Abstract
Objective: To compare pain scores and complication rates between a novel disposable intrauterine infusion catheter (Yikai Medical) and a conventional balloon-tipped 12F double-lumen catheter during hysterosalpingo-contrast sonography (HyCoSy). Methods: This single-center retrospective cohort study included 128 infertile women who underwent HyCoSy between January 2022 and June 2023. Patients were divided into observation group (disposable catheter, n = 64) and control group (balloon catheter, n = 64). Primary outcomes were Visual Analog Scale (VAS) pain scores (intraoperative, 30 min, 2 h postoperative) and complication rates. Secondary outcomes included procedure time and imaging success rate. Multivariate logistic regression was used to identify independent predictors of pain. Results: Baseline characteristics were generally comparable between the groups, although patients in the observation group were slightly younger (P = 0.041). VAS scores were significantly lower in the observation group at all time points (all P<0.001). Overall complication rate was lower in the observation group (15.6% vs. 29.7%) but not statistically significant (P = 0.091). Procedure time was shorter in the observation group (8.28 ± 1.44 vs. 10.49 ± 2.32 min, P<0.001). Imaging success was similar (100% vs. 98.4%, P = 1.000). Catheter type was an independent predictor of moderate-to-severe pain (OR = 0.115, 95% CI 0.050-0.267, P<0.001). Conclusion: The disposable intrauterine infusion catheter significantly reduces pain and procedure time during HyCoSy compared to the conventional balloon-tipped catheter, with comparable imaging success and a trend toward fewer complications.
Keywords: Hysterosalpingo-contrast sonography, pain, complications, catheter, retrospective study
Introduction
Hysterosalpingo-contrast sonography (HyCoSy) is a non-invasive imaging modality that assesses fallopian tube patency by instilling an ultrasound contrast agent into the uterine cavity and dynamically observing its flow through the tubes in real time [1]. Compared with traditional X-ray hysterosalpingography using iodinated oil, HyCoSy offers advantages such as the absence of ionizing radiation, relative procedural simplicity, and good reproducibility [2]. It has thus been widely adopted for the initial assessment of tubal function in infertile patients and is currently considered one of the first-line clinical diagnostic modalities [3].
In HyCoSy, the catheter is essential for effective contrast delivery, and its design significantly affects procedural workflow, patient experience, and complication rates [4]. Currently, the most commonly used device in clinical practice is the conventional balloon-tipped 12F double-lumen catheter [5]. This catheter is fixed at the internal cervical os via balloon inflation to prevent contrast leakage and ensure adequate intrauterine pressure for imaging [6,7]. However, the process of balloon inflation itself exerts mechanical dilation and pressure stimulation on the internal os and lower uterine segment [8]. This often triggers significant lower abdominal distension and colicky pain; in some cases, patients are even forced to discontinue the examination due to pain intolerance. Furthermore, mechanical friction between the balloon and the cervical mucosa may lead to mucosal injury and prolonged vaginal bleeding postoperatively [9,10]. Although less frequent, complications such as balloon rupture and vasovagal reflex also occur sporadically. Collectively, these factors adversely affect the patient experience and postoperative recovery.
In recent years, driven by advancements in material science and device engineering, a novel single-use intrauterine infusion and hysterosalpingography catheter (e.g., produced by Yikai Medical) has been introduced into clinical practice [11,12]. This catheter eliminates the balloon design and instead optimizes the flexibility of the catheter body and the morphology of the tip [13]. The aim is to maintain effective contrast infusion while reducing mechanical irritation to the cervix and uterine cavity [14,15]. Theoretically, this balloon-free design is expected to lower the risk of procedure-related pain and complications. However, high-quality evidence supporting its clinical superiority remains relatively limited, particularly regarding systematic comparative data - either retrospective or prospective - against traditional catheters.
Pain management is a key factor affecting patient tolerance and compliance with HyCoSy. Pain not only imposes physiological and psychological burdens but may also compromise image quality and diagnostic accuracy due to patient movement or tension [16,17]. Therefore, selecting a catheter that ensures imaging success while enhancing patient comfort is of clear clinical significance. Simultaneously, reducing procedure-related complications is a crucial factor in improving medical safety and patient satisfaction [18-20].
