Abstract
Although measurement of thyroid hormones (THs), TSH and thyroid autoantibodies, is usually straightforward, erroneous results are occasionally generated, raising the possibility of incorrect or delayed diagnoses, unwarranted investigations, and provision of inappropriate treatments. Assay interference can be acquired (e.g. antibodies), genetic (e.g. variant TH-binding proteins), or pharmaceutical (e.g. displacing agents and biotin) in nature. In those with discordant thyroid function, assay interference must be excluded prior to the diagnosis of rare inherited (resistance to thyroid hormone) or tumoural conditions (TSHoma). Liaison between the clinician and clinical biochemistry teams is essential when assay interference is being investigated; appropriate further biochemical testing depends on the clinical picture, initial biochemistry, and local biochemical expertise. In a subset of cases, access to a regional centre where patients can be clinically assessed, in conjunction with performance of specialised clinical laboratory testing (biochemical, +/− genetic), may be required. When assay interference is identified, clear communication with the patient (ideally aided by reliable, accessible patient literature) and documentation to their physician are necessary. Professionals must stay updated on advances in testing and be alert for evolving assay interferences. An accompanying plain language summary is intended to provide an accessible overview of the key messages and recommendations contained in this guideline for non-specialist readers, including patients, carers, healthcare professionals outside the speciality, and the wider public.
Keywords: thyroid function tests, free T4, immunoassays, assay interference
Methodology and grading
The executive committee of the European Thyroid Association (ETA), following consultation with its guidelines and publications committee, commissioned the development of this guideline, led by one chairperson (CM). The guideline task force, appointed by the chairperson and ETA guidelines and publications committee, consisted of three endocrinologists (IC, UFR, and CM) and four specialists in laboratory medicine (DJH, DG, VR, and SvdB). All panellists had clinical and/or clinical laboratory expertise in the investigation of discordant function tests. The manuscript and specific recommendations were developed by reviewing the best available research evidence, with the knowledge and technical experience of the panel. The Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) framework was used for grading the quality of evidence and making recommendations (1, 2). The quality of the evidence was graded as level 1: high (RCT evidence/meta-analysis – high-quality evidence (ØØØØ)); level 2: moderate (intervention short of RCT or large observational studies – moderate quality (ØØØO)); level 3: low quality (case–control studies and case series – low quality (ØØOO)); and level 4: very low quality (case reports and expert opinion – very low-quality (ØOOO)) using the modified GRADE criteria (3). The strength of a recommendation was categorised as either strong (S; a recommendation) or weak (W; a suggestion, not a recommendation). The final draft of the manuscript was sent to the guidelines and publications committee for comments and thereafter posted on the ETA website for four weeks for feedback from the ETA members. All proposed changes and comments were taken into consideration, and the resulting revisions were incorporated into the final documents submitted to the European Thyroid Journal after approval by the guideline board.
Introduction
Accurate measurement of thyroid hormones (THs), including thyroid-stimulating hormone (TSH), free and total fractions of triiodothyronine (T3) and thyroxine (T4), and testing for thyroid autoantibodies and cancer-associated biomarkers (thyroglobulin and calcitonin), is key for diagnosing and managing thyroid disorders. Immunoassays remain the standard in clinical laboratories because of their sensitivity, specificity, automation, and rapid turnaround (4). However, their reliability can be compromised by analytical interferences that may lead to diagnostic errors and inappropriate management (5).
Interference arises when endogenous or exogenous substances disrupt antibody–antigen binding or signal detection. Common sources include heterophilic antibodies, biotin, and autoantibodies directed against THs or assay components. Heterophilic antibodies may link capture and detection antibodies, generating spurious signals, while biotin (widely used as a supplement) has previously been described to cause falsely high or low hormone results in streptavidin–biotin-based assays (6).
These interferences can result in misclassification of thyroid status, underscoring the need for clear clinical guidelines and mitigation strategies, such as blocking reagents, assay redesign, or alternative technologies less prone to interference (6). Advances in immunoassay platforms, automated mass spectrometry, and AI-supported assay development may further enhance test robustness.
Here, we summarise the current understanding of immunoassay interference, outline approaches for detection and resolution, and offer guidance for clinicians and clinical laboratory professionals managing patients with confirmed or suspected assay interference in tests used to measure THs or related substances.
General principles
Thyroid disease is common, affecting over 8% of the population (2.5% thyrotoxicosis; 5.8% hypothyroidism) (7). Although TH measurement is generally straightforward, interfering substances can cause inaccurate results, leading to misdiagnosis, inappropriate investigations, or unnecessary interventions. In a review of >150 cases, ≥50% of patients experienced negative clinical consequences from assay interference, whereas only 8% had no impact (6). Thus, interference is clinically significant and may cause patient harm if unrecognised.
People with discordant thyroid function tests (TFTs) due to assay interference or genuine causes may present in any clinical setting. Identification of assay interference as a cause of discordant TFTs can be challenging. Specialist laboratories will have expertise in investigation of discordant TFTs, while others will have limited capability to investigate such patients. We acknowledge that different clinics/laboratories have varying levels of expertise; some basic tests for interference may be done locally, while more specialised testing will require regional specialist services. It will not be possible or practical for all laboratories to develop assays for specialist investigation of thyroid function, so each clinical setting/laboratory should have access to a regional specialist centre where a biochemical workup can be performed and relayed back to the referring centre, and where patients can be referred for specialist clinical evaluation if indicated. We acknowledge that in resource-limited settings, access to such testing may be extremely limited.
Commercially available TSH and TH assays vary in susceptibility to interference, which may be assay-specific or genetically influenced (e.g. albumin or TSHβ variants). Assay architecture is continually updated to reduce interference, such as from biotin.
There are key steps required when assessing a patient with discordant TFTs, including i) careful history taking to note, in particular, thyroid symptoms, family history, and physiological states that affect THs (e.g. pregnancy and severe illness); ii) review of medications or supplements used (especially biotin and amiodarone); and iii) clinical examination for goitre and thyroid status (Fig. 1). Following the assessment, it may be possible to identify which assay (if any) seems erroneous. This is helpful information to the clinical biochemist.
Figure 1.
Algorithm demonstrating general approach to detection, investigation, and confirmation of assay interference. Assay interference may initially be suspected by a clinician or a clinical biochemist. In either instance, liaison between the clinician and clinical biochemist will ensure appropriate further testing. In some circumstances, specialist testing and/or clinical assessment may be required at a tertiary referral centre. Interdisciplinary discussion (specialists in laboratory medicine and clinicians +/− genetics), at either local or specialist centres, should occur in order to optimise testing and identification of interference. Created in BioRender by Moran C (2026) https://BioRender.com/uv0w3ur.
Following assessment of the patient with discordant TFTs, liaison with the clinical biochemist is essential. Initial steps may include repeat testing using an alternate assay (preferably a two-step method) or dilution studies. Advanced methods, such as polyethylene glycol precipitation, may be needed, as outlined below.
If assay interference is identified, the findings should be clearly reported back to the patient and their referring physician. If an inherited cause of assay interference is identified (such as familial dysalbuminaemic hyperthyroxinaemia (FDH) or variant thyroxine-binding globulin (TBG)/transthyretin (TTR)/TSHβ subunit), screening should be offered to first-degree relatives. Clear, written information regarding the presence of assay interference should be provided to the patient, who can provide this to any treating physicians in future to avoid duplication of investigation or unnecessary treatment. It can be helpful to inform referring physicians which (if any) assays are reliable; if TSH assays are unaffected by assay interference, for example, the TSH concentration could be relied upon during a future assessment of thyroid status, if required. A summary of the task force recommendations is supplied in Table 1.
Table 1.
Summary of recommendations.
