Macroscopic and microscopic findings following thulium laser exposure

Group Peak temperature (°C) Charring depth (µm) Histology (H&E, 40x)
Dry 28 ± 2 0 Intact, uniform fibres
Water-soaked 32 ± 3 0 Intact, uniform fibres
Blood-soaked 65 ± 5 250 ±50 Superficial carbonisation, deeper structures intact on H&E but not formally characterised