Abstract
Objective
To systematically characterise United States Food and Drug Administration (FDA) authorised urology‐specific artificial intelligence (AI)‐enabled medical devices and to evaluate the publicly available evidence supporting their authorisation.
Methods
We reviewed the FDA AI–Enabled Medical Devices List. Devices cleared by September 2025 were manually screened for urological relevance using FDA clearance documents (e.g., 510(k), De Novo) to confirm clinical application. Eligible devices were categorised by FDA approval pathway, review panel, indication, and year of approval. FDA decision summaries were reviewed to characterise study design, training/testing and external validation features, and reporting of patient demographics. Descriptive statistics were used to summarise distributions and temporal trends.
Results
Of 1358 FDA‐cleared AI devices, 30 (2.2%) were urology specific. Approvals increased over time, with 26.7% cleared in 2025 (through September). The predominant FDA medical device category was Radiology (70.0%), with indications led by prostate imaging (43.3%). Information on study design, recruitment, and validation was frequently unavailable. Race and ethnicity were reported in one device (3.3%) for training/testing and four devices (13.3%) for external validation. Among these, aggregated proportions showed that the majority of subjects were White (86.1% for training/testing; 87.6% for external validation).
Conclusions
The FDA‐authorised urological AI devices are dominated by radiology‐based, prostate‐focused applications and have increased in recent years. However, publicly available evidence supporting their authorisation is often limited, with incomplete reporting of study design and patient demographics. Greater transparency and post‐authorisation evaluation are needed to support interpretation of these tools in clinical practice.
Keywords: artificial intelligence, urology, data reporting, representation, United States Food and Drug Administration
Abbreviations
- AI
artificial intelligence
- FDA
United States Food and Drug Administration
- PRISMA(‐ScR)
Preferred Reporting Items for Systematic Reviews and Meta‐Analyses (Extension for Scoping Reviews)
Introduction
Artificial intelligence (AI) applications in medicine have expanded rapidly in recent years and are increasingly incorporated into clinical care [1]. In urology, AI holds promise across multiple domains, including diagnostic imaging, pathology interpretation, risk stratification, and procedural planning [2, 3]. Urology has historically been an early adopter of new technologies, most notably robot‐assisted surgery, as well as advanced imaging and minimally invasive techniques [4, 5, 6]. Consistent with this tradition, there has been substantial interest in applying AI‐based tools to various aspects of urological care, with a growing number of products entering clinical and regulatory pipelines [2, 3].
Despite this rapid growth, the development and evaluation of AI‐enabled medical devices have drawn increasing attention to issues of transparency, validation, and generalisability. There is growing interest in clearer reporting of the data used to train, test, and validate AI systems, including study design, data sources, and patient characteristics [7, 8]. In 2021, regulatory agencies including the United States Food and Drug Administration (FDA), Health Canada, and the UK's Medicines and Healthcare Products Regulatory Agency issued guidance emphasising the importance of understanding how AI models are developed and evaluated, given implications for performance and clinical use [9]. Incomplete reporting of these elements can limit interpretation of the supporting evidence and how these tools may perform in routine clinical use.
To date, there has been no systematic evaluation of the landscape of FDA‐authorised AI‐enabled medical devices specific to urology, nor a focused examination of the publicly accessible evidence supporting their regulatory clearance or approval. In this study, we aimed to review FDA‐approved AI‐enabled medical devices in urological practice. Our objectives were to describe temporal trends, regulatory pathways, and clinical indications of these devices, and to characterise the publicly available data supporting their authorisation, with particular attention to study design features and reporting of patient demographic characteristics.
