Retinal fluid segmentation in real-world OCT imaging of neovascular AMD using hybrid deep learning and graph-based optimization

Zhi Chen 1,2,*, Bernardo Bach 3, Honghai Zhang 1,2, Andreas Wahle 1,2, Ian C Han 3, H C Boldt 3, Stephen R Russell 3, Milan Sonka 1,2, Elliott H Sohn 3
PMCID: PMC13592241  PMID: 42770020

Abstract

Accurate segmentation and quantification of retinal fluid in optical coherence tomography (OCT) images are important for assessing disease activity and treatment response in neovascular age-related macular degeneration (nvAMD). Yet, manual delineation of intraretinal fluid (IRF), subretinal fluid (SRF), and pigment epithelial detachment (PED) is labor-intensive and subject to inter-grader variability. We developed a hybrid framework for robust retinal fluid segmentation in clinical-grade longitudinal OCT, combining anatomically constrained retinal-region extraction using Deep LOGISMOS with nnU-Net–based 2-D and 3-D fluid segmentation. To improve adaptation to heterogeneous clinical data while reducing annotation burden, an iterative expert-guided labeling strategy was used, starting with 70 RETOUCH OCT scans and 100 scans from subjects with macular neovascularization at the University of Iowa and expanding to a final aggregated training set of 291 scans. The method was validated on a held-out test set of 50 OCT volumes from 50 nvAMD subjects, with a subset of 20 volumes independently annotated by a second expert for inter-observer analysis. It was further applied to a longitudinal dataset of 14,500 Heidelberg Spectralis macular OCT scans from 221 independent subjects. The ensemble model achieved Dice similarity coefficients (%) of 89.5±11.7 for IRF, 88.0±10.7 for SRF, and 82.0±13.0 for PED against the primary expert annotations, with performance approaching expert-level agreement. In longitudinal inference, our developed framework supports robust large-scale quantification of fluid burden and spatial extent across heterogeneous scan protocols and extended follow-up. The artificial intelligence-based approach has the potential to enable scalable and clinically meaningful longitudinal analysis of retinal fluid in nvAMD.

1. Introduction

Age-related macular degeneration (AMD) is the most common cause of moderate to severe irreversible vision impairment in individuals over 50 years of age in the western world [1]. In the United States, AMD accounts for 54.4% of blindness cases among patients of Caucasian ancestral background older than 40 [2]. Globally, an estimated 196 million people were affected in 2020, and the projection is to reach 288 million by 2040 [3]. The most severe forms of AMD are classified as advanced AMD, characterized by the development of either neovascular AMD (nvAMD) or geographic atrophy [4]. While initial steps toward treating geographic atrophy have been recently taken with the FDA approval of complement factor inhibitors, such as the C3 inhibitor pegcetacoplan [5], treatment for nvAMD is well established with intravitreal injections of vascular endothelial growth factor inhibitors (anti-VEGF), which have been widely used for years and significantly reduced the incidence of blindness in patients with nvAMD [6–8].

Optical coherence tomography (OCT) is routinely applied as a non-invasive imaging method for diagnosis, evaluation of disease activity, and assessment of treatment responsiveness in nvAMD [9–11]. The quantification metrics derived from OCT have been proven to be important biomarkers for final prognosis after treatment [11–13].

Patients with nvAMD have an unpredictable treatment responsiveness, meaning that some show limited improvement and require more anti-VEGF injections at shorter intervals, while others greatly improve and stabilize with fewer shots at longer intervals. Consequently, there is increasing interest in evaluating OCT biomarkers that can potentially predict treatment burden and outcomes [12–14].

Some studies suggest that intraretinal fluid (IRF) may be the main biomarker in predicting visual acuity (VA) at the initial OCT evaluation [14–16]. Moreover, the presence of IRF, as well as its persistence, volume, extent, and foveal position, have been associated with worse treatment response and prognosis [14,17–24]. Subretinal fluid (SRF) shows more conflicting results regarding its prognostic relevance, with some studies suggesting that its presence may be related to better final VA [15,25–29], while others report no significant difference [30,31]. However, the presence of SRF has been associated with disease activity, and rapid SRF resolution and repeated visits without SRF have been associated with better outcomes [20,32].

Fluctuations in SRF volume and the coexistence of SRF with IRF, pigment epithelial detachment (PED), or subretinal hyperreflective material have also been described as poor prognostic biomarkers [13,33]. Regarding PED, its presence, as well as increased width and size, have been associated with worse prognosis, particularly when PED is accompanied by IRF or SRF [17,33–35].

Most studies evaluating OCT biomarkers rely on qualitative assessments or manual segmentation to quantify IRF, SRF, and PED (Fig. 1). Such approaches are time-consuming, prone to inter-grader variability, and difficult to scale for large datasets. In addition, many analyses use 2-D OCT measurements even though these fluid compartments are inherently 3-D and are better represented volumetrically. In nvAMD, visible exudation as IRF, SRF, and or PED on OCT is the clinical standard for diagnosing and monitoring disease activity, but measurement methods have been inconsistent, ranging from presence/absence grading to manual delineation on selected B-scans, particularly in longitudinal studies with many follow-up visits.

Fig. 1.

Fig. 1.

Representative OCT B-scan demonstrating typical appearances of retinal fluid. Intraretinal fluid (IRF) appears as hyporeflective cystic spaces within the neurosensory retina, subretinal fluid (SRF) is located between the neurosensory retina and the retinal pigment epithelium, and pigment epithelial detachment (PED) is characterized by elevation of the retinal pigment epithelium from the underlying tissue.

To improve scalability, prior work introduced computational pipelines for volumetric fluid analysis, including classical model-based approaches such as fuzzy level-set segmentation [36]. More recently, deep learning methods have become the dominant approach for automated fluid detection and quantification across devices and exudative macular diseases [37–40]. For AMD, Sappa et al. developed RetFluidNet, a convolutional neural network incorporating skip connections and atrous spatial pyramid pooling to capture multi-scale contextual information, enabling automated segmentation of major fluid-related abnormalities from SD-OCT images [38]. Guo et al. proposed ReF-Net for volumetric retinal fluid segmentation in DME and further demonstrated that combining structural OCT with OCT angiography information improved segmentation performance, highlighting the potential value of multimodal information for automated fluid quantification [39]. Hsu et al. developed a modified U-Net framework to segment intraretinal cystoid fluid and SRF together with photoreceptor-related features in DME, and further examined associations between these automatically quantified OCT biomarkers and visual acuity [40]. Deep learning-based fluid quantification has also been applied at scale in nvAMD to characterize anti-VEGF response using compartment-specific fluid volumes [26], with repeatability studies supporting their potential clinical utility [41]. However, these methods still face key limitations. First, most supervised segmentation pipelines require extensive expert labeling [38–40]; consequently, training sets can limit dataset size and under-represent severe or heavy-fluid presentations, which are precisely the scans that pose the greatest challenge to models. Second, domain shift across devices, scan protocols, and real-world image artifacts can degrade performance, and deep-learning approaches can be sensitive to the input distribution — raising concerns about stability when deployed outside the development cohort [37,41]. Third, many methods operate primarily on individual 2-D B-scans [38–40], which can yield anatomically inconsistent masks across adjacent B-scans and unstable volumetric estimates, especially in extensive SRF/IRF or complex PED morphology.

