Skip to main content
Biomedical Optics Express logoLink to Biomedical Optics Express
. 2026 Aug 26;17(9):4854–4869. doi: 10.1364/BOE.607950

Optimal excitation polarization schemes to boost the precision of single-molecule orientation imaging

Kaizhi A Nie 1, Yuanxin Qiu 1, Matthew D Lew 1,*
PMCID: PMC13592251  PMID: 42770015

Abstract

Modulating excitation polarization across camera frames can improve the precision of single-molecule orientation-localization microscopy (SMOLM), but principled methods for designing these excitation sequences remain underdeveloped. Here, we introduce a Fisher-information framework for optimizing linearly and circularly polarized excitation schemes under a fixed total illumination photon budget. For the multi-view reflector microscope, 3000 illumination photons, and 7.5 detected background photons per pixel, an optimized four-frame linear scheme improves median orientation measurement precision by 21.2% relative to the conventional six-frame equal-pumping scheme and achieves Cramér–Rao-bound-limited median precisions of σθ,50=2.59∘ and σϕ,50=3.22∘ . Increasing to six optimized linear frames provides only modest additional improvement, demonstrating that four frames offer an effective balance between precision, uniformity across molecular orientations, and temporal resolution. With only two frames, circularly polarized excitation yields more uniform performance and 27% better median precision than two-frame linear excitation, whereas linear schemes perform better with four or more frames. Applying the framework to four engineered dipole-spread functions shows that excitation sequences should be tailored to the detection system. These results provide practical strategies for balancing orientation precision, photon budget, and acquisition speed in SMOLM.

1. Introduction

For decades, polarized fluorescence microscopy [1] has been utilized to measure the orientations of dipole-like emitters, e.g., fluorescent molecules [2,3]. Combining polarization sensing with single-molecule localization microscopy, termed single-molecule orientation-localization microscopy (SMOLM) [4,5], has shed light on a variety of biological structures and processes at the nanoscale, from the organization of the cytoskeleton [6,7] and cellular membranes [8] to the architectures of self-assembling peptide fibers [9,10] and proteins at the interfaces of biomolecular condensates [11]. Polarization-resolved single-molecule imaging has also proven essential for mapping the local field near the focus of a high numerical aperture objective lens [12–14].

Since only hundreds to thousands of photons are typically collected from single fluorophores before they irreversibly photobleach, numerous approaches have been developed [3] to measure molecular orientations more precisely and robustly in the presence of severe shot noise [15]. Most of these techniques use phase- and/or polarization-modulating optical elements to create new dipole-spread functions (DSFs) [16], i.e., the image of a dipole emitter created by a microscope analogous to the point-spread function for isotropically emitting point sources. On the other hand, many studies have used multiple images of molecules excited by light of various polarization states to measure their orientations [17–19]. These techniques leverage the properties that 1) the probability Pa of a molecule to absorb light depends on the relative alignment between its absorption transition dipole moment μa and the excitation electric field Eex as Pa∝|Eex⋅μa|2 , and 2) the number of fluorescence photons emitted by a molecule is linearly proportional to Pa below saturation [20]. Thus, if one is not careful to vary Eex during an experiment, then only a subset of molecules may be excited, and selection bias can occur [21,22]. Importantly, the additional contrast yielded by varying excitation polarization across multiple camera images can boost the precision of measuring molecular orientations by an order of magnitude [23]. In addition, modulating both illumination polarization and the detection DSF enables one to distinguish a single wobbling molecule from a pair of fixed molecules [24], which is impossible to accomplish using either approach alone. However, optimal strategies for designing illumination polarizations, especially in the context of specific DSFs, remain unexplored.

Here, we develop a framework for designing optimal excitation polarization schemes for SMOLM using Fisher information. We design linear and circular polarization schemes for the radially and azimuthally polarized multi-view reflector (MVR) microscope [8], quantify performance for each design, and compare how performance varies with the number of excitations. We further extend the optimization to four engineered DSFs, double helix [25,26], pixOL [27], vortex [28,29], and CHIDO [30], and show that tailoring the excitation sequence to each DSF consistently improves orientation estimation performance, particularly for orientations that are poorly measured by DSF engineering alone.

2. Designing optimal illumination polarizations

2.1. Forward model

We model a fluorescent molecule’s orientation using a real-valued unit vector μ=[μx,μy,μz]⊺ that represents its transition dipole moment (Fig. 1(a) inset), where ⊺ represents a matrix/vector transpose. To excite fluorescence from the molecule, it is illuminated by pumping photons with a polarization state described by Eex=[Ex,Ey,Ez]⊺ ; we assume this state is constant over the detector integration time. As shown previously [2,3,19,23], the probability Pabs of a molecule to absorb an incoming polarized photon is given by Pabs∝cT∫0T|Eex⊺μ(t)|2dt , where c is a constant that accounts for the molecule’s absorption cross-section and quantum yield and T is the detector integration time. This time average yields an absorption probability that is proportional to the second-order moments m of the molecule’s transition dipole, such that mij=1T∫0Tμi(t)μj(t)dt and i,j∈{x,y,z} . We therefore define these so-called second moments as

m=[mxxmyymzzmxymxzmyz]⊺. (1)

While the precise trajectory μ(t) is arbitrary, we assume for simplicity that the molecule “wobbles” uniformly within a cone oriented along μ¯=[μ¯x,μ¯y,μ¯z]⊺=[sin⁡θcos⁡ϕ,sin⁡θsin⁡ϕ,cos⁡θ]⊺ , such that [19,24,31–33]

m=γ[μ¯x2μ¯y2μ¯z2μ¯xμ¯yμ¯xμ¯zμ¯yμ¯z]⊺+(1−γ)[131313000]⊺, (2)

where (θ,ϕ) specifies the mean dipole orientation in spherical coordinates. The parameter γ∈[0,1] describes the degree of rotational confinement [31–33], with γ=1 corresponding to a fixed dipole and γ=0 corresponding to a complete rotational average (i.e., an isotropic emitter). This confinement parameter γ is directly related to the surface area Ω of the cone via γ=1−3Ω/(4π)+Ω2/(8π2) [2,31,34].

Fig. 1.

Fig. 1.

Excitation modulation and engineered dipole-spread functions in single-molecule orientation-localization microscopy (SMOLM). (a) Optical system schematic. A laser beam passes through an electro-optic modulator (EOM) and a quarter-wave plate (QWP) to facilitate real-time control of excitation polarization, which is focused by L1 onto the pupil plane of an objective lens (OL) to create widefield illumination in the sample. Fluorescence is collected by the same objective, separated from the excitation by a dichroic mirror (DM), and split by a vortex waveplate (VWP) and polarizing beam splitter (PBS) into azimuthally and radially polarized channels. The OL pupil is imaged by TL and L2 onto the multi-view reflector (MVR) mirrors (MVM1 and MVM2), which split the fluorescence further into four separate beams. Lenses L3 and L4 finally create four images on non-overlapping regions of each camera excited using frame 4 of the optimized scheme (red circle surrounding the Pabs,ℓ map in (b.ii)). The molecule’s transition dipole moment μ (blue arrow) with its associated polar angle θ and azimuthal angle ϕ is shown above OL. Middle inset: Excitation electric field Eℓ for frame ℓ (yellow arrow), parameterized by angles αℓ and βℓ . Right inset: (top) azimuthally and (bottom) radially polarized MVR images of a test dipole with (θ=45∘,ϕ=0) . (b) Absorption probability Pabs,ℓ(θ,ϕ) for each pumping modulation Eℓ of the (i) L=6 equal-pumping (EP) scheme and (ii) L=6 optimized excitation scheme, respectively. Red dots mark the orientation of a test dipole at (θ=45∘,ϕ=0) . (c) Normalized Fisher information matrices (FIM, Eq. (11)) corresponding to (b) for a test dipole at (θ=45∘,ϕ=0) . Rows and columns correspond to the six orientational second moments {mxx,myy,mzz,mxy,mxz,myz} (Eq. (2)). Colorbars: (a, inset right) MVR image intensity (normalized), (b) normalized absorption probability Pabs,ℓ(θ,ϕ) , (c) Fisher information, normalized to the maximum entry of the EP scheme (c.i). Scalebar: (a, inset right) 500 nm.

