Feasibility and Safety of Cross‐Regional 5G‐Enabled Remote Robot‐Assisted Laparoscopic Surgery: A Retrospective Case Series of 21 Patients

Di Pan 1, Yong Gu 2, Dahong Zhang 1,✉
PMCID: PMC13417746  PMID: 42522914

ABSTRACT

Background

To address surgical resource scarcity in remote areas, we assessed the feasibility, safety, network stability, and training impact of 5G‐enabled remote robot assisted laparoscopic surgery.

Methods

Retrospective analysis of 21 consecutive patients (Oct 2023–Sep 2025) undergoing remote surgery from Zhejiang to Xinjiang (3700 km) using the Toumai robot with a redundant 5G network. Primary endpoint was technical success; secondary endpoints included operative outcomes, complications, network metrics, and OSATS scores.

Results

Technical success was 100%. Procedures included partial nephrectomy (8), cholecystectomy (4), hysterectomy (3), prostatectomy (2), colorectal resection (2), and liver segmentectomy (2). Mean operative time 158.5 ± 42.0 min, console time 91.7 ± 26.6 min, blood loss 69.8 ± 40.2 mL. Two minor complications occurred. Mean round‐trip latency 37.7 ± 5.6 ms, no packet loss. OSATS improved from 64.6 ± 7.1 to 87.9 ± 4.6.

Conclusions

With redundant 5G and local backup, remote robotic surgery is feasible and safe and may facilitate training.

Keywords: 5G‐enabled remote surgery, cross‐regional healthcare, laparoscopic surgery, network performance, robotic surgery, surgical training

1. Introduction

Inadequate access to surgical services and the uneven distribution of surgical resources are long‐standing challenges for health systems worldwide. The Lancet Commission on Global Surgery 2030 highlighted that large numbers of patients bear avoidable disease and economic burdens because they cannot obtain timely, safe, and affordable surgical care [1, 2]. Although traditional short‐term surgical outreach programs can address some immediate surgical needs, their sustainability, cost‐effectiveness, and effectiveness in transferring skills remain debated [3]. Robot‐assisted laparoscopic surgery enhances a surgeon's ability to perform delicate manoeuvres in anatomically complex regions by providing three‐dimensional vision, tremor filtration, motion scaling, and wristed instrumentation [4, 5]. However, robotic systems and the specialised teams that operate them are largely concentrated in high‐volume medical centres, leaving patients in remote areas with limited access to advanced minimally invasive procedures.

Remote robotic surgery is regarded as a potential strategy to overcome this spatial mismatch. In 2001, Marescaux et al. performed a transatlantic robot‐assisted cholecystectomy, demonstrating that telesurgery over long distances can be accomplished under rigorous technical conditions [6]. Subsequently, Anvari et al. established a telerobotic surgical service for a rural community, showing that such a model can have clinical utility within a defined healthcare system [7]. However, broader adoption of remote surgery has been constrained by communication latency, network instability, video transmission quality, equipment reliability, data security, and liability allocation [8, 9]. The emergence of 5G communication technology has renewed interest in remote robotic surgery. Compared with public networks, 5G dedicated lines or private medical networks can provide higher bandwidth, lower latency, and a more stable data transmission environment, thereby supporting the real‐time transfer of high‐definition video, voice communication, and robotic control signals. Early clinical applications of 5G‐assisted telementoring and 5G‐enabled remote robot assisted laparoscopic surgery (5G‐RRLS) have been reported [10, 11, 12, 13]. However, most published studies are small case series, single‐procedure evaluations, or technical feasibility assessments, and few have systematically quantified network performance or examined capacity building for the patient‐side team.

Since 2023, Zhejiang Provincial People's Hospital and the First People's Hospital of Aksu Prefecture, Xinjiang, have established a cross‐regional remote robotic surgery collaboration based on a redundant 5G dedicated‐line network. This programme extends beyond the question of whether remote surgery can be successfully completed and also incorporates local team training, patient‐side workflow coordination, and emergency takeover capability into a comprehensive evaluation. The present study retrospectively analysed the initial 21 cases of cross‐regional 5G‐RRLS, aiming to assess the technical feasibility, perioperative safety, network stability, and preliminary capacity‐building outcomes of this model.

2. Methods

2.1. Study Design and Ethics

This retrospective case series consecutively enroled all 5G‐RRLSs performed through the collaboration between Zhejiang Provincial People's Hospital and the First People's Hospital of Aksu Prefecture, Xinjiang, from October 2023 to September 2025. The study protocol was approved by the Ethics Committee of the First People's Hospital of Aksu Prefecture, Xinjiang (approval number 2023‐K‐046) and registered with the Ethics Committee of Zhejiang Provincial People's Hospital. All patients provided written informed consent for both robotic surgery and the specific remote surgical procedure.

