Abstract
Background
Consumer wearables are increasingly being integrated into health research for data collection. Although they are attractive to use, the accuracy of their photoplethysmography (PPG)-based measurements can be influenced by user characteristics such as sex, age, BMI, and skin tone. However, our knowledge regarding the validity of these measurements in certain populations, such as those with darker skin tones, seems limited. This is concerning as uncorrected differences in measurement accuracy can lead to health disparities when consumer wearable measurements are used more frequently. A potential cause for the gap in our knowledge regarding consumer wearable validity is the underrepresentation of certain population groups in studies validating PPG-based consumer wearables.
Objective
This scoping review aimed to map the representation of different sex, age, BMI, and skin tone groups in studies assessing the validity of PPG-based pulse rate, heart rate variability, blood pressure, peripheral blood oxygen saturation (SpO2), and respiratory rate measurements of consumer wearables.
Methods
A literature search was conducted in Scopus, PubMed, and IEEE Xplore in July 2025. Papers were eligible if they assessed the validity of consumer wearable PPG-based measurements, expressed as the agreement with a reference method. From the included papers, the study population distribution of sex, age, BMI, and Fitzpatrick scale was extracted. To evaluate the representation, percentages of people in specific age, BMI, and skin tone groups were estimated based on reported means and SDs. The median percentage of participants in each population group, as well as the total percentage, is reported.
Results
After the removal of duplicates, 734 papers were screened for eligibility. Following title and abstract screening, 238 papers remained, of which 186 passed full-text screening and were included in the review. Most of the studies (n=160) focused on pulse rate. Sex, age, BMI, and Fitzpatrick scale were reported by 179 (96.0%), 178 (96.0%), 101 (54.0%), and 35 (19.0%) out of 186 studies, respectively. While the median representation was 0% (IQR 0%-8%) for both older adults (>65 y) and individuals with obesity (BMI>30 kg/m2; IQR 0%-13%), aggregate participation across all studies was higher (1290/6367, 20.0% and 473/3428, 14.0%, respectively). Individuals with underweight (BMI<18.5 kg/m2) remained rare (median 3%, IQR 0%-7%), and the aggregate was 7.0% (225/3428). The median percentage of people with darker skin tones (Fitzpatrick type V and VI) participating in a study was 0%.
Conclusions
Based on our results, it can be concluded that older adults and people with underweight, obesity, or darker skin tones are generally underrepresented in studies assessing the validity of consumer wearable PPG-based measurements. Future validation studies should focus more on the representativeness of the study population. This can be achieved by setting a benchmark for representativeness and including study population representatives during the study design process.
Introduction
In 2012, the first commercial wrist-worn pulse rate sensor was released via a Kickstarter campaign [1]. It was developed for athletes to keep track of their pulse rate during training without having to wear a chest strap. Now, a little more than a decade later, the sensing functionalities of consumer wearables have expanded significantly. In addition to pulse rate, these wearables measure, among others, heart rate variability (HRV), peripheral blood oxygen saturation (SpO2), respiratory rate, blood pressure, skin temperature, and skin conductance. Using these measurements, composite measures such as stress and sleep quality scores are also derived. The added sensing functionalities have broadened the use of consumer wearables from a tool to enhance exercise performance in athletes to health and well-being monitoring for consumers. It is forecasted that the number of smartwatch users will reach 740.53 million in 2029 [2].
Not only have consumers embraced wearables, but the use of consumer wearables in health research is also increasing. A scoping review published in 2022, mapping the use of wearables in health research, identified 179 studies, of which 163 used a consumer wearable [3]. An increase in the number of studies using wearables published per year was observed, rising from 1 in 2013 to 21 in 2017 and doubling to 56 in 2020. In these studies, consumer wearables were implemented for various purposes, ranging from examining correlations between wearable-derived outcomes and other physiological outcomes to developing risk prediction models, disease screening, and monitoring.
The integration of consumer wearables into health research is attractive, as they enable continuous and remote measurements of diverse outcomes while being more comfortable for participants and often more affordable than medical-grade or research-grade sensors. It should be noted, however, that many consumer wearable measurements, including pulse rate, HRV, SpO2, blood pressure, and respiratory rate, are based on photoplethysmography (PPG). PPG is an optical measurement technique in which light is used to detect changes in blood volume in the skin microvascular bed [4]. The signal produced by this technique is sensitive to external factors, such as environmental light and movement, as well as user characteristics that influence the optical properties of the skin, such as sex, age, BMI, and skin tone [5-7]. For example, melanin in the skin, a determinant of skin tone, absorbs and scatters light, which influences the PPG signal quality [5]. The validity of PPG-based measurements thus varies between measurement conditions and users.
In the last few years, concerns have been raised about our limited knowledge regarding the validity of PPG-based consumer wearable measurements in certain population groups, such as people with darker skin tones or older individuals [8-10]. As manufacturers often do not report the validity of their consumer wearable measurements, researcher-initiated validation studies are being conducted to compare consumer wearable measurements with a reference method. Thus far, these validation studies have reported the accuracy of consumer wearable measurements under different conditions, such as during rest, physical activity, or sleep, in both controlled laboratory settings and free-living conditions [11,12]. However, the influence of characteristics such as sex, age, BMI, and skin tone on measurement validity remains largely unclear, and studies investigating this influence appear to be scarce [13-15]. If potential differences in measurement validity remain unidentified, these differences could contribute to variations in the accuracy of consumer wearable–based risk prediction, disease screening, and monitoring tools across populations. In turn, this may lead to an increase in health disparities—defined as preventable differences in the opportunity to achieve optimal health—as the integration of consumer wearables into health research and, eventually, health care progresses [8-10].
It is hypothesized that the underrepresentation of certain population groups in studies assessing the validity of PPG-based consumer wearable measurements contributes to the gaps in our knowledge about the validity of these measurements in different populations. To improve our understanding of these topics, the potential gaps in the representation of different population groups need to be identified and addressed. The aim of this scoping review is to systematically map the representation of different sex, age, BMI, and skin tone population groups in studies assessing the validity of PPG-based consumer wearable measurements, including pulse rate, HRV, blood pressure, blood oxygen saturation, and respiratory rate. Additionally, we will evaluate the representation of these different population groups to determine which groups require increased representation in future validation studies to enable the investigation of differences in measurement validity between population groups.
Methods
Study Design
This study was performed between May 2024 and August 2025 using the methodological framework for scoping reviews proposed by Arksey and O’Malley [16]. The registration and protocol for this review are available on the Open Science Framework (OSF) [17]. This review is reported according to the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) extension for scoping reviews (PRISMA-ScR; Checklist 1) guidelines [18].
Eligibility Criteria
Papers eligible for inclusion in this scoping review were those assessing the criterion validity, the agreement of a new measurement method with a criterion measure, of one or more commercial wearable sensors for measuring pulse rate, HRV, blood pressure, respiratory rate, and/or oxygen saturation in humans. The following exclusion criteria were applied:
The wearable sensor that was validated is a chest strap, as chest straps are considered a criterion measure [19].
The wearable sensor that was validated is not and was not at any point commercially available to consumers, as this scoping review focuses on consumer wearables only.
The wearable sensor that was validated is not meant to be worn continuously during daily activities, as we consider this an important characteristic of a true consumer wearable device.
The wearable measurement that was validated was measured using a method other than PPG, as this scoping review focuses on PPG-based measurements only.
N=1 or case studies are excluded as this review has a focus on the distribution of characteristics within study populations.
Animal studies are excluded, as this scoping review focuses on human participants.
Reviews, editorials, opinion pieces, and protocols are excluded, as these are not primary research papers.
No full text could be obtained, or insufficient information was provided in the full text. At least the sample size, the reference method used, and the name and brand of the wearable used should be mentioned to be able to verify if other exclusion criteria apply.
Publication languages other than English are not considered, as English is the language in which all involved reviewers are proficient.
Search Strategy
A literature search was conducted on June 11, 2024, in the databases Scopus, PubMed, and IEEE Xplore. To update the results before publication, the same search was repeated on July 16, 2025. The main concepts included in the search strategy were “PPG-based measurements,” “validity,” and “consumer wearable.” The keywords for these concepts were selected through pilot searches. The search strategy for all 3 databases can be found in Multimedia Appendix 1. Additionally, the reference lists of reviews returned by the database search were reviewed to identify additional papers.
Screening and Selection
The database search results were uploaded to EndNote 20 (Clarivate) for duplicate removal, after which the remaining papers were uploaded to Rayyan for screening. Prior to initiating the official screening process, the screening guide was tested by 2 reviewers (RMS and FOD) during a pilot screening round with 10 papers. There were no conflicts between the 2 reviewers, and no adjustments were made to the screening guide. The reviewers, RMS and FOD, then proceeded to independently screen the papers for eligibility based on title and abstract, after which a full-text screening took place. Conflicts were resolved through discussion between the 2 reviewers.
Data Charting
Extracted data included information on the type of publication, wearable, study context, recruitment and eligibility, and study population. A complete overview of the data items extracted is presented in Table 1. Most studies do not report how they assessed sex; therefore, it is often unclear whether they report sex or gender. For consistency within this review, we will refer to this characteristic as sex. For skin tone, it was decided to extract the Fitzpatrick scale as this is how most studies report skin tone [13], and as it is the measure recommended by several guidelines, including the US Food and Drug Administration (FDA) guideline on pulse oximeter evaluation, the guidelines for the evaluation of consumer wearable pulse rate validity published by the INTERLIVE network, and the ANSI/CTA (American National Standards Institute/Consumer Technology Association) standard [19-21].
Table 1. Extracted data items from 186 studies assessing the validity of consumer wearable photoplethysmography-based measurements.
| Category | Extracted information |
|---|---|
| Publication information | Title, year of publication, first author, and DOI |
| Wearable | Brand, model, version of operating system, and intended wear location |
| Context | Study location, reference device, and level of physical activity |
| Recruitment and eligibility | Recruitment location, recruitment channels, and eligibility criteria |
| Study population | Sample size, distribution of sex, age, height, weight, BMI, and Fitzpatrick scale, and disease status |
The data charting form was created by RMS in Microsoft Excel. This form was used independently by 2 data extractors in a data charting pilot (RMS and FOD) with 10% of the included papers. For the main data items of interest, for example, study population characteristics, the percentage agreement between the 2 extractors was 95.5%. However, the agreement on the level of physical activity was low (40%). Therefore, additional guidelines for extracting the level of physical activity were added to the data extraction form, after which RMS proceeded with charting the data of the remaining papers independently.
Synthesis of Results
With the extracted data, the number of studies per PPG-based measurement, the number of wearable brands and models, studies per year, and per study location were determined. Additionally, the recruitment locations and eligibility criteria were summarized.
Evaluating Representation
To provide an overview of the representation of population groups, 2 different approaches were applied. First, for the characteristic sex, the percentage of women in a study was extracted or calculated based on the number of women and the sample size. Based on this percentage, studies were categorized as equal distribution (at least 40% men or women), men-dominant or women-dominant (more than 60% men or women, respectively), men only, or women only. As none of the studies reported the participation of intersex individuals, this review can therefore only report the distribution of men and women.
Second, for age, weight, height, BMI, and Fitzpatrick scale, the weighted mean (SD) was calculated to capture the distribution of these characteristics in the included studies. Additionally, for age, BMI, and Fitzpatrick scale, the percentage of participants in commonly used groups of these characteristics was estimated to better visualize the representation of these different groups (Textbox 1) [22-24]. The percentages were estimated based on probability, using the mean and SD reported by the included studies [25]. First, the z-scores for the boundaries of each group were calculated using the provided mean and SD. Second, to estimate the probability of a value equal to or less than the boundary z-scores occurring, known as the cumulative distribution function, the pnorm function was used in RStudio (version 2023.6.1.524) with R (version 4.4.1). The percentage of participants in a group was then estimated by subtracting the cumulative distribution function of the lower boundaries from the cumulative distribution function of the upper boundaries. A threshold was applied to ensure that percentages lower than the contribution of 1 participant were rounded to 0. For age, the distribution was often truncated at 18 years of age, as studies only included adults. In these cases, the ptruncnorm function with a lower bound of 18 was used instead of the pnorm function.
Textbox 1. Age, BMI, and skin tone categories used in a scoping review of 186 studies assessing the validity of consumer PPG-based measurements.
