Skip to main content
letter
. 2005 Dec;43(12):6212–6213. doi: 10.1128/JCM.43.12.6212-6213.2005

TABLE 2.

Inter- and intralaboratory comparisons of retapamulin zone diameter resultsa

Zone diam (mm) No. of occurrences of organism in laboratory:
A B C D E F G H Total
S. aureus ATCC 25923
    21 0
    22 1 1
    23 6 2 12 7 27b
    24 3 9 11 28 24 75b
    25 21 4 21 3 17 20 8 94b
    26 36 8 21 20 18 16 119b
    27 20 3 1 32 6 4 66b
    28 17 14 5 5 1 42b
    29 11 19 30b
    30 21 21b
    31 5 5
    32 0
S. pneumoniae ATCC 49619
    11 0
    12 4 4
    13 28 28c
    14 26 9 13 19 67c
    15 2 19 29 30 1 21 102c
    16 18 6 13 21 10 3 71c
    17 6 11 12 14 6 13 62c
    18 8 1 5 15 4 18 51c
    19 9 19 28c
    20 6 6
    21 1 1
    22 0
a

Results represent testing of two QC strains for an eight-medical-center protocol meeting the study design guidelines found in CLSI M23-A2 (4).

b

Proposed QC range (23 to 30 mm) that included 98.8% of reported zone diameters from eight laboratories.

c

Proposed QC range (13 to 19 mm) that included 97.4% of reported zone diameters from seven laboratories (laboratory H results were omitted from analysis).