Based on the background outlined above, this study aims to systematically compare the clinical performance of the single-use intrauterine infusion catheter and the conventional balloon-tipped 12F double-lumen catheter in HyCoSy through a retrospective cohort study. We focus specifically on analyzing differences in procedure-related pain scores (intraoperative and short-term postoperative) and complication rates between the two groups, with the goal of providing evidence-based guidance for the rational selection of catheters in clinical practice. We hypothesize that, compared with the traditional balloon catheter, the single-use intrauterine infusion catheter will significantly alleviate patient pain and reduce the incidence of related complications.
Materials and methods
Study design
This was a single-center, retrospective cohort study designed to compare the clinical performance of a single-use intrauterine infusion catheter with that of the conventional balloon-tipped 12F double-lumen catheter in hysterosalpingo-contrast sonography (HyCoSy). The study protocol was reviewed and approved by the Ethics Committee of Jingmen Hospital of Traditional Chinese Medicine. All data were extracted from previously collected clinical records and anonymized to protect patient privacy. All procedures involving human participants were conducted in accordance with the ethical standards of the Declaration of Helsinki (as revised in 2013).
Study participants
Clinical data of infertile patients who underwent HyCoSy at the Reproductive Medicine Center of our hospital between January 2022 and June 2023 were retrospectively collected via the hospital’s electronic medical record system. All participants met the following inclusion criteria: (1) age between 20 and 45 years; (2) diagnosis of infertility; (3) regular menstrual cycles (28-35 days); and (4) no history of uterine cavity procedures within 3 months prior to the examination. The following exclusion criteria were applied: (1) severe dysfunction of the heart, liver, or kidneys; (2) known allergy to ultrasound contrast agents; (3) coagulation disorders; (4) acute genital tract infections, such as acute pelvic inflammatory disease or vaginitis, at the time of examination; or (5) confirmed intrauterine pregnancy at the time of examination. Ultimately, all cases included in the analysis possessed complete clinical records.
Sample size description
This study employed a retrospective cohort design; therefore, no prospective sample size calculation was performed. Through systematic screening of medical records, a total of 128 patients who met the eligibility criteria were identified. These patients were naturally divided into an observation group and a control group based on the actual catheter type used, with 64 patients in each group. This sample size is consistent with the typical scale for comparative studies involving two independent groups. Furthermore, referencing previous similar studies and pilot data, this cohort possesses sufficient statistical power to detect intergroup differences in the primary outcome (pain score) (Figure 1).
Figure 1.
Patient screening flowchart.
Grouping and catheter types
Patients were divided into two groups according to the type of catheter used: Observation group: A single-use intrauterine infusion and hysterosalpingography catheter (Yikai Medical, Model: GK-3.5, Specification: 3.5Fr, Length: 35 cm) was utilized.
Control group: A traditional balloon-tipped 12F double-lumen catheter (Cook Medical, Model: J-ASSC-155000, Specification: 12Fr, Balloon capacity: 1.5-3.0 mL) was utilized.
Examination equipment and contrast procedure
All examinations were performed using standardized equipment and protocols: Ultrasound equipment: GE Voluson E8 ultrasound system equipped with a RIC5-9-D transvaginal 3D volumetric probe. Contrast agent: SonoVue (Bracco, Italy), prepared as a suspension according to the manufacturer’s instructions. Perfusion system: ACIST CVi power injector, set at a constant flow rate of 3-5 mL/min. Operators: All examinations were performed by the same team of physicians with over 5 years of experience in HyCoSy procedures.
The examination procedure consisted of the following steps: The patient was placed in the lithotomy position, followed by routine disinfection and draping. Catheter insertion under ultrasound guidance (In the observation group, the balloon was not inflated; in the control group, the balloon was inflated with 1.5-2.0 mL of normal saline). Infusion of the contrast suspension (5 mL SonoVue + 15 mL normal saline) through the catheter. Real-time ultrasound monitoring of the flow of the contrast agent within the uterine cavity and fallopian tubes. Catheter removal after the examination; patients were discharged after 30 minutes of observation with no adverse events.
Outcome measures and data extraction
The following information was extracted from the medical records: Primary outcome measures: Pain score: Assessed using the Visual Analog Scale (VAS, 0-10), recorded at three time points - immediately during the procedure, 30 minutes post-procedure, and 2 hours post-procedure. Complications: Including abdominal pain (>2 hours duration), vaginal bleeding (>3 days duration), infection (body temperature >38°C), vasovagal reaction, and uterine perforation.