Age categories (years)
Children: <13
Adolescents: 13‐17
Adults: 18‐65
Older adults: ≥65
BMI categories (kg/m2)
Underweight: <18.5
Normal weight: 18.5‐25
Overweight: 25‐30
Obesity: ≥30
Skin tone categories
Fitzpatrick type I: always burns, never tans
Fitzpatrick type II: usually burns, tans less than average
Fitzpatrick type III: sometimes burns mildly, tans about average
Fitzpatrick type IV: rarely burns, tans more than average
Fitzpatrick type V: rarely burns, brown tone
Fitzpatrick type VI: rarely burns, dark brown or black tone
To visualize the distribution of the representation across studies, the median (IQR) of the percentage of participants within each category was reported. To quantify the representation in all studies combined, the total number of participants within each category was calculated based on the estimated percentages. All plots and quantitative analyses were performed in RStudio (version 2023.6.1.524) with R (version 4.4.1).
To evaluate whether a population group was sufficiently represented, the median and total percentage of participants in this population group were compared to a predetermined benchmark [26]. For the purpose of this review, the selected benchmark is equal representation of all population groups. For example, for BMI with 4 groups, the percentage of participants in each BMI group should approach 25%. This benchmark was selected as equal representation of all population groups is desirable when statistically comparing group results.
Results
Study Selection
A total of 1403 papers were identified and extracted from PubMed, Scopus, and IEEE Xplore (Figure 1), of which 695 were identified as duplicates and removed, leaving 708 papers for title and abstract screening. Based on title and abstract screening, 500 papers were excluded, and an additional 5 papers were excluded because the full text could not be retrieved. Screening the reference lists of relevant reviews identified 29 additional papers for full-text screening. In the end, 233 papers were included in full-text screening. Based on full-text screening, 187 papers remained for inclusion, providing results from 186 individual studies.
Figure 1. PRISMA flowchart. PPG: photoplethysmography; PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses.

General Study Characteristics
Of the 186 included studies (n=8786 participants), 160 (n=6461 participants) examined pulse rate, 25 (n=2053) oxygen saturation, 19 (n=887) HRV, 7 (n=326) blood pressure [27-33], and 5 (n=329) respiration rate [34-38]. The median (IQR) sample size was 30 (19-49) participants [19,39]. The publication year ranged from 2015 to 2025 (Figure 2), and studies were performed in 32 countries (Table 2). Across all studies, 33 wearable brands and 129 different models were used. Most of the examined wearables were wrist-worn (119 models), and some were worn on the upper arm (3 models), the finger (2 models) [40-44], or in the ear (5 models) [42,45-47]. Most studies (106 studies) validated 1 wearable device; the remaining 80 studies validated 2 to 7 wearable devices. Additional information about the study characteristics of studies for each examined PPG-based measurement can be found in Multimedia Appendix 2. The extracted data for each paper can be found in Multimedia Appendix 3.
Figure 2. Number of studies validating consumer wearable measurements per photoplethysmography-based measurement per year, including studies published until July 16, 2025. HRV: heart rate variability; SpO2: oxygen saturation.

Table 2. Number of studies validating consumer wearable photoplethysmography-based measurements by study location at the country level.
| Country | Number of studies, n | References |
|---|---|---|
| United States | 54 | [6,32,39,42,43,45,47-94] |
| Italy | 13 | [27,30,35,95-104] |
| Germany | 12 | [34,36,46,105-113] |
| Australia | 11 | [41,114-123] |
| Canada | 10 | [124-133] |
| Republic of Korea | 9 | [37,134-141] |
| United Kingdom | 9 | [142-150] |
| Spain | 8 | [151-158] |
| Belgium | 6 | [28,38,159-162] |
| Brazil | 6 | [29,33,163-166] |
| Denmark | 6 | [167-172] |
| Switzerland | 6 | [44,173-177] |
| Finland | 4 | [40,178-180] |
| France | 3 | [181-184] |
| Ireland | 3 | [185-187] |
| Singapore | 3 | [188-190] |
| Austria | 3 | [100,191,192] |
| China | 2 | [31,193] |
| Czech Republic | 2 | [194,195] |
| Netherlands | 2 | [196,197] |
| Saudi Arabia | 2 | [198,199] |
| Taiwan | 2 | [200,201] |
| Turkey | 2 | [202,203] |
| India | 1 | [204] |
| Japan | 1 | [205] |
| Malaysia | 1 | [206] |
| New Zealand | 1 | [207] |
| Norway | 1 | [208] |
| Philippines | 1 | [209] |
| Poland | 1 | [210] |
| Sweden | 1 | [211] |
| Thailand | 1 | [212] |
Participant Eligibility and Recruitment
Eligibility Criteria
Of the 186 studies, 150 studies provided eligibility criteria for their study populations, and 69 (37.0%; n=3640 participants) studies stated that adults were eligible for participation. Moreover, there were 31 of 186 studies (7.0%; n=1001 participants) that only included younger adults, 10 (5.0%; n=684 participants) studies that only included children and/or adolescents, 4 (2.0%; n=97 participants) that only included older adults [38,92,163,185], and 1 (n=60 participants) that included children, adolescents, and adults [178]. Eight studies used BMI-based exclusion criteria, either excluding people who were underweight (3 studies) [78,116,188], people with overweight (1 study) [176], people with obesity (2 studies) [77,177], people with clinical obesity (BMI≥40 kg/m2, 1 study) [108], or both people who were underweight and overweight (1 study) [131].
Health Status
According to the description of the populations, out of 186 studies, most of them (n=137, 74.0%) included healthy populations. There were 47 (25.0%) studies that used the Physical Activity Readiness Questionnaire to screen participants for inclusion. Exclusion criteria for cardiovascular diseases (n=81, 44.0%), musculoskeletal injuries, diseases, or disabilities (n=46, 25.0%), and pregnancy (n=11, 6.0%) were common. Of 186 studies, 49 (26.0%; n=4921 participants) were performed on patient populations with various conditions. These conditions included cardiovascular diseases (21 studies), respiratory diseases (8 studies) [38,49,53,82,101,166,169,202], surgeries (4 studies) [33,36,98,108], disabilities (3 studies) [167,197,208], breast cancer (1 study) [92], COVID-19 (1 study) [123], chronic musculoskeletal pain (1 study) [211], and cystic fibrosis (1 study) [116].
Recruitment Location
Of the 186 studies, 91 (49.0%) studies provided information on the recruitment location; 46 (25.0%) studies recruited participants from a hospital, and 26 (14.0%) studies recruited participants from a university or research group exclusively. Other recruitment settings included local communities and associations (n=9, 5.0%) [55,80,92,132,136,140,142,185,187], patient organizations (n=3, 2.0%) [167,197,208], professional environments such as the military (n=3, 2.0%) [121,174,176], and a physical activity challenge (n=1, 1.0%) [188]. One study included participants from a previous trial [148].
Reporting of Study Population Characteristics
In the 186 studies reviewed, the most reported study population characteristics were age and sex (n=178, 96.0%, and n=179, 96.0%, respectively), followed by BMI (n=101, 54.0%), weight (n=102, 55.0%), height (n=98, 53.0%), and finally the Fitzpatrick scale (n=35, 19.0%; Table 3). Over time, the percentage of studies reporting weight and height remained stable at ~50% (Figure 3). The percentage of studies reporting BMI fluctuated over time, with a minimum of 0% (0/2) in 2015 and a maximum of 77.0% (10/13) in 2025. An upward trend in the reporting of the Fitzpatrick scale was observed from 2019 to 2025 (1/24, 4.0% to 5/13, 38.0%).
Table 3. Number (%) of consumer wearable photoplethysmography (PPG)-based measurement validation studies reporting sex, age, BMI, weight, height, and Fitzpatrick scale.
| Characteristics | All | PRa | HRVb | SpO2c | BPd | RRe |
|---|---|---|---|---|---|---|
| Studies, N | 186 | 160 | 19 | 25 | 7 | 5 |
| Sex, n (%) | 179 (96.0) | 157 (99.0) | 16 (84.0) | 23 (92) | 7 (100) | 5 (100) |
| Age, n (%) | 178 (96.0) | 156 (98.0) | 16 (84.0) | 22 (88) | 7 (100) | 4 (80) |
| BMI, n (%) | 101 (54.0) | 87 (54.0) | 7 (37.0) | 13 (52) | 3 (43.0) | 4 (80) |
| Weight, n (%) | 102 (55.0) | 93 (58.0) | 6 (32.0) | 12 (48) | 1 (14.0) | 1 (20) |
| Height, n (%) | 98 (53.0) | 89 (56.0) | 5 (26.0) | 12 (48) | 1 (14.0) | 1 (20) |
| Fitzpatrick scale, n (%) | 35 (19.0) | 31 (19.0) | 3 (16.0) | 9 (36) | 1 (14.0) | 2 (40) |
| All, n (%) | 15 (8.0) | 14 (9.0) | 2 (11.0) | 3 (12) | 0 (0) | 1 (20) |
PR: pulse rate.
HRV: heart rate variability.
SpO2: oxygen saturation.
BP: blood pressure.
RR: respiratory rate.
Figure 3. The percentage of consumer wearable photoplethysmography-based measurement validation studies per year reporting BMI (A), Fitzpatrick scale (B), weight (C), and height (D).

Age, BMI, weight, and height were mostly reported as mean or median combined with an SD, SE, or IQR (age: 158/178, 89.0%; BMI: 98/101, 97.0%; weight: 98/102, 96.0%; height: 94/98, 96.0%). A lower percentage of studies additionally reported the range (age: 44/178, 25.0%; BMI: 25/101, 25.0%; weight: 25/102, 25.0%; height: 23/98, 23.0%). Reporting of the Fitzpatrick scale was highly variable, with studies either reporting a combination of mean, median, SD, or IQR (12/35, 34.0%), the percentage of participants in each level or in a combination of at least 2 levels (6/35, 17.0%, for both), the range (5/35, 14.0%), or a combination of these options (6/35, 17.0%).
Representation of Study Population Characteristics
Representation of Men and Women
In 179 studies (n=8203 participants), the percentage of women in a study was 43±19%. Most of the studies (102 studies, n=4570 participants) included an approximately equal percentage of men and women. There were 46 men-dominant studies (n=2637 participants), 17 women-dominant studies (n=688 participants), 11 men-only (n=277 participants) [59,121,135,146,157,164,167,174,199], and 3 women-only (n=31 participants) studies [35,136,213].
Representation of Age Groups
The weighted mean (SD) of age across 149 studies (n=6367) reporting these statistics was 44 (11.9) years. The lowest reported age was 1.2 months [127], and the highest reported age was 94 years [126,202]. In the 112 studies (n=5088), which did not exclude older adults and reported the mean (SD), the median (IQR) percentage of older adults in a study was 0% (0%-8%). The total estimated number of older adults participating across the 149 studies that reported the mean (SD) was 1290 of 6367 participants (20.0%). The results of each examined PPG-based measurement separately are presented in Table 4 and Multimedia Appendix 4. Of these 1290 estimated older adults, 1247 (97.0%) participated in studies with a focus on patient populations. From 2021 to 2024, the total percentage of older adults participating each year increased from 23% to 38% (Multimedia Appendix 5).
Table 4. Percentage (median, IQR) of older adults and participants belonging to BMI and skin tone groups in consumer wearable photoplethysmography-based measurement validation studies.
| Participant characteristics | PRa,b, median (IQR) | HRVc,d, median (IQR) | BPe,f, median (IQR) | SpO2g,h, median (IQR) | RRi,j, median (IQR) |
|---|---|---|---|---|---|
| Age | |||||
| Older adults | 0 (0‐0) | 0 (0‐0) | 11 (2-28) | 6 (0‐27) | 11 (6-14) |
| BMI | |||||
| Underweight | 4 (0‐7) | 6 (3-7) | 4 (2-6) | 3 (0‐3) | 3 (1-3) |
| Normal weight | 58 (41-72) | 42 (36-46) | 23 (30-37) | 34 (26-50) | 37 (26-68) |
| Overweight | 31 (17-36) | 34 (27-36) | 34 (30-37) | 40 (28-41) | 27 (13-35) |
| Obesity | 0 (0‐12) | 13 (9-19) | 37 (36-38) | 20 (0‐33) | 15 (8-34) |
| Fitzpatrick scale | |||||
| I | 15 (7-24) | 18 (16-20) | —k | 15 (3-21) | — |
| II | 35 (19-45) | 32 (24-41) | — | 34 (26-46) | — |
| III | 25 (14-30) | 20 (20,21) | — | 21 (15-30) | — |
| IV | 9 (3-21) | 9 (4-13) | — | 10 (2-18) | — |
| V | 0 (0‐9) | 9 (4-13) | — | 1 (0‐3) | — |
| VI | 0 (0‐0) | 9 (5-14) | — | 0 (0‐2) | — |
PR: pulse rate.
Age: n=132; BMI: n=78; Fitzpatrick scale: n=16.
HRV: heart rate variability.
