Results of M. pneumoniae detection in 2004

Sample no. and test description Input (CCU/ 100 μl), pool no.a Laboratory no. and result
% Positive
1 1b 2 3 4 5 6 8 9 11 11c 14 16 17 20 18 18c 19c
4 5,000, p2 + + + + + + + + + + + + + + + + + + 100
6 5,000, p1 + + + + + + + + + + + + + + + + + 94.4
1 500, p1 + + + + + + + + + + + + + + + + + 94.4
7 500, p2 + + + + + + + + + + + + + + + + 88.9
8 5,000, p3 + + + + + + + + + + + + + + + + 88.9
9 500, p3 + + + + + + + + + + + + + + + + 88.9
2 0, p3 0
3 0, p1 0
5 0, p2 0
10 0, p1 0
% Correct results 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 90 70 60
Routine application NAATd Y Y N Y N Y Y Y N Y Y N Y N Y Y ? N
No. of samples tested monthlye 3 3 0 1 0 1 2 2 0 3 ? 0 2 0 1 1 ? 0
a

p1, clear BAL pool; p2, more viscous BAL pool; p3, BAL pool blood stained.

b

Results of multiplex NASBA.

c

Results of multiplex PCR.

d

All operations were performed in separate rooms.

e

0 = 0; 1 = 1 to 10 samples monthly, 2 = 11 to 45; 3 means more than 50.