88, 8 | Occipital Mid (L) | ||
| 12–28 Hz | –4.02 | 3.14 | 16.1% | –0.02 | –12.00 | –40, –80, 16 | Occipital Mid (L) | ||
| 28–46 Hz | 3.74 | 0.00 | 0.0% | 0.01 | 3.74 | 24, –8, 64 | Frontal Sup (R) | ||
| 55–70 Hz | 3.45 | 2.13 | 0.7% | 0.01 | 4.72 | 16, –80, 8 | Calcarine (R) | ||
| HF | 64–116 Hz | 3.74 | 1.07 | 2.0% | –0.01 | 5.19 | 0, –88, 16 | Cuneus (L) | |
| BB | 0.5–116 Hz | –4.01 | 3.48 | 15.1% | –0.02 | –12.73 | –16, –88, 32 | Occipital Sup (L) | |
| OPM to iEEG | 0.5–4 Hz | 3.32 | 1.15 | 6.6% | 0.03 | 4.74 | –37, –45, 69 | Parietal Sup (L) | |
| 4–8 Hz | 3.68 | 1.23 | 9.9% | 0.04 | 7.10 | –21, –55, 78 | Parietal Sup (L) | ||
| 8–12 Hz | 3.57 | 1.89 | 16.3% | 0.05 | 8.11 | –72, –38, 9 | Temporal Mid (L) | ||
| 12–28 Hz | 3.55 | 1.09 | 7.9% | 0.02 | 5.68 | –54, –75, 3 | Occipital Mid (L) | ||
| 28–46 Hz | – | – | – | – | – | – | – | ||
| 55–70 Hz | – | – | – | – | – | – | – | ||
| HF | 64–116 Hz | – | – | – | – | – | – | – | |
| BB | 0.5–116 Hz | 3.71 | 1.62 | 24.2% | 0.03 | 8.16 | –29, –49, 73 | Parietal Sup (L) | |
| fMRI to iEEG | 0.5–4 Hz | –4.68 | 1.27 | 3.8% | –0.09 | –7.40 | –11, –64, 75 | Precuneus (L) | |
| 4–8 Hz | –3.76 | 2.12 | 7.6% | –0.08 | –7.09 | –74, –28, 3 | Temporal Mid (L) | ||
| 8–12 Hz | –4.37 | 1.20 | 6.1% | –0.08 | –7.96 | –46, –88, 17 | Occipital Mid (L) | ||
| 12–28 Hz | –4.11 | 1.13 | 4.1% | –0.05 | –7.48 | –35, –91, 22 | Occipital Mid (L) | ||
| 28–46 Hz | 3.77 | 1.72 | 4.0% | 0.03 | 5.96 | –72, –38, 9 | Temporal Mid (L) | ||
| 55–70 Hz | 4.53 | 1.26 | 3.5% | 0.05 | 7.54 | –21, –55, 78 | Parietal Sup (L) | ||
| HF | 64–116 Hz | 4.97 | 1.44 | 5.9% | 0.07 | 7.71 | –46, –83, 7 | Occipital Mid (L) | |
| BB | 0.5–116 Hz | –4.40 | 1.30 | 8.5% | –0.08 | –7.57 | –46, –88, 17 | Occipital Mid (L) | |
Note. All statistics are reported for data surviving correction for MC. For each comparison and frequency band, we report the mean and standard deviation of -scores (, ), the percentage of virtual channels/voxels/electrodes surviving correction (% MC), the maximum absolute correlation coefficient (), and the peak -score with its MNI coordinate and anatomical label. Statistics for within-viewing analyses are reported in Supplementary Table S3.
Of the frequency bands that demonstrated broader topographies (, , and ), peak -scores were located in the left superior parietal gyrus ( and ) and the left middle occipital gyrus ( and ), with a notable focal peak observed for the HF band in left cuneus. These spatial patterns are consistent with prior findings demonstrating that MEG and fMRI exhibit the greatest intersubject consistency in occipital areas during naturalistic movie viewing (Lankinen et al., 2018). Moreover, MNE yielded visual peaks in the 55-116 Hz range (Supplementary Fig. S5A), which were also observed under the more spatially constrained LCMV solution (Supplementary Table S5, OPM-fMRI).
Though OPM-fMRI reliability patterns largely converged on regions reliable within each modality, supplemental analyses identified effects exclusive to each comparison (see Supplementary Tables S6 and S8). Interestingly, when contrasting within-subject OPM and within-subject fMRI, OPM-specific effects were observed in the cerebellum, providing convergent support for previous work demonstrating that OPM-MEG can capture cerebellar activity (Lin et al., 2019). This finding is notable given the prior claims that the architecture of the cerebellum restricts its contributions to SQUID-MEG signal, (Freeman et al., 2009) relative to fMRI; however, cerebellar effects in OPM may also reflect spatial leakage from visual regions due to limitations of the source reconstruction. Compared to SQUIDs, improved sensor-to-scalp proximity may afford OPMs greater sensitivity to cerebellar sources. Likewise, structures such as the basal ganglia have been shown to contribute weakly to SQUID-MEG signals due to their radial, non-columnar organization (Schomer & da Silva, 2018). Yet, cross-modal-specific peaks emerged in regions of the basal ganglia for the OPM-fMRI comparison (Supplementary Fig. S6A). Additionally, prefrontal regions, which have previously exhibited weak intersubject coherence during naturalistic viewing (Hasson et al., 2004), demonstrated fMRI-specific reliability within subjects, but emerged as even more reliable in the OPM-fMRI comparison, with higher mean and maximum -scores, as well as a larger proportion of channels surviving correction for MC (see Supplementary Table S6). These patterns were largely consistent in the between-subject versus cross-modal contrasts, with one notable difference in the band, where OPM-specific peak reliability was observed in the temporal pole that was absent in between-subject fMRI (see Supplementary Fig. S7). This finding may reflect continuous language comprehension during the movie that is uniquely captured with the greater temporal resolution of OPM (Mesulam, 2022). Though more spatially focal, these effects were also present in the band of the OPM-fMRI comparison. Collectively, these results suggest that while each modality has distinct measurement strengths, cross-modal approaches can enhance sensitivity to shared activity that may be difficult to capture with either modality independently.
3.3.2. OPM to iEEG
Cross-modal consistency between OPM and iEEG was more modest, with significant effects observed in lower frequency bands, spanning frontal, parietal, and ventral visual regions (see Fig. 5B). As reported in Table 3, OPM to iEEG, peak -scores ranged from 4.74 () to 8.11 (), with the greatest proportion of correspondence observed in the band (% MC = 16.3, = 3.57, = 1.89).