Assuming that the illumination intensity is well below saturation, the expected number s of fluorescence photons detected in frame ℓ of an experiment may be written as

sℓ=aEℓ⊺m, (3)

where we have introduced the second-order moments of the excitation field as

Eℓ⊺=[|Ex,ℓ|2|Ey,ℓ|2|Ez,ℓ|22Re(Ex,ℓ∗Ey,ℓ)2Re(Ex,ℓ∗Ez,ℓ)2Re(Ey,ℓ∗Ez,ℓ)] (4)

and a brightness-scaling constant a , and ∗ represents complex conjugation. To increase orientation sensitivity, we choose to sequentially excite the molecule using L distinct excitation polarization states, one per camera frame, such that

s=aE⊺m, (5)

where s∈RL represents the photons detected from the molecule across the L frames and E∈RL×6 represents a sequence of L excitation second moments. To compare fairly schemes with different numbers L of modulation frames, we fix the total number of illumination photons to Npump=3000 , i.e., the total illumination photon budget is held constant across all schemes. Thus, the per-frame excitation photon budget is given by a=Npump/L .

To facilitate implementation, we parameterize any sequence of linearly polarized fields using two angles (βℓ,αℓ) in spherical coordinates per modulation, such that

Eℓ=[sin⁡βℓcos⁡αℓsin⁡βℓsin⁡αℓcos⁡βℓ]⊺. (6)

The two angles define the directions of the linearly polarized electric field, and the corresponding excitation second moments are then obtained via Eq. (4). In practice, these fields may be modulated by using an electro-optic modulator and quarter waveplate, as shown in Fig. 1(a).

To model circularly polarized excitation, we use the angular parameters (βℓ,αℓ) to define the propagation direction of the excitation beam, which is perpendicular to the electric field. We construct orthonormal basis vectors sˆℓ=[−sin⁡αℓ,cos⁡αℓ,0]⊺ and pˆℓ=[cos⁡αℓcos⁡βℓ,sin⁡αℓcos⁡βℓ,−sin⁡βℓ]⊺ to represent the s- and p-polarization directions, respectively. Decomposing circularly polarized light into s- and p-polarized fields of equal intensities, we may represent second-order moments of the electric field in the lab frame using

Eℓ,circ=[pˆℓ,12+sˆℓ,122pˆℓ,22+sˆℓ,222pˆℓ,32+sˆℓ,322pˆℓ,1pˆℓ,2+sˆℓ,1sˆℓ,2pˆℓ,1pˆℓ,3+sˆℓ,1sˆℓ,3pˆℓ,2pˆℓ,3+sˆℓ,2sˆℓ,3]. (7)

Finally, the expected number of fluorescence photons may be computed by substituting Eℓ,circ into Eq. (3).

2.2. Image formation

We follow closely the image formation model in [2]. We model the emission collection optics within a microscope using a set of basis images B∈RP×6 [35], where P camera pixels are used to capture images of the microscope’s (possibly engineered [3,16]) DSF. The ℓth image Iℓ∈RP (Fig. 1(a)) collected from a molecule exhibiting orientation second-moments m pumped by excitation second-moments Eℓ is given by

Iℓ=sℓBm+b, (8)

where b denotes a spatially uniform background flux. The total number of collected background photons per pixel Lb is held constant across all schemes. We depict a simplified architecture for the multi-view reflector microscope [8] in Fig. 1(a), but our mathematical framework can implement any engineered DSF [3].

Importantly, the precise sequence Eℓ of illumination polarizations will affect the total number Nabs=∑ℓsℓ of fluorescence photons detected from any given molecule, depending on its orientation m (Eq. (3)). For all performance analyses and design optimizations, we set Npump=3000 , Lb=7.5 background photons/pixel split across all channels, λ=596 nm, NA=1.5 , immersion medium refractive index n1=1.515 , sample refractive index n2=1.33 , the pixel size to 66.85 nm in object space, and the molecule’s position to lie at the glass-water interface. Thus, our model includes supercritical angle fluorescence (SAF) collected from each molecule [7,27,36,37], and our design optimizations are tailored to harness this information.

2.3. Quantifying performance

Before we can design an optimal set of excitation polarizations E , we must choose a performance metric. We elect to use Fisher information (FI) [38], which has been utilized extensively to compare the performance of various methods [2,39] and to develop optimal microscope designs for SMLM [40] and SMOLM [16,27,37,41]. FI has also been used recently to design optimal excitation patterns for MINFLUX [42]. Notably, the associated Cramér-Rao bound (CRB) provides a best-possible limit on the variance of any unbiased estimator such that Cov(Θˆ)⪰F−1 , where Θˆ represents an estimate of the parameters of interest, F is the Fisher information matrix (FIM), and ⪰ denotes a generalized inequality such that Cov(Θˆ)−F−1 is positive semidefinite. Thus, imaging systems can be optimized directly using their statistical information content, i.e., maximizing FI, thereby sidestepping the need to simultaneously design and evaluate an inverse estimation algorithm.

Assuming Poisson-distributed photon counts, the FIM for measuring orientational second moments ( Θˆ=mˆ ) of a molecule with orientation (θ,ϕ) and wobble γ is given by

Fjk(Eℓ;θ,ϕ,γ)=∑p=1P1Iℓ,p(Eℓ;θ,ϕ,γ)∂Iℓ,p(Eℓ;θ,ϕ,γ)∂mj∂Iℓ,p(Eℓ;θ,ϕ,γ)∂mk, (9)

where Iℓ,p(Eℓ;θ,ϕ,γ) denotes the expected number of photons collected in the pth camera pixel during the ℓth frame of the excitation sequence E and j,k∈{xx,yy,zz,xy,xz,yz} . Taking the partial derivative of Iℓ,p with respect to each second moment mj , we find

∂Iℓ,p∂mj=a(Bp,:m)Eℓ,j+a(Eℓ⊺m)Bp,j, (10)

where Eℓ,j denotes the excitation second moment j for frame ℓ , Bp,j denotes the pth pixel of basis image j , and j∈{xx,yy,zz,xy,xz,yz} . The first term arises from changes in the number of signal photons detected due to polarized illumination, where for each pixel p , contributions from all second-moment basis images are summed together. The second term arises from changes in the spatial distribution of fluorescence photons due to the microscope’s DSF (i.e., its basis images B ), where for each excitation modulation frame ℓ , contributions from all second moments of the illuminating electric field are summed together.

