Skip to main content
. 2006 Mar;44(3):1101–1104. doi: 10.1128/JCM.44.3.1101-1104.2006

TABLE 3.

Evaluation of accuracies of automated systems for susceptibility testing of 100 P. aeruginosa strains against β-lactam antimicrobial agents

Antimicrobial and error typea Error rate (%)
MicroScan WalkAway VITEK 2 VITEK
Aztreonam
    Very major 0 0 2
    Major 0 1 2
    Minor 28 31 28
Cefepime
    Very major 0 0 0
    Major 3 1 0
    Minor 32 18 26
Ceftazidime
    Very major 0 1 2
    Major 0 1 0
    Minor 13 9 11
Imipenem
    Very major 0 1 0
    Major 2 2 2
    Minor 10 8 11
Piperacillin-tazobactam
    Very major 19 27 21
    Major 1 0 0
a

Error rates were calculated based on the consensus result among the broth microdilution (frozen dry-form panels), agar dilution, and disk diffusion methods.