Skip to main content
. 2005 Dec;12(12):1425–1428. doi: 10.1128/CDLI.12.12.1425-1428.2005

TABLE 1.

Performance of kits with known specimens

Specimen (total no.) No. of specimens identified correctly by:
Enzaids Murex Genscreen HIV-CheX Comb Aids
Western blot confirmed, HIV-I positive (n = 100) 100 100 100 89 100
HIV-I negative, in FDA-approved ELISA (n = 100) 100 100 100 100 100
BBI panel (n = 6) 4a 6 NDb ND ND
NEW LAV BLOT II confirmed, HIV-II positive (n = 4) 4 4 4 4 4
In-house seroconversion panel (n = 9) 9 9 9 7 9
a

The two initial specimens, from 0 and 9 days, were not detected.

b

ND, not done.