Abstract
Background:
Robotic-assisted total knee arthroplasty (RA-TKA) may disproportionately affect revision rates based on a surgeon’s revision profile. This national registry study (1) stratified surgeons by pre–RA-TKA revision rates as low, medium, or high and (2) compared the change in revision rates and modes of failure within each cohort after commencing RA-TKA.
Methods:
All TKA procedures performed for osteoarthritis and recorded by the Australian Orthopaedic Association National Joint Replacement Registry from September 1999 to December 2023 were screened for eligibility. A total of 178 surgeons were included, performing 85,075 TKAs since commencing RA-TKA. TKA with hinged, constrained or medial pivot bearings, and American Society of Anesthesiologists (ASA) Scores 4 and 5 were excluded. Surgeons were stratified into quartiles based on their 2-year cumulative-percent-revision (CPR) rates before commencing RA-TKA into low (2-year CPR ≤1%), middle (1%-2.5%), and high revision cohorts (≥2.5%). The primary outcome was CPR with comparisons between time-matched RA-TKA and non-RA-TKA for each surgeon cohort after commencing RA-TKA. Secondary outcomes were changes in failure modes. Cox proportional hazards modelling controlled for age, sex, body mass index, ASA, patellar resurfacing, surgical year, fixation, polyethylene, and bearing type.
Results:
After commencing RA-TKA, there was no difference in CPR between RA-TKA and non-RA-TKA in low (hazard ratio [HR] 1.27; 95% CI 0.84-1.91; p = 0.26) and middle-revision rate surgeon groups (HR 0.99; 95% CI 0.83-1.19; p = 0.95). High-revision rate surgeons had decreased three-month CPR (HR 0.51; 95% CI 0.34-0.75; p < 0.001) but increased >9-month CPR (HR 2.28; 95% CI 1.74-2.98; p < 0.001). The mean follow-up was 2.4, 2.4 and 2.3 years, respectively, with maximum 7-year follow-up. Failure modes including infection, loosening, and instability did not change within any revision rate cohort after commencing RA-TKA (p > 0.05).
Conclusions:
At a grouped cohort level, commencing RA-TKA does not improve revision rates nor change failure modes. Surgeons with higher revision rates before commencing RA-TKA experienced further increases in revision rates.
Level of Evidence:
Level III. See Instructions for Authors for a complete description of levels of evidence.
Introduction
Robotic-assisted total knee arthroplasty (RA-TKA) is increasing1. In 2024, 41.9% of TKA in Australia were recorded as RA-TKA by the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR)2. Adoption has likely benefited from the support of orthopaedic device companies3, and the pressure surgeons feel to adopt robotics4. The technology is compelling to many, but nearly all studies that compare RA-TKA with conventional techniques involve financial conflicts5.
Despite clear theoretical and physical benefits of RA-TKA (accurate bone cuts, precision alignment, quantifiable soft tissue balancing and intra-operative feedback)6-16, clinical advantages of RA-TKA have neither been apparent in randomized controlled trials17,18 nor meta-analyses19,20. However, increased cost21, longer operative time,22,23 and unique complications have been reported24,25. Accordingly, the role of robotics in contemporary TKA is unclear.
Surgeons experience a range of revision rates and complication profiles26. Acknowledging that RA-TKA can standardize surgical technique and improve technical performance, could the real advantage of RA-TKA be for surgeons with higher revision rates? If only a subgroup of surgeons benefits from RA-TKA, its advantages may have been obscured in previous studies. If RA-TKA reduces the disparity in revision outcomes, then patients, surgeons and health care systems would all benefit to an extent that might prove cost effective27. This study, accordingly, compared the effect of RA-TKA on the revision rates and modes of failure of surgeons divided into 3 groups according to their revision rates before commencing RA-TKA.
Methods
Details of the AOANJRR data set, including its national coverage, data validation processes, and linkage methods, have been described28. In brief, the AOANJRR captures over 98% of arthroplasty procedures in Australia and links revision procedures to their corresponding primaries using a validated matching process.