Spatially, peak correspondence in the lower bands was localized to the superior parietal gyrus ( and ), middle temporal gyrus (), and middle occipital gyrus (). Although no effects were found in the , , and HF frequency bands, significant cross-modal consistency between OPM and iEEG was further observed in the broadband signal, with 24.2% of electrodes surviving correction for MC, and a consistent peak in the superior parietal gyrus (see Fig. 5B).
When contrasting within-subject OPM and within-subject iEEG (see Supplementary Table S7), OPM-specific reliability was broadly observed primarily in visual regions, whereas no regions were uniquely reliable in iEEG. The between-method OPM-iEEG comparison nevertheless revealed sparse, primarily frontal peaks (Supplementary Fig. S6B), mirroring the OPM-fMRI results and further supporting the idea that cross-modal analyses can reveal overlapping dynamics that are less apparent within individual modalities.
3.3.3. fMRI to iEEG
To provide a comprehensive evaluation of the consistency between all three imaging modalities, we reproduced the results of Haufe et al. (2018) comparing iEEG to fMRI.
We observed significant cross-modal correlations between fMRI and iEEG across all frequency bands, with effects spanning temporal, parietal, and occipital regions (Fig. 5C), consistent with Haufe et al. (2018). As in our fMRI to OPM comparison, negative correlations were observed in lower-frequency bands (-), while () and high-frequency bands showed positive associations with the BOLD signal (see Table 3, fMRI to iEEG). Peak -scores ranged from -7.96 () to 7.71 (HF), with the largest proportion of channels surviving correction for MC in the band (7.6%, = -3.76, = 2.12).
Peak effects were largely located in areas associated with auditory and linguistic processing (Bhaya-Grossman & Chang, 2022; Wernicke, 1974), including the middle temporal ( and ) and occipital (, , HF, and BB) gyri. This pattern is consistent with prior work demonstrating strong BOLD coupling to oscillatory activity in the auditory cortex during movie viewing (Mukamel et al., 2005).
3.4. Signal-to-noise (SNR) comparisons
3.4.1. Comparison of within-subject SNR to between-subject SNR in OPM
We observed clusters of significantly reduced between-subject SNR relative to within-subject SNR for OPM across all frequency bands ( , controlling the false discovery rate at , as per cluster-based permutation testing; see Methods). Maximum negative clusters were spatially distributed across sensorimotor, parietal, and frontal regions, reflecting areas with greater variability between subjects (Fig. 6). The most pronounced SNR loss was observed in the band, which yielded the largest significant negative cluster (85 channels, sum = -557.77), corresponding in the observed data to a region centered on the right postcentral gyrus (peak = -8.84, = -6.56, = 1.00). As shown in the top panel of Table 4, significant decreases in SNR were also found across the (sum = -313.80, max in inferior temporal gyrus), (sum = -72.94, max in cuneus), (sum = -87.73, max in postcentral gyrus), (sum = -331.11, max in superior temporal gyrus), (sum = -283.24, max in gyrus rectus), HF (sum = -137.95, max in middle temporal gyrus), and broadbands (sum = -263.46, max in precentral gyrus).
Fig. 6.
SNR differences in within- and between-subject comparisons for OPM. Within-modality SNR differences in OPM were computed by subtracting within-subject from between-subject -scores at each channel. Negative values (blue) indicate greater signal consistency within individuals than across individuals; positive values (green) indicate greater signal consistency across individuals than within individuals. Significant clusters surviving MC correction (FDR ) are outlined in the respective color. SNR changes are shown across frequency bands from (0.5–4 Hz) to high-frequency (64–116 Hz), as well as the broadband (0.5–116 Hz). Diagrams below illustrate the comparison logic. Maps of uncorrected clusters identified by the initial cluster-forming threshold are shown in Supplementary Figure S3A.
Table 4.
Cluster statistics for SNR differences in within- and between-subject OPM comparisons.
| Cluster Maxima | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Cluster Type | Frequency Band | Size | Sum | MNI Coord. | AAL Label (L/R) | ||||
| Negative | 0.5–4 Hz | 54 | 313.80 | 5.81 | 0.80 | 8.32 | 56, 24, 16 | Temporal Inf (R) | |
| 4–8 Hz | 17 | 72.94 | 4.29 | 0.59 | 5.28 | 0, 88, 24 | Cuneus (L) | ||
| 8–12 Hz | 85 | 557.77 | 6.56 | 1.00 | 8.84 | 24, 32, 80 | Postcentral (R) | ||
| 12–28 Hz | 24 | 87.73 | 3.66 | 1.00 | 5.82 | 48, 24, 56 | Postcentral (R) | ||
| 28–46 Hz | 60 | 331.11 | 5.52 | 0.99 | 8.84 | 48, 24, 8 | Temporal Sup (L) | ||
| 55–70 Hz | 50 | 283.24 | 5.66 | 0.84 | 8.31 | 0, 48, 16 | Rectus (L) | ||
| HF | 64–116 Hz | 29 | 137.95 | 4.76 | 0.72 | 6.79 | 64, 40, 8 | Temporal Mid (L) | |
| BB | 0.5–116 Hz | 59 | 263.46 | 4.47 | 1.31 | 7.51 | 40, 24, 64 | Precentral (R) | |
| Positive | 0.5–4 Hz | 34 | 171.90 | 5.06 | 0.45 | 5.94 | 64, 8, 0 | Temporal Sup (L) | |
| 4–8 Hz | 17 | 84.16 | 4.95 | 0.58 | 5.96 | 40, 8, 16 | Amygdala (R)a | ||
| 8–12 Hz | 25 | 140.77 | 5.63 | 0.68 | 7.31 | 56, 64, 8 | Temporal Mid (L) | ||
| 12–28 Hz | 23 | 151.02 | 6.57 | 0.71 | 8.94 | 56, 48, 40 | Parietal Inf (L) | ||
| 28–46 Hz | 28 | 130.74 | 4.67 | 0.77 | 7.38 | 8, 8, 32 | Cingulum Mid (L)a | ||
| 55–70 Hz | 21 | 96.44 | 4.59 | 0.66 | 6.11 | 16, 8, 16 | Caudate (R)a | ||
| HF | 64–116 Hz | 17 | 75.89 | 4.46 | 0.48 | 5.30 | 16, 40, 24 | Cerebellum IV/V (L)a | |
| BB | 0.5–116 Hz | 46 | 309.45 | 6.73 | 0.62 | 7.85 | 32, 72, 56 | Parietal Sup (L) | |
Note. All statistics are reported for the largest cluster per frequency band surviving MC correction (). Negative clusters (top) reflect higher within- than between-subject reliability. Positive clusters (bottom) reflect higher between- than within-subject reliability. For each cluster, we report its size (number of virtual channels), the sum, , and of -scores, and the peak -score along with its MNI coordinate and anatomical label. Superscript a denotes that the first label was found in the channel with the third highest -score.