Building on the multi-vendor, multi-disease benchmarking established by RETOUCH [42], we developed a robust nnU-Net-based framework [43] that is tailored for real-world longitudinal OCT data. First, standardized retinal-region preprocessing is applied to ensure anatomically consistent retinal localization across scans, combining deep learning-based region detection with Deep LOGISMOS surface optimization [44,45]. This step reduces variability introduced by different devices and evolving scan protocols, which is common in long-term clinical follow-up. Second, the model is pretrained on diverse public OCT data (RETOUCH) and subsequently fine-tuned to the target clinical nvAMD dataset, allowing adaptation to site-specific imaging characteristics and longitudinal acquisition patterns. Finally, an iterative labeling strategy is used in which automated segmentations are reviewed and selectively refined, with particular attention to scans containing extensive or complex fluid. This process enables efficient improvement of annotation quality and model performance in clinically challenging cases. In this study, we further validate the proposed framework by explicitly accounting for inter-observer variability and demonstrate its applicability to large-scale longitudinal OCT data acquired in routine clinical practice.

2. Methods

As shown in Fig. 2, the proposed method consists of two interconnected stages: pre-segmentation of the retina and final segmentation of fluids. During the pre-segmentation stage, the retina and the top and bottom background regions are segmented first; subsequently, the target fluids, including IRF, SRF, and PED, are simultaneously identified in 3-D during the final segmentation stage. The pre-segmentation stage is implemented using a 2-D U-Net network followed by a two-surface 3-D LOGISMOS approach [44], while the final segmentation stage employs a 3-D U-Net network.

Fig. 2.

Fig. 2.

Workflow of the OCT fluid analysis framework. The retina extraction stage identifies the whole retina region using deep learning and optimal graph-based surface extraction, while the fluid segmentation stage identifies individual fluids, including PED (blue), IRF (red), and SRF (green), within focal retina region using U-Net models. An adaptive inference strategy is applied to accommodate heterogeneous real-world clinical OCT data.

To support efficient analysis of longitudinal clinical OCT, the framework was designed to reuse prior knowledge while adapting to the heterogeneity of real-world OCT. Rather than relying entirely on de-novo manual annotation, the segmentation models are initialized from previously learned representations. When needed, the models are further refined for the target nvAMD cohort through an iterative labeling strategy, in which automated results are selectively reviewed and corrected before added to the next-iteration training set. This improves annotation efficiency, focuses expert effort on challenging cases, and progressively enhances model robustness. Another challenge for longitudinal data is the substantial variation in scan protocols across visits, including A-scan counts ranging from 512 to 1536, B-scan counts from 7 to 121, and differences in voxel spacing across acquisition settings (as shown in Table 1). To address this variability, the framework first normalizes the retinal region through pre-segmentation and anatomically constrained retinal extraction, thereby reducing variation in image dimensions. The inference stage then employs an adaptive strategy that leverages 3-D contextual information when B-scan continuity is adequate, while remaining robust and flexible for scans acquired under suboptimal settings.

Table 1. Summary of datasets used in this study.

Dataset Device Images Image Size a Voxel Size ( μ m) B-scans
RETOUCH Cirrus b 24 512×1024×128 11.72×1.95×46.88 3,072
Total=70 Spectralis c 24 512×496×49 11.72×3.87×123.72 1,176
Topcon d 22 512×885×128 11.72×2.60×46.88 2,816

UI_AMD Training Spectralis 169 512×496×19 11.80×3.87×250.8 3211
Total=221 2 512×496×25 11.3×3.87×241.4 50
1 512×496×49 10.74×3.87×241.4 49
10 768×496×19 10.7×3.87×226.4 190
20 768×496×25 11.1×3.87×237.2 500
12 768×496×31 11.2×3.87×239.3 372
1 768×496×121 11.6×3.87×61.89 121
1 1024×496×19 5.3×3.87×226.8 73
1 1024×496×73 5.83×3.87×62.15 73
1 1024×496×97 5.83×3.87×62.15 97
1 1536×496×7 5.41×3.87×230.9 7
2 1536×496×19 5.8×3.87×246.9 38

UI_AMD Testing f Spectralis 38 512×496×19 11.80×3.87×250.8 722
Total=50 1 512×496×25 11.3×3.87×241.4 25
2 768×496×19 10.7×3.87×226.4 38
7 768×496×25 11.1×3.87×237.2 175
2 768×496×31 11.2×3.87×239.3 62

UI_AMD Longitudinal Spectralis 14,500 Multiple e Multiple e ∼ 290,000
a

X × Y × Z. X: number of A-scans per B-scan, Y: number of voxels per A-scan, Z: number of B-scans.

b

Carl Zeiss Meditec, Dublin, CA, USA.

c

Heidelberg Engineering, Heidelberg, Germany;

d

Topcon Corporation, Tokyo, Japan.

e

Comprising all acquisition protocols and voxel resolutions above.

f

20 were independently annotated by a second expert for inter-observer variability analysis.

Note: Datasets denoted bold have reference tracing available.

2.1. Retina pre-segmentation

OCT B-scans exhibit a pronounced imbalance between retinal tissue and background, as the retina typically occupies only a small fraction of the image. Training deep learning models directly on full B-scans can therefore bias segmentation toward background regions and increase computational cost. To mitigate this issue, we employed a retinal pre-segmentation stage to localize the retinal region prior to fluid segmentation. This stage was implemented using a 2-D U-Net from the nnU-Net framework to classify voxels into 3 regions — top background, retina, and bottom background — where the retinal region is defined as the area between the inner limiting membrane (ILM) and the outer aspect of the retinal pigment epithelium (RPE) complex.

The pre-segmentation model was not trained on AMD datasets but was reused from our prior work on OCT layer segmentation [44], which was developed using datasets without retinal fluid. When applied to unseen AMD datasets, this model may produce region-level inaccuracies, such as holes, irregular attachments, or misclassified regions, due to domain differences in pathology and imaging characteristics. Direct conversion of such region-based predictions into anatomically consistent surfaces is challenging and would typically require ad hoc post-processing. Retraining a dedicated pre-segmentation model for AMD would require extensive manual annotation of retinal layers across large volumes of data, imposing a substantial annotation burden.

In contrast, the Deep LOGISMOS framework leverages in-region costs derived from the three-region probability maps together with edge-based gradient information of ILM and RPE to impose robust surface-based constraints. In addition, our approach explicitly incorporates 3-D contextual information, including inter-layer ordering, smoothness constraints across neighboring layers, and continuity across adjacent B-scans. Together, these properties of our approach enable fast and reliable retinal layer identification despite region-level imperfections from deep learning model [44,45].

As a result, accurate retinal region extraction is achieved even in the presence of fluid-related abnormalities of retinal morphology. The extracted retinal region was subsequently cropped based on the maximum retinal thickness and padded with additional margins to preserve contextual information, including potential PED regions below the RPE, before being passed to the downstream U-Net network for fluid segmentation.