All TKA procedures performed for osteoarthritis recorded by the AOANJRR from September 1999 to December 2023 were screened for eligibility. To mitigate learning curves and capture greater experience, surgeons and their cases were included if they performed ≥50 TKAs with a minimum 2-year follow-up both before and after adopting RA-TKA. Procedures with hinged, constrained, or medial pivot bearings and American Society of Anesthesiologists (ASA) Scores 4 and 5 were excluded. Patient, surgeon, and implants variables were collected before and after commencing RA-TKA. Robotic commencement was identified per surgeon using standard AOANJRR procedure forms that record technology use for every case. All procedures performed after a surgeon’s first recorded robotic case were eligible for inclusion in the study cohorts. Surgeons were stratified into low (2-year CPR ≤1%), middle (1%-2.5%), and high revision cohorts (≥2.5%) based on their 2-year cumulative-percent-revision (CPR) rate quartiles before commencing RA-TKA. This established surgeon groupings of comparable standards for subsequent study. This methodology is consistent with prior AOANJRR Research26,29. The 2-year CPR was used solely to establish stable baseline performance groups immediately before robotic-adoption, reflecting surgical complications that typically manifest early which are influenced by surgeon factors [22[. The subsequent comparison of RA-TKA versus non–RA-TKA procedures performed by the same surgeons after they commenced robotic use, stratified by their pre–RA-TKA revision profile was not restricted to 2 years but used the entire available follow-up. Prerobotic data (1999-2023) were used solely for surgeon stratification; comparative analyses included only procedures performed from 2016 onward, with follow-up to 31 December 2023. Surgeons were not restratified after commencing robotic-assistance; the prerobotic cohorts were retained for all subsequent analyses.
Having established groups with surgeons of similar revision profiles, the primary outcome was within-group difference in all-cause revision for the 3 surgeon groups when comparing RA-TKA with non–RA-TKA after the commencement of RA-TKA. To minimize confounding, these results were adjusted for the effects of age, sex, ASA Score, Body Mass Index (BMI), patella resurfacing, year of surgery (to account for secular trends impacting revision rates), component fixation, bearing surface, and insert type. The primary outcome assessed whether the adoption of RA-TKA affects CPR within each group of surgeons. The secondary outcomes were within-group changes in modes of failure for the different causes of revision that included infection, loosening, and instability.
Statistical Analysis
Time-to-first revision was estimated with Kaplan-Meier functions, and the unadjusted (raw) CPR was estimated with 95% confidence intervals (CI) presented in figures. Hazard ratios (HRs) from Cox proportional hazard models were used to compare CPR between each surgeon cohort and between RA-TKA and non–RA-TKA procedures. The assumption of proportional hazards was checked analytically for each model. If the interaction between the predictor and the log-of-time was statistically significant in the standard Cox model, then a time-varying model was estimated. Time-points were iteratively chosen until the assumption of proportionality was met, and the HRs were calculated for each selected period. In the following results, if no period is specified, then the HR was proportional over the entire follow-up period. All tests were two-tailed at the 5% level of significance. All analyses and figures were performed using SAS version 9.4 (SAS Institute).
Results
A total of 178 surgeons were included for analysis, performing a total of 85,075 TKAs since their commencement of robotics (Fig. 1). Before RA-TKA, 31 (17.4%) surgeons were classified as low-revision rate surgeons, 103 (57.9%) as middle-revision rate surgeons, and 44 (24.7%) as high-revision rate surgeons. High-revision rate surgeons performed less mean TKA volume before the commencement of RA-TKA (Table I), but each surgeon group was similar with respect to TKA volume after commencing RA-TKA and performed similar proportions of RA-TKA overall (Table II). The distribution of robotic systems used is summarized in Table III.
Fig. 1.

Summary of inclusion and exclusions for primary total knee replacement procedures.
TABLE I.
Summary of Surgeons Performing Primary Total Knee Replacement by Surgeon Revision Rate Before Robotic Assistance
| Variable | Low Revision Rate Surgeons | Middle Revision Rate Surgeons | High Revision Rate Surgeons | Total |
|---|---|---|---|---|
| No. of procedures 365 days before use of Robotic assistance (%) | ||||
| Mean (SD) | 96.9 (79.1) | 90.8 (62.7) | 68.2 (40.8) | 86.3 (62) |
| Median (IQR) | 60 (45, 149) | 71 (50, 121) | 62.5 (39.5, 86) | 68.5 (46, 115) |
| Minimum | 9 | 8 | 12 | 8 |
| Maximum | 277 | 401 | 174 | 401 |
| No. of Robotic procedures 365 days after first use of Robotic assistance (%) | ||||
| Mean (SD) | 44.2 (31.9) | 48 (67.8) | 48.3 (49.8) | 47.4 (58.5) |
| Median (IQR) | 40 (16, 63) | 30 (17, 53) | 36 (23, 53) | 33 (19, 54) |
| Minimum | 2 | 1 | 6 | 1 |
| Maximum | 127 | 520 | 317 | 520 |
| No. of all procedures 365 days after first use of Robotic Assistance (%) | ||||
| Mean (SD) | 104.1 (82.9) | 101.4 (80.)5 | 85.5 (51.2) | 98 (74.8) |
| Median (IQR) | 77 (46, 157) | 82 (47, 134) | 75 (50.5, 106) | 79.5 (49, 124) |
| Minimum | 5 | 10 | 28 | 5 |
| Maximum | 317 | 600 | 334 | 600 |
| Robotics commencing year (%) | ||||
| 2016 | 1 (3.2) | 1 (1) | 2 (1.1) | |
| 2017 | 4 (12.9) | 15 (14.6) | 6 (13.6) | 25 (14) |
| 2018 | 8 (25.8) | 38 (36.9) | 13 (29.5) | 59 (33.1) |
| 2019 | 5 (16.1) | 26 (25.2) | 11 (25) | 42 (23.6) |
| 2020 | 9 (29) | 17 (16.5) | 11 (25) | 37 (20.8) |
| 2021 | 4 (12.9) | 6 (5.8) | 3 (6.8) | 13 (7.3) |
| Surgeon Class before first used of Robotics (%) | ||||
| Regular using navigation | 17 (54.8) | 73 (70.9) | 31 (70.5) | 121 (68) |
| Rear or not using navigation | 14 (45.2) | 30 (29.1) | 13 (29.5) | 57 (32) |
| Total | 31 | 103 | 44 | 178 |
Abbreviation: IQR = interquartile range.