| General considerations | |
| 1 | We recommend that clinicians interpreting thyroid function tests and all clinical laboratory staff are aware of the possibility of assay interference in immunoassays. (ØØOO, S) |
| 2 | We recommend that endocrinologists and specialists in laboratory medicine be aware of how reference intervals are generated and how these are affected by physiological and pathological states and understand the limitations of TSH, FT4, and FT3 measurements. (ØØOO, S) |
| 3 | We recommend that endocrinologists and specialists in laboratory medicine know which type of assay was used to measure TSH, (F)T4, and (F)T3 in a particular patient and the limitations of these assays. (ØØOO, S) |
| 4 | We suggest that endocrinologists and specialists in laboratory medicine be aware of the general causes of assay interference in TSH and thyroid hormone measurements, such as antibody interference, thyroid hormone-binding protein abnormalities, anti-reagent antibodies, displacing agents, and others. (ØOOO, W) |
| 5 | We recommend that endocrinologists and clinical biochemists be aware that the susceptibility of assays to various forms of interference will change over time, with newer generation of assays being developed to overcome observed forms of interference. (ØØOO, S) |
| 6 | We recommend consideration of assay interference when thyroid function tests (TSH, (F)T4 +/− (F)T3) are incongruent with the clinical picture or with each other (e.g. elevated T4 but TSH within the reference interval). (ØØOO, S) |
| 7 | Following identification of a patient with suspected TSH, thyroid hormone, or thyroid autoantibody assay interference by a clinician, we recommend liaison with clinical laboratories to determine the most appropriate further biochemical test(s). (ØØOO, S) |
| 8 | If suspected TSH, thyroid hormone, or thyroid autoantibody assay interference is first identified by the specialist in laboratory medicine, we recommend liaison with the requesting clinician as the first step and thereafter performing further testing as clinically indicated. (ØOOO, S) |
| 9 | We recommend that all patients with suspected assay interference in TFTs have a clinical review, including medication history, family history, and clinical examination for signs of thyroid disease. (ØØOO, S) |
| 10 | We recommend that endocrinologists and clinical biochemists establish a communication pathway within their hospital/clinic to relay results of suspected assay interference and facilitate discussion of such cases. (ØØOO, S) |
| 11 | We acknowledge that availability of testing for assay interference locally will vary depending on the clinical laboratory expertise and may be limited. (ØØOO, S) |
| 12 | We recommend that all medical scientists and endocrinologists have access to a referral/specialist clinical laboratory where specialist testing for assay interference can be performed if indicated, such as method comparison, TSH dilution studies, immunosubtraction studies (typically with polyethylene glycol), gel filtration chromatography, equilibrium dialysis, testing for thyroid hormone-binding abnormalities, measurement of thyroid hormones by mass spectrometry, and others. (ØØOO, S) |
| 13 | We recommend that each local centre have access to a regional specialist centre where patients can be referred for specialist clinical evaluation if indicated. (ØØOO, S) |
| 14 | We recommend that centres offering specialised interference testing are adequately supported to perform ongoing specialist testing. (ØØOO, S) |
| 15 | We recommend ongoing research in development of new methods to identify and delineate causes of assay interference. (ØØOO, S) |
| 16 | Where assay interference is identified, we recommend that this information is formally communicated in writing to the primary care physician and the referring physician (if different to the primary care physician). The patient should also receive this information. (ØOOO, S) |
| 17 | If assay interference is identified, we suggest that clinical biochemists and endocrinologists provide advice to primary care physicians regarding measurements (TSH or FT4) that are reliable and could be used to aid assessment of thyroid status in future. This can be helpful in a patient with coexistent thyroid disease, or if symptoms suggestive of thyroid disease later arise. (ØOOO, W) |
| 18 | We suggest that accessible plain language information on assay interference, in written and/or online formats, be available to patients in whom assay interference is identified. (ØOOO, W) |
| 19 | Where an inherited cause of assay interference is identified (such as familial dysalbuminaemic hyperthyroxinaemia, TBG/transthyretin abnormalities, and TSHB subunit variant), we recommend that screening be offered to first-degree relatives who may be affected. (ØOOO, S) |
| TSH assays | |
| 20 | Where TSH assay interference is suspected, we recommend that further biochemical tests for TSH interference be performed. The method chosen will be dependent on the capabilities of the local clinical laboratory but may include TSH measurement in alternative assay(s) and TSH dilution studies. For some clinical laboratories, direct referral for more specialised testing at a referral laboratory at the outset may be most appropriate (Ab neutralisation, PEG precipitation, gel filtration chromatography; see Table 2). (ØØOO, S) |
| 21 | If TSH assay interference is confirmed, we suggest that the laboratory report contains information on the assay(s) affected and what tests have been performed to investigate the cause of the interference. Depending on the nature of the interference, the assessment of thyroid status is based on clinical status and FT4 measurement +/− an alternative assay for TSH. (ØOOO, W) |
| T4/T3 assays | |
| 22 | Where interference in FT4 and/or FT3 is suspected, we recommend further clinical laboratory testing, with the method chosen dependent on the capabilities of the local laboratory, such as FT4/FT3 measurement in alternative assay(s), total T4/total T3 measurement, and/or more specialised testing at a referral clinical laboratory. (ØØOO, S) |
| 23 | Where interference in FT4 and/or FT3 is suspected to be due to medications (e.g. biotin and displacing agents), we recommend repeat analyses of FT4 and/or FT3 after omission of the suspected agent, if possible. (ØØOO, S) |
| 24 | Where an inherited form of FT4 and/or FT3 assay interference is suspected (e.g. familial dysalbuminaemic hyperthyroxinaemia (FDH), transthyretin abnormalities (TTR), or TBG abnormalities (TBG)), we recommend familial TSH, FT4, and FT3 measurements. We recommend confirmation of diagnosis by genetic testing. (ØØOO, S) |
| 25 | If FT4/FT3 assay interference is confirmed, we suggest that the assessment of thyroid status is based on clinical status and TSH measurement. In some cases, quantification of true endogenous T4/T3 concentration may be made by measuring FT4/FT3 by an alternative assay/a two-step assay, following equilibrium dialysis, or TT4/TT3 by mass spectrometry. (ØOOO, W) |
| 26 | We recommend proceeding to specialist investigations for rare inherited or acquired causes of discordant TFTs (such as resistance to thyroid hormone b- or TSH-secreting pituitary tumours) only after assay interference has been definitively excluded by a specialist in laboratory medicine. (ØØOO, S) |
| Other assays – thyroglobulin, anti- TSH receptor antibody, anti- thyroglobulin antibody, anti- TPO antibody, and calcitonin | |
| 27 | We recommend that clinicians and clinical biochemists be aware of the possibility of thyroglobulin, thyroglobulin antibody, and calcitonin assay interference, which is especially relevant in the management of thyroid cancer patients. (ØØOO, S) |
| 28 | Although anti-TSH receptor antibodies have high sensitivity and specificity for Graves’ disease, we recommend that clinicians and specialists in laboratory medicine are aware of rare occurrences of assay interference in these assays, with the resultant possibility of misdiagnosis. (ØOOO, S) |
| 29 | We recommend that clinicians and clinical biochemists consider the possibility of a high-dose hook effect when unexpectedly low thyroglobulin or calcitonin concentrations are observed in patients with differentiated or medullary thyroid cancer, respectively, especially when results are particularly discordant with the clinical picture. (ØOOO, S) |
| Future considerations | |
| 30 | Assays for TSH, thyroid hormones and thyroid autoantibodies change over time, and we recommend that endocrinologists and specialists in laboratory medicine remain up to date in changes in assays and advances in measurement in order to implement these into their practice, as appropriate. (ØOOO, S) |
Fundamental principles of assay performance and interference
Large molecules, such as TSH, can be measured using an immunometric (‘sandwich’) assay, where the TSH molecule is bound by a capture antibody (immobilised on a solid phase) and a detection antibody (the ‘tracer’, labelled with enzymes, fluorescence, radiation, or chemiluminescence) (Fig. 2). The signal generated is directly proportional to the TSH concentration (5). Assay interference occurs most commonly due to antibody interference, which can cause either falsely high or low TSH concentrations. Rarely, genetic variations in the TSH molecule can occur, which can cause falsely low TSH concentrations.
Figure 2.
Mechanisms of antibody interference in TSH immunoassays. Schematic demonstrating mechanisms of assay interference in TSH immunometric immunoassays. A positive signal occurs when the chemiluminescent signal is immobilised to the solid phase; this occurs by sandwiching TSH between a capture and detection antibody (1). Falsely low TSH results (negative interference) can occur due to the presence of an antibody that binds directly to the capture antibody (2), or the labelled, detection, antibody (3). Falsely high TSH results (positive interference) occur when antibodies form a bridge between capture and detection antibodies (4), or endogenous TSH autoantibodies form immunoreactive but biologically inactive hormone complexes (macro-TSH) that are detected in the assay (5). Created in BioRender by Moran C (2026) https://BioRender.com/uv0w3ur.
T4 and T3 molecules are smaller and more difficult to capture between two antibodies; thus, single-site immunoassays are typically used for thyroid hormone measurement. In this assay format, the concentration of hormones is estimated by competition between the hormone present in the sample and a labelled hormone or hormone analogue for available binding sites on an anti-T4/T3 antibody. Some assays use a labelled antibody and immobilised analogue. Thyroid hormones are also extensively protein bound with the small fraction of free hormone being biologically active. The single-site assay can be used for both total and free hormone assays depending on whether the protein-binding equilibrium is preserved (in free hormone assays) or disrupted (in total hormone assays). Unfortunately, preserving the in vivo equilibrium between free and bound hormone to enable accurate free hormone assessment is challenging, and any factor that disturbs this equilibrium may lead to assay interference (6).
The generated signal is inversely proportional to the concentration of T4/T3 present in the serum (5) (Fig. 3). Any factor that disturbs the in vitro equilibrium between the free THs (FT4/FT3) and thyroid hormone-binding proteins (THBPs) may lead to assay interference or shift the balance between bound and free TH fractions. This includes, for example, medications that displace THs from THBPs, or the presence of variant THBPs with altered affinity for THs or abnormal concentrations.
Figure 3.
Mechanisms of interference in free T4 assays. Free T4 (FT4) concentrations are determined by indirect, competitive immunoassays. Labelled T4 and ‘free’ (endogenous) T4 compete for available anti-T4 antibody-binding sites; the more endogenous fT4 is present in the sample, the less labelled T4 is captured (1). The amount of T4 captured can be affected by (i) thyroid hormone-binding protein variants or displacing agents affecting the T4:FT4 equilibrium, (ii) endogenous anti-T4 autoantibodies binding the T4 tracer, or (iii) interfering antibodies blocking the capture antibody (2). FT4 assay interference is typically positive (i.e. less tracer is bound, and hence a falsely high FT4 result is generated), although weakly binding displacing agents can cause negative interference. Created in BioRender by Moran C (2026) https://BioRender.com/uv0w3ur.
In addition, multiple forms of antibody interference can occur and may lead to elevated or decreased T4/T3 results. The impact of these interferences varies between immunoassays, as assay formats and designs differ across manufacturers, resulting in different susceptibilities to matrix effects and binding perturbations.
Various forms of antibody interference exist and may cause falsely high or low FT4/FT3 concentrations.
TSH assay interference
Prevalence of TSH assay interference
Numerous causes of analytical interference in the TSH assay are reported in the scientific literature, including medications (biotin), genetic variation in TSHB (8), antibodies to components of the assay (anti-streptavidin or anti-ruthenium antibodies), autoantibodies (anti-TSH autoantibodies, also termed macro-TSH), and others (human anti-animal antibodies (HAMAs) or heterophilic antibodies) (6). Prevalence of such causes of artefactual results varies from 0.6–1.6% for macro-TSH to 0.4% for anti-reagent antibodies. The TSHβ variant is particularly prevalent (1%) in the South Asian population (4, 9).
Overview of TSH assays
The advent of monoclonal antibodies allowed development of immunometric assay methodology (sandwich assay). This improved the specificity and drastically increased the functional sensitivity by dropping the limit of quantification from approximately 1 mIU/L with first-generation immunoassays to 0.01 mIU/L with third-generation immunoassays, which are now widely available (10). Depending on the manufacturer, the capture antibodies may be linked to magnetic bead covalently or with a system involving biotin/streptavidin affinity. The detection antibody is linked to the signal. TSH immunoassays are unlikely to be susceptible to the hook effect.
Unlike bioassays that measure the bioactivity of TSH, immunoassays cannot provide information on variations in glycosylation and bioactivity of TSH. Mass spectrometry TSH assays are currently not available for clinical use, although being developed. They may be able to distinguish isoform differences, but assay design and analytical sensitivity are still major issues (8).
The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Committee for Standardization of Thyroid Function Tests (C-STFT) concluded that TSH cannot be fully standardised because no reference measurement procedure exists. Instead, the committee developed a harmonisation strategy using multi-assay comparisons with native, clinically relevant samples across multiple manufacturers (11). Although this approach improves agreement among commercial assays, it does not address immunoassay-related analytical interferences.