The informed consent documents covered the tele‐surgery workflow, the location of the remote operating surgeon, the potential risks related to the network, the local team's emergency takeover plan, possible conversion to conventional laparoscopy or open surgery, and measures to protect audiovisual data and patient privacy. This report follows the STROBE statement for observational studies and the PROCESS guideline for surgical case series [14, 15].

2.2. Patient Selection

Inclusion criteria were: age 18–80 years; American Society of Anaesthesiologists (ASA) physical status I–III; diagnosis of benign or malignant disease in urology, hepatobiliary surgery, gynaecology, or gastrointestinal surgery with an indication for elective laparoscopic surgery; and, for malignant tumours, clinical stage T1–3N0M0. Exclusion criteria were: emergency surgery, severe coagulopathy, severe cardiopulmonary dysfunction, multiple prior major abdominal operations, suspected extensive abdominal adhesions on preoperative evaluation, body mass index > 35 kg/m2, or cases in which, in the judgement of the local team, the patient could not be safely managed by bedside takeover, conversion to conventional laparoscopy, or conversion to open surgery in the event of a network or equipment failure.

2.3. Remote Robotic Surgical System

All procedures used the Toumai endoscopic surgical robot equipped with a telesurgery module. The remote surgeon console was located at Zhejiang Provincial People's Hospital, and the patient cart was situated in the operating room of the First People's Hospital of Aksu Prefecture, Xinjiang, approximately 3700 km apart (Figure 1).

FIGURE 1.

FIGURE 1

Schematic diagram of the cross‐regional 5G‐enabled remote robot‐assisted laparoscopic surgery system. Schematic of the 5G‐enabled remote robotic surgery system spanning 3700 km between Hangzhou (surgeon console) and Aksu, Xinjiang (patient cart). A redundant 5G dedicated network ensures low‐latency transmission of video, control, and audio signals. The remote surgeon operated the robotic instruments, while the patient‐side team managed all local intraoperative tasks and emergency backup.

The remote operating surgeon performed the core robotic manoeuvres—instrument control and camera manipulation—from the console in Hangzhou. The patient‐side team at Aksu was responsible for anaesthesia management, patient positioning, port placement, robot docking, instrument exchange, bedside suction and retraction, specimen extraction, wound closure, and intraoperative emergency management.

2.4. Network Architecture and Performance Monitoring

The remote surgeries were carried out over a redundant 5G dedicated‐line network. A 5G dedicated line from China Telecom served as the primary link, with 5G dedicated lines from China Mobile and China Unicom as redundant backup links. Dedicated quality‐of‐service (QoS) strategies were applied to the transmission of surgical control signals and video streams. Real‐time voice and video communication between the two sites was maintained via a dedicated audiovisual conferencing system.

The preset network requirements were: round‐trip latency < 100 ms, with the minimum safe threshold set at < 200 ms; bandwidth > 100 Mbps; packet loss rate < 1%, with the minimum safe threshold < 2%; and network jitter < 10 ms. The day before each operation, system integration testing, network stress testing, audiovisual testing, and emergency disconnection drills were performed.

Recorded network metrics included mean round‐trip latency, maximum latency, packet loss rate, jitter, number of network interruptions, intraoperative link switching events, and whether the local team had to take over because of a network problem. The safety of remote robotic surgery depends not only on average latency but also on peak latency, jitter, packet loss rate, and the stability of video feedback. Previous studies suggest that communication latency can affect telerobotic performance and surgeon workload [13].

2.5. Preoperative Preparation and Team Division

All patients underwent a preoperative remote multidisciplinary assessment involving the remote surgeon, the local surgical team, an anaesthesiologist, an operating room nurse, and, when necessary, specialists in imaging, oncology, or critical care. Preoperative evaluations jointly reviewed imaging studies, laboratory results, anaesthetic risk, surgical indications, port placement plans, contingency plans, and postoperative management protocols.

A standardized checklist was executed for every procedure, including robotic system calibration, network connectivity testing, audiovisual system testing, preparation of surgical instruments and consumables, readiness of open surgical instruments, and confirmation of the local emergency surgical team. Intraoperative safety checks were performed before skin incision, before initiating remote manipulation, before major vascular transactions, and before specimen extraction. In the event of a network failure, robotic system anomaly, uncontrollable bleeding, or any situation compromising patient safety, the local team was to take over immediately and convert to conventional laparoscopy or open surgery as appropriate.

