Abstract
Purpose
Corneal guttae are the hallmark pathologic phenotype of Fuchs endothelial corneal dystrophy (FECD). Tools performing quantitative analysis of guttae and their changes over time are needed as medical therapies advance. We developed and validated a novel automated software platform that reports the guttae area ratio (GAR) in FECD specular microscopy images.
Methods
Patients with FECD with modified Krachmer grade (mKg) 2 to 5 disease and no prior keratoplasty were enrolled and imaged prospectively with one specular microscope unit. In a validation study, expert manual grading and automated software grading of GAR were compared for agreement. Additional images were analyzed to correlate automated GAR with disease grade and subclinical edema by Scheimpflug imaging.
Results
Nineteen images were included in the validation study. Mean manual GAR (0.400 ± 0.213) and mean automated GAR (0.395 ± 0.182) demonstrated equivalence (P = 0.003). Additionally, 446 images (109 eyes) demonstrated that automated GAR increases with worsening disease in mKg 2 to 5. Most mKg 4 images (n = 172, 93.5%; 49 eyes) and mKg 5 images (n = 59, 55.1%; 19 eyes) had no evidence of definite subclinical edema. Repeat image analysis demonstrated high repeatability of GAR (intraclass correlation = 0.85; 95% confidence interval, 0.78–0.90).
Conclusions
This study validates that automated software for guttae quantification in FECD specular microscopy images is equivalent to manual expert grading. GAR, an objective metric that increases in direct proportion to mKg, may precede development of subclinical edema.
Translational Relevance
This software supports precise quantification of FECD disease secondary to guttae, allowing for objective measurement of disease progression and therapeutic efficacy.
Keywords: Fuchs endothelial corneal dystrophy, specular microscopy, guttae, medical therapeutics, diagnostic imaging
Introduction
Fuchs endothelial corneal dystrophy (FECD) has been the leading indication for corneal transplantation in the United States for the past decade.1 In 2024, over 17,000 transplants were performed for FECD in the United States.1 Currently, FECD is a disease diagnosed clinically and phenotypically by detecting the presence of guttae in the endothelium of the posterior cornea.2,3 Guttae may be visualized grossly on examination using slit-lamp biomicroscopy and in more minute detail using specular microscopy imaging. As these excrescences result from abnormal secretions of diseased corneal endothelial cells and tend to grow in both size and number over years, guttae are important to detect and track for both patients with FECD and their clinicians. The principal grading schema used clinically to describe FECD severity, a scale developed by Jay Krachmer at Iowa in the 1970s, is based almost entirely on the diameter of corneal endothelium occupied by guttae.4
Although Descemet membrane endothelial keratoplasty remains the standard of care for FECD treatment, alternative approaches to corneal transplantation are evolving rapidly. Novel therapeutic approaches target earlier stages of FECD. Advancements include topical Rho-associated kinase inhibitors to promote endothelial cell migration and wound closure after Descemet stripping only, intracameral injection of cultured corneal endothelial cell transplants, and identification of pharmacotherapeutics and gene therapy targets to slow or suppress disease expression.5–14 These therapeutic advancements may prevent the development of secondary structural changes from primary corneal endothelial cell pathology that precede corneal edema, including subclinical corneal edema from pump dysfunction and/or guttae accumulation from abnormal extracellular matrix deposition. These interventions may also limit the development of subclinical anterior stromal haze and subepithelial fibrosis, changes that correlate with daily activity impairment and often persist despite surgical intervention.15–20 As such, there is a more immediate and practical need to quantify corneal endothelial cell morphology and corneal guttae objectively as markers of disease progression. Although the modified Krachmer grading scale widely used to characterize FECD progression reflects the increased quantity and confluence of guttae as disease severity worsens, this scale lacks the objective precision needed to quantify disease progression at the cellular level, with moderate interobserver agreement.4,21
Given these considerations, ancillary testing modalities that yield high-resolution images of the corneal endothelium as the primary site of disease in endotheliopathies such as FECD are indicated to permit quantification and monitoring of disease-specific cellular and extracellular alterations. Studies have evaluated automated models designed to quantify endothelial cell density in FECD using confocal imaging in humans and specular microscopy in mouse models, but to date, they have not expanded this work to widely available specular microscopy imaging modalities or human models, respectively.22,23 In 2014, McLaren et al.23 introduced effective endothelial cell density (EECD), using automated thresholding with manual adjustments to calculate regions in confocal images free of guttae. Sierra et al.24 expanded on this work in 2023 with UNet-based segmentation and the inverse of EECD, reporting the ratio of guttae to total analyzed specular microscopy area (guttae area ratio [GAR], reported as a percentage of occupancy per image). We have built upon this work and concentrated efforts to quantify guttae using specular microcopy imaging to directly capture disease progression characterized by the accumulation of guttae. In this study, we present a novel automated software application (GuttEye; ADCIS, Saint-Contest, France) used to report GAR25,26 in FECD eyes with modified Krachmer scale disease grades (mKg) 2 to 5, representing FECD disease stages without clinically detectable corneal edema most likely to be monitored for presurgical disease progression. In addition to performing software validation testing, this study determined the correlation between clinical grading of patients with FECD using the modified Krachmer scale and automated grading of those same patients using this software tool to establish the utility of this automated approach for grading and tracking the extracellular matrix pathology of FECD disease progression. Furthermore, this study reports on the relationship between GAR and tomographic markers of subclinical edema to investigate the relationship between disease burden visualized at a cellular level and structural changes appreciated on tomography.20 We anticipate that this software tool will be useful in tracking endothelial pathology and permit more robust quantification of disease progression in patients with FECD in both research and clinical applications.