TABLE II.
Summary of Primary Total Knee Replacement since Commencement of Robotics by Prerobotics Surgeon Revision Rate
| Variable | Low Revision Rate Surgeons | Middle Revision Rate Surgeons | High Revision Rate Surgeons | Total |
|---|---|---|---|---|
| Age | ||||
| Mean (SD) | 69.5 (8.4) | 68.4 (8.8) | 67.8 (9.2) | 68.5 (8.9) |
| Median (IQR) | 70 (64, 75) | 69 (62, 75) | 68 (61, 74) | 69 (62, 75) |
| Age group (%) | ||||
| <55 | 716 (4.6) | 3,111 (6.2) | 1,574 (8) | 5,401 (6.3) |
| 55-64 | 3,530 (22.9) | 13,418 (26.8) | 5,343 (27.2) | 22,291 (26.2) |
| 65-74 | 6,727 (43.6) | 20,687 (41.3) | 7,973 (40.7) | 35,387 (41.6) |
| ≥75 | 4,449 (28.8) | 12,829 (25.6) | 4,718 (24.1) | 21,996 (25.9) |
| Sex (%) | ||||
| Male | 6,992 (45.3) | 22,362 (44.7) | 9,221 (47) | 38,575 (45.3) |
| Female | 8,430 (54.7) | 27,683 (55.3) | 10,387 (53) | 46,500 (54.7) |
| BMI category (%) | ||||
| Underweight (<18.50) | 32 (0.2) | 83 (0.2) | 37 (0.2) | 152 (0.2) |
| Normal (18.50-24.99) | 2,029 (13.2) | 5,525 (11) | 2,025 (10.3) | 9,579 (11.3) |
| Pre obese (25.00-29.99) | 5,308 (34.4) | 15,993 (32) | 6,041 (30.8) | 27,342 (32.1) |
| Obese Class 1 (30.00-34.99) | 4,690 (30.4) | 15,389 (30.8) | 6,340 (32.3) | 26,419 (31.1) |
| Obese Class 2 (35.00-39.99) | 2,146 (13.9) | 8,174 (16.3) | 3,266 (16.7) | 13,586 (16) |
| Obese Class 3 (≥40.00) | 1,217 (7.9) | 4,881 (9.8) | 1,899 (9.7) | 7,997 (9.4) |
| ASA (%) | ||||
| 1 | 745 (4.8) | 2,427 (4.8) | 993 (5.1) | 4,165 (4.9) |
| 2 | 8,981 (58.2) | 25,125 (50.2) | 10,726 (54.7) | 44,832 (52.7) |
| 3 | 5,696 (36.9) | 22,493 (44.9) | 7,889 (40.2) | 36,078 (42.4) |
| Fixation (%) | ||||
| Cemented | 8,616 (55.9) | 23,225 (46.4) | 9,063 (46.2) | 40,904 (48.1) |
| Cementless | 2,399 (15.6) | 17,816 (35.6) | 5,927 (30.2) | 26,142 (30.7) |
| Hybrid (Tibial Cemented) | 4,394 (28.5) | 8,793 (17.6) | 4,401 (22.4) | 17,588 (20.7) |
| Hybrid (tibial cementless) | 13 (0.1) | 211 (0.4) | 217 (1.1) | 441 (0.5) |
| Patella usage (%) | ||||
| Patella used | 13,230 (85.8) | 42,023 (84) | 12,883 (65.7) | 68,136 (80.1) |
| No patella | 2,192 (14.2) | 8,022 (16) | 6,725 (34.3) | 16,939 (19.9) |
| Bearing surface (%) | ||||
| Non XLPE | 1,437 (9.3) | 3,852 (7.7) | 1,178 (6) | 6,467 (7.6) |
| XLPE | 10,549 (68.4) | 37,043 (74) | 11,771 (60) | 59,363 (69.8) |
| XLPE + antioxidant | 3,436 (22.3) | 9,150 (18.3) | 6,659 (34) | 19,245 (22.6) |
| Stability (%) | ||||
| Medial Pivot design | 248 (1.6) | 1,360 (2.7) | 628 (3.2) | 2,236 (2.6) |
| Minimally stabilized | 12,202 (79.1) | 43,839 (87.6) | 16,699 (85.2) | 72,740 (85.5) |
| Posterior stabilized | 2,972 (19.3) | 4,846 (9.7) | 2,281 (11.6) | 10,099 (11.9) |
| Robot Type (%) | ||||
