Quantification of a known amount of HIV-1 RNA subtype C in pooled CVS lavage fluids and plasma samples from HIV-negative women spiked with a known amount of HIV-1 subtype C reference control by HIV-1 RNA real-time RT-PCR using a Light Cycler apparatusa

Expected result (log10 copies)b Cervicovaginal secretion
Plasma
Observed result (log10 copies) Δlog10 Observed result (log10 copies) Δlog10
6.00 5.90 0.10 5.73 0.27
5.00 4.95 0.05 5.02 −0.02
4.00 4.27 −0.27 3.99 0.01
3.00 3.21 −0.21 3.02 −0.02
2.70 2.92 −0.22 2.67 0.03
2.40 2.62 −0.22 2.71 −0.31
2.10 2.37 −0.28 2.05 0.04
1.80 2.13 −0.33 2.10 −0.31
Mean −0.17 −0.04
a

Δlog10 is the difference between expected log10 values and observed log10 values obtained with a real-time RT-PCR assay.

b

HIV-1 RNA subtype C at 109 copies/ml was obtained from the Viral Diagnostic Quality Control Programme, Amsterdam, The Netherlands.