Methods
The University of Iowa Institutional Review Board approved this prospective study (IRB 202301213), which complied with the Health Insurance Portability and Accountability Act and adhered to the tenets of the Declaration of Helsinki. Informed consent was obtained from all volunteers who participated in this study after an explanation of the nature and possible consequences of participation.
Participant Enrollment
Participants meeting the inclusion criteria for this study were between 18 and 99 years old, with a diagnosis of FECD and modified Krachmer grades of 2 to 5. Participants with FECD modified Krachmer grade 6 disease were excluded from analysis in this study, as the corneal edema present in this patient group, by definition, tends to drive the decision for surgery and degrade specular microscopy image quality.23 FECD diagnosis and modified Krachmer grade assignment (mKg) were determined by a cornea fellowship-trained ophthalmologist (MAG).27 Participants who had prior keratoplasty performed in both eyes were excluded from participation.
Software Design
GuttEye software was developed by authors (JAA, JMS, PD, FP, MAG) using C# and the .NET Framework (v4.7.2). The software incorporates an advanced image-processing algorithm built with the Aphelion SDK (v4.7.0). The guttae detection algorithm begins by correcting light intensity variations to ensure uniform distribution across the image. This is followed by applying a threshold to isolate the darker regions representing the guttae. Subsequently, morphologic shape corrections are performed to minimize false positives and refine the contours. Measurements are then taken from these contours based on the selected resolution. The software allows for the definition of multiple resolution profiles, accommodating various specular microscopes and configurations (Fig. 1).
Figure 1.
Patient images, organized by date and eye, are imported into this software platform as a JPEG file. (A) The user may indicate the device and magnification used to capture the images. (B) The user clicks on the “Process” button to capture guttae automatically. Manual edits may be made if desired. (C) Reported results include the total area and detected area of guttae in µm². GAR is reported, and the image is categorized into a range of guttae quantity. The user may mark image quality, indicate the presence of edema in the patient, or flag it as needed. All reported results may be exported for multiple images in a CSV file.
Image Analysis
FECD eyes were imaged using the Konan NSP-9900 specular microscope (Konan Medical, Hyogo, Japan). Central images were obtained for all participant eyes meeting study criteria. For the validation study, the left eye image was analyzed only if the right eye image was not of sufficient quality for analysis. For the clinical correlation study, right eye and left eye images meeting the inclusion criteria were analyzed. Specular microscopy images were graded for quality (MAG) and included for analysis only if cell boundaries were clearly identifiable in regions void of guttae. Images were exported from the specular microscope as JPEG files. Using initial datasets of prospectively collected specular microscopy images from patients with FECD, expert graders (JMS, MAG) manually analyzed the images using the software's tool palette to determine the GAR as a percentage of disease involvement per image. This manual analysis was used to support a pilot study for sample size determination and an analysis comparing manually graded and automated software graded values (Fig. 2). Regions of guttae were outlined using a mouse cursor in the manual tracing mode available on the software. Graders were instructed that guttae identified should meet the definitional criteria—circular, hypointense areas surrounded by discrete foci of hyperreflective cells—as guided by published methods.24,28,29 Graders were also advised that the confluence of guttae may present as large, geographic areas of hypointensity and that narrow regions of hypointensity between endothelial cells should be distinguished from guttae as normal intercellular junctions. Subsequently, using additional datasets of prospectively collected specular microscopy images from patients with FECD, images were processed by the automated software tool (SRS) to calculate the GAR.25
Figure 2.
(A) Unmarked specular microscopy image (mKg 4). (B) Manual guttae grading by MAG (GAR 54.4%). (C) Manual guttae grading by JMS (GAR 55.1%). (D) Automated software grading (GAR 59.3%).
For the clinical correlation study, all participants were assessed by an expert grader (MAG) for their mKg clinically.4 Scheimpflug corneal imaging (Pentacam HR; Oculus, Arlington, WA, USA) obtained concurrently with specular microscopy and clinical examination for each participant was assessed for signs of subclinical edema, defined as the loss of regular isopachs, displacement of the thinnest point of the cornea from the corneal apex, and focal posterior corneal depression on the four-map refractive display.20 Participants received a score between 0 and 3, corresponding to the number of features present in tomographic scans. A score of 0 to 1 was classified as a lack of definite subclinical edema, and a score of 2 to 3 was classified as the presence of definite subclinical edema.20
Statistical Analysis
A pilot study of prospectively collected images was used to determine the sample size of this study for a one-sample equivalence test of means for the paired difference between manual and automated grading of GAR. Pilot study data included 25 images from participants who met the above inclusion and exclusion criteria with image quality determined to be suitable for analysis. Images were manually analyzed by the same graders (JMS, MAG) and independently analyzed with the automated software (SRS). This pilot study yielded an observed mean difference of 0.023 between SRS and manual, with a standard deviation (SD) of the difference of 0.062. Using an equivalence bound greater than the observed mean and smaller than the observed SD of the difference, a sample size of 19 images was determined to be necessary, one image per eye, for a prospective study with 0.05 bounds of equivalence, SD of the paired difference of 0.062, α of 0.05, and power of 0.90. This sample size also ensures, with 90% probability, that the 95% confidence interval (CI) of the paired mean difference will be no wider than mean ± 0.04.