2.2. Fluid segmentation

Fluid segmentation was performed using a 3-D deep learning framework based on nnU-Net, applied to anatomically normalized retinal volumes from the pre-segmentation stage. As part of this normalization, all OCT volumes were resampled to a uniform in-plane pixel resolution, ensuring consistent input dimensions despite the heterogeneous scan protocols used across datasets (Table 1). This standardization maps scans of varying resolution and field size into a common coordinate space, minimizing protocol-induced variability and enabling the model to focus on fluid morphology rather than acquisition-related differences. These novel aspects of our approach markedly enhance robustness across patients and longitudinal visits, where changes in imaging protocols are common in clinical practice.

The segmentation task was formulated as a multi-class voxel-wise classification problem with four labels: background, IRF, SRF, and PED. Ground truth labels were constructed iteratively as described in Section 3.3. Following size normalization, all images and corresponding labels underwent nnU-Net’s standard preprocessing pipeline, including dataset-adaptive intensity normalization, before training [43].

Segmentation was implemented within the nnU-Net framework using residual encoder U-Net architectures [46]. The residual encoder design facilitates stable training of deeper networks by improving gradient flow and feature reuse, which is particularly beneficial for AMD fluid segmentation. Retinal fluid in AMD exhibits large variability in size, shape, and appearance, often coexisting with subtle anatomical distortions and low-contrast boundaries. This residual encoder design thus enables the network to capture both fine structural details and broader contextual patterns, supporting accurate delineation of heterogeneous fluid morphologies while preserving retinal layer continuity.

To evaluate the influence of dimensionality and volumetric context, multiple configurations were trained and compared: 2-D, 3-D low-resolution, 3-D full-resolution, and an ensemble of these variants. The 3-D models exploit continuity across adjacent B-scans, which is critical for accurately segmenting fluid structures that span multiple slices, while the 2-D model provides complementary slice-wise detail. All models followed nnU-Net’s standard patch-based training strategy and default optimization settings, including a Dice-based loss function to mitigate class imbalance. Model development and hyperparameter tuning relied on five-fold cross-validation.

3. Experimental methods

3.1. University of Iowa AMD (UI_AMD) datasets

In this study, OCT volume scans were retrospectively collected from patients with neovascular age-related macular degeneration (nvAMD) followed at the University of Iowa Health Care between 2015 and 2021. Patients with at least one visit for nvAMD during this period were enrolled in this IRB-approved study. OCT scans from all available visits, including pre-treatment, during treatment, and post-treatment follow-up, were included to capture longitudinal disease progression. Patients with ocular conditions that could confound or obscure nvAMD-related OCT findings were excluded from analysis.

The resulting UI_AMD dataset comprises longitudinal spectral-domain OCT volumes consistently acquired for each patient using the Heidelberg Spectralis platform, with repeated imaging across multiple clinical visits (Table 1). Imaging was performed in a registered manner using the instrument’s commercially available system, allowing direct comparison of scans across visits with prior imaging. As summarized in the table, the dataset includes scans acquired with heterogeneous protocols, encompassing varying numbers of A-scans per B-scan (512, 768, 1024, 1536) and B-scan counts (7, 19, 25, 31, 49, 73, 97, 121). This reflects real-world clinical practice, where scan protocols are adapted over time and across visits based on disease status and clinical needs.

The UI_AMD dataset therefore spans a wide range of image dimensions, voxel sizes, and scan densities, as well as diverse stages of neovascular disease severity and treatment response. This heterogeneity provides a real world example of representative cohort for model development and evaluation, enabling assessment of algorithm robustness across variable imaging settings, and differing disease burdens.

3.2. RETOUCH public datasets

The RETOUCH dataset was utilized in this study as a curated source of expert-labeled retinal OCT data for fluid segmentation [42]. In this work, only the RETOUCH training dataset was used to support supervised model development. As described in the section 2.1, a retinal pre-segmentation stage was first applied to localize and normalize the retinal region, providing a consistent anatomical context for subsequent fluid analysis. This preprocessing step enables the fluid segmentation model to focus on clinically relevant retinal structures while reducing the impact of abundant background pixels and potential imaging artifacts.

The RETOUCH training dataset consists of 3-D spectral-domain OCT volumes with voxel-wise expert annotations generated by experienced graders from independent clinical centers. The images were acquired using 3 widely used clinical OCT platforms — Heidelberg Spectralis, Zeiss Cirrus, and Topcon — resulting in substantial variability in imaging characteristics and lesion presentation (Table 1). This dataset encompasses a broad spectrum of lesion sizes, shapes, and spatial distributions. When combined with the pre-segmented and anatomically normalized retinal regions, the training data support learning of clinically representative fluid morphologies and facilitate robust generalization to independent patient cohorts.

3.3. Iterative labeling

Using the RETOUCH training dataset as an initial source of expert-labeled OCT volumes, we adopted an iterative, semi-automatic labeling workflow for fluid segmentation. Retinal pre-segmentation was first applied to localize and normalize the retinal region, after which model-generated segmentations were subjected to systematic quality control and selectively refined by expert graders. This strategy enabled efficient construction of high-quality ground truth annotations while preserving clinical reliability. Because the nvAMD cases in this longitudinal cohort spanned a wider range of disease severity and fluid burden than those represented in the RETOUCH training data, and were acquired using different imaging protocols and devices (Section 3.1), additional expert-guided refinement was required to adapt the labels to the target clinical dataset.

In the first iteration, an initial segmentation model was trained at an early stage of data collection for the longitudinal cohort, using 100 randomly selected UI_AMD scans from the cohort together with 70 RETOUCH scans. This model was then applied to generate 3-fluid (IRF/SRF/PED) segmentation results for 837 scans acquired at later stages of the same longitudinal cohort.

To facilitate efficient quality control at scale, an automated anomaly detection step was applied to the segmented fluids across all B-scans. Deep feature embeddings were extracted from a pretrained ResNet-50 network [47]. An Isolation Forest algorithm [48] was then applied in the embedding space to identify atypical cases by recursively partitioning the feature space, such that anomalous samples are isolated with shorter path lengths in randomly constructed trees. Scans flagged as anomalous by this unsupervised screening were prioritized for manual review. Following further expert review, 3 scans were excluded due to image quality problems and 173 scans were identified with substantial segmentation errors (as shown in Fig. 3).

Fig. 3.

Fig. 3.

Examples illustrating the iterative refinement process for fluid segmentation. For each case, the original OCT B-scan is shown alongside segmentation outputs from an earlier model iteration and a later iteration after additional expert-guided refinement. Progressive improvement is observed in the delineation of PED (blue), IRF (red), and SRF (green), particularly in scans with extensive fluid. White arrows indicate regions where segmentation errors were improved or corrected in the later iteration.

Then a subset of 22 scans was randomly selected for manual correction and incorporated as the first refined batch. The segmentation model was then retrained in an iterative manner using progressively expanded training sets that combined the original UI_AMD and RETOUCH data with expert-refined labels and high-quality automated predictions. At each iteration, model outputs on previously problematic scans were reviewed, with some accepted without modification and others selectively refined by expert graders.