Note that Eq. (9) quantifies Fisher information for the ℓth excitation polarization state. Since Fisher information is additive across measurements, the total FIM for an entire excitation sequence E becomes

Fjk(E;θ,ϕ,γ)=∑ℓ=1LFjk(Eℓ;θ,ϕ,γ). (11)

Because the orientational second moments m depend on the mean orientation angles (θ,ϕ) and the rotational-confinement parameter γ (Eq. (2)), the FIM jointly quantifies measurement sensitivity to all orientational degrees of freedom. Performance limits are obtained from the CRB, i.e., inverting the FIM, with the diagonal elements providing lower bounds on the variances of unbiased estimates.

To summarize the performance of an imaging system for measuring “first-moment” orientation parameters, i.e., mean orientation (θ,ϕ) and rotational confinement γ , we compute the 3×3 Fisher information matrix Fθϕγ(E;θ,ϕ,γ) (see Supplement 1 (3.6MB, pdf) ). The best-possible precisions of measuring the first-moment orientation parameters are then given by

σθ=[Fθϕγ−1]11, (12)
σϕ=[Fθϕγ−1]22,and (13)
σγ=[Fθϕγ−1]33. (14)

2.4. Optimization strategy

To design optimal excitation polarization schemes, we adopt the strategy of maximizing the FIM associated with the orientational second moments (Eq. (11)). The goal is to identify a set of L excitation polarizations that yield high-precision measurements of m over all possible dipole orientations. Because FI quantifies measurement sensitivity locally within a parameter space, we densely sample dipole orientations using a Fibonacci lattice [43] of Nlattice=300 points (Eq. (16)). This sampling provides approximately uniform coverage over the hemisphere (θ∈[0,90∘],ϕ∈[0,360∘)) . One must also choose a value for rotational confinement γ (or wobble cone surface area Ω ) to properly tailor the polarization modulation scheme to the molecules of interest. Since a mean orientation (θ,ϕ) is not defined for an isotropic emitter ( γ=0 ), we choose a modest degree of wobble γ=0.9 (corresponding to a wobble cone solid angle Ω≈0.14π ) for all optimization runs, unless noted otherwise.

Following standard information-theoretic design criteria, we adopt a determinant (D-optimal) objective given by

J(E;θi,ϕi,γ)=det(F(E;θi,ϕi,γ)+ϵI), (15)

where ϵ=10−6 is a small regularization constant used to ensure numerical stability and I is the identity matrix. This regularization ensures that det(F(E;θi,ϕi,γ)+ϵI) remains strictly positive even when F is poorly conditioned or nearly singular for a given excitation sequence E and orientation (θi,ϕi,γ) , which can otherwise occur during optimization and make the square-root objective J numerically unstable.

Maximizing J seeks to minimize the generalized variance Cov(mˆ) of measuring the second moments, which in turn improves joint sensitivity to mean orientation (θ,ϕ) and wobble γ (Eq. (2)). To develop a scheme optimized for all possible orientations, we minimize

L(β1,α1,…,βL,αL)=−∑i=1NlatticeJ(E(β1,α1,…,βL,αL);θi,ϕi,γ) (16)

using gradient-based optimization (L-BFGS-B via scipy.optimize.minimize, SciPy v1.17.1), with gradients computed analytically via PyTorch autograd ( torch.Tensor.grad, PyTorch v2.11.0). For each optimization run, i.e., each combination of DSF and L , we run 5 independent restarts with parameters (βl,αl) each initialized randomly as unconstrained real values and keep the best result. Each restart runs for at most 200 iterations using default convergence tolerances.

Thus, for L polarization states, there are 2L angular parameters to optimize (Eqs. (6),(7)). The excitation modulation designs featured in Secs. 3 and 4 are optimized for the MVR microscope [8]. Section 5 describes optimal excitation modulation designs for the double-helix [25,26], pixOL [27], vortex [28,29], and CHIDO ( c=π/2 ) [30,37] DSFs.

To compare performance across various numbers of frames L , we compute the fractional improvement in performance using

ΔL=J(EL;θ10,L,ϕ10,L,γ)−J(EL−1;θ10,L−1,ϕ10,L−1,γ)J(EL;θ10,L,ϕ10,L,γ), (17)

where EL denotes a modulation scheme that uses L excitation polarizations and (θ10,L,ϕ10,L) refers to the specific molecular orientation that generates a tenth-percentile ( P10 ) performance score J . We use the tenth-percentile score J rather than the median to quantify performance improvement for difficult-to-measure orientations. Note that the tenth-percentile orientation (θ10,L,ϕ10,L) for L modulations is typically not the same as that (θ10,L−1,ϕ10,L−1) for L−1 modulations, because the specific polarization states Eℓ can vary widely as the number of illuminations L increases.

All modeling, evaluation, and optimization code, performance data, and optimized excitation polarization designs are available via OSF (Ref. [44]).

3. Optimal linearly polarized illumination

Previous studies [23] have shown that taking multiple images of a molecule, each using a unique polarization of excitation light, enhances orientation measurement sensitivity even when using engineered DSFs. For example, if molecules of all possible orientations are excited equally during a single imaging frame, then we observe a median performance score log⁡Jsingle2=13.8 . In contrast, one may collect 6 images of a molecule, each using a different linearly polarized electric field (βℓ,αℓ)={(90∘,90∘) , (35.26∘,180∘) , (90∘,30∘) , (35.26∘,−60∘) , (90∘,−30∘) , (35.26∘,60∘)} (Fig. 1(b).i) such that the total absorption probability Pabs(θ,ϕ) over all 6 images is uniform for all orientations (θ,ϕ) [24]. We term this scheme equal pumping (EP). As a performance baseline, EP yields an average performance score log⁡JEP2=16.9 , corresponding to a median improvement in measurement precision JEP/Jsingle of 35.0 (Fig. S1(a).i).

We now optimize L (Eq. (16)) to find L=6 linearly polarized excitation states that maximize the MVR microscope’s precision for measuring molecular orientation and wobble (Sec. 2.4). The resulting polarization states are shown in Fig. 1(b).ii as absorption probabilities Pabs,ℓ(θ,ϕ) for each Eℓ . Note that by modulating the illumination polarization, the total number of excitation photons absorbed by a molecule over L frames will now depend on its orientation (θ,ϕ) , which will then be linearly proportional to the number of fluorescence photons detected on the camera. The EP scheme has a uniform absorption probability for all molecular orientations, but the optimized scheme pumps in-plane molecules more strongly ( Nabs=963 for z-oriented molecules vs. Nabs=1020 for molecules in the xy plane, see Fig. S2(a)). This variation directly impacts orientation-measurement sensitivity as quantified by the entries Fjk of the FIM (Fig. 1(c) and Eq. (11)). Fxy,xy , Fxz,xz , and Fyz,yz in particular are boosted in amplitude by 46.4%, 28.6%, and 54.8%, respectively. Boosting information in these so-called cross-moments leads to a median performance score log⁡Jopt,L=62=17.0 , corresponding to a median 23.1% improvement in measurement precision Jopt,L=6/JEP over that of EP (Fig. S1(b).iii).

