Skip to main content
. 2026 Jun 4;21:606920. doi: 10.2147/IJN.S606920

Table 4.

Detection Results of Real Samples Using the SERS Platform and qRT-PCR

The Healthy (pM) LSIL (pM) HSIL (pM) Cancer (pM)
SERS miR103a 0.426 1.402 6.292 15.246
miR221 43.251 75.614 108.16 516.513
qRTPCR miR103a 0.397 1.538 6.649 14.531
miR221 41.148 71.917 104.07 530.74
Relative error [%] miR103a 7.3 −8.84 −5.37 4.92
miR221 5.11 5.14 3.93 −2.68