Table 1.
Type of alarm | Alarm category | Note | Standard |
Electric or pneumatic failure | High priority | EN 794-1 [12] | |
FIO2 high or low | At least medium priority | Is applicable as soon as O2 concentration is | EN 794-1 |
different from that of ambient air | |||
Paw high | High priority | EN 794-1 | |
VE low* or VT low* | At least medium priority | EN 794-1 | |
Apnoea | At least medium priority | EN 794-1 | |
Disconnection | At least medium priority | Could be detected for example from a low Paw, | EN 794-1 |
a low ETCO2 and a low tidal volume | |||
Continuous pressure | High priority | Is relative to a continuous pressure kept over a | EN 794-1 |
given limit during more than 15 ± 1.5 s | |||
ETCO2 | |||
High | Medium priority | EN 864 [14] | |
Low | Medium priority | EN 864 | |
FICO2 high | Medium priority | EN 864 | |
SpO2 | |||
High | No priority indicated | For neonatology | EN 865 [15] |
Low | No priority indicated | EN 865 | |
Sensor failure | Low or medium priority | EN 865 |
*According to these standards, except for the ventilators used in neonatology, the measurement of expiratory tidal volume (VT) or minute ventilation (VE) must be provided. Only the parameters and events listed in the standards are reported here. The values of high and low alarm limits are set by the medical staff. An alarm of high priority implies an immediate response from the staff; an alarm of medium priority implies a prompt response from the staff; an alarm of low priority is used to attract staff's attention. ETCO2, end tidal CO2; FICO2, concentration of carbon dioxide during inspiration; FIO2, concentration of dioxygen during inspiration; Paw, airway pressure; SpO2, saturation of oxyhemoglobin determined by pulse oximetry.