Following the pilot study, a separate group of participants and images meeting the same inclusion, exclusion, and image quality criteria were analyzed for the software validation and clinical correlation arms of this study. Data were collected and plotted using Excel (version 16.82; Microsoft Corporation, Redmond, WA, USA) and analyzed using SAS (version 9.4; SAS/STAT 15.2, Cary, NC, USA) and SPSS (version 29.0.1.1; IBM Corporation, Armonk, NY, USA). In the manual versus automated GAR analysis, agreement between manual and automated grading of GAR values was assessed using the two one-sided t-test (TOST) for equivalence of means for paired data. This analysis was powered at 90% for a paired equivalence TOST, with a 0.05 bound for equivalence, to detect a zero-mean difference between manual and automated grading, where P ≤ 0.05 was considered statistically significant. In addition, a Bland–Altman plot with 95% limits of agreement was constructed to show the distribution of the difference between manual and automated grading. For the analysis of the association of automated GAR with clinical mKg disease grade, differences in mean GAR between eyes with mKg disease grade were examined using the generalized mixed model fitted by the generalized estimating equation (GEE) method to account for correlation of GAR from multiple images from the same eye with the test of linear mean contrast to assess directional trend in mean GAR in relation to mKg disease grade, as well as the Tukey post hoc test for pairwise mean GAR comparisons between grades. The generalized mixed model fitted by the GEE method was also used to assess the association between GAR and subclinical edema. Variability of GAR between repeat images within a single imaging session was estimated using a random-effects model, and the distribution of the GAR differences between these repeat images was presented by constructing a Bland–Altman plot with 95% limits of agreement.
Results
Nineteen FECD images were included for analysis in the validation study to test the equivalence of automated software compared to manual grading of guttae and GAR (Fig. 2). The mean age was 66.8 ± 10.1 years, and 13 participants (68.4%) were female. Fifteen (75%) images were obtained from the right eye. The mean mKg was 4.4 ± 0.84, and 15 images were obtained with the participant fixated on the central target (Table 1).
Table 1.
Validation Study Demographics, Manual Versus Automated Software GAR (n = 19 Images)
| Characteristic | Value |
|---|---|
| Age, mean ± SD, y | 66.8 ± 10.1 |
| Sex (female) | 13 (68.4) |
| Laterality (OD) | 15 (78.9) |
| Phakic | 12 (63.2) |
| Modified Krachmer grade | |
| Grade 3 | 2 (10.5) |
| Grade 4 | 9 (47.4) |
| Grade 5 | 6 (31.6) |
| Grade 6 | 2 (10.5) |
| Central fixation target | 15 (78.9) |
Values are presented as number (%) unless otherwise indicated.
The mean GAR by manual grader was 0.397 ± 0.208 (MAG) vs. 0.404 ± 0.219 (JMS). Manual analysis conducted by both graders differed by less than 5% in 18 images (94.7%), with a mean difference of 0.007 (95% CI, −0.005 to 0.019). The mean combined manual GAR was 0.400 ± 0.213. Mean automated GAR was 0.395 ± 0.182. The TOST for equivalence resulted in a lower bound of −0.018 and an upper bound of 0.030, both within the 0.05 equivalence bound, indicating equivalence in mean GAR between manual and automated (P = 0.003) (Table 2). A Bland–Altman plot was created to show the distribution between manual and automated GAR. The 95% limits of agreement were −0.093 and 0.104. Seventeen (89.5%) images were within the bounds of agreement (Fig. 3). The largest outlier was of a mKg 5 image, demonstrating that the automated GAR was 0.29 lower than the mean GAR by the two manual graders (Fig. 4).
Table 2.
Manual Versus Automated Software GAR Analysis
| GAR, Mean ± SD | ||||
|---|---|---|---|---|
| MAG | JMS | Mean Difference (95% CI) | P Value | |
| Manual expert grading | 0.397 ± 0.208 | 0.404 ± 0.219 | 0.007 (−0.005 to 0.019) | 0.25 |
| Combined manual vs. automated grading | MAG + JMS | Automated | TOST bounds | |
| 0.400 ± 0.213 | 0.395 ± 0.182 | −0.018 to 0.030 | 0.003 | |
Figure 3.
Bland–Altman plots depicting the agreement between manual and automated GAR with 95% limits of agreement of −0.093 and 0.104. (A) GAR distribution for both manual graders (MAG and JMS) individually compared to automated GAR. (B) GAR distribution for both manual graders combined compared to GAR. Both plots illustrate that 17 of 19 images are within the 95% bounds of agreement.
Figure 4.
(A) Unmarked specular microscopy image (mKg 5). (B) Manual guttae grading by MAG (GAR 75.8%). (C) Manual guttae grading by JMS (GAR 79.5%). (D) Automated software grading (GAR 48.7%).