Forms of TSH assay interference
Antibody interference
Antibodies are the main source of analytical interference in TSH immunoassays (Fig. 2). HAMAs and heterophilic antibodies, anti-streptavidin, or anti-ruthenium may cause under- or over-estimation of TSH concentration (12). Anti-TSH antibodies that are complexed with TSH are termed ‘macro-TSH’; this molecule is biologically inactive due to confinement to the intravascular compartment because of its high molecular size. None of the available immunoassays can discriminate all macro-TSH forms from bioactive free TSH, although some assays are less sensitive to its presence (e.g. Abbott assays, e.g. Alinity®) than others (e.g. Roche Cobas®) (6, 13, 14). Notably, transplacental transfer of maternal macro-TSH may result in biochemical findings that mimic transient congenital hypothyroidism in newborns, potentially leading to misdiagnosis and unnecessary treatment for congenital hypothyroidism (15).
Genetic
A single nucleotide change in TSHB encodes a variant TSHβ subunit that retains bioactivity but alters an epitope necessary to generate a TSH ‘sandwich’ in some (but not all) assays (8, 9).
Others
Non-specific interference has been observed in patients tested shortly after iodine-based contrast administration, causing under- or overestimation of results (16). Increasing use of therapeutic monoclonal antibodies may also raise the risk of antibody-related interference, although some studies report no such effects across multiple assays (17).
Methods to identify TSH assay interference
The first and most simple strategy is the use of an alternative immunoassay employing antibodies with different epitopes and host species (6, 9). Depending on the suspected nature of interference, several other tests may be used (Fig. 4). Polyethylene glycol will differentially precipitate proteins based on solubility; in serum, immunoglobulins are readily precipitated (anti-reagent and anti-TSH antibodies). However, this test should be interpreted with caution as up to 70% of TSH immunoreactivity can be precipitated by polyethylene glycol in the absence of antibody interference; for this reason, platform and population specific cut-offs need to be established in each laboratory (18). Mixing studies can be carried out using sera with high TSH values; the presence of interference is likely to result in a non-linear reduction in TSH concentration. Heterophilic blocking tubes (HBTs) can be used to preabsorb interfering antibodies, before remeasuring TSH, but are relatively ineffective in most cases. More specialised methods are still available in some laboratories; these include gel filtration chromatography to identify macro-TSH and genetic testing (TSHB) when necessary (9). When the interference is thought to be related to a pharmaceutical product (e.g. biotin), the simplest method is to draw blood for TFT after cessation of the medication.
Figure 4.
Summary of tools available to investigate suspected TSH and free T4 assay interference. Tools available to investigate TSH and free T4 assay interference are listed. These will not be available in the majority of laboratories, but are available in specialist referral laboratories. The tests performed will depend on the assay in which interference in suspected, the suspected cause of assay interference, and availability of testing in a given centre. Created in BioRender by Moran C (2026) https://BioRender.com/uv0w3ur.
T4 assay interference
Most laboratories routinely measure FT4 when assessing thyroid function, and most reports of T4 assay interference relate to FT4 assays. Therefore, the guidance here focuses primarily on FT4 measurement. Some information on total T4 (TT4) assay interference is provided separately below.
Prevalence of FT4 assay interference
The prevalence of free T4 (FT4) assay interference varies by assay type. Single-step competitive immunoassays (widely used in routine laboratories) are more prone to interference than reference methods, such as equilibrium dialysis or ultrafiltration, which are typically confined to specialist centres. Inter-method variability in FT4 immunoassays exists, particularly when FT4 concentrations are high (19).
Genetic variants in THBPs also cause assay artefacts. Albumin variants occur in up to 1.8% of the population (7); transthyretinaemic hyperthyroxinaemia (TTH) has an estimated prevalence of 1:6,300 (7); and X-linked TBG variants occur in about 1:5,000 individuals, occasionally yielding falsely elevated results (9, 20).
Thyroid hormone autoantibodies (THAAs) are found in ∼1% of the general population (21) and up to 40% of patients with autoimmune thyroid disease (6, 22) but interfere in only 1–2% of affected samples (23, 24, 25). Although modern assays have reduced interference from heterophilic and anti-reagent antibodies, these remain unpredictable and continue to cause diagnostic confusion.
Monoclonal paraproteins may cause assay interference in up to 8% of people with discordant TFTs due to assay interference (26).
Overview of FT4 assays
FT4 can be measured in clinical practice using two main approaches.
Indirect immunoassays
Using indirect methods of FT4 measurement, free hormone is inferred by competitive assay responses. These automated, high-throughput methods are widely used but are vulnerable to interference from autoantibodies, displacing agents, and genetic variation in THBPs. Because T4 is small, both current assay designs use single-antibody formats. In one-step assays, the sample, detection antibody, and competitive tracer are incubated together, whereas two-step assays add a wash step before tracer addition to limit interference from serum components.
Direct methods
In direct methods, bound T4 and FT4 are physically separated, using equilibrium dialysis (27) or ultrafiltration (28), and then, FT4 is measured using an immunoassay or LC–MS. These require more time, manual handling, and larger sample volumes, making them less suitable for routine use. However, they are more resistant to interference. The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) working group for standardisation of Thyroid Function Tests (WG-STFT) has established a reference measurement procedure for FT4 (equilibrium dialysis–isotope dilution–liquid chromatography/tandem mass spectrometry, ED–ID–LC/tandem MS) (29). Recently, alternative candidate reference methods for FT4 have been proposed (27, 30). None of these are readily available for routine clinical use.
Forms of FT4 assay interference
Displacing agents
Some medications affect the accuracy of FT4 measurement, often due to displacement of T4 from THBPs (31). Heparin is the best described example: it stimulates lipoprotein lipase, generating free fatty acids ex vivo that displace T4 and falsely elevate FT4 (32). Markedly increased endogenous fatty acids can produce a similar effect. High doses of phenytoin and furosemide can also generate incorrect FT4 results if the assay design permits readjustment of the T4-binding equilibrium during measurement.
THBP variation
Although FT4 assays aim to minimise THBP dependence, many immunoassays are affected by genetic and physiological variation, including pregnancy (33). FDH, caused by albumin variants with increased T4 affinity (8), typically leads to FT4 overestimation in indirect assays, with the degree of interference being assay-specific (34). Direct methods are less affected but remain sensitive to assay conditions (35, 36). Interference from TTH is likewise assay-dependent and may fluctuate within individuals due to illness-related changes in TTR structure (8). TBG variants mainly affect TT4, but TBG deficiency can occasionally yield falsely high FT4 (9, 20).
Antibody interference
Thyroid hormone autoantibodies (THAAs)
THAAs can increase FT4 in one-step assays by binding the tracer, whereas two-step assays are less susceptible because THAAs are removed before tracer addition (6). Despite improved design of one-step methods, THAA-related interference (though rare) remains clinically significant.
Anti-reagent antibodies
Antibodies against assay components, including ruthenium, biotin, and streptavidin, may block tracer binding and produce spuriously high FT4 (12). Manufacturers incorporate blocking reagents, but complete prevention is not achievable (37).
Medications
Biotin supplements interfere with biotin–streptavidin-based competitive assays, typically causing falsely high FT4 (38). Chloral hydrate can cause erroneously raised TH concentrations (9).
Others
Markedly increased free fatty acid concentrations can raise FT4 artefactually (39), although concentrations required for interference are unlikely in patients with normal lipase activity (40). Clonal paraproteins may cause assay interference (26).
Methods to investigate FT4 assay interference
Reviewing prescribed and over-the-counter medications that may alter T4 metabolism or displace T4 from binding proteins is a recommended first step before investigating assay interference. If displacement is suspected, TT4 measurement can help, as concentrations are expected to remain normal in euthyroid individuals without THBP variation. Serum protein electrophoresis and immunofixation may be helpful if a lymphoproliferative disorder is suspected (26). Direct methods, such as equilibrium dialysis, are resistant to most forms of interference, but require careful interpretation due to factors such as buffer composition, pH, and temperature (41). Notably, in cases of displacement, dialysis methods return a high FT4 concentration, since the FT4 concentration in vitro is genuinely elevated.
Given the cost, complexity, and limited availability of direct assays, repeating measurement using an alternative immunoassay can serve as a practical alternative, particularly since the performance characteristics of commonly used commercial assays are well established (6, 9). For laboratories using one-step assays, confirming results with a two-step method is advisable. Awareness of method bias is essential when comparing assays, especially as manufacturers continuously update immunoassays, potentially altering susceptibility to interference.
Although radioimmunoprecipitation for identifying anti-T3 and anti-T4 autoantibodies is rarely performed in clinical laboratories today, it remains a valuable tool (42). Specialist laboratories may perform radiolabelled T4-binding studies or electrophoretic techniques to assess abnormal T4–THBP interactions, although DNA sequencing now largely replaces older assays for detecting THBP variants (43). When inherited interference (e.g. FDH) is suspected, measuring THs in first-degree relatives can be a cost-effective and informative approach (Fig. 4).
Total T4 assay interference
The IFCC WG-STFT uses isotope dilution–liquid chromatography/tandem mass spectrometry (ID–LC/tandem MS) as the reference measurement procedure for TT4 (44). TT4 measurement is strongly influenced by THBP variation and is therefore less reliable in settings such as pregnancy or critical illness. Its current use is largely confined to newborn screening programmes (45) and to evaluating suspected FT4 assay interference.
TT4 immunoassays remain the most common method of TT4 measurement, but can be affected by THAAs and heterophilic antibodies directed against the hormone or assay reagents, so TT4 measurement by mass spectrometry may be more accurate.
T3 assay interference
Prevalence of T3 assay interference
Total T3 or FT3 assay interference is less commonly reported than FT4 assay interference. While the exact prevalence is unknown, one study reported a prevalence of total T3 (TT3) assay interference of 1.2% of patients, with a higher frequency in females (1.4 vs 0.5%) (46).
Overview of T3 assays
The assessment of TT3 and FT3 is technically more difficult than TT4 and FT4 measurements, due to their lower concentration in the serum (23). TT3 concentration is dependent on THBP and thus less reliable in conditions such as pregnancy and critical illness. In most laboratories, FT3 is measured by automated competitive immunoassays with a one-step or two-step architecture (22).
Indirect methods, such as free T3 index (FT3I) and T3/TBG ratio (47), are not frequently used and are not suitable for patients with THBP defects. Direct equilibrium dialysis and ultrafiltration coupled with RIA methods (48, 49) are no longer available in most laboratories.