Additionally, all frequencies revealed clusters of significant positive SNR, representing areas where neural responses were more consistent across the OPM cohort than within individuals. These clusters were largely left-lateralized and predominately localized posterior association and limbic areas (see Fig. 6), consistent with previous findings of intersubject synchrony during emotionally engaging movie viewing (Jääskeläinen et al., 2008; Maffei, 2020). Posterior peaks spanned the superior temporal (, sum = 171.90), middle temporal (, sum = 140.77), inferior parietal (, sum = 151.02), and superior parietal gyri (BB, sum = 309.45) (see Table 4, bottom panel). Additional positive clusters were found in subcortical regions, including (sum = 84.16, max in amygdala) and (sum = 96.44, max in caudate nucleus), as well as in the cerebellum (HF: sum = 75.89), and middle cingulate gyrus (: sum = 130.74). Such higher SNR between different subjects can stem from a decrease in shared noise that exceeds the decrease in shared signal.
3.4.2. SNR comparison of intersubject correlations in OPM and inter-method correlations between OPM and fMRI
Cluster-based permutation testing revealed significant negative clusters surviving MC correction in all frequency bands, indicating greater signal consistency within modality (OPM) than across modalities (OPM to fMRI). These effects were distributed primarily across frontal, parietal, temporal, and occipital regions (see Fig. 7), with the largest cluster appearing in the band (84 channels, sum = -1000.63), corresponding to a region centered on the left lingual gyrus ( = -14.98, = -11.91, = 0.94; Table 5, top panel). Across frequency bands, negative clusters mainly comprised posterior dorsal regions, including the precuneus (, sum = -499.24), superior parietal gyrus (, sum = -568.91), and middle occipital gyrus (BB, sum = -1329.52), highlighting robust reliability within OPM that was not strongly reflected in fMRI.
Fig. 7.
SNR differences within and across OPM and fMRI. Between-modality SNR differences contrast the SNR from between-subject OPM comparisons with the SNR from between-method OPM-to-fMRI comparisons. Negative values (green) indicate greater signal reliability within the OPM sample than in the cross-modal comparison. Positive values (red) indicate greater signal reliability in the cross-modal comparison (fMRI-to-OPM) than within the OPM sample. Significant clusters surviving MC correction (FDR ) are outlined in the respective color. SNR changes are shown across frequency bands from (0.5- 4 Hz) to high-frequency (64–116 Hz), as well as the broadband (0.5–116 Hz). Diagrams below illustrate the comparison logic for each panel. Maps of uncorrected clusters identified by the initial cluster-forming threshold are shown in Supplementary Figure S3B.
Table 5.
Cluster statistics for SNR differences in between-subject OPM and between-method OPM to fMRI comparisons.
| Cluster Maxima | |||||||||
|---|---|---|---|---|---|---|---|---|---|
| Cluster Type | Frequency Band | Size | Sum | MNI Coord. | AAL Label (L/R) | ||||
| Negative | 0.5–4 Hz | 43 | 456.72 | 10.62 | 1.69 | 14.83 | 32, 16, 40 | Temporal Pole Mid (L) | |
| 4–8 Hz | 84 | 1000.63 | 11.91 | 0.94 | 14.98 | 16, 88, 8 | Lingual (R) | ||
| 8–12 Hz | 49 | 499.24 | 10.19 | 0.95 | 12.58 | 8, 72, 40 | Precuneus (R) | ||
| 12–28 Hz | 57 | 568.91 | 9.98 | 0.70 | 11.85 | 32, 72, 56 | Parietal Sup (L) | ||
| 28–46 Hz | 19 | 71.51 | 3.76 | 0.46 | 4.85 | 32, 16, 24 | Frontal Inf Tri (R)a | ||
| 55–70 Hz | 26 | 88.11 | 3.39 | 0.60 | 4.70 | 64, 8, 24 | Precentral (R)a | ||
| HF | 64–116 Hz | 19 | 69.84 | 3.68 | 0.28 | 4.30 | 24, 8, 0 | Putamen (R) | |
| BB | 0.5–116 Hz | 83 | 1329.52 | 16.02 | 1.09 | 20.03 | 24, 96, 8 | Occipital Mid (R) | |
| Positive | 0.5–4 Hz | 37 | 90.08 | 2.43 | 0.72 | 4.67 | 16, 24, 40 | Frontal Sup (L) | |
| 4–8 Hz | 35 | 90.53 | 2.59 | 0.61 | 4.07 | 16, 56, 8 | Frontal Sup Orb (L)a | ||
| 8–12 Hz | 37 | 91.56 | 2.47 | 0.45 | 3.45 | 32, 56, 24 | Frontal Mid (R) | ||
| 12–28 Hz | 24 | 46.09 | 1.92 | 0.86 | 4.57 | 40, 48, 16 | Frontal Mid (R) | ||
| 28–46 Hz | 24 | 101.47 | 4.23 | 0.63 | 5.72 | 64, 24, 0 | Temporal Mid (L) | ||
| 55–70 Hz | 18 | 81.95 | 4.55 | 0.60 | 6.31 | 32, 88, 24 | Occipital Mid (R) | ||
| HF | 64–116 Hz | 17 | 63.04 | 3.71 | 0.62 | 5.39 | 16, 80, 40 | Occipital Sup (L) | |
| BB | 0.5–116 Hz | – | – | – | – | – | – | – | |
Note. All statistics are reported for the largest cluster per frequency band surviving MC correction (). Negative clusters (top) reflect greater reliability between OPM subjects than when compared to fMRI. Positive clusters (bottom) reflect greater cross-modal correspondence between OPM and fMRI than within the OPM group alone. For each cluster, we report its size (number of channels), the sum, , and of -scores, and the peak -score along with its MNI coordinate and anatomical label. Superscript a denotes that the first label was found in the channel with the second highest -score. Superscript b denotes that the first label was found in the channel with the third highest -score.
Surprisingly, significant positive clusters also emerged across all frequency bands, indicating that, in some regions, signal correspondence between OPM and fMRI exceeded within-modality reliability. In the lower frequencies, these clusters were largely concentrated in frontal regions (see Fig. 7), likely indicating areas where activity is integrated over longer timescales, captured by both OPM and fMRI. Such clusters were observed in superior frontal ( and ) and middle frontal gyri ( and ). Moreover, and consistent with prior findings on high-frequency oscillations in visual processing (Martinovic & Busch, 2011; Mukamel et al., 2005), positive clusters were found in the occipital cortex in both the (sum = 81.95) and HF bands (sum = 63.04) (see Table 5, bottom panel), where, despite OPM’s fragmentary spatial pattern, alignment with fMRI was robust. Complementary to these effects, a positive cluster was identified in , localized to the middle temporal gyrus (sum = 101.47), in line with proposed oscillatory dynamics during speech processing (Giraud & Poeppel, 2012). These findings indicate that, although within-modality reliability is generally higher, cross-modal correspondence may exceed it in anatomically and spectrally specific patterns.