We next investigate how the number of excitation polarization states ( L=2 to 7) affects orientation-measurement performance while holding the total number of pumping photons constant (Sec. 2.1). Any particular scheme will have FI scores that vary with orientation (θ,ϕ) ( log⁡J2 shown in Fig. 2(a)); these scores can vary as much as an order of magnitude over the orientation hemisphere, e.g., for L=5 . To compare schemes, we quantify the fractional improvement in the tenth-percentile performance ΔL (Eq. (17), Sec. 2.4) as the number of pumping modulations L increases (Fig. 2(b)). This metric directly measures how much each scheme improves measurements of the most difficult orientations, making it a stringent indicator of design robustness.

Fig. 2.

Fig. 2.

Orientation measurement performance of optimized linearly polarized excitation schemes. (a) Orientation measurement performance of (gray) the EP and optimized linear (orange) L=4 , (green) L=5 , and (red) L=6 schemes reported as log⁡J2 (Eq. (15)) for all possible orientations. Markers denote the median and shaded regions span the P10 – P90 percentiles. (b) Fractional improvement in estimation performance ΔL when using L modulations compared to L−1 modulations, calculated using the tenth percentile performance score (Eq. (17)). Dashed line: ΔL=5% . (c) Performance score J2(θ,ϕ) for the (i) EP and linear (ii) L=4 , (iii) L=5 , and (iv) L=6 schemes. (d) Information gain contributed by each excitation polarization state, represented as ∑ℓdet(F(Eℓ;θ,ϕ,γ=0.9)) for each excitation scheme corresponding to (c). (e) Best-possible precision of measuring the azimuthal angle σϕ (weighted by sin⁡θ ), reported as the median over all angles θ . (Dotted black) EP, (solid orange) linear L=4 , (solid green) L=5 , and (solid red) L=6 . (f) Best-possible precision of measuring the polar angle σθ , reported as the median over all angles ϕ . Precisions are computed using the square root of the CRB (Eqs. (12), (13)). Colorbars: (c,d) Fisher information (normalized to the L=4 scheme).

Using L=4 modulations strikes a surprisingly favorable balance. Its median FI score is comparable to schemes with larger L , and crucially, it exhibits more uniform measurement performance across all dipole orientations than other schemes (Figs. 2(a), (c).ii). The 4 linearly polarized states all lie at β=54∘ spaced at 90∘ intervals along the azimuth (Figs. 2(d).ii), in contrast to the 3 out-of-plane and 3 xy-plane illumination polarizations for EP (Figs. 2(d).i). Despite having 2 fewer illumination polarizations, L=4 achieves superior median performance over EP (21.2% median improvement in measurement precision Jopt,L=4/JEP , Fig. S1(b).i).

Interestingly, adding one illumination polarization ( L=5 ) dramatically changes performance relative to L=4 : FI now varies widely for various (θ,ϕ) , with molecules in the xy plane ( θ=90∘ ) exhibiting 1.56 times worse J2 scores than molecules aligned along the optical axis ( θ=0 ) (Fig. 2(c).iii). Comparing the configuration of illumination polarizations, we find that the optimizer chooses to strengthen the relatively weak performance of z-oriented molecules for L=4 (Fig. 2(d).ii) by adding z-polarized excitation (Fig. 2(d).iii). With the other four pumping directions relatively unchanged, in-plane orientation precision remains relatively weak.

The L=6 scheme recovers from this imbalance. Its tenth-percentile FI score is highest of the schemes shown (Figs. 2(a),(b)), and it exhibits relatively strong FI scores for molecules oriented within the xz and yz planes (Fig. 2(c).iv). Using six polarization states gives the optimizer freedom to design pairs of complementary states satisfying E(βℓ,αℓ)⋅E(βcomp,αcomp)≈0 , resulting in broadened and balanced orientation coverage (Fig. 2(d).iv). Interestingly, the optimizer has converged to use only 2 xy-polarized illuminations (Fig. 2(d).iv) compared to 3 for EP (Fig. 2(d).i), thereby using 4 out-of-plane excitations to boost information collected from intermediate θ angles.

We may more directly compare measurement performance by computing the best-possible precision for measuring (θ,ϕ) using the CRB (Figs. 2(e,f) and Eqs. (12), (13)). Detailed distributions of orientation precisions σϕ and σθ are shown in Fig. S3(a-c). The optimized L=4 scheme achieves median precisions σθ,50=2.59∘ and σϕ,50=3.22∘ as well as relatively uniform performance across all orientations given by σθ,10--90∈[2.47∘,2.71∘] and σϕ,10--90∈[3.01∘,5.23∘] ( P10 – P90 ). However, increasing to L=5 improves the median polar angle precision to σθ,50=2.41∘ but worsens the azimuthal angle precision to σϕ,50=3.43∘ and widens the performance spread significantly ( σθ,10--90∈[2.17∘,2.85∘] , σϕ,10--90∈[3.13∘,5.56∘] ). The L=6 scheme brings polar and azimuthal angle estimation back on par with one another ( σθ,50=2.50∘ , σϕ,50=3.19∘ ). The variation in performance is also reduced across all orientations ( σθ,10--90∈[2.39∘,2.73∘] , σϕ,10--90∈[2.79∘,5.80∘] ) compared to L=5 . Further, the optimized L=6 polarization states achieve better orientation precision than EP ( Δσθ,50=−0.06∘ , Δσϕ,50=−0.24∘ at the median), thereby confirming that achieving superior performance scores J (Fig. 2(a)) directly translates to superior orientation precision (σθ,σϕ) .

To assess design and performance stability, we repeated the polarization optimization process for molecules exhibiting greater wobble ( γ=0.5 corresponding to a wobble cone Ω≈0.76π ). One may interpret such a molecule as emitting approximately half of its photons identically to an isotropic emitter, and the other half as a fixed dipole (Eq. (2)). We consequently observe orientation sensitivity to decrease for molecules with greater wobble (Fig. S4(a)) relative to more confined molecules (Fig. 2(a)) as expected, but our optimization strategy produces the same performance trends (Figs. S4(b,c)) as before (Figs. 2(b,c)). Importantly, the optimized designs using both L=4 and 6 pumping polarizations are nearly identical for molecules of both small ( γ=0.9 ) and moderate ( γ=0.5 ) wobble (Figs. 2(d).ii,(d).iv and S4(d).ii,(d).iv), thus demonstrating that our design strategy is robust to changes in wobble. Interestingly, designing a scheme that uses L=5 polarized excitations is difficult for both small and moderate wobble, represented by the negative performance improvements in Figs. 2(b) and S4(b), and is likely due to non-complementarity between L=5 modulations and the 4-fold symmetry of the MVR microscope. Notably, both L=4 and 6 modulations significantly improve orientation precision relative to EP ( >20% , Figs. S4e,f), similar to molecules with small wobble.

Overall, the best linear scheme for a particular application depends on the relative importance of maximizing measurement performance for specific orientations vs. the entire sphere (Figs. 2(e,f)) and maximizing temporal resolution, i.e., minimizing the number of excitation modulations L . We can prioritize temporal resolution by adopting a stopping criterion for ΔL (Eq. (17)); e.g., we find the smallest L that does not cross the threshold ΔL≤5% . We thus choose L=4 with its excellent uniformity in orientation measurement performance, best median azimuthal angle precision, and superior polar angle precision for small θ . Moving to L=6 slightly improves overall measurement precision, especially for intermediate polar angles θ , at the cost of requiring 50% additional excitation modulations, i.e., 50% more images to acquire to measure orientation.

