Table 3. Troubleshooting guide for α-Syn purification issues.

Problem Potential causes Solutions

Low purity after anion exchange chromatography

(1) Overloading column

(2) Insufficient lysate clarification

(1) Load ≤10 mg total protein per run

(2) Centrifuge lysate at 40,000g for 1 h before loading

Protein aggregation during size exclusion chromatography (SEC)

(1) High protein concentration

(2) Improper buffer pH or contaminants

Dilute the sample to ≤2 mg/mL before SEC

Low yield of purified α-Syn

(1) Inefficient lysis

(2) Loss of protein during purification steps

(1) Ensure complete bacterial lysis using 4–5 cycles of high-pressure homogenization

(2) Pool fractions containing α-Syn to maximize yield