Over 4 refinement cycles, additional manually corrected scans and expert-approved automated predictions were incorporated, yielding a final aggregated training set of 291 scans, comprising 70 RETOUCH scans, 100 early-stage and 121 late-stage scans from the longitudinally-imaged cohort. In total, 59 scans were manually refined by experts, and an additional 62 scans with improved automated segmentations were accepted without further correction. Further refinement was not pursued, as additional iterations resulted in limited incremental improvement on the validation sets during model training, while expert graders considered the latest model outputs visually satisfactory.

Figure 3 shows representative examples from the iterative labeling process. Early model outputs demonstrate incomplete or fragmented fluid segmentation in complex cases, whereas later iterations achieve more complete and anatomically consistent delineation after targeted expert refinement. This iterative human-in-the-loop strategy progressively improved segmentation performance while reducing the manual annotation burden by focusing expert review on scans with automated segmentation errors.

3.4. Model training parameters

For the retinal pre-segmentation stage, we reused a previously developed Deep LOGISMOS-based retinal-region extraction model, which combines 2-D U-Net-based region detection with graph-based surface optimization to identify anatomically constrained retinal boundaries [44,45]. This component required no additional fine-tuning in the present study and was used to generate anatomically standardized retinal regions before fluid segmentation.

For the fluid segmentation stage, nnU-Net models with residual encoder were trained on 291 OCT volumes labeled for background, IRF, SRF, and PED. Five-fold cross-validation was performed during the training. The B-scan image input size was standardized to 256×512 . The 2-D model used a patch size of 256×512 and a batch size of 70, whereas the 3-D model used a patch size of 112×160×256 and a batch size of 2. Training used 1000 epochs, Dice and cross-entropy loss with deep supervision, foreground oversampling of 0.33, and stochastic gradient descent optimization with an initial learning rate of 0.01, momentum of 0.99, Nesterov acceleration, weight decay of 3×10−5 , and a polynomial learning-rate schedule.

3.5. Validation metrics

Evaluation was performed on a held-out test set of 50 subject OCT volumes, randomly selected from the longitudinal cohort, with one time point per patient and manual expert tracings as ground truth. One expert (M1) annotated all 50 subjects, while a second expert (M2) independently annotated a subset of 20 subjects to enable assessment of inter-observer variability (Table 1). For subjects annotated by both experts, a consensus mask was also generated by retaining only voxels for which both experts identified the same fluid label, providing a conservative reference restricted to regions of expert agreement.

Automated fluid segmentation performance was evaluated using the Dice similarity coefficient as the primary metric. Dice scores were computed separately for each fluid type (IRF, SRF and PED) to quantify spatial agreement between automated predictions and expert annotations. Automated segmentations were compared against individual expert annotations, and the consensus masks. Inter-observer agreement was quantified by comparing annotations from the two experts directly. Patient-level Dice scores were summarized across the cohort by reporting the mean and standard deviation for each fluid type and comparison, enabling robust assessment of model performance relative to expert variability and annotation uncertainty. In addition, agreement between automated and expert-derived fluid 3-D volumes was evaluated using linear regression.

3.6. Inference on longitudinal datasets

In addition to above small datasets used for training and validation, we collected a separate longitudinal test dataset (Table 1) consisting of 14,500 Heidelberg Spectralis macular OCT scans from 221 independent subjects with nvAMD at the University of Iowa Health Care. This dataset was used to evaluate the robustness of the proposed method in longitudinal settings. On average, each subject contributed 62.2 ± 52.0 scans acquired over 95.8 ± 43.3 months from the first to the most recent visit, reflecting real-world longitudinal follow-up with variable imaging frequency and duration.

Because OCT scans acquired in routine clinical practice vary in scan density and continuity between B-scans, an adaptive inference approach was used (Fig. 2). Structural similarity between adjacent B-scans was assessed using the structural similarity index (SSIM) [49] to determine whether a scan exhibited adequate 3-D continuity. Scans with consistent inter–B-scan structure were analyzed using a 3-D approach to take advantage of volumetric information, whereas scans with disrupted continuity were analyzed using a 2-D approach to avoid error propagation caused by unreliable slice-to-slice relationships. This adaptive strategy was applied only during inference on real-world longitudinal data; during validation, a fixed model configuration was used to ensure consistent and unbiased performance comparison across models.

To assess deployment efficiency, we conducted an additional runtime experiment using 100 randomly selected OCT volumes from the above dataset. The complete inference workflow was executed for each scan, and runtime was recorded at both the full-pipeline level and the neural-network inference level. For each experiment, we summarized the mean inference time per OCT volume, total processing time, and overall throughput in volumes per hours. We also documented the GPU configuration, including GPU model, memory capacity, and CUDA version, to support reproducibility and facilitate interpretation of runtime performance. This experiment was designed to quantify the practical feasibility of applying the proposed workflow to large-scale longitudinal OCT datasets in a clinical research setting.

3.7. Quantitative indices of AMD fluid, visualization of results

The following quantitative indices were derived from the automated 3-D fluid segmentations to characterize fluid burden and spatial distribution for IRF, SRF, and PED.

  • •

    Fluid volume: Total 3-D volume of each fluid type.

  • •

    En-face fluid area: 3-D fluid labels were projected onto the en-face plane to quantify the lateral area of fluid involvement.

  • •

    En-face fluid extent: Maximum horizontal (lateral) and vertical (inter–B-scan) extents were calculated from the en-face projections to characterize the spatial spread of fluid.

  • •

    En-face largest area: En-face area metrics were additionally refined for the largest connected component to emphasize clinically dominant 3-D fluid regions and reduce sensitivity to small isolated areas.

  • •

    Max extent of the largest area: The maximum linear extent within the en-face projection of the largest connected component was calculated automatically as a compact descriptor.

Visualization outputs presenting all quantitative fluid indices were generated to support qualitative assessment and clinical interpretation. As illustrated in Fig. 4, the automated 3-D fluid segmentation results are summarized using representative B-scans, en-face projections, and geometric measurements for each fluid type. These visualizations provide an intuitive overview of fluid burden and its spatial distribution, and enable direct inspection of the correspondence between volumetric indices and their anatomical manifestations.

Fig. 4.

Fig. 4.

Two representative visualizations (a,b) of automated 3-D fluid segmentation results and derived quantitative indices for intraretinal fluid (IRF), subretinal fluid (SRF), and pigment epithelial detachment (PED). Top rows: Representative B-scans with color-coded fluid overlays (each showing one individual type of fluid) and corresponding total 3-D fluid volumes. Middle rows: En-face projections of each fluid type overlaid on the retinal en-face image, with corresponding en-face fluid area measurements. Bottom rows: En-face projections of the largest connected fluid component, with the maximum linear extent indicated by the yellow lines.

4. Results

Retinal fluid was present in 290 of 291 training OCT volumes and in all 50 testing OCT volumes. In the training dataset, PED was the most frequent fluid compartment, present in 248 volumes, followed by SRF in 195 and IRF in 178; coexisting IRF, SRF, and PED was the most common pattern, observed in 96 volumes, followed by SRF + PED in 82 and IRF + PED in 42. The median total labeled fluid burden was 54,411 voxels per positive training volume, with median compartment-specific burdens of 9,963 voxels for IRF, 9,683 for SRF, and 32,460 for PED. In the testing dataset, PED was present in 48 volumes, SRF in 39, and IRF in 32; coexisting IRF, SRF, and PED was again the most common pattern, observed in 21 volumes, followed by SRF + PED in 17 and IRF + PED in 10. The median total labeled fluid burden was 56,305.5 voxels, with median compartment-specific burdens of 4,413 voxels for IRF, 15,113 for SRF, and 32,109 for PED.