4. Optimal circularly polarized illumination

We next extend our design framework to find the set of optimal circularly polarized illuminations that yield the best orientation precision, keeping the total number of excitation photons consistent between linearly and circularly polarized schemes (Fig. S1(b)). To do this, we relate the design parameters (βℓ,αℓ) to the second moments of circularly polarized electric fields Eℓ,circ (Eqs. (6),(7)). This facile modification does not change the number of parameters to be optimized and thus holds the computational cost of optimization constant. We optimize for a range of frame numbers Lcirc and quantify relative performance (Fig. 3(a)) and tenth-percentile performance improvements ΔL with increasing Lcirc (Fig. 3(b)).

Fig. 3.

Fig. 3.

Orientation measurement performance of optimized circularly polarized excitation schemes. (a) Orientation measurement performance of optimal (blue) linear L=2 and circular (orange) Lcirc=2 , (green) Lcirc=3 , and (red) Lcirc=4 schemes reported as log⁡J2 (Eq. (15)) for all possible orientations. Markers denote the median and shaded regions span the P10 – P90 percentiles. (b) Fractional improvement in estimation performance ΔL when using Lcirc modulations compared to Lcirc−1 modulations, calculated using the tenth percentile performance score (Eq. (17)). Dashed line: ΔL=5% . (c) Performance score J2(θ,ϕ) for the (i) linear L=2 and circular (ii) Lcirc=2 , (iii) Lcirc=3 , and (iv) Lcirc=4 schemes. (d) Information gain contributed by each excitation polarization state, represented as ∑ℓdet(F(Eℓ,circ;θ,ϕ,γ)) for each excitation scheme corresponding to (c). (e) Best-possible precision of measuring the azimuthal angle σϕ (weighted by sin⁡θ ), reported as the median over all angles θ . (Blue) linear L=2 , circular (orange) Lcirc=2 , (green) Lcirc=3 , and (red) Lcirc=4 . (f) Best-possible precision of measuring the polar angle σθ , reported as the median over all angles ϕ . Precisions are computed using the square root of the CRB (Eqs. (12), (13)). Colorbars: (c,d) Fisher information (normalized to the Lcirc=2 scheme).

We observe that the performance of linear L=2 varies by several orders of magnitude over all possible orientations, but that of circular Lcirc=2 is much more uniform (Fig. 3(a),(c).i,(c).ii). This result is expected, because a single linearly polarized state excites only a focused spot in orientation space (Fig. 3(d).i), and collecting very few photons from dipoles outside this region causes poor measurement precision. In contrast, a single circularly polarized beam pumps an entire ring of dipole orientations simultaneously (Fig. 3(d).ii), at the cost of reduced pumping efficiency per excitation photon. Together, these features enable circularly polarized excitation to achieve superior performance over linear, with a median 27% improvement in measurement precision ( JL=2,circ/JL=2,linear ). For Lcirc=3 and 4, measurement precision improves in terms of both median and tenth-percentile performance (Fig. 3(a),(b),(c).iii,(c).iv). Further increases in Lcirc beyond 5 produce mixed results. Due to the ring-shaped excitation pattern, odd Lcirc values cause asymmetric performance in orientation space (Fig. 3(d).iii), whereas even values of Lcirc restore symmetry (Fig. 3(d).iv).

We again use CRB to characterize orientation measurement performance. Detailed distributions of orientation precisions σϕ and σθ are shown in Fig. S3(d-f). Circularly polarized Lcirc=2 achieves median precisions of σθ,50=3.84∘ and σϕ,50=4.53∘ with spreads of σθ,10--90∈[2.69∘,6.10∘] and σϕ,10--90∈[3.83∘,9.59∘] ( P10 – P90 ). In comparison, linear L=2 has nearly identical median polar angle precision ( σθ,50=3.82∘ ) but a dramatically wider spread ( σθ,10--90∈[2.32∘,11.59∘] , σϕ,10--90∈[2.11∘,17.62∘] ). Interestingly, linear L=2 also has superior median azimuthal precision ( σϕ,50=3.40∘ ) compared to all circular polarization schemes explored here (Fig. 3(e)). Increasing to Lcirc=3 improves the median precisions to σθ,50=3.65∘ and σϕ,50=4.57∘ and narrows the performance variations considerably ( σθ,10--90∈[2.87∘,4.73∘] , σϕ,10--90∈[3.92∘,9.24∘] ). Lcirc=4 achieves the best median and tightest distribution among all circular schemes, with σθ,50=3.49∘ , σϕ,50=4.61∘ , σθ,10--90∈[3.03∘,4.10∘] , and σϕ,10--90∈[3.85∘,8.58∘] .

Importantly, these trends reveal a different tradeoff as L increases for circularly versus linearly polarized excitations. Overall, fewer frames Lcirc are needed to achieve decent performance across all orientations; for L=2 , circular is superior to linear. We attribute this behavior to the ring vs. spot excitation patterns of circularly and linearly polarized illuminations, respectively, where circularly polarized light pumps a large range of dipole orientations more efficiently. However, if there is sufficient time to use a large number of modulations before photobleaching, combining L≥4 linear polarizations for orientation measurement yields superior precision over using circular (median 6.83-fold improvement, JL=4,linear/JL=4,circ , Fig. S1(c).iii). Each individual linearly polarized excitation provides increased Fisher information for measuring orientation, but more are required to cover orientation space sufficiently.

5. Designing optimal polarization modulations for specific DSFs

Up to this point, optimal excitation modulation schemes were derived for the MVR microscope [8], which encodes orientational information through a combination of radial and azimuthal polarization splitting across multiple views (Fig. 1(a), inset right). Because Fisher information depends on both the excitation scheme and the detection DSF (parameterized using the basis images B , Eq. (10)), excitation strategies must be tailored to specific microscopes for optimal performance. Here we extend optimal excitation modulation design to four additional DSFs: double helix [25,26], pixOL [27], polarized vortex [29], and CHIDO ( c=π/2 ) [30] (see Fig. S5 for representative DSF images).

To prioritize temporal resolution, we use the stopping criterion ΔL≤5% (Eq. (17)), thereby ensuring that additional excitation modulations are considered only when they provide a meaningful improvement in worst-case orientation precision. Either L=5 or 6 modes are optimal for all the DSFs we tested (Figs. 4(a) and S5), but the polarization schemes themselves are uniquely customized to each DSF (Figs. 4(b) and S6). For example, the optimal scheme for the double-helix DSF uses two closely-spaced pumping polarizations near the z-axis, with the 4 remaining modes oriented closer to the xy plane (Fig. 4(b).i). In contrast, CHIDO uses a highly symmetric scheme with a set of 3 pumping polarizations at β=76∘ , another set at β=39∘ , and the modes within each set spaced at 120∘ intervals along α (Fig. 4(b).iv).

Fig. 4.

Fig. 4.