2.6. Surgical Technique

All operations were performed under general anaesthesia, using either total intravenous anaesthesia or combined intravenous–inhalational anaesthesia. Pneumoperitoneum pressure was maintained at 12–14 mmHg. Patient positioning, port placement, and robot docking were carried out by the Aksu patient‐side team according to the pre‐operative plan. The remote operating surgeon completed the key robotic steps from the console in Hangzhou. The patient‐side assistant performed suction, retraction, clipping, stapler assistance, specimen handling, and instrument exchange as required. The intraoperative steps for each specialty followed the standard principles of the corresponding robot‐assisted laparoscopic procedure.

2.7. Structured Training Programme

Concurrently, eight robotic surgery teams at the First People's Hospital of Aksu Prefecture underwent a structured training programme. Each team consisted of one lead surgeon, one assistant, and one operating room nurse, totaling 24 participants.

The training comprised 8 hours of didactic lectures and staged clinical mentoring during actual cases. Training outcomes were evaluated using the OSATS score, a widely used tool for objective assessment of surgical technical skill that evaluates dimensions such as tissue handling, instrument use, operative flow, and overall technical performance [16]. In this study, the OSATS score was converted to a 100‐point scale, and a score of ≥ 80 was considered satisfactory. OSATS scores were independently assigned by two trained evaluators who were blinded to the training phase.

2.8. Outcome Measures

The primary endpoint was the technical success rate. Technical success was defined as completion of the planned procedure under remote robotic control without conversion to open surgery, conversion to conventional local laparoscopy, or termination of remote robotic manipulation due to network, equipment, or intraoperative safety issues. If failover between network links occurs without causing perceptible delay or interruption in surgical control, the case is still considered a technical success. Secondary endpoints included total operative time, robotic console time, estimated intraoperative blood loss, intraoperative complications, conversion rate, length of postoperative hospital stay, 30‐day postoperative complications according to the Clavien–Dindo classification [17], late complications during follow‐up, network performance metrics, and changes in OSATS scores before and after training.

2.9. Statistical Analysis

Statistical analysis was performed using SPSS 26.0. Continuous variables that were normally or approximately normally distributed are presented as mean ± standard deviation, while non‐normally distributed variables are presented as median and range. Categorical variables are expressed as frequencies and percentages. Pre‐ and post‐training OSATS scores were compared using the paired t‐test. A two‐tailed p value < 0.05 was considered statistically significant.

3. Results

3.1. Patient Characteristics and Operative Outcomes

A total of 21 consecutive patients underwent cross‐regional 5G‐RRLS between October 2023 and September 2025. All procedures were completed as planned, yielding a technical success rate of 100%. No procedure required conversion to open or conventional laparoscopic surgery, and no procedure was terminated because of network, equipment, or safety issues. The cohort included 14 male patients and 7 female patients, accounting for 66.7% and 33.3% of the cohort, respectively. The mean age was 56.4 ± 9.9 years, with a range of 38–75 years. The mean BMI was 24.5 ± 1.9 kg/m2. ASA physical status was classified as grade I in 8 patients (38.1%), grade II in 10 patients (47.6%), and grade III in 3 patients (14.3%).

The procedures included partial nephrectomy in 8 patients (38.1%), cholecystectomy in 4 patients (19.0%), total hysterectomy in 3 patients (14.3%), radical prostatectomy in 2 patients (9.5%), radical resection of colorectal cancer in 2 patients (9.5%), and liver segmentectomy in 2 patients (9.5%). The mean total operative time was 158.5 ± 42.0 min, with a median of 165 min and a range of 88–225 min. The mean console time was 91.7 ± 26.6 min, with a median of 92 min and a range of 50–140 min. The mean estimated blood loss was 69.8 ± 40.2 mL, with a median of 55 mL and a range of 20–150 mL. The mean postoperative hospital stay was 5.2 ± 1.5 days, with a median of 5 days and a range of 3–8 days. No intraoperative complications occurred. Two patients experienced 30‐day postoperative complications, including one wound infection classified as Clavien‐Dindo grade I and one fever classified as Clavien‐Dindo grade II, corresponding to an overall 30‐day complication rate of 9.5%. The mean follow‐up duration was 13.0 ± 3.9 months, with a range of 6–18 months. Table 1 summarises the detailed patient information.

TABLE 1.

Individual patient data for 21 cases of 5G‐enabled remote robot‐assisted laparoscopic surgery.