As shown in Table 2, across all models, automated segmentation (Auto) achieved high agreement with expert annotations for IRF, SRF, and PED. Dice scores for Auto vs. M1 consistently exceeded 88% for IRF and SRF and exceeded 80% for PED. When compared against expert consensus, Dice values were further increased, reaching 90% or higher for IRF across all models, indicating strong agreement in regions of expert concurrence.

Table 2. Dice similarity coefficients (mean ± SD) for U-Net models. Performance is reported against expert annotations (M1, M2), expert consensus, and inter-observer agreement (M1 vs. M2).

Model Fluid Auto vs. M1 Auto vs. M2 Auto vs. consensus M1 vs. M2
n=50 n=20 n=20 n=20
2-D U-Net IRF 88.0±12.4 85.9±17.3 90.0±12.0 87.5±16.6
SRF 87.6±11.0 84.4±14.4 83.5±14.1 82.7±14.5
PED 81.4±13.4 76.6±13.3 80.3±13.6 80.4±11.7

3-D U-Net IRF 88.6±11.2 87.5±16.3 91.4±10.2 87.5±16.6
SRF 87.7±9.9 85.1±13.1 84.2±12.5 82.7±14.5
PED 81.2±11.5 78.2±11.8 81.4±11.8 80.4±11.7

IRF 89.5±11.7 87.0±17.5 91.4±11.0 87.5±16.6
Ensemble SRF 88.0±10.7 84.6±14.0 83.9±13.4 82.7±14.5
(2-D+3-D) PED 82.0±13.0 77.4±12.9 80.8±13.0 80.4±11.7

The 3-D U-Net achieved the highest overall performance for most comparisons, particularly for Auto vs. M2 and Auto vs. consensus, reflecting the benefit of incorporating volumetric context. The ensemble model combining 2-D and 3-D predictions also demonstrated improved performance, with the best Dice scores observed for IRF (89.5%), SRF (88.0%), and PED (82.0%) when compared against M1.

Inter-observer variability (M1 vs. M2) yielded Dice values of 87.5% for IRF, 82.7% for SRF, and 80.4% for PED. Notably, automated segmentation performance across all models fell within or exceeded this range, demonstrating expert-level accuracy. Segmentation accuracy was lower for PED than for IRF and SRF, which is expected given the greater anatomical complexity of PED. In particular, extensive overlying fluid can obscure the lower boundary of the PED, making its delineation more challenging and contributing to increased inter-observer variability.

Besides, as shown in Fig. 5, volume correlations between automated and manual measurements were high across all 3 fluid compartments. Against M1, the highest R2 values were 0.993 for IRF, 0.976 for SRF, and 0.959 for PED. Against M2, the highest R2 values were 0.949, 0.925, and 0.997, respectively. These results were comparable to inter-observer volume agreement between M1 and M2 ( R2=0.978 , 0.845, and 0.995 for IRF, SRF, and PED, respectively), further supporting the reliability of the proposed models.

Fig. 5.

Fig. 5.

Correlation between automated and expert-derived 3-D fluid volumes for IRF, SRF, and PED. Scatter plots show automated volume estimates versus expert measurements M1 (top) and M2 (middle) for 2-D, 3-D, and ensemble models, as well as the inter-observer agreement between M1 and M2 (bottom). Lines indicate linear regression fits.

Although the 3-D U-Net performed best overall in most comparisons, the ensemble model achieved the highest Dice scores for several Auto vs. M1 evaluations. However, the differences among models were not statistically significant, so the variation in ranking likely reflects annotations variability rather than a fundamental inconsistency in model performance. The 3-D model uses volumetric context from neighboring B-scans, which can improve spatial consistency and may better match M2 and the consensus masks. In some cases, however, the ensemble may better capture slice-level details and therefore align more closely with M1.

Moreover, comparable models without a residual encoder were also evaluated. Consistent improvements were observed with the residual encoder architecture, particularly for the 3-D models, where Dice scores increased by approximately 1–2%. These gains are consistent with our hypothesis that the residual encoder helps the network learn more effectively by retaining both fine structural details and broader contextual information. This design enables more accurate segmentation of diverse fluid patterns while preserving the anatomical continuity of retinal layers.

Figure 6 illustrates representative qualitative examples of automated fluid segmentation across a range of nvAMD presentations. The examples demonstrate accurate delineation of IRF, SRF, and PED under diverse anatomical configurations and disease severities. The method successfully captures small, focal fluid pockets as well as extensive, multilobulated lesions, including cases with coexisting fluid compartments and disrupted retinal architecture. Notably, segmentation remains anatomically consistent in challenging scenarios such as steep PEDs, shadowing, and heterogeneous reflectivity, which are commonly encountered in routine clinical imaging. These examples highlight the robustness of the proposed framework in handling complex real-world OCT scans and visually support the quantitative performance reported above.

Fig. 6.

Fig. 6.

Representative qualitative examples of retinal fluid segmentation in nvAMD. For each case, the original OCT B-scan is shown alongside the manual consensus annotation from two experts and the corresponding predictions from the 2-D and 3-D models. Segmentation overlays indicate intraretinal fluid (IRF, red), subretinal fluid (SRF, green), and pigment epithelial detachment (PED, blue). Examples span a range of disease severities and anatomical patterns, including focal and extensive fluid, coexisting fluid compartments, and complex retinal morphology.

To further assess model robustness in the presence of vitreoretinal interface abnormalities, we performed a qualitative evaluation using representative OCT scans with epiretinal membrane (ERM) or posterior vitreous detachment (PVD). The trained segmentation framework was applied without additional fine-tuning. Representative examples of model performance in scans with ERM and PVD are shown in Fig. 7. Overall, the model produced anatomically plausible fluid segmentations in these challenging scans, localizing IRF, SRF, and PED while preserving the expected spatial relationships between fluid compartments and retinal anatomy.

Fig. 7.

Fig. 7.

Representative qualitative examples of fluid segmentation in OCT scans with epiretinal membrane (ERM) and posterior vitreous detachment (PVD). Each example is shown as the original B-scan and the corresponding segmentation overlay. Despite vitreoretinal interface abnormalities and distorted retinal morphology, the model produced anatomically plausible segmentations of IRF, SRF, and PED.

To qualitatively demonstrate the clinical utility of the proposed framework, PED was used as an example to illustrate how the method captures longitudinal fluid dynamics at both the individual and population levels.

Figure 8 presents representative longitudinal fluid trajectories for 3 subjects, summarizing changes in PED 3-D volume, en-face area, and maximum en-face extent of the largest PED across multiple treatment phases. These examples illustrate distinct patterns of PED response over time, including regression, persistence, and fluctuation, demonstrating the ability of the proposed framework to capture clinically meaningful longitudinal dynamics in real-world follow-up data.