Age: n=12; BMI: n=5; Fitzpatrick scale: n=2.
BP: blood pressure.
Age: n=6; BMI: n=2.
SpO2: peripheral blood oxygen saturation.
Age: n=16; BMI: n=8; Fitzpatrick scale: n=6.
RR: respiratory rate.
Age: n=5; BMI: n=3.
Not applicable.
Representation of BMI Categories
Based on 89 studies (n=3428), the weighted mean (SD) of BMI was 24.8 (4.2) kg/m2, with BMIs ranging from 16.3 [118] to 52.0 kg/m2 [34]. In these studies, the median (IQR) percentage of people with underweight, normal weight, overweight, and obesity in a study was 3% (0%-7%), 51% (38%-70%), 31% (18%-37%), and 0% (0%-13%), respectively (Figure 4). Of the 3428 participants, it was estimated that 225 (7.0%) were underweight, 1602 (47.0%) had normal weight, 1048 (31.0%) were overweight, and 473 (14.0%) had obesity. Of the 473 estimated participants with obesity, 364 (77.0%) participated in studies that focused on patient populations. The representation of people with obesity increased over time from 4% in 2016 to 22% in 2022, after which it decreased to 15% in both 2023 and 2024 (Multimedia Appendix 5).
Figure 4. Boxplot of the distribution of the estimated percentage of participants in 4 BMI (kg/m2) categories in 89 studies (n=3428 participants) validating consumer wearable photoplethysmography-based measurements.

Representation of Fitzpatrick Scale Categories
In the 19 studies (n=876) that reported the mean (SD) or percentage of participants in each Fitzpatrick scale category, the majority of participants fell in type II (median 33%, IQR 20%-42%), while the minority were in types V and VI (median 0%, IQR 0%-9% and median 0%, IQR 0%-1%, respectively; Figure 5). Over the years, the representation of the different categories fluctuated, with no clear trend toward increasing or decreasing representation of any category (Multimedia Appendix 5).
Figure 5. Boxplot of the distribution of the estimated percentage of participants in the Fitzpatrick scale categories in 19 studies (n=876 participants) validating consumer wearable photoplethysmography-based measurements.

Discussion
Principal Findings
This scoping review aimed to provide an overview of the representation of different sex, age, BMI, and skin tone groups in studies assessing the validity of PPG-based consumer wearable measurements. Based on the inclusion of 186 studies, our results indicate that the representation of the different population groups is not equal. In particular, older adults, people who were underweight, people living with obesity, and people with darker skin tones (Fitzpatrick type V or VI) are underrepresented compared to other groups. When interpreting our findings, it is important to consider the following points of discussion: (1) the reporting of user characteristics by the included studies; (2) the benchmark used for evaluating representation; and (3) the relevance of evaluating both study-level and total representation.
First, the results of this study revealed that a large percentage of studies assessing the validity of consumer wearables’ PPG-based measurements did not report BMI and skin tone. Not only does this limit our ability to fully evaluate the representation of these characteristics in this review, but it also indicates a critical gap in the available evidence. The nonreporting of these characteristics provides a barrier to investigating the influence of these characteristics on measurement validity. To increase the reporting of study population characteristics like BMI and skin tone, reporting guidelines for consumer wearable validation studies, such as those provided by the INTERLIVE network and the ANSI/CTA, could help. These reporting guidelines recommend the reporting of study populations’ sex, age, weight, height, BMI, and skin tone distribution for studies validating pulse rate measurements and could be extended toward other PPG-based validation studies [19,20]. To truly achieve improved reporting, guidelines like those from the INTERLIVE network should additionally be endorsed by journals that publish validation studies [214,215].
Second, the benchmark used to evaluate representation impacts what is considered sufficient representation and underrepresentation. For example, using equal representation as a benchmark, we consider people with underweight, with a median representation of 3%, to be underrepresented. However, using the prevalence of underweight in the general population, namely 1.6%, as a benchmark, the representation of people who are underweight would be considered sufficient [216,217]. The most suitable benchmark depends on the study’s aim. When the aim is to investigate the influence of sex, age, BMI, and skin tone on wearable measurement validity, equal representation of these groups is a suitable benchmark. However, if a validation study aims to reflect performance in a specific target population or the general population, then equal representation of all groups would not be a suitable benchmark. For this purpose, the prevalence of sex, age, BMI, and skin tone groups in the target or general population should be used as a benchmark. The Clinical Trial Diversity Rating framework, developed by the Institute for Clinical and Economic Review, can then be used to evaluate if this benchmark is sufficiently met [26]. Selecting a suitable benchmark is an important step toward achieving study population representativeness.
Third, in this review, we evaluated representation at the level of individual studies as well as total representation. There are population groups that had low representation at both levels, such as children and adolescents, people who are underweight, and people with darker skin tones (Fitzpatrick types V and VI). However, for older adults and people with obesity, although the representation at the study level was very low, the overall representation was higher. This discrepancy can occur if there is a particular subset of studies in which a group is highly represented, while in other studies, the representation is low. For example, in this scoping review, there was a subset of studies with large sample sizes conducted in patient populations in which older adults were highly represented, while in the remaining studies with smaller sample sizes, almost no older adults participated. In these validation studies, the older adults represented are thus mostly older adults living with disease, who might not be an accurate representation of healthy older adults. This pattern was only observed because we evaluated representation both at the level of individual studies and at the overall level, indicating the relevance of evaluating both levels. In addition to evaluating representation at these 2 levels, it would have been valuable to examine the representation of demographic characteristics intersectionally. Based on the information provided by the included studies, this was not feasible. In reality, the 4 characteristics investigated in this review coexist in one individual, and their cumulative effect is what influences signal quality. An older individual with obesity and dark skin, for example, could present an “intersectional stress test” that current algorithms frequently fail. This necessitates stratified sampling and reporting in future validation studies.
Increasing the Representation of Underrepresented Population Groups
There are 2 explanations for the underrepresentation observed in this review. The first cause is the active exclusion of some population groups through exclusion criteria. For example, many of the validation studies investigating pulse rate measurement validity included in this review had a vigorous exercise component, which can be harmful to some population groups, including older adults. Likely due to this ethical consideration, 31 studies excluded older adults. The second cause is barriers to participation related to other elements of the study. These elements included the research question, the composition of the research team, the recruitment methods, the complexity of research processes, and the required investment of time and resources from the participants [218]. For example, none of the studies had skin tone-based eligibility criteria, yet the representation of people with darker skin tones was very low, which suggests that some other study elements could have prevented their participation.
Irrespective of the cause, the underrepresentation of older adults, people who are underweight, people with obesity, and people with darker skin tones currently provides a barrier for us to gain a better understanding of the differences in measurement validity between population groups. If there are relevant differences in consumer wearable measurement validity between these groups, as suspected, we need to take them into account when analyzing consumer wearable measurement data and interpreting the results. If not, differences in the accuracy of consumer wearable-based risk prediction, disease screening, and monitoring tools between populations may be created, exacerbating already existing health disparities as the implementation of these tools progresses. Additionally, a reduced accuracy of these tools in older adults and people with obesity poses a public health risk, as these tools can be used to screen for conditions that are more prevalent in these population groups (eg, atrial fibrillation and sleep apnea) [219]. Increasing the representation of the “missing” population groups in consumer wearable validation studies is therefore vital. To increase the representation of these groups, a suitable benchmark for representation should be selected during the study design process, and the representativeness of the study population should be evaluated based on that benchmark throughout the entire study. Ideally, this evaluation should not only be conducted by the research team but also by funding agencies, journals, peer reviewers, and other stakeholders in the research process. Additionally, researchers should involve their study populations in the study design process to reduce barriers to participation.
Strengths and Limitations
This scoping review is the first to provide a systematic overview of the representation of sex, age, BMI, and skin tone groups in consumer wearable validation studies. Through an extensive search in 3 electronic bibliographic databases and the reference lists of 58 reviews, we were able to include 186 relevant papers. To ensure the transparency of our methods, the entire review process was guided by a protocol that is openly available on OSF. Nevertheless, this review has some limitations that need to be considered.
The representation of age, BMI, and skin tone groups was estimated based on the reported mean (SD), assuming a normal distribution. Though this approach allowed us to visualize the representation of these population groups, there is likely some overestimation or underestimation in these percentages. We hypothesize, based on our results, that the distribution of age, BMI, and skin tone is more likely to be skewed to the right in the included validation studies. An overestimation of the percentage of older adults, people with obesity, and people with darker skin tones in this review is therefore more likely. This suggests that the true representation of these population groups could be even lower than the percentages presented in this review. This is important to take into account when interpreting our findings, as is the fact that there is likely still variation in the representation within the categories applied in this review. Especially within the category of adults, which includes both young adults (18‐40 y) and middle-aged adults (41‐64 y), the high representation of young adults might mask the underrepresentation of middle-aged adults. Similarly, as no breakdown of the representation of the demographic groups per brand or model is reported in this review, it is important to note that there may be differences in the adequacy of representation between different devices, which are not reported in this review.
Another limitation of this review is the use of the Fitzpatrick scale to assess the representation of skin tone. The Fitzpatrick scale was included in this review because alternative measures are currently limited, and this scale is recommended by current reporting guidelines. However, it does not accurately depict the true representation of different skin tones with different melanin content, which truly impacts the PPG signal by absorbing and scattering the light [5]. Therefore, we recommend that future validation studies consider the use of melanometry, a method using reflected light that has been suggested as an objective quantitative measurement of skin melanin [220]. Future studies should also consider using arterial blood gas as a reference standard for measuring SpO2 instead of finger pulse oximeters. It is a critical limitation that many included studies used finger pulse oximeters, as these devices also exhibit pigment-related bias. Comparing wearables against them, rather than arterial blood gas, may obscure the true extent of skin tone inaccuracies. Additionally, it should be noted that BMI, like the Fitzpatrick scale, is a crude proxy for what it aims to measure, namely, adiposity. Although BMI is convenient for reporting and provides a general overview of the representation of different body sizes, future validation studies focusing on the mechanistic determinants of the validity of PPG-based measurements should consider measuring body composition.
Additionally, it is important to consider that this review only included studies that were published in English. This could have resulted in an underrepresentation of studies conducted in Asia, Africa, and South America, thereby reducing the representation of darker skin tones observed in this review. During the screening of the search results, only 3 studies were excluded based on the language criterion. This suggests that our search strategy, in general, was not very effective in retrieving non-English validation studies. If a similar review is performed in the future, we therefore recommend investigating methods to improve the retrieval of non-English studies, such as including local scientific literature databases.
Conclusions
With this scoping review, we aimed to identify underrepresented population groups by mapping the representation of sex, age, BMI, and skin tone groups in 186 consumer wearable PPG-based measurement validation studies. Based on the results, we conclude that older adults, people with underweight or obesity, and people with darker skin tones are generally strongly underrepresented in consumer wearable validation studies. Researchers making use of PPG-based consumer wearable measurements in their studies should, therefore, be aware of the limited knowledge of the measurement validity in these population groups and take this uncertainty into account when interpreting their findings. To avoid the creation of further health disparities, there should be more focus on the representativeness of consumer wearable validation study populations. More transparent reporting of study population characteristics and the use of benchmarks for representation will allow researchers and other stakeholders in the research process to evaluate the representativeness of a study population. To achieve study populations that properly represent the target population, representatives from the target population should be included in the study design process.
Supplementary material
Acknowledgments
In the process of this review, ChatGPT (free version) was used in the creation of the R script to provide suggestions for the code. These suggestions were then verified by testing the code and reading other sources. Additionally, Grammarly was used to improve the spelling and grammar of the written text.
Abbreviations
- ANSI
American National Standards Institute
- CTA
Consumer Technology Association
- HRV
heart rate variability
- PPG
photoplethysmography
- PRISMA
Preferred Reporting Items for Systematic Reviews and Meta-Analyses
- PRISMA-ScR
PRISMA extension for scoping review
- SpO2
oxygen saturation
Footnotes
Funding: This work was supported by the RECENTRE HTSF program, financed by the 4TU.Federation (grant 4TU-UIT-468).
Data Availability: After the publication of this manuscript, a CSV file with the extracted data and R scripts will be uploaded to the related project in OSF [17]. A CC-BY-4.0 license will apply.
Authors’ Contributions: Conceptualization: RMS, AAMB, EH
Data curation: RMS
Formal analysis: RMS
Funding acquisition: AAMB, AW, LW
Investigation: RMS, FOD
Methodology: RMS, AAMB, EH
Project administration: RMS
Resources: AAMB
Software: RMS
Supervision: AAMB, EH
Validation: FOD
Visualization: RMS
Writing – original draft: RMS, AAMB, EH
Writing – review & editing: RMS, FOD, AJ, MR, LW, EH, AW, AAMB
Conflicts of Interest: None declared.
