Table 4. Troubleshooting guide for RT-QuIC assay issues.

Problem Potential causes Solutions

Delayed or no ThT fluorescence increase

(1) Insufficient seeding activity in TCA-enriched samples

(2) Suboptimal SDS or ThT concentration

(1) Increase seed concentration to 2 μg/mL or extend reaction time to 72 h

(2) Optimize SDS concentration (0.001%–0.003%) to balance fibril formation

High background fluorescence in negative controls

(1) Contamination from reagents or equipment

(2) Premature fibrillization due to temperature fluctuations

(1) Use fresh, sterile-filtered buffers and UV-treated consumables

(2) Pre-chill plates on ice and maintain 37°C during incubation

Inconsistent fibril formation

(1) Variability in recombinant α-Syn batch

(2) Uneven plate shaking

(1) Use monomeric, endotoxin-free α-Syn validated by SEC

(2) Ensure consistent 600 r/min shaking during RT-QuIC