. 2006 Feb;44(2):423–432. doi: 10.1128/JCM.44.2.423-432.2006

TABLE 3.

Quantification of a known amount of HIV-1 RNA subtype A in pooled CVS lavage fluids and plasma samples from HIV-negative women spiked with a known amount of HIV-1 subtype A reference control by HIV-1 RNA real-time RT-PCR using a Light Cycler apparatusa

Expected result (log10 copies)b Cervicovaginal secretion
Plasma
Observed result (log10 copies) Δlog10 Observed result (log10 copies) Δlog10
6.00 5.96 0.04 5.77 0.23
5.00 5.15 −0.15 4.80 0.20
4.00 4.21 −0.21 3.91 0.09
3.00 3.15 −0.15 2.83 0.17
2.70 2.82 −0.12 2.70 0.00
2.40 2.45 −0.05 2.36 0.04
2.10 2.20 −0.10 2.04 0.06
1.80 1.53 0.27 1.70 0.10
Mean −0.06 0.11
a

Δlog10 is the difference between expected log10 values and observed log10 values obtained with a real-time RT-PCR assay.

b

HIV-1 RNA subtype A at 109 copies/ml was obtained from the Viral Diagnostic Quality Control Programme, Amsterdam, The Netherlands.