References
- 1.Dickinson L. ALPHA—the Holy Grail of Heart Rate (Look Ma, no hands!) Kickstarter. 2012. [15-11-2024]. https://www.kickstarter.com/projects/alphaheartrate/alpha-the-holy-grail-of-heart-rate-look-ma-no-hand/description URL. Accessed.
- 2.Number of user of smartwatches world wide from 2017 to 2029 (in millions) Statista. 2025. [14-07-2026]. https://tinyurl.com/3srbkba4 URL. Accessed.
- 3.Huhn S, Axt M, Gunga HC, et al. The impact of wearable technologies in health research: scoping review. JMIR mHealth uHealth. 2022 Jan 25;10(1):e34384. doi: 10.2196/34384. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. 2007 Mar;28(3):R1–39. doi: 10.1088/0967-3334/28/3/R01. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 5.Fine J, Branan KL, Rodriguez AJ, et al. Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors (Basel) 2021 Apr 16;11(4):126. doi: 10.3390/bios11040126. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Bent B, Goldstein BA, Kibbe WA, Dunn JP. Investigating sources of inaccuracy in wearable optical heart rate sensors. NPJ Digit Med. 2020;3(1):18. doi: 10.1038/s41746-020-0226-6. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Boonya-Ananta T, Rodriguez AJ, Du Le VN, Ramella-Roman JC. Monte Carlo analysis of optical heart rate sensors in commercial wearables: the effect of skin tone and obesity on the photoplethysmography (PPG) signal. Biomed Opt Express. 2021 Dec 1;12(12):7445–7457. doi: 10.1364/BOE.439893. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Colvonen PJ, DeYoung PN, Bosompra NOA, Owens RL. Limiting racial disparities and bias for wearable devices in health science research. Sleep. 2020 Oct 13;43(10):zsaa159. doi: 10.1093/sleep/zsaa159. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Zinzuwadia A, Singh JP. Wearable devices-addressing bias and inequity. Lancet Digit Health. 2022 Dec;4(12):e856–e857. doi: 10.1016/S2589-7500(22)00194-7. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 10.Ibrahim H, Liu X, Zariffa N, Morris AD, Denniston AK. Health data poverty: an assailable barrier to equitable digital health care. Lancet Digit Health. 2021 Apr;3(4):e260–e265. doi: 10.1016/S2589-7500(20)30317-4. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 11.Doherty C, Baldwin M, Keogh A, Caulfield B, Argent R. Keeping pace with wearables: a living umbrella review of systematic reviews evaluating the accuracy of consumer wearable technologies in health measurement. Sports Med. 2024 Nov;54(11):2907–2926. doi: 10.1007/s40279-024-02077-2. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Fuller D, Colwell E, Low J, et al. Reliability and validity of commercially available wearable devices for measuring steps, energy expenditure, and heart rate: systematic review. JMIR mHealth uHealth. 2020 Sep 8;8(9):e18694. doi: 10.2196/18694. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Koerber D, Khan S, Shamsheri T, Kirubarajan A, Mehta S. Accuracy of heart rate measurement with wrist-worn wearable devices in various skin tones: a systematic review. J Racial Ethn Health Disparities. 2023 Dec;10(6):2676–2684. doi: 10.1007/s40615-022-01446-9. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Musa SB, Ellis R, Chafe B, et al. Wearable device validity in measuring steps, energy expenditure, and heart rate across age, gender, and body mass index: data analysis from a systematic review. J Phys Act Health. 2022 Feb 1;20(2):100–105. doi: 10.1123/jpah.2022-0160. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 15.Singh S, Bennett MR, Chen C, Shin S, Ghanbari H, Nelson BW. Impact of skin pigmentation on pulse oximetry blood oxygenation and wearable pulse rate accuracy: systematic review and meta-analysis. J Med Internet Res. 2024 Oct 10;26:e62769. doi: 10.2196/62769. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Arksey H, O’Malley L. Scoping studies: towards a methodological framework. Int J Soc Res Methodol. 2005 Feb;8(1):19–32. doi: 10.1080/1364557032000119616. doi. [DOI] [Google Scholar]
- 17.Schipper RM, Oumar Djibrillah F, Roosink M, et al. The representation of different study populations in studies assessing the validity of consumer wearables for measuring vital signs: a scoping review. OSF. 2024 Nov 27; doi: 10.17605/OSF.IO/48V7G. doi. [DOI] [PMC free article] [PubMed]
- 18.Tricco AC, Lillie E, Zarin W, et al. PRISMA extension for scoping reviews (PRISMA-ScR): checklist and explanation. Ann Intern Med. 2018 Oct 2;169(7):467–473. doi: 10.7326/M18-0850. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 19.Mühlen JM, Stang J, Lykke Skovgaard E, et al. Recommendations for determining the validity of consumer wearable heart rate devices: expert statement and checklist of the INTERLIVE network. Br J Sports Med. 2021 Jul;55(14):767–779. doi: 10.1136/bjsports-2020-103148. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Consumer Technology Association (CTA); 2023. [14-07-2026]. Physical activity monitoring for heart rate—real world analysis (ANSI/CTA-2065.1)https://www.cta.tech/standards/ansicta-20651/ URL. Accessed. [Google Scholar]
- 21.U.S. Food and Drug Administration; 2024. [14-07-2026]. FDA executive summary: performance evaluation of pulse oximeters taking into consideration skin pigmentation, race and ethnicity.https://www.fda.gov/media/175828/download URL. Accessed. [Google Scholar]
- 22.Age groups. National Library of Medicine. 1998. [20-10-2025]. https://www.ncbi.nlm.nih.gov/mesh/68009273 URL. Accessed.
- 23.Fitzpatrick TB. The validity and practicality of sun-reactive skin types I through VI. Arch Dermatol. 1988 Jun;124(6):869–871. doi: 10.1001/archderm.124.6.869. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 24.Obesity and overweight. World Health Organisation. 2025. [17-09-2025]. https://www.who.int/news-room/fact-sheets/detail/obesity-and-overweight URL. Accessed.
- 25.Konrat C, Boutron I, Trinquart L, Auleley GR, Ricordeau P, Ravaud P. Underrepresentation of elderly people in randomised controlled trials. The example of trials of 4 widely prescribed drugs. PLoS One. 2012;7(3):e33559. doi: 10.1371/journal.pone.0033559. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Agboola F, Wright AC. A framework for evaluating the diversity of clinical trials. J Clin Epidemiol. 2024 May;169:111299. doi: 10.1016/j.jclinepi.2024.111299. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 27.Antognoli L, Panni L, Cosoli G, Scalise L. Metrological characterization of commercial smartwatches: are these sensors suitable for the assessment of well-being?. 2023 IEEE International Workshop on Metrology for Living Environment (MetroLivEnv); May 29-31, 2023; Milano, Italy. Presented at. doi. [DOI] [Google Scholar]
- 28.Falter M, Budts W, Goetschalckx K, Cornelissen V, Buys R. Accuracy of Apple Watch measurements for heart rate and energy expenditure in patients with cardiovascular disease: cross-sectional study. JMIR mHealth uHealth. 2019 Mar 19;7(3):e11889. doi: 10.2196/11889. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Lins LFTDS, do Nascimento EGC, da Silva Júnior JA, de Medeiros Fernandes TAA, de Andrade MF, de Mesquita Andrade C. Accuracy of wearable electronic device compared to manual and automatic methods of blood pressure determination. Med Biol Eng Comput. 2023 Oct;61(10):2627–2636. doi: 10.1007/s11517-023-02869-0. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 30.Poli A, Cosoli G, Iadarola G, Spinsante S, Scalise L. Feasibility of blood pressure measurement through wearable devices: analysis of smartwatches performance. 2022 IEEE International Symposium on Medical Measurements and Applications (MeMeA); Jun 22-24, 2022; Messina, Italy. Presented at. doi. [DOI] [Google Scholar]
- 31.Yi L, Lv ZH, Hu SY, et al. Validating the accuracy of a multifunctional smartwatch sphygmomanometer to monitor blood pressure. J Geriatr Cardiol. 2022 Nov 28;19(11):843–852. doi: 10.11909/j.issn.1671-5411.2022.11.004. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Hahnen C, Freeman CG, Haldar N, et al. Accuracy of vital signs measurements by a smartwatch and a portable health device: validation study. JMIR mHealth uHealth. 2020 Feb 12;8(2):e16811. doi: 10.2196/16811. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Monteiro R, Rabello GCM, Moreno CR, et al. Enhancing cardiac postoperative care: a smartwatch-integrated remote telemonitoring platform for health screening with ECG analysis. Front Cardiovasc Med. 2024;11:1443998. doi: 10.3389/fcvm.2024.1443998. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Budig M, Stoohs R, Keiner M. Validity of two consumer multisport activity tracker and one accelerometer against polysomnography for measuring sleep parameters and vital data in a laboratory setting in sleep patients. Sensors (Basel) 2022 Dec 6;22(23):9540. doi: 10.3390/s22239540. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Cosoli G, Antognoli L, Scalise L. Indirect estimation of breathing rate through wearable devices. 2022 IEEE International Symposium on Medical Measurements and Applications (MeMeA); Jun 22-24, 2022; Messina, Italy. Presented at. doi. [DOI] [Google Scholar]
- 36.Helmer P, Hottenrott S, Wienböker K, et al. Reliability of continuous vital sign monitoring in post-operative patients employing consumer-grade fitness trackers: a randomised pilot trial. Digit Health. 2024;10:20552076241254026. doi: 10.1177/20552076241254026. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Jung HC, Kang M, Lee NH, Jeon S, Lee S. Impact of placement of Fitbit HR under laboratory and free-living conditions. Sustainability. 2020;12(16):6306. doi: 10.3390/su12166306. doi. [DOI] [Google Scholar]
- 38.Hermans F, Arents E, Blondeel A, et al. Validity of a consumer-based wearable to measure clinical parameters in patients with chronic obstructive pulmonary disease and healthy controls: observational study. JMIR mHealth uHealth. 2024 Nov 6;12:e56027. doi: 10.2196/56027. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Bai Y, Hibbing P, Mantis C, Welk GJ. Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR. J Sports Sci. 2018 Aug;36(15):1734–1741. doi: 10.1080/02640414.2017.1412235. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 40.Cao R, Azimi I, Sarhaddi F, et al. Accuracy assessment of Oura Ring nocturnal heart rate and heart rate variability in comparison with electrocardiography in time and frequency domains: comprehensive analysis. J Med Internet Res. 2022 Jan 18;24(1):e27487. doi: 10.2196/27487. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Miller DJ, Sargent C, Roach GD. A validation of six wearable devices for estimating sleep, heart rate and heart rate variability in healthy adults. Sensors (Basel) 2022 Aug 22;22(16):6317. doi: 10.3390/s22166317. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Navalta JW, Montes J, Bodell NG, Salatto RW, Manning JW, DeBeliso M. Concurrent heart rate validity of wearable technology devices during trail running. PLoS One. 2020;15(8):e0238569. doi: 10.1371/journal.pone.0238569. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Stone JD, Ulman HK, Tran K, et al. Assessing the accuracy of popular commercial technologies that measure resting heart rate and heart rate variability. Front Sports Act Living. 2021;3:585870. doi: 10.3389/fspor.2021.585870. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Abdalazim N, Larraza JAA, Alchieri L, Alecci L, Santini S, Gashi S. In: Pervasive Computing Technologies for Healthcare—16th EAI International Conference, PervasiveHealth 2022, Thessaloniki, Greece, December 12-14, 2022, Proceedings. Tsanas A, Triantafyllidis A, editors. Springer Nature; 2023. Heart rate during sleep measured using finger-, wrist- and chest-worn devices: a comparison study; pp. 18–32. doi. [DOI] [Google Scholar]
- 45.Boudreaux BD, Hebert EP, Hollander DB, et al. Validity of wearable activity monitors during cycling and resistance exercise. Med Sci Sports Exerc. 2018 Mar;50(3):624–633. doi: 10.1249/MSS.0000000000001471. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 46.Passler S, Müller N, Senner V. In-ear pulse rate measurement: a valid alternative to heart rate derived from electrocardiography? Sensors (Basel) 2019 Aug 21;19(17):3641. doi: 10.3390/s19173641. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Reece JD, Bunn JA, Choi M, Navalta JW. Assessing heart rate using Consumer Technology Association standards. Technologies (Basel) 2021;9(3):46. doi: 10.3390/technologies9030046. doi. [DOI] [Google Scholar]