| Robotically assisted | 6,551 (42.5) | 34,181 (68.3) | 12,862 (65.6) | 53,594 (63) |
| No robotic assistance | 8,871 (57.5) | 15,864 (31.7) | 6,746 (34.4) | 31,481 (37) |
| Total | 15,422 | 50,045 | 19,608 | 85,075 |
Abbreviations: ASA = American Society of Anesthesiologists, BMI = Body Mass Index (kg/m2), and IQR = interquartile range.
Note: Commencement of Robotic assistance is at individual level.
TABLE III.
Summary of Robotic Systems Used
| Variable | Low Revision Rate Surgeons | Middle Revision Rate Surgeons | High Revision Rate Surgeons | Total |
|---|---|---|---|---|
| Robotic system used (%) | ||||
| CORI | 352 (2.3) | 631 (1.3) | 281 (1.4) | 1,264 (1.5) |
| MAKO | 4,778 (31.0) | 26,063 (52.1) | 6,076 (31.0) | 36,917 (43.4) |
| NAVIO | 155 (1.0) | 1,410 (2.8) | 1,894 (9.7) | 3,459 (4.1) |
| Not defined | 4 (0.03) | 5 (0.01) | 1 (0.01) | 10 (0.01) |
| ROSA | 1,247 (8.1) | 6,051 (12.1) | 4,610 (23.5) | 11,908 (14.0) |
| VELYS knee | 15 (0.1) | 21 (0.04) | 36 (0.04) | |
| Non-Robotic assistance | 8,871 (57.5) | 15,864 (31.7) | 6,746 (34.4) | 31,481 (37.0) |
| Total | 15,422 | 50,045 | 19,608 | 85,075 |
A summary of the patient demographics and implant types used by the 178 included surgeons (stratified by their pre–RA-TKA CPR) after commencing RA-TKA is shown in Table II. Each surgeon group was similar with respect to patient age, sex, BMI, ASA, and use of highly cross-linked polyethylene. Low revision rate surgeons used more cemented fixation (55.9%), more posterior stabilized (PS) (19.3%) bearings and less robotics (42.5%). High-revision rate surgeons had lower rates of patellar resurfacing (34.3%). With the adoption of RA-TKA, surgeons across all revision cohorts more frequently utilized cementless fixation and highly cross-linked polyethylene and resurfaced the patella, although decreased use of PS bearings.
Within-Group Changes After Commencing RA-TKA
After adopting RA-TKA, there was no change in CPR for all-causes among low (HR 1.27; 95% CI 0.84-1.91; p = 0.26), or middle-revision rates surgeons (HR 0.99; 95% CI 0.83-1.19; p = 0.95). However, as a group, high-revision surgeons showed a lower early revision rate (first 3 months: HR 0.51; 95% CI 0.34-0.75; p < 0.001) post–RA-TKA, but an increased rate from 9 months onward (HR 2.28; 95% CI 1.74-2.98; p < 0.001). The increased late revision rate in high-revision surgeons using RA-TKA was not driven by any specific mode of failure (p > 0.05 for all modes). No significant difference was found between 3 and 9 months (Fig. 2).
Fig. 2.

Cumulative percent revision of primary total knee replacement since use of robotics by prerobotics revision rate and robotic assistance (primary diagnosis OA).