Secondary outcome measures: Procedure time: Duration from catheter insertion to the end of imaging. Imaging success rate: Clear visualization of bilateral fallopian tubes; first-attempt catheterization success rate. Contrast reflux: Incidence of contrast agent reflux.
Statistical analysis
All statistical analyses were performed using SPSS software (version 26.0). Continuous variables were tested for normality using the Shapiro-Wilk test. Normally distributed data were presented as mean ± standard deviation (SD) and were compared between groups using the independent samples t-test. Non-normally distributed data were presented as median (interquartile range) and were compared using the Mann-Whitney U test. Categorical variables were expressed as number (percentage) and were compared using the chi-square test or Fisher’s exact test when the expected frequency was <5. To identify independent factors influencing intraoperative pain, multivariate logistic regression analysis was performed. The dependent variable was the occurrence of moderate-to-severe pain (Visual Analog Scale [VAS] ≥4), and the independent variables included catheter type (observation vs. control), age, duration of infertility, and history of pelvic surgery. Odds ratios (OR) and their 95% confidence intervals (CI) were calculated. The Hosmer-Lemeshow test was used to assess the goodness-of-fit of the model. Furthermore, to describe the trend of pain scores over time, repeated measures analysis of variance (ANOVA) was conducted to compare VAS scores at the three time points (intraoperative, 30 minutes postoperative, and 2 hours postoperative) between the two groups, and a trend graph was plotted. All hypothesis tests were two-sided, and a P-value <0.05 was considered statistically significant.
Results
Baseline characteristics
A total of 128 patients were included in this study, with 64 assigned to the observation group and 64 to the control group. There were no statistically significant differences between the two groups in terms of duration of infertility, proportion of primary infertility, history of pelvic surgery, or uterine position (all P>0.05), indicating good baseline comparability. The only significant difference was in mean age: patients in the observation group (29.36 ± 3.74 years) were slightly younger than those in the control group (30.73 ± 3.78 years; t = -2.069, P = 0.041). Detailed baseline characteristics are presented in Table 1.
Table 1.
Baseline characteristics
| Variable | Observation Group (n = 64) | Control Group (n = 64) | t/χ2 | P value |
|---|---|---|---|---|
| Age (years) | 29.36 ± 3.74 | 30.73 ± 3.78 | -2.069 | 0.041 |
| Duration of infertility (years) | 3.27 ± 1.67 | 3.59 ± 2.16 | -0.933 | 0.353 |
| Primary infertility (n, %) | 29 (45.3%) | 32 (50.0%) | 0.125 | 0.723 |
| History of pelvic surgery (n, %) | 21 (32.8%) | 21 (32.8%) | 0.000 | 1.000 |
| Uterus position (n, %) | 0.144 | 0.931 | ||
| Anterior | 32 (50.0%) | 30 (46.9%) | ||
| Middle | 24 (37.5%) | 25 (39.1%) | ||
| Posterior | 8 (12.5%) | 9 (14.1%) |
Procedure-related outcomes
The mean procedural time was significantly shorter in the observation group (8.28 ± 1.44 minutes) compared with the control group (10.49 ± 2.32 minutes; t = -6.454, P<0.001) (Figure 2). The first-attempt catheter insertion success rate was 100% in both groups. Imaging success rates were 100.0% (64/64) in the observation group and 98.4% (63/64) in the control group, with no statistically significant difference (χ2 = 0.000, P = 1.000) (Table 2).
Figure 2.
Procedure time comparison.
Table 2.
Procedure-related outcomes
| Variable | Observation Group (n = 64) | Control Group (n = 64) | t/χ2 | P value |
|---|---|---|---|---|
| Procedure time (min) | 8.28 ± 1.44 | 10.49 ± 2.32 | -6.454 | <0.001 |
| Successful imaging (n, %) | 64 (100.0%) | 63 (98.4%) | 0.000 | 1.000 |
Comparison of pain scores
Pain scores assessed by the Visual Analog Scale (VAS) were significantly lower in the observation group than those in the control group at all three time points - intraoperatively, 30 minutes post-procedure, and 2 hours post-procedure (all P<0.001) (Figure 3). Specifically, VAS scores in the observation group were 2.92 ± 1.03, 1.78 ± 0.83, and 0.89 ± 0.56, respectively, compared with 4.63 ± 1.57, 3.27 ± 1.25, and 1.69 ± 0.86 in the control group. The trend analysis of pain scores over time demonstrated more rapid and pronounced pain relief in the observation group (Figure 4).
