Skip to main content
. 2026 Jul 19;16:22551. doi: 10.1038/s41598-026-62387-5

Table 3.

Per-Record QRS Detection Performance—MIT-BIH Arrhythmia Database.

Record Condition SoC BER Sens. Base [%] Sens. BAFE [%] Spec. BAFE [%] F1 BAFE [%] ΔSens. [pp] SQNR Base [dB] SQNR BAFE [dB] Clin. OK
MITBIH-001 Normal Sinus Rhythm 20% 1.78 × 10⁻² 37.5 97.8 97.7 97.8 + 60.3 18.9 12.1 Yes
MITBIH-001 Normal Sinus Rhythm 40% 7.33 × 10⁻³ 48.3 98.1 97.8 98.0 + 49.8 22.7 13.8 Yes
MITBIH-002 Atrial Fibrillation 20% 1.78 × 10⁻² 42.1 96.4 96.3 96.3 + 54.3 18.6 11.6 Yes
MITBIH-003 Sinus Bradycardia 20% 1.78 × 10⁻² 39.4 98.1 97.9 98.0 + 58.7 20.0 11.5 Yes
MITBIH-004 Sinus Tachycardia 20% 1.78 × 10⁻² 36.8 97.4 97.2 97.3 + 60.6 18.6 11.6 Yes
MITBIH-005 PVC 20% 1.78 × 10⁻² 31.2 72.3 81.6 76.7 + 41.1 20.1 12.3 No
MITBIH-006 LBBB 20% 1.78 × 10⁻² 28.9 96.2 96.1 96.2 + 67.3 18.5 12.1 Yes
MITBIH-007 Atrial Flutter 20% 1.78 × 10⁻² 23.8 95.6 96.0 95.8 + 71.8 20.3 12.9 Yes
MITBIH-008 Normal Sinus Rhythm 20% 1.78 × 10⁻² 14.1 97.7 97.9 97.8 + 83.6 20.5 13.1 Yes
MITBIH-009 Ventricular Tachycardia 20% 1.78 × 10⁻² 37.4 69.1 82.7 75.3 + 31.7 18.6 12.5 No
MITBIH-010 Normal Sinus Rhythm 20% 1.78 × 10⁻² 59.3 98.3 98.3 98.3 + 39.0 18.9 11.8 Yes
Mean (20% battery) — 20% 1.78 × 10⁻² 35.1 91.9 94.2 92.9 + 56.8 19.3 12.1 8/10
Mean (40% battery) — 40% 7.33 × 10⁻³ 43.0 91.8 93.5 92.6 + 48.9 23.4 14.3 8/10
Mean (100% battery) — 100% 1.18 × 10⁻³ 86.4 93.3 96.4 94.7 + 7.0 31.4 17.2 8/10

ΔSens.: sensitivity gain (BAFE − baseline) in percentage points. Clin. OK: Se ≥ 95% (ANSI/AAMI EC5711). 8 of 10 MIT-BIH records meet this threshold at 20% SoC. ‡SQNR values represent 10 log₁₀(Pₛᴵᴳₙₐₗ/Pₙₒᴵₛᴵ). At ≤ 0% SoC, BAFE improves SQNR (+ 3.2 dB at 5% SoC) by suppressing severe bit-flip noise. At ≥ 20% SoC, BPF morphological distortion dominates: BAFE reduces SQNR by 7.2 dB—the approximate-computing trade-off33where wideband fidelity is sacrificed for QRS detectability.