Table 3.

Effect of partial humeral head inclusion on radiation dose

Patient characteristics All scans (n = 90) Scans arms out (n = 35) Scans arms in (= 55) P-value
Body mass index (kg/m2) 90 21.1 (19.5; 22.4) 55 21.0 (19.6; 22.1) 35 21.6 (19.0; 23.0) 0.634
 Weight (kg) 90 62.0 (57.5; 67.9) 55 61.2 (56.9; 67.5) 35 62.7 (58.2; 69.0) 0.317
 Height (cm) 90 172.0 (168.0; 177.5) 55 171.0 (167.0; 177.0) 35 174.0 (170.0; 178.0) 0.081
Scan characteristics
 CT scan length (cm) 90 4.9 (4.2; 5.4) 35 4.9 (4.2; 5.2) 55 4.8 (4.4; 5.5) 0.904
 Scan length group, n (%) 90 35 55
≤ 4 cm 26 (28.9) 10 (28.6) 16 (29.1)
> 4 cm 64 (71.1) 25 (71.4) 39 (70.9) 0.958
 Total humeral head inclusion (R&L) (cm) 90 0.4 (0.0; 2.3) 35 0.0 (0.0; 0.0) 55 1.8 (0.8; 3.60 < 0.001
Right humeral head inclusion (cm) 90 0.0 (0.0; 1.2) 35 0.0 (0.0; 0.0) 55 0.9 (0.2; 1.9) < 0.001
Left humerus head inclusion (cm) 90 0.1 (0.0; 1.0) 35 0.0 (0.0; 0.0) 55 0.9 (0.3; 1.70 < 0.001
 Humeral head position, n (%) 90 35 55
Arms in 55 (61.1) 0 (0.0) 55 (100.0)
Arms out 35 (38.9) 35 (100.0) 0 (0.0) < 0.001
 Humeral head inclusion/scan length, n (%) 90 35 55
Arms out, SL: ≤4 cm 10 (11.1) 10 (28.6) 0 (0.0)
Arms in, SL: ≤4 cm 16 (17.8) 0 (0.0) 16 (29.1)
Arms out, SL: >4 cm 25 (27.8) 25 (71.4) 0 (0.0) < 0.001
Arms in, SL: >4 cm 39 (43.3) 0 (0.0) 39 (70.9)
Radiation doses
 Effective CT tube current (mAs) 90 65.5 (58.0; 81.0) 35 58.0 (50.0; 74.0) 55 68.0 (61.0; 84.0) < 0.001
 CT dose index (mGy) 90 4.5 (3.9; 5.5) 35 3.9 (3.4; 5.0) 55 4.7 (4.2; 5.7) 0.001
 Dose length product (mGy*cm) 90 30.0 (27.0; 38.0) 35 29.0 (24.0; 32.0) 55 33.0 (28.0; 42.0) 0.003
 Effective dose (mSv) 90 0.4 (0.4; 0.5) 35 0.4 (0.3; 0.4) 55 0.5 (0.4; 0.6) 0.003

Continuous variables are described with median (IQR), p-values obtained by Wilcoxon rank sum test

Categorical variables are shown with number (%) in each category, p-values obtained by Chi-squared test or by Fisher’s exact test if appropriate