Methods
Data Source and Device Identification
We used the FDA AI–Enabled Medical Devices List as the primary data source. This publicly available resource identifies medical devices incorporating AI or machine learning that are authorised for marketing in the United States and is intended to promote transparency regarding the use of AI technologies in medical devices. The list includes devices that have met applicable FDA pre‐market requirements, including review of safety and effectiveness and evaluation of supporting evidence relative to the device's intended use. Each device entry links to the corresponding FDA database record, which contains publicly releasable information such as summaries of safety and effectiveness. Devices are identified by the FDA based on AI‐related terminology in public authorisation documents and device classifications, which may not represent a complete inventory of all AI‐enabled medical devices. We used the AI‐Enabled Medical Devices List last updated on 5 December 2025 and reviewed the list and associated FDA documentation between December 2025 and January 2026. The dataset included devices with FDA authorisation dates through to 30 September 2025.
Eligibility Criteria and Screening
All devices included in the FDA AI‐Enabled Medical Devices List were independently reviewed by two urologists (Z.Q. and P.B.) to determine relevance to urology. Devices were eligible for inclusion if their intended use was specific to a urological organ, disease, or clinical application within urological practice. These included devices designed for diagnosis, risk stratification, image interpretation, procedural guidance, or physiological assessment explicitly targeting the prostate, bladder, kidney, ureter, urethra, or male reproductive system. Devices were excluded if their intended use was not specific to urology or if they represented general‐purpose platforms applicable across multiple organ systems or disease states without a urological focus. For example, radiation oncology planning or contouring software intended for broad oncological use, without urology‐specific indications or labelling, was not included. In situations where there was inter‐reviewer disagreement about urology specificity, the discrepancies were reviewed and resolved.
Data Abstraction and Classification
For each eligible device, we abstracted the year of FDA authorisation, regulatory approval pathway (510(k) or De Novo), FDA review panel, and primary clinical application. Devices were categorised based on their stated intended use as described in FDA clearance documents, including 510(k) summaries and De Novo decision summaries, supplemented by publicly available manufacturer descriptions when necessary.
To investigate the demographic representation of the data supporting FDA authorisation for each device, we systematically reviewed FDA decision summaries to identify reported characteristics of the datasets used for training, testing, and external validation. Specifically, we abstracted information on study design (retrospective vs prospective), sample size, study setting (single centre vs multicentre), recruitment strategy, and availability of patient demographic data. Demographic variables of interest included age, sex, race, ethnicity, and education level, when reported. For sex reporting, devices limited to male‐specific clinical applications were identified and reporting proportions were calculated only among devices for which sex reporting was applicable. When demographic data were reported across multiple datasets for a single device, aggregated proportions were calculated.
Statistical Analysis
Descriptive statistics were used to summarise device characteristics, regulatory pathways, clinical indications, and features of the supporting evidence. Categorical variables were reported as counts and percentages. Continuous variables were summarised using means and ranges when available. Temporal trends in FDA approvals were assessed descriptively. The study used publicly available, de‐identified data and was therefore exempt from Institutional Review Board review. The P values were two‐sided and all statistics were analysed using Python 3.11 (Python Software Foundation, Wilmington, DE, USA). We abided by the Preferred Reporting Items for Systematic Reviews and Meta‐Analyses (PRISMA) Extension for Scoping Reviews (PRISMA‐ScR) reporting guideline.
Results
Overall Landscape of FDA‐Authorised Urological AI Devices
Among 1358 AI and machine learning‐enabled medical devices authorised by the FDA, 1328 were excluded for lack of urological relevance and 30 devices (2.2%) were clinically used in urological practice. The PRISMA flow diagram is included in the Supporting Information. FDA authorisation of urological AI devices increased over time, with approvals spanning from 2017 to 2025. Eight devices (26.7%) were authorised by the end of September in 2025 alone (Fig. 1). Most devices were authorised through the 510(k) pathway (28 devices [93.3%]), while two devices (6.7%) received De Novo authorisation. Devices were most commonly reviewed under the Radiology panel (21 devices [70.0%]), followed by Haematology (five devices [16.7%]), Pathology (three devices [10.0%]), and General and Plastic Surgery (one device [3.3%]) (Table 1). Details of included devices and resolved inter‐reviewer discrepancies are listed in Tables S1 and S2.
Fig. 1.

The FDA‐approved urology‐related AI device over time (cut‐off 30 September 2025).
Table 1.