The IFCC C-STFT has established a reference measurement procedure for FT3 (equilibrium dialysis–isotope dilution–mass spectrometry, ED–ID–MS) (50, 51) and TT3 (isotope dilution–liquid chromatography/tandem mass spectrometry, ID–LC/MS/MS) (44, 52, 53), but neither are widely available for routine testing.
Forms of T3 assay interference
Similar to T4 assay interference, FT3/TT3 interference may occur due to antibody interference, medications, and THBP variations.
Displacing agents
Several drugs can displace THs from binding proteins that may increase both FT4 and FT3 serum concentrations, as described above (6).
THBP variation
Patients with FDH due to variants in codons 218 and 222 of the albumin gene have falsely raised FT3 and TT3 concentrations in many immunoassays, including one two-step platform (54). Isolated raised TT3 (but not TT4, FT4, or FT3) concentrations were described in a family harbouring a different (L66P) heterozygous mutation in ALB (55).
Antibody interference
Antibodies are the main source of analytical interference in TT3/FT3 immunoassay (6). HAMAs and heterophilic antibodies ((56, 57), new), anti-T3 autoantibodies (24), anti-streptavidin (12), anti-ruthenium (58), paraproteins (59), and rheumatoid factor (60) can cause overestimation of TT3 and/or FT3 concentrations. The frequency of interference with these antibodies is low, variable among different assays, and the relationship between antibody concentrations and severity of interference is not linear (24, 25).
Medications
Therapeutic TH analogues, such as triiodothyroacetic acid (TRIAC) (61) and diiodothyropropionic acid (DITPA) (62), cross-react to variable extents on different FT3 and TT3 assays. Biotin treatment is associated with spuriously high FT3 (63) with a mechanism similar to anti-streptavidin antibodies. The sedative agent chloral hydrate may cause falsely elevated FT3 (9).
Others
A pre-analytical cause of spuriously increased TT3 due to surfactant agents in blood collection tubes has been reported (64).
Methods to investigate T3 assay interference
To investigate possible TT3/FT3 interference, the initial and most simple strategy is the use of an alternative one-step, or ideally two-step, immunoassay (6) or, if available, an alternative method such as LC-MS/MS. Remeasurement of both TT3 and FT3 can be helpful, since discrepant results are suggestive of interference. In one study of 120 samples with spuriously increased FT4, 71 patients had raised FT3 too, but TT3 concentrations were normal in more than 95% (65). HBTs can be useful, but the results can be assay-dependent and false positives may occur (66). PEG precipitation has been used in suspected FT3 interference due to paraproteins, but dilution of the sample changes the ratio of bound to unbound T3 (67), so it is advisable to compare the results with those of controls and genuine hyperthyroid patients (68).
Interference in other assays used in the assessment of thyroid disease
Thyroglobulin assay interference
Prevalence of thyroglobulin assay interference
Measurement of thyroglobulin (Tg) is primarily used to identify residual thyroid tissue or recurrence/metastases from a previously treated differentiated thyroid cancer (69, 70). The most common type of interference in the immunoassays is the presence of thyroglobulin antibodies (TgAbs) (71). The prevalence of these antibodies is high in normal individuals (in females up to 15–20%) (70) and also in patients with thyroid cancer (25%), tending to disappear after total thyroidectomy (72, 73) and remission (70, 74).
Other causes of Tg assay interference are likely only present in 1–2% of measurements, but considering the large volume of Tg measurements globally, and the emphasis placed on these results in cancer follow-up, this results in a large burden for healthcare systems and patients (70).
Overview of thyroglobulin assays
Tg is measured by an immunometric assay, which can be susceptible to interference (69, 70, 75). More recently, measurement by liquid chromatography–tandem mass spectrometry (LC–MS/MS) assay has been introduced (76), but these are not universally available due to the cost and low sensitivity, despite lesser influence from autoantibodies (77). In general, the variability between assays is very high, even after introduction of an international reference calibrator CRM457 (78), which reduced the inter-assay variability from approximately 60 to 30% (79, 80).
Forms of thyroglobulin assay interference
As stated above, the most common type of interference in the immunoassays is the presence of thyroglobulin antibodies (TgAbs). Tg assay interference may also be due to anti-thyroid hormone antibodies, heterophilic antibodies, anti-ruthenium and anti-streptavidin antibodies (74), rheumatoid factor (81), and biotin (63). It is also likely that other immunoglobulins, autoantibodies, and proteins interfere in Tg assays; for example, therapeutic infusions of intravenous immunoglobulin may contain TgAbs. A high-dose hook effect has been described in Tg measurements (82).
Methods to investigate thyroglobulin assay interference
TgAb measurement is recommended for all samples submitted for thyroglobulin (Tg) testing to help identify potential assay interference. However, TgAb assays are variable: antibodies may be present but undetected or detected without affecting the Tg assay. Heterophilic antibodies pose a greater challenge and may require specific approaches, including PEG precipitation, antigen blocking, or Tg measurement using an alternative assay. LC-MS/MS can be used when Tg assay interference is suspected, although falsely low Tg concentrations have been reported in some patients with structural disease and positive TgAbs (76).
Calcitonin assay interference
Prevalence of calcitonin assay interference
Calcitonin (CT) assays are important diagnostic tools primarily used for the detection and monitoring of medullary thyroid carcinoma. Although less frequent, interference has been reported in CT assays (83).
Overview of calcitonin assays
These assays typically employ two-site immunoassay techniques, with chemiluminescence signalling. The limit of quantification of CT immunoassays has been decreasing over the years, resulting in increased sensitivity.
Forms of calcitonin assay interference
The presence of heterophilic antibodies can lead to falsely elevated CT concentrations (84), as can autoantibodies that cross-react with the assay components. For example, in one case, calcitonin measurements were confounded by the presence of rheumatoid factor, leading to significantly elevated calcitonin concentrations despite the absence of medullary thyroid carcinoma (81). Macro-calcitonin interferes with CT assays by mimicking elevated CT concentrations without actual increases in the hormone (85). Another issue is the high-dose hook effect, where extremely high analyte concentrations saturate the assay’s detection antibodies, resulting in falsely low readings (83).
Methods to investigate calcitonin assay interference
CT interference should be considered if the result is at variance with the clinical picture. Heterophilic blocking tubes, dilution studies, or CT measurement using alternative assays could be employed to investigate CT assay interference.
Anti-receptor antibody (TRAb) assay interference
Prevalence of TRAb assay interference
The specificity and sensitivity of anti-TSH receptor autoantibodies (TRAbs) for Graves’ disease are excellent such that a positive TRAb result is almost diagnostic for Graves’ disease (86). However, as TRAb methods are based on immunoassay techniques, they are likely to suffer the same interferences as described above, although the prevalence has not been determined.
Overview of TRAb assays
In clinical practice, most TRAb assays are based on the displacement of a labelled monoclonal anti-TSH receptor antibody from immobilised recombinant TSH receptors.
As both blocking and stimulating TRAbs exist (87), a bioassay is required to discriminate between these species. Recent development of rapid and robust TRAb bioassays may enable clinical laboratories to distinguish between blocking and stimulating TRAbs more readily (88, 89).
Forms of TRAb assay interference
Biotin (90) and anti-streptavidin antibodies (91) have been described to cause TRAb interference, resulting in falsely elevated TRAbs.
Methods to investigate TRAb assay interference
TRAb interference should be considered if the result is at variance with the clinical picture or TH concentrations. Repeat testing following biotin withdrawal, or use of an alternative assay (or bioassay), if available, is advised.
Anti-thyroglobulin antibody (TgAb) assay interference
Prevalence of TgAb assay interference
The prevalence of anti-thyroglobulin antibodies (TgAbs) is high in most populations, as mentioned above. The prevalence of TgAb assay interference is not known.
Overview of TgAb assays
TgAb measurement is cumbersome, with high inter-method variability and wide variation in limits of detection and reference intervals (70, 92).
Forms of TgAb assay interference
Very high concentrations of thyroglobulin in serum can interfere with the autoantibody measurement in some assays (93). Interference from thyroglobulin is assumed to vary with the initial concentration, i.e. low TgAb concentrations only need mildly elevated Tg to interfere with the measurement, while very high TgAb concentrations need very high Tg concentrations to interfere with the quantification (94, 95). It is also likely that other immunoglobulins (including IvIg treatment), autoantibodies and proteins may interfere in TgAb assays (96).
Methods to investigate TgAb assay interference
TgAb interference should be considered if the result is at variance with the clinical picture. Repeat testing using an alternative method could be considered.
Anti-TPO antibody (TPO-Ab) assay interference
Prevalence of TPO-Ab assay interference
TPO-Abs are mainly used for the diagnosis of autoimmune thyroid disease but are often used as markers of autoimmunity in general (96, 97). The prevalence of TPO-Ab assay interference is not known.
Overview of TPO-Ab assays
TPO-Ab assays used in Europe are mainly third-generation automated immunoassay platforms (98). The upper reference limit is method-dependent and often arbitrarily established, in part due to uncertainty of criteria used to define the reference population (99). The sensitivity and specificity of TPO-Abs for autoimmune thyroid disease are relatively poor (97).
Forms of TPO-Ab assay interference
TPO-Ab immunoassays are vulnerable to interference from exogenous antibodies (e.g. intravenous immunoglobulin) and endogenous factors, such as heterophilic antibodies and rheumatoid factor (91). In addition, structural homology between thyroid peroxidase and myeloperoxidase permits antibody cross-reactivity (100), particularly in denatured or solid-phase assay conditions, although this effect appears limited in native systems (101). These mechanisms should be considered when laboratory findings are inconsistent with the clinical picture.
Methods to investigate TPO-Ab assay interference
If TPO-Ab assay interference is suspected, repeat testing using an alternative method, or measurement of other thyroid autoantibodies, could be considered.
Information on investigation of assay interference for specialists in laboratory medicine
Numerous laboratory methods are available to detect and investigate interference. Table 2 outlines these, providing a valuable resource for specialists in laboratory medicine.
Table 2.