ID Age (y) Gender BMI ASA Surgery OPT COT EBL ITOC PHS 30‐day complication Follow‐up (months)
1 62 Male 23.8 II Partial nephrectomy 165 92 50 None 5 None 12
2 48 Female 22.1 I Cholecystectomy 98 55 30 None 3 None 18
3 55 Male 26.3 II Partial nephrectomy 180 105 80 None 6 Wound infection (CD Grade I) 10
4 70 Male 24.9 III Radical prostatectomy 210 125 120 None 7 None 8
5 42 Female 21.4 I Total hysterectomy 135 80 40 None 4 None 15
6 58 Male 27.1 II Partial nephrectomy 175 100 65 None 5 None 11
7 67 Male 25.5 III Radical resection of colorectal cancer 225 140 150 None 8 Fever (CD Grade II) 9
8 51 Female 23.2 I Cholecystectomy 88 50 20 None 3 None 18
9 63 Male 26.8 II Partial nephrectomy 190 110 85 None 5 None 13
10 45 Female 20.9 I Total hysterectomy 118 68 35 None 4 None 16
11 59 Male 25.7 II Partial nephrectomy 162 90 55 None 6 None 12
12 75 Male 24.1 III Segmentectomy (liver) 195 115 130 None 7 None 6
13 38 Female 21.7 I Cholecystectomy 92 52 25 None 3 None 18
14 54 Male 27.4 II Partial nephrectomy 168 95 70 None 5 None 14
15 60 Female 24.6 II Total hysterectomy 128 72 45 None 5 None 17
16 68 Male 25.3 II Cholecystectomy 105 58 30 None 4 None 15
17 53 Male 26.1 II Segmentectomy (liver) 178 105 95 None 6 None 11
18 49 Female 22.5 I Radical resection of colorectal cancer 205 120 110 None 8 None 10
19 66 Male 24.4 II Radical prostatectomy 215 130 140 None 7 None 7
20 44 Male 23.9 I Partial nephrectomy 142 78 40 None 4 None 18
21 57 Male 25.9 I Partial nephrectomy 155 85 50 None 5 None 16

Abbreviations: BMI: Body Mass Index, kg/m2; CD: Clavien‐Dindo classification; COT: Console time, min; EBL: Estimated blood loss, mL; ITOC: Intraoperative complication; OPT: Operative time, min; PHS: Postoperative hospital stays, days.

3.2. Network Performance

Network performance remained stable throughout all 21 procedures (Table 2). The mean round‐trip latency was 37.7 ± 5.6 ms, with a median of 38 ms and a range of 28–47 ms. The mean maximum latency was 47.5 ± 5.4 ms, with a median of 48 ms and a range of 38–55 ms. The mean jitter was 2.2 ± 0.5 ms, with a median of 2.1 ms and a range of 1.5–3.3 ms. No packet loss or network interruption occurred during any procedure.

TABLE 2.

Individual network performance data for 21 surgeries.

ID Mean round‐trip latency (ms) Maximum latency (ms) Packet loss rate (%) Jitter (ms) Interruption
1 35 44 0 2.1 None
2 42 52 0 1.8 None
3 38 48 0 2.4 None
4 45 55 0 2.5 None
5 33 42 0 1.9 None
6 30 40 0 2.0 None
7 46 53 0 3.1 None
8 41 50 0 2.2 None
9 37 47 0 2.3 None
10 28 38 0 1.5 None
11 36 45 0 2.0 None
12 43 55 0 2.8 None
13 39 49 0 2.1 None
14 31 43 0 1.7 None
15 40 51 0 2.4 None
16 44 54 0 2.9 None
17 38 48 0 2.2 None
18 47 55 0 3.3 None
19 34 44 0 2.0 None
20 29 39 0 1.6 None
21 36 46 0 2.0 None

3.3. Training Outcomes

A total of 24 local trainees from 8 surgical teams completed the structured training programme, including surgeons, assistants, and nurses. The mean OSATS score improved from 64.6 ± 7.1 before training to 87.9 ± 4.6 after training, with a mean increase of 23.3 ± 2.8 points. These results suggest that the cross‐regional 5G‐RRLS programme was associated not only with successful clinical implementation but also with measurable improvement in local team technical performance. Individual score data are presented in Table 3.

TABLE 3.

Individual pre‐ and post‐training OSATS scores for 24 trainees.

Trainee ID Team Role Pre‐training OSATS Post‐training OSATS
1 1 Surgeon 68 90
2 1 Assistant 72 92
3 1 Nurse 55 82
4 2 Surgeon 65 88
5 2 Assistant 70 91
6 2 Nurse 58 85
7 3 Surgeon 62 87
8 3 Assistant 74 93
9 3 Nurse 60 84
10 4 Surgeon 58 83
11 4 Assistant 69 89
12 4 Nurse 52 80
13 5 Surgeon 71 94
14 5 Assistant 66 88
15 5 Nurse 63 86
16 6 Surgeon 75 95
17 6 Assistant 68 90
18 6 Nurse 57 83
19 7 Surgeon 64 89
20 7 Assistant 73 93
21 7 Nurse 59 84
22 8 Surgeon 61 86
23 8 Assistant 77 96
24 8 Nurse 54 81

Note: The original total score of 35 was converted to a 100‐point scale as follows: 100‐point score = (raw total/35) × 100.