Characteristics of urological AI and machine learning‐based medical devices cleared or approved by the FDA.
| Variable | N (%) |
|---|---|
| Year of approval | |
| 2017 | 2 (6.7) |
| 2018 | 1 (3.3) |
| 2019 | 1 (3.3) |
| 2020 | 5 (16.7) |
| 2021 | 2 (6.7) |
| 2022 | 3 (10.0) |
| 2023 | 3 (10.0) |
| 2024 | 5 (16.7) |
| 2025 (by September) | 8 (26.7) |
| Approval pathway | |
| 510(k) | 28 (93.3) |
| De Novo | 2 (6.7) |
| FDA review panel | |
| Radiology | 21 (70.0) |
| Haematology | 5 (16.7) |
| Pathology | 3 (10.0) |
| General and Plastic Surgery | 1 (3.3) |
| Indications | |
| Prostate MRI image contouring and cancer detection | 13 (43.3) |
| Prostate ultrasonography contouring and fusion | 2 (6.7) |
| Semen analyser | 4 (13.3) |
| Urine analyser | 1 (3.3) |
| Bladder urine volume estimation | 5 (16.7) |
| Prostate cancer pathology analyser and risk stratifier | 3 (10.0) |
| Bladder outlet treatment robotic system | 1 (3.3) |
| Kidney tumour cross sectional image contouring | 1 (3.3) |
| Total | 30 (100.0) |
Clinical Application of Authorised Devices
The most common clinical application involved prostate imaging. In all, 13 devices (43.3%) were indicated for prostate MRI contouring or prostate cancer detection, and two devices (6.7%) were designed for prostate ultrasonography contouring or image fusion. Additional indications included bladder urine volume estimation (five devices [16.7%]), semen analysis (four devices [13.3%]), prostate cancer pathology analysis and risk stratification (three devices [10.0%]), urine analysis (one device [3.3%]), kidney tumour cross‐sectional image contouring (one device [3.3%]), and robotic systems for bladder outlet treatment (one device [3.3%]) (Table 1).
Characteristics of Supporting Evidence
Publicly available reporting of study design and data sources supporting FDA authorisation varied across devices. Information on sample size for training or testing datasets was available for five devices (16.7%), with a mean (range) sample size of 10 755 (90–50 255) patients. For external validation datasets, sample size information was reported for 13 devices (43.3%), with a mean (range) of 210 (50–700) patients (Table 2). Study setting was frequently unreported in publicly available decision summaries. For training or testing datasets, one device (3.3%) reported single‐centre data, seven devices (23.3%) reported multicentre data, and 22 devices (73.3%) did not specify study setting. For external validation datasets, three devices (10.0%) reported single‐centre data, nine devices (30.0%) reported multicentre data, and 18 devices (60.0%) did not report study setting. Recruitment strategy was inconsistently described. Among training or testing datasets, 10 devices (33.3%) reported retrospective recruitment, no devices reported prospective recruitment, and 20 devices (66.7%) did not specify recruitment methods. For external validation datasets, 11 devices (36.7%) reported retrospective recruitment, one device (3.3%) reported prospective recruitment, and 18 devices (60.0%) did not report recruitment strategy.
Table 2.
Characteristics and data quality of accessible evidence supporting FDA clearance or approval of AI and machine learning‐based medical devices for urological diseases.