Comprehensive overview of detection strategies for assay interference. Summary of laboratory methods used to identify and investigate assay interference in TSH, T4, thyroglobulin, and thyroid autoantibody assays.
| Method | Detection mechanism | Effectiveness | Notes | Potential limits | References |
|---|---|---|---|---|---|
| Dilution testing | Serial dilution of sample | Effective for identifying non-linearity | Helps to confirm the presence of macro-analytes or interference by revealing non-linearity | Will not identify the nature of the interference | (13, 107) |
| Remeasurement using an alternative assay | Comparing analyte measurements using one or more alternative assays | Effective for confirmation of interference | Discrepancies between different methods can indicate interference | Requires access to multiple platforms; cost and time-intensive | (5, 108) |
| Polyethylene glycol treatment | Precipitation with polyethylene glycol | Good for detecting macro-aggregates | Useful to differentiate true analyte from bound forms causing interference | Does not eliminate all forms of interference | (109, 110) |
| Pre-treatment | Use of heterophilic blocking tubes | Highly effective for heterophilic antibody interference | Directly targets and neutralises interfering antibodies prior to analysis | Limited to known heterophilic interactions | (56, 111) |
| Preventing biotin interference | Biotin supplementation awareness and assay design | Highly effective | Advising patients to withhold biotin supplements before testing and using assays not affected by biotin | Patient compliance needed; not all assays may be biotin-free | (6, 112) |
| Antibody neutralisation | Use of specific neutralising agents against known interferences | Varies with the target interference | Employed to directly tackle known interferences such as biotin, rheumatoid factor, or animal antibodies | Specificity required; may not work for unidentified interferences | (6) |
Dilution testing is a straightforward approach to suspected TSH assay interference, where serial dilution reveals non-linearity, indicative of macro-analytes or other interferences. Although effective, it does not pinpoint the exact nature of the interference, making it a preliminary rather than conclusive test, and requiring adequate dilution controls.
Analysing the same sample using alternative immunoassays can help identify assay interference, particularly when discrepant results are obtained across assays. This process requires access to multiple platforms and can be both cost- and time-intensive but is effective in confirmation.
Polyethylene glycol (PEG) treatment helps differentiate ‘true’ analytes from bound forms causing interference by precipitating macro-aggregates. While good for detecting certain types of interference, it cannot address all forms.
Pre-treatment with heterophilic blocking tubes offers a targeted approach to neutralise heterophilic antibody interference before analysis. This method is effective but limited to known heterophilic interactions; these reagents are usually already incorporated in modern immunoassays.
Finally, preventing biotin interference involves pre-treatment solution or advising patients to withhold biotin supplements before testing and using assays designed to be unaffected by biotin. Over time, assay redesign by manufacturers can reduce assay susceptibility to specific forms of interference. Together, these methods form a comprehensive toolkit for specialists in laboratory medicine to address and manage assay interferences, ensuring that TFT results are both accurate and reliable in reflecting the thyroid status of patients.
Access to these methods and level of expertise in detecting and investigating assay interference will vary between laboratories.
Investigation, management, and follow-up of people in whom interference is identified
Detection and confirmation of assay interference is important; in patients without thyroid disease, it allows the physician to clearly avoid unnecessary further investigation and treatment, and in patients with thyroid disease (for example, a patient with Hashimoto’s thyroiditis with FT4 assay interference), it facilitates appropriate interpretation of thyroid function tests. In those with genuinely discordant thyroid function (such as TSH-secreting pituitary tumours or resistance to thyroid hormone β), assay interference should always be excluded prior to progressing to further, specialised, investigations, as suggested in specific ETA guidelines (102, 103).
In patients with evidence of FT4 assay interference, TSH alone can be used to guide thyroid status. If an inherited cause of assay interference is identified, screening should be offered to first-degree family members. For individuals with confirmed TSH assay interference, an alternative assay that accurately measures TSH may be used to evaluate thyroid function. However, in certain cases of TSH interference, such as macro-TSH, obtaining a reliable TSH measurement may not be feasible. In these situations, thyroid assessment relies solely on clinical context along with FT4 and FT3 values.
If assay interference has been identified, clear documentation should be provided to the patient and primary care practitioner and recorded in the patient’s medical record. The concept of assay interference can be difficult to explain to the patient, so access to high-quality, reliable patient information from national or international patient organisations would be valuable. The accompanying plain language summary of this guideline (Appendix 1) is intended to provide an accessible overview of the key messages and recommendations contained in this guideline for non-specialist readers, including patients, carers, healthcare professionals outside the speciality, and the wider public.
Notably, as assays change and evolve, assay interference may not be present indefinitely in an individual, so patients need to be re-assessed by an endocrinologist if clinical features of thyroid disease emerge in future.
Conclusion and future directions
As we look to the future of thyroid function tests (TFTs), a convergence of emerging technologies stands poised to significantly reduce assay interferences and enhance diagnostic accuracy. Advances in omics technologies and automated mass spectrometry (MS) are at the forefront, promising to deepen our understanding of thyroid biochemistry and pathophysiology (76). Automated MS, in particular, may offer a robust solution to the prevalent issue of antibody interferences in traditional immunoassays, by providing a more specific and reliable means of measuring thyroid markers. Concurrently, the rise of artificial intelligence (AI) is set to transform TFTs through the in silico design of assays (104). By employing machine learning algorithms, these AI-enhanced designs can anticipate and circumvent potential interferences before the assays are even deployed in a clinical setting. Furthermore, integrating these AI models with electronic medical records enables the analysis of large datasets to detect patterns that predict likelihood of interference, thereby improving the predictability and reliability of thyroid testing. Integration of AI not only aims to refine the accuracy of TFTs but also facilitates a shift towards personalised medicine. By tailoring TFTs to the individual biochemical profiles of patients, these technologies hold the potential to greatly improve disease management and patient outcomes (105). Together, these advancements herald a new era in endocrinology, where technology-driven insights and innovations lead to more precise, interference-resistant thyroid function testing. However, the adoption of emerging technologies should be rigorously validated and ethically managed to ensure that they meet high standards of effectiveness and fairness (106).
Declaration of interest
The task force had no commercial support, and the task force members declare no conflicts of interest.
Funding
UFR’s research salary was funded by Kirsten and Freddy Johansen’s Fund.
Acknowledgement
The authors thank the British Thyroid Foundation for reviewing the plain language summary draft and providing valuable feedback on its accessibility and readability.
Appendix 1
Understanding the 2026 ETA Guideline on Interference in Immunoassay Measurements Used in the Assessment of Thyroid Function: A Plain Language Summary
Thyroid blood tests are commonly used to diagnose and monitor thyroid conditions, including an underactive thyroid (hypothyroidism), an overactive thyroid (hyperthyroidism) and thyroid cancer. These tests usually provide reliable information and are essential for guiding treatment decisions.
Most thyroid tests are performed using laboratory methods called immunoassays, which measure hormones, such as thyroid stimulating hormone (TSH), thyroxine (T4), triiodothyronine (T3), and thyroid-related antibodies, in blood samples. Although these tests are generally accurate, results can occasionally be misleading because of assay interference. Assay interference occurs when substances in the blood, inherited biological differences, or medications affect how the laboratory test performs, leading to results that do not accurately reflect thyroid gland function.
Incorrect thyroid test results may lead to delayed diagnosis, unnecessary investigations, inappropriate treatment, or confusion for patients and healthcare professionals. In some cases, misleading test results may resemble rare thyroid disorders, making interpretation particularly challenging.
This guideline summarises the causes of assay interference in thyroid-related laboratory testing and provides recommendations for healthcare professionals investigating unexpected or conflicting thyroid test results.
The guideline highlights that interference may arise from several sources. These include naturally occurring antibodies that affect laboratory measurements, medicines and some supplements (particularly biotin, commonly used in hair and nail supplements), and inherited differences in proteins that transport thyroid hormones in the bloodstream. Pregnancy, severe illness, and some medical treatments may also influence the interpretation of thyroid test results.
The susceptibility to interference varies between different laboratory methods. Some machines used for thyroid hormone levels are more resistant to interference than others, meaning that the same blood sample may occasionally produce different results depending on the method used. For this reason, unexpected or unusual thyroid blood tests should be interpreted carefully and always in the context of a person's symptoms, examination findings, medications, and medical history.
A major recommendation of the guideline is that healthcare professionals should consider assay interference when thyroid blood test results do not fit the clinical picture.
The guideline also emphasises the importance of collaboration between clinicians and laboratory medicine specialists when interference is suspected. Investigation may involve repeating tests using an alternative laboratory method, reviewing medications and supplements, or performing more specialised laboratory analyses. In selected cases, referral to specialist centres with expertise in complex thyroid function testing may be required.
The guideline also discusses interference affecting other thyroid-related tests, including thyroglobulin, calcitonin, and thyroid antibody measurements, which are commonly used in thyroid cancer and autoimmune thyroid disease. Awareness of possible interference in these tests is important because inaccurate results may influence diagnosis, treatment decisions, and long-term follow-up.
Importantly, the guideline does not suggest that thyroid blood tests are unreliable. On the contrary, thyroid testing is highly accurate for most people. However, recognising the small number of situations in which results may be misleading is important to avoid unnecessary investigations or inappropriate treatment.
Advances in laboratory technology, including improved assay design and more sophisticated methods of thyroid hormone measurement, are expected to reduce interference in the future and improve test accuracy further. Increased awareness among healthcare professionals and close communication between clinical and laboratory teams remain central to ensuring thyroid test results are interpreted correctly.
This guideline provides practical recommendations to help healthcare professionals recognise, investigate, and manage assay interference, with the aim of improving the safety, reliability, and interpretation of thyroid function testing.