Figure 3.
Pain scores at different time points. VAS, Visual Analog Scale.
Figure 4.
Pain score changes over time. VAS, Visual Analog Scale.
Comparison of complication rates
The overall complication rate was 15.6% (10/64) in the observation group, compared with 29.7% (19/64) in the control group; however, this difference did not reach statistical significance (χ2 = 2.853, P = 0.091). Regarding specific complications, the incidence of abdominal pain (>2 hours) (7.8% vs. 17.2%) and prolonged vaginal bleeding (>3 days) (4.7% vs. 9.4%) was lower in the observation group, but neither difference was statistically significant (abdominal pain: χ2 = 1.786, P = 0.181; vaginal bleeding: Fisher’s exact test, P = 0.492). Rates of infection, vasovagal reactions, and uterine perforation were low in both groups and showed no statistically significant differences (Table 3).
Table 3.
Complication rates
| Complication | Observation Group (n, %) | Control Group (n, %) | χ2 | P value |
|---|---|---|---|---|
| Abdominal pain | 5 (7.8%) | 11 (17.2%) | 1.786 | 0.181 |
| Vaginal bleeding >3 days | 3 (4.7%) | 6 (9.4%) | - | 0.492* |
| Infection | 2 (3.1%) | 1 (1.6%) | - | 1.000* |
| Vagal response | 0 (0.0%) | 2 (3.1%) | - | 0.496* |
| Uterine perforation | 0 (0.0%) | 1 (1.6%) | - | 1.000* |
| Total complications | 10 (15.6%) | 19 (29.7%) | 2.853 | 0.091 |
indicates the use of Fisher’s exact test.
Multivariate regression analysis of factors associated with pain
Multivariate logistic regression analysis revealed that, after adjusting for age, duration of infertility, and history of pelvic surgery, catheter type was an independent predictor of moderate-to-severe intraoperative pain (defined as VAS ≥4). Patients in the observation group (using the single-use intrauterine infusion catheter) had a significantly lower risk of moderate-to-severe pain compared with those in the control group (using the traditional balloon-tipped double-lumen catheter) (β = -2.163, OR = 0.115, 95% CI: 0.050-0.267, P<0.001). In contrast, age, duration of infertility, and pelvic surgery history were not significantly associated with pain scores (Table 4).
Table 4.
Multivariate logistic regression analysis
| Variable | β coefficient | Odds ratio (95% CI) | P value |
|---|---|---|---|
| Catheter type (Observation vs. Control) | -2.163 | 0.115 (0.050-0.267) | <0.001 |
| Age | 0.074 | 1.077 (0.963-1.204) | 0.196 |
| Duration of infertility | 0.015 | 1.015 (0.819-1.257) | 0.893 |
| History of pelvic surgery | 0.256 | 1.291 (0.529-3.150) | 0.574 |
| Constant | -1.742 | 0.175 (0.006-5.540) | 0.323 |
Discussion
This study employed a retrospective cohort design to systematically compare the clinical performance of a novel single-use intrauterine infusion catheter with that of the conventional balloon-tipped double-lumen catheter in hysterosalpingo-contrast sonography (HyCoSy). The main findings can be summarized as follows: compared with the traditional balloon catheter, the use of the single-use intrauterine infusion catheter significantly reduced pain scores during and shortly after the procedure, shortened procedural time, and showed a trend toward lower overall complication rates while maintaining comparable imaging success.
In our study, VAS scores in the observation group were significantly lower than those in the control group at all three time points - intraoperatively, 30 minutes post-procedure, and 2 hours post-procedure - consistent with our hypothesis and suggesting a clear advantage of the balloon-free design in alleviating patient discomfort. This benefit may be explained by multiple mechanisms. First, from a mechanical perspective, the conventional catheter relies on balloon inflation (typically with 1.5-3.0 mL of saline) to anchor at the internal cervical os, which directly exerts traction and compression on the densely innervated cervical tissue, often triggering spastic pain [21]. Second, from a hydrodynamic standpoint, the presence of the balloon may distort normal uterine cavity morphology, leading to uneven contrast flow and localized pressure elevation, thereby exacerbating pain perception [22]. Notably, vagal reactions occurred in 3.1% of patients in the control group but were absent in the observation group, further supporting the hypothesis that balloon-induced stimulation may activate pelvic autonomic reflexes [23]. Moreover, multivariate logistic regression confirmed catheter type as an independent predictor of moderate-to-severe pain (OR = 0.115, P<0.001), underscoring the pivotal role of catheter design in pain management [24].