Orientation measurement performance of optimized linearly polarized excitation schemes for various dipole-spread functions (DSFs). (a) Fractional improvement in estimation performance ΔL when using L modulations compared to L−1 modulations, calculated using the tenth percentile performance score (Eq. (17)) for the (i) double-helix (optimal L=6 ) [25,26], (ii) pixOL (optimal L=5 ) [27], (iii) vortex (optimal L=5 ) [29], and (iv) CHIDO (optimal L=6 ) [30] DSFs. Dashed line: ΔL=5% . (b) Information gain contributed by each excitation polarization state, represented as ∑ℓdet(F(Eℓ;θ,ϕ,γ)) for each DSF corresponding to (a). (c) Best-possible precision of measuring the azimuthal angle σϕ (weighted by sin⁡θ ), reported as the median over all angles θ . (d) Best-possible precision of measuring the polar angle σθ , reported as the median over all angles ϕ . Solid lines: optimized scheme. Dashed lines: EP. Precisions are computed using the square root of the CRB (Eqs. (12), (13)). Colorbars: (b) Fisher information.

Analyzing the best-possible precisions using the CRB, we find that the optimized linear excitation schemes for each DSF generally exhibit improved median orientation precisions over those of EP (double-helix: σθ,50=3.61∘ and σϕ,50=4.42∘ , pixOL: σθ,50=2.52∘ and σϕ,50=2.62∘ , vortex: σθ,50=2.70∘ and σϕ,50=3.00∘ , CHIDO: σθ,50=2.72∘ and σϕ,50=3.00∘ , Figs. 4(c),(d)). Each optimized scheme also shows smaller performance variations across orientation space compared to EP. Detailed distributions of measurement performance σϕ and σθ are shown in Fig. S7. Although not shown here, we expect this strategy, if applied to the unpolarized vortex DSF [28], to produce a different set of excitation polarizations and yield different performance tradeoffs.

Overall, this analysis shows that using optimized excitation modulation schemes improves orientation precision, reduces the number of modulations required relative to the baseline EP scheme ( L=6 ), or both. The degree of improvement certainly depends on the DSF design itself. Each DSF has its own weaknesses; for example when using EP, double helix, pixOL, and vortex have worse precision σϕ at ϕ={0,90∘,180∘,270∘} , while CHIDO measures ϕ with worse precision at ϕ={60∘,180∘,300∘} (Figs. 4(c) and S7). Accordingly, our optimized schemes specifically compensate for these weaknesses.

We can gain insight into the optimization strategy by examining how excitation modulation and DSF engineering individually contribute to Fisher information in Eq. (10). The first term, which measures sensitivity gained by using different polarized excitations Eℓ , is largely shaped by each DSF and shows very similar responses across (θ,ϕ) for both EP and optimized schemes (Fig. S8i,iii). However, excitation modulation greatly reshapes the information conveyed by DSF engineering (Fig. S8ii,iv). For each DSF, orientations with weak information yielded by excitation modulation (Fig. S8iii) are specifically complemented by strong information from DSF engineering (Fig. S8iv). Our optimization strategy specifically pumps certain dipole orientations strongly (Fig. 4(b)) to achieve this complementarity and achieve overall high performance (Fig. S7).

6. Discussion and conclusion

We have presented a systematic framework for designing sequences of polarized excitations for single-molecule orientation-localization microscopy (SMOLM). By maximizing a Fisher information objective, our approach produces optimal excitation schemes for any number L of illumination polarizations using a fixed number of total pumping photons. Due to the fixed photon budget, we observe that increasing the number of polarized illuminations has rapidly diminishing benefits (Fig. S9), regardless of DSF, and find that the optimal number for most DSFs is L=5 or 6. Using circular polarizations has the benefit of illuminating a wider range of dipole orientations with fewer modes, thus yielding superior performance over linear for as few as Lcirc=2 modulations. This scheme is beneficial when the fewest number of illumination modulations is desired, thus maximizing temporal resolution while maintaining good performance (Fig. 3(e,f)).

However, linearly polarized modes produce superior orientation performance when L≥4 . Moreover, pumping all possible orientations equally does not produce performance superior to those of our optimized designs. Notably, the optimal scheme for MVR uses only L=4 modulations (Fig. 2(a)), which is fewer than all other DSFs, yet still achieves comparable median orientation precisions ( σθ,50=2.59∘ , σϕ,50=3.22∘ ) (Fig. 2(e,f)). The fewer number of modes implies that the MVR microscope efficiently encodes molecular orientation information into its fluorescence images (Fig. 1(a) right inset).

In this work, we assume that the molecules of interest wobble uniformly within a cone (Eq. (2)). Our design methodology can be easily adapted for molecules rotating within a harmonic or arbitrary potential well [32] or exhibiting anisotropic wobble [29,34]. Optimal polarization modulation schemes for these types of rotational dynamics will undoubtedly be different from the ones shown here. Further, we assume that molecules undergo fast rotational diffusion, i.e., the fluorescence lifetime is much longer than the rotational correlation time, thus resulting in absorption and emission transition dipole moments that are uncorrelated [24,32]. Our framework may also be adapted for slow rotational diffusion, where correlations between absorption and emission introduce fourth-order dipole moments [24]. These correlations can be harnessed for greater orientation-measurement sensitivity, but doing so requires careful optimization of polarization modulations to the absorption dipole dynamics as well as a DSF tailored to the emission dipole dynamics.

Our design approach has limitations. For example, performance improvements for measuring rotational confinement γ are mixed, where changing the number of polarizations L can improve median performance while worsening performance variations, or vise-versa (Fig. S10). It is possible that our optimization criterion (Eq. (11)) is not optimal for specifically improving wobble measurement performance, and further theoretical developments are needed [24,33,34,37,45,46]. In particular, there may be opportunities to improve wobble precision by jointly engineering excitation modulations and DSFs simultaneously. Since excitation and detection contribute additively to FI (Eq. (10)), joint optimization of excitation polarizations and a DSF could reveal phase-mask symmetries and polarization-channel splitting schemes that synergize with specific excitation polarization states. We believe this is a promising direction for future work.

Importantly, our design approach is restricted to optimizing linearly and circularly polarized excitation fields using parameters (β,α) . A general treatment by Thorsen et al. [23] analyzes FI for any fully polarized electric field by parameterizing its propagation direction (β,α) and the ellipticity angle τ and long axis ξ of its polarization ellipse. Our findings here may be viewed as special cases of this framework. Expanding to elliptical polarization states could overcome some limitations of linearly and circularly polarized excitation, especially when considering the wobble of a molecule, by enabling Eℓ (4 degrees of freedom) to be directly tailored to (θ,ϕ,γ) (3 degrees of freedom). However, optimizing FI over all possible polarization states represents a significantly more challenging design problem and is the subject of a future study.

Finally, our design study is strictly computational and optimizes the information theoretic Cramér–Rao bound, which does not guarantee that practical estimators or neural networks can achieve this performance in an experiment. Realizing experimentally the performance benefits of optimal excitation modulation will require careful control and calibration of illumination polarizations [23,47], calibration of the detection DSF [48,49], and robust algorithms for estimating molecular orientations using multiple illuminations while accounting for possible blinking photophysics during an acquisition [50,51]. We reserve these tasks for future work.

