. 2026 Jul 13;13:1878351. doi: 10.3389/fmed.2026.1878351

Table 2.

Segmentation performance on internal 2D ultrasound and CEUS domain-shift tests.

Model Evaluation set N images Dice IoU Boundary F1 HD95 Pixel ECE
Residual U-Net Internal 2D ultrasound test 382 0.750 (0.191) 0.632 (0.216) 0.167 (0.123) 97.986 (50.627) 0.039 (0.044)
Residual U-Net CEUS domain-shift test 170 0.468 (0.190) 0.326 (0.167) 0.037 (0.035) 199.683 (65.159) 0.314 (0.143)
Combined augmentation Internal 2D ultrasound test 382 0.745 (0.196) 0.628 (0.221) 0.178 (0.133) 93.448 (50.716) 0.037 (0.042)
Combined augmentation CEUS domain-shift test 170 0.476 (0.192) 0.334 (0.171) 0.041 (0.032) 197.557 (64.357) 0.297 (0.138)
Blur augmentation Internal 2D ultrasound test 382 0.748 (0.196) 0.631 (0.220) 0.168 (0.126) 99.917 (52.583) 0.037 (0.042)
Blur augmentation CEUS domain-shift test 170 0.477 (0.212) 0.338 (0.187) 0.050 (0.037) 203.073 (64.607) 0.275 (0.132)
Photometric augmentation Internal 2D ultrasound test 382 0.752 (0.190) 0.635 (0.217) 0.173 (0.127) 96.156 (49.826) 0.038 (0.043)
Photometric augmentation CEUS domain-shift test 170 0.465 (0.190) 0.323 (0.165) 0.043 (0.033) 201.750 (64.956) 0.302 (0.124)
Noise augmentation Internal 2D ultrasound test 382 0.746 (0.193) 0.627 (0.217) 0.167 (0.118) 98.379 (51.834) 0.039 (0.045)
Noise augmentation CEUS domain-shift test 170 0.429 (0.158) 0.287 (0.136) 0.017 (0.019) 208.041 (68.450) 0.439 (0.126)
Attention U-Net Internal 2D ultrasound test 382 0.749 (0.190) 0.631 (0.217) 0.168 (0.129) 98.487 (53.943) 0.038 (0.042)
Attention U-Net CEUS domain-shift test 170 0.429 (0.175) 0.289 (0.146) 0.035 (0.028) 204.505 (65.856) 0.328 (0.132)
U-Net Internal 2D ultrasound test 382 0.752 (0.189) 0.635 (0.214) 0.166 (0.123) 98.914 (51.314) 0.039 (0.044)
U-Net CEUS domain-shift test 170 0.439 (0.162) 0.296 (0.141) 0.024 (0.028) 202.188 (66.009) 0.374 (0.114)

Values are mean (SD).