4. Discussion
We systematically evaluated the reliability of time-resolved neural signals in 8 frequency bands captured by OPM, and their cross-modal alignment with fMRI and iEEG during repeated viewings of a movie stimulus. OPM demonstrated widespread within-subject reliability, especially at lower frequencies; these results were consistent with SQUID-MEG studies of naturalistic movie viewing (Lankinen et al., 2018), but showed a broader spatial distribution that extended beyond occipital regions. We further observed reliable correlations between subjects in our OPM sample; their SNR exceeded, in some cases, that of within-subject analyses, which suggests that a shared signal consistently exceeded noise across individuals. Moreover, spatial maps became increasingly focused to sensory regions, mirroring patterns reported in previous intersubject analyses (Chang et al., 2015; Hasson et al., 2004; Haufe et al., 2018; Lankinen et al., 2014).
Correlations of OPM signals produced SNR estimates comparable in magnitude to those observed in fMRI and iEEG and with convergent spatial patterns in visual and auditory regions. OPM signals also directly correlated with both modalities, with the largest effects localized to visual areas. Cross-modal comparisons indicated that OPM showed slightly stronger alignment with fMRI than with iEEG in terms of peak -score magnitude; however, mean -scores and the proportion of channels surviving correction for MC were comparable, and sometimes greater between OPM and iEEG. Still, these differences may have been driven by differences in the number of channels in each comparison (i.e., OPM-fMRI: n = 3,581; OPM-iEEG: n = 656). Contrary to previous SQUID–MEG based findings (Lankinen et al., 2018), OPM achieved reliable correlations with fMRI, and showed frequency-dependent shifts in correlation-direction that are consistent with the inverse relationship between low-frequency oscillatory power and BOLD signals (Zumer et al., 2010). Collectively, these results demonstrate that OPM reliably captures stimulus-driven neural dynamics shared across individuals and imaging modalities.
Still, shared activity between two modalities does not imply uniform cross-modal congruence across all frequency bands. For instance, correspondence between all three modalities was comparable in the lower frequency range (LF; 0.5–28 Hz). At higher frequencies (HF; 28–116 Hz), however, only a few channels survived correction for multiple comparisons in the OPM-fMRI comparison, despite reliable iEEG-fMRI correspondence. This pattern may arise from frequency-dependent differences in the spatial and temporal sensitivity between OPM and iEEG. LF activity unfolds over longer timescales and spreads more broadly (Buzsáki & Draguhn, 2004), affording a higher SNR in the recorded signal. This high SNR in LFs likely supports the observed correspondence between OPM and iEEG, despite potential modality-specific sensitivities to sulcal and gyral sources.
In contrast, transient and local HF activity is reliably detected with intracranial recordings (Parvizi & Kastner, 2018). While direct comparisons between iEEG and SQUID-MEG are sparse and have largely been evaluated in clinical contexts, it has been shown that MEG reliably captures LF dynamics observed in iEEG, but nonetheless, has reduced sensitivity to highly localized high-gamma activity and seizure onset zones (Dalal et al., 2009; Kim et al., 2016). This discrepancy may be driven by a combination of differential sensitivity to sulcal versus gyral sources, the limited spatial distribution of HF activity, and the proximity of intracranial electrodes to the neural sources that produce them. Thus, while OPM captures fragments of high-frequency activity, OPM and iEEG might not capture the same neural generators, which can be addressed in future studies. Likewise, the reduced magnitude of HF OPM-fMRI correspondence may be at least partially attributed to the lower within-modality reliability in OPM compared to other frequency bands.
Although reliability within individual modalities was generally convergent between them, supplemental analyses revealed modality-specific effects as well as regions that emerged as reliable only in cross-modal comparisons. Across analyses, OPM exhibited modality-specific sensitivity to regions that are typically difficult to detect with SQUID-MEG, including the cerebellum, consistent with prior work (Lin et al., 2019) and likely reflecting improved sensor proximity and temporal resolution. In contrast, within-subject OPM–iEEG comparisons showed OPM-exclusive reliability primarily in visual regions, with no regions uniquely reliable in iEEG, consistent with the idea that OPM captures large-scale synchronized activity that may not dominate individual intracranial contacts. Several regions, such as prefrontal cortex and deep structures, were weak or inconsistent in within-modality contrasts but emerged more reliably in OPM-fMRI and OPM-iEEG comparisons, suggesting that cross-modal alignment enhances sensitivity to shared activity.
We note, however, that spatial leakage inherent to MEG source reconstruction warrants caution in interpreting the spatial extent of these effects. Consistent with this, although reliability patterns were largely consistent across source reconstruction approaches, supplemental analyses using minimum-norm-estimation yielded less focal spatial distributions. As such, we recommend viewing these reliability patterns as “heatmaps” of where neural sources are commensurable with the measured data and focusing on peak locations as the most plausible generators of the observed signals. More broadly, this underscores that within-modality reliability alone is insufficient to establish that the measured signal represents meaningful neural activity, highlighting the importance of cross-modal comparisons. Our results demonstrate that OPM signal is not only reliable within and across individuals, but critically, that it captures neural activity measured with independent imaging modalities.
One advantage of our analysis approach is that it directly quantifies the reliability of shared signals relative to shared noise. In our analyses, Pearson correlation coefficients (-values) increased substantially when down-sampling to 0.67 Hz. The computed -scores, on the other hand, account for temporal autocorrelation and noise structure, offering an appropriate metric of signal reliability relative to surrogate data (Haufe et al., 2018). Furthermore, by conducting pairwise analyses at a common sampling rate of 20 Hz, we preserved intersubject variance and temporal precision and were able to demonstrate the reliability of OPM both within and between individuals. Significant effects emerged despite minimal preprocessing and early-generation sensor hardware, which underscores the efficacy of OPM that can likely be improved with future advances in sensor design and denoising techniques (Rier et al., 2023).
Importantly, estimating SNR via -scores places all modalities on a shared statistical scale, allowing for comparisons across datasets with differing temporal resolution, spatial coverage, and sample sizes. Leveraging this approach, we observed that within-subject analyses yielded more widespread reliability than between-subject comparisons. Indeed, between-subject comparisons appeared spatially focused in primary sensory regions, consistent with neural responses commonly evoked during movie viewing (Hasson et al., 2004; Lankinen et al., 2014). That is, while it is often considered a limitation of a design when data are acquired from separate participant groups, because of the introduction of subject-specific variability into comparisons, our findings suggest that intersubject analyses can also mitigate idiosyncratic biases by attenuating shared noise (e.g., spatial leakage in source reconstruction) and task-unrelated fluctuations. In our data, between-subject comparisons effectively isolated stimulus-driven responses generalizable across individuals (Hasson et al., 2004, 2010; Nastase et al., 2019). We acknowledge that residual misalignment arising from spatial normalization may reduce anatomical correspondence, and that cross-modal reliability with iEEG may be further affected by partial volume effects in fMRI at cortical boundaries (Ballester et al., 2002) or depth bias in OPM source reconstruction. However, because such effects are expected to attenuate rather than inflate correlation estimates, the reported results should be interpreted as a lower bound on shared signal. Our SNR comparisons further suggest that this principle can extend across imaging modalities. Speculatively, within-subject designs with cross-modal comparisons may also overestimate reliability by capturing stable but characteristic individual signal components such as subject-specific lateralization.