An additional 446 images from 109 eyes were included in the analysis of the association between automated GAR, expressed as a percentage, and mKg (Table 3). Most images analyzed (n = 388, 87.0%) were from participants with mKg 4 and 5 FECD. Over 85% of the images (n = 384) were obtained with the participant in central fixation. Based on GAR from eyes with both central and peripheral fixation images from the same session, linear mixed-model analysis showed that the difference in GAR between central and peripheral location did not vary by mKg grade (P = 0.88). GAR was significantly greater in central fixation images than peripheral fixation images, with a mean difference of 33.88% (95% CI, 13.46%–54.29%; P = 0.005). Mean GARs at each mKg grade are presented in Table 4.
Table 3.
Correlation Study Demographics, Automated Software GAR Versus FECD Clinical Grade (n = 446 Images)
| Characteristic | n (%) |
|---|---|
| Laterality (OD) | 228 (51.1) |
| Modified Krachmer grade: participant, eyes, images | |
| Grade 2 | 3, 5, 15 (3.4) |
| Grade 3 | 9, 14, 43 (9.6) |
| Grade 4 | 33, 56, 254 (57.0) |
| Grade 5 | 21, 34, 134 (30.0) |
| Central fixation target | 384 (86.1) |
| Grade 2 | 15 (100) |
| Grade 3 | 41 (95.3) |
| Grade 4 | 201 (79.1) |
| Grade 5 | 127 (94.8) |
| Phakic | 278 (62.3) |
| Grade 2 | 15 (100) |
| Grade 3 | 19 (44.2) |
| Grade 4 | 176 (69.3) |
| Grade 5 | 68 (53.5) |
Table 4.
Automated Software Analysis of Central Versus Peripheral Fixation GAR by FECD Clinical Grade
| Central Fixation (n = 387) | Peripheral Fixation (n = 62) | ||
|---|---|---|---|
| Modified Krachmer grade | Mean GAR ± SD (n) | Mean GAR ± SD (n) | P Value |
| Grade 2 | 8.15 ± 10.93 (15) | 0 (0) | — |
| Grade 3 | 27.00 ± 22.65 (41) | 9.00 ± 9.61 (2) | 0.27 |
| Grade 4 | 44.97 ± 22.20 (201) | 21.73 ± 21.35 (53) | <0.001 |
| Grade 5 | 61.82 ± 17.71 (127) | 33.47 ± 19.39 (7) | <0.001 |
Mixed-model GEE analysis of central fixation images showed a significant linear trend with increased mean automated GAR associated with a higher mKg grade (χ2 = 6.52; P = 0.011) (Fig. 5). Tukey post hoc pairwise mean comparison showed statistically significant differences, with the mean GAR for the higher mKg grades being significantly greater compared to the mean at the lower mKg grades (Fig. 5).
Figure 5.
A total of 384 central fixation images were analyzed for automated GAR (images per grade: mKg 2, n = 15; mKg 3, n = 41; mKg 4, n = 201; mKg 5, n = 127). A statistically significant difference between groups was demonstrated by mixed-model GEE analysis (χ2 = 6.52; P = 0.011). Tukey post hoc pairwise mean comparison indicated significance between all mKg group comparisons (mKg 2 vs. mKg 3, P = 0.056; mKg 3 vs. 4, P = 0.018; mKg 4 vs. 5, P = 0.004; all other mKg, P < 0.001).
A total of 184 images from 53 eyes with FECD mKg 4 and 107 images from 36 eyes with FECD mKg 5 were analyzed using a mixed-effects model fitted with GEE for the presence of subclinical edema on Scheimpflug imaging. Most of the mKg 4 images (n = 172, 93.5%; 49 eyes) did not have evidence of definite subclinical edema (score, 0–1), and only 12 images (6.5%; 4 eyes) had definite subclinical edema (score, 2–3), with a larger mean GAR (45.87 vs. 58.95, respectively) significantly associated with definite subclinical edema (mean GAR difference 13.1; 95% CI, 0.3–25.9; P = 0.046; Fig. 6). Similarly, most mKg 5 images (n = 59, 55.1%; 19 eyes) did not have evidence of definite subclinical edema (score, 0–1), and 48 images (44.9%; 17 eyes) had definite subclinical edema (score, 2–3) but did not show a significant difference in mean GAR (62.02 vs. 61.30, respectively; mean GAR difference −0.7; 95% CI, −8.9 to 7.5; P = 0.86; Fig. 6).
Figure 6.
Scheimpflug images were analyzed for the presence or absence of subclinical edema (loss of regular isopachs, displacement of the thinnest point, and presence of focal posterior depression, 1 point per feature). (A) A total of 184 images from 53 eyes with mKg 4 were analyzed; 172 images had a score of 0 to 1, and 12 images had a score of 2 to 3. GAR was significantly greater when all features were present (45.87 ± 2.95 for scores 0–1 vs. 58.95 ± 5.73 for scores 2–3, P = 0.046). (B) A total of 107 images from 36 eyes with mKg 5 were analyzed; 59 images had a score of 0 to 1, and 48 images had a score of 2 to 3. GAR was not significantly different between groups (62.02 ± 2.82 for scores 0–1 vs. 61.30 ± 3.69 for scores 2–3, P = 0.86).
Eyes with multiple images taken during one session were analyzed to assess variability among repeated images. A total of 365 images (56 participants, 96 eyes, with two to four repeat images within a session) were analyzed. A random-effects model to estimate within-session repeat image variation in GAR yielded a within-session SD of 9.51, with an intraclass correlation (ICC) of 0.85 (95% CI, 0.78–0.90), indicating good to excellent repeatability of GAR from multiple images of the same eye within a session. The distribution of GAR difference between images from the same session is shown in the Bland–Altman plot, with most differences contained within the 95% bounds of agreement of ±26.36 (Fig. 7).