References
- 1.Atkins D, Best D, Briss PA, et al. Grading quality of evidence and strength of recommendations. BMJ 2004. 328 1490. ( 10.1136/bmj.328.7454.1490) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Swiglo BA, Murad MH, Schünemann HJ, et al. A case for clarity, consistency, and helpfulness: state-of-the-art clinical practice guidelines in endocrinology using the grading of recommendations, assessment, development, and evaluation system. J Clin Endocrinol Metab 2008. 93 666–673. ( 10.1210/jc.2007-1907) [DOI] [PubMed] [Google Scholar]
- 3.Guyatt GH, Oxman AD, Vist GE, et al. GRADE: an emerging consensus on rating quality of evidence and strength of recommendations. BMJ 2008. 336 924–926. ( 10.1136/bmj.39489.470347.AD) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Van Uytfanghe K, Ehrenkranz J, Halsall D, et al. Thyroid stimulating hormone and thyroid hormones (triiodothyronine and thyroxine): an American Thyroid Association-Commissioned Review of Current Clinical and Laboratory Status. Thyroid 2023. 33 1013–1028. ( 10.1089/thy.2023.0169) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Koulouri O, Moran C, Halsall D, et al. Pitfalls in the measurement and interpretation of thyroid function tests. Best Pract Res Clin Endocrinol Metab 2013. 27 745–762. ( 10.1016/j.beem.2013.10.003) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Favresse J, Burlacu MC, Maiter D, et al. Interferences with thyroid function immunoassays: clinical implications and detection algorithm. Endocr Rev 2018. 39 830–850. ( 10.1210/er.2018-00119) [DOI] [PubMed] [Google Scholar]
- 7.Taylor PN, Albrecht D, Scholz A, et al. Global epidemiology of hyperthyroidism and hypothyroidism. Nat Rev Endocrinol 2018. 14 301–316. ( 10.1038/nrendo.2018.18) [DOI] [PubMed] [Google Scholar]
- 8.Pappa T, Ferrara AM & Refetoff S. Inherited defects of thyroxine-binding proteins. Best Pract Res Clin Endocrinol Metab 2015. 29 735–747. ( 10.1016/j.beem.2015.09.002) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Moran C, Schoenmakers N, Halsall D, et al. Approach to the patient with raised thyroid hormones and nonsuppressed TSH. J Clin Endocrinol Metab 2024. 109 1094–1108. ( 10.1210/clinem/dgad681) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Spencer CA. Laboratory thyroid tests: a historical perspective. Thyroid 2023. 33 407–419. ( 10.1089/thy.2022.0397) [DOI] [PubMed] [Google Scholar]
- 11.Thienpont LM, Van Uytfanghe K, De Grande LAC, et al. Harmonization of serum thyroid-stimulating hormone measurements paves the way for the adoption of a more uniform reference interval. Clin Chem 2017. 63 1248–1260. ( 10.1373/clinchem.2016.269456) [DOI] [PubMed] [Google Scholar]
- 12.Favresse J, Lardinois B, Nassogne MC, et al. Anti-streptavidin antibodies mimicking heterophilic antibodies in thyroid function tests. Clin Chem Lab Med 2018. 56 e160–e163. ( 10.1515/cclm-2017-1027) [DOI] [PubMed] [Google Scholar]
- 13.Hattori N, Ishihara T & Shimatsu A. Variability in the detection of macro TSH in different immunoassay systems. Eur J Endocrinol 2016. 174 9–15. ( 10.1530/EJE-15-0883) [DOI] [PubMed] [Google Scholar]
- 14.Larsen CB, Petersen ERB, Overgaard M, et al. Macro-TSH: a diagnostic challenge. Eur Thyroid J 2021. 10 93–97. ( 10.1159/000509184) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Hattori N, Aisaka K, Yamada A, et al. Prevalence and pathogenesis of macro-thyrotropin in neonates: analysis of umbilical cord blood from 939 neonates and their mothers. Thyroid 2023. 33 45–52. ( 10.1089/thy.2022.0457) [DOI] [PubMed] [Google Scholar]
- 16.van der Molen AJ, Krabbe JG, Dekkers IA, et al. Analytical interference of intravascular contrast agents with clinical laboratory tests: a joint guideline by the ESUR Contrast Media Safety Committee and the Preanalytical Phase Working Group of the EFLM Science Committee. Clin Chem Lab Med 2023. 62 608–614. ( 10.1515/cclm-2023-1184) [DOI] [PubMed] [Google Scholar]
- 17.Galior KD, Ladwig PM, Snyder MR, et al. Lack of observed interference by therapeutic monoclonal antibodies in select commonly utilized immunoassays. Clin Biochem 2023. 121–122 110685. ( 10.1016/j.clinbiochem.2023.110685) [DOI] [PubMed] [Google Scholar]
- 18.Piticchio T, Chiardi I, Tumminia A, et al. PEG precipitation to detect macro-TSH in clinical practice: a systematic review. Clin Endocrinol 2025. 102 235–244. ( 10.1111/cen.15179) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Westbye AB, Aas FE, Dahl SR, et al. Large method differences for free thyroid hormone assays in the hyperthyroid range can affect assessment of hyperthyroid status: Comparison of Abbott Alinity to Roche Cobas, Siemens Centaur and equilibrium dialysis LC-MS/MS. Clin Biochem 2023. 121–122 110676. ( 10.1016/j.clinbiochem.2023.110676) [DOI] [PubMed] [Google Scholar]
- 20.Refetoff S. Inherited thyroxine-binding globulin abnormalities in man. Endocr Rev 1989. 10 275–293. ( 10.1210/edrv-10-3-275) [DOI] [PubMed] [Google Scholar]
- 21.Sakata S, Matsuda M, Ogawa T, et al. Prevalence of thyroid hormone autoantibodies in healthy subjects. Clin Endocrinol 1994. 41 365–370. ( 10.1111/j.1365-2265.1994.tb02558.x) [DOI] [PubMed] [Google Scholar]
- 22.Bhagat CI, Garcia-Webb P, Watson F, et al. Interference in radioimmunoassay of total serum thyroxin and free thyroxin due to thyroxin-binding autoantibodies. Clin Chem 1983. 29 1324–1325. ( 10.1093/clinchem/29.6.1324) [DOI] [PubMed] [Google Scholar]
- 23.Spencer CA. Assay of thyroid hormones and related substances. In Endotext. South Dartmouth, MA, USA: MDText.com, Inc, 2017. (https://www.ncbi.nlm.nih.gov/books/NBK279113/) [Google Scholar]
- 24.Ni J, Long Y, Zhang L, et al. High prevalence of thyroid hormone autoantibody and low rate of thyroid hormone detection interference. J Clin Lab Anal 2022. 36 e24124. ( 10.1002/jcla.24124) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Sapin R, Gasser F & Schlienger JL. Familial dysalbuminemic hyperthyroxinemia and thyroid hormone autoantibodies: interference in current free thyroid hormone assays. Horm Res 1996. 45 139–141. ( 10.1159/000184776) [DOI] [PubMed] [Google Scholar]
- 26.Musolino A, Grifoni GFA, Rodolfi S, et al. Variable diagnoses in cases referred for discordant thyroid function tests: focus on lymphoproliferative disorders. Thyroid 2025. 35 1003–1012. ( 10.1177/10507256251364557) [DOI] [PubMed] [Google Scholar]
- 27.Jansen HI, van der Steen R, Brandt A, et al. Description and validation of an equilibrium dialysis ID-LC-MS/MS candidate reference measurement procedure for free thyroxine in human serum. Clin Chem Lab Med 2023. 61 1605–1611. ( 10.1515/cclm-2022-1134) [DOI] [PubMed] [Google Scholar]
- 28.Kushnir MM, Nelson HA & Doyle K. Clinical utility and analytical aspects of direct measurements of free hormones using mass spectrometry-based methods. J Appl Lab Med 2022. 7 945–970. ( 10.1093/jalm/jfac010) [DOI] [PubMed] [Google Scholar]
- 29.Van Houcke SK, Van Uytfanghe K, Shimizu E, et al. IFCC international conventional reference procedure for the measurement of free thyroxine in serum: International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Working Group for Standardization of Thyroid Function Tests (WG-STFT). Clin Chem Lab Med 2011. 49 1275–1281. ( 10.1515/CCLM.2011.639) [DOI] [PubMed] [Google Scholar]
- 30.Ribera A, Zhang L, Dabbs-Brown A, et al. Development of an equilibrium dialysis ID-UPLC-MS/MS candidate reference measurement procedure for free thyroxine in human serum. Clin Biochem 2023. 116 42–51. ( 10.1016/j.clinbiochem.2023.03.010) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Stockigt J. Assessment of thyroid function: towards an integrated laboratory – clinical approach. Clin Biochem Rev 2003. 24 109–122. [PMC free article] [PubMed] [Google Scholar]
- 32.Mendel CM, Frost PH, Kunitake ST, et al. Mechanism of the heparin-induced increase in the concentration of free thyroxine in plasma. J Clin Endocrinol Metab 1987. 65 1259–1264. ( 10.1210/jcem-65-6-1259) [DOI] [PubMed] [Google Scholar]
- 33.Jansen HI, van Herwaarden AE, Huijgen HJ, et al. Pregnancy disrupts the accuracy of automated fT4 immunoassays. Eur Thyroid J 2022. 11 e220145. ( 10.1530/ETJ-22-0145) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Khoo S, Lyons G, McGowan A, et al. Familial dysalbuminaemic hyperthyroxinaemia interferes with current free thyroid hormone immunoassay methods. Eur J Endocrinol 2020. 182 533–538. ( 10.1530/EJE-19-1021) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Refetoff S, Scherberg NH, Yuan C, et al. Free thyroxine concentrations in sera of individuals with familial dysalbuminemic hyperthyroxinemia: comparison of three methods of measurement. Thyroid 2020. 30 37–41. ( 10.1089/thy.2019.0058) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Choudhary A, Sriphrapradang C, Refetoff S, et al. Familial dysalbuminemic hyperthyroxinemia in a 4-year-old girl with hyperactivity, palpitations and advanced dental age: how gold standard assays may be misleading. J Pediatr Endocrinol Metab 2015. 28 241–245. ( 10.1515/jpem-2014-0019) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Takahashi S, Nishikawa M, Nishihara E, et al. Interference against a newly labeled substance with ruthenium sulfonate complexes showing discrepant thyroid function test results. Clin Chim Acta 2024. 553 117706. ( 10.1016/j.cca.2023.117706) [DOI] [PubMed] [Google Scholar]
