System Hardware Parameters.

Component Parameter Symbol Value Unit Reference
Battery Chemistry — Li-ion — 37
Battery Nominal voltage V_nom 3.7 V Typical
Battery Rated capacity C_bat 500 mAh Typical
Battery Internal resistance R_int 0.15 Ω Typical
Battery Peukert exponent k 1.05 — 37
BLE 5.0 Radio Center frequency f_c 2.440 GHz 3
BLE 5.0 Radio Channel bandwidth B 2 MHz 3
BLE 5.0 Radio PHY data rate R_b 1.0 Mbps 3
BLE 5.0 Radio Max payload (DLE) L_pkt 251 bytes 3
BLE 5.0 Radio Rx sensitivity P_sens −96 dBm 32
BLE 5.0 Radio Noise figure (Rx) NF 7.0 dB 32
BLE 5.0 Radio Design distance d 3.0 m Assumed
BLE 5.0 Radio Body+indoor path loss PL_add 35 dB 7,8
ECG Sensor IC model — ADS1293 — 36
ECG Sensor Sampling freq. (MIT-BIH) f_s 360 Hz 34
ECG Sensor Sampling freq. (PTB-XL) f_s 500 Hz 35
ECG Sensor ADC resolution N_bits 12 bits 36
MCU (nRF52840) MCU current (Full mode) I_MCU, full 5.20 mA 32
MCU (nRF52840) MCU current (Critical) I_MCU, crit 2.10 mA 32

* Peukert correction applied with k = 1.0537.