Figure 7.

Bland–Altman plots depicting the agreement between GAR measurements of multiple images taken on eyes during a single session (n = 96 eyes; 365 images, with 2–4 images per eye). The plot shows the distribution of GAR differences between images from the same session, of which most are expected to fall within the 95% bounds of agreement of ±26.36.
Discussion
This study validates a novel automated software platform and tool that reports guttae occupancy per specular micrograph as a quantifiable metric, GAR, with high accuracy and agreement between manual expert and automated software grading in patients with FECD prior to developing clinically detectable corneal edema. Current automated analysis using commercially available specular microscopy software such as programming embedded into a microscope's unit and display is a reliable method to quantify healthy corneal endothelial cell density. However, such automated measurements tend to overestimate corneal endothelial cell density in patients with FECD, particularly in regions encompassing guttae and in the context of cellular polymegathism and pleomorphism of surrounding diseased cells.30,31 Notably, automated measurements in this study reported that more than one-third of mKg 4 and nearly all mKg 5 images were “unanalyzable” for endothelial cell density measurement. Manual measurement of guttae is a time-consuming, unrealistic alternative. Advantages of an automated tool, such as the one we present and validate in this study, include rapid image analysis and the ability for users to make manual alterations to the image area and segmentation if needed. Additional advantages of this software include compatibility with all specular microscopy units currently available, including wide-field imaging (data not shown). Validation of this automated software tool supports its use as a viable alternative to manual grading and analysis for rapid, accurate, and quantifiable measurement of FECD disease impact, making it possible to obtain detailed extracellular and cellular measurements of FECD disease progression over time. Immediately useful in the conduct of clinical research and clinical trial testing, including FECD drug safety and efficacy studies, automated guttae quantification using this software platform may also be used in the future with appropriate regulatory clearance to support clinical decision-making and surgical planning.
Our findings incorporate and build upon previous research reporting GAR as a clinically relevant metric for tracking FECD disease impact and severity. McLaren et al.23 used image thresholding to calculate guttae-free areas of confocal imaging (EECD). They also analyzed the ratio of guttae, R, and demonstrated a high degree of correlation between regions covered by guttae and the clinical severity of FECD. Sierra et al.25 developed UNet-based segmentation software that quantified GAR in FECD grades 1 to 3. Further analysis of this model demonstrated a significant correlation between GAR and FECD modified Krachmer disease grade.25,26 Multivariate regression analysis by Prada et al.26 indicated that of several morphometric parameters (EECD, coefficient of variation, percent hexagonality, and percent GAR), GAR was the strongest predictor of FECD severity. Results from our study support and expand upon these findings by analyzing the relationship between GAR and higher grades of disease, particularly mKg 4 and 5, more commonly found in clinical practice. Results using GuttEye software analysis demonstrate a statistically significant correlation between increasing GAR and modified Krachmer scale disease grade. While Prada et al.26 used an artificial intelligence–based approach (convolutional neural networks) to quantify guttae, our software platform provides an automated output that utilizes thresholding and morphologic corrections to identify and report guttae that can be manually adjusted if deemed necessary by the user.26 Of note, in analyzing central versus peripheral images, Prada et al.26 found a marginally positive correlation between GAR and inferior image location. Our subanalysis of central versus peripheral images, in contrast, demonstrated increased central GAR compared to peripheral GAR in mKg 4 and 5, supporting clinical findings and previous quantification of central guttae that spread peripherally.32
Given the tedious nature of the manual quantification of guttae in FECD, previous studies have investigated automated measurements of guttae progression visualized using slit-lamp biomicroscopy images and retroillumination.33–35 Retroillumination image analytical techniques may permit a broader field of view for objective analysis and comparison to modified Krachmer scale grading.34,36 However, retroillumination-based disease characterization is limited by the lack of specificity at the cellular level. Numerical quantification of individual guttae fails to characterize variation in size and confluence of guttae, which have been determined to influence local cell behavior in FECD.37 Additionally, a variety of technical challenges associated with slit-lamp photography, including photographer experience and skill, patient positioning and gaze, and setting consistency (e.g., equipment and illumination), may introduce significant artifacts that limit the quality, accuracy, and reproducibility of FECD disease grading. In contrast, specular microscopy–based GAR quantification captures disease burden with a greater sensitivity to changes at the cellular level. While overall guttae burden may be objectively quantifiable using other techniques, the technique we report also allows the user to assess guttae patterns occurring at the cellular level that may not be apparent from retroillumination imaging.