Although the difference in overall complication rates between the two groups (15.6% in the observation group vs. 29.7% in the control group) did not reach statistical significance (P = 0.091), the observation group showed consistently lower incidences of common complications - specifically abdominal pain (7.8% vs. 17.2%) and prolonged vaginal bleeding (>3 days; 4.7% vs. 9.4%). This trend is likely attributable to the balloon-free design, which avoids mechanical friction against the cervical mucosa and excessive stimulation of the myometrium [25]. Regarding procedural efficiency, the mean operation time in the observation group was approximately 2.2 minutes shorter than that in the control group, primarily due to the elimination of balloon inflation and deflation steps, thereby streamlining the workflow. This reduction not only enhances clinical throughput but may also indirectly improve patient experience by minimizing exposure time to anxiety and discomfort [26].
Our findings align with existing literature demonstrating the advantages of non-balloon catheters. For instance, Xu et al. [27] demonstrated in a prospective study that non-balloon catheters significantly reduced pain during intrauterine procedures. A small-sample domestic study also suggested a similar trend [28]. The contribution of our study lies in providing more robust retrospective evidence through a larger sample size (n = 64 per group), systematic multi-timepoint pain assessments, and comprehensive documentation of various complications. Although the overall complication rate was lower in the observation group, the difference did not reach statistical significance, possibly due to insufficient sample size to detect rare events (Type II error). Future prospective studies with larger cohorts are warranted to confirm these trends.
Strengths of this study include the use of standardized procedural protocols to minimize operator bias, multivariate adjustment for potential confounders such as age and duration of infertility, and dynamic evaluation of pain over time. Nevertheless, inherent limitations of a retrospective design must be acknowledged. First, despite statistical adjustments, residual confounding or selection bias cannot be entirely ruled out. Second, data were derived from a single center, potentially limiting generalizability. Third, pain assessment relied solely on subjective VAS scores; future studies could incorporate objective physiological indicators (e.g., heart rate variability) for validation [8]. Finally, long-term outcomes (e.g., subsequent pregnancy rates) and health economic analyses were not evaluated.
Based on our findings, we propose that the single-use intrauterine infusion catheter represents a valuable tool for optimizing the HyCoSy experience - particularly beneficial for patients with heightened pain sensitivity, those with special cervical conditions (e.g., stenosis or incompetence), or high-volume outpatient settings where procedural efficiency is critical [29]. Looking ahead, we recommend conducting multicenter, prospective, randomized controlled trials to generate higher-level evidence and further explore the utility of this catheter in specific populations (e.g., patients with uterine malformations) and its long-term clinical impact.
Conclusion
In summary, in hysterosalpingo-contrast sonography (HyCoSy), the use of a single-use intrauterine infusion catheter, compared with the traditional balloon-tipped double-lumen catheter, significantly reduces patient pain, improves procedural efficiency, and maintains high imaging success rates with a favorable safety profile. This approach offers a new evidence-based option for enhancing patient comfort and optimizing clinical workflows in routine practice.
Disclosure of conflict of interest
None.