Overall, our easy-to-use framework facilitates optimal measurements of absorption dipole orientations. Tailored optimization schemes for various DSFs powerfully compensate for any performance weaknesses they may exhibit. Tradeoffs between temporal resolution (e.g., the number of excitation frames L ) and orientation precision are easily adjusted alongside polarization constraints. Here, we optimized excitation schemes to achieve excellent median and tenth-percentile performance across all orientations, but the flexibility of our framework also enables bespoke designs customized for specific samples (i.e., local regions of orientation space). With these benefits, we believe our approach enables more facile and performant SMOLM for cutting-edge biophysical and chemical imaging.

Supplemental information

Supplement 1. Supplement 1.
boe-17-9-4854-s001.pdf (3.6MB, pdf)

Acknowledgments

The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH. It is subject to the NIH Public Access Policy. Through acceptance of this federal funding, NIH has been given a right to make this manuscript publicly available in PubMed Central upon the Official Date of Publication, as defined by NIH.

Funding

National Institute of General Medical Sciences https://ror.org/04q48ey07 ( R35GM124858).

Disclosures

The excitation modulation optimization methodology mentioned in this work was invented by K.A.N., Y.Q., and M.D.L., and Washington University has filed a patent application covering the technology (application number 64/034,172).

Data availability

Data underlying the results presented in this paper are available via OSF (Ref. [44]).

Supplemental document

See Supplement 1 (3.6MB, pdf) for supporting content.