- 48.Rozanski GM, Aqui A, Sivakumaran S, Mansfield A. Consumer wearable devices for activity monitoring among individuals after a stroke: a prospective comparison. JMIR Cardio. 2018 Jan 4;2(1):e1. doi: 10.2196/cardio.8199. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Browne SH, Vaida F, Umlauf A, Kim J, DeYoung P, Owens RL. Performance of a commercial smart watch compared to polysomnography reference for overnight continuous oximetry measurement and sleep apnea evaluation. J Clin Sleep Med. 2024 Sep 1;20(9):1479–1488. doi: 10.5664/jcsm.11178. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Bai Y, Tompkins C, Gell N, Dione D, Zhang T, Byun W. Comprehensive comparison of Apple Watch and Fitbit monitors in a free-living setting. PLoS One. 2021;16(5):e0251975. doi: 10.1371/journal.pone.0251975. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Gillinov S, Etiwy M, Wang R, et al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med Sci Sports Exerc. 2017 Aug;49(8):1697–1703. doi: 10.1249/MSS.0000000000001284. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 52.Haghayegh S, Khoshnevis S, Smolensky MH, Diller KR. Accuracy of PurePulse photoplethysmography technology of Fitbit Charge 2 for assessment of heart rate during sleep. Chronobiol Int. 2019 Jul;36(7):927–933. doi: 10.1080/07420528.2019.1596947. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 53.Jiang Y, Spies C, Magin J, Bhosai SJ, Snyder L, Dunn J. Investigating the accuracy of blood oxygen saturation measurements in common consumer smartwatches. PLoS Digit Health. 2023 Jul;2(7):e0000296. doi: 10.1371/journal.pdig.0000296. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Pasadyn SR, Soudan M, Gillinov M, et al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther. 2019 Aug;9(4):379–385. doi: 10.21037/cdt.2019.06.05. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Shcherbina A, Mattsson CM, Waggott D, et al. Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort. J Pers Med. 2017 May 24;7(2):2017–5. doi: 10.3390/jpm7020003. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Stahl SE, An HS, Dinkel DM, Noble JM, Lee JM. How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc Med. 2016;2(1):e000106. doi: 10.1136/bmjsem-2015-000106. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Cadmus-Bertram L, Gangnon R, Wirkus EJ, Thraen-Borowski KM, Gorzelitz-Liebhauser J. The accuracy of heart rate monitoring by some wrist-worn activity trackers. Ann Intern Med. 2017 Apr 18;166(8):610–612. doi: 10.7326/L16-0353. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Dooley EE, Golaszewski NM, Bartholomew JB. Estimating accuracy at exercise intensities: a comparative study of self-monitoring heart rate and physical activity wearable devices. JMIR mHealth uHealth. 2017 Mar 16;5(3):e34. doi: 10.2196/mhealth.7043. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.El-Amrawy F, Nounou MI. Are currently available wearable devices for activity tracking and heart rate monitoring accurate, precise, and medically beneficial? Healthc Inform Res. 2015 Oct;21(4):315–320. doi: 10.4258/hir.2015.21.4.315. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Nash D, Shah MJ, Shehab O, et al. “But for the blind spot”: accuracy and diagnostic performance of smart watch cardiac features in pediatric patients. Heart Rhythm. 2024 May;21(5):581–589. doi: 10.1016/j.hrthm.2024.01.021. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 61.Wang R, Blackburn G, Desai M, et al. Accuracy of wrist-worn heart rate monitors. JAMA Cardiol. 2017 Jan 1;2(1):104–106. doi: 10.1001/jamacardio.2016.3340. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 62.Yu H, Kotlyar M, Dufresne S, Thuras P, Pakhomov S. Feasibility of using an armband optical heart rate sensor in naturalistic environment. Pac Symp Biocomput. 2022:43–54. doi: 10.1142/9789811270611_0005. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Dondzila CJ, Lewis C, Lopez JR, Parker T. Congruent accuracy of wrist-worn activity trackers during controlled and free-living conditions. Int J Exerc Sci. 2018;11(7):575–584. doi: 10.70252/UBBT6886. doi. [DOI] [Google Scholar]
- 64.Saghir N, Aggarwal A, Soneji N, Valencia V, Rodgers G, Kurian T. A comparison of manual electrocardiographic interval and waveform analysis in lead 1 of 12-lead ECG and Apple Watch ECG: a validation study. Cardiovasc Digit Health J. 2020;1(1):30–36. doi: 10.1016/j.cvdhj.2020.07.002. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Seshadri DR, Bittel B, Browsky D, et al. Accuracy of the Apple Watch 4 to measure heart rate in patients with atrial fibrillation. IEEE J Transl Eng Health Med. 2020;8:2700204. doi: 10.1109/JTEHM.2019.2950397. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Spierer DK, Rosen Z, Litman LL, Fujii K. Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol. 2015;39(5):264–271. doi: 10.3109/03091902.2015.1047536. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 67.Thiebaud RS, Funk MD, Patton JC, et al. Validity of wrist-worn consumer products to measure heart rate and energy expenditure. Digit Health. 2018;4:2055207618770322. doi: 10.1177/2055207618770322. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Thomson EA, Nuss K, Comstock A, et al. Heart rate measures from the Apple Watch, Fitbit Charge HR 2, and electrocardiogram across different exercise intensities. J Sports Sci. 2019 Jun;37(12):1411–1419. doi: 10.1080/02640414.2018.1560644. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 69.Turki A, Behbehani K, Ding K, Zhang R, Li M, Bell K, et al. Estimation of heart rate variability measures using Apple Watch and evaluating their accuracy: estimation of heart rate variability measures using Apple Watch. PETRA ’21:The 14th PErvasive Technologies Related to Assistive Environments Conference; Jun 29 to Jul 2, 2021; Corfu, Greece. Presented at. doi. [DOI] [Google Scholar]
- 70.Turki AF, Jani MB, Ding K, Zhang R, Behbehani K. An investigation of heartrate sensing accuracy by wrist-worn fitness tracking devices. Annu Int Conf IEEE Eng Med Biol Soc. 2019 Jul;2019:3337–3340. doi: 10.1109/EMBC.2019.8856375. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 71.Lauterbach CJ, Romano PA, Greisler LA, Brindle RA, Ford KR, Kuennen MR. Accuracy and reliability of commercial wrist-worn pulse oximeter during normobaric hypoxia exposure under resting conditions. Res Q Exerc Sport. 2021 Sep;92(3):549–558. doi: 10.1080/02701367.2020.1759768. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 72.Merrigan JJ, Stovall JH, Stone JD, Stephenson M, Finomore VS, Hagen JA. Validation of Garmin and Polar devices for continuous heart rate monitoring during common training movements in tactical populations. Meas Phys Educ Exerc Sci. 2023 Jul 3;27(3):234–247. doi: 10.1080/1091367X.2022.2161820. doi. [DOI] [Google Scholar]
- 73.Montalvo S, Martinez A, Arias S, et al. Commercial smart watches and heart rate monitors: a concurrent validity analysis. J Strength Cond Res. 2023 Sep 1;37(9):1802–1808. doi: 10.1519/JSC.0000000000004482. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 74.Montes J, Tandy R, Young J, Lee SP, Navalta J. A comparison of multiple wearable technology devices heart rate and step count measurements during free motion and treadmill based measurements. Int J Kinesiol Sports Sci. 2019;7(2):30–39. doi: 10.7575/aiac.ijkss.v.7n.2p.30. doi. [DOI] [Google Scholar]
- 75.Montoye AHK, Mitrzyk JR, Molesky MJ. Comparative accuracy of a wrist-worn activity tracker and a smart shirt for physical activity assessment. Meas Phys Educ Exerc Sci. 2017 Oct 2;21(4):201–211. doi: 10.1080/1091367X.2017.1331166. doi. [DOI] [Google Scholar]
- 76.Navalta JW, Davis DW, Carrier B, et al. Validity and reliability of wearable devices during self-paced walking, jogging and overground skipping. Sport Mont. 2023 Oct 1;21(3):23–29. doi: 10.26773/smj.231004. doi. [DOI] [Google Scholar]
- 77.Nuss KJ, Sanford JL, Archambault LJ, et al. Accuracy of heart rate and energy expenditure estimations of wrist-worn and arm-worn Apple watches. J Meas Phys Behav. 2019;2(3):166–175. doi: 10.1123/jmpb.2018-0061. doi. [DOI] [Google Scholar]
- 78.Pope ZC, Lee JE, Zeng N, Gao Z. Validation of four smartwatches in energy expenditure and heart rate assessment during exergaming. Games Health J. 2019 Jun;8(3):205–212. doi: 10.1089/g4h.2018.0087. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 79.Powierza CS, Clark MD, Hughes JM, Carneiro KA, Mihalik JP. Validation of a self-monitoring tool for use in exercise therapy. PM R. 2017 Nov;9(11):1077–1084. doi: 10.1016/j.pmrj.2017.03.012. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 80.de Zambotti M, Baker FC, Willoughby AR, et al. Measures of sleep and cardiac functioning during sleep using a multi-sensory commercially-available wristband in adolescents. Physiol Behav. 2016 May;158:143–149. doi: 10.1016/j.physbeh.2016.03.006. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 81.Etiwy M, Akhrass Z, Gillinov L, et al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther. 2019 Jun;9(3):262–271. doi: 10.21037/cdt.2019.04.08. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.Huynh P, Shan R, Osuji N, et al. Heart rate measurements in patients with obstructive sleep apnea and atrial fibrillation: prospective pilot study assessing Apple Watch’s agreement with telemetry data. JMIR Cardio. 2021 Feb 8;5(1):e18050. doi: 10.2196/18050. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 83.Reddy RK, Pooni R, Zaharieva DP, et al. Accuracy of wrist-worn activity monitors during common daily physical activities and types of structured exercise: evaluation study. JMIR mHealth uHealth. 2018 Dec 10;6(12):e10338. doi: 10.2196/10338. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Jo E, Lewis K, Directo D, Kim MJ, Dolezal BA. Validation of biofeedback wearables for photoplethysmographic heart rate tracking. J Sports Sci Med. 2016 Sep;15(3):540–547. Medline. [PMC free article] [PubMed] [Google Scholar]
- 85.Brazendale K, Decker L, Hunt ET, et al. Validity and wearability of consumer-based fitness trackers in free-living children. Int J Exerc Sci. 2019;12(5):471–482. doi: 10.70252/BBTU4889. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.AbiMansour JP, Kaur J, Velaga S, Vatsavayi P, Vogt M, Chandrasekhara V. Accuracy and role of consumer facing wearable technology for continuous monitoring during endoscopic procedures. Front Digit Health. 2024;6:1422929. doi: 10.3389/fdgth.2024.1422929. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 87.Bonneval L, Wing D, Sharp S, et al. Validity of heart rate variability measured with Apple Watch Series 6 compared to laboratory measures. Sensors (Basel) 2025 Apr 9;25(8):2380. doi: 10.3390/s25082380. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 88.Carrier B, Marten Chaves S, Navalta JW. Validation of aerobic capacity (VO2max) and pulse oximetry in wearable technology. Sensors (Basel) 2025 Jan 6;25(1):275. doi: 10.3390/s25010275. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Chase SE, Liddell RG, McGonagle CL, Ives SJ. Accuracy of the Apple Watch Series 9 for measures of energy expenditure and heart rate at rest and during exercise: impact of skin pigmentation. J Funct Morphol Kinesiol. 2024 Dec 17;9(4):275. doi: 10.3390/jfmk9040275. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Icenhower A, Murphy C, Brooks AK, et al. Investigating the accuracy of Garmin PPG sensors on differing skin types based on the Fitzpatrick scale: cross-sectional comparison study. Front Digit Health. 2025;7:1553565. doi: 10.3389/fdgth.2025.1553565. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 91.Navalta JW, Carrier B, Blank M, et al. Validity and reliability of wearable technology devices during simulated pickleball game play. Sports (Basel) 2024 Aug 28;12(9):234. doi: 10.3390/sports12090234. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 92.Page LL, Fanning J, Phipps C, Berger A, Reed E, Ehlers D. Heart rate monitoring among breast cancer survivors: quantitative study of device agreement in a community-based exercise program. JMIR Cancer. 2024 Jun 20;10:e51210. doi: 10.2196/51210. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Pierick AR, Burke KJ, Prusi M, et al. Validity of wrist-worn activity tracker heart rate detection in Fontan patients during exercise. Med Sci Sports Exerc. 