Although conventional Scheimpflug corneal tomography reflects structural changes in corneal architecture that precede clinically detectable corneal edema, this imaging modality lacks the objective resolution to detect pathologic changes at the cellular and tissue levels, where function needs to be tracked more precisely. These limitations are particularly relevant in the context of advancements in clinical trials and therapeutics that may target moderate FECD (mKg 3 and 4) and advanced FECD (mKg 5), disease stages representing a large spectrum of patients who may be progressing but may not have developed clinically detectable or subclinical corneal edema, or may have symptoms due to posterior corneal pathology without clinically detectable edema.33,38–40 Furthermore, a linear correlation has been well established between guttae accumulation, visual dysfunction, and advancement of corneal edema.4,27 Guttae are also highly relevant at the molecular level, particularly because these extracellular matrix by-products are implicated in oxidative damage and cell death mechanisms in the corneal endothelium, including ferroptosis and apoptosis.9,11,41–43 As such, quantification of guttae accumulation and the functional anatomy of the posterior cornea are highly relevant to understanding FECD disease progression at the clinical, cellular, and molecular levels. Our comparative analysis of GAR against Scheimpflug image evidence for subclinical corneal edema indicated that almost all modified Krachmer FECD grade 4 disease eyes and over 50% of grade 5 disease eyes do not have evidence for subclinical edema. These findings indicate that tracking changes in GAR may detect individuals who are progressing before they develop subclinical edema and therefore represent an opportunity to detect and potentially treat medically an important population at risk for developing corneal edema and symptoms that drive surgery other than edema that may not be detected without reporting of guttae impact and evolution.
Limitations of this study include the small area of image analysis inherent with specular microscopy, which may affect GAR calculations more with older microscopes than newer devices, which tend to have larger image areas. Limited sample size represents an additional limitation, although the size of this study was determined from a prospective pilot study and was designed to reach 90% power. Application of this software for future use in clinical settings may need to be expanded for increased validation in more mild disease stages with lower participant numbers in this study (e.g., mKg 2 and 3). Additionally, while this study was conducted at a single tertiary care center, the population was largely homogeneous, consisting predominantly of Caucasian females, in line with other major studies of FECD.44 Future applications utilizing this software tool could include multicenter studies to assess the relationship between GAR and mKg and integration of corneal endothelial cell morphometric data to better define disease progression risk. This automated software platform may also be used to assess treatment monitoring in clinical trials and natural history progression analysis by FECD genotype. Software advancements, such as a tracking feature that permits individual patient disease progression analysis over time (Fig. 8) and deep learning training of expanded datasets to further improve the accuracy of automated guttae segmentation, will require additional testing. ICC of repeat images (Fig. 7) indicates good to excellent repeatability of GAR among repeat images from the same eye within a session, further supporting future work tracking progression over time. Modified Krachmer grade 6 images were excluded from this study due to anticipated limitations of applying specular microscopy analytical tools in eyes with clinically detectable corneal edema. Indeed, the presence of corneal edema resulted in uneven phase contrast that did not permit reporting GAR in high fidelity with expert grading, in line with previous limitations noted in confocal imaging of the corneal endothelium.23 Such image variability in high disease burden highlights the importance of the ongoing integration of deep learning segmentation to quantify disease in advanced FECD with clinical edema, although the utility of this tool for central image analysis may be limited by disease severity.
Figure 8.
Representative patient (mKg 4) specular microscopy images used to calculate the change in GAR, ΔGAR. (A) Patient image captured at baseline, automatic GAR calculated at 7.58%. (B) Patient image captured 71 days later, automatic GAR calculated at 14.04%. (C) Aligned image ΔGAR analysis demonstrating a calculated ΔGAR of 6.06% over a period of 71 days.
In conclusion, in this prospective study, we validated a software analysis tool that can be used to report GAR in FECD specular microscopy images over a wide range of disease severity. Automated results were statistically equivalent compared to manually measured guttae. Increased GAR was associated with increased modified Krachmer grade and central image location. Findings support the use of this automated software platform to track guttae expansion as a marker of FECD disease progression in research and ultimately clinical applications.
Acknowledgments
The authors thank the Robert and Joell Brightfelt Professorship in Cornea Research, Lloyd and Betty Schermer, M.D. Wagoner & M.A. Greiner Cornea Excellence Fund, Cornea Transplant Research Fund, Beulah and Florence Usher Chair in Cornea/External Disease and Refractive Surgery, Research to Prevent Blindness, and our patients.
Supported by a departmental grant from Research to Prevent Blindness.