Modes of Failure
After adopting RA-TKA, there was no change in CPR due to infection for low (HR 1.37; 95% CI 0.77-2.44; p = 0.29), middle (HR 0.84; 95% CI 0.65-1.09; p = 0.20), or high-revision rate surgeons (HR 0.99; 95% CI 0.65-1.51; p = 0.97). For revision due to implant loosening or instability, adjustments were only made for age and sex due to the small number of revision cases. Subsequently, there was no change in CPR due to implant loosening for low (HR 1.27; 95% CI 0.35-4.53; p = 0.72), middle (HR 1.04; 95% CI 0.61-1.76; p = 0.89), or high-revision rate surgeons (HR 1.17; 95% CI 0.56-2.46; p = 0.67). There was also no change in CPR due to instability among low (HR 1.11; 95% CI 0.38-3.22; p = 0.85), middle (HR 1.05; 95% CI 0.59-1.87; p = 0.88), or high-revision rate surgeons (HR 1.19; 95% CI 0.55-2.55; p = 0.66).
Discussion
This large national registry study does not support that RA-TKA improves the revision rates of surgeons stratified by their revision rates before commencing RA-TKA, nor changes their modes of failure. Rather, surgeon groups appear to persist with the same revision profile as before their commencement of RA-TKA. However, RA-TKA was associated with elevated rates of revision from 9 months in high-revision rate surgeons compared with non-RA-TKA. This finding is despite the increased selection of confounding prosthetic factors recognized to be associated with lower revision rates, except for cementless TKA fixation. Factors beyond robotic-assistance should be investigated to attempt to reduce the revision burden, and potential confounding factors associated with RA-TKA also require investigation to determine why revision rates increase for some surgeons. Further long-term follow-up on the continued impact of RA-TKA on revision rates is required when a greater number of procedures and potentially increased surgeon uptake has occurred.
Our results showing no improvement for any group of surgeons’ revision rates after commencement of RA-TKA are supported by a number of comparative studies17,18,21 and registry-based research30,31, albeit with shorter follow-up. The American Joint Replacement Registry (AJRR), after controlling for confounders including the surgeon, found no difference in revision rates between RA-TKA and TKA at two-years30, nor when solely examining cementless TKA between a cohort of patients undergoing RA-TKA compared with conventional techniques31. Their findings are important as we highlight the increased use of cementless fixation by all surgeon groups after commencing RA-TKA. However, the AJRR did find significantly increased odds for revision due to instability with robotic-assistance30, and a trend towards increased odds of mechanical loosening with cementless TKA (OR 3.2; p = 0.09). When the Medicare dataset from the AJRR was assessed, robotic-assistance did not improve five-year CPR, mechanical loosening, or other mechanical complications32. Along with adoption of RA-TKA there may be other confounding factors beyond robotic-assistance that explain the results of this study. Although cementless fixation increased, there were changes made by surgeons that should reduce revision rates, including better polyethylene choices, less PS bearings and more patella resurfacing. It remains unclear if any other concurrent change made at the time of commencing RA-TKA, including selection bias with case complexity, technique, alignment strategy and/or a learning curve, are important confounders for results.
We hypothesized that RA-TKA surgery would reduce the revision rates for specific causes, particularly component loosening and instability, which may relate to technical factors including accuracy of bony cuts, limb alignment and/or soft tissue balancing. However, we found this not to be the case, and no improvement was demonstrated. This perhaps is not entirely surprising given robotic-assistance still cannot improve many important technical issues associated with TKA survivorship including cementing technique, component insertion, patella cuts, aspects of patella tracking and component rotation, intra-operative decision-making including appropriateness for cementless fixation and the appropriate insert level of constraint, as well as preoperative decisions regarding indications for surgery and patient optimization. As mentioned, other changes could influence revision rates including fixation choice, bearings and patellar resurfacing. Moreover, errors may occur with robotic planning, registration or execution33. What remains unknown is how robotic-surgery is used by surgeons and how their alignment technique may or may not differ34. The initial premise of navigation, or computer assisted surgery (CAS), was that benefit would be achieved by both increased accuracy and by minimizing outliers. However, CAS began in an era where alignment targets and techniques were almost exclusively mechanical. Now, with the precision of robotic-assistance, multiple personalized anatomical-based alignment targets and techniques have evolved. Where CAS minimized outliers, robotic-assistance may facilitate outliers from traditional targets34. How RA-TKA and alignments are intertwined with revision outcomes, and the relative importance that component alignment is as a confounder to outcomes, remains unknown, and may explain some increase in revision rates35. Only enhanced data collection from joint replacement registries can assess this36.