Fig. 8.

Fig. 8.

Longitudinal quantitative PED characteristics for representative nvAMD subjects. Time-series plots show PED volume, en-face area, and maximum en-face extent of the largest PED across clinical follow-up, aligned by days after first anti-VEGF injection. Different colors indicate treatment phases, illustrating heterogeneous PED responses over time among 3 representative subjects.

Figure 9 shows population-level changes in PED volume between the first and final anti-VEGF injections. The distribution demonstrates that the proposed method can consistently quantify PED volume changes across a large and heterogeneous clinical dataset, capturing both minimal change and substantial progression or regression.

Fig. 9.

Fig. 9.

Distribution of changes in PED volume from the first to the final anti-VEGF injection. The histogram shows the frequency of PED volume change across subjects, with negative values indicating volume reduction and positive values indicating volume growth.

Overall, these longitudinal analyses show that the proposed automated approach supports consistent, subject-level tracking of fluid changes over extended follow-up. By jointly summarizing volume, en-face area, and spatial extent, the framework provides a comprehensive description of fluid evolution that is well suited for longitudinal assessment of treatment response and disease progression.

Runtime performance was evaluated on 100 randomly selected OCT volumes comprising 2068 total B-scans, with a mean of 20.68 B-scans per volume and a range of 19 to 61 B-scans per volume. The complete inference workflow processed all 100 scans successfully in 1168.68 seconds, corresponding to 11.69 seconds per OCT volume and a throughput of 308.04 volumes per hour. The nnU-Net inference steps required 1040.69 seconds in total, averaging 10.41 seconds per OCT volume. Specifically, pre-segmentation of the retina required 3.62 seconds per volume, whereas final segmentation of fluids, including IRF, SRF, and PED, required 6.78 seconds per volume. Experiments were performed on an NVIDIA A100-SXM4-40GB GPU with CUDA 12.2, supporting the feasibility of applying the proposed framework to large-scale longitudinal clinical OCT datasets.

5. Discussion

We present a robust and scalable framework for automated 3-D segmentation of retinal fluid in nvAMD, tailored specifically for real-world longitudinal OCT data. A core strength of this approach lies in its ability to handle the substantial heterogeneity that is typical for routine clinical imaging, including different scan protocols at each visit, and variable inter–B-scan continuity during long-term follow-up.

The framework integrates anatomically constrained preprocessing with nnU-Net–based segmentation and adaptive inference strategies, delivering consistent performance across diverse imaging conditions while remaining computationally feasible for large-scale clinical deployment. A pivotal component is the incorporation of Deep LOGISMOS for retinal-region extraction from our prior work [44], which requires no additional fine-tuning, and enforces anatomically meaningful surface constraints and 3-D smoothness. This preprocessing step markedly reduces sensitivity to region-level errors from deep learning models trained on heterogeneous data — particularly in severe disease cases where extensive fluid, shadowing, and disrupted retinal architecture can confound purely data-driven region segmentation.

Subsequent segmentation employs a hybrid of 2-D and 3-D nnU-Net models to balance robustness and volumetric contextual awareness: 3-D models exploit inter-slice continuity when reliable, while 2-D models ensure stable performance under limited or inconsistent B-scan alignment. An iterative human-in-the-loop labeling strategy further enhances reliability by prioritizing scans with extensive or complex fluid accumulations. This dynamic process progressively reduces segmentation errors in challenging clinical cases, achieving expert-level performance that closely approximates inter-observer variability. Validation against independent expert annotations, combined with large-scale longitudinal application, underscores the framework’s emphasis on clinical reliability over real-world datasets.

Furthermore, this segmentation approach allows reliable segmentation of multiple fluid compartments, including IRF, SRF, and PED. This is particularly notable as most previously published algorithms focus on a specific type of fluid, which emphasizes the overall robustness of our algorithm. Additionally, the iterative labeling process proved essential for resolving challenging anatomical features, as demonstrated in our qualitative results. For instance, the volumetric quantification of PED is often complicated by indistinct, low-contrast lower boundaries and the presence of extensive overlying fluid that obscures the true margins. By selectively incorporating expert-guided corrections on these difficult cases, the iterative approach effectively stabilized the models and yielded more anatomically consistent boundary delineations.

Several limitations warrant attention. First, the framework does not explicitly model subretinal hyperreflective material (SHRM) or drusen as separate pathological features. The presence of SHRM can create substantial ambiguity on OCT, particularly around PED-related boundaries, which makes consistent annotation and automated segmentation more difficult. Similarly, large drusen or drusenoid RPE elevations may exhibit imaging features that overlap with shallow PED, potentially leading to ambiguity in PED delineation. This may constrain analyses of SHRM- or drusen-specific biomarkers. Second, vitreoretinal interface abnormalities such as ERM and PVD were not explicitly labeled or quantitatively analyzed. Although qualitative examples showed anatomically plausible segmentation in representative ERM and PVD cases, severe cases may distort retinal contour or introduce hyperreflective structures that complicate automated segmentation. Future work will include dedicated subgroup analyses in eyes with ERM, PVD, and other coexisting ocular pathologies. Third, while comprehensive volumetric and en-face fluid metrics are provided, quantifications span the entire macular scan rather than being anchored to the foveal center. Future iterations will incorporate fovea localization derived from layered retinal segmentation to enable more precise, focal analyses that better correlate with visual function and inform clinical decision-making. Fourth, while the proposed model demonstrates robust performance on a large, real-world clinical dataset from our institution, the longitudinal evaluation was predominantly based on scans acquired using the Heidelberg Spectralis system. Although our training procedure incorporated data from multiple device vendors via the RETOUCH dataset, further validation on independent external cohorts across a wider range of imaging systems and diverse patient populations will be valuable to comprehensively confirm its broad generalizability.

6. Conclusion

We present a clinically oriented, technically robust framework for automated 3-D fluid segmentation and longitudinal analysis in nvAMD. By combining anatomically constrained preprocessing, adaptive 2-D and 3-D deep learning models, and iterative expert-guided labeling, the proposed approach achieves expert-level accuracy while remaining resilient to the heterogeneity of real-world OCT data. Validation against inter-observer variability and application to large longitudinal cohorts demonstrate its suitability for routine clinical imaging and large-scale research studies.

Funding

National Eye Institute https://ror.org/03wkg3b53 ( NEI R01-EY035435).

Disclosures

Z. Chen, None; B. Bach, None; H. Zhang, None; A. Wahle, None; Ian C. Han, None; H. C. Boldt, None; Stephen R. Russell, None; M. Sonka, University of Iowa (Patents), Medical Imaging Applications, LLC (Co-Founder), VIDA Diagnostics, Inc. (Co-Founder); E. Sohn, None.

Data Availability

The data used to generate the results in this study are not available to the public at this time.