Finally, we observed that SNR can—in some regions—increase when comparisons are drawn between different subjects undergoing different imaging modalities. While differences in modality sensitivity often reduce overall alignment, they may also support the suppression of shared noise and can consequently sometimes lead to increased SNR. In particular, while high-frequency OPM activity appeared spatially diffuse, its correlation with fMRI was tightly localized to regions expected to support visual processing. That is, while high-frequency measurements in OPM may not be stable enough to be reliably localized across subjects or measurements within OPM, its similarity to fMRI signal allowed for the demonstration that it is reliably present. Notably, such preserved shared signal can only be detected in regions where the modalities functionally align. In the future, cross-modal convergence may be leveraged to guide denoising and source reconstruction, effectively integrating the spatial precision of fMRI with the temporal resolution of MEG.
In conclusion, our findings demonstrate that during audio-visual processing, OPM achieves reliability levels comparable to fMRI and iEEG; these effects were most robust in low-frequency bands. OPM showed strong correspondence with BOLD signals, following the established patterns of inverse relationships between high- and low-frequency neural activity. These results underscore the potential of OPM as a flexible, noninvasive tool capable of resolving complex brain activity across multiple temporal and spatial scales.
Supplementary Material
Acknowledgments
We would like to thank Uri Hasson and Christopher J. Honey, who generously shared the data that made this work possible, as well as Stefan Haufe, who made the preprocessed fMRI data from Arcaro et al. (2015) available and pioneered the analysis approach. We also thank Lukas Rier and coauthors, notably Elena Boto and Matthew J. Brookes, for making the OPM data publicly available and for valuable comments on early conceptualization and analyses of this study.
Data and Code Availability
Data and analysis code underlying the conclusions of the manuscript are available at the Open Science Framework (OSF): https://osf.io/urjvb. The raw OPM data from Rier et al. (2023) are available on Zenodo: https://doi.org/10.5281/zenodo.7525341
Author Contributions
Olivia R. Christiano: Conceptualization; Data curation; Formal analysis; Project administration; Software; Visualization; and Writing—original draft. Sebastian Michelmann: Conceptualization; Data curation; Formal analysis; Methodology; Resources; Software; Supervision; and Writing—review & editing.
Declaration of Competing Interest
The authors declare that there are no competing interests.
Supplementary Materials
Supplementary material for this article is available here: https://doi.org/10.1162/IMAG.a.1218#supplementary-data.
References
- An, K. M., Shim, J. H., Kwon, H., Lee, Y. H., Yu, K. K., Kwon, M., Chun, W. Y., Hirosawa, T., Hasegawa, C., Iwasaki, S., Kikuchi, M., & Kim, K. (2022). Detection of the 40 Hz auditory steady-state response with optically pumped magnetometers. Scientific Reports, 12(1), 17993. 10.1038/s41598-022-21870-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Andersen, L. M., Oostenveld, R., Pfeiffer, C., Ruffieux, S., Jousmäki, V., Hämäläinen, M., Schneiderman, J. F., & Lundqvist, D. (2017). Similarities and differences between on-scalp and conventional in-helmet magnetoencephalography recordings. PLoS One, 12(7), e0178602. 10.1371/journal.pone.0178602 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Arcaro, M. J., Honey, C. J., Mruczek, R. E., Kastner, S., & Hasson, U. (2015). Widespread correlation patterns of fMRI signal across visual cortex reflect eccentricity organization. eLife, 4, e03952. 10.7554/eLife.03952 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ashburner, J. (2007). A fast diffeomorphic image registration algorithm. NeuroImage, 38(1), 95–113. 10.1016/j.neuroimage.2007.07.007 [DOI] [PubMed] [Google Scholar]
- Ballester, M. Á. G., Zisserman, A. P., & Brady, M. (2002). Estimation of the partial volume effect in MRI. Medical Image Analysis, 6(4), 389–405. 10.1016/s1361-8415(02)00061-0 [DOI] [PubMed] [Google Scholar]
- Bhaya-Grossman, I., & Chang, E. F. (2022). Speech computations of the human superior temporal gyrus. Annual Review of Psychology, 73(1), 79–102. 10.1146/annurev-psych-022321-035256 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boto, E., Bowtell, R., Krüger, P., Fromhold, T. M., Morris, P. G., Meyer, S. S., Barnes, G. R., & Brookes, M. J. (2016). On the potential of a new generation of magnetometers for MEG: A beamformer simulation study. PLoS One, 11(8), e0157655. 10.1371/journal.pone.0157655 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boto, E., Hill, R. M., Rea, M., Holmes, N., Seedat, Z. A., Leggett, J., Shah, V., Osborne, J., Bowtell, R., & Brookes, M. J. (2021). Measuring functional connectivity with wearable MEG. NeuroImage, 230, 117815. 10.1016/J.NEUROIMAGE.2021.117815 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boto, E., Holmes, N., Leggett, J., Roberts, G., Shah, V., Meyer, S. S., Muñoz, L. D., Mullinger, K. J., Tierney, T. M., Bestmann, S., Barnes, G. R., Bowtell, R., & Brookes, M. J. (2018). Moving magnetoencephalography towards real-world applications with a wearable system. Nature, 555(7698), 657–661. 10.1038/nature26147 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Boto, E., Meyer, S. S., Shah, V., Alem, O., Knappe, S., Kruger, P., Fromhold, T. M., Lim, M., Glover, P. M., Morris, P. G., Bowtell, R., Barnes, G. R., & Brookes, M. J. (2017). A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers. NeuroImage, 149, 404–414. 10.1016/j.neuroimage.2017.01.034 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brookes, M. J., Boto, E., Rea, M., Shah, V., Osborne, J., Holmes, N., Hill, R. M., Leggett, J., Rhodes, N., & Bowtell, R. (2021). Theoretical advantages of a triaxial optically pumped magnetometer magnetoencephalography system. NeuroImage, 236, 118025. 10.1016/j.neuroimage.2021.118025 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Buzsáki, G., & Draguhn, A. (2004). Neuronal oscillations in cortical networks. Science, 304(5679), 1926–1929. 10.1126/science.1099745 [DOI] [PubMed] [Google Scholar]