Disclosure: S.R. Sarin, None; J. Abou Assi, ADCIS Evolucare (E, P); J.M. Skeie, None; P. Deman, ADCIS Evolucare (E, P); M.B. Zimmerman, None; D.M. McGranahan, None; J.M. Troyer, None; G.A. Schmidt, None; F. Potevin, ADCIS Evolucare (E, P); M.A. Greiner, None
References
- 1. EBAA. 2024 Eye Banking Statistical Report. Chantilly, VA; 2025: 134. [Google Scholar]
- 2. Wilson SE, Bourne WM.. Fuchs’ dystrophy. Cornea. 1988; 7(1): 2–18. [PubMed] [Google Scholar]
- 3. Altamirano F, Ortiz-Morales G, O'Connor-Cordova MA, Sancen-Herrera JP, Zavala J, Valdez-Garcia JE. Fuchs endothelial corneal dystrophy: an updated review. Int Ophthalmol. 2024; 44(1): 61. [DOI] [PubMed] [Google Scholar]
- 4. Krachmer JH, Purcell JJ Jr, Young CW, Bucher KD. Corneal endothelial dystrophy. A study of 64 families. Arch Ophthalmol. 1978; 96(11): 2036–2039. [DOI] [PubMed] [Google Scholar]
- 5. Koizumi N, Okumura N, Ueno M, Nakagawa H, Hamuro J, Kinoshita S.. Rho-associated kinase inhibitor eye drop treatment as a possible medical treatment for Fuchs corneal dystrophy. Cornea. 2013; 32(8): 1167–1170. [DOI] [PubMed] [Google Scholar]
- 6. Kinoshita S, Koizumi N, Ueno M, et al.. Injection of cultured cells with a ROCK inhibitor for bullous keratopathy. N Engl J Med. 2018; 378(11): 995–1003. [DOI] [PubMed] [Google Scholar]
- 7. Kinoshita S, Colby KA, Kruse FE.. A close look at the clinical efficacy of Rho-associated protein kinase inhibitor eye drops for Fuchs endothelial corneal dystrophy. Cornea. 2021; 40(10): 1225–1228. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Davies E, Jurkunas U, Pineda R II. Pilot study of corneal clearance with the use of a Rho-kinase inhibitor after descemetorhexis without endothelial keratoplasty for Fuchs endothelial corneal dystrophy. Cornea. 2021; 40(7): 899–902. [DOI] [PubMed] [Google Scholar]
- 9. Saha S, Skeie JM, Schmidt GA, et al.. TCF4 trinucleotide repeat expansions and UV irradiation increase susceptibility to ferroptosis in Fuchs endothelial corneal dystrophy. Redox Biol. 2024; 77: 103348. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Czugala M, Mykhaylyk O, Bohler P, et al.. Efficient and safe gene delivery to human corneal endothelium using magnetic nanoparticles. Nanomedicine (Lond). 2016; 11(14): 1787–1800. [DOI] [PubMed] [Google Scholar]
- 11. Tomioka Y, Ueno M, Yamamoto A, et al.. Guttae morphology after cultured corneal endothelial cell transplant in Fuchs endothelial corneal dystrophy. JAMA Ophthalmol. 2024; 142(9): 818–826. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Dirisamer M, Ham L, Dapena I, van Dijk K, Melles GR.. Descemet membrane endothelial transfer: “free-floating” donor Descemet implantation as a potential alternative to “keratoplasty.” Cornea. 2012; 31(2): 194–197. [DOI] [PubMed] [Google Scholar]
- 13. Van Meter W, Mathews P, Philippy B, DeMatteo J.. 2023 Eye Banking Statistical Report—executive summary. Eye Banking Corneal Transplantation. 2024; 3(4): e0034. [Google Scholar]
- 14. Schlotzer-Schrehardt U, Zenkel M, Strunz M, et al.. Potential functional restoration of corneal endothelial cells in Fuchs endothelial corneal dystrophy by ROCK inhibitor (ripasudil). Am J Ophthalmol. 2021; 224: 185–199. [DOI] [PubMed] [Google Scholar]
- 15. Sarin SR, Kigin M, Balk E, et al.. Functional assessment of FECD in the National Advanced Driving Simulator: initial study of nighttime glare and Scheimpflug imaging. Cornea. 2025; 45: 595–601. [DOI] [PubMed] [Google Scholar]
- 16. Watanabe A, Oie Y, Maeda S, et al.. Quantitative analysis of glare symptom in patients with Fuchs endothelial corneal dystrophy. Invest Ophthalmol Vis Sci. 2020; 61(7): 3332–3332. [Google Scholar]
- 17. Shah K, Eghrari AO, Vanner EA, O'Brien TP, Koo EH. Scheimpflug corneal densitometry values and severity of guttae in relation to visual acuity in Fuchs endothelial corneal dystrophy. Cornea. 2022; 41(6): 692–698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Wacker K, Grewing V, Fritz M, Bohringer D, Reinhard T. Morphological and optical determinants of visual disability in Fuchs endothelial corneal dystrophy. Cornea. 2020; 39(6): 726–731. [DOI] [PubMed] [Google Scholar]
- 19. Kai C, Oie Y, Nishida N, et al.. Associations between visual functions and severity gradings, corneal scatter, or higher-order aberrations in Fuchs endothelial corneal dystrophy. Invest Ophthalmol Vis Sci. 2024; 65(6): 15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Sun SY, Wacker K, Baratz KH, Patel SV.. Determining subclinical edema in Fuchs endothelial corneal dystrophy: revised classification using Scheimpflug tomography for preoperative assessment. Ophthalmology. 2019; 126(2): 195–204. [DOI] [PubMed] [Google Scholar]
- 21. Repp DJ, Hodge DO, Baratz KH, McLaren JW, Patel SV.. Fuchs’ endothelial corneal dystrophy: subjective grading versus objective grading based on the central-to-peripheral thickness ratio. Ophthalmology. 2013; 120(4): 687–694. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Okumura N, Yamada S, Nishikawa T, et al.. U-Net convolutional neural network for segmenting the corneal endothelium in a mouse model of Fuchs endothelial corneal dystrophy. Cornea. 2022; 41(7): 901–907. [DOI] [PubMed] [Google Scholar]
