Table 1:

Acquisition Parameters for the Imaging Protocol, Including T1-weighted, T2-weighted, Kidney MRF, DWI, and ASL MRI

Parameter T1-weighted T2-weighted MRF DWI* ASL
Acquisition sequence Gradient echo HASTE FISP Multishot EPI Gradient echo
Field of view (mm) 400 × 400 400 × 400 400 × 400 240 × 240 320 × 160
Matrix size 300 × 300 300 × 300 256 × 256 160 × 160 64 × 32
Spatial resolution (mm) 1.3 × 1.3 1.3 × 1.3 1.6 × 1.6 1.5 × 1.5 5.0 × 5.0
Section thickness (mm) 3 5 5 4 5
Repetition time (msec) 9.0 1050 5.7 2400 5000
Echo time (msec) 1.1 82 1 55/91 19.2
Flip angle (degrees) 4 … 5–12 … …
Duration (min) 0.25 0.32 0.25 per section … 5.3

Note.—ASL = arterial spin labeling, DWI = diffusion-weighted imaging, EPI = echo-planar imaging, FISP = fast imaging with steady-state free precession, HASTE = half-Fourier acquisition single-shot turbo spin echo, MRF = MR fingerprinting.

*

Two b values (0 and 400 sec/mm2) were applied with 12 diffusion directions.