- Chang, W. T., Jääskeläinen, I. P., Belliveau, J. W., Huang, S., Hung, A. Y., Rossi, S., & Ahveninen, J. (2015). Combined MEG and EEG show reliable patterns of electromagnetic brain activity during natural viewing. NeuroImage, 114, 49–56. 10.1016/j.neuroimage.2015.03.066 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Cohen, D. (1968). Magnetoencephalography: Evidence of magnetic fields produced by alpha-rhythm currents. Science, 161(3843), 784–786. 10.1126/science.161.3843.784 [DOI] [PubMed] [Google Scholar]
- Cohen, D. (1972). Magnetoencephalography: Detection of the brain’s electrical activity with a superconducting magnetometer. Science, 175(4022), 664–666. 10.1126/science.175.4022.664 [DOI] [PubMed] [Google Scholar]
- Cohen, M. X. (2014). Analyzing neural time series data: Theory and Practice. The MIT Press. 10.7551/mitpress/9609.001.0001 [DOI] [Google Scholar]
- Cox, R. W. (1996). AFNI: Software for analysis and visualization of functional magnetic resonance neuroimages. Computers and Biomedical Research, 29(3), 162–173. 10.1006/cbmr.1996.0014 [DOI] [PubMed] [Google Scholar]
- Dalal, S. S., Baillet, S., Adam, C., Ducorps, A., Schwartz, D., Jerbi, K., Bertrand, O., Garnero, L., Martinerie, J., & Lachaux, J.-P. (2009). Simultaneous MEG and intracranial EEG recordings during attentive reading. NeuroImage, 45(4), 1289–1304. 10.1016/j.neuroimage.2009.01.017 [DOI] [PubMed] [Google Scholar]
- Freeman, W. J., Ahlfors, S. P., & Menon, V. (2009). Combining fMRI with EEG and MEG in order to relate patterns of brain activity to cognition. International Journal of Psychophysiology, 73(1), 43–52. 10.1016/j.ijpsycho.2008.12.019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ghinst, M. V., Bourguignon, M., de Beeck, M. O., Wens, V., Marty, B., Hassid, S., Choufani, G., Jousmäki, V., Hari, R., Bogaert, P. V., Goldman, S., & Tiège, X. D. (2016). Left superior temporal gyrus is coupled to attended speech in a cocktail-party auditory scene. Journal of Neuroscience, 36(5), 1596–1606. 10.1523/JNEUROSCI.1730-15.2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Giraud, A.-L., & Poeppel, D. (2012). Cortical oscillations and speech processing: Emerging computational principles and operations. Nature Neuroscience, 15(4), 511–517. 10.1038/nn.3063 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hämäläinen, M., Hari, R., Ilmoniemi, R. J., Knuutila, J., & Lounasmaa, O. V. (1993). Magnetoencephalography—Theory, instrumentation, and applications to noninvasive studies of the working human brain. Reviews of Modern Physics, 65(2), 413–497. 10.1103/RevModPhys.65.413 [DOI] [Google Scholar]
- Hasson, U., Malach, R., & Heeger, D. J. (2010). Reliability of cortical activity during natural stimulation. Trends in Cognitive Sciences, 14(1), 40–48. 10.1016/j.tics.2009.10.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hasson, U., Nir, Y., Levy, I., Fuhrmann, G., & Malach, R. (2004). Intersubject synchronization of cortical activity during natural vision. Science, 303(5664), 1634–1640. 10.1126/science.1089506 [DOI] [PubMed] [Google Scholar]
- Haufe, S., DeGuzman, P., Henin, S., Arcaro, M., Honey, C. J., Hasson, U., & Parra, L. C. (2018). Elucidating relations between fMRI, ECOG, and EEG through a common natural stimulus. NeuroImage, 179, 79–91. 10.1016/j.neuroimage.2018.06.016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hill, R., Boto, E., Rea, M., Holmes, N., Leggett, J., Coles, L., Papastavrou, M., Everton, S., Hunt, B., Sims, D., Bowtell, R., & Brookes, M. (2020). Multi-channel whole-head OPM-MEG: Helmet design and a comparison with a conventional system. NeuroImage, 219, 116995. 10.1016/j.neuroimage.2020.116995 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Honey, C. J., Thesen, T., Donner, T. H., Silbert, L. J., Carlson, C. E., Devinsky, O., Doyle, W. K., Rubin, N., Heeger, D. J., & Hasson, U. (2012). Slow cortical dynamics and the accumulation of information over long timescales. Neuron, 76(2), 423–434. 10.1016/j.neuron.2012.08.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Iivanainen, J., Carter, T. R., Trumbo, M. C., McKay, J., Taulu, S., Wang, J., Stephen, J. M., Schwindt, P. D., & Borna, A. (2023). Single-trial classification of evoked responses to auditory tones using OPM- and SQUID-MEG. Journal of Neural Engineering, 20(5), 056032. 10.1088/1741-2552/acfcd9 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Iivanainen, J., Stenroos, M., & Parkkonen, L. (2017). Measuring MEG closer to the brain: Performance of on-scalp sensor arrays. NeuroImage, 147, 542–553. 10.1016/J.NEUROIMAGE.2016.12.048 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jääskeläinen, I. P., Koskentalo, K., Balk, M. H., Autti, T., Kauramäki, J., Pomren, C., & Sams, M. (2008). Inter-subject synchronization of prefrontal cortex hemodynamic activity during natural viewing. The Open Neuroimaging Journal, 2(1), 14–19. 10.2174/1874440000802010014 [DOI] [PMC free article] [PubMed] [Google Scholar]
- James, O., Park, H., & Kim, S. G. (2019). Impact of sampling rate on statistical significance for single subject fMRI connectivity analysis. Human Brain Mapping, 40(11), 3321–3337. 10.1002/hbm.24600 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim, D., Joo, E. Y., Seo, D.-W., Kim, M.-Y., Lee, Y.-H., Kwon, H. C., Kim, J.-M., & Hong, S. B. (2016). Accuracy of MEG in localizing irritative zone and seizure onset zone: Quantitative comparison between MEG and intracranial EEG. Epilepsy Research, 127, 291–301. 10.1016/j.eplepsyres.2016.08.013 [DOI] [PubMed] [Google Scholar]
- Lankinen, K., Saari, J., Hari, R., & Koskinen, M. (2014). Intersubject consistency of cortical MEG signals during movie viewing. NeuroImage, 92, 217–224. 10.1016/j.neuroimage.2014.02.004 [DOI] [PubMed] [Google Scholar]
- Lankinen, K., Saari, J., Hlushchuk, Y., Tikka, P., Parkkonen, L., Hari, R., & Koskinen, M. (2018). Consistency and similarity of MEG- and fMRI-signal time courses during movie viewing. NeuroImage, 173, 361–369. 10.1016/j.neuroimage.2018.02.045 [DOI] [PubMed] [Google Scholar]