| Study design | Training/testing | External validation |
|---|---|---|
| Device | Device | |
| Devices with sample size information, n (%) | 5 (16.7) | 13 (43.3) |
| Sample size, median (IQR) | 1407 (137–1887) | 139 (81–227) |
| Study site, n (%) | ||
| Single centre | 1 (3.3) | 3 (10.0) |
| Multicentre | 7 (23.3) | 9 (30.0) |
| Unknown | 22 (73.3) | 18 (60.0) |
| Study recruitment, n (%) | ||
| Retrospective | 10 (33.3) | 11 (36.7) |
| Prospective | 0 (0.0) | 1 (3.3) |
| Unknown | 20 (66.7) | 18 (60.0) |
| Study demographics | Device, n (%) | Aggregated*, % | Device, n (%) | Aggregated*, % |
|---|---|---|---|---|
| Age, years, mean (range) | 1 (3.3) | 67 (67–67) | 4 (13.3) | 66.2 (65–67) |
| Sex † , n (%) | ||||
| Female | 2 (28.6) | 44.1 | 1 (14.2) | 69.0 |
| Male | 2 (28.6) | 55.9 | 1 (14.2) | 31.0 |
| Education | 0 (0.0) | 0 (0.0) | ||
| Race/Ethnicity, n (%) | ||||
| Asian | 1 (3.3) | 3.4% | 4 (13.3) | 1.4 |
| Black | 1 (3.3) | 8.6% | 4 (13.3) | 8.6 |
| Hispanic | 0 (0.0) | n/a | 1 (3.3) | 3.1 |
| White | 1 (3.3) | 86.1% | 4 (13.3) | 87.6 |
IQR, interquartile range.
Aggregated proportions were calculated for devices with reported sex, education, and race/ethnicity. Age was reported as mean and range.
In all, 23 of the 30 products in total were male patient specific, the percentage of reporting of patient sex was based on the remaining seven products.
Reporting of Patient Demographic Characteristics
Reporting of patient demographic characteristics in publicly available FDA decision summaries was limited across devices. Age was reported for one device (3.3%, mean age 67 years) in training or testing datasets and for four devices (13.3%, mean age 66.2 years) in external validation datasets. Among the seven devices that apply to both sexes, female patients comprised a mean of 44.1% of training or testing datasets and 69.0% of external validation datasets, respectively. Race and ethnicity were infrequently reported. For training or testing datasets, race or ethnicity information was available for one device (3.3%). For external validation datasets, race or ethnicity data were reported for four devices (13.3%). Within the one device reporting race and ethnicity for training/testing and four devices for external validation, aggregated proportions showed White subjects comprised 86.1% and 87.6%, respectively (Table 2).
Discussion
In this comprehensive review of FDA‐authorised urological AI‐enabled medical devices, we identified three key findings. First, AI devices in urology represented a small proportion of all FDA‐authorised AI devices, although the number of approvals increased over time, particularly in recent years. Second, authorised urological AI devices were predominantly radiology based and most commonly focused on prostate‐related clinical applications. Third, publicly accessible FDA decision summaries frequently lacked detailed reporting on study design, validation characteristics, and patient demographic composition. Collectively, our findings suggest the rapid growth of AI in urology that may outpace the requisite reporting and validation studies to support appropriate interpretation and use.
The small but growing presence of urological AI devices within the broader FDA AI ecosystem likely reflects both specialty‐specific factors and broader trends in medical AI development. Urology has historically adopted new technologies early, particularly those related to imaging, minimally invasive procedures, and workflow optimisation [5, 6, 10]. Prostate cancer care, which relies increasingly on imaging and standardised diagnostic pathways, represents a natural early target for AI‐based tools [6, 11, 12, 13]. The observed increase in approvals over time is consistent with expanding interest in AI across medicine and with increased regulatory familiarity with AI‐enabled technologies [1, 9]. We also observed that most authorised urological AI devices were reviewed under the Radiology panel and were focused on imaging‐based prostate applications. This pattern aligns with the relative maturity of AI methods in medical imaging and the availability of large, labelled imaging datasets suitable for model development. Similar trends have been reported across other specialties, where AI development has been most rapid under the Radiology device panel of the FDA, accounting for >75% of all approved devices [14]. In contrast, fewer devices were identified for non‐imaging urological applications, such as procedural guidance or functional assessment, which may reflect greater technical complexity or higher regulatory barriers.