References
- 1.Dreisler E, Kjer JJ. Asherman’s syndrome: current perspectives on diagnosis and management. Int J Womens Health. 2019;11:191–198. doi: 10.2147/IJWH.S165474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Silvinato A, Bernardo WM. Hysterosalpingography: balloon catheter or metal cannula? Rev Assoc Med Bras (1992) 2020;66:589. doi: 10.1590/1806-9282.66.5.589. [DOI] [PubMed] [Google Scholar]
- 3.Kılıç KK, Gürses C, Karadağ C, Sözel YK, Özdemir Ö. When a balloon catheter or tenaculum is required for cervical traction during hysterosalpingography. J Obstet Gynaecol. 2023;43:2171777. doi: 10.1080/01443615.2023.2171777. [DOI] [PubMed] [Google Scholar]
- 4.Youssef Y, Raju R, Sanya R, Abuzeid O, Hebert J, Abuzeid M. Perfecting hysterosalpingography: limitations to the use of disposable balloon catheter and its solution. Int J Gynaecol Obstet. 2022;158:736–741. doi: 10.1002/ijgo.14118. [DOI] [PubMed] [Google Scholar]
- 5.Setúbal A, Alves J, Osório F, Sidiropoulou Z. Demonstration of isthmocele surgical repair. J Minim Invasive Gynecol. 2021;28:389–390. doi: 10.1016/j.jmig.2020.09.007. [DOI] [PubMed] [Google Scholar]
- 6.Li Y, Chen X, Wang Q, Wang J. Factors influencing the visualization of fallopian tubes in hysterosalpingo-contrast sonography (HyCoSy)-the value of multimodal HyCoSy in visualizing the fallopian tubes. J Ultrasound Med. 2024;43:1957–1967. doi: 10.1002/jum.16530. [DOI] [PubMed] [Google Scholar]
- 7.Ahmadi F, Jahangiri N, Zafarani F, Vosough A. Pain perception and side effects during saline infusion sonohysterography with a balloon catheter: a randomized comparative study of cervical versus intrauterine catheter placement. J Ultrasound Med. 2020;39:1829–1837. doi: 10.1002/jum.15292. [DOI] [PubMed] [Google Scholar]
- 8.Ma Y, Du L, Li F, Zhu Z. Use of atropine in four-dimensional hysterosalpingo-contrast sonography: does it suppress pain during infertility examination? Clin Exp Pharmacol Physiol. 2018;45:1334–1340. doi: 10.1111/1440-1681.12919. [DOI] [PubMed] [Google Scholar]
- 9.Maheux-Lacroix S, Li F, Laberge PY, Abbott J. Imaging for polyps and leiomyomas in women with abnormal uterine bleeding: a systematic review. Obstet Gynecol. 2016;128:1425–1436. doi: 10.1097/AOG.0000000000001776. [DOI] [PubMed] [Google Scholar]
- 10.Elsayed M, Amer MI, Gamal R, Haggag A, Mashaly AR, Absi TE, Elsayed W. Measuring uterine cavity volume with sonohysterography: a new objective method. Ultrasound. 2024;32:94–101. doi: 10.1177/1742271X231215502. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Jin BB, Ma Y, Zhao XH, Teng YL, Zhu SY. Operator’s influence on contrast agent intravasation during hysterosalpingo-contrast sonography: explanation based on a physics model. Int J Gen Med. 2022;15:7709–7718. doi: 10.2147/IJGM.S380917. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Matsubayashi H, Takaya Y, Takeuchi T, Doshida M, Ohara Y, Ishikawa T. Falloposcopic tuboplasty: an easy, quick, and safe technique. Fertil Steril. 2021;116:1669–1672. doi: 10.1016/j.fertnstert.2021.08.019. [DOI] [PubMed] [Google Scholar]
- 13.Ludwin A, Ludwin I, Bhagavath B, Lindheim SR. Pre-, intra-, and postoperative management of Robert’s uterus. Fertil Steril. 2018;110:778–779. doi: 10.1016/j.fertnstert.2018.05.033. [DOI] [PubMed] [Google Scholar]
- 14.Pace C, Argirò R, Casadei L, Cesareni M, Orlacchio A. Comparison between X-ray-hysterosalpingography and 3 Tesla magnetic resonance-hysterosalpingography in the assessment of the tubal patency in the cause of female infertility. Radiol Med. 2022;127:1373–1382. doi: 10.1007/s11547-022-01556-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Javaheri A, Kianfar K, Pourmasumi S, Eftekhar M. Platelet-rich plasma in the management of Asherman’s syndrome: an RCT. Int J Reprod Biomed. 2020;18:113–120. doi: 10.18502/ijrm.v18i2.6423. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Keltz M, Brown EC, Frishman GN, Sauerbrun-Cutler MT. Fluoroscopically-guided hysteroscopic tubal cannulation: a procedure for proximal tubal obstruction. JSLS. 2022;26:e2022.00047. doi: 10.4293/JSLS.2022.00047. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Sabry ASA, Fadl SA, Szmigielski W, Alobaidely A, Ahmed SSH, Sherif H, R H Yousef R, Mahfouz A. Diagnostic value of three-dimensional saline infusion sonohysterography in the evaluation of the uterus and uterine cavity lesions. Pol J Radiol. 2018;83:e482–e490. doi: 10.5114/pjr.2018.80132. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Saini A, Ghuman NK, Raikar S, Bhati M, Singh P. Comparison of pain score between Foleys catheter and Leich-Wilkinson cannula in hysterosalpingography by numerical rating scale - a randomized control trial. Health Care Women Int. 2023;44:487–495. doi: 10.1080/07399332.2022.2134389. [DOI] [PubMed] [Google Scholar]
- 19.Xie H, Zhang X, Tian D, Liang D, He C, Mao H. Effect of acupoint application of Chinese medicine combined with a hot compress on pain of uterine and fallopian tube contrast agent in infertile women. Altern Ther Health Med. 2024;30:65–71. [PubMed] [Google Scholar]
- 20.Tanaka K, Chua J, Cincotta R, Ballard EL, Duncombe G. Hysterosalpingo-foam sonography (HyFoSy): tolerability, safety and the occurrence of pregnancy post-procedure. Aust N Z J Obstet Gynaecol. 2018;58:114–118. doi: 10.1111/ajo.12716. [DOI] [PubMed] [Google Scholar]
- 21.Li PC, Ding DC. The feasibility of reduced sutures in 3D laparoscopic tubal sterilization reversal. J Gynecol Obstet Hum Reprod. 2022;51:102473. doi: 10.1016/j.jogoh.2022.102473. [DOI] [PubMed] [Google Scholar]
- 22.Bar G, Harlev A, Alfayumi-Zeadna S, Zeadna A, Bord I, Har-Vardi I, Lunenfeld E, Levitas E. Recurrent implantation failure: which patients benefit from endometrial scratching prior to IVF? Arch Gynecol Obstet. 2020;301:817–822. doi: 10.1007/s00404-019-05424-1. [DOI] [PubMed] [Google Scholar]
- 23.Kwa M, Karabakhtsian RG, Cotzia P, Rotenberg O. Ultrasound-guided endometrial aspiration and immunohistochemistry for detecting focal metastatic cancer. J Clin Ultrasound. 2026;54:731–735. doi: 10.1002/jcu.70120. [DOI] [PubMed] [Google Scholar]
- 24.Cosmatos A, McCormick B, Blew B, Brown PA. Pre-peritoneal dialysis peritonitis after saline infusion sonohysterogram in a patient with an embedded catheter: a case report. Can J Kidney Health Dis. 2023;10:20543581231156854. doi: 10.1177/20543581231156854. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Ludwin I, Martins WP, Nastri CO, Ludwin A. Pain intensity during ultrasound assessment of uterine cavity and tubal patency with and without painkillers: prospective observational study. J Minim Invasive Gynecol. 2017;24:599–608. doi: 10.1016/j.jmig.2017.01.015. [DOI] [PubMed] [Google Scholar]
- 26.Keltz M, Rubin SC, Brown E, Bibi M, Sauerbrun-Cutler MT. Fluoroscopic-guided hysteroscopic tubal cannulation results in high technical success and pregnancy rates comparable with the more traditional laparoscopically guided hysteroscopic tubal cannulation. F S Rep. 2024;5:205–210. doi: 10.1016/j.xfre.2024.02.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Xu AX, Zhang DW, Wang M, Long JF. The analysis of factors influencing the success of diagnosing salpingemphraxis by x-ray hysterosalpingography. Eur Rev Med Pharmacol Sci. 2016;20:2764–2768. [PubMed] [Google Scholar]
- 28.Melcer Y, Pekar-Zlotin M, Youngster M, Gat I, Maymon R. Fallopian tube catheterization under 3D vaginal ultrasound guidance followed by highly selective hysterosalpingo-foam sonography: an outpatient procedure. Ultraschall Med. 2025;46:285–290. doi: 10.1055/a-2384-4254. [DOI] [PubMed] [Google Scholar]
- 29.Jin BB, Ma Y, Zhao XH, Teng YL, Zhu SY. Risk factors of contrast agent intravasation with sulfur hexafluoride microbubble during hysterosalpingo-contrast sonography. J Ultrasound Med. 2023;42:1587–1594. doi: 10.1002/jum.16179. [DOI] [PubMed] [Google Scholar]