2025 Feb 1;57(2):280–284. doi: 10.1249/MSS.0000000000003567. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 94.Yasuda M, Silverman J, Best AM. Can caregivers reliably assess their child’s heart rate and respiration rate using smartphone or smartwatch applications? Pediatr Emerg Care. 2025 Aug 1;41(8):e67–e72. doi: 10.1097/PEC.0000000000003400. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 95.Benedetti D, Olcese U, Frumento P, et al. A validation study versus portable polysomnography. J Sleep Res. 2021;30(6):e13346. doi: 10.1111/jsr.13346. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Benedetto S, Caldato C, Bazzan E, Greenwood DC, Pensabene V, Actis P. Assessment of the Fitbit Charge 2 for monitoring heart rate. PLoS One. 2018;13(2):e0192691. doi: 10.1371/journal.pone.0192691. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.Cosoli G, Antognoli L, Veroli V, Scalise L. Accuracy and precision of wearable devices for real-time monitoring of swimming athletes. Sensors (Basel) 2022 Jun 23;22(13):4726. doi: 10.3390/s22134726. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Pelizzo G, Guddo A, Puglisi A, et al. Accuracy of a wrist-worn heart rate sensing device during elective pediatric surgical procedures. Children (Basel) 2018 Mar 8;5(3):38. doi: 10.3390/children5030038. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Rho G, Di Rienzo F, Marinai C, et al. In: MEDICON’23 and CMBEBIH’23—Proceedings of the Mediterranean Conference on Medical and Biological Engineering and Computing (MEDICON) and International Conference on Medical and Biological Engineering (CMBEBIH), September 14–16, 2023, Sarajevo, Bosnia and Herzegovina—Volume 1: Imaging, Engineering and Artificial Intelligence in Healthcare. Badnjević A, Gurbeta Pokvić L, editors. Springer; 2024. Preliminary assessment of the Samsung Galaxy Watch 5 accuracy for the monitoring of heart rate and heart rate variability parameters; pp. 22–30. doi. [DOI] [Google Scholar]
- 100.Schiefer LM, Treff G, Treff F, et al. Validity of peripheral oxygen saturation measurements with the Garmin Fēnix 5X plus wearable device at 4559 m. Sensors (Basel) 2021 Sep 23;21(19):6363. doi: 10.3390/s21196363. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Spaccarotella C, Polimeni A, Mancuso C, Pelaia G, Esposito G, Indolfi C. Assessment of non-invasive measurements of oxygen saturation and heart rate with an Apple smartwatch: comparison with a standard pulse oximeter. J Clin Med. 2022 Mar 8;11(6):1467. doi: 10.3390/jcm11061467. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 102.Spinsante S, Porfiri S, Scalise L. Accuracy of heart rate measurements by a smartwatch in low intensity activities. 2019 IEEE International Symposium on Medical Measurements and Applications (MeMeA); Jun 26-28, 2019; Istanbul, Turkey. Presented at. doi. [DOI] [Google Scholar]
- 103.Cosoli G, Antognoli L, Scalise L. Wearable electrocardiography for physical activity monitoring: definition of validation protocol and automatic classification. Biosensors (Basel) 2023 Jan 18;13(2):154. doi: 10.3390/bios13020154. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 104.Coelli S, Carrara M, Bianchi AM, De Tommaso M, Actis-Grosso R, Reali P. Comparing consumer and research-grade wristbands for inter-beat intervals monitoring. 2024 IEEE International Conference on Metrology for eXtended Reality, Artificial Intelligence and Neural Engineering (MetroXRAINE); Oct 21-23, 2024; St Albans, United Kingdom. Presented at. doi. [DOI] [Google Scholar]
- 105.Budig M, Keiner M, Stoohs R, Hoffmeister M, Höltke V. Heart rate and distance measurement of two multisport activity trackers and a cellphone app in different sports: a cross-sectional validation and comparison field study. Sensors (Basel) 2021 Dec 28;22(1):180. doi: 10.3390/s22010180. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Düking P, Giessing L, Frenkel MO, Koehler K, Holmberg HC, Sperlich B. Wrist-worn wearables for monitoring heart rate and energy expenditure while sitting or performing light-to-vigorous physical activity: validation study. JMIR mHealth uHealth. 2020 May 6;8(5):e16716. doi: 10.2196/16716. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 107.Hänsel K, Poguntke R, Haddadi H, Alomainy A, Schmidt A. What to put on the user: sensing technologies for studies and physiology aware systems. CHI '18: Proceedings of the 2018 CHI Conference on Human Factors in Computing Systems; Apr 21-26, 2018; Montreal, QC, Canada. Presented at. doi. [DOI] [Google Scholar]
- 108.Helmer P, Hottenrott S, Rodemers P, et al. Accuracy and systematic biases of heart rate measurements by consumer-grade fitness trackers in postoperative patients: prospective clinical trial. J Med Internet Res. 2022 Dec 30;24(12):e42359. doi: 10.2196/42359. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Nissen M, Slim S, Jäger K, et al. Heart rate measurement accuracy of Fitbit Charge 4 and Samsung Galaxy Watch Active2: device evaluation study. JMIR Form Res. 2022 Mar 1;6(3):e33635. doi: 10.2196/33635. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Pätz C, Michaelis A, Markel F, et al. Accuracy of the Apple Watch oxygen saturation measurement in adults and children with congenital heart disease. Pediatr Cardiol. 2023 Feb;44(2):333–343. doi: 10.1007/s00246-022-02987-w. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 111.Mink F, Lutz W, Rafaeli E, Hehlmann M, Schwartz B, Uhl J. Validation and application of wrist-worn wearable physiological measures in psychotherapy. Cogn Ther Res. 2025 Oct;49(5):902–919. doi: 10.1007/s10608-025-10601-5. doi. [DOI] [Google Scholar]
- 112.Weis A, Leroy M, Jux C, Rupp S, Backhoff D. Oxygen saturation measurement in cyanotic heart disease with the Apple Watch. Cardiol Young. 2024 Oct 8;34(9):1–3. doi: 10.1017/S1047951124025216. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 113.Wolf S, Seidel P, Ockenga TA, Schoder D. Proceedings of the 57th Hawaii International Conference on System Sciences. IEEE Computer Society; 2024. Heart-to-wear: assessing the accuracy of heart rate sensor measurements of wearable devices in uncontrolled environments; pp. 3183–3191. doi. [DOI] [Google Scholar]
- 114.Al-Kaisey AM, Koshy AN, Ha FJ, et al. Accuracy of wrist-worn heart rate monitors for rate control assessment in atrial fibrillation. Int J Cardiol. 2020 Feb 1;300:161–164. doi: 10.1016/j.ijcard.2019.11.120. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 115.Climstein M, Alder JL, Brooker AM, et al. Reliability of the Polar Vantage M sports watch when measuring heart rate at different treadmill exercise intensities. Sports (Basel) 2020 Aug 23;8(9):117. doi: 10.3390/sports8090117. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 116.Gaynor M, Sawyer A, Jenkins S, Wood J. Variable agreement between wearable heart rate monitors during exercise in cystic fibrosis. ERJ Open Res. 2019 Oct;5(4):00006-2019. doi: 10.1183/23120541.00006-2019. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Ge Z, Prasad PWC, Costadopoulos N, Alsadoon A, Singh AK, Elchouemi A. Evaluating the accuracy of wearable heart rate monitors. 2016 2nd International Conference on Advances in Computing, Communication, & Automation (ICACCA) (Fall); Sep 30 to Oct 1, 2016; Bareilly, India. Presented at. doi. [DOI] [Google Scholar]
- 118.Hettiarachchi IT, Hanoun S, Nahavandi D, Nahavandi S. Validation of Polar OH1 optical heart rate sensor for moderate and high intensity physical activities. PLOS One. 2019;14(5):e0217288. doi: 10.1371/journal.pone.0217288. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 119.Koshy AN, Sajeev JK, Nerlekar N, et al. Smart watches for heart rate assessment in atrial arrhythmias. Int J Cardiol. 2018 Sep 1;266:124–127. doi: 10.1016/j.ijcard.2018.02.073. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 120.Shumate T, Link M, Furness J, Kemp-Smith K, Simas V, Climstein M. Validity of the Polar Vantage M watch when measuring heart rate at different exercise intensities. PeerJ. 2021;9:e10893. doi: 10.7717/peerj.10893. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 121.Uphill A, Kendall KL, Fogarty A, et al. Validity of Apple Watch, Garmin Forerunner 935 and GENEActiv for estimating energy expenditure during close quarter battle training in Special Forces soldiers. Eur J Sport Sci. 2024 May;24(5):614–622. doi: 10.1002/ejsc.12080. doi. [DOI] [Google Scholar]
- 122.Wallen MP, Gomersall SR, Keating SE, Wisløff U, Coombes JS. Accuracy of heart rate watches: implications for weight management. PLoS One. 2016;11(5):e0154420. doi: 10.1371/journal.pone.0154420. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Rajakariar K, Buntine P, Ghaly A, et al. Accuracy of smartwatch pulse oximetry measurements in hospitalized patients with coronavirus disease 2019. Mayo Clin Proc Digit Health. 2024 Mar;2(1):152–158. doi: 10.1016/j.mcpdig.2024.02.001. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 124.Gagnon J, Khau M, Lavoie-Hudon L, Vachon F, Drapeau V, Tremblay S. Comparing a Fitbit wearable to an electrocardiogram gold standard as a measure of heart rate under psychological stress: a validation study. JMIR Form Res. 2022 Dec 21;6(12):e37885. doi: 10.2196/37885. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 125.Kroll RR, Boyd JG, Maslove DM. Accuracy of a wrist-worn wearable device for monitoring heart rates in hospital inpatients: a prospective observational study. J Med Internet Res. 2016 Sep 20;18(9):e253. doi: 10.2196/jmir.6025. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 126.Lee C, Lee C, Fernando C, Chow CM. Comparison of Apple Watch vs KardiaMobile: a tale of two devices. CJC Open. 2022 Nov;4(11):939–945. doi: 10.1016/j.cjco.2022.07.011. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 127.Littell L, Roelle L, Dalal A, et al. Assessment of Apple Watch Series 6 pulse oximetry and electrocardiograms in a pediatric population. PLoS Digit Health. 2022 Aug;1(8):e0000051. doi: 10.1371/journal.pdig.0000051. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Moayedi Y, Abdulmajeed R, Duero Posada J, et al. Assessing the use of wrist-worn devices in patients with heart failure: feasibility study. JMIR Cardio. 2017 Dec 19;1(2):e8. doi: 10.2196/cardio.8301. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 129.Ray I, Liaqat D, Gabel M, de Lara E. Skin tone, confidence, and data quality of heart rate sensing in WearOS smartwatches. 2021 IEEE International Conference on Pervasive Computing and Communications Workshops and other Affiliated Events (PerCom Workshops); Mar 22-26, 2021; Kassel, Germany. Presented at. doi. [DOI] [Google Scholar]
- 130.Sequeira N, D’Souza D, Angaran P, Aves T, Dorian P. Common wearable devices demonstrate variable accuracy in measuring heart rate during supraventricular tachycardia. Heart Rhythm. 2020 May;17(5 Pt B):854–859. doi: 10.1016/j.hrthm.2020.02.018. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 131.Horton JF, Stergiou PRO, Fung TS, Katz L. Comparison of Polar M600 optical heart rate and ECG heart rate during exercise. Med Sci Sports Exerc. 2017 Dec;49(12):2600–2607. doi: 10.1249/MSS.0000000000001388. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 132.Hung SH, Serwa K, Rosenthal G, Eng JJ. Validity of heart rate measurements in wrist-based monitors across skin tones during exercise. PLoS One. 2025;20(2):e0318724. doi: 10.1371/journal.pone.0318724. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 133.Held NJ, Perrotta AS, Mueller T, Pfoh-MacDonald SJ. Agreement of the Apple Watch and Fitbit Charge for recording step count and heart rate when exercising in water. Med Biol Eng Comput. 2022 May;60(5):1323–1331. doi: 10.1007/s11517-022-02536-w. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 134.Ahn J, Ra HK, Yoon HJ, Son SH, Ko J. On-device filter design for self-identifying inaccurate heart rate readings on wrist-worn PPG sensors. IEEE Access. 2020;8:184774–184784. doi: 10.1109/ACCESS.2020.3025776. doi. [DOI] [Google Scholar]
- 135.Baek S, Ha Y, Park HW. Accuracy of wearable devices for measuring heart rate during conventional and Nordic walking. PM R. 2021 Apr;13(4):379–386. doi: 10.1002/pmrj.12424. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 136.Choi JH, Kim HJ. Accuracy verification of heart rate and energy consumption tracking devices to develop forest-based customized health care service programs. J People Plants Environ. 2019 Apr;22(2):219–229. doi: 10.11628/ksppe.2019.22.2.219. doi. [DOI] [Google Scholar]