- 38.Balzer AHA & Whitehurst CB. An analysis of the Biotin-(Strept)avidin system in immunoassays: interference and mitigation strategies. Curr Issues Mol Biol 2023. 45 8733–8754. ( 10.3390/cimb45110549) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Liewendahl K, Helenius T, Näveri H, et al. Fatty acid-induced increase in serum dialyzable free thyroxine after physical exercise: implication for nonthyroidal illness. J Clin Endocrinol Metab 1992. 74 1361–1365. ( 10.1210/jcem.74.6.1592881) [DOI] [PubMed] [Google Scholar]
- 40.Mendel CM, Frost PH & Cavalieri RR. Effect of free fatty acids on the concentration of free thyroxine in human serum: the role of albumin. J Clin Endocrinol Metab 1986. 63 1394–1399. ( 10.1210/jcem-63-6-1394) [DOI] [PubMed] [Google Scholar]
- 41.D’Aurizio F, Kratzsch J, Gruson D, et al. Free thyroxine measurement in clinical practice: how to optimize indications, analytical procedures, and interpretation criteria while waiting for global standardization. Crit Rev Clin Lab Sci 2023. 60 101–140. ( 10.1080/10408363.2022.2121960) [DOI] [PubMed] [Google Scholar]
- 42.D’Aurizio F, Biasotto A, Cipri C, et al. Thyroid function tests, incongruent internally and with thyroid status, both in a pregnant woman and in her newborn daughter. Eur Thyroid J 2022. 11 e210088. ( 10.1530/ETJ-21-0088) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Pandian MR, Morgan C, Nelson JC, et al. Differentiating various abnormalities of thyroxin binding to serum proteins by radioelectrophoresis of thyroxin and immunoassay of binding proteins. Clin Chem 1990. 36 457–461. ( 10.1093/clinchem/36.3.457) [DOI] [PubMed] [Google Scholar]
- 44.Thienpont LM, Van Uytfanghe K, Beastall G, et al. Report of the IFCC working group for standardization of thyroid function tests; part 3: total thyroxine and total triiodothyronine. Clin Chem 2010. 56 921–929. ( 10.1373/clinchem.2009.140228) [DOI] [PubMed] [Google Scholar]
- 45.Boelen A, Zwaveling-Soonawala N, Heijboer AC, et al. Neonatal screening for primary and central congenital hypothyroidism: is it time to go Dutch? Eur Thyroid J 2023. 12 e230041. ( 10.1530/ETJ-23-0041) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Rezaei-Ghaleh N, Hedayati M, Ordookhani A, et al. Positive interference in triiodothyronine (T3) assay using a radioimmunoassay kit. Int J Endocrinol Metab 2007. 4 142–146. ( 10.5812/ijem.37792) [DOI] [Google Scholar]
- 47.Konno N, Nakazato T, Hagiwara K, et al. A comparison of measurements of serum free T3 concentration by equilibrium dialysis, free T3 index, and T3:TBG ratio in thyroidal and non-thyroidal illnesses. Acta Endocrinol 1983. 103 501–508. ( 10.1530/acta.0.1030501) [DOI] [PubMed] [Google Scholar]
- 48.Wang YS, Hershman JM & Pekary AE. Improved ultrafiltration method for simultaneous measurement of free thyroxin and free triiodothyronine in serum. Clin Chem 1985. 31 517–522. ( 10.1093/clinchem/31.4.517) [DOI] [PubMed] [Google Scholar]
- 49.Weeke J, Boye N & Orskov H. Ultrafiltration method for direct radioimmunoassay measurement of free thyroxine and free tri-iodothyronine in serum. Scand J Clin Lab Invest 1986. 46 381–389. ( 10.3109/00365518609083686) [DOI] [PubMed] [Google Scholar]
- 50.Thienpont LM, Van Uytfanghe K, Beastall G, et al. Report of the IFCC working group for standardization of thyroid function tests; part 2: free thyroxine and free triiodothyronine. Clin Chem 2010. 56 912–920. ( 10.1373/clinchem.2009.140194) [DOI] [PubMed] [Google Scholar]
- 51.Gu J, Soldin OP & Soldin SJ. Simultaneous quantification of free triiodothyronine and free thyroxine by isotope dilution tandem mass spectrometry. Clin Biochem 2007. 40 1386–1391. ( 10.1016/j.clinbiochem.2007.08.007) [DOI] [PubMed] [Google Scholar]
- 52.Thienpont LM, Fierens C, De Leenheer AP, et al. Isotope dilution-gas chromatography/mass spectrometry and liquid chromatography/electrospray ionization-tandem mass spectrometry for the determination of triiodo-L-thyronine in serum. Rapid Commun Mass Spectrom 1999. 13 1924–1931. ( 10.1002/(SICI)1097-0231(19991015)13:19<1924::AID-RCM734>3.0.CO;2-U) [DOI] [PubMed] [Google Scholar]
- 53.Soukhova N, Soldin OP & Soldin SJ. Isotope dilution tandem mass spectrometric method for T4/T3. Clinica Chim Acta 2004. 343 185–190. ( 10.1016/j.cccn.2004.01.007) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Schoenmakers N, Moran C, Campi I, et al. A novel albumin gene mutation (R222I) in familial dysalbuminemic hyperthyroxinemia. J Clin Endocrinol Metab 2014. 99 E1381–E1386. ( 10.1210/jc.2013-4077) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Sunthornthepvarakul T, Likitmaskul S, Ngowngarmratana S, et al. Familial dysalbuminemic hypertriiodothyroninemia: a new, dominantly inherited albumin defect. J Clin Endocrinol Metab 1998. 83 1448–1454. ( 10.1210/jcem.83.5.4815) [DOI] [PubMed] [Google Scholar]
- 56.Bolstad N, Warren DJ & Nustad K. Heterophilic antibody interference in immunometric assays. Best Pract Res Clin Endocrinol Metab 2013. 27 647–661. ( 10.1016/j.beem.2013.05.011) [DOI] [PubMed] [Google Scholar]
- 57.Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999. 45 942–956. ( 10.1093/clinchem/45.7.942) [DOI] [PubMed] [Google Scholar]
- 58.Buijs MM, Gorgels JP & Endert E. Interference by antiruthenium antibodies in the Roche thyroid-stimulating hormone assay. Ann Clin Biochem 2011. 48 276–281. ( 10.1258/acb.2010.010160) [DOI] [PubMed] [Google Scholar]
- 59.Mandal K, Ashorobi D, Lee A, et al. Factitiously elevated total triiodothyronine in a euthyroid patient with multiple myeloma. Case Rep Endocrinol 2021. 2021 8479193. ( 10.1155/2021/8479193) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Martel J, Després N, Ahnadi CE, et al. Comparative multicentre study of a panel of thyroid tests using different automated immunoassay platforms and specimens at high risk of antibody interference. Clin Chem Lab Med 2000. 38 785–793. ( 10.1515/CCLM.2000.112) [DOI] [PubMed] [Google Scholar]
- 61.Chan SL, Refetoff S, Babic N, et al. Triiodothyroacetic acid cross-reacts with measurement of triiodothyronine (T3) on various immunoassay platforms. Am J Clin Pathol 2022. 157 156–158. ( 10.1093/ajcp/aqab124) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Leung EK, Yi X, Refetoff S, et al. Diiodothyropropionic acid (DITPA) cross-reacts with thyroid function assays on different immunoassay platforms. Clin Chim Acta 2016. 453 203–204. ( 10.1016/j.cca.2015.12.008) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Ylli D, Soldin SJ, Stolze B, et al. Biotin interference in assays for thyroid hormones, thyrotropin and thyroglobulin. Thyroid 2021. 31 1160–1170. ( 10.1089/thy.2020.0866) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Bowen RA, Chan Y, Ruddel ME, et al. Immunoassay interference by a commonly used blood collection tube additive, the organosilicone surfactant silwet L-720. Clin Chem 2005. 51 1874–1882. ( 10.1373/clinchem.2005.055400) [DOI] [PubMed] [Google Scholar]
- 65.Külz M, Fellner S, Rocktäschel J, et al. Dubiously increased FT4 and FT3 levels in clinically euthyroid patients: clinical finding or analytical pitfall? Clin Chem Lab Med 2022. 60 877–885. ( 10.1515/cclm-2021-1211) [DOI] [PubMed] [Google Scholar]
- 66.Cantor D. Reason for limitations of heterophilic blocking tube use on certain beckman coulter access assays. Clin Chem 2005. 51 1311. ( 10.1373/clinchem.2005.052142) [DOI] [PubMed] [Google Scholar]
- 67.Geiseler D & Ritter M. Effect of sample dilution on measurements of free (unbound) hormones. Clin Chem 1984. 30 28–32. ( 10.1093/clinchem/30.1.28) [DOI] [PubMed] [Google Scholar]
- 68.Beato-Víbora PI & Alejo-González S. Avoiding misdiagnosis due to antibody interference with serum free thyroxin. Int J Endocrinol Metabol 2016. 15 e37792. ( 10.5812/ijem.37792) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Giovanella L, Feldt-Rasmussen U, Verburg FA, et al. Thyroglobulin measurement by highly sensitive assays: focus on laboratory challenges. Clin Chem Lab Med 2015. 53 1301–1314. ( 10.1515/cclm-2014-0813) [DOI] [PubMed] [Google Scholar]
- 70.Giovanella L, D’Aurizio F, Petranović Ovčariček P, et al. Diagnostic, Theranostic and prognostic value of thyroglobulin in thyroid cancer. J Clin Med 2024. 13 2463. ( 10.3390/jcm13092463) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.McLachlan SM & Rapoport B. Breaking tolerance to thyroid antigens: changing concepts in thyroid autoimmunity. Endocr Rev 2014. 35 59–105. ( 10.1210/er.2013-1055) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Tuttle RM, Ahuja S, Avram AM, et al. Controversies, consensus, and collaboration in the use of 131I therapy in differentiated thyroid cancer: a joint statement from the American Thyroid Association, the European Association of nuclear medicine, the Society of Nuclear Medicine and Molecular Imaging, and the European Thyroid Association. Thyroid 2019. 29 461–470. ( 10.1089/thy.2018.0597) [DOI] [PubMed] [Google Scholar]
- 73.Giovanella L, Clark PM, Chiovato L, et al. Thyroglobulin measurement using highly sensitive assays in patients with differentiated thyroid cancer: a clinical position paper. Eur J Endocrinol 2014. 171 R33–R46. ( 10.1530/EJE-14-0148) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 74.Paczkowska K, Otlewska A, Loska O, et al. Laboratory interference in the thyroid function test. Endokrynol Pol 2020. 71 551–560. ( 10.5603/EP.a2020.0079) [DOI] [PubMed] [Google Scholar]
- 75.Evans C, Tennant S & Perros P. Thyroglobulin in differentiated thyroid cancer. Clin Chim Acta 2015. 444 310–317. ( 10.1016/j.cca.2014.10.035) [DOI] [PubMed] [Google Scholar]