- Lew, B. J., Fitzgerald, E. E., Ott, L. R., Penhale, S. H., & Wilson, T. W. (2021). Three-year reliability of MEG resting-state oscillatory power. NeuroImage, 243, 118516. 10.1016/j.neuroimage.2021.118516 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lin, C.-H., Tierney, T. M., Holmes, N., Boto, E., Leggett, J., Bestmann, S., Bowtell, R., Brookes, M. J., Barnes, G. R., & Miall, R. C. (2019). Using optically pumped magnetometers to measure magnetoencephalographic signals in the human cerebellum. The Journal of Physiology, 597(16), 4309–4324. 10.1113/JP277899 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Liu, X., Dai, Y., Xie, H., & Zhen, Z. (2022). A studyforrest extension, MEG recordings while watching the audio-visual movie “forrest gump”. Scientific Data, 9(1), 206. 10.1038/s41597-022-01299-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lumet, S. (1975). Dog day afternoon [Film]. Warner Bros. https://www.imdb.com/title/tt0072890/ [Google Scholar]
- MacEvoy, S. P., & Epstein, R. A. (2011). Constructing scenes from objects in human occipitotemporal cortex. Nature Neuroscience, 14(10), 1323–1329. 10.1038/nn.2903 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Maffei, A. (2020). Spectrally resolved EEG intersubject correlation reveals distinct cortical oscillatory patterns during free-viewing of affective scenes. Psychophysiology, 57(11), e13652. 10.1111/psyp.13652 [DOI] [PubMed] [Google Scholar]
- Marhl, U., Jodko-Władzińska, A., Brühl, R., Sander, T., & Jazbinšek, V. (2022). Transforming and comparing data between standard squid and OPM-MEG systems. PLoS One, 17(1), e0262669. 10.1371/journal.pone.0262669 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martín-Buro, M. C., Garcés, P., & Maestú, F. (2016). Test-retest reliability of resting-state magnetoencephalography power in sensor and source space. Human Brain Mapping, 37(1), 179–190. 10.1002/hbm.23027 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Martinovic, J., & Busch, N. A. (2011). High frequency oscillations as a correlate of visual perception. International Journal of Psychophysiology, 79(1), 32–38. 10.1016/j.ijpsycho.2010.07.004 [DOI] [PubMed] [Google Scholar]
- Mejia, A. F., Nebel, M. B., Shou, H., Crainiceanu, C. M., Pekar, J. J., Mostofsky, S., Caffo, B., & Lindquist, M. A. (2015). Improving reliability of subject-level resting-state fMRI parcellation with shrinkage estimators. NeuroImage, 112, 14–29. 10.1016/J.NEUROIMAGE.2015.02.042 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mesulam, M. M. (2022). Temporopolar regions of the human brain. Brain, 146(1), 20–41. 10.1093/brain/awac339 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Michelmann, S., Bowman, H., & Hanslmayr, S. (2016). The temporal signature of memories: Identification of a general mechanism for dynamic memory replay in humans. PLoS Biology, 14(8), e1002528. 10.1371/journal.pbio.1002528 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Mukamel, R., Gelbard, H., Arieli, A., Hasson, U., Fried, I., & Malach, R. (2005). Neuroscience: Coupling between neuronal firing, field potentials, and fMRI in human auditory cortex. Science, 309(5736), 951–954. 10.1126/science.1110913 [DOI] [PubMed] [Google Scholar]
- Nakatani, H., Nakamura, Y., & Okanoya, K. (2022). Respective involvement of the right cerebellar crus I and II in syntactic and semantic processing for comprehension of language. The Cerebellum, 22(4), 739–755. 10.1007/s12311-022-01451-y [DOI] [PubMed] [Google Scholar]
- Nastase, S. A., Gazzola, V., Hasson, U., & Keysers, C. (2019). Measuring shared responses across subjects using intersubject correlation. Social Cognitive and Affective Neuroscience, 14(6), 667–685. 10.1093/scan/nsz037 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nolte, G. (2003). The magnetic lead field theorem in the quasi-static approximation and its use for magnetoencephalography forward calculation in realistic volume conductors. Physics in Medicine and Biology, 48(22), 3637–3652. 10.1088/0031-9155/48/22/002 [DOI] [PubMed] [Google Scholar]
- Nourski, K. V., Reale, R. A., Oya, H., Kawasaki, H., Kovach, C. K., Chen, H., Howard, M. A., & Brugge, J. F. (2009). Temporal envelope of time-compressed speech represented in the human auditory cortex. The Journal of Neuroscience, 29(49), 15564–15574. 10.1523/jneurosci.3065-09.2009 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Nugent, A. C., Andonegui, A. B., Holroyd, T., & Robinson, S. E. (2022). On-scalp magnetocorticography with optically pumped magnetometers: Simulated performance in resolving simultaneous sources. NeuroImage: Reports, 2(2), 100093. 10.1016/j.ynirp.2022.100093 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Oostenveld, R., Fries, P., Maris, E., & Schoffelen, J. M. (2010). Fieldtrip: Open source software for advanced analysis of MEG, EEG, and invasive electrophysiological data. Computational Intelligence and Neuroscience, 2011, 1–9. 10.1155/2011/156869 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Pajula, J., & Tohka, J. (2014). Effects of spatial smoothing on inter-subject correlation based analysis of fMRI. Magnetic Resonance Imaging, 32(9), 1114–1124. 10.1016/J.MRI.2014.06.001 [DOI] [PubMed] [Google Scholar]
- Park, H., Ince, R. A., Schyns, P. G., Thut, G., & Gross, J. (2018). Representational interactions during audiovisual speech entrainment: Redundancy in left posterior superior temporal gyrus and synergy in left motor cortex. PLoS Biology, 16(8), e2006558. 10.1371/journal.pbio.2006558 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Parvizi, J., & Kastner, S. (2018). Promises and limitations of human intracranial electroencephalography. Nature Neuroscience, 21(4), 474–483. 10.1038/s41593-018-0108-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Proudfoot, M., Woolrich, M. W., Nobre, A. C., & Turner, M. R. (2014). Magnetoencephalography. Practical Neurology, 14(5), 336–343. 10.1136/practneurol-2013-000768 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ray, S., & Maunsell, J. H. R. (2011). Different origins of gamma rhythm and high-gamma activity in macaque visual cortex (L. Ungerleider, Ed.). PLoS Biology, 9(4), e1000610. 