Abbreviation: OSATS, Objective Structured Assessment of Technical Skill.

4. Discussion

This study demonstrates that, with stringent patient selection, standardized preoperative preparation, a redundant 5G dedicated‐line network, and a competent patient‐side team, 5G‐RRLS over approximately 3700 patients is technically feasible. All 21 procedures were completed without conversion to open surgery, intraoperative complications, or network interruptions, and no major (≥Grade III) postoperative complications occurred. Network performance was generally stable, with a mean round‐trip latency of 37.7 ms, a maximum latency of 55 ms, zero packet loss, and low jitter. Meanwhile, the structured training programme significantly improved the local team's OSATS scores.

The clinical results should first be interpreted as an initial feasibility assessment and cannot be directly equated with outcomes that are comparable to those of conventional onsite robotic surgery. The absence of intraoperative complications and the occurrence of only two mild‐to‐moderate postoperative complications are encouraging, but the sample size is small and the patients were carefully selected, indicating that the overall surgical risk was relatively well controlled. Robotic surgery has been applied extensively across several specialties, and previous randomized and large‐scale clinical studies have shown its clear clinical value in prostate cancer, rectal cancer, and other fields; however, learning curves, risk profiles, and endpoint definitions differ substantially among procedures [18, 19, 20]. The present series included urological, hepatobiliary, gynaecological, and gastrointestinal procedures. While this heterogeneity suggests multi‐specialty potential for the remote robotic platform, it also limits in‐depth evaluation of the efficacy and safety of any specific procedure.

Network reliability is central to the safe performance of remote robotic surgery. Unlike general telemedicine consultation or telementoring, even brief communication anomalies during telesurgery can affect instrument control, visual feedback, and intraoperative decision‐making. Therefore, evaluating the safety of a remote surgical network requires attention not only to average latency but also to peak latency, jitter, packet loss rate, video transmission delay, link‐switching capability, and whether the surgeon perceives control lag. Previous studies have shown that communication latency can impair telerobotic performance and increase the cognitive workload of the surgeon [9, 21]. Accordingly, the assessment of network safety should extend beyond mean latency to include peak latency, jitter, packet loss rate, and the stability of visual feedback. In the current series, all network metrics remained within the preset safety ranges, and no clinically perceptible image freezing or operative delay occurred. Low jitter and zero packet loss may have been important factors in maintaining instrument movement continuity and stable video feedback.

The favourable network performance observed here may be attributable to the 5G dedicated lines, QoS assurance, and multi‐carrier redundant links. Compared with public internet connections, private medical lines can reduce the risks of network congestion, routing fluctuations, and unpredictable packet loss. In recent years, 5G‐RRLS has been reported in spinal surgery, thoracic surgery, gastrointestinal surgery, and urology, indicating that 5G networks can support real‐time telesurgery under defined conditions [7, 8, 9, 10]. Nonetheless, remote surgery requires not only low latency but also fault tolerance. A relatively safe remote surgical network architecture should therefore include preoperative stress testing, real‐time network monitoring, clearly defined link‐switching protocols, executable disconnection contingency plans, and local team takeover capability.

Another point that warrants emphasis is that the safety of remote surgery does not depend solely on the remote operating surgeon but heavily relies on the patient‐side team as well. Although the remote surgeon performs the core robotic manipulations, the patient‐side team remains responsible for anaesthesia, patient positioning, port placement, robot docking, instrument exchange, suction and retraction, specimen extraction, and intraoperative crisis management. In the event of bleeding, equipment malfunction, or communication loss, the local team must be able to stabilise the patient and convert to conventional laparoscopy or open surgery. Thus, remote robotic surgery should not be understood as a replacement for local surgical capability but rather as a structured collaboration between a remote expert centre and the local team.

The training component of this study may represent an important contribution that distinguishes it from reports that focus exclusively on remote surgery completion. Many telesurgery studies primarily examine whether the procedure can be performed across a distance, with less attention given to whether such activities foster sustainable local technical capacity. Surgical skills training research indicates that structured curricula, simulation, and objective assessment improve technical performance and shorten the learning curve [16, 22, 23]. Moreover, the association between surgeon technical skill and postoperative complication rates further underscores the importance of standardized training and objective evaluation [24]. In this programme, didactic teaching, simulation, animal experiments, graded clinical mentoring, and phased delegation were integrated and validated with OSATS scoring. The significant improvement in the local team's OSATS score after training suggests that the remote robotic platform can serve not only as a vehicle for single‐procedure assistance but also as a tool for continuous surgical training and skill transfer.