Across devices, publicly available information describing the evidence supporting FDA authorisation was frequently limited in publicly accessible decision summaries. Many decision summaries did not report key features such as sample size, study setting, recruitment strategy, or validation approach, limiting external assessment of how these models were developed and evaluated. This lack of transparency is increasingly recognised as important in the AI literature, where model performance and reliability are known to depend strongly on the composition and structure of training and validation datasets [7, 8]. In particular, incomplete reporting of patient demographic characteristics is relevant given extensive prior work demonstrating that models trained on imbalanced or non‐representative data can propagate or amplify existing disparities [15, 16, 17]. Prior studies across medicine have shown that AI systems may underdiagnose disease or misestimate risk in racial and ethnic minority populations when those groups are underrepresented in development datasets [15, 16, 18]. In imaging‐ and pathology‐based applications, which comprised most of the devices in this study, concerns have also been raised regarding ‘shortcut learning’, in which models rely on spurious correlations or site‐specific artefacts rather than clinically meaningful features, a phenomenon that is more likely when training data are narrow, homogeneous, or poorly characterised [17, 19]. Similar challenges with transparency have been documented across various specialties [8, 20]. For example, within radiology devices, which comprises the largest share of FDA‐authorised AI devices, 95.5% lack open supporting evidence of clinical utility [21, 22, 23]. Urology, and prostate cancer in particular, is a field with well‐documented racial disparities in incidence, stage at diagnosis, treatment, and outcomes [24, 25]. Among the limited devices that did report race and ethnicity, White individuals comprised 86.1% of training/testing datasets and 87.6% of external validation datasets—notably higher than the ~75% White composition of the United States population per the 2020 Census, raising concerns of overrepresentation even within the small subset of devices that disclosed demographic data [26]. Together, limited demographic reporting and the well‐documented risk of bias in medical AI models calls for greater transparency to support interpretation of these tools across diverse urological populations. We need collaborative and ongoing efforts to generate real‐world evidence on how these technologies perform once deployed in clinical care.
The transparency gaps identified in this study carry particular significance in the context of evolving reimbursement policy. Recent legislative efforts have sought to facilitate Medicare payment for FDA‐authorised digital and AI‐enabled medical devices, including temporary reimbursement pathways intended to support early adoption while cost data are generated [27, 28]. Expedited coverage will translate directly into increased healthcare spending, and financial incentives created by reimbursement will likely intensify industry efforts to drive practitioner adoption. In this environment, tools with incompletely characterised evidence bases may become embedded in routine clinical workflows not because of demonstrated value, but because of financial momentum. If those tools carry undetected bias or perform poorly in underrepresented populations, the consequences extend beyond wasted expenditure—patients may be harmed by diagnostic errors or suboptimal risk stratification driven by algorithms that were never rigorously validated in populations like them. This underscores the need for robust post‐authorisation surveillance and transparent evidence standards before large‐scale reimbursement decisions are made.
This study has several limitations. First, our analysis was based on publicly available FDA documentation, including decision summaries and device listings, which do not capture the full technical or clinical data submitted as part of regulatory review. As a result, the amount and quality of evidence available to us likely underestimate what was evaluated by the FDA during authorisation. Second, the FDA AI‐Enabled Medical Devices List is not intended to be a comprehensive inventory of all AI‐enabled devices and relies on identification through AI‐related terminology in public summaries and classifications, which may lead to under‐sampling of all relevant products. Third, most urological AI devices were authorised through the 510(k) pathway, which is based on substantial equivalence to previously cleared predicate devices; we did not evaluate the underlying evidence supporting those predicate devices. Fourth, this study focused on regulatory authorisation rather than real‐world implementation, and we did not evaluate post‐market use, uptake, or patient outcomes. Urological AI devices currently represent only 2.2% of all FDA‐authorised AI devices, reflecting the early stage of this field, with further growth anticipated in coming years. Future studies are needed to address these limitations.
Conclusion
Our study provides an overview of FDA‐authorised urological AI devices and highlights critical gaps in publicly accessible supporting evidence. Mandatory demographic reporting and details on algorithm and device development in FDA decision summaries and standardised public reporting templates for AI authorisation documents should be considered to strengthen the evidence base supporting regulatory authorisation. As urological AI proliferates and reimbursement pathways expand, ensuring that authorised tools are transparent, reproducible, and representative of the populations they serve becomes increasingly critical.
