TABLE 4.
Test and result | No. of samples with the indicated result by the following test:
|
|||||||
---|---|---|---|---|---|---|---|---|
rOMP90-4
|
rOMP91B iELISA
|
cELISA
|
CFT
|
|||||
Positive | Negative | Positive | Negative | Positive | Negative | Positive | Negative | |
rOMP90-3 ELISA | ||||||||
Positive | 115 | 1 | 78 | 0 | 32 | 6 | 111 | 5 |
Negative | 40 | 281 | 15 | 161 | 54 | 91 | 141 | 180 |
% Concordancea | 90.6 | 94.1 | 67.2 | 66.6 | ||||
rOMP90-4 ELISA | ||||||||
Positive | 87 | 2 | 47 | 19 | 143 | 12 | ||
Negative | 6 | 159 | 39 | 78 | 109 | 173 | ||
% Concordance | 96.9 | 68.3 | 72.3 | |||||
rOMP91B iELISA | ||||||||
Positive | 86 | 7 | ||||||
Negative | 36 | 125 | ||||||
% Concordance | 83.1 | |||||||
cELISA | ||||||||
Positive | 86 | 0 | ||||||
Negative | 44 | 53 | ||||||
% Concordance | 76.0 |
Concordance is the sum of positive-positive values and negative-negative values expressed as a percentage of the total number of serum samples.