There are limitations in this study. We only included revision outcomes, and not patient reported outcomes (PRO’s). However, the AOANJRR has previously demonstrated that RA-TKA has no significant association with change in Oxford Knee Scores or EQ-5D, nor does the surgeon performing TKA37. RA-TKA may make improvement to other measures of patient function or knee performance, that are not recorded by the AOANJRR and cannot be tested in this study. Other limitations are the limited follow-up available. Despite up to 7 years of follow-up, until long-term data are available, the true difference between surgery performed with and without robotic-assistance is unknown. The accrual of more procedures, performed by more surgeons, is required. This is particularly the case for low revision surgeons who had a non-statistical increase in revision rates with the commencement of RA-TKA. Another limitation is that we also stipulated in our inclusion criteria that surgeons needed to have a 2-year CPR. Whether the results can be extrapolated to very low volume surgeons is unknown. In addition, we do not know the amount or level of training, with conventional or robotic techniques, as these may be important factors influencing revision rates19,38. All robotic systems were also grouped together. Different systems have varying alignment philosophies, workflows, learning curves, and system specific complications, which represents a limitation of this analysis. Of note, approximately 70% of RA-TKA were Mako (Stryker) and 25% ROSA (Zimmer-Biomet). Finally, there is the potential for selection bias with RA-TKA used in selected cases.
Conclusions
RA-TKA failed to demonstrate any reduction in revision rates relative to standard TKA across 3 cohorts of surgeons performing more than 50 cases annually, irrespective of their baseline revision rate classification - low, medium, or high. Surgeons with higher preadoption revision rates experienced an increase in late (>9 months) revision risk with RA-TKA. Modes of failure were unchanged. Although the benefits of RA-TKA may potentially extend beyond revision rates, with the data currently available, the exact role of RA-TKA remains unknown.
Footnotes
Investigation performed at the Australian Orthopaedic Association National Joint Replacement Registry, Adelaide, South Australia, Australia
Disclosure: The Disclosure of Potential Conflicts of Interest forms are provided with the online version of the article (http://links.lww.com/JBJSOA/B212).
Contributor Information
Charles Gusho, Email: cagghd@health.missouri.edu.
Roger Bingham, Email: rogerbingham@mac.com.
Peiyao Du, Email: andie.xu@sahmri.com.
Helena Oakey, Email: helena.oakey@sahmri.com.
Kelly G Vince, Email: kellyvince@mac.com.
Michael McAuliffe, Email: mmcauliffe@aoanjrr.org.au.
References
- 1.Inabathula A, Semerdzhiev DI, Srinivasan A, Amirouche F, Puri L, Piponov H. Robots on the stage: a snapshot of the American robotic total knee arthroplasty market. JBJS Open Access. 2024;9(3):e24. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Lewis PL, Gill DR, McAuliffe MJ, Stoney JD, Vertullo CJ, Wall CJ, Corfield S, Esiain R, Moylan S, Du P, Holder C, Edwards S, Xu Q, Oakey H, Lorimer MF, Smith PN. Hip, Knee and Shoulder Arthroplasty: 2025 Annual Report; 2025. [Google Scholar]
- 3.Sousa PL, Sculco PK, Mayman DJ, Jerabek SA, Ast MP, Chalmers BP. Robots in the operating room during hip and knee arthroplasty. Curr Rev Musculoskelet Med. 2020;13(3):309-17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sherman WF, Wu VJ. Robotic surgery in total joint arthroplasty: a survey of the AAHKS membership to understand the utilization, motivations, and perceptions of total joint surgeons. J Arthroplasty. 2020;35(12):3474-81. e2. [DOI] [PubMed] [Google Scholar]
- 5.DeFrance MJ, Yayac MF, Courtney PM, Squire MW. The impact of author financial conflicts on robotic-assisted joint arthroplasty research. J Arthroplasty. 2021;36(4):1462-9. [DOI] [PubMed] [Google Scholar]
- 6.Banger M, Doonan J, Rowe P, Jones B, MacLean A, Blyth MJ. Robotic arm-assisted versus conventional medial unicompartmental knee arthroplasty: five-year clinical outcomes of a randomized controlled trial. Bone Jt J. 2021;103-B(6):1088-95. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Bell SW, Anthony I, Jones B, MacLean A, Rowe P, Blyth M. Improved accuracy of component positioning with robotic-assisted unicompartmental knee arthroplasty: data from a prospective, randomized controlled study. J Bone Jt Surg. 2016;98(8):627-35. [DOI] [PubMed] [Google Scholar]