Disclosure of Interests
Quoc‐Dien Trinh reports consulting fees from Astellas, Bayer, Intuitive Surgical, Janssen, Novartis, Pfizer, and research funding from the American Cancer Society, Pfizer Global Medical Grants (Prostate Cancer Disparities #63354905). Quoc‐Dien Trinh reports funding from Health Disparity Research Award from the Department of Defence Congressionally Directed Medical Research Program (#PC220551). Francesco Giganti reports consulting fees from DeepHealth, SpectraCure and Procept outside of the submitted work. He has received speaker fees from Bayer and Siemens. Alexander P. Cole reports research funding from the Bruce A. Beal and Robert L. Beal surgical fellowship of the BWH Department of Surgery, from the Prostate Cancer Foundation and American Cancer Society (#23YOUN25) and from a Physician Research Award from the Department of Defence Congressionally Directed Medical Research Program (#PC220342). The remaining authors have no disclosures.
Funding
Not applicable.
Supporting information
Table S1. Complete list of FDA‐authorised urological AI‐enabled medical devices included in this review.
Table S2. List of excluded FDA‐authorised AI‐enabled medical devices upon inter‐reviewer discussion and reason for exclusion.
[Correction added on 13 July 2026, after first online publication: The author affiliations have been revised in this version.]
Data Availability Statement
Data used in this study are publicly available at https://www.fda.gov/medical‐devices/software‐medical‐device‐samd/artificial‐intelligence‐enabled‐medical‐devices.
References
- 1. Haug CJ, Drazen JM. Artificial intelligence and machine learning in clinical medicine, 2023. N Engl J Med 2023; 388: 1201–1208 [DOI] [PubMed] [Google Scholar]
- 2. Suarez‐Ibarrola R, Hein S, Reis G, Gratzke C, Miernik A. Current and future applications of machine and deep learning in urology: a review of the literature on urolithiasis, renal cell carcinoma, and bladder and prostate cancer. World J Urol 2020; 38: 2329–2347 [DOI] [PubMed] [Google Scholar]
- 3. Abid R, Hussein AA, Guru KA. Artificial intelligence in urology: current status and future perspectives. Urol Clin North Am 2024; 51: 117–130 [DOI] [PubMed] [Google Scholar]
- 4. Bhatt NR, Davis NF, Dalton DM et al. Quantitative analysis of technological innovation in urology. Urology 2018; 111: 230–237 [DOI] [PubMed] [Google Scholar]
- 5. Korn SM, Qian Z, Zurl H et al. Claims‐based analysis of national utilization patterns and regional variability of prostate cancer ablative therapies: a medicare study. J Endourol 2026; 40: 72–78 [DOI] [PubMed] [Google Scholar]
- 6. Korn SM, Qian Z, Zurl H et al. Geographic variability in contemporary utilization of PET imaging for prostate cancer: a medicare claims cohort study. Cancer Imaging 2025; 25: 86 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Mehta V, Komanduri A, Bhadouriya RS et al. Evaluating transparency in AI/ML model characteristics for FDA‐reviewed medical devices. NPJ Digit Med 2025; 8: 673 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Lin JC, Jain B, Iyer JM et al. Benefit‐risk reporting for FDA‐cleared artificial intelligence−enabled medical devices. JAMA Health Forum 2025; 6: e253351‐e [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Warraich HJ, Tazbaz T, Califf RM. FDA perspective on the regulation of artificial intelligence in health care and biomedicine. JAMA 2025; 333: 241–247 [DOI] [PubMed] [Google Scholar]
- 10. Biasatti A, Soputro NA, Porpiglia F et al. The current landscape of single‐port robotic surgery in urology. Nat Rev Urol 2025; 23: 156–173 [DOI] [PubMed] [Google Scholar]
- 11. Goldenberg SL, Nir G, Salcudean SE. A new era: artificial intelligence and machine learning in prostate cancer. Nat Rev Urol 2019; 16: 391–403 [DOI] [PubMed] [Google Scholar]
- 12. Korn SM, Qian Z, Zurl H et al. Geographic variability in germline genetic testing for prostate cancer: a Nationwide Medicare claims study with evidence of low uptake. J Natl Compr Cancer Netw 2025; 23: e257074 [DOI] [PubMed] [Google Scholar]
