Table 2.

General laboratory setup, North West Ethiopia, 2024

Malaria microscopy laboratory format
Variables Response category Percent of study participants
Presence of a standard laboratory request paper Yes 4(100%)
No 0
Availability of SOP Yes 1 (25%)
No 3 (75%)
Presence of guideline for reagent preparation Yes 2 (50%)
No 2 (50%)
Availability of Internal quality control (IQC)log sheet Yes 2 (50%)
No 2 (50%)
Maintenance logbook for microscope Yes 0
No 4 (100%)
Malaria microscopy skill assessment
type of blood film they use Thin 0
Thick 4 (100%)
Both 0
Blood smear drying system Heat 4 (100%)
Air 0
Number of fields of examined to report a negative result ≥ 100 1 (25%)
< 100 3 (75%)
Perform a parasite count Yes 0
No 4 (100%)
Reporting species & stages Yes 4 (100%)
No 0
number of malaria slides examined daily < 5 0
5–10 0
> 10 4 (100%)
Does the work load allow the recommended reading time Yes 0
No 4 (100%)
Information related to staining reagent & equipment
Method of diagnosis for malaria Microscope 3 (75%)
Microscope &RDT 1 (25%)
Labeling of stock solution Yes 3 (75%)
No 1 (25%)
Recommended Storage of staining solutions Yes 4 (100%)
No 0
How prepare/dilute working solution Tape water 3 (75%)
Distilled water 1 (25%)
Cleaning of Lense of microscope with recommended solution Yes 2 (50%)
No 2 (50%)
General Information on quality Assurance
Internal quality control programs conducted regularly Yes 0
No 4 (100%)
Slide rechecked when result released Yes 0
No 4 (100%)
Supervision from regional or national laboratories within a year Yes 3 (75%)
No 1 (25%)
Regular training program for microscopists available Yes 1 (25%)
No 3 (75%)
Participating in EQA program Yes 4 (100%)
No 0