- 8.Bouché P-A, Corsia S, Dechartres A, Resche-Rigon M, Nizard R. Are there differences in accuracy or outcomes scores among navigated, robotic, patient-specific instruments or standard cutting guides in TKA? A network meta-analysis. Clin Orthop Relat Res. 2020;478(9):2105-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kayani B, Tahmassebi J, Ayuob A, Konan S, Oussedik S, Haddad FS. A prospective randomized controlled trial comparing the systemic inflammatory response in conventional jig-based total knee arthroplasty versus robotic-arm assisted total knee arthroplasty. Bone Jt J. 2021;103-B(1):113-22. [DOI] [PubMed] [Google Scholar]
- 10.Li Z, Chen X, Wang X, Zhang B, Wang W, Fan Y, Yan J, Zhang X, Zhao Y, Lin Y, Liu J, Lin J. HURWA robotic-assisted total knee arthroplasty improves component positioning and Alignment–A prospective randomized and multicenter study. J Orthop Transl. 2022;33:31-40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Liow MHL, Goh GSH, Wong MK, Chin PL, Tay DKJ, Yeo SJ. Robotic‐assisted total knee arthroplasty may lead to improvement in quality‐of‐life measures: a 2‐year follow‐up of a prospective randomized trial. Knee Surg Sports Traumatol Arthrosc. 2017;25(9):2942-51. [DOI] [PubMed] [Google Scholar]
- 12.Liow MHL, Xia Z, Wong MK, Tay KJ, Yeo SJ, Chin PL. Robot-assisted total knee arthroplasty accurately restores the joint line and mechanical axis. A prospective randomised study. J Arthroplasty. 2014;29(12):2373-7. [DOI] [PubMed] [Google Scholar]
- 13.Motesharei A, Rowe P, Blyth M, Jones B, Maclean A. A comparison of gait one year post operation in an RCT of robotic UKA versus traditional Oxford UKA. Gait Posture. 2018;62:41-5. [DOI] [PubMed] [Google Scholar]
- 14.Song E-K, Seon J-K, Yim J-H, Netravali NA, Bargar WL. Robotic-assisted TKA reduces postoperative alignment outliers and improves gap balance compared to conventional TKA. Clin Orthop Relat Res. 2013;471(1):118-26. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Song EK, Seon JK, Park SJ, Jung WB, Park HW, Lee GW. Simultaneous bilateral total knee arthroplasty with robotic and conventional techniques: a prospective, randomized study. Knee Surg Sports Traumatol Arthrosc. 2011;19(7):1069-76. [DOI] [PubMed] [Google Scholar]
- 16.Vaidya NV, Deshpande AN, Panjwani T, Patil R, Jaysingani T, Patil P. Robotic‐assisted TKA leads to a better prosthesis alignment and a better joint line restoration as compared to conventional TKA: a prospective randomized controlled trial. Knee Surg Sports Traumatol Arthrosc. 2022;30(2):621-6. [DOI] [PubMed] [Google Scholar]
- 17.Gilmour A, MacLean AD, Rowe PJ, Banger MS, Donnelly I, Jones BG, Blyth MJ. Robotic-arm–assisted vs conventional unicompartmental knee arthroplasty. The 2-year clinical outcomes of a randomized controlled trial. J Arthroplasty. 2018;33(7):S109-S115. [DOI] [PubMed] [Google Scholar]
- 18.Kim Y-H, Yoon S-H, Park J-W. Does robotic-assisted TKA result in better outcome scores or long-term survivorship than conventional TKA? A randomized, controlled trial. Clin Orthop Relat Res. 2020;478(2):266-75. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Ali M, Phillips D, Kamson A, Nivar I, Dahl R, Hallock R. Learning curve of robotic-assisted total knee arthroplasty for non-fellowship-trained orthopedic surgeons. Arthroplasty Today. 2022;13:194-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ruangsomboon P, Ruangsomboon O, Pornrattanamaneewong C, Narkbunnam R, Chareancholvanich K. Clinical and radiological outcomes of robotic-assisted versus conventional total knee arthroplasty: a systematic review and meta-analysis of randomized controlled trials. Acta Orthop. 2023;94:60-79. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Tompkins GS, Sypher KS, Li H-F, Griffin TM, Duwelius PJ. Robotic versus manual total knee arthroplasty in high volume surgeons: a comparison of cost and quality metrics. J Arthroplasty. 2022;37(8):S782-S789. [DOI] [PubMed] [Google Scholar]
- 22.Tay ML, Carter M, Bolam SM, Zeng N, Young SW. Robotic-arm assisted unicompartmental knee arthroplasty system has a learning curve of 11 cases and increased operating time. Knee Surg Sports Traumatol Arthrosc. 2023;31(3):793-802. [DOI] [PubMed] [Google Scholar]
- 23.Zak SG, Cieremans D, Tang A, Schwarzkopf R, Rozell JC. Intraoperative technology increases operating room times in primary total knee arthroplasty. Arch Orthop Trauma Surg. 2022;143(4):2113-9. [DOI] [PubMed] [Google Scholar]