- 23. McLaren JW, Bachman LA, Kane KM, Patel SV.. Objective assessment of the corneal endothelium in Fuchs’ endothelial dystrophy. Invest Ophthalmol Vis Sci. 2014; 55(2): 1184–1190. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Sierra JS, Pineda J, Rueda D, et al.. Corneal endothelium assessment in specular microscopy images with Fuchs’ dystrophy via deep regression of signed distance maps. Biomed Opt Express. 2023; 14(1): 335–351. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Sierra JS, Pineda J, Rueda D, et al.. Corneal endothelium assessment in specular microscopy images with Fuchs’ dystrophy via deep regression of signed distance maps. Biomed Opt Express. 2022; 14(1): 335–351. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Prada AM, Quintero F, Mendoza K, et al.. Assessing Fuchs corneal endothelial dystrophy using artificial intelligence–derived morphometric parameters from specular microscopy images. Cornea. 2022:10: 1097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Louttit MD, Kopplin LJ, Igo RP Jr, et al.. A multicenter study to map genes for Fuchs endothelial corneal dystrophy: baseline characteristics and heritability. Cornea. 2012; 31(1): 26–35. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Chiou AG, Kaufman SC, Beuerman RW, Ohta T, Soliman H, Kaufman HE.. Confocal microscopy in cornea guttata and Fuchs’ endothelial dystrophy. Br J Ophthalmol. 1999; 83(2): 185–189. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Ong Tone S, Kocaba V, Bohm M, Wylegala A, White TL, Jurkunas UV.. Fuchs endothelial corneal dystrophy: the vicious cycle of Fuchs pathogenesis. Prog Retin Eye Res. 2021; 80: 100863. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Huang J, Maram J, Tepelus TC, et al.. Comparison of manual & automated analysis methods for corneal endothelial cell density measurements by specular microscopy. J Optom. 2018; 11(3): 182–191. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Huang J, Maram J, Tepelus TC, Sadda SR, Chopra V, Lee OL.. Comparison of noncontact specular and confocal microscopy for evaluation of corneal endothelium. Eye Contact Lens. 2018; 44(suppl 1): S144–S150. [DOI] [PubMed] [Google Scholar]
- 32. Fujimoto H, Maeda N, Soma T, et al.. Quantitative regional differences in corneal endothelial abnormalities in the central and peripheral zones in Fuchs’ endothelial corneal dystrophy. Invest Ophthalmol Vis Sci. 2014; 55(8): 5090–5098. [DOI] [PubMed] [Google Scholar]
- 33. Soh YQ, Peh GS, Naso SL, Kocaba V, Mehta JS.. Automated clinical assessment of corneal guttae in Fuchs endothelial corneal dystrophy. Am J Ophthalmol. 2021; 221: 260–272. [DOI] [PubMed] [Google Scholar]
- 34. Eghrari AO, Mumtaz AA, Garrett B, et al.. Automated retroillumination photography analysis for objective assessment of Fuchs corneal dystrophy. Cornea. 2017; 36(1): 44–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Gottsch JD, Sundin OH, Rencs EV, et al.. Analysis and documentation of progression of Fuchs corneal dystrophy with retroillumination photography. Cornea. 2006; 25(4): 485–489. [DOI] [PubMed] [Google Scholar]
- 36. Soh YQ, Peh GSL, Naso SL, Kocaba V, Mehta JS.. Automated clinical assessment of corneal guttae in Fuchs endothelial corneal dystrophy. Am J Ophthalmol. 2021; 221: 260–272. [DOI] [PubMed] [Google Scholar]
- 37. Kocaba V, Katikireddy KR, Gipson I, Price MO, Price FW, Jurkunas UV.. Association of the gutta-induced microenvironment with corneal endothelial cell behavior and demise in Fuchs endothelial corneal dystrophy. JAMA Ophthalmol. 2018; 136(8): 886–892. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Oie Y, Yamaguchi T, Nishida N, et al.. Systematic review of the diagnostic criteria and severity classification for Fuchs endothelial corneal dystrophy. Cornea. 2023; 42(12): 1590–1600. [DOI] [PubMed] [Google Scholar]
- 39. Patel SV, Gupta N, Bhogal M, et al.. Delphi-based global consensus on Fuchs endothelial corneal dystrophy. An Endothelial Keratoplasty Learners Group initiative. Am J Ophthalmol. 2025; 280: 130–143. [DOI] [PubMed] [Google Scholar]
- 40. Yesilirmak N, Souedan V, Pison A, Bourges JL.. Predicting corneal decompensation in Fuchs endothelial corneal dystrophy with Scheimpflug tomography and clinical parameters. Indian J Ophthalmol. 2025; 73(1): 52–58. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Patel SV. Towards clinical trials in Fuchs endothelial corneal dystrophy: classification and outcome measures—the Bowman Club Lecture 2019. BMJ Open Ophthalmol. 2019; 4(1): e000321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Kocaba V, Katikireddy KR, Gipson I, MO Price, Price FW, Jurkunas UV.. Association of the gutta-induced microenvironment with corneal endothelial cell behavior and demise in Fuchs endothelial corneal dystrophy. JAMA Ophthalmol. 2018; 136(8): 886–892. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Hartness EM, Shevalye H, Skeie JM, et al.. Iron-sulfur clusters and iron responsive element binding proteins mediate iron accumulation in corneal endothelial cells in Fuchs dystrophy. Invest Ophthalmol Vis Sci. 2025; 66(4): 23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44. Price MO, Szczotka-Flynn LB, Bauza CE, et al.. Diabetes endothelial keratoplasty study: methods and impact on the use of corneas from donors with diabetes for Descemet membrane endothelial keratoplasty. Cornea. 2026; 45(3): 312–321. [DOI] [PMC free article] [PubMed] [Google Scholar]