- 13. Qian Z, Chen Y‐J, Feldman J et al. Prostate magnetic resonance imaging utilization and its relationship with advanced prostate cancer detection. Urol Oncol Semin Orig Investig 2024; 42: 370‐e1 [DOI] [PubMed] [Google Scholar]
- 14. Sivakumar R, Lue B, Kundu S. FDA approval of artificial intelligence and machine learning devices in radiology: a systematic review. JAMA Netw Open 2025; 8: e2542338‐e [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Seyyed‐Kalantari L, Zhang H, McDermott MBA, Chen IY, Ghassemi M. Underdiagnosis bias of artificial intelligence algorithms applied to chest radiographs in under‐served patient populations. Nat Med 2021; 27: 2176–2182 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Ferryman K, Mackintosh M, Ghassemi M. Considering biased data as informative artifacts in AI‐assisted health care. N Engl J Med 2023; 389: 833–838 [DOI] [PubMed] [Google Scholar]
- 17. Colacci M, Huang YQ, Postill G et al. Sociodemographic bias in clinical machine learning models: a scoping review of algorithmic bias instances and mechanisms. J Clin Epidemiol 2025; 178: 111606 [DOI] [PubMed] [Google Scholar]
- 18. Meng C, Trinh L, Xu N, Enouen J, Liu Y. Interpretability and fairness evaluation of deep learning models on MIMIC‐IV dataset. Sci Rep 2022; 12: 7166 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Banerjee I, Bhattacharjee K, Burns JL et al. “Shortcuts” causing bias in radiology artificial intelligence: causes, evaluation, and mitigation. J Am Coll Radiol 2023; 20: 842–851 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Chen KY, Ross JS, Cohen AB, Karlawish J, Oh ES, Gupta R. Demographic data supporting FDA authorization of AI devices for Alzheimer disease and related dementias. JAMA 2025; 334: 1196–1199 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Windecker D, Baj G, Shiri I et al. Generalizability of FDA‐approved AI‐enabled medical devices for clinical use. JAMA Netw Open 2025; 8: e258052 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. van Leeuwen KG, Schalekamp S, Rutten MJ, van Ginneken B, de Rooij M. Artificial intelligence in radiology: 100 commercially available products and their scientific evidence. Eur Radiol 2021; 31: 3797–3804 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Antonissen N, Tryfonos O, Houben IB, Jacobs C, de Rooij M, van Leeuwen KG. Artificial intelligence in radiology: 173 commercially available products and their scientific evidence. Eur Radiol 2025; 36: 526–536 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Qian Z, Alexander J, Daniels D et al. Racial differences in postpandemic trends in prostate‐specific antigen screening. JNCI Cancer Spectr 2024; 8: pkae016 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Zurl H, Qian Z, Pohl KK et al. Rising metastatic prostate cancer rates but narrowing racial gap. BMC Med 2025; 23: 1–10 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Dicennial Census . US Census Bureau US Census Bureau. 2020.
- 27. Moneer O, Kadakia KT, Dhruva SS et al. Hospital acquisition costs for breakthrough‐designated devices awarded supplemental medicare reimbursement. JAMA Health Forum 2025; 6: e255183‐e [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Fleisher LA. Reimbursement for AI—balancing innovation, adoption, and the Medicare trust fund. JAMA Cardiol 2025; 10: 647–648 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1. Complete list of FDA‐authorised urological AI‐enabled medical devices included in this review.
Table S2. List of excluded FDA‐authorised AI‐enabled medical devices upon inter‐reviewer discussion and reason for exclusion.
Data Availability Statement
Data used in this study are publicly available at https://www.fda.gov/medical‐devices/software‐medical‐device‐samd/artificial‐intelligence‐enabled‐medical‐devices.