References

  • 1.Bourne R., Steinmetz J. D., Flaxman S., et al. , “Trends in prevalence of blindness and distance and near vision impairment over 30 years: an analysis for the global burden of disease study,” The Lancet Global Health 9(2), e130–e143 (2021). 10.1016/S2214-109X(20)30425-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Congdon N., O’Colmain B., Klaver C., et al. , “Causes and prevalence of visual impairment among adults in the United States,” Arch. Ophthalmol. 122(4), 477–485 (2004). 10.1001/archopht.122.4.477 [DOI] [PubMed] [Google Scholar]
  • 3.Wong W. L., Su X., Li X., et al. , “Global prevalence of age-related macular degeneration and disease burden projection for 2020 and 2040: a systematic review and meta-analysis,” The Lancet Global Health 2(2), e106–e116 (2014). 10.1016/S2214-109X(13)70145-1 [DOI] [PubMed] [Google Scholar]
  • 4.Ferris III F. L., Wilkinson C., Bird A., et al. , “Clinical classification of age-related macular degeneration,” Ophthalmology 120(4), 844–851 (2013). 10.1016/j.ophtha.2012.10.036 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Heier J. S., Lad E. M., Holz F. G., et al. , “Pegcetacoplan for the treatment of geographic atrophy secondary to age-related macular degeneration (OAKS and DERBY): two multicentre, randomised, double-masked, sham-controlled, phase 3 trials,” The Lancet 402(10411), 1434–1448 (2023). 10.1016/S0140-6736(23)01520-9 [DOI] [PubMed] [Google Scholar]
  • 6.Bakall B., Folk J. C., Boldt H. C., et al. , “Aflibercept therapy for exudative age-related macular degeneration resistant to bevacizumab and ranibizumab,” Am. J. Ophthalmol. 156(1), 15–22.e1 (2013). 10.1016/j.ajo.2013.02.017 [DOI] [PubMed] [Google Scholar]
  • 7.Bressler N. M., Doan Q. V., Varma R., et al. , “Estimated cases of legal blindness and visual impairment avoided using ranibizumab for choroidal neovascularization: non-Hispanic white population in the United States with age-related macular degeneration,” Arch. Ophthalmol. 129(6), 709–717 (2011). 10.1001/archophthalmol.2011.140 [DOI] [PubMed] [Google Scholar]
  • 8.Reyes A. G., Shah T. J., Linton E. F., et al. , “Real-world short-term outcomes after switching to fixed-interval anti-VEGF therapy for neovascular AMD during the COVID-19 pandemic,” Int. J. Retin. Vitr. 11(1), 103 (2025). 10.1186/s40942-025-00734-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Elsharkawy M., Elrazzaz M., Ghazal M., et al. , “Role of optical coherence tomography imaging in predicting progression of age-related macular disease: A survey,” Diagnostics 11(12), 2313 (2021). 10.3390/diagnostics11122313 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Gualino V., Tadayoni R., Cohen S. Y., et al. , “Optical coherence tomography, fluorescein angiography, and diagnosis of choroidal neovascularization in age-related macular degeneration,” Retina 39(9), 1664–1671 (2019). 10.1097/IAE.0000000000002220 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Schmidt-Erfurth U., Klimscha S., Waldstein S., et al. , “A view of the current and future role of optical coherence tomography in the management of age-related macular degeneration,” Eye 31(1), 26–44 (2017). 10.1038/eye.2016.227 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Metrangolo C., Donati S., Mazzola M., et al. , “OCT biomarkers in neovascular age-related macular degeneration: a narrative review,” J. Ophthalmol. 2021(1), 9994098 (2021). 10.1155/2021/9994098 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Rispoli M., Cennamo G., Di Antonio L., et al. , “Practical guidance for imaging biomarkers in exudative age-related macular degeneration,” Surv. Ophthalmol. 68(4), 615–627 (2023). 10.1016/j.survophthal.2023.02.004 [DOI] [PubMed] [Google Scholar]
  • 14.Schmidt-Erfurth U., Bogunovic H., Sadeghipour A., et al. , “Machine learning to analyze the prognostic value of current imaging biomarkers in neovascular age-related macular degeneration,” Ophthalmology Retina 2(1), 24–30 (2018). 10.1016/j.oret.2017.03.015 [DOI] [PubMed] [Google Scholar]
  • 15.Pokroy R., Mimouni M., Barayev E., et al. , “Prognostic value of subretinal hyperreflective material in neovascular age-related macular degeneration treated with bevacizumab,” Retina 38(8), 1485–1491 (2018). 10.1097/IAE.0000000000001748 [DOI] [PubMed] [Google Scholar]
  • 16.Waldstein S. M., Wright J., Warburton J., et al. , “Predictive value of retinal morphology for visual acuity outcomes of different ranibizumab treatment regimens for neovascular AMD,” Ophthalmology 123(1), 60–69 (2016). 10.1016/j.ophtha.2015.09.013 [DOI] [PubMed] [Google Scholar]
  • 17.Zhang X., Lai T. Y., “Baseline predictors of visual acuity outcome in patients with wet age-related macular degeneration,” BioMed Res. Int. 2018, 9640131 (2018). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Lai T.-T., Hsieh Y.-T., Yang C.-M., et al. , “Biomarkers of optical coherence tomography in evaluating the treatment outcomes of neovascular age-related macular degeneration: a real-world study,” Sci. Rep. 9(1), 529 (2019). 10.1038/s41598-018-36704-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Waldstein S. M., Philip A.-M., Leitner R., et al. , “Correlation of 3-dimensionally quantified intraretinal and subretinal fluid with visual acuity in neovascular age-related macular degeneration,” JAMA Ophthalmol. 134(2), 182–190 (2016). 10.1001/jamaophthalmol.2015.4948 [DOI] [PubMed] [Google Scholar]
  • 20.Jaffe G. J., Ying G.-S., Toth C. A., et al. , “Macular morphology and visual acuity in year five of the comparison of age-related macular degeneration treatments trials,” Ophthalmology 126(2), 252–260 (2019). 10.1016/j.ophtha.2018.08.035 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Segal O., Barayev E., Nemet A. Y., et al. , “Predicting response of exudative age-related macular degeneration to bevacizumab based on spectralis optical coherence tomography,” Retina 36(2), 259–263 (2016). 10.1097/IAE.0000000000000690 [DOI] [PubMed] [Google Scholar]
  • 22.Kang E. C., Choi S., Koh H. J., “Inner nuclear layer cystoid spaces are a poor prognostic factor in typical age-related macular degeneration and polypoidal choroidal vasculopathy,” Graefe’s Arch. Clin. Exp. Ophthalmol. 255(11), 2157–2163 (2017). 10.1007/s00417-017-3776-9 [DOI] [PubMed] [Google Scholar]
  • 23.Riedl S., Vogl W.-D., Waldstein S. M., et al. , “Impact of intra-and subretinal fluid on vision based on volume quantification in the HARBOR trial,” Ophthalmology Retina 6(4), 291–297 (2022). 10.1016/j.oret.2021.12.007 [DOI] [PubMed] [Google Scholar]
  • 24.Chakravarthy U., Pillai N., Syntosi A., et al. , “Association between visual acuity, lesion activity markers and retreatment decisions in neovascular age-related macular degeneration,” Eye 34(12), 2249–2256 (2020). 10.1038/s41433-020-0799-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Waldstein S. M., Simader C., Staurenghi G., et al. , “Morphology and visual acuity in aflibercept and ranibizumab therapy for neovascular age-related macular degeneration in the VIEW trials,” Ophthalmology 123(7), 1521–1529 (2016). 10.1016/j.ophtha.2016.03.037 [DOI] [PubMed] [Google Scholar]