- 76.Barbesino G, Algeciras-Schimnich A & Bornhorst J. Thyroglobulin assay interferences: clinical usefulness of mass-spectrometry methods. J Endocr Soc 2022. 7 bvac169. ( 10.1210/jendso/bvac169) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Clarke NJ, Zhang Y & Reitz RE. A novel mass spectrometry-based assay for the accurate measurement of thyroglobulin from patient samples containing antithyroglobulin autoantibodies. J Invest Med 2012. 60 1157–1163. ( 10.2310/JIM.0b013e318276deb4) [DOI] [PubMed] [Google Scholar]
- 78.Feldt-Rasmussen U, Profilis C, Colinet E, et al. Human thyroglobulin reference material (CRM 457). 2nd part: physicochemical characterization and certification. Ann Biol Clin 1996. 54 343–348. [PubMed] [Google Scholar]
- 79.Feldt-Rasmussen U & Schlumberger M. European interlaboratory comparison of serum thyroglobulin measurement. J Endocrinological Invest 1998. 11 175–181. ( 10.1007/BF03350129) [DOI] [PubMed] [Google Scholar]
- 80.Spencer CA, Bergoglio LM, Kazarosyan M, et al. Clinical impact of thyroglobulin (Tg) and Tg autoantibody method differences on the management of patients with differentiated thyroid carcinomas. J Clin Endocrinol Metab 2005. 90 5566–5575. ( 10.1210/jc.2005-0671) [DOI] [PubMed] [Google Scholar]
- 81.Lupoli GA, Barba L, Liotti A, et al. Falsely elevated thyroglobulin and calcitonin due to rheumatoid factor in non-relapsing thyroid carcinoma: a case report. Medicine 2019. 98 e14178. ( 10.1097/MD.0000000000014178) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Hillebrand JJ, Siegelaar SE & Heijboer AC. Falsely decreased thyroglobulin levels in a patient with differentiated thyroid carcinoma. Clin Chim Acta 2020. 509 217–219. ( 10.1016/j.cca.2020.06.027) [DOI] [PubMed] [Google Scholar]
- 83.Censi S, Cavedon E, Fernando SW, et al. Calcitonin measurement and immunoassay interference: a case report and literature review. Clin Chem Lab Med 2016. 54 1861–1870. ( 10.1515/cclm-2015-1161) [DOI] [PubMed] [Google Scholar]
- 84.Bories PN, Broutin A, Delette A, et al. Comparison of the Elecsys calcitonin assay with the Immulite 1000 assay. Describing one case with heterophilic antibody interference. Clin Chem Lab Med 2016. 54 e45–e47. ( 10.1515/cclm-2015-0587) [DOI] [PubMed] [Google Scholar]
- 85.Alves TG, Kasamatsu TS, Yang JH, et al. Macrocalcitonin is a novel pitfall in the routine of serum calcitonin immunoassay. J Clin Endocrinol Metab 2016. 101 653–658. ( 10.1210/jc.2015-3137) [DOI] [PubMed] [Google Scholar]
- 86.Jensen EA, Petersen PH, Blaabjerg O, et al. Establishment of reference distributions and decision values for thyroid antibodies against thyroid peroxidase (TPOAb), thyroglobulin (TgAb) and the thyrotropin receptor (TRAb). Clin Chem Lab Med 2006. 44 991–998. ( 10.1515/CCLM.2006.166) [DOI] [PubMed] [Google Scholar]
- 87.Evans M, Sanders J, Tagami T, et al. Monoclonal autoantibodies to the TSH receptor, one with stimulating activity and one with blocking activity, obtained from the same blood sample. Clin Endocrinol 2010. 73 404–412. ( 10.1111/j.1365-2265.2010.03831.x) [DOI] [PubMed] [Google Scholar]
- 88.George A, Lotz J, Luffy M, et al. A novel bioassay for thyroid-blocking immunoglobulins. Front Endocrinol 2014. 15 1463379. ( 10.3389/fendo.2024.1463379) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Olivo PD, Kim H, Miao L, et al. Analytical validation of a novel bioassay for thyroid-stimulating immunoglobulin. Front Endocrinol 2025. 15 1468768. ( 10.3389/fendo.2024.1468768) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Elston MS, Sehgal S, Du Toit S, et al. Factitious Graves’ disease due to biotin immunoassay interference-a case and review of the literature. J Clin Endocrinol Metab 2016. 101 3251–3255. ( 10.1210/jc.2016-1971) [DOI] [PubMed] [Google Scholar]
- 91.Al-Bahadili H, Powers Carson J, Markov A, et al. The complex web of interferences with thyroid function tests. Endocr Pract 2025. 31 92–101. ( 10.1016/j.eprac.2024.10.007) [DOI] [PubMed] [Google Scholar]
- 92.D’Aurizio F, Metus P, Ferrari A, et al. Definition of the upper reference limit for thyroglobulin antibodies according to the National Academy of Clinical Biochemistry Guidelines: comparison of eleven different automated methods. Auto Immun Highlights 2017. 8 8. ( 10.1007/s13317-017-0096-3) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Bech K, Feldt-Rasmussen U, Bliddal H, et al. The acute changes in thyroid stimulating immunoglobulins, thyroglobulin and thyroglobulin antibodies following subtotal thyroidectomy. Clin Endocrinol 1982. 16 235–242. ( 10.1111/j.1365-2265.1982.tb00712.x) [DOI] [PubMed] [Google Scholar]
- 94.Feldt-Rasmussen U & Rasmussen AK. Serum thyroglobulin (Tg) in presence of thyroglobulin autoantibodies (TgAb). Clinical and methodological relevance of the interaction between Tg and TgAb in vitro and in vivo. J Endocrinological Invest 1985. 8 571–576. ( 10.1007/BF03348564) [DOI] [PubMed] [Google Scholar]
- 95.Cubero JM, Rodríguez-Espinosa J, Gelpi C, et al. Thyroglobulin autoantibody levels below the cut-off for positivity can interfere with thyroglobulin measurement. Thyroid 2003. 13 659–661. ( 10.1089/105072503322240013) [DOI] [PubMed] [Google Scholar]
- 96.Jiménez-Legido M, Cantarín-Extremera V, López-Guio ME, et al. Value of thyroid peroxidase antibodies in neuroimmune diseases: analysis of interference during treatment with intravenous immunoglobulins. Lab Med 2022. 53 426–432. ( 10.1093/labmed/lmac013) [DOI] [PubMed] [Google Scholar]
- 97.Bliddal S, Nielsen CH & Feldt-Rasmussen U. Recent advances in understanding autoimmune thyroid disease: the tallest tree in the forest of polyautoimmunity. F1000Res 2017. 6 1776. ( 10.12688/f1000research.11535.1) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Laurberg P, Cerqueira C, Ovesen L, et al. Iodine intake as a determinant of thyroid disorders in populations. Best Pract Res Clin Endocrinol Metab 2010. 24 13–27. ( 10.1016/j.beem.2009.08.013) [DOI] [PubMed] [Google Scholar]
- 99.Tozzoli R, D’Aurizio F, Ferrari A, et al. The upper reference limit for thyroid peroxidase autoantibodies is method-dependent: a collaborative study with biomedical industries. Clin Chim Acta 2016. 452 61–65. ( 10.1016/j.cca.2015.10.025) [DOI] [PubMed] [Google Scholar]
- 100.Haapala AM, Hyöty H, Parkkonen P, et al. Antibody reactivity against thyroid peroxidase and myeloperoxidase in autoimmune thyroiditis and systemic vasculitis. Scand J Immunol 2016. 46 78–85. ( 10.1046/j.1365-3083.1997.d01-90.x) [DOI] [PubMed] [Google Scholar]
- 101.Freire BA, Paula ID, Paula F, et al. Absence of cross-reactivity to myeloperoxidase of anti-thyroid microsomal antibodies in patients with autoimmune thyroid diseases. Am J Med Sci 2001. 321 109–112. ( 10.1097/00000441-200102000-00001) [DOI] [PubMed] [Google Scholar]
- 102.Persani L, Rodien P, Moran C, et al. 2024 European Thyroid Association Guidelines on diagnosis and management of genetic disorders of thyroid hormone transport, metabolism and action. Eur Thyroid J 2024. 13 e240125. ( 10.1530/ETJ-24-0125) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 103.Beck-Peccoz P, Lania A, Beckers A, et al. 2013 European thyroid association guidelines for the diagnosis and treatment of thyrotropin-secreting pituitary tumors. Eur Thyroid J 2013. 2 76–82. ( 10.1159/000351007) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Gruson D, Dabla P, Stankovic S, et al. Artificial intelligence and thyroid disease management: considerations for thyroid function tests. Biochem Med 2022. 32 020601. ( 10.11613/BM.2022.020601) [DOI] [PMC free article] [PubMed] [Google Scholar]
- 105.Yang HS, Pan W, Wang Y, et al. Generalizability of a machine learning model for improving utilization of parathyroid hormone-related peptide testing across multiple clinical centers. Clin Chem 2023. 69 1260–1269. ( 10.1093/clinchem/hvad141) [DOI] [PubMed] [Google Scholar]
- 106.Greaves R, Kricka L, Gruson D, et al. Toolkit for emerging technologies in laboratory medicine. Clin Chem Lab Med 2023. 61 2102–2114. ( 10.1515/cclm-2023-0571) [DOI] [PubMed] [Google Scholar]
- 107.Tate J & Ward G. Interferences in immunoassay. Clin Biochem Rev 2004. 25 105–120. [PMC free article] [PubMed] [Google Scholar]
- 108.Ismail AA. Identifying and reducing potentially wrong immunoassay results even when plausible and “not-unreasonable”. Adv Clin Chem 2014. 66 241–294. ( 10.1016/b978-0-12-801401-1.00007-4) [DOI] [PubMed] [Google Scholar]
- 109.Fahie-Wilson M & Smith TP. Determination of prolactin: the macroprolactin problem. Best Pract Res Clin Endocrinol Metab 2013. 27 725–742. ( 10.1016/j.beem.2013.07.002) [DOI] [PubMed] [Google Scholar]
- 110.Favresse J, Stoefs A, Bayart JL, et al. A challenging case of falsely elevated free thyroid hormones. J Appl Lab Med 2020. 5 406–411. ( 10.1093/jalm/jfz027) [DOI] [PubMed] [Google Scholar]
- 111.Lippi G, Giavarina D, Gelati M, et al. Reference range of hemolysis index in serum and lithium-heparin plasma measured with two analytical platforms in a population of unselected outpatients. Clin Chim Acta 2014. 429 143–146. ( 10.1016/j.cca.2013.12.010) [DOI] [PubMed] [Google Scholar]
- 112.Piketty ML, Prie D, Sedel F, et al. High-dose biotin therapy leading to false biochemical endocrine profiles: validation of a simple method to overcome biotin interference. Clin Chem Lab Med 2017. 55 817–825. ( 10.1515/cclm-2016-1183) [DOI] [PubMed] [Google Scholar]

This work is licensed under a 