References

  • 1.Brasselet S., Alonso M. A., “Polarization microscopy: From ensemble structural imaging to single-molecule 3D orientation and localization microscopy,” Optica 10(11), 1486 (2023). 10.1364/OPTICA.502119 [DOI] [Google Scholar]
  • 2.Zhang O., Lew M. D., “Single-molecule orientation-localization microscopy: Applications and approaches,” Q. Rev. Biophys. 57, e17 (2024). 10.1017/S0033583524000167 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Brasselet S., Lew M. D., “Single-molecule orientation and localization microscopy,” Nat. Photonics 19(9), 925–937 (2025). 10.1038/s41566-025-01724-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lelek M., Gyparaki M. T., Beliu G., et al. , “Single-molecule localization microscopy,” Nat. Rev. Methods Primers 1(1), 39 (2021). 10.1038/s43586-021-00038-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Hugelier S., Colosi P., Lakadamyali M., “Quantitative Single-Molecule Localization Microscopy,” Annu. Rev. Biophys. 52(1), 139–160 (2023). 10.1146/annurev-biophys-111622-091212 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Rimoli C. V., Valades-Cruz C. A., Curcio V., et al. , “4polar-STORM polarized super-resolution imaging of actin filament organization in cells,” Nat. Commun. 13(1), 301 (2022). 10.1038/s41467-022-27966-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Senthil Kumar C. S., Valades Cruz C. A., Sison M., et al. , “4polar3D single molecule imaging of 3D orientation in dense actin networks using ratiometric polarization splitting,” Nat. Commun. 17(1), 4246 (2026). 10.1038/s41467-026-70852-y [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Zhang O., Guo Z., He Y., et al. , “Six-dimensional single-molecule imaging with isotropic resolution using a multi-view reflector microscope,” Nat. Photonics 17(2), 179–186 (2022). 10.1038/s41566-022-01116-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Sun B., Ding T., Zhou W., et al. , “Single-Molecule Orientation Imaging Reveals the Nano-Architecture of Amyloid Fibrils Undergoing Growth and Decay,” Nano Lett. 24(24), 7276–7283 (2024). 10.1021/acs.nanolett.4c01263 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhou W., O’Neill C. L., Ding T., et al. , “Resolving the Nanoscale Structure of β-Sheet Peptide Self-Assemblies Using Single-Molecule Orientation–Localization Microscopy,” ACS Nano 18(12), 8798–8810 (2024). 10.1021/acsnano.3c11771 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Wu T., King M. R., Qiu Y., et al. , “Single-fluorogen imaging reveals distinct environmental and structural features of biomolecular condensates,” Nat. Phys. 21(5), 778–786 (2025). 10.1038/s41567-025-02827-7 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Novotny L., Beversluis M. R., Youngworth K. S., et al. , “Longitudinal Field Modes Probed by Single Molecules,” Phys. Rev. Lett. 86(23), 5251–5254 (2001). 10.1103/PhysRevLett.86.5251 [DOI] [PubMed] [Google Scholar]
  • 13.Beversluis M. R., Novotny L., Stranick S. J., “Programmable vector point-spread function engineering,” Opt. Express 14(7), 2650 (2006). 10.1364/OE.14.002650 [DOI] [PubMed] [Google Scholar]
  • 14.Marx D., Gligonov I., Malsbenden D., et al. , “Mapping Optical Chirality with Single Fluorescent Molecules,” Nano Lett. 26(9), 3012–3017 (2026). 10.1021/acs.nanolett.5c05316 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Fazel M., Grussmayer K. S., Ferdman B., et al. , “Fluorescence microscopy: A statistics-optics perspective,” Rev. Mod. Phys. 96(2), 025003 (2024). 10.1103/RevModPhys.96.025003 [DOI] [Google Scholar]
  • 16.Wu T., Lew M. D., “Dipole-Spread Function Engineering for Six-Dimensional Super-Resolution Microscopy,” in Coded Optical Imaging , Liang J., ed. (Springer International Publishing, Cham, 2024), pp. 207–223. [Google Scholar]
  • 17.Sosa H., Peterman E. J., Moerner W. E., et al. , “ADP-induced rocking of the kinesin motor domain revealed by single-molecule fluorescence polarization microscopy.,” Nat. Struct Biol. 8(6), 540–544 (2001). 10.1038/88611 [DOI] [PubMed] [Google Scholar]
  • 18.Forkey J. N., Quinlan M. E., Alexander Shaw M., et al. , “Three-dimensional structural dynamics of myosin V by single-molecule fluorescence polarization,” Nature 422(6930), 399–404 (2003). 10.1038/nature01529 [DOI] [PubMed] [Google Scholar]
  • 19.Backer A. S., Lee M. Y., Moerner W. E., “Enhanced DNA imaging using super-resolution microscopy and simultaneous single-molecule orientation measurements,” Optica 3(6), 659 (2016). 10.1364/OPTICA.3.000659 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Valeur B., Berberan-Santos M. N., Molecular Fluorescence: Principles and Applications (Wiley-VCH, 2012), 2nd edition ed. [Google Scholar]
  • 21.Zhang O., Zhou W., Lu J., et al. , “Resolving the Three-Dimensional Rotational and Translational Dynamics of Single Molecules Using Radially and Azimuthally Polarized Fluorescence,” Nano Lett. 22(3), 1024–1031 (2022). 10.1021/acs.nanolett.1c03948 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Munger E., Sison M., Brasselet S., “Influence of the excitation polarization on single molecule 3D orientation imaging,” Opt. Commun. 541, 129480 (2023). 10.1016/j.optcom.2023.129480 [DOI] [Google Scholar]
  • 23.Ø. Thorsen R., Hulleman C. N., Rieger B., et al. , “Photon efficient orientation estimation using polarization modulation in single-molecule localization microscopy,” Biomed. Opt. Express 13(5), 2835 (2022). 10.1364/BOE.452159 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Chen Y., Qiu Y., Lew M. D., “Resolving the Orientations of and Angular Separation Between a Pair of Dipole Emitters,” Phys. Rev. Lett. 134(9), 093805 (2025). 10.1103/PhysRevLett.134.093805 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Pavani S. R. P., Thompson M. A., Biteen J. S., et al. , “Three-dimensional, single-molecule fluorescence imaging beyond the diffraction limit by using a double-helix point spread function,” Proc. Natl. Acad. Sci. 106(9), 2995–2999 (2009). 10.1073/pnas.0900245106 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Backlund M. P., Lew M. D., Backer A. S., et al. , “Simultaneous, accurate measurement of the 3D position and orientation of single molecules,” Proc. Natl. Acad. Sci. 109(47), 19087–19092 (2012). 10.1073/pnas.1216687109 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Wu T., Lu J., Lew M. D., “Dipole-spread-function engineering for simultaneously measuring the 3D orientations and 3D positions of fluorescent molecules,” Optica 9(5), 505 (2022). 10.1364/OPTICA.451899 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Hulleman C. N., Ø. Thorsen R., Kim E., et al. , “Simultaneous orientation and 3D localization microscopy with a Vortex point spread function,” Nat. Commun. 12(1), 5934 (2021). 10.1038/s41467-021-26228-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Ding T., Lew M. D., “Single-Molecule Localization Microscopy of 3D Orientation and Anisotropic Wobble Using a Polarized Vortex Point Spread Function,” J. Phys. Chem. B 125(46), 12718–12729 (2021). 10.1021/acs.jpcb.1c08073 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Curcio V., Alemán-Casta neda L. A., Brown T. G., et al. , “Birefringent Fourier filtering for single molecule coordinate and height super-resolution imaging with dithering and orientation,” Nat. Commun. 11(1), 5307 (2020). 10.1038/s41467-020-19064-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Backer A. S., Moerner W. E., “Determining the rotational mobility of a single molecule from a single image: a numerical study,” Opt. Express 23(4), 4255–4276 (2015). 10.1364/OE.23.004255 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Stallinga S., “Effect of rotational diffusion in an orientational potential well on the point spread function of electric dipole emitters,” J. Opt. Soc. Am. A 32(2), 213 (2015). 10.1364/JOSAA.32.000213 [DOI] [PubMed] [Google Scholar]
  • 33.Zhang O., Lew M. D., “Fundamental Limits on Measuring the Rotational Constraint of Single Molecules Using Fluorescence Microscopy,” Phys. Rev. Lett. 122(19), 198301 (2019). 10.1103/PhysRevLett.122.198301 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Zhou W., Wu T., Lew M. D., “Fundamental Limits in Measuring the Anisotropic Rotational Diffusion of Single Molecules,” J. Phys. Chem. A 128(28), 5808–5815 (2024). 10.1021/acs.jpca.4c03160 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Backer A. S., Moerner W. E., “Extending Single-Molecule Microscopy Using Optical Fourier Processing,” J. Phys. Chem. B 118(28), 8313–8329 (2014). 10.1021/jp501778z [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Ding T., Wu T., Mazidi H., et al. , “Single-molecule orientation localization microscopy for resolving structural heterogeneities between amyloid fibrils,” Optica 7(6), 602 (2020). 10.1364/OPTICA.388157 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Herrera I., Alemán-Casta neda L. A., Brasselet S., et al. , “Stokes-based analysis for the estimation of 3D dipolar emission,” J. Opt. Soc. Am. A 41(11), 2134 (2024). 10.1364/JOSAA.538706 [DOI] [PubMed] [Google Scholar]
  • 38.Chao J., Sally Ward E., Ober R. J., “Fisher information theory for parameter estimation in single molecule microscopy: Tutorial,” J. Opt. Soc. Am. A 33(7), B36 (2016). 10.1364/JOSAA.33.000B36 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Zhang O., Lew M. D., “Single-molecule orientation localization microscopy II: A performance comparison,” J. Opt. Soc. Am. A 38(2), 288 (2021). 10.1364/JOSAA.411983 [DOI] [PubMed] [Google Scholar]
  • 40.Shechtman Y., Sahl S. J., Backer A. S., et al. , “Optimal Point Spread Function Design for 3D Imaging,” Phys. Rev. Lett. 113(13), 133902 (2014). 10.1103/PhysRevLett.113.133902 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Jusuf J. M., Lew M. D., “Towards optimal point spread function design for resolving closely spaced emitters in three dimensions,” Opt. Express 30(20), 37154 (2022). 10.1364/OE.472067 [DOI] [PubMed] [Google Scholar]
  • 42.Liu Y., Dong J., Maya J. A., et al. , “Point-spread-function engineering in MINFLUX: Optimality of donut and half-moon excitation patterns,” Opt. Lett. 50(1), 37–40 (2025). 10.1364/OL.543882 [DOI] [PubMed] [Google Scholar]
  • 43.González Á., “Measurement of Areas on a Sphere Using Fibonacci and Latitude–Longitude Lattices,” Math. Geosci. 42(1), 49–64 (2009). 10.1007/s11004-009-9257-x [DOI] [Google Scholar]
  • 44.Nie K. A., Qiu Y., Lew M. D., “Optimal excitation polarization schemes to boost the precision of single-molecule orientation imaging,” OSF (2026). https://osf.io/hzw5x/overview?view_only=765d89e12f1140d1afc26840daa357c0.
  • 45.Chandler T., Shroff H., Oldenbourg R., et al. , “Spatio-angular fluorescence microscopy III. Constrained angular diffusion, polarized excitation, and high-NA imaging,” J. Opt. Soc. Am. A 37(9), 1465 (2020). 10.1364/JOSAA.389217 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Zhang O., Lew M. D., “Single-molecule orientation localization microscopy I: Fundamental limits,” J. Opt. Soc. Am. A 38(2), 277 (2021). 10.1364/JOSAA.411981 [DOI] [PubMed] [Google Scholar]
  • 47.Qiu Y., Nie K. A., Lew M. D., “Sensing negative-cone rotational diffusion of dipole-like emitters,” in Single Molecule Spectroscopy and Superresolution Imaging XIX , vol. 13858 (SPIE, 2026), pp. 15–22. [Google Scholar]
  • 48.Ferdman B., Nehme E., Weiss L. E., et al. , “VIPR: Vectorial implementation of phase retrieval for fast and accurate microscopic pixel-wise pupil estimation,” Opt. Express 28(7), 10179 (2020). 10.1364/OE.388248 [DOI] [PubMed] [Google Scholar]
  • 49.Liu S., Chen J., Hellgoth J., et al. , “Universal inverse modeling of point spread functions for SMLM localization and microscope characterization,” Nat. Methods 21(6), 1082–1093 (2024). 10.1038/s41592-024-02282-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Pennacchietti F., Gould T. J., Hess S. T., “The Role of Probe Photophysics in Localization-Based Superresolution Microscopy,” Biophys. J. 113(9), 2037–2054 (2017). 10.1016/j.bpj.2017.08.054 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Bourgeois D., “Single molecule imaging simulations with advanced fluorophore photophysics,” Commun. Biol. 6(1), 53 (2023). 10.1038/s42003-023-04432-x [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplement 1. Supplement 1.
boe-17-9-4854-s001.pdf (3.6MB, pdf)

Data Availability Statement

Data underlying the results presented in this paper are available via OSF (Ref. [44]).


Articles from Biomedical Optics Express are provided here courtesy of Optica Publishing Group

RESOURCES