  • 26.Schmidt-Erfurth U., Vogl W.-D., Jampol L. M., et al. , “Application of automated quantification of fluid volumes to anti-VEGF therapy of neovascular age-related macular degeneration,” Ophthalmology 127(9), 1211–1219 (2020). 10.1016/j.ophtha.2020.03.010 [DOI] [PubMed] [Google Scholar]
  • 27.Kim J. H., Chang Y. S., Kim J. W., “Natural course of patients discontinuing treatment for age-related macular degeneration and factors associated with visual prognosis,” Retina 37(12), 2254–2261 (2017). 10.1097/IAE.0000000000001494 [DOI] [PubMed] [Google Scholar]
  • 28.Chatziralli I., Nicholson L., Vrizidou E., et al. , “Predictors of outcome in patients with neovascular age-related macular degeneration switched from ranibizumab to 8-weekly aflibercept,” Ophthalmology 123(8), 1762–1770 (2016). 10.1016/j.ophtha.2016.05.002 [DOI] [PubMed] [Google Scholar]
  • 29.Ashraf M., Souka A., Adelman R. A., “Age-related macular degeneration: using morphological predictors to modify current treatment protocols,” Acta Ophthalmol. 96(2), 120–133 (2018). 10.1111/aos.13565 [DOI] [PubMed] [Google Scholar]
  • 30.Orlin A., Hadley D., Chang W., et al. , “Association between high-risk disease loci and response to anti–vascular endothelial growth factor treatment for wet age-related macular degeneration,” Retina 32(1), 4–9 (2012). 10.1097/IAE.0b013e31822a2c7c [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Sadda S., Holekamp N. M., Sarraf D., et al. , “Relationship between retinal fluid characteristics and vision in neovascular age-related macular degeneration: HARBOR post hoc analysis,” Graefe’s Arch. Clin. Exp. Ophthalmol. 260(12), 3781–3789 (2022). 10.1007/s00417-022-05716-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Tuerksever C., Pruente C., Hatz K., “High frequency SD-OCT follow-up leading to up to biweekly intravitreal ranibizumab treatment in neovascular age-related macular degeneration,” Sci. Rep. 11(1), 6816 (2021). 10.1038/s41598-021-86348-2 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Azar G., Wolff B., De Bats F., et al. , “Morphological predictive features on spectral-domain optical coherence tomography for visual outcomes in neovascular age-related macular degeneration treated with ranibizumab,” BioMed Res. Int. 2018, 1–9 (2018). 10.1155/2018/7438083 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Clemens C. R., Wolf A., Alten F., et al. , “Response of vascular pigment epithelium detachment due to age-related macular degeneration to monthly treatment with ranibizumab: the prospective, multicentre RECOVER study,” Acta Ophthalmol. 95(7), 683–689 (2017). 10.1111/aos.13359 [DOI] [PubMed] [Google Scholar]
  • 35.Cho H. J., Kim K. M., Kim H. S., et al. , “Response of pigment epithelial detachment to anti–vascular endothelial growth factor treatment in age-related macular degeneration,” Am. J. Ophthalmol. 166, 112–119 (2016). 10.1016/j.ajo.2016.03.039 [DOI] [PubMed] [Google Scholar]
  • 36.Wang J., Zhang M., Pechauer A. D., et al. , “Automated volumetric segmentation of retinal fluid on optical coherence tomography,” Biomed. Opt. Express 7(4), 1577–1589 (2016). 10.1364/BOE.7.001577 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Schlegl T., Waldstein S. M., Bogunovic H., et al. , “Fully automated detection and quantification of macular fluid in OCT using deep learning,” Ophthalmology 125(4), 549–558 (2018). 10.1016/j.ophtha.2017.10.031 [DOI] [PubMed] [Google Scholar]
  • 38.Sappa L. B., Okuwobi I. P., Li M., et al. , “Retfluidnet: Retinal fluid segmentation for sd-oct images using convolutional neural network,” J Digit Imaging 34(3), 691–704 (2021). 10.1007/s10278-021-00459-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Guo Y., Hormel T. T., Xiong H., et al. , “Automated segmentation of retinal fluid volumes from structural and angiographic optical coherence tomography using deep learning,” Translational vision science technology 9(2), 54 (2020). 10.1167/tvst.9.2.54 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Hsu H.-Y., Chou Y.-B., Jheng Y.-C., et al. , “Automatic segmentation of retinal fluid and photoreceptor layer from optical coherence tomography images of diabetic macular edema patients using deep learning and associations with visual acuity,” Biomedicines 10(6), 1269 (2022). 10.3390/biomedicines10061269 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Mantel I., Mosinska A., Bergin C., et al. , “Automated quantification of pathological fluids in neovascular age-related macular degeneration, and its repeatability using deep learning,” Translational Vision Science Technology 10(4), 17 (2021). 10.1167/tvst.10.4.17 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Bogunović H., Venhuizen F., Klimscha S., et al. , “Retouch: The retinal OCT fluid detection and segmentation benchmark and challenge,” IEEE Transactions on Medical Imaging 38(8), 1858–1874 (2019). 10.1109/TMI.2019.2901398 [DOI] [PubMed] [Google Scholar]
  • 43.Isensee F., Jaeger P. F., Kohl S. A., et al. , “nnU-Net: a self-configuring method for deep learning-based biomedical image segmentation,” Nat. Methods 18(2), 203–211 (2021). 10.1038/s41592-020-01008-z [DOI] [PubMed] [Google Scholar]
  • 44.Chen Z., Zhang H., Linton E. F., et al. , “Hybrid deep learning and optimal graph search method for optical coherence tomography layer segmentation in diseases affecting the optic nerve,” Biomed. Opt. Express 15(6), 3681–3698 (2024). 10.1364/BOE.516045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Garvin M. K., Abramoff M. D., Wu X., et al. , “Automated 3-D intraretinal layer segmentation of macular spectral-domain optical coherence tomography images,” IEEE Transactions on Medical Imaging 28(9), 1436–1447 (2009). 10.1109/TMI.2009.2016958 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Isensee F., Wald T., Ulrich C., et al. , “nnU-Net revisited: A call for rigorous validation in 3D medical image segmentation,” in International Conference on Medical Image Computing and Computer-Assisted Intervention, (Springer, 2024), pp. 488–498. [Google Scholar]
  • 47.He K., Zhang X., Ren S., et al. , “Deep residual learning for image recognition,” in Proceedings of the IEEE conference on computer vision and pattern recognition, (2016), pp. 770–778. [Google Scholar]
  • 48.Liu F. T., Ting K. M., Zhou Z.-H., “Isolation forest,” in 2008 eighth IEEE international conference on data mining, (IEEE, 2008), pp. 413–422. [Google Scholar]
  • 49.Wang Z., Bovik A. C., Sheikh H. R., et al. , “Image quality assessment: from error visibility to structural similarity,” IEEE Transactions on Image Processing 13(4), 600–612 (2004). 10.1109/TIP.2003.819861 [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data used to generate the results in this study are not available to the public at this time.


Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group