- 137.Hwang J, Kim J, Choi KJ, Cho MS, Nam GB, Kim YH. Assessing accuracy of wrist-worn wearable devices in measurement of paroxysmal supraventricular tachycardia heart rate. Korean Circ J. 2019 May;49(5):437–445. doi: 10.4070/kcj.2018.0323. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Jung H, Kim D, Choi J, Joo EY. Validating a consumer smartwatch for nocturnal respiratory rate measurements in sleep monitoring. Sensors (Basel) 2023 Sep 19;23(18):7976. doi: 10.3390/s23187976. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Jung H, Kim D, Lee W, et al. Performance evaluation of a wrist-worn reflectance pulse oximeter during sleep. Sleep Health. 2022 Oct;8(5):420–428. doi: 10.1016/j.sleh.2022.04.003. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 140.Kang S, Kim Y, Byun W, Suk J, Lee JM. Comparison of a wearable tracker with actigraph for classifying physical activity intensity and heart rate in children. Int J Environ Res Public Health. 2019 Jul 25;16(15):2663. doi: 10.3390/ijerph16152663. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 141.Kim C, Song JH, Kim SH. Validation of wearable digital devices for heart rate measurement during exercise test in patients with coronary artery disease. Ann Rehabil Med. 2023 Aug;47(4):261–271. doi: 10.5535/arm.23019. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 142.Abt G, Bray J, Benson AC. The validity and inter-device variability of the Apple Watch for measuring maximal heart rate. J Sports Sci. 2018 Jul;36(13):1447–1452. doi: 10.1080/02640414.2017.1397282. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 143.Collins T, Woolley SI, Oniani S, et al. Version reporting and assessment approaches for new and updated activity and heart rate monitors. Sensors (Basel) 2019 Apr 10;19(7):1705. doi: 10.3390/s19071705. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 144.Hough P, Glaister M, Pledger A. The accuracy of wrist-worn heart rate monitors across a range of exercise intensities. J Phys Act Res. 2017 Jan 21;2(2):112–116. doi: 10.12691/jpar-2-2-8. doi. [DOI] [Google Scholar]
- 145.Jarchi D, Salvi D, Velardo C, Mahdi A, Tarassenko L, Clifton DA. Estimation of HRV and SpO2 from wrist-worn commercial sensors for clinical settings. 2018 IEEE 15th International Conference on Wearable and Implantable Body Sensor Networks (BSN); Mar 4-7, 2018; Las Vegas, NV, USA. Presented at. doi. [DOI] [Google Scholar]
- 146.Khushhal A, Nichols S, Evans W, et al. Validity and reliability of the Apple Watch for measuring heart rate during exercise. Sports Med Int Open. 2017 Oct;1(6):E206–E211. doi: 10.1055/s-0043-120195. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 147.Muggeridge DJ, Hickson K, Davies AV, et al. Measurement of heart rate using the Polar OH1 and Fitbit Charge 3 wearable devices in healthy adults during light, moderate, vigorous, and sprint-based exercise: validation study. JMIR mHealth uHealth. 2021 Mar 25;9(3):e25313. doi: 10.2196/25313. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.O’Driscoll R, Turicchi J, Hopkins M, et al. The validity of two widely used commercial and research-grade activity monitors, during resting, household and activity behaviours. Health Technol. 2020 May;10(3):637–648. doi: 10.1007/s12553-019-00392-7. doi. [DOI] [Google Scholar]
- 149.Williams K, Jamieson A, Chaturvedi N, Hughes A, Orini M. Validation of wearable derived heart rate variability and oxygen saturation from the Garmin’s health snapshot. 50th Computing in Cardiology Conference (Computing in Cardiology 2023, CinC 2023); Sep 24-27, 2023; Atlanta, Georgia, USA. [14-07-2026]. Presented at. URL. Accessed. [Google Scholar]
- 150.Jamieson A, Jones S, Chaturvedi N, Hughes AD, Orini M. Accuracy of smartwatches for the remote assessment of exercise capacity. Sci Rep. 2024 Oct 3;14(1):22994. doi: 10.1038/s41598-024-74140-x. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 151.Alfonso C, Garcia-Gonzalez MA, Parrado E, Gil-Rojas J, Ramos-Castro J, Capdevila L. Agreement between two photoplethysmography-based wearable devices for monitoring heart rate during different physical activity situations: a new analysis methodology. Sci Rep. 2022 Sep 14;12(1):15448. doi: 10.1038/s41598-022-18356-9. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 152.Hernando D, Roca S, Sancho J, Alesanco Á, Bailón R. Validation of the Apple Watch for heart rate variability measurements during relax and mental stress in healthy subjects. Sensors (Basel) 2018 Aug 10;18(8):2619. doi: 10.3390/s18082619. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 153.Martín-Escudero P, Cabanas AM, Dotor-Castilla ML, et al. Are activity wrist-worn devices accurate for determining heart rate during intense exercise? Bioengineering (Basel) 2023 Feb 15;10(2):254. doi: 10.3390/bioengineering10020254. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 154.Meza C, Juega J, Francisco J, et al. Accuracy of a smartwatch to assess heart rate monitoring and atrial fibrillation in stroke patients. Sensors (Basel) 2023 May 10;23(10):4632. doi: 10.3390/s23104632. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 155.Ruiz-Malagón EJ, Castro-Infantes S, Ritaco-Real M, Soto-Hermoso VM. Concurrent validity of the Polar Precision Prime photoplethysmographic system to measure heart rate during a tennis training session. Proc Inst Mech Eng P J Sports Eng Technol. 2025 Sep;239(3):545–551. doi: 10.1177/17543371231165102. doi. [DOI] [Google Scholar]
- 156.Ruiz-Malagón EJ, Ruiz-Alias SA, García-Pinillos F, Delgado-García G, Soto-Hermoso VM. Comparison between photoplethysmographic heart rate monitor from Polar Vantage M and Polar V800 with H10 chest strap while running on a treadmill: validation of the Polar Precision Prime photoplethysmographic system. Proc Inst Mech Eng P J Sports Eng Technol. 2021 Sep;235(3):212–218. doi: 10.1177/1754337120976659. doi. [DOI] [Google Scholar]
- 157.Sañudo B, De Hoyo M, Muñoz-López A, Perry J, Abt G. Pilot study assessing the influence of skin type on the heart rate measurements obtained by photoplethysmography with the Apple Watch. J Med Syst. 2019 May 22;43(7):195. doi: 10.1007/s10916-019-1325-2. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 158.Marzano-Felisatti JM, De Lucca L, Priego-Quesada JI, Pino-Ortega J. Heart rate measurement accuracy during intermittent efforts under laboratory conditions: a comparative analysis between chest straps and armband. Appl Sci. 2024;14(24):11872. doi: 10.3390/app142411872. doi. [DOI] [Google Scholar]
- 159.Claes J, Buys R, Avila A, et al. Validity of heart rate measurements by the Garmin Forerunner 225 at different walking intensities. J Med Eng Technol. 2017 Aug;41(6):480–485. doi: 10.1080/03091902.2017.1333166. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 160.Falter M, Scherrenberg M, Driesen K, et al. Smartwatch-based blood pressure measurement demonstrates insufficient accuracy. Front Cardiovasc Med. 2022;9:958212. doi: 10.3389/fcvm.2022.958212. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161.Van Oost N, Masci F, Schyvens AM, et al. A new perspective on validation of wearables for stress monitoring of occupational drivers. 2024 46th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); Jul 15-19, 2024; Orlando, FL, USA. Presented at. doi. [DOI] [PubMed] [Google Scholar]
- 162.Vermunicht P, Makayed K, Buyck C, et al. Continuous heart rate measurements in patients with cardiac disease: device comparison and development of a novel artefact removal procedure. Digit Health. 2025;11:20552076251337598. doi: 10.1177/20552076251337598. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 163.Rodrigues E, Lima D, Barbosa P, et al. HRV monitoring using commercial wearable devices as a health indicator for older persons during the pandemic. Sensors (Basel) 2022 Mar 4;22(5):2001. doi: 10.3390/s22052001. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 164.Damasceno VDO, Costa ADS, Campello MCA, et al. Criterion validity and accuracy of a heart rate monitor. Hum Mov. 2022;23(1):60–68. doi: 10.5114/hm.2021.104188. doi. [DOI] [Google Scholar]
- 165.Lima CS, Bertocco FC, de Oliveira JIV, de Souza TMF, da Silva EP, Von Zuben FJ. Assessment of Samsung Galaxy Watch4 PPG-based heart rate during light-to-vigorous physical activities. IEEE Sens Lett. 2024;8(7):1–4. doi: 10.1109/LSENS.2024.3408089. doi. [DOI] [Google Scholar]
- 166.Pipek LZ, Nascimento RFV, Acencio MMP, Teixeira LR. Comparison of SpO2 and heart rate values on Apple Watch and conventional commercial oximeters devices in patients with lung disease. Sci Rep. 2021 Sep 23;11(1):18901. doi: 10.1038/s41598-021-98453-3. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 167.Jørgensen A, Toftager M, Eghøj M, Ried-Larsen M, Bjørk Petersen C. Heart rate responses, agreement and accuracy among persons with severe disabilities participating in the indirect movement program: Team Twin—an observational study. Front Sports Act Living. 2023;5:1213655. doi: 10.3389/fspor.2023.1213655. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 168.Leth S, Hansen J, Nielsen OW, Dinesen B. Evaluation of commercial self-monitoring devices for clinical purposes: results from the Future Patient Trial, phase I. Sensors (Basel) 2017 Jan 22;17(1):211. doi: 10.3390/s17010211. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 169.Støve MP, Graversen AH, Sørensen J. Assessment of noninvasive oxygen saturation in patients with COPD during pulmonary rehabilitation: smartwatch versus pulse oximeter. Respir Care. 2023 Aug;68(8):1041–1048. doi: 10.4187/respcare.10760. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 170.Støve MP, Hansen ECK. Accuracy of the Apple Watch Series 6 and the Whoop Band 3.0 for assessing heart rate during resistance exercises. J Sports Sci. 2022 Dec;40(23):2639–2644. doi: 10.1080/02640414.2023.2180160. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 171.Støve MP, Haucke E, Nymann ML, Sigurdsson T, Larsen BT. Accuracy of the wearable activity tracker Garmin Forerunner 235 for the assessment of heart rate during rest and activity. J Sports Sci. 2019 Apr;37(8):895–901. doi: 10.1080/02640414.2018.1535563. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 172.Støve MP, Holm RS, Kjaersgaard AS, Duncker K, Jensen MR, Larsen BT. Measurement latency significantly contributes to reduced heart rate measurement accuracy in wearable devices. J Med Eng Technol. 2020 Apr;44(3):125–132. doi: 10.1080/03091902.2020.1753836. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 173.Barrios L, Oldrati P, Santini S, Lutterotti A. Evaluating the accuracy of heart rate sensors based on photoplethysmography for in-the-wild analysis. PervasiveHealth’19: The 13th EAI International Conference on Pervasive Computing Technologies for Healthcare; May 20-23, 2019; Trento Italy. Presented at. doi. [DOI] [Google Scholar]
- 174.Gilgen-Ammann R, Roos L, Wyss T, et al. Validation of ambulatory monitoring devices to measure energy expenditure and heart rate in a military setting. Physiol Meas. 2021 Aug 27;42(8) doi: 10.1088/1361-6579/ac19f9. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 175.Kozak D, Burgermeister S, De Chassey JDB, Naef A, Maringue A, Dietrich D. A functionality, safety and validity study of innovative REACH devices. Gerontechnology. 2017 Nov 7;16(3):181–188. doi: 10.4017/gt.2017.16.3.007.00. doi. [DOI] [Google Scholar]