10.1371/journal.pbio.1000610 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rea, M., Boto, E., Holmes, N., Hill, R., Osborne, J., Rhodes, N., Leggett, J., Rier, L., Bowtell, R., Shah, V., & Brookes, M. J. (2022). A 90-channel triaxial magnetoencephalography system using optically pumped magnetometers. Annals of the New York Academy of Sciences, 1517(1), 107–124. 10.1111/nyas.14890 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rhodes, N., Rea, M., Boto, E., Rier, L., Shah, V., Hill, R. M., Osborne, J., Doyle, C., Holmes, N., Coleman, S. C., Mullinger, K., Bowtell, R., & Brookes, M. J. (2023). Measurement of frontal midline theta oscillations using OPM-MEG. NeuroImage, 271, 120024. 10.1016/j.neuroimage.2023.120024 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Rier, L., Michelmann, S., Ritz, H., Shah, V., Hill, R. M., Osborne, J., Doyle, C., Holmes, N., Bowtell, R., Brookes, M. J., Norman, K. A., Hasson, U., Cohen, J. D., & Boto, E. (2023). Test-retest reliability of the human connectome: An OPM-MEG study. Imaging Neuroscience, 1, 1–20. 10.1162/imag_a_00020 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Safar, K., Vandewouw, M. M., Sato, J., Devasagayam, J., Hill, R. M., Rea, M., Brookes, M. J., & Taylor, M. J. (2024). Using optically pumped magnetometers to replicate task-related responses in next generation magnetoencephalography. Scientific Reports, 14(1), 6513. 10.1038/s41598-024-56878-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schaworonkow, N., Blythe, D. A., Kegeles, J., Curio, G., & Nikulin, V. V. (2015). Power-law dynamics in neuronal and behavioral data introduce spurious correlations. Human Brain Mapping, 36(8), 2901–2914. 10.1002/hbm.22816 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Schomer, D.L., & Lopes da Silva, F. H. (Eds). (2018). Niedermeyer’s electroencephalography: Basic principles, clinical applications, and related fields (7th edn). Oxford University Press. 10.1093/med/9780190228484.001.0001 [DOI] [Google Scholar]
- Schreiber, T., & Schmitz, A. (1996). Improved surrogate data for nonlinearity tests. Physical Review Letters, 77(4), 635–638. 10.1103/PhysRevLett.77.635 [DOI] [PubMed] [Google Scholar]
- Sinai, A., Crone, N. E., Wied, H. M., Franaszczuk, P. J., Miglioretti, D., & Boatman-Reich, D. (2009). Intracranial mapping of auditory perception: Event-related responses and electrocortical stimulation. Clinical Neurophysiology, 120(1), 140–149. 10.1016/j.clinph.2008.10.152 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Stelzer, J., Chen, Y., & Turner, R. (2013). Statistical inference and multiple testing correction in classification-based multi-voxel pattern analysis (MVPA): Random permutations and cluster size control. NeuroImage, 65, 69–82. 10.1016/j.neuroimage.2012.09.063 [DOI] [PubMed] [Google Scholar]
- Sun, T., Chi, X., Peng, Y., Zhang, Q., Liu, K., Ma, Y., Ding, M., Ji, N., & Zhang, Y. (2024). A first-in-human application of OPM-MEG for localizing motor activity area: Compared to functional MRI. NeuroImage, 304, 120953. 10.1016/j.neuroimage.2024.120953 [DOI] [PubMed] [Google Scholar]
- Tierney, T. M., Holmes, N., Meyer, S. S., Boto, E., Roberts, G., Leggett, J., Buck, S., Duque-Muñoz, L., Litvak, V., Bestmann, S., Baldeweg, T., Bowtell, R., Brookes, M. J., & Barnes, G. R. (2018). Cognitive neuroscience using wearable magnetometer arrays: Non-invasive assessment of language function. NeuroImage, 181, 513–520. 10.1016/J.NEUROIMAGE.2018.07.035 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Tzourio-Mazoyer, N., Landeau, B., Papathanassiou, D., Crivello, F., Etard, O., Delcroix, N., Mazoyer, B., & Joliot, M. (2002). Automated anatomical labeling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single-subject brain. NeuroImage, 15(1), 273–289. 10.1006/nimg.2001.0978 [DOI] [PubMed] [Google Scholar]
- Veen, B. V., Drongelen, W. V., Yuchtman, M., & Suzuki, A. (1997). Localization of brain electrical activity via linearly constrained minimum variance spatial filtering. IEEE Transactions on Biomedical Engineering, 44(9), 867–880. 10.1109/10.623056 [DOI] [PubMed] [Google Scholar]
- Wang, X., Teng, P., Meng, Q., Jiang, Y., Wu, J., Li, T., Wang, M., Guan, Y., Zhou, J., Sheng, J., Gao, J. H., & Luan, G. (2024). Performance of optically pumped magnetometer magnetoencephalography: Validation in large samples and multiple tasks. Journal of Neural Engineering, 21(6), 066033. 10.1088/1741-2552/ad9680 [DOI] [PubMed] [Google Scholar]
- Wernicke, C. (1974). Der aphasische symptomenkomplex. Springer Berlin Heidelberg. 10.1007/978-3-642-65950-8_1 [DOI] [Google Scholar]
- Woolrich, M. W., Ripley, B. D., Brady, M., & Smith, S. M. (2001). Temporal autocorrelation in univariate linear modeling of fMRI data. NeuroImage, 14(6), 1370–1386. 10.1006/nimg.2001.0931 [DOI] [PubMed] [Google Scholar]
- Yang, A. I., Wang, X., Doyle, W. K., Halgren, E., Carlson, C., Belcher, T. L., Cash, S. S., Devinsky, O., & Thesen, T. (2012). Localization of dense intracranial electrode arrays using magnetic resonance imaging. NeuroImage, 63(1), 157–165. 10.1016/j.neuroimage.2012.06.039 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Ye, Z., Rüsseler, J., Gerth, I., & Münte, T. F. (2017). Audiovisual speech integration in the superior temporal region is dysfunctional in dyslexia. Neuroscience, 356, 1–10. 10.1016/j.neuroscience.2017.05.017 [DOI] [PubMed] [Google Scholar]
- Zarahn, E., Aguirre, G. K., & D’esposito, M. (1997). Empirical analyses of bold fmri statistics i. spatially unsmoothed data collected under null-hypothesis conditions. NeuroImage, 5(3), 179–197. 10.1006/nimg.1997.0263 [DOI] [PubMed] [Google Scholar]
- Zumer, J. M., Brookes, M. J., Stevenson, C. M., Francis, S. T., & Morris, P. G. (2010). Relating bold fMRI and neural oscillations through convolution and optimal linear weighting. NeuroImage, 49(2), 1479–1489. 10.1016/J.NEUROIMAGE.2009.09.020 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
Data and analysis code underlying the conclusions of the manuscript are available at the Open Science Framework (OSF): https://osf.io/urjvb. The raw OPM data from Rier et al. (2023) are available on Zenodo: https://doi.org/10.5281/zenodo.7525341