From a health system perspective, traditional short‐term expert visits or itinerant surgery can meet some patient needs over a short period but may have limited long‐term impact if continuous training and local team building are lacking. In contrast, the remote robotic surgery model allows experts to participate in surgical guidance and team development without frequent cross‐regional travel, theoretically improving the continuity of skill transfer and reducing waiting times for advanced minimally invasive surgery in remote areas. However, whether this model is superior to conventional approaches in terms of economic cost, patient experience, expert workload, hospital management, and long‐term skill retention requires further investigation.

In recent years, telemedicine has experienced exponential growth, with its clinical applications extending well beyond telephone triage and ambulatory e‐visits to encompass telesurgery, telerehabilitation, remote intensive care, and chronic disease management [25, 26]. The ethical, legal, and regulatory dimensions of remote robotic surgery should not be overlooked. This technology involves cross‐regional licencing, the division of medical liability, attribution of liability for network failures, informed consent, data security, surgical video storage, remote device control authorisation, and quality assurance. Ethical guidelines for telemedicine stress that patients should be fully informed about the nature of the tele‐service, its potential risks, alternative options, and privacy protections [27, 28]. In the context of remote surgery, additional clarity is needed regarding cross‐regional practice privileges, liability allocation, network failure attribution, and data security management [29, 30]. Patients should also explicitly understand that the operating surgeon is not physically present in the operating room and that, in an emergency, the local team will assume control. Institutions should establish written policies before initiating a remote surgery programme that clearly define the responsibilities of the remote surgeon, the local surgeon, the anaesthesiologist, the robotic device manufacturer, and the network service provider.

This study has several limitations. First, it is a retrospective case series with a small sample size. Second, all cases came from a single inter‐institutional collaborative project, limiting external generalisability. Third, multiple specialties and procedures were included, precluding reliable procedure‐specific analyses. Fourth, the patients were strictly selected—emergency cases, severe obesity, severe cardiopulmonary dysfunction, and suspected extensive adhesions were excluded—so the results cannot be extrapolated to high‐risk or complex cases. Fifth, no control group of conventional onsite robotic surgery or traditional laparoscopic surgery was available for comparison. Sixth, training outcomes were mainly evaluated by OSATS scores, and long‐term competence, learning curves, and the quality of independently performed local cases have not yet been tracked longitudinally. Seventh, health‐economic indicators, patient‐reported outcomes, operator workload, and team psychological stress were not systematically assessed. Finally, network metrics need to be reported with finer technical granularity in future studies, including sampling frequency, video latency, control‐signal latency, and link‐switching performance.

5. Conclusions

In this selected cohort, cross‐regional 5G‐RRLS was technically feasible and was not associated with clinically significant perioperative complications. A stable low‐latency network, standardized preoperative workflows, and a patient‐side team that is prepared to take over in an emergency are essential prerequisites for safe implementation. Larger prospective, multicenter studies are needed to confirm the clinical efficacy, cost‐effectiveness, and generalisability of this model, and to evaluate its role in training and sustaining local robotic surgical proficiency.

Author Contributions

D.P., Y.G. and D.H.Z. conceived and designed the study. D.P. and Y.G. collected the clinical and network data and performed the statistical analysis. D.P. drafted the original manuscript. D.P. and Y.G. critically reviewed and edited the manuscript. D.H.Z. supervised the overall project. All authors read and approved the final manuscript, and agreed to be accountable for all aspects of the work.

Funding

The authors have nothing to report.

Ethics Statement

This study was a retrospective case series and was conducted in accordance with the Declaration of Helsinki. Ethical approval was obtained from the Ethics Committee of the First People's Hospital of Aksu Prefecture (No.: 2023‐K‐046) and was registered with the Ethics Committee of Zhejiang Provincial People's Hospital.

Consent

Written informed consent for both robotic surgery and the specific remote surgical procedure was obtained from all individual participants included in the study.

Conflicts of Interest

The authors declare no conflicts of interest.

Acknowledgements

We thank all participating patients and their families for their trust in this technology. We also acknowledge the surgical staff, anesthesiologists, and operating room nurses at the First People's Hospital of Aksu Prefecture for their indispensable contributions.

Data Availability Statement

The data analyzed during the current study are available within the article.