- 24.St Mart J-P, De Steiger RN, Cuthbert A, Donnelly W. The three-year survivorship of robotically assisted versus non-robotically assisted unicompartmental knee arthroplasty: a study from the Australian Orthopaedic Association National Joint Replacement Registry. Bone Jt J. 2020;102-B(3):319-28. [DOI] [PubMed] [Google Scholar]
- 25.Thomas TL, Goh GS, Nguyen MK, Lonner JH. Pin-related complications in computer longer follow-up is required to better assess the true impacts of RA-TKA.navigated and robotic-assisted knee arthroplasty: a systematic review. J Arthroplasty. 2022;37(11):2291-307. e2. [DOI] [PubMed] [Google Scholar]
- 26.Hoskins W, Rainbird S, Lorimer M, Graves SE, Bingham R. What can we learn from surgeons who perform THA and TKA and have the lowest revision rates? A study from the Australian orthopaedic association national joint replacement registry. Clin Orthop Relat Res. 2022;480(3):464-81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Rajan PV, Khlopas A, Klika A, Molloy R, Krebs V, Piuzzi NS. The cost-effectiveness of robotic-assisted versus manual total knee arthroplasty: a Markov model–based evaluation. J Am Acad Orthop Surg. 2022;30(4):168-76. [DOI] [PubMed] [Google Scholar]
- 28.Zhou Y, Wall CJ, Stevens J, Fraval A, Lewis PL, McAuliffe MJ, Vertullo CJ, Gill DRJ, Stoney JD, Bastiras D, Corfield S, Harvey B, Lorimer MF, Hill K, Smith PN. Data resource profile: the Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR). Int J Epidemiol. 2025;54(4):dyaf078. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Hoskins W, Bingham R, Graves SE, Harries D, Cuthbert AR, Corfield S, Smith P, Vince KG. What are the relative associations of surgeon performance and prosthesis quality with THA revision rates? Clin Orthop Relat Res. 2025;483(2):237-49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Kirchner GJ, Stambough JB, Jimenez E, Nikkel LE. Robotic-assisted TKA is not associated with decreased odds of early revision: an analysis of the American Joint Replacement Registry. Clin Orthop Relat Res. 2024;482(2):303-10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Kirchner GJ, Stambough JB, Jimenez E, Mullen K, Nikkel LE. Robotic assistance is not associated with decreased early revisions in cementless TKA: an analysis of the American Joint Replacement Registry. Clin Orthop Relat Res. 2025;483(3):431-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Pius AK, Sporer SM, Sterling O, De M, Jahan M, Browne JA, Springer BD, Huddleston JI. Navigated and robotic total knee arthroplasty do not confer improved 5-Year survivorship compared to conventional TKA: an analysis from the American joint replacement registry. J Arthroplast. 2025. [DOI] [PubMed] [Google Scholar]
- 33.Pagano NR, Menendez ME, Moverman MA, Puzzitiello RN, Gordon MR. Adverse events associated with robotic-assisted joint arthroplasty: an analysis of the US Food and Drug Administration MAUDE Database. J Arthroplasty. 2022;37(8):1526–1533. [DOI] [PubMed] [Google Scholar]
- 34.Morcos MW, Uhuebor D, Vendittoli PA. Overview of the different personalized total knee arthroplasty with robotic assistance, how choosing? Front Surg. 2023;10:1120908. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Bae K, Lee BS, Kim JM, Bin SI, Lee J, Kim D, Jeon T. Effect of joint-line obliquity on long-term survivorship of total knee arthroplasty: a postoperative phenotype analysis. Knee Surg Sports Traumatol Arthrosc. 2024;32(12):3230-8. [DOI] [PubMed] [Google Scholar]
- 36.Hoskins W, Bingham R, Vince KG. A systematic review of data collection by national joint replacement registries: what opportunities exist for enhanced data collection and analysis?. JBJS Rev. 2023;11(10):e23. [DOI] [PubMed] [Google Scholar]
- 37.Hoskins W, Bingham R, Corfield S, Harries D, Harris IA, Vince KG. Do the revision rates of arthroplasty surgeons correlate with postoperative patient-reported outcome measure scores? A study from the Australian Orthopaedic Association National Joint Replacement Registry. Clin Orthop Relat Res. 2022;10:1097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Katanbaf R, Hoveidaei AH, Swartz GN, Misch M, Stokes L, Ihekweazu UN, Mont MA, Nace J, Delanois RE. Fellowship training in adult reconstruction is associated with decreased complications up to two years following total knee arthroplasty. J Arthroplasty. 2025;S0883-5403(25):01229-X. [DOI] [PubMed] [Google Scholar]