- 176.Stucky B, Clark I, Azza Y, et al. Validation of Fitbit Charge 2 sleep and heart rate estimates against polysomnographic measures in shift workers: naturalistic study. J Med Internet Res. 2021 Oct 5;23(10):e26476. doi: 10.2196/26476. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 177.Schweizer T, Gilgen-Ammann R. Wrist-worn and arm-worn wearables for monitoring heart rate during sedentary and light-to-vigorous physical activities: device validation study. JMIR Cardio. 2025 Mar 21;9:e67110. doi: 10.2196/67110. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 178.Kinnunen H, Rantanen A, Kenttä T, Koskimäki H. Feasible assessment of recovery and cardiovascular health: accuracy of nocturnal HR and HRV assessed via ring PPG in comparison to medical grade ECG. Physiol Meas. 2020 May 7;41(4):04NT01. doi: 10.1088/1361-6579/ab840a. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 179.Nuuttila OP, Korhonen E, Laukkanen J, Kyröläinen H. Validity of the wrist-worn Polar Vantage V2 to measure heart rate and heart rate variability at rest. Sensors (Basel) 2021 Dec 26;22(1):137. doi: 10.3390/s22010137. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Sarhaddi F, Kazemi K, Azimi I, et al. A comprehensive accuracy assessment of Samsung smartwatch heart rate and heart rate variability. PLOS One. 2022;17(12):e0268361. doi: 10.1371/journal.pone.0268361. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 181.Hermand E, Coll C, Richalet JP, Lhuissier FJ. Accuracy and reliability of pulse O2 saturation measured by a wrist-worn oximeter. Int J Sports Med. 2021 Dec;42(14):1268–1273. doi: 10.1055/a-1337-2790. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 182.Bretonneau Q, Peruque-Gayou E, Wolfs E, Bosquet L. Accuracy of heart-rate-recovery parameters assessed from a wrist-worn photoplethysmography monitor (Polar Unite) Int J Sports Physiol Perform. 2023 Jan 1;19(1):13–18. doi: 10.1123/ijspp.2023-0120. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 183.Bretonneau Q, Peruque-Gayou E, Wolfs E, Bosquet L. Parameters influencing the accuracy of a wrist photoplethysmography heart-rate monitor (Polar Unite) during exercise. Int J Sports Physiol Perform. 2023 Apr 1;18(4):440–443. doi: 10.1123/ijspp.2022-0288. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 184.Ceugniez M, Devanne H, Hermand E. Reliability and accuracy of the Fitbit Charge 4 photoplethysmography heart rate sensor in ecological conditions: validation study. JMIR mHealth uHealth. 2025 Jan 8;13:e54871. doi: 10.2196/54871. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 185.Tedesco S, Sica M, Ancillao A, Timmons S, Barton J, O’Flynn B. Accuracy of consumer-level and research-grade activity trackers in ambulatory settings in older adults. PLoS One. 2019;14(5):e0216891. doi: 10.1371/journal.pone.0216891. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 186.Giggins OM, Doyle J, Sojan N, et al. Accuracy of wrist-worn photoplethysmography devices at measuring heart rate in the laboratory and during free-living activities. 2021 43rd Annual International Conference of the IEEE Engineering in Medicine & Biology Society (EMBC); Nov 1-5, 2021; Mexico. Presented at. doi. [DOI] [PubMed] [Google Scholar]
- 187.O’Grady B, Lambe R, Baldwin M, Acheson T, Doherty C. The validity of Apple Watch Series 9 and Ultra 2 for serial measurements of heart rate variability and resting heart rate. Sensors (Basel) 2024 Sep 26;24(19):6220. doi: 10.3390/s24196220. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 188.Müller AM, Wang NX, Yao J, et al. Heart rate measures from wrist-worn activity trackers in a laboratory and free-living setting: validation study. JMIR mHealth uHealth. 2019 Oct 2;7(10):e14120. doi: 10.2196/14120. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 189.Gorny AW, Liew SJ, Tan CS, Müller-Riemenschneider F. Fitbit Charge HR wireless heart rate monitor: validation study conducted under free-living conditions. JMIR mHealth uHealth. 2017 Oct 20;5(10):e157. doi: 10.2196/mhealth.8233. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 190.Ravanelli N, Lefebvre K, Brough A, Paquette S, Lin W. Validation of an open-source smartwatch for continuous monitoring of physical activity and heart rate in adults. Sensors (Basel) 2025 May 6;25(9):2926. doi: 10.3390/s25092926. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 191.Theurl F, Schreinlechner M, Sappler N, et al. Smartwatch-derived heart rate variability: a head-to-head comparison with the gold standard in cardiovascular disease. Eur Heart J Digit Health. 2023 May;4(3):155–164. doi: 10.1093/ehjdh/ztad022. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 192.Neudorfer M, Kumar D, Smeddinck JD, et al. Validity of four consumer‐grade optical heart rate sensors for assessing volume and intensity distribution of physical activity. Scand J Med Sci Sports. 2024 Nov;34(11) doi: 10.1111/sms.14756. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 193.Yang J, Lu Z, Li X, Tao K, Qiu J. Validity of Hera Leto consumer earbuds for heart rate monitoring during stationary cycling. Sci Rep. 2025;15(1):15944. doi: 10.1038/s41598-025-00559-5. doi. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 194.Rafl J, Bachman TE, Rafl-Huttova V, Walzel S, Rozanek M. Commercial smartwatch with pulse oximeter detects short-time hypoxemia as well as standard medical-grade device: validation study. Digit Health. 2022;8:20552076221132127. doi: 10.1177/20552076221132127. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 195.Walzel S, Mikus R, Rafl-Huttova V, Rozanek M, Bachman TE, Rafl J. Evaluation of leading smartwatches for the detection of hypoxemia: comparison to reference oximeter. Sensors (Basel) 2023 Nov 14;23(22):9164. doi: 10.3390/s23229164. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 196.Haveman ME, van Rossum MC, Vaseur RME, et al. Continuous monitoring of vital signs with wearable sensors during daily life activities: validation study. JMIR Form Res. 2022 Jan 7;6(1):e30863. doi: 10.2196/30863. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 197.Hoevenaars D, Yocarini IE, Paraschiakos S, et al. Accuracy of heart rate measurement by the Fitbit Charge 2 during wheelchair activities in people with spinal cord injury: instrument validation study. JMIR Rehabil Assist Technol. 2022 Jan 19;9(1):e27637. doi: 10.2196/27637. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 198.Turki AF. An investigation into the accuracy of blood oxygen saturation measurement using smart watch devices among pediatric users in Saudi Arabia. 2023 IEEE International Biomedical Instrumentation and Technology Conference (IBITeC); Nov 9-10, 2023; Yogyakarta, Indonesia. Presented at. doi. [DOI] [Google Scholar]
- 199.Khushhal AA, Mohamed AA, Elsayed ME. Accuracy of Apple Watch to measure cardiovascular indices in patients with chronic diseases: a cross sectional study. J Multidiscip Healthc. 2024;17:1053–1063. doi: 10.2147/JMDH.S449071. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 200.Ho WT, Yang YJ, Li TC. Accuracy of wrist-worn wearable devices for determining exercise intensity. Digit Health. 2022;8:20552076221124393. doi: 10.1177/20552076221124393. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 201.Chow HW, Yang CC. Accuracy of optical heart rate sensing technology in wearable fitness trackers for young and older adults: validation and comparison study. JMIR mHealth uHealth. 2020 Apr 28;8(4):e14707. doi: 10.2196/14707. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 202.Arslan B, Sener K, Guven R, et al. Accuracy of the Apple Watch in measuring oxygen saturation: comparison with pulse oximetry and ABG. Ir J Med Sci. 2024 Feb;193(1):477–483. doi: 10.1007/s11845-023-03456-w. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 203.Paslı S, Topçuoğlu H, Yılmaz M, Yadigaroğlu M, İmamoğlu M, Karaca Y. Diagnostic accuracy of Apple Watch ECG outputs in identifying dysrhythmias: a comparison with 12-lead ECG in emergency department. Am J Emerg Med. 2024 May;79:25–32. doi: 10.1016/j.ajem.2024.01.046. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 204.Jaiswal P, Sahu NK, Lone HR. Comparative assessment of smartwatch photoplethysmography accuracy. IEEE Sens Lett. 2024;8(1):1–4. doi: 10.1109/LSENS.2023.3342292. doi. [DOI] [Google Scholar]
- 205.Imai H, Shimada J. Evaluation of telemetric single-lead cardiac transmitter and Apple Watch for heart rate monitoring: implications for heart failure management in home care. Cureus. 2025 Mar;17(3):e80232. doi: 10.7759/cureus.80232. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 206.Ibrahim NS, Rampal S, Lee WL, Pek EW, Suhaimi A. Evaluation of wrist-worn photoplethysmography trackers with an electrocardiogram in patients with ischemic heart disease: a validation study. Cardiovasc Eng Tech. 2024 Feb;15(1):12–21. doi: 10.1007/s13239-023-00693-z. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 207.Flett IL, Su Y, Cameron C, Zhou C, Holder-Pearson L, Chase JG. Assessing the accuracy of a wrist-worn wearable device’s heart rate measurement over changing levels of activity. SSRN. 2025 Sep 10; doi: 10.2139/ssrn.5412841. Preprint posted online on. doi. [DOI]
- 208.Danielsson ML, Vergeer M, Plasqui G, Baumgart JK. Accuracy of the Apple Watch Series 4 and Fitbit Versa for assessing energy expenditure and heart rate of wheelchair users during treadmill wheelchair propulsion: cross-sectional study. JMIR Form Res. 2024 May 7;8:e52312. doi: 10.2196/52312. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 209.Ferrera LE, Caballero JM, Kurniawan Y, Tedjasaputra JA, Sari E. Proceedings of the 4th International Conference on Human-Computer Interaction and User Experience in Indonesia, CHIuXiD '18. Association for Computing Machinery; 2018. Health quadrant mHealth application: a performance analysis of the heart rate accuracy vis a vis a pulse oximeter; pp. 54–60. doi. [DOI] [Google Scholar]
- 210.Jachymek M, Jachymek MT, Kiedrowicz RM, Kaźmierczak J, Płońska-Gościniak E, Peregud-Pogorzelska M. Wristbands in home-based rehabilitation-validation of heart rate measurement. Sensors (Basel) 2021 Dec 23;22(1):60. doi: 10.3390/s22010060. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 211.Sjöberg V, Westergren J, Monnier A, et al. Wrist-worn activity trackers in laboratory and free-living settings for patients with chronic pain: criterion validity study. JMIR mHealth uHealth. 2021 Jan 12;9(1):e24806. doi: 10.2196/24806. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 212.Choksatchawathi T, Ponglertnapakorn P, Ditthapron A, et al. Improving heart rate estimation on consumer grade wrist-worn device using post-calibration approach. IEEE Sensors J. 2020;20(13):7433–7446. doi: 10.1109/JSEN.2020.2979191. doi. [DOI] [Google Scholar]
- 213.Maijala A, Kinnunen H, Koskimäki H, Jämsä T, Kangas M. Nocturnal finger skin temperature in menstrual cycle tracking: ambulatory pilot study using a wearable Oura Ring. BMC Womens Health. 2019 Nov 29;19(1):150. doi: 10.1186/s12905-019-0844-9. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 214.Stahl AC, Tietz AS, Kendziora B, Dewey M. Has the STARD statement improved the quality of reporting of diagnostic accuracy studies published in European Radiology? Eur Radiol. 2023 Jan;33(1):97–105. doi: 10.1007/s00330-022-09008-7. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 215.Plint AC, Moher D, Morrison A, et al. Does the CONSORT checklist improve the quality of reports of randomised controlled trials? A systematic review. Med J Aust. 2006 Sep 4;185(5):263–267. doi: 10.5694/j.1326-5377.2006.tb00557.x. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 216.Fryar CD, Carroll MD, Afful J. National Center for Health Statistics (NCHS), Centers for Disease Control and Prevention (CDC); 2020. [15-07-2026]. Prevalence of Underweight Among Adults Aged 20 and Over: United States, 1960–1962 Through 2017–2018.https://www.cdc.gov/nchs/data/hestat/underweight-adult-17-18/ESTAT-Underweight-Adult-H.pdf URL. Accessed. [Google Scholar]
- 217.Waist circumference and BMI. Australian Bureau of Statistics. 2022. [15-07-2026]. https://www.abs.gov.au/statistics/health/health-conditions-and-risks/waist-circumference-and-bmi/latest-release URL. Accessed.
- 218.National Academies of Sciences, Engineering, and Medicine . Improving Representation in Clinical Trials and Research: Building Research Equity for Women and Underrepresented Groups. The National Academies Press; 2022. [15-07-2026]. Barriers to representation of underrepresented and excluded populations in clinical research.https://www.ncbi.nlm.nih.gov/books/NBK584407/ URL. Accessed. [PubMed] [Google Scholar]
- 219.Warburton DER, Gledhill N, Jamnik VK, et al. Evidence-based risk assessment and recommendations for physical activity clearance: Consensus Document 2011. Appl Physiol Nutr Metab. 2011 Jul;36 Suppl 1:S266–S298. doi: 10.1139/h11-062. doi. Medline. [DOI] [PubMed] [Google Scholar]
- 220.Vasudevan S, Vogt WC, Weininger S, Pfefer TJ. Melanometry for objective evaluation of skin pigmentation in pulse oximetry studies. Commun Med (Lond) 2024 Jul 11;4(1):138. doi: 10.1038/s43856-024-00550-7. doi. Medline. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.