References

  • 1. Meara J. G., Leather A. J., Hagander L., et al., “Global Surgery 2030: Evidence and Solutions for Achieving Health, Welfare, and Economic Development,” Lancet 386, no. 9993 (2015): 569–624, 10.1016/S0140-6736(15)60160-X. [DOI] [PubMed] [Google Scholar]
  • 2. Alkire B. C., Raykar N. P., Shrime M. G., et al., “Global Access to Surgical Care: A Modelling Study,” Lancet Global Health 3, no. 6 (2015): e316–e323, 10.1016/S2214-109X(15)70115-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Shrime M. G., Sleemi A., and Ravilla T. D., “Charitable Platforms in Global Surgery: A Systematic Review of Their Effectiveness, Cost‐Effectiveness, Sustainability, and Role Training,” World Journal of Surgery 39, no. 1 (2015): 10–20, 10.1007/s00268-014-2516-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Lanfranco A. R., Castellanos A. E., Desai J. P., and Meyers W. C., “Robotic Surgery: A Current Perspective,” Annals of Surgery 239, no. 1 (2004): 14–21, 10.1097/01.sla.0000103020.19595.7d. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Peters B. S., Armijo P. R., Krause C., Choudhury S. A., and Oleynikov D., “Review of Emerging Surgical Robotic Technology,” Surgical Endoscopy 32, no. 4 (2018): 1636–1655, 10.1007/s00464-018-6079-2. [DOI] [PubMed] [Google Scholar]
  • 6. Marescaux J., Leroy J., Gagner M., et al., “Transatlantic Robot‐Assisted Telesurgery,” Nature 413, no. 6854 (2001): 379–380, 10.1038/35096636. [DOI] [PubMed] [Google Scholar]
  • 7. Anvari M., McKinley C., and Stein H., “Establishment of the World's First Telerobotic Remote Surgical Service: For Provision of Advanced Laparoscopic Surgery in a Rural Community,” Annals of Surgery 241, no. 3 (2005): 460–464, 10.1097/01.sla.0000154456.69815.ee. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Nguan C. Y., Morady R., Wang C., et al., “Robotic Pyeloplasty Using Internet Protocol and Satellite Network‐Based Telesurgery,” International Journal of Medical Robotics 4, no. 1 (2008): 10–14, 10.1002/rcs.173. [DOI] [PubMed] [Google Scholar]
  • 9. Haidegger T., Sándor J., and Benyó Z., “Surgery in Space: The Future of Robotic Telesurgery,” Surgical Endoscopy 25, no. 3 (2011): 681–690, 10.1007/s00464-010-1243-3. [DOI] [PubMed] [Google Scholar]
  • 10. Lacy A. M., Bravo R., Otero‐Piñeiro A. M., et al., “5G‐Assisted Telementored Surgery,” British Journal of Surgery 106, no. 12 (2019): 1576–1579, 10.1002/bjs.11364. [DOI] [PubMed] [Google Scholar]
  • 11. Tian W., Fan M., Zeng C., Liu Y., He D., and Zhang Q., “Telerobotic Spinal Surgery Based on 5G Network: The First 12 Cases,” Neurospine 17, no. 1 (2020): 114–120, 10.14245/ns.1938454.227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Zheng J., Wang Y., Zhang J., et al., “5G Ultra‐Remote Robot‐Assisted Laparoscopic Surgery in China,” Surgical Endoscopy 34, no. 11 (2020): 5172–5180, 10.1007/s00464-020-07823-x. [DOI] [PubMed] [Google Scholar]
  • 13. Zhou L. P., Zhang X. L., Zhang H. Q., et al., “5G‐Enabled Remote Robotic‐Assisted Percutaneous Pedicle Screw Fixation in Single Thoracolumbar Fractures: Initial Clinical Feasibility and Safety Evaluation,” European Spine Journal 35, no. 2 (2026): 944–954, 10.1007/s00586-025-09478-5. [DOI] [PubMed] [Google Scholar]
  • 14. von E. E., Altman D. G., Egger M., et al., “The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) Statement: Guidelines for Reporting Observational Studies,” Lancet 370, no. 9596 (2007): 1453–1457, 10.1016/S0140-6736(07)61602-X. [DOI] [PubMed] [Google Scholar]
  • 15. Agha R. A., Sohrabi C., Mathew G., et al., “The PROCESS 2020 Guideline: Updating Consensus Preferred Reporting of CasESeries in Surgery (PROCESS) Guidelines,” International Journal of Surgery 84 (2020): 231–235, 10.1016/j.ijsu.2020.11.005. [DOI] [PubMed] [Google Scholar]
  • 16. Martin J. A., Regehr G., Reznick R., et al., “Objective Structured Assessment of Technical Skill (OSATS) for Surgical Residents,” British Journal of Surgery 84, no. 2 (1997): 273–278, 10.1046/j.1365-2168.1997.02502.x. [DOI] [PubMed] [Google Scholar]
  • 17. Dindo D., Demartines N., and Clavien P. A., “Classification of Surgical Complications: A New Proposal With Evaluation in a Cohort of 6336 Patients and Results of a Survey,” Annals of Surgery 240, no. 2 (2004): 205–213, 10.1097/01.sla.0000133083.54934.ae. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Yaxley J. W., Coughlin G. D., Chambers S. K., et al., “Robot‐Assisted Laparoscopic Prostatectomy Versus Open Radical Retropubic Prostatectomy: Early Outcomes From a Randomised Controlled Phase 3 Study,” Lancet 388, no. 10049 (2016): 1057–1066, 10.1016/S0140-6736(16)30592-X. [DOI] [PubMed] [Google Scholar]
  • 19. Coughlin G. D., Yaxley J. W., Chambers S. K., et al., “Robot‐Assisted Laparoscopic Prostatectomy Versus Open Radical Retropubic Prostatectomy: 24‐Month Outcomes From a Randomised Controlled Study,” Lancet Oncology 19, no. 8 (2018): 1051–1060, 10.1016/S1470-2045(18)30357-7. [DOI] [PubMed] [Google Scholar]
  • 20. Jayne D., Pigazzi A., Marshall H., et al., “Effect of Robotic‐Assisted vs Conventional Laparoscopic Surgery on Risk of Conversion to Open Laparotomy Among Patients Undergoing Resection for Rectal Cancer: The ROLARR Randomized Clinical Trial,” JAMA 318, no. 16 (2017): 1569–1580, 10.1001/jama.2017.7219. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Ye Y., Zhou T., Zhu Q., Vann W., and Du J., “Brain Functional Connectivity Under Teleoperation Latency: A fNIRS Study,” Frontiers in Neuroscience 18 (2024): 1416719: Published 2024 Nov 13, 10.3389/fnins.2024.1416719. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Reznick R. K. and MacRae H., “Teaching Surgical Skills—changes in the Wind,” New England Journal of Medicine 355, no. 25 (December 2006): 2664–2669: PMID: 17182991, 10.1056/NEJMra054785. [DOI] [PubMed] [Google Scholar]
  • 23. Grantcharov T. P. and Reznick R. K., “Teaching Procedural Skills,” BMJ 336, no. 7653 (2008): 1129–1131, 10.1136/bmj.39517.686956.47. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Birkmeyer J. D., Finks J. F., O'Reilly A., et al., “Surgical Skill and Complication Rates After Bariatric Surgery,” New England Journal of Medicine 369, no. 15 (2013): 1434–1442, 10.1056/NEJMsa1300625. [DOI] [PubMed] [Google Scholar]
  • 25. Feng D., Shi X., Wang J., et al., “Telemedicine‐Medical “Snack Community”‐PHS Ecosystem: Insights Into the Double‐Edged Sword Role of Telemedicine in Clinical Practice and Medical Education During the COVID‐19 Pandemic and Beyond,” Exploration (Beijing) 4, no. 4 (2024): 20230111, 10.1002/EXP.20230111. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26. Li D., Wang J., Wang Z., et al., “Next‐Generation Surgery: Integrating Artificial Intelligence, Genetic Technologies, Bioengineering and Rehabilitation Into Modern Practices,” Exploration 6, no. 3 (2026): 20240417, 10.1002/EXP.20240417. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Chaet D., Clearfield R., Sabin J. E., and Skimming K., and Council on Ethical and Judicial Affairs American Medical Association , “Ethical Practice in Telehealth and Telemedicine,” Journal of General Internal Medicine 32, no. 10 (2017): 1136–1140, 10.1007/s11606-017-4082-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Stanberry B., “Legal and Ethical Aspects of Telemedicine,” Journal of Telemedicine and Telecare 12, no. 4 (2006): 166–175, 10.1258/135763306777488825. [DOI] [PubMed] [Google Scholar]
  • 29. Kichloo A., Albosta M., Dettloff K., et al., “Telemedicine, the Current COVID‐19 Pandemic and the Future: A Narrative Review and Perspectives Moving Forward in the USA,” Family Medicine and Community Health 8, no. 3 (2020): e000530, 10.1136/fmch-2020-000530. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Hollander J. E. and Carr B. G., “Virtually Perfect? Telemedicine for Covid‐19,” New England Journal of Medicine 382, no. 18 (2020): 1679–1681, 10.1056/NEJMp2003539. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data analyzed during the current study are available within the article.


Articles from The International Journal of Medical Robotics + Computer Assisted Surgery are provided here courtesy of Wiley

RESOURCES