Skip to main content
. 2026 May 2;418(14):4481–4495. doi: 10.1007/s00216-026-06539-6

Table 2.

δ(61Ni/62Ni) values obtained in a series of 61Ni-enriched/natural Ni mixtures in various ratios. U (δ(61Ni/62Ni)) is the standard deviation of 900 instrumental replicates

xspike/% δ(61Ni/62Ni)/‰ δ(61Ni/62Ni)/‰ U (δ(61Ni/62Ni))/‰
Theoretical Measured
0.00 0.00 −0.1514 0.0008
0.01 7.87 5.136 0.028
0.02 15.6 19.23 0.20
0.05 38.2 35.28 0.21
0.10 73.5 63.96 0.50
0.15 106 96.63 0.43
1.50 545 521.0 11.0
2.25 643 619.0 3.3
20.0 945 939.8 5.1
50.0 980 978.9 7.3
90.0 